Beruflich Dokumente
Kultur Dokumente
(Received 7 February 2017; revised 14 June 2017; accepted for publication 28 June 2017;
published 12 August 2017)
Purpose: The aim of in vivo skin dosimetry was to measure the absorbed dose to the skin during
radiotherapy, when treatment planning calculations cannot be relied on. It is of particularly impor-
tance in hypo-fractionated stereotactic modalities, where excessive dose can lead to severe skin toxic-
ity. Currently, commercial diodes for such applications are with water equivalent depths ranging
from 0.5 to 0.8 mm. In this study, we investigate a new detector for skin dosimetry based on a silicon
epitaxial diode, referred to as the skin diode.
Method: The skin diode is manufactured on a thin epitaxial layer and packaged using the “drop-in”
technology. It was characterized in terms of percentage depth dose, dose linearity, and dose rate
dependence, and benchmarked against the Attix ionization chamber. The response of the skin diode
in the build-up region of the percentage depth dose (PDD) curve of a 6 MV clinical photon beam
was investigated. Geant4 radiation transport simulations were used to model the PDD in order to esti-
mate the water equivalent measurement depth (WED) of the skin diode. Measured output factors
using the skin diode were compared with the MOSkin detector and EBT3 film at 10 cm depth and at
surface at isocenter of a water equivalent phantom. The intrinsic angular response of the skin diode
was also quantified in charge particle equilibrium conditions (CPE) and at the surface of a solid water
phantom. Finally, the radiation hardness of the skin diode up to an accumulated dose of 80 kGy
using photons from a Co-60 gamma source was evaluated.
Results: The PDD curve measured with the skin diode was within 0.5% agreement of the equivalent
Geant4 simulated curve. When placed at the phantom surface, the WED of the skin diode was esti-
mated to be 0.075 0.005 mm from Geant4 simulations and was confirmed using the response of a
corrected Attix ionization chamber placed at water equivalent depth of 0.075 mm, with the measure-
ment agreement to within 0.3%. The output factor measurements at 10 cm depth were within 2% of
those measured with film and the MOSkin detector down to a field size of 2 9 2 cm2. The dose–re-
sponse for all detector samples was linear and with a repeatability within 0.2%. The skin diode intrin-
sic angular response showed a maximum deviation of 8% at 90 degrees and from 0 to 60 degree is
less than 5%. The radiation sensitivity reduced by 25% after an accumulated dose of 20 kGy but
after was found to stabilize. At 60 kGy total accumulated dose the response was within 2% of that
measured at 20 kGy total accumulated dose.
Conclusions: This work characterizes an innovative detector for in vivo and real-time skin dose mea-
surements that is based on an epitaxial silicon diode combined with the Centre for Medical Radiation
Physics (CMRP) “drop-in” packaging technology. The skin diode proved to have a water equivalent
depth of measurement of 0.075 0.005 mm and the ability to measure doses accurately relative to
reference detectors. © 2017 American Association of Physicists in Medicine [https://doi.org/10.1002/
mp.12469]
Key words: external beam radiotherapy skin dose, silicon detectors, skin dosimetry
5402 Med. Phys. 44 (10), October 2017 0094-2405/2017/44(10)/5402/11 © 2017 American Association of Physicists in Medicine 5402
5403 Vicoroski et al.: Silicon diode for skin dosimetry 5403
(a) (b)
FIG. 1. Simplified diagram (not to scale) of the skin diode Epi-5B topology (a) top view and (b) cross-section view. [Color figure can be viewed at wileyonlineli-
brary.com]
measurement, a second epitaxial silicon diode was fabricated aluminum layers is illustrated in Fig. 3. The build-up mate-
by CMRP on a silicon P-type 50 lm thick epitaxial layer rial thickness above the detector sensitive layer is repro-
grown on a 0.001 Ohm-cm substrate (named Epi-1C – ducible and already proved to be able to provide a WED of
Fig. 2). The Epi-1C has N+ phosphorus implanted active approximately 0.07 mm.16 The “drop-in” technology is
area of 0.6 9 0.6 mm2 and overall die dimensions of compatible with standard flexible printed circuit manufac-
1.5 9 1.5 mm2. This diode has been adopted for the fabrica- turing technologies and avoids the use of high atomic num-
tion of the 11 9 11 2D diode array MagicPlate.22 ber materials for packaging of semiconductor radiation
Both epitaxial diodes have been embedded in kapton detectors, minimizing the perturbation of the beam. The
pigtails with 0.6 9 3 mm2 cross section and 350 mm kapton pigtail outside of the diode is shielded by a thin
