Model AMX – 4
Manufactured 1989
WARNING
- Stand at least 6 feet away from the machine when taking an x-ray
To move the unit, press fingers toward yourself on the drive bar and an
audible unlock noise will indicate that the unit can move. The machine
eases operator labor by self-propelling in the direction pushed or
pulled.
To lift the tube arm, press on the handles of the x-ray tube head and
pull up. The arm can then be telescoped up, around and out to desired
location. Please note that if the machine is to be moved while arm is
not in locked dock, movement speed will be dramatically reduced.
For safety, there is a bar on the front that stops the machine when it is
depressed in order to prevent damage to the equipment, facility, and
individuals should they be struck by the front of the machine.
If the battery power goes lower than 10%, the unit will not be able to
create x-rays and the display will read, RECHARGE IMMEDIATELY - X-
RAY INHIBITED.
The message, CHARGE COMPLETE, will display when the unit is fully
charged.
The tube head provides all the controls necessary to the radiologic
technologist. 2 knobs control the field size both longitudinally and
horizontally. A tape measurer provides accurate SID for proper
exposure precautions. The center black button provides light to
visualize the approximate field of exposure.
Be sure to check and set kVp and mAs settings appropriate for the
exam. Recommended settings have been included and asked to be
mounted on the machine arm for technologist reference.
Use the corded exposure switch to take the x-ray. A black button
illuminates the light for visual at a distance from the operating station
and the top button is a two-step button that initiates the exposure. The
first step rotors the machine up, while the second completes the
exposure when the button is fully depressed.