Sie sind auf Seite 1von 13

Development of a silicon diode detector for skin dosimetry in radiotherapy

Nikolina Vicoroski, Anthony Espinoza, and Mitchell Duncan


Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2500, Australia
Bradley M. Oborn and Martin Carolan
Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2500, Australia
Illawarra Cancer Care Centre, Wollongong Hospital, Wollongong, NSW 2500, Australia
Peter Metcalfe
Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2500, Australia
Illawarra Health and Medical Research Institute – IHMRI, Wollongong, NSW 2500, Australia
David Menichelli
IBA Dosimetry GmbH, Bahnhofstr, 5 90592 Schwarzenbruck, Germany
Vladimir L. Perevertaylo
SPA BIT, Kiev 08296, Ukraine
Michael L. F. Lerch, Anatoly B. Rosenfeld, and Marco Petaseccaa)
Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2500, Australia
Illawarra Health and Medical Research Institute – IHMRI, Wollongong, NSW 2500, Australia

(Received 7 February 2017; revised 14 June 2017; accepted for publication 28 June 2017;
published 12 August 2017)
Purpose: The aim of in vivo skin dosimetry was to measure the absorbed dose to the skin during
radiotherapy, when treatment planning calculations cannot be relied on. It is of particularly impor-
tance in hypo-fractionated stereotactic modalities, where excessive dose can lead to severe skin toxic-
ity. Currently, commercial diodes for such applications are with water equivalent depths ranging
from 0.5 to 0.8 mm. In this study, we investigate a new detector for skin dosimetry based on a silicon
epitaxial diode, referred to as the skin diode.
Method: The skin diode is manufactured on a thin epitaxial layer and packaged using the “drop-in”
technology. It was characterized in terms of percentage depth dose, dose linearity, and dose rate
dependence, and benchmarked against the Attix ionization chamber. The response of the skin diode
in the build-up region of the percentage depth dose (PDD) curve of a 6 MV clinical photon beam
was investigated. Geant4 radiation transport simulations were used to model the PDD in order to esti-
mate the water equivalent measurement depth (WED) of the skin diode. Measured output factors
using the skin diode were compared with the MOSkin detector and EBT3 film at 10 cm depth and at
surface at isocenter of a water equivalent phantom. The intrinsic angular response of the skin diode
was also quantified in charge particle equilibrium conditions (CPE) and at the surface of a solid water
phantom. Finally, the radiation hardness of the skin diode up to an accumulated dose of 80 kGy
using photons from a Co-60 gamma source was evaluated.
Results: The PDD curve measured with the skin diode was within 0.5% agreement of the equivalent
Geant4 simulated curve. When placed at the phantom surface, the WED of the skin diode was esti-
mated to be 0.075  0.005 mm from Geant4 simulations and was confirmed using the response of a
corrected Attix ionization chamber placed at water equivalent depth of 0.075 mm, with the measure-
ment agreement to within 0.3%. The output factor measurements at 10 cm depth were within 2% of
those measured with film and the MOSkin detector down to a field size of 2 9 2 cm2. The dose–re-
sponse for all detector samples was linear and with a repeatability within 0.2%. The skin diode intrin-
sic angular response showed a maximum deviation of 8% at 90 degrees and from 0 to 60 degree is
less than 5%. The radiation sensitivity reduced by 25% after an accumulated dose of 20 kGy but
after was found to stabilize. At 60 kGy total accumulated dose the response was within 2% of that
measured at 20 kGy total accumulated dose.
Conclusions: This work characterizes an innovative detector for in vivo and real-time skin dose mea-
surements that is based on an epitaxial silicon diode combined with the Centre for Medical Radiation
Physics (CMRP) “drop-in” packaging technology. The skin diode proved to have a water equivalent
depth of measurement of 0.075  0.005 mm and the ability to measure doses accurately relative to
reference detectors. © 2017 American Association of Physicists in Medicine [https://doi.org/10.1002/
mp.12469]

Key words: external beam radiotherapy skin dose, silicon detectors, skin dosimetry

5402 Med. Phys. 44 (10), October 2017 0094-2405/2017/44(10)/5402/11 © 2017 American Association of Physicists in Medicine 5402
5403 Vicoroski et al.: Silicon diode for skin dosimetry 5403