length using CMRP proprietary “drop-in” packaging tech- aluminum foil and grounded to minimize radiofrequency
nology. This packaging technique is based on tab bonding interference.23 Further details on the “drop-in” packaging
of the silicon die underneath a stack of polyamide and technology can be found elsewhere.24
(a) (b)
FIG. 2. (a) Top view diagram of the Epi-1C diode topology and (b) cross-section view, not in scale. [Color figure can be viewed at wileyonlinelibrary.com]
1C and compared to the Attix ionization chamber. All these The reference field size for this study was 10 9 10 cm2,
detectors were placed, one at a time, in a recess machined in which was measured at isocenter and at a depth of 10 cm
a Virtual Water (VW) slab to ensure consistency of the scat- (SSD is 90 cm), with delivered 100 MU each time. The out-
tering conditions. A 10 cm thick slab of 30 9 30 cm2 VW put factor was measured by placing the preirradiated skin
phantom is used for backscattering and several slabs ranging diodes (60 kGy irradiation dose by Co-60) in the center of
from 0.1 cm to 20 cm are used to obtain the depth dose pro- the field for field sizes and varying the jaws’ aperture from
files. The samples are irradiated with 100 MU of a 6 MV 2 9 2 cm2 to 30 9 30 cm2. A careful alignment of the
photon beam with a 10 9 10 cm2 field size at 100 cm SSD detector was performed by maximizing the detector response
for each measurement. for the 2 9 2 cm2 beam by a scan in the SUP-INF and lateral
Additionally, PDD were obtained for the skin diode Epi- directions using the couch positioning system (spatial resolu-
5B with the epitaxial layer of the skin diode facing down- tion of 1 mm). A measurement of the field factor at surface
wards and the diode being exposed to the incident radiation of the phantom was also performed to evaluate the capabili-
beam. This was conducted under the same experimental con- ties of the detector to measure the surface dose as a function
ditions as when the epitaxial layer of the skin diode was fac- of the field size defined by the jaws. The silicon epitaxial
ing up, to determine whether orientation of the detector diode was benchmarked against Gafchromic EBT3 film, the
would affect the PDD. Attix ionization chamber, and the MOSkin, which are fre-
Monte Carlo Geant4 version 10.0p0133 simulations were quently used for interface measurements and described in
performed to determine the dose distribution at the build-up several studies.16,37–40
region generated by the specific Linac used for this experi- The films were calibrated using 1.5 9 1.5 cm² squares
ment. The beam 10 9 10 cm2, 6 MV x-ray, passing through and scanned using a flatbed scanner (Epson 10000XL scan-
a 30 9 30 9 30 cm3 water equivalent block phantom was ner; Epson America, Inc. Long Beach, CA, USA) 24 h after
simulated. irradiation to allow for postirradiation development. The
The source to surface distance (SSD) was 100 cm, and films were scanned in transmission mode with color correc-
the beams were fired from phase space files created from tion turned off, at a resolution of 72 dpi and 48 bit RGB data
an EGSnrc Monte Carlo based system that models the Var- format. Analysis of three scanning repetitions for each film
ian 2100C Linac at Illawarra Cancer Centre in Wollon- placed with the same orientation were performed using Ima-
gong (Australia). The model used to represent the Linac geJ 1.46r software (National Institute of Health, Bethesda,
and the phantom setup, along with the physics of trans- MD, USA) and one standard deviation of the optical density
port, has been already validated by previous experimental was used as uncertainty. A template was used to scan the
results.34 The physics processes modeled in the simulation films at the center of the scanner bed to avoid nonuniformity
are from the Geant4 electromagnetic standard physics associated with the position of the film in respect to the scan-
package and include photoelectric effect, Compton scat- ner lamp or sensor. Only the red channel was used for analy-
tering and gamma conversion (photons), ionization, sis of a region of interest (ROI) at the center of the film. The
Bremsstrahlung, and positron annihilation. The particle films used to perform the measurements have been pres-
range cut was set to 0.1 mm and the electron/positron canned to evaluate the background optical density and to take
maximum step length was 0.1 mm. Dose was scored care of the nonuniformities associated with the specific film
inside the phantom at a voxel size of 0.02 9 1 9 1 mm3. cut. Butson et al.10 provided in-depth details of the prepara-
Each simulation was split into 10 parallel jobs each with tion and scanning procedure adopted for the analysis of the
unique seeds and the mean dose reported. The standard EBT3 film.
