® DeFelsko Corporation DeFelsko 802 Proctor Avenue Ogdensburg, New York 13669-2205 USA Certificate of Calibration Certificate Number: 18-306349 Nomenclature: PosiTest Adhesion Tester Laboratory Environment Manufacturer: DeFelsko Corporation ‘Temperature: 23 + 5°C Model: Positest AT-A Relative Humidity: Up to 95% Serial No: ATO8668 Date of Calibration: July 3, 2018 Test Method: This Adhesion Tester was calibrated to manufacturer’s specifications according to procedure MP 2571 using a Load Cell calibrated by an accredited laboratory and traceable to NIST by certificate 76003. ‘Adhesion Tester Full Measurement Gage (psi) _ | Load Cell (psi) | Range Accuracy (%) 498 999 1507 1999 2505 3012 Uncertainty = 0.08% FS. © 2.5psi) Measurement Performed by: Andrew Lauson (by seen Technician DeFelsko Corporation operates under Management Procedures intended to implement the requirements of ISO 9001, ISO 10012-1, ISO 17025 and ANSI/NCSL Z540-1. This document certifies that the Adhesion Tester identified above has met the published specification of: +£1% of full measurement range Calibration interval will vary based on usage, handling and storage conditions. This certificate shall not be reproduced, except in full, without the written approval of DeFelsko Corporation. Page 1 of | Management Form 2007,03-1/2009 Client: NBTC - Technical Service Division