1. INTRODUCTION sensitivity and small sensitive volumes. The MOSkin, based


on the MOSFET technology and recently introduced by
Quality assurance of radiotherapy treatments involves several CMRP, are also able to produce real-time readout and
techniques and diversified approaches to verify the dose provide a reproducible water equivalent depth of dose mea-
delivered to the patient. Monitoring of the skin dose helps to surement of 0.07 mm.15 They are the only detector for real-
prevent and minimize the side effects associated with inevita- time in vivo skin dosimetry currently available. MOSkin
ble irradiation of healthy tissues such as the skin. About 20% has demonstrated good agreement with the Attix IC for
of women undergoing 3D conformal radiotherapy for breast surface dose measurements and angular dependence in MV
cancer treatments develop acute skin toxicity which affects and kVp photon beams.16 However, due to their limited
dramatically the quality of life of the patient.1–3 ICRP defines linear dynamic range (up to 50 Gy) they are not practical
skin dose as the dose absorbed and measured at the water to use in a 2D array.
equivalent depth (WED) of 0.07 mm which makes advance- Silicon diode detectors could represent a good candidate
ment of skin dosimetry challenging.4 to develop a new technology to measure in real-time in vivo
An ideal skin dosimeter should have extremely thin sensi- skin dose. The technology development of silicon diodes
tive volume covered with tissue equivalent build-up providing makes them the standard reference dosimeter for a variety of
a WED of 0.07 mm for skin dosimetry or even less if surface application in radiotherapy from quality assurance of stereo-
dose measurement is required. tactic modalities to in vivo dosimetry due to their small size
Although the reference instrument for surface dosimetry at allowing for high spatial resolution and high sensitivity.17,18
the air-to-water interface is the extrapolation ion chamber, it However, current designs of silicon diodes die and their pack-
is not used routinely for skin dosimetry.5 In this work, the aging, which provides electronic equilibrium conditions for
Attix ionization chamber (IC) is adopted as reference for MV field of interest, do not make them suitable for direct
measurements in a phantom.6 Cylindrical IC are normally skin dosimetry on medical linacs.
unsuitable for this application because of their large diameter, In this paper, a full characterization of a silicon epitaxial
thick entrance window, and large beam perturbation. diode detector for surface dosimetry in a phantom was per-
Unguarded parallel plate ICs show an over-response at very formed. It includes: (a) detector calibration, linearity and
shallow depths due to the lateral secondary electrons scatter- reproducibility; (b) percentage depth dose; (c) field size
ing into the sensitive volume.7 For this reason, only large dependence; (d) angular dependence; (e) dose per pulse
diameter ring-guarded ICs such as the Attix ionizing chamber dependence; and (f) radiation hardness. Comparisons were
(Model 449 CNMC Co. Nashville, TN – USA) are com- made with Monte Carlo simulations, ionization chambers,
monly considered as the reference instrument for interface and film.
dose measurement. Although very accurate for in-phantom
measurements, the Attix IC can be used for skin dosimetry in
2. MATERIALS AND METHODS
a phantom only; they are limited to measure field size beams
of 5 9 5 cm2 and above and they have been never manufac- Characterization of dosimeters is normally performed
tured in array for skin 2D dose mapping.8 where charged particle equilibrium (CPE) condition exists.18–21
Radiochromic films have an excellent spatial resolution, However, characterization of a skin dosimeter should be
tissue equivalency, and can provide 2D dosimetric informa- performed at the surface and at very shallow depth in the
tion. However, as radiochromic film is a passive detector, it build-up region, where strong electronic disequilibrium and
does not allow for real-time measurements. Its response steep dose gradient occurs. Such conditions are typical for
may also be affected by a WED of measurements much in vivo skin dosimetry.16
larger than 0.07 mm, improper handling, and scanner
performance.9,10
2.A. Silicon diode detectors
Thermal luminescence dosimeters (TLDs) have been
used11 clinically for skin dose measurements in their carbon- Two types of epitaxial diodes were fabricated, and five
loaded version. They are small in size and tissue equivalent, samples for each prototype were tested. The skin diode
however, they require a long series of pre- and postirradiation (named Epi-5B), fabricated by Fondazione Bruno Kessler
processes and a refined calibration procedure to be readout (FBK - Trento, Italy) on an N-type silicon 5 um thick epitax-
accurately. They are labor-intensive and not real-time dosime- ial layer grown on a 0.001 Ohm-cm N-type silicon substrate,
ters, rendering them impractical for use in real-time in vivo were originally designed for high energy physics applica-
skin dosimetry.12,13 Zhuang and Olch showed that a nano-dot tions. The detector structure is represented in Fig. 1 and is
optically stimulated luminescence dosimeter can be adopted composed of a boron implanted P+ sensitive area of
as skin dosimeter with stabilization delay of less than 10 min 0.5 9 0.5 mm2, with a total size of the die of
after irradiation. However, it has an effective WED of 0.75 9 0.75 mm2. The effects of the superficial leakage cur-
0.8 mm which is larger than WED of 0.07 mm accepted for rent are minimized by using a guard-ring surrounding the
skin dosimetry.14 entire pixel sensitive area.
Semiconductor detectors such as MOSFETs are able to In order to estimate the effect of the thickness of the epi-
achieve excellent spatial resolution because of their high taxial layer (i.e., sensitive volume) on the WED of dose

Medical Physics, 44 (10), October 2017


5404 Vicoroski et al.: Silicon diode for skin dosimetry 5404

(a) (b)

FIG. 1. Simplified diagram (not to scale) of the skin diode Epi-5B topology (a) top view and (b) cross-section view. [Color figure can be viewed at wileyonlineli-
brary.com]

measurement, a second epitaxial silicon diode was fabricated aluminum layers is illustrated in Fig. 3. The build-up mate-
by CMRP on a silicon P-type 50 lm thick epitaxial layer rial thickness above the detector sensitive layer is repro-
grown on a 0.001 Ohm-cm substrate (named Epi-1C – ducible and already proved to be able to provide a WED of
Fig. 2). The Epi-1C has N+ phosphorus implanted active approximately 0.07 mm.16 The “drop-in” technology is
area of 0.6 9 0.6 mm2 and overall die dimensions of compatible with standard flexible printed circuit manufac-
1.5 9 1.5 mm2. This diode has been adopted for the fabrica- turing technologies and avoids the use of high atomic num-
tion of the 11 9 11 2D diode array MagicPlate.22 ber materials for packaging of semiconductor radiation
Both epitaxial diodes have been embedded in kapton detectors, minimizing the perturbation of the beam. The
pigtails with 0.6 9 3 mm2 cross section and 350 mm kapton pigtail outside of the diode is shielded by a thin
length using CMRP proprietary “drop-in” packaging tech- aluminum foil and grounded to minimize radiofrequency
nology. This packaging technique is based on tab bonding interference.23 Further details on the “drop-in” packaging
of the silicon die underneath a stack of polyamide and technology can be found elsewhere.24

(a) (b)

FIG. 2. (a) Top view diagram of the Epi-1C diode topology and (b) cross-section view, not in scale. [Color figure can be viewed at wileyonlinelibrary.com]

Medical Physics, 44 (10), October 2017


5405 Vicoroski et al.: Silicon diode for skin dosimetry 5405

when the beam is on, by synchronization of the detector sam-


pling and the analog to digital conversion with the linac sync
signal available from the linac stand. This modality has the
double advantage of minimizing the effects of electronic
noise and parasitic currents, and managing the dead time by
reading out the data while the beam is off. It readouts the
integral charge on a 16 bit scale of 4.8 pC, which corre-
sponds to the measured charge resolution of 7 9 105 pC.
The electrometer is readout by a USB2.0 interface managed
by a graphical interface installed into a standard laptop. For
further details on the electrometer architecture please refer to
Ref. [30].