deviation is taken across 10 simulations to evaluate uncer-
tainties. In total, 4 9 1013 primary histories are simulated
2.H. Dose per pulse dependence
where the primary histories represent the number of elec-
trons hitting the x-ray target in the linac head. The dose The measurement of dose per pulse dependence (DPP)
uncertainty is approximately 0.5% of the dose at 1.5 cm refers to the evaluation of the change of the detector sensitiv-
depth at the beam central axis (CAX) for the 10 9 10 cm2 ity due to variation of the instantaneous dose rate in a pulsed
field size. radiation beam typical of a medical linear accelerator.41 It
was reported by Rikner and Grusell that n-type silicon diodes
2.G. Output factor and surface field size are more sensitive to dose per pulse variation than compared
dependence to p-type silicon detectors, thus showing an increase in sensi-
tivity with increasing dose per pulse.41
The output factor is defined as the ratio of dose in water Therefore, an evaluation of these characteristics is vital to
(Dw) for a given beam collimator aperture (A) at a reference determine if the technology adopted for the Epi-5B skin
depth (d) to the dose at the same point and depth for the refer- diode can be used for quality assurance on medical linac. The
ence collimator aperture (Aref)35,36: DPP were performed by measuring the diode and the Attix
chamber’s response placed at a 1.5 cm depth and irradiated
Dw ðA; dÞ with a 10 9 10 cm2 field of 6 MV photon beam at 600 MU/
Output Factor ¼
Dw ðAref ; dÞ min within the phantom, varying SSD from 100 cm up to
250 cm. The ratio between the charge measured by the sili-
con diode to the charge measured by the Attix ionization
chamber represents the DPP dependence of the detector and
is normalized to the dose per pulse of 2.78 104 Gy/pulse
corresponding to the reference SSD of 100 cm. This method
avoids variation of the spectrum of the radiation and can be
used to characterize only the dose rate dependence assuming
that the ion chamber’s dose per pulse dependence is
negligible.42
3. RESULTS
FIG. 5. Dose linearity response for the Epi-5B (squared symbols) and Epi-
3.A. Radiation hardness 1C (dotted) epitaxial silicon diodes. [Color figure can be viewed at wileyonli-
Figure 4 shows the variation of the response of the n-type nelibrary.com]
skin diode (Epi-5B) as a function of the dose accumulated
from the Co-60 gamma source. As expected there is a TABLE I. Dose calibration factors of the Epi-5B and Epi-1C.
response decrease with accumulated dose; however, the
response becomes stable with a variation of approximately Diodes Response (pC/cGy)
0.05%/kGy after 60 kGy.
Epi-1C 115.18 0.14
Epi-5B 53.12 0.03
3.B. Dose linearity
Figure 5 displays the dose linearity of the accumulated
dose varying from 50 cGy to 500 cGy of the Epi-5B and
3.C. Dose per pulse dependence
Epi-1C epitaxial diodes after preirradiation by 60 kGy.
