Beruflich Dokumente
Kultur Dokumente
When you are finished with setting these parameters and you can click and hold A1-button in
2 seconds to save and quit the parameter-menus.
When you are finished to set the parameters and calibrated the sensor you can continue to
step 3 or 4, depending on which system you are using.
Do the following:
Make sure the conveyor is empty and no objects are inside the reading beam. Put in the
«Analog In» value (DV) which the PLC-program is reading, under Test Minimum -X1: ANA. IN.
Then put in 0mm in Y1:INDICATION. Set an object under the sensor which is approximately or
higher than 1200mm and put in the currently reading «Analog In» value (DV) into Test
Maximum – X2: ANA. IN and write in the height on the object(s) into Y2: INDICATION.
Change Maximum Pos. In between 1400 and 1450mm and put in a tolerance between 10 and
50mm. These values are for giving warnings if the foam is above or under the range you are
trying to produce and above the maximum height of production.
Example:
Value directly from sensor: 7.85 volts = Real measurement on the conveyor 0mm
Value directly from sensor: 1,23 volts = Real measurement on the conveyor 1375mm
7,85 volts = 0mm = Min. Input values on instrument (Lo.D/Lo/Lo.E)
1,23 volts = 1375mm = Max. Input values on instrument (Hi.D/Hi/Hi.E)
You would obviously have to do the opposite if you have the smallest value from the sensor
when it reads directly onto the conveyor.
In relations to setting the parameter in a digital instrument you would have to follow the
instructions and manual for these.