2.D. Response linearity


FIG. 3. Simplified schematic of the kapton pigtail cross section accommo-
dating the skin diode die using the “drop-in” packaging technology. [Color The detectors’ linear response is investigated under the
figure can be viewed at wileyonlinelibrary.com] standard conditions of 6 MV photon beam with field size
10 9 10 cm2 and 1.5 cm depth in a Virtual Water phantom
(Standard Imaging Inc., Middleton, WI, USA). The diodes’
response was recorded from 50 cGy up to 500 cGy. Uncer-
2.B. Attix ionization chamber
tainty is calculated by two standard deviations over
Ionization chambers are considered the reference tool for three consecutive measurement of each of three samples
calibration and comparison of other dosimeter detectors in from the same production batch and packaged with the same
radiotherapy. The reason behind their popularity is due to technology.
their small variation in response to energy, dose, dose rate,
and their high response reproducibility.25
2.E. Irradiations and radiation hardness
However, measuring small or narrow fields becomes a
characterization
challenge when using ionization chambers because of the
lack of lateral electronic equilibrium and the volume of the All measurements were performed using the Varian Cli-
chamber perturbing the field. nac 21Xi medical linac at the Illawarra Cancer Care Centre
The ionization chamber used in this work is the Attix of Wollongong Hospital (NSW, Australia) 6 MV photon
ionization chamber (Model 449 CNMC Co. Nashville, TN beams with a flattening filter. The irradiation of the epitax-
– USA). The reason for using a parallel plate ionization ial diode probes for studying the radiation damage effects
chamber over a cylindrical or Markus ionization chamber is on charge collection efficiency of the skin diodes has been
that the Attix ionization chamber is flat and thin with the performed by at GATRI, a Co-60 source at the Australian
front electrode consisting of a window of Kapton 0.025 mm National Science and Technology Organisation (ANSTO,
thick including a conductive coating and only 2 mm separa- Lucas Heights, NSW, Australia). The effect of radiation
tion between the electrode plates.26,27 This provides a suit- damage on degradation of radiation sensitivity of Epi-1C
able geometry for surface dosimetry, although at the epitaxial diode is presented in Ref. [31]. As proved by
expense of the chamber’s directional dependence. The Attix Bruzzi et al., passive epitaxial diodes are usually less
ionization chamber has a large guard-ring 13.5 mm wide affected by accumulated dose due to the thin sensitive
that reduces the over-response in the build-up region to less volume.32
than 1%.28 To stabilize the skin diodes response, they were preirradi-
ated with dose increments of 20 kGy up to a total accumu-
lated dose of 80 kGy with response measurement at each
2.C. Electrometer
irradiation step. The response of the skin diodes was mea-
The extreme thin sensitive volume adopted for the skin sured under standard conditions (1 MU equals 1 cGy) by
diode Epi-5B requires the use of a high sensitivity electrome- delivering 100 MU on 6 MV photon beam, field size of
ter. The multichannel electrometer used for this work is based 10 9 10 cm2, at depth of 1.5 cm in the Virtual Water phan-
on the AFE0064 which is an analog front end designed and tom at a source to surface distance (SSD) of 100 cm. Error
commercialized by Texas Instruments29 to readout flat panel bars are calculated as one standard deviation over five repeti-
amorphous silicon detectors. The AFE0064 is equipped with tions of the same measurement.
an optional double sampler for the subtraction of the signal
baseline — a feature that has been adopted in this work for
2.F. Percentage depth dose
accurate measurement of the charge collected in the volume
of 1.25 9 103 mm3 of the Epi-5B. The AFE0064 also has The percentage depth dose (PDD) measurements are
the major advantage to be able to sample the detector only obtained for both the silicon epitaxial diodes Epi-5B and Epi-