Table I summarized the dose calibration factors from charge Figure 6 demonstrates the dose per pulse dependence
to dose extracted from the linearity measurements. The con- (DPP) response of the skin diode Epi-5B normalized to the
version factors demonstrate that despite the sensitivity of the dose measured by the ion chamber at 2.78 9 104 Gy/pulse
Epi-5B being very low in comparison with the thicker epitax- corresponding to a 6 MV photon beam, 600 MU/min, with a
ial diode Epi-1C, its reproducibility is still within 0.2% with field size of 10 9 10 cm2 and at 1.5 cm depth in solid water
excellent linearity. The reproducibility of the diode detectors and SSD 100 cm.
is determined based on two standard deviations of five The skin diode exhibits a decrease of 50% in sensitivity
repeated measurements of each accumulated dose. for a 10 times fold dose rate (Gy/pulse) reduction.
(a) (b)
FIG. 7. (a) Percentage depth dose measured by the skin diode in comparison to the Attix ionization chamber with no build-up cap and no correction factors
applied; (b) Comparison of the same dataset measured with Attix and Epi-5B with the simulations performed by the means of Geant4 in water and represented in
logarithmic scale of the depth in water. [Color figure can be viewed at wileyonlinelibrary.com]
(a) (b)
FIG. 8. PDD measured by Epi-1C in comparison to Attix IC in a solid water phantom; (a) Epi-1C is “face-up” (b) Epi-5B is “face-down”.
intrinsic angular dependence), the film and the Attix ionizing Dose rate dependence in silicon diodes can potentially
chamber, as the beam incidence angle increases. The mea- affect the capability of the detector to reconstruct accurately
sured dose increases due to Dmax shifting toward the surface. dose profiles and depth dose distributions. The skin diode
shows a variation of less than 5% within the range between
104 and 2.78 9 104 Gy/pulse confirmed by the PDD mea-
4. DISCUSSION
surements that show an agreement within 1% compared to
The skin diode (Epi-5B) is manufactured on n-type epitax- the parallel plate ion chamber.
ial silicon substrate which shows a radiation damage effect as The Geant4 simulations of the PDD in build-up region have
variation of the detector response as a function of the accu- been used to derive the effective WED of 0.07 mm and
mulated dose. Figure 6 shows that the response of the skin 0.03 mm of the skin diode and the Attix chamber, respectively.
diode decreases by a factor of 25% in the first 20 kGy of The thickness of the sensitive volume of the detector in
accumulated dose. For higher doses the response variation is step dose gradients is a major factor for accuracy in skin
within 2% only that is explained by the saturation of the dif- dosimetry at water depth of 0.07 mm. The Epi-1C epitaxial
fusion length of the minority carriers. This is in agreement diode with 50 lm thick epitaxial layer was found to have a
with previous results on thinned and epitaxial silicon detector WED of 1 mm, making this diode unsatisfactory for skin
studies on substrate with similar resistivity.32,45 dosimetry.
(a) (b)
FIG. 9. (a) Output factor measured by MOSkin, EBT3 film, and skin diode Epi-5B at 10 cm depth in Virtual Water phantom and (b) field size dependence of the
skin diode placed at surface of the phantom in comparison with Attix with different build-up thickness. [Color figure can be viewed at wileyonlinelibrary.com]
FIG. 10. Normalized intrinsic angular response of the skin diode placed in
center of a cylindrical PMMA phantom of radius of 15 cm as a function of
the angle. Center of the phantom is placed at isocenter. Normal incident FIG. 11. Comparison of the normalized surface dose angular dependence
angle is 0 degree. measured by the skin diode, EBT3 film, and Attix chamber at surface of a
Virtual Water phantom for a 6 MV photon beam. [Color figure can be
viewed at wileyonlinelibrary.com]
25. Das IJ, Cheng CW, Watts RJ, et al. Accelerator beam data commission- 35. Sauer OA, Wilbert J. Measurement of output factors for small photon
ing and procedures. Report of the Task Group 106 Therapy Physics beams. Med Phys. 2007;34:1983–1990.
