Process Calibration Tools:

Temperature Applications
Application Note

Introduction
Process Calibration Tools (PCT) is
a family of Fluke tools that enable
users to calibrate temperature,
pressure, flow and electrical
sensors, transmitters and gages
products in-situ and in I&C and Mounting means
(thread)
I&E shops.
Terminal block
Terminals Sensing resistor Protective sheath
Temperature
Flange
Electronic thermometers such
as RTDs (resistance rempera-
ture detectors), PRTs (platinum
resistance thermometers) and
thermocouples are often used to
monitor temperatures in pro-
cesses. RTDs and PRTs are used
when accuracy is particularly
important. Thermocouples are
The sensing resistor of an RTD is measured at the terminals of the assembly.
inexpensive and can be used over
wider temperature ranges than
of the thermometers into numeri- For thermocouples the procedure
PRTs and RTDs,but they are less cal temperature. There are also is the same, except voltage is
accurate. Failure modes for these handheld temperature calibrators measured across the terminal
1/4 in

sensors include open circuit, short


3/16 in

that, in addition to displaying rather than resistance. 3/8 in 1/4 in

circuit, and drift. Sensor drift is temperature, enable the electrical 1/4 in
1/8 in

the most common failure mode output of electronic thermometers Insert “A”
and is corrected by calibration. to be calibrated.
Electronic thermometers Temperature sources are 1/4 in
3/16 in

include a protective sheath ovens that generate tempera- 3/8 in 3/8 in

which allows them to be inserted tures for calibrating electronic


3/16 in
1/4 in

directly into the medium to be thermometers. There are two Insert “B”
measured, or they can be placed types: dry-blocks (also called dry
in a thermal well. This is used to wells) that are typically portable, 1/4 in 1/4 in

protect the sensor from vibration and baths, which use liquid 1/4 in 1/4 in

and to allow it to be placed in media to achieve more stable 1/4 in 1/4 in

environments under pressure or temperatures. Baths are normally Insert “C”


where it would otherwise not be larger than dry wells and are not
safe to insert the sheath directly. portable. 4 mm 3 mm

Temperature displays are Dry-block calibrators are 6 mm 6 mm

digital readouts that convert the designed with inserts that contain 3 mm 4 mm

changes in electrical resistance holes bored into them match- Insert “D”
ing standard diameters used for
RTDs, PRTs, and thermocouples. 3 mm
0.25 in

To calibrate an RTD or PRT 10 mm


8 mm

the probe is removed from the 4 mm


6 mm

thermal well, or other location Insert “E”


where it is mounted, and inserted
into the dry-block calibrator. 3 mm 4 mm

The resistance of the probes 6 mm 0.25 in

is measured at the terminal at 4 mm 3 mm

pre-determined temperature set-


RTDs may be inserted into thermal wells. Insert “F”

Some RTDs have large protective sheaths


points supplied by the calibrator. RTDs are placed in the calibrator in holes bored into
and flanges that make calibration difficult. interchangeable metal inserts.

From the Fluke Digital Library @ www.fluke.com/library


Conduit
connection
Connection head

Process piping
or vessel wall

Fluid
stream

RTD Fluid level


RTDs are often
inserted in process
Applications Why calibrate?
piping or a vessel Mechanical thermometers are The need to achieve consistent
wall. Temperature often used to measure the tem- results with interchangeable
transmitters convert perature in pressurized vessels
the resistance or
The 914X Dry-Well Calibrators automate measurement equipment is one
voltage supplied by such as boilers or pipes. They heating or cooling the sensor. Process
of the most important reasons
electronics supply loop power, measure
the RTD sensor to a may need to be calibrated for resistance, voltage, and 4-20 milliamp sig- why we calibrate.
4-20 milliamp cur- compliance with ISO 9001 or nals, and document the results. Accuracy is an important fea-
rent signal.
ISO 50001 (energy management), ture of a calibrator. You may need
FDA regulation, safety, good man- two items to be calibrated: the a particular level of accuracy
ufacturing practices, or to manage temperature sensor (75 % of the to comply with standards that
energy costs. In some cases, test error) and the transmitter. specify a test accuracy ratio (TAR)
wells are built into a process so Calibrating the sensor requires or test uncertainty ratio (TUR). For
that a reference thermometer can a temperature source such as example many standards require
be placed in line with the ther- a dry-block calibrator or Micro- a 4:1 ratio between the speci-
mometer to be calibrated in situ. Bath. Calibrating the transmitter fied tolerance of the device under
Mechanical thermometers requires a Field Metrology Well, test (DUT), and the accuracy or
(bi-metallic, fluid filled, liquid in multi-function calibrator or docu- uncertainty of the calibration
glass) don’t require multifunction menting process calibrator. equipment.
calibrators. However, they do Some temperature transmit- However, accuracy is also
require a dry-block or Micro-Bath. ters can be adjusted digitally important because when accurate
Also, if flanges, sockets, or bulbs using industry standard digital standards are used, downtime
prevent sufficient immersion in communication protocols such typically only needs to be long
a dry-block calibrator, or if the as HART, Fieldbus, or Profibus. enough to verify that the instru-
stem is too short for sufficient Other temperature transmitters ments are still in tolerance. In
immersion in a dry block calibra- only require a screw driver to contrast, with inaccurate calibra-
tor, use a Micro-Bath instead. adjust a trim pot on face of the tion standards more borderline
Typically four to six inches and transmitter inside the connection and out-of-tolerance indications
snug contact with the insert is head that houses it. are found. As a result, a routine
required for sufficient immersion If the sensor or transmitter verification turns into an addi-
in a dry-block calibrator. is found to be out of tolerance, tional adjustment procedure and a
Sensors and transmitters: In the span and offset have to be final verification at each of the test
many installations, a tempera- adjusted on the transmitter. If points to prove the “as left” condi-
ture transmitter is connected to the transmitter uses a digital tion is in tolerance. This more
the terminal of the RTD, PRT, or communication protocol the span than doubles the down time, and
thermocouple. The transmitter and offset is adjusted through a the technician time involved in
converts the resistance, or volt- digital communicator or with a completing the calibration. This
age supplied by the sensor, to documenting process calibrator is due to the fact that inaccurate
a 4-20 milliamp signal that can capable of digital communica- standards tend to not be consis-
be transmitted to a control panel tion. Otherwise, trim pots on the tent with each other, requiring
or PLC. In this case there are transmitter are used to adjust the more adjustments to correct phan-
span and offset. tom errors.

°F °F
180
Scale
170

160

150 Glass thermometer tube


140
Fluke. The Most Trusted Tools in the World.
130

120 Case
110

100 Fluke Corporation Fluke Europe B.V.


90 PO Box 9090, PO Box 1186, 5602 BD
80 Everett, WA 98206 U.S.A. Eindhoven, The Netherlands
70

60 For more information call:


50
In the U.S.A. (800) 443-5853 or Fax (425) 446-5116
40 Stem
30
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Fluke documenting process calibrators work Web access: http://www.fluke.com
together with Fluke dry-block calibrators to
©2013 Fluke Corporation. Specifications subject to change without notice.
Liquid in glass thermometer stem has a automate and document loop calibrations. Printed in U.S.A. 1/2013 4265025A_EN
bulb.
Modification of this document is not permitted without written permission from
2 Fluke Corporation Process Calibration Tools: Temperature Applications Fluke Corporation.