Medical Physics, 44 (10), October 2017


5406 Vicoroski et al.: Silicon diode for skin dosimetry 5406

1C and compared to the Attix ionization chamber. All these The reference field size for this study was 10 9 10 cm2,
detectors were placed, one at a time, in a recess machined in which was measured at isocenter and at a depth of 10 cm
a Virtual Water (VW) slab to ensure consistency of the scat- (SSD is 90 cm), with delivered 100 MU each time. The out-
tering conditions. A 10 cm thick slab of 30 9 30 cm2 VW put factor was measured by placing the preirradiated skin
phantom is used for backscattering and several slabs ranging diodes (60 kGy irradiation dose by Co-60) in the center of
from 0.1 cm to 20 cm are used to obtain the depth dose pro- the field for field sizes and varying the jaws’ aperture from
files. The samples are irradiated with 100 MU of a 6 MV 2 9 2 cm2 to 30 9 30 cm2. A careful alignment of the
photon beam with a 10 9 10 cm2 field size at 100 cm SSD detector was performed by maximizing the detector response
for each measurement. for the 2 9 2 cm2 beam by a scan in the SUP-INF and lateral
Additionally, PDD were obtained for the skin diode Epi- directions using the couch positioning system (spatial resolu-
5B with the epitaxial layer of the skin diode facing down- tion of 1 mm). A measurement of the field factor at surface
wards and the diode being exposed to the incident radiation of the phantom was also performed to evaluate the capabili-
beam. This was conducted under the same experimental con- ties of the detector to measure the surface dose as a function
ditions as when the epitaxial layer of the skin diode was fac- of the field size defined by the jaws. The silicon epitaxial
ing up, to determine whether orientation of the detector diode was benchmarked against Gafchromic EBT3 film, the
would affect the PDD. Attix ionization chamber, and the MOSkin, which are fre-
Monte Carlo Geant4 version 10.0p0133 simulations were quently used for interface measurements and described in
performed to determine the dose distribution at the build-up several studies.16,37–40
region generated by the specific Linac used for this experi- The films were calibrated using 1.5 9 1.5 cm² squares
ment. The beam 10 9 10 cm2, 6 MV x-ray, passing through and scanned using a flatbed scanner (Epson 10000XL scan-
a 30 9 30 9 30 cm3 water equivalent block phantom was ner; Epson America, Inc. Long Beach, CA, USA) 24 h after
simulated. irradiation to allow for postirradiation development. The
The source to surface distance (SSD) was 100 cm, and films were scanned in transmission mode with color correc-
the beams were fired from phase space files created from tion turned off, at a resolution of 72 dpi and 48 bit RGB data
an EGSnrc Monte Carlo based system that models the Var- format. Analysis of three scanning repetitions for each film
ian 2100C Linac at Illawarra Cancer Centre in Wollon- placed with the same orientation were performed using Ima-
gong (Australia). The model used to represent the Linac geJ 1.46r software (National Institute of Health, Bethesda,
and the phantom setup, along with the physics of trans- MD, USA) and one standard deviation of the optical density
port, has been already validated by previous experimental was used as uncertainty. A template was used to scan the
results.34 The physics processes modeled in the simulation films at the center of the scanner bed to avoid nonuniformity
are from the Geant4 electromagnetic standard physics associated with the position of the film in respect to the scan-
package and include photoelectric effect, Compton scat- ner lamp or sensor. Only the red channel was used for analy-
tering and gamma conversion (photons), ionization, sis of a region of interest (ROI) at the center of the film. The
Bremsstrahlung, and positron annihilation. The particle films used to perform the measurements have been pres-
range cut was set to 0.1 mm and the electron/positron canned to evaluate the background optical density and to take
maximum step length was 0.1 mm. Dose was scored care of the nonuniformities associated with the specific film
inside the phantom at a voxel size of 0.02 9 1 9 1 mm3. cut. Butson et al.10 provided in-depth details of the prepara-
Each simulation was split into 10 parallel jobs each with tion and scanning procedure adopted for the analysis of the
unique seeds and the mean dose reported. The standard EBT3 film.
deviation is taken across 10 simulations to evaluate uncer-
tainties. In total, 4 9 1013 primary histories are simulated
2.H. Dose per pulse dependence
where the primary histories represent the number of elec-
trons hitting the x-ray target in the linac head. The dose The measurement of dose per pulse dependence (DPP)
uncertainty is approximately 0.5% of the dose at 1.5 cm refers to the evaluation of the change of the detector sensitiv-
depth at the beam central axis (CAX) for the 10 9 10 cm2 ity due to variation of the instantaneous dose rate in a pulsed
field size. radiation beam typical of a medical linear accelerator.41 It
was reported by Rikner and Grusell that n-type silicon diodes
2.G. Output factor and surface field size are more sensitive to dose per pulse variation than compared
dependence to p-type silicon detectors, thus showing an increase in sensi-
tivity with increasing dose per pulse.41
The output factor is defined as the ratio of dose in water Therefore, an evaluation of these characteristics is vital to
(Dw) for a given beam collimator aperture (A) at a reference determine if the technology adopted for the Epi-5B skin
depth (d) to the dose at the same point and depth for the refer- diode can be used for quality assurance on medical linac. The
ence collimator aperture (Aref)35,36: DPP were performed by measuring the diode and the Attix
chamber’s response placed at a 1.5 cm depth and irradiated
Dw ðA; dÞ with a 10 9 10 cm2 field of 6 MV photon beam at 600 MU/
Output Factor ¼
Dw ðAref ; dÞ min within the phantom, varying SSD from 100 cm up to

Medical Physics, 44 (10), October 2017


5407 Vicoroski et al.: Silicon diode for skin dosimetry 5407

250 cm. The ratio between the charge measured by the sili-
con diode to the charge measured by the Attix ionization
chamber represents the DPP dependence of the detector and
is normalized to the dose per pulse of 2.78 104 Gy/pulse
corresponding to the reference SSD of 100 cm. This method
avoids variation of the spectrum of the radiation and can be
used to characterize only the dose rate dependence assuming
that the ion chamber’s dose per pulse dependence is
negligible.42

2.I. Angular dependence


The angular dependence of the skin diode (Epi-5B)
response was evaluated in two different conditions. The con-
dition (a) corresponds to the detector placed in a cylindrical
PMMA phantom at 15 cm depth (aligned with the central
FIG. 4. Response variation of the skin diode (Epi-5B) as a function of accu-
axis of the phantom) and the detector placed at isocenter of a
mulated dose from a Co-60 gamma source.
10 9 10 cm2 field of 6 MV photon beam. 100 MU are deliv-
ered at each angle and the response normalized to the 0
degree, corresponding to the beam perpendicular to the
detector’s face-up configuration. The condition (b) corre-
sponds to the detectors placed at surface of a 30 9 30 cm2
Virtual Water phantom with 10 cm of backscattering. In this
configuration, the Epi-5B is benchmarked, after correction
for its intrinsic angular dependence, to the ion chamber Attix
using 10 9 10 cm2 field size 6 MV photon beam at SSD of
100 cm and to EBT3 film. The response of the film is sam-
pled at the center of the field and averaging the response from
9 9 9 voxels corresponding to an equivalent area of
2.7 9 2.7 mm2. The gantry has been rotated from  60o to
+ 60o in 15 degree increments and responses normalized to
the 0 degree gantry position.