Committee of the AAPM, August; 2008. 36. Khan FM. The Physics of Radiation Therapy, 3rd edn. Philadelphia: Lip-
26. Cross P. The use of parallel plate ion chambers to determine surface pincott Williams & Wilkins; 2003.
dose of a 6 MV photon beam. Australas Phys Eng Sci Med. 37. Kwan IS, Wilkinson D, Cutajar D, et al. The effect of rectal heterogene-
1992;15:208–213. ity on wall dose in high dose rate brachytherapy. Med Phys.
27. Gerbi BJ, Khan FM. Plane-parallel ionization chamber response in the 2009;36:224–232.
buildup region of obliquely incident photon beams. Med Phys. 38. Hardcastle N, Cutajar DL, Metcalfe PE, et al. In vivo real-time rectal wall
1997;24:873–878. dosimetry for prostate radiotherapy. Phys Med Biol. 2010;55:3859–3871.
28. Gerbi BJ. The response characteristics of a newly designed plane-paral- 39. Qi ZY, Deng XW, Huang SM, et al. Real-time in vivo dosimetry with
lel ionization chamber in high-energy photon and electron beams. Med MOSFET detectors in serial tomotherapy for head and neck cancer
Phys. 1993;20:1411–1415. patients. Int J Radiat Oncol Biol Phys. 2011;80:1581–1588.
29. AFE0064. Texas Instruments, 64 Channel Analog Front End for Digital 40. Alnawaf H, Butson M, Yu PK. Measurement and effects of MOSKIN
X-Ray Detector. http://www.ti.com/lit/ds/symlink/afe0064.pdf. Accessed detectors on skin dose during high energy radiotherapy treatment. Aus-
September 3; 2009. tralas Phys Eng Sci Med. 2012;35:321–328.
30. Fuduli I, Newall MK, Espinoza AA, et al. Multichannel data acquisition 41. Rikner G, Grusell E. Effects of radiation damage on p-type silicon detec-
systems comparison for quality assurance in external beam radiation tors. Phys Med Biol. 1983;28:1261–1267.
therapy. Rad Meas. 2014;71:338–341. 42. Wang Y, Easterling SB, Ting JY. Ion recombination corrections of ion-
31. Aldosari AH, Espinoza A, Robinson D, et al. Characterization of an ization chambers in flattening filter-free photon radiation. Appl Clin
innovative p-type epitaxial diode for dosimetry in modern external beam Med Phy. 2012;13:3578.
radiotherapy. IEEE Trans Nucl Sci. 2013;60:4705–4712. 43. Devic S. Accurate skin dose measurements using radiochromic film in
32. Bruzzi M, Bucciolini M, Casati M, et al. Epitaxial silicon devices for clinical applications. Med Phys. 2006;33:1116–1124.
dosimetry applications. App Phys Lett. 2007;90:172109. 44. Aldosari AH, Petasecca M, Espinoza A, et al. A two dimensional silicon
33. Agostinelli S, Allison J, Amako K, et al. Geant4-a simulation toolkit. detectors array for quality assurance in stereotactic radiotherapy: Magic-
Nucl Instrum Methods Phys Res, Sect A. 2003;506:250–303. Plate-512. Med Phys. 2014;41:091707.
34. Oborn BM, Williams M, Bailey M, Carolan MG. IMRT treatment moni- 45. Petasecca M, Pignatel GU, Moscatelli F, Passeri D, Caprai G. Numerical
tor unit verification using absolute calibrated BEAMnrc and Geant4 analysis of thinned silicon detectors. Nucl Instrum Meth A.
Monte Carlo simulations. J Phys Conf Ser. 2014;489:012020. 2007;572:319–320.