Beruflich Dokumente
Kultur Dokumente
791
CALIBRATION AND FUNCTION TESTING 621.317.72
Published by NORDTEST Tekniikantie 12, FIN-02150 ESPOO FINLAND Phone +358 9 455 4600 Fax +358 9 455 4272
ISSN 0283-717X Proj. 1124-93 www.nordtest.org
NORDTEST METHOD NT ELEC 028 2
A reading from an instrument where a calibration takes place. True value (T):
Electrical current.
Uncertainty (a):
The calibration results shall be given in a table containing the 6 CALIBRATION OF AMMETERS
readings of the standard, the readings of the instrument which
6.1 Principle
is calibrated, the error at each calibration point (except for zero)
and the standard deviation at each calibration point if relevant. The ammeter which is to be calibrated is compared with the
output from a current standard. The uncertainty of the output
At each calibration point the arithmetic mean value shall be
from the current standard must not be greater than ±0.5 % of
calculated by using Equation (1) and the standard deviation
the setting.
shall be calculated by using Equation (3).
If the uncertainty of the standard is greater than ±0.5 % of the
The error q at each calibration point (except for zero) shall be
setting, an ammeter with uncertainty less than ±0.5 % of the
calculated by using Equation (2) and the total uncertainty for
reading can be used as a standard. The short term stability of
the calibration shall be calculated by using Equation (5).
the current source must be sufficient to give stable readings
on this ammeter. Three different methods of calibration are
described below. The method to be used depends on the
5 CALIBRATION OF VOLTMETERS
current standard and the ammeter which shall be calibrated.
5.1 Principle
The voltmeter which is to be calibrated is compared with the
6.2 Direct calibration of ammeters
output from a voltage standard. The uncertainty of the output
from the voltage standard must not be greater than ±0.2 % of The instrument set-up for this calibration is shown in Fig. 2.
setting.
The output from the current standard is connected to the input
If the uncertainty of the standard is greater than ±0.2 % of the of the current shunt or with one turn through the clamp-on
setting, a voltmeter with uncertainty less than ±0.2 % can be current probe.
used as standard. The short term stability of the voltage source
This method shall be used if the current standard can supply
must be sufficient to give stable readings on this voltmeter. The
enough current to calibrate the ammeter to 100 % of the range
instrument set-up is shown in Fig. 1.
which shall be calibrated. If this is not the case, methods
described in 6.3 or 6.4 shall be used.
VOLTAGE VOLTMETER
STANDARD
CURRENT AMMETER
Fig. 1. Calibration of voltmeter. STANDARD
standard can supply sufficient current to perform a direct The calibration points should be equally spaced over the range
calibration described in 6.2, this method should be used since (0-20-40-60-80-100 % of the range).
this will give a calibration of the complete current measuring
The ammeter shall be calibrated in at least one point on the
equipment.
other ranges which are to be calibrated. On ranges which are
For indirect calibration of ammeters with a voltage standard calibrated in only one point this calibration shall be between
the reading (R) on the ammeter will be the voltage input (Vin) 80 % and 100 % of the range. Ammeters which can measure
multiplied by the scaling for the shunt (A/V): current in both directions shall be calibrated in both directions.
(H - L)100
sp = U or I
(H + L)√3 (7)
, _ .._._....................... _...______...____............ .
7.2.2 Testing of averaging time for clamp on ammeters shorter than recommended. The following criteria and equa-
with no removable clamp on probe and shunt tion shall be used for classifying averaging times.
ammeters with no removable shunt Calculated variations ±20-40 %: averaging time is within
The averaging time is normally not related to the range, and it recommended range.
is therefore adequate to test the instrument on one range, Calculated variations <±20 %: averaging time is longer than
preferably the lowest. To have an impression of how the recommended.
instrument responds to input variations the instrument should Calculated variations >±40 %: averaging time is shorter than
be exposed to a known input signal. A principle diagram of a recommended.
circuit which can produce such a signal is shown in Fig. 4. Set
(RH - RL)100 %
the pulse width to approximately 1.0 s, and the frequency to c = (8)
(RH+RL)
approximately 0.5 Hz. The amplitude should preferably be set
to the range of the ammeter. For a mean value measuring where
ammeter the mean value of the readings shall be the pulse
c is the calculated percentage variation which shall be used
amplitude divided by 2. For an instrument with an averaging
to classify the averaging time
time of 1-2 s the variations in the reading will be ±20-40 % of
the mean value. Record the highest and lowest values during RH is the highest reading during the averaging time test
a period of at least 30 s. RL is the lowest reading during the averaging time test.
and the frequency to approximately 0.5 Hz. Set the amplitude Fig. 5. Principle circuit for amplitude and frequency response
of the pulse generator to give readings on the ammeter around testing of voltmeters.
50 % of the range. For an average measuring instrument the
mean value of the readings should be the pulse amplitude
scaled to ampere and divided by 2. For an instrument with an
averaging time of 1-2 s the variations in the readings will be 7.3.2 Test procedure
±20-40 % of the mean value. Record the highest and lowest
Connect the voltmeter to the test circuit shown in Fig.5.
values during a period of at least 30s.
The output from the test circuit shall be set to direct voltage
output only. Adjust the voltage until the reading on the
7.2.4 Expression of results from testing of averaging voltmeter is approximately 90 % of the range of the voltmeter.
times
Record the reading of the voltmeter (d).
The test results given as percentage variations from the mean
values shall be calculated and recorded in the test results. Set the AC generator to 50 Hz and adjust the amplitude of the
Equation (8) below shall be used for the calculations. alternating voltage to the same voltage as the direct voltage
and record the reading of the voltmeter (a).
The test results shall classify the averaging time as being
within the recommended range, larger than recommended or Calculate the ratio m = a/d
NORDTEST METHOD NT ELEC 028 6
A mean value measuring voltmeter will indicate the same 7.4.2 Test procedure
voltage as when only the direct voltage was connected. An
Connect the ammeter to the test circuit shown in Fig. 6.
r.m.s. measuring voltmeter will indicate the direct voltage
multiplied by 1.2. Sweep the frequency of the alternating The output from the test circuit shall be set to direct current
voltage between 20 Hz and 2000 Hz. output only. Adjust the current until the reading on the ammeter
is approximately 90 % of the range of the ammeter.
7.3.3 Expression of results Set the AC generator to 50 Hz and adjust the amplitude of the
alternating current to the same current as the direct current
The following shall be given as the result of test 7.3:
and record the reading of the ammeter (a).
The ratio (m) calculated in 7.3.2. Calculate the ratio m = a/d
The maximum deviation as a percentage of the reading at 50 A mean value measuring ammeter will indicate the same
Hz in the range 20-1000 Hz and the frequency for the reading. current as when only the direct current was connected. An
Recommended maximum deviation less than ±2 %. r.m.s. measuring ammeter will indicate the direct current mul-
tiplied by 1.2.
The maximum deviation as a percentage of the reading at 50
Hz in the range 1000-2000 Hz and the frequency for the Sweep the frequency of the alternating current between 20 Hz
reading. Recommended maximum deviation less than ±10 %. and 2000 Hz.
U or l
OSCILLATOR POWER
20-2000 H z DC-POWER Direct voltage or
AMPLIFIER
SUPPLY direct current
Table 1
1.3
7 1.3
8 1.2
9 1.2