3. RESULTS
FIG. 5. Dose linearity response for the Epi-5B (squared symbols) and Epi-
3.A. Radiation hardness 1C (dotted) epitaxial silicon diodes. [Color figure can be viewed at wileyonli-
Figure 4 shows the variation of the response of the n-type nelibrary.com]
skin diode (Epi-5B) as a function of the dose accumulated
from the Co-60 gamma source. As expected there is a TABLE I. Dose calibration factors of the Epi-5B and Epi-1C.
response decrease with accumulated dose; however, the
response becomes stable with a variation of approximately Diodes Response (pC/cGy)
0.05%/kGy after 60 kGy.
Epi-1C 115.18  0.14
Epi-5B 53.12  0.03
3.B. Dose linearity
Figure 5 displays the dose linearity of the accumulated
dose varying from 50 cGy to 500 cGy of the Epi-5B and
3.C. Dose per pulse dependence
Epi-1C epitaxial diodes after preirradiation by 60 kGy.
Table I summarized the dose calibration factors from charge Figure 6 demonstrates the dose per pulse dependence
to dose extracted from the linearity measurements. The con- (DPP) response of the skin diode Epi-5B normalized to the
version factors demonstrate that despite the sensitivity of the dose measured by the ion chamber at 2.78 9 104 Gy/pulse
Epi-5B being very low in comparison with the thicker epitax- corresponding to a 6 MV photon beam, 600 MU/min, with a
ial diode Epi-1C, its reproducibility is still within 0.2% with field size of 10 9 10 cm2 and at 1.5 cm depth in solid water
excellent linearity. The reproducibility of the diode detectors and SSD 100 cm.
is determined based on two standard deviations of five The skin diode exhibits a decrease of 50% in sensitivity
repeated measurements of each accumulated dose. for a 10 times fold dose rate (Gy/pulse) reduction.

Medical Physics, 44 (10), October 2017


5408 Vicoroski et al.: Silicon diode for skin dosimetry 5408

3.E. Output factor and surface field size


dependence
Figure 9(a) displays the output factor measured by the
skin diode and compared to EBT3 film and MOSkin detector
at 10 cm depth within the Virtual Water phantom. As
expected, MOSkin and EBT3 film agree within 1.2% from a
square field size (SFS) of 0.5 up to 30 cm, confirming the
results obtained previously for small field dosimetry stud-
ies.35,44 The skin diode in the face-up configuration has been
tested from a SFS of 2 up to 30 cm with an agreement within
2% in respect to EBT3 film. The lack of data below a SFS of
2 cm is due to a poor signal-to-noise ratio as the dose is low
at very small field sizes and at a depth of 10 cm in the solid
water phantom.
FIG. 6. Dose per pulse (DPP) response of the skin diode Epi-5B estimated at In addition, Fig. 9(b) shows the response’s dependence to
Dmax (1.5 cm) in solid water with a 6 MV photon beam of 10 9 10 cm2 field size variations at surface of the Attix chamber and the
field size varying SSD from 100 cm to 250 cm.
skin diode. The responses are normalized to a field size of
10 9 10 cm2 and WED of the Attix chamber adjusted using
extra thin sheets of kapton above the IC. The configuration
with 0.05 mm of build-up that is corresponding to a total
3.D. Percentage depth dose
WED of 0.075 mm matches the skin diode response within
Figure 7(a) shows the comparison of the percentage 3% for all the field sizes investigated.
depth dose (PDD) measurements from surface to 20 cm
depth in Virtual Water with the skin diode and the Attix
3.F. Angular dependence
chamber normalized to dmax for a 6 MV photon beam with
10 9 10 cm2 radiation field area. The observed maximum The surface dose increases when beam’s incidence angle
difference between the PDD measured with the skin diode increases (tangential beams); correct measurements of the
and calculated with Geant4, excluding the outsider point at skin dose at different angles of beam incidence are valuable
1 cm depth, is within 2% with an agreement within for both treatment planning system (TPS) verification36 and
+ 0.5% at a water equivalent depth of 0.07 mm. The differ- for studies correlating acute skin toxicity to dose delivered to
ence between skin diode and the uncorrected Attix cham- the skin in head and neck or breast cancer treatments. In order
ber response at 0.07 mm depth is approximately + 2%. to measure accurately the skin dose from tangential beams,
This difference reduces to 0.2% if the ion chamber the intrinsic angular dependence of the skin diode must be
response is corrected for skin dosimetry, as suggested by compensated for. The angular response of the skin diode in
Gerbi and Khan implying an effective depth of measure- face-up configuration was investigated by placing the device
ment for the uncorrected response of approximately in a central axis of a cylindrical PMMA phantom of radius
0.03 mm.27,28,43 15 cm and aligned with the linac’s isocenter. Figure 10
Figure 8(a) displays the comparison of the PDD measured shows the result for the intrinsic angular dependence of the
in standard conditions between the thick epitaxial Epi-1C Epi-5B normalized to the response of the detector at gantry
diode and the ion chamber Attix in Virtual Water. The angle 0 degree (beam incident normally). The behavior of the
observed maximum difference between the PDD measure- device confirms what is expected from a planar device with a
ments is + 5.5% representing an over-response of the epitax- very thin sensitive volume: the largest variation ( 8%)
ial Epi-1C diode on a surface of the phantom while its occurs at 90 degree and the attenuation of the beam incident
agreement with the Attix ionization chamber at depths greater from the backside of the detector is approximately  2%.
than 5 mm is within 1.5%. The correction factor for a tangential beam with angle of 60
This measurement was performed to show how sensitive degree, as expected in a 3D conformal treatment of breast
the WED is to the thickness of the epitaxial layer and the fab- cancer, will be 5% at largest.
rication of an effective skin diode, even if the diodes utilize In order to evaluate the skin diode ability to measure skin
the same “drop-in” packaging technology. Another example dose at various beam incidence angles, 100 MU of a 6 MV
of the accuracy required in the design of a skin dosimeter photon beam, field size 10 9 10 cm2, was delivered with the
using silicon is given by the results showed in Fig. 8(b) gantry position rotated from  60o to + 60o in increments of
where the skin diode has been irradiated face-down. The Epi- 15 degree. The skin diode and the EBT3 film were placed on
5B face-down over-responses with + 20% more dose a surface of the phantom and their readings were normalized
recorded at surface in comparison to the Attix ionization to the reading obtained at 0o gantry angle with a fixed SSD of
chamber. Obtained results are expected due to the steep dose 100 cm. Figure 11 shows the comparison of the angular
gradient close to the surface. dependence of the skin diode response (corrected for its

