BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
V. CALIBRATION
- Click “CALIBRATION”
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
- Click “CALIBRATION REQUEST”
- Select the TEST to be calibrated and click “OK”
- Click the “START” to start calibration.
Note: It can be done simultaneously in QC running.
VI. QC RUNNING
- Click the icon “QC REQUEST”
- Select the TEST for QC and click “OK”
- Click “START” to start running.
- Click “STATUS” for monitoring.
a. BY NUMBERS
- Click “SAMPLE REQUEST”
- Select the test to be analyzed and click “OK”. If
there’s more than 1 sample, just select the test for
the corresponding patient number and click “OK” and
if all samples are loaded on the sample rack, click
“START”
Note: Numbering is automatically generated.
b. BY DETAILS
- Click “SAMPLE REQUEST”
- Click “DETAILS”, fill out the following;
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
Sent From (WARD)
Sent By (Requesting Physician, scroll down and
select)
Tested in (select LAB)
Tested by (Medical Technologist on duty, scroll
down and select)
Patient (Complete Name)
Age
Gender
MRN (HOSPITAL NUMBER)
Bed No. (OR NUMBER), if patient’s details are
supplied click “OK”
- Then click “CLOSE”
- Select the test to be analyzed
- Click “OK”
- Click “START”
- Click “STATUS” after clicking start.
NOTE: Clear any obstacles that might block the
movement of the probe to avoid PROBE damage
before clicking “START”…!!! Do not forget to
UNSCREW the cap of the reagents…!!!
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
again the position from 15-40 positions. Position of
calibrators and controls are not included.
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
- Click “QC” icon, and select Day to Day QC.
- For “Tabular Data”
- Select the test you want to view for the QC result.
- Select the Date.
- Select Westgard RULE.
- Select the type of control you want to view.
- Then click “REFRESH”.
- For “Graphical Data”
- Select the test you want to view for the QC result.
- Select the Date
- Select Westgard RULE.
- Select the type of control you want to view.
- Then click “REFRESH” (green dot=normal control, red
dot=pathological/abnormal control).
- Select “QC SUM” to see the actual concentration of
the control (select date, select control, click refresh
- Click results
- Select patient you want to view
XII. MAINTENANCE
- Click “Maintenance”
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
- Click “Daily Maintenance”
XIII. OTHERS
1. FOR HBA1c;
REAGENTS:
LATEX – R1
Antibody – R2
*Antibody Reagent Preparation: Add entire
content of R2(b) Antibody to bottle of HBA1c Buffer
(R2a). Mix antibody working reagent well before use.
Stable for 30 days when properly stored at 2-8C
CALIBRATORS:
*Reconstitute each calibrator vial using 0.5 mL
of distilled water.
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
*Store at 2-8C.
*Protect from light and heat.
*The calibrator retains it assigned value for at least 30
days at 2-8C.
Note: There are 4 Calibrators (values are listed on the
vial label).
CONTROLS:
SAMPLE PREPARATION:
1. Dispense 0.5 ml Hemolysis Reagent into tubes.
2. Place 10uL of well mixed whole blood into the
tube. Mix.
3. Allow to stand for 5 minutes or until complete
lysis is evident.
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
Same procedure in Running Samples, the only
difference is to change the specimen with CSF.
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
- Click Open
- Click “OK”
- *Go to Parameters from main menu
- Click “Off-system”
- Click “Results” from the main menu
- Select patient
- Click “Add”
- Select “Test”, scroll down
- Type result
- Click “OK”, then Click “Close” (Repeat from asterisk
above until you complete the test result, depends on
how many tests), then click “PRINT”
CALIBRATORS:
- Click “CALIBRATION” from the main menu
- Click”CALIBRATOR”
- Select CAL (left box)
- Change Expiration Date
- Enter Lot Number
- Select analyte you want to change CAL. Value (right box)
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
- Click “OK”
Note: There are 4 sets of CALIBRATORS.
CONTROLS:
- Click “QC” from the main menu
- Click”CONTROL”
- Select CONTROL (left box)
- Change Expiration Date
- Enter Lot Number
- Select analyte you want to change CONTROL Value (right
box)
- Type the values on the “Conc”
- Click “OK”
Note: There are 3 sets of CONTROLS.
7. COMMON ERRORS
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
b. Sample or Reagent is EXHAUSTED
- Click “STOP PROBE”
- Click “STATUS”
- Click “SAMPLE DISK” or “REAGENT DISK” where
exhausted error appear
- Check the sample that is empty (exhausted sample or
reagent appear with a white circle at the center)
- Immediately ADD SAMPLE or REAGENT
- After addition, Click “REFRESH” or if more than one
click “REFRESH ALL”
- Then click “START”
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
8. CONVERTION FACTORS
ALBUMIN = 10
AMYLASE = 1.85
TOTAL BILIRUBIN = 17.10
DIRECT BILIRUBIN = 17.10
BUN = 0.357
CALCIUM = 0.25
CHOLESTEROL = 0.026
CREATININE = 88.4
GLUCOSE = 0.0555
HDL-CHOLE = 0.026
MAGNESIUM = 0.5
PHOSPHORUS = 0.323
TRIGLYCERIDE = 0.0113
TOTAL PROTEIN = 10
URIC ACID = 0.0595
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
9. OUTER RING POSITIONS
Position 1 = Water
Position 2 = SB MULTICAL
Position 3 = SB NORMAL CONTROL
Position 4 = SB PATHOLOGIC CONTROL
Position 5 = HDL CALIBRATOR
Position 6 = SR CALIBRATOR
Position 7 = SR NORMAL CONTROL
Position 8 = SR PATHOLOGIC CONTROL
Position 9 = SB HBA1c CAL 1
Position 10=SB HBA1c CAL 2
Position 11=SB HBA1c CAL 3
Position 12=SB HBA1c CAL 4
Position 13=SB HBA1c NORMAL
Position 14=SB HBA1c PATHOLOGIC
Position 15 – 40 = SAMPLES
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
11. HOW TO EDIT OR ADD NORMAL VALUES
- Click “PARAMETERS” from the main menu
- Click TEST
- Select analyte you want to EDIT or ADD normal values
- Click “REFERENCE”
- To EDIT; click existing, type the values on the lower
right part, just below the LOW and HIGH, click “OK”.
- To ADD; click “ADD”, select GENDER (if general,select
blank), select Sample Type, Age (from 0-100), LOW
(lowest range), HIGH (highest range), click “OK”.
Fcc_042016
BS-200 QUICK GUIDE FOR TONDO MEDICAL CENTER
- Plug in, then click the power button at the back of the
machine
- Select “MEASURE” from the menu
- Click Pg D/U to select test
Note: Be sure that you already prepare the standard,
control and sample before clicking the test, you need
3 test tubes for each. See Quick Guide Preparation of
the reagents provided).
- Selcet analyte to run (start from number 28-50)
- Click Zero for blanking (use distilled water)
Feed the distilled water, press the button
- Click standard, feed the standard, press button
- Press control if you are running control if not go
directly to sample, feed the sample, press button