Medical Physics, 44 (10), October 2017


5409 Vicoroski et al.: Silicon diode for skin dosimetry 5409

(a) (b)

FIG. 7. (a) Percentage depth dose measured by the skin diode in comparison to the Attix ionization chamber with no build-up cap and no correction factors
applied; (b) Comparison of the same dataset measured with Attix and Epi-5B with the simulations performed by the means of Geant4 in water and represented in
logarithmic scale of the depth in water. [Color figure can be viewed at wileyonlinelibrary.com]

(a) (b)

FIG. 8. PDD measured by Epi-1C in comparison to Attix IC in a solid water phantom; (a) Epi-1C is “face-up” (b) Epi-5B is “face-down”.

intrinsic angular dependence), the film and the Attix ionizing Dose rate dependence in silicon diodes can potentially
chamber, as the beam incidence angle increases. The mea- affect the capability of the detector to reconstruct accurately
sured dose increases due to Dmax shifting toward the surface. dose profiles and depth dose distributions. The skin diode
shows a variation of less than 5% within the range between
104 and 2.78 9 104 Gy/pulse confirmed by the PDD mea-
4. DISCUSSION
surements that show an agreement within 1% compared to
The skin diode (Epi-5B) is manufactured on n-type epitax- the parallel plate ion chamber.
ial silicon substrate which shows a radiation damage effect as The Geant4 simulations of the PDD in build-up region have
variation of the detector response as a function of the accu- been used to derive the effective WED of 0.07 mm and
mulated dose. Figure 6 shows that the response of the skin 0.03 mm of the skin diode and the Attix chamber, respectively.
diode decreases by a factor of 25% in the first 20 kGy of The thickness of the sensitive volume of the detector in
accumulated dose. For higher doses the response variation is step dose gradients is a major factor for accuracy in skin
within 2% only that is explained by the saturation of the dif- dosimetry at water depth of 0.07 mm. The Epi-1C epitaxial
fusion length of the minority carriers. This is in agreement diode with 50 lm thick epitaxial layer was found to have a
with previous results on thinned and epitaxial silicon detector WED of 1 mm, making this diode unsatisfactory for skin
studies on substrate with similar resistivity.32,45 dosimetry.

Medical Physics, 44 (10), October 2017


5410 Vicoroski et al.: Silicon diode for skin dosimetry 5410

(a) (b)

FIG. 9. (a) Output factor measured by MOSkin, EBT3 film, and skin diode Epi-5B at 10 cm depth in Virtual Water phantom and (b) field size dependence of the
skin diode placed at surface of the phantom in comparison with Attix with different build-up thickness. [Color figure can be viewed at wileyonlinelibrary.com]

FIG. 10. Normalized intrinsic angular response of the skin diode placed in
center of a cylindrical PMMA phantom of radius of 15 cm as a function of
the angle. Center of the phantom is placed at isocenter. Normal incident FIG. 11. Comparison of the normalized surface dose angular dependence
angle is 0 degree. measured by the skin diode, EBT3 film, and Attix chamber at surface of a
Virtual Water phantom for a 6 MV photon beam. [Color figure can be
viewed at wileyonlinelibrary.com]

The skin diode with epitaxial layer 0.005 mm has demon-


strated to be suitable for skin dosimetry with a WED of because the WED of the IC is about 0.025 mm, in contrast to
0.075 mm. The angular response at surface of the phantom the WED of the skin diode, estimated to be approximately
agrees within 3% of relative variation between the skin diode 0.075 mm.28 Adding two kapton sheets, each one 25 lm
and EBT3 film with beam incident angles in the range thick, and matching field size response of the IC to the skin
between 60 degree. The skin diode shows a minimal angu- diode confirms that the skin diode WED is approximately
lar dependence comparable with the film response, but it also between 0.07 and 0.08 mm, as estimated by the Geant4
measures the dose at the WED of 0.07 mm which cannot be simulations.
achieved by film. Film has a WED of dose measurement at The use of the skin diode for low dose rate measurements
approximately 0.1 mm due to the polyamide protection layer is challenging due to the trade-off between thin sensitive vol-
above the radiation sensitive polymeric substrate (0.05 mm ume and signal-to-noise ratio. This was primarily observed
thick).19,43 The Attix IC response shows a large over-response during output factor measurements, where measurements in
up to + 20% at 60 degree as confirmed by other studies.24 fields below 2 9 2 cm2 were impossible with the electrome-
Skin dose increases also as a function of the field size and ter adopted. Adopting a larger area for the p+ top junction
depends on WED of measurement. In this sense, the skin without enlarging the overall die’s size by reducing the exter-
diode and Attix IC’s response with field size are different nal guard-ring would double the sensitivity of the sensor

Medical Physics, 44 (10), October 2017


5411 Vicoroski et al.: Silicon diode for skin dosimetry 5411

solving the issue of the signal-to-noise ratio. This will REFERENCES


increase the amount of charge collected without perturbing 1. Fisher B, Anderson S, Bryant J, et al. Twenty-year follow-up of a ran-
the beam with a larger silicon die. Although variation of the domized trial comparing total mastectomy, lumpectomy, and lumpec-
detector response as a function of temperature in a thin epi- tomy plus irradiation for the treatment of invasive breast cancer. N Engl
taxial diode is expected to be negligible because diffusion J Med. 2002;347:1233–1241.
2. Veronesi U, Cascinelli N, Mariani L, et al. Twenty-year follow-up of a
length of the charge carriers is shorter than thickness of the randomized study comparing breast-conserving surgery with radical
epitaxial layer, the skin diode response variation with tem- mastectomy for early breast cancer. N Engl J Med. 2002;347:1227–
perature will be investigated in a separate work. For a clini- 1232.
cal application of the skin diode also the electron 3. Early Breast Cancer Trialists’ Collaborative Group. (EBCTCG), Effect
of radiotherapy after breast-conserving surgery on 10-year recurrence
contamination effects due to different collimator configura- and 15-year breast cancer death: meta-analysis of individual patient data
tions and presence of the flattening filter will be investi- for 10 801 women in 17 randomised trials. Lancet. 2011;378:771–784.
gated and reported. 4. ICRP Task Group. The biological basis for dose limitation in the skin. A
report of a Task Group of Committee 1 of the International Commission
on Radiological Protection. Ann ICRP. 1992;22:1–104.
5. CONCLUSION 5. Mellenberg DE. Determination of build-up region over-response correc-
tions for a Markus-type chamber. Med Phys. 1990;17:1041–1044.
This work presents a feasibility study toward the develop- 6. Shani G. Radiation Dosimetry Instrumentation and Methods, 2nd Edn.
Boca Raton, FL: CRC Press LLC; 2001, pp. 205. ISBN 0-8493-1505-0.
ment of a silicon diode that can measure the skin dose accu- 7. Nilsson B, Montelius A. Fluence perturbation in photon beams under
rately in respect to clinical standards for QA of medical nonequilibrium conditions. Med Phys. 1986;13:191–195.
linacs. The use of a diode for skin dose measurements is of 8. Kawachi T, Saitoh H, Inoue M, et al. Reference dosimetry condition and
interest due to the possibility of developing an array of such beam quality correction factor for CyberKnife beam. Med Phys.
2008;35:4591–4598.
detectors for real-time, in vivo skin dose mapping in several 9. Hardcastle N, Soisson E, Metcalfe P, Rosenfeld AB, Tome WA. Dosi-
radiotherapy and diagnostic imaging modalities, where the metric verification of helical tomotherapy for total scalp irradiation.
risk of acute skin toxicity is a serious issue for the patients’ Med Phys. 2008;35:5061–5068.
quality of life. The skin diode is fabricated on a 0.005 mm 10. Butson MJ, Cheung T, Yu PKN. Scanning orientation effects on Gaf-
chromic EBT film dosimetry. Australasian Phys Eng Sci Med.
thick epitaxial silicon layer placed above 0.4 mm thick sili- 2006;29:281–284.
con substrate and embedded in a thin water equivalent pack- 11. Ostwald PM, Kron T, Hamilton CS, Denham JW. Clinical use of carbon-
age using “drop-in” packaging technology. The skin diode loaded thermoluminescent dosimeters for skin dose determination. Int J
has a WED of 0.075  0.005 mm, as recommended by the Radiat Oncol Biol Phys. 1995;33:943–950.
12. Mijnheer B, Beddar S, Izewska J, Reft C. In vivo dosimetry in external
ICRP and can be used to measure the skin dose delivered by beam radiotherapy. Med Phys. 2013;40:070903.
oblique radiation beams. 13. Colussi VC, Beddar S, Kinsella TJ, Sibata CH. In vivo dosimetry using
The skin diode can also be used for conventional QA a single diode for megavoltage photon beam radiotherapy: implementa-
tion and response characterisation. J Appl Clin Med Phys. 2001;2:210–
dosimetry in radiation therapy because it has a linear dose–
218.
response over a dynamic range from 25 to 500 cGy and mini- 14. Zhuang AH, Olch AJ. Validation of OSLD and a treatment planning sys-
mal intrinsic angular dependence, in conjunction with the tem for surface dose determination in IMRT treatments. Med Phys.
ability to measure the PDD within 2% compared to IC. 2014;41:081720.
15. Jong WL, Wong JHD, Ng K, et al. Characterisation of MOSkin detector
The challenge to be addressed is the improvement of sig- for in vivo dose measurement during megavoltage radiotherapy. J Appl
nal-to-noise ratio for low dose rate applications, which could Clin Med Phys. 2014;15:120–130.
be achieved by increasing the sensitive area of the skin diode 16. Kwan IS, Rosenfeld AB, Qi ZY, et al. Skin dosimetry with new MOS-
and/or electrometer sensitivity, and could possibly lead to a FET detectors. Rad Meas. 2008;43:929–932.
17. Jursinic PA. Angular dependence of dose sensitivity of surface diodes.
universal diode for QA in radiation therapy. Med Phys. 2009;36:2165–2171.
18. AAPM Report No. 87. Diode in vivo dosimetry for patients receiving
external beam radiation therapy. Report of Task Group 62 of the Radia-
ACKNOWLEDGMENTS tion Therapy Committee; 2005: 1-64.
19. Meiler RJ, Podgorsak MB. Characterization of the response of commer-
Authors thank Dr Justin Davies for his support with irradi- cial diode detectors used for in vivo dosimetry. Med Dosim. 1997;22:31–
ating the detectors at the GATRI facility (ANSTO, Lucas 37.
20. Yorke E, Alecu R, Ding L, et al. Diode in vivo dosimetry for patients
Heights, NSW, Australia). Thank you to FBK Foundation receiving external beam radiation therapy. Report of Task Group 62 of
Trento (Italy) for providing the n-type silicon epitaxial the Radiation Therapy Committee; 2005. Madison, WI.
samples for testing throughout the project. 21. Scalchi P, Francescon P. Calibration of a MOSFET detection system for
6-MV in vivo dosimetry. Int J Radiat Oncol Biol Phys. 1998;40:987–
993.
CONFLICT OF INTEREST 22. Wong JHD, Fuduli I, Carolan M, et al. Characterization of a novel two
dimensional diode array the “magic plate” as a radiation detector for
No conflict of interest or financial involvement to be radiation therapy treatment. Med Phys. 2012;39:2544–2558.
disclosed. 23. Petasecca M, Cullen A, Fuduli I, et al. X-Tream: a novel dosimetry sys-
tem for synchrotron microbeam radiation therapy. J Instrum. 2012;
P07022:1–13.
24. Petasecca M, Alhujaili S, Aldosari AH, et al. Angular independent sili-
a)
Author to whom correspondence should be addressed. Electronic mail: con detector for dosimetry in external beam radiotherapy. Med Phys.
marcop@uow.edu.au. 2015;42:4708–4718.

Medical Physics, 44 (10), October 2017


5412 Vicoroski et al.: Silicon diode for skin dosimetry 5412

25. Das IJ, Cheng CW, Watts RJ, et al. Accelerator beam data commission- 35. Sauer OA, Wilbert J. Measurement of output factors for small photon
ing and procedures. Report of the Task Group 106 Therapy Physics beams. Med Phys. 2007;34:1983–1990.
Committee of the AAPM, August; 2008. 36. Khan FM. The Physics of Radiation Therapy, 3rd edn. Philadelphia: Lip-
26. Cross P. The use of parallel plate ion chambers to determine surface pincott Williams & Wilkins; 2003.
dose of a 6 MV photon beam. Australas Phys Eng Sci Med. 37. Kwan IS, Wilkinson D, Cutajar D, et al. The effect of rectal heterogene-
1992;15:208–213. ity on wall dose in high dose rate brachytherapy. Med Phys.
27. Gerbi BJ, Khan FM. Plane-parallel ionization chamber response in the 2009;36:224–232.
buildup region of obliquely incident photon beams. Med Phys. 38. Hardcastle N, Cutajar DL, Metcalfe PE, et al. In vivo real-time rectal wall
1997;24:873–878. dosimetry for prostate radiotherapy. Phys Med Biol. 2010;55:3859–3871.
28. Gerbi BJ. The response characteristics of a newly designed plane-paral- 39. Qi ZY, Deng XW, Huang SM, et al. Real-time in vivo dosimetry with
lel ionization chamber in high-energy photon and electron beams. Med MOSFET detectors in serial tomotherapy for head and neck cancer
Phys. 1993;20:1411–1415. patients. Int J Radiat Oncol Biol Phys. 2011;80:1581–1588.
29. AFE0064. Texas Instruments, 64 Channel Analog Front End for Digital 40. Alnawaf H, Butson M, Yu PK. Measurement and effects of MOSKIN
X-Ray Detector. http://www.ti.com/lit/ds/symlink/afe0064.pdf. Accessed detectors on skin dose during high energy radiotherapy treatment. Aus-
September 3; 2009. tralas Phys Eng Sci Med. 2012;35:321–328.
30. Fuduli I, Newall MK, Espinoza AA, et al. Multichannel data acquisition 41. Rikner G, Grusell E. Effects of radiation damage on p-type silicon detec-
systems comparison for quality assurance in external beam radiation tors. Phys Med Biol. 1983;28:1261–1267.
therapy. Rad Meas. 2014;71:338–341. 42. Wang Y, Easterling SB, Ting JY. Ion recombination corrections of ion-
31. Aldosari AH, Espinoza A, Robinson D, et al. Characterization of an ization chambers in flattening filter-free photon radiation. Appl Clin
innovative p-type epitaxial diode for dosimetry in modern external beam Med Phy. 2012;13:3578.
radiotherapy. IEEE Trans Nucl Sci. 2013;60:4705–4712. 43. Devic S. Accurate skin dose measurements using radiochromic film in
32. Bruzzi M, Bucciolini M, Casati M, et al. Epitaxial silicon devices for clinical applications. Med Phys. 2006;33:1116–1124.
dosimetry applications. App Phys Lett. 2007;90:172109. 44. Aldosari AH, Petasecca M, Espinoza A, et al. A two dimensional silicon
33. Agostinelli S, Allison J, Amako K, et al. Geant4-a simulation toolkit. detectors array for quality assurance in stereotactic radiotherapy: Magic-
Nucl Instrum Methods Phys Res, Sect A. 2003;506:250–303. Plate-512. Med Phys. 2014;41:091707.
34. Oborn BM, Williams M, Bailey M, Carolan MG. IMRT treatment moni- 45. Petasecca M, Pignatel GU, Moscatelli F, Passeri D, Caprai G. Numerical
tor unit verification using absolute calibrated BEAMnrc and Geant4 analysis of thinned silicon detectors. Nucl Instrum Meth A.
Monte Carlo simulations. J Phys Conf Ser. 2014;489:012020. 2007;572:319–320.

Medical Physics, 44 (10), October 2017

Das könnte Ihnen auch gefallen