Sie sind auf Seite 1von 55

Heinen + Löwenstein GmbH & Co.

KG

Most Welcome

Juli 16 1
Location „Bad Ems“

Juli 16 2
Anaesthesia Heinen + Löwenstein

• History
• Product Range
• Safety Concept
• Concept Ventilation Modes
• Concept User Interface
• Hardware / Configurations

Juli 16 3
History
• 1986 Reinhard Löwenstein assumed 50% of the shares in
Heinen GmbH Meckenheim from Mr. Felix F. Marx
whose company had gained already a good name in
neonatologies. Under the new co-operate management, the
company became an exclusive partner to AirShields, Bear
and Healthdyne selling products such as incubators, thermal
beds, child ventilators and SIDS monitors thus remaining a
specialist for neonatologies.

• 1989 Development and manufacture of the first


Heinen + Löwenstein anaesthesia unit, the
Modumat Anaesthesia Workstation.

• 1991 Reinhard Löwenstein assumed the remaining shares


from Mr. Felix F. Marx becoming thus the General
Manager.

• 1992 Development and manufacture of the second Heinen +


Löwenstein anaesthesia unit, the Tizian Anaesthesia
Workstation.

.
Juli 16 4
History
• 1998 Development and manufacture of the third Heinen +
Löwenstein anaesthesia unit, the Sinus MAV Anaesthesia
Workstation. And the Sinus TR as a standard cicuit
system.

• 2002 Extension of the Heinen + Löwenstein development


department for software, electronics and mechanics

• 2005 Completion and manufacture of the Heinen +


Löwenstein anaesthesia unit Leon plus.

• 2006 Extension of the range of self-manufactured products


by the anaesthesia unit Leon.

• 2009 Extension of the „Anaesthesia Family“ Leon plus,


Leon and new … Leon basic version

• 2010 Extension of the „Anaesthesia Family“, new


product … Leon MRI

• 2014 New Dimensions in Anaesthesia, upcoming Project

Juli 16 5
Product Range

1. Leon plus
2. Leon
3. Leon basic
4. Leon MRI
5. Sinus TR
6. New Project

Juli 16 6
Product Range

1. Leon plus
• High Premium Product
• Adult / Pediatrics / Neonates (optional)
• Semi closed / Half closed / Opened
• Minimal Flow / Low Flow / High Flow
• Electronic Mixer / Ratio Function
• Integrated Gas Measurement
• Integrated Aspiration System
• Touchscreen 15“
• O2 consumption tool
• Ecometer

Juli 16 7
Product Range

2. Leon
• Premium Product
• Adult / Pediatrics
• Semi closed / Half closed / Opened
• Minimal Flow / Low Flow / High Flow
• Rotameter / Ratio Function
• Integrated Gas Measurement
• Integrated Aspiration System
• Touchscreen 12“

Juli 16 8
Product Range

3. Leon basic
• Basic Standard Product
• Adult / Pediatrics
• Semi closed / Half closed / Opened
• Minimal Flow / Low Flow / High Flow
• Rotameter / Ratio Function
• Integrated Gas Measurement (optional)
• Aspiration system (optional)
• Touchscreen 12“

Juli 16 9
Product Range

4. Leon MRI
• Premium Product
• MRI compatible
• Adult / Pediatrics / Neonates
• Semi closed / Half closed / Opened
• Minimal Flow / Low Flow / High Flow
• Rotameter / Ratio Function
• Integrated Gas Measurement (optional)
• Integrated Aspiration System
• Touchscreen 12“
• Alarm light
• Integrated Tesla Sensor
Juli 16 10
Product Range

5. Sinus TR
• Standard Product
• Adult / Pediatrics
• Half closed / Opened (optional)
• Low Flow / High Flow
• Rotameter / Ratio Function (optional)
• Aspiration System (optional)
• Transportable System
• Cart Version (optional)

Juli 16 11
Product Range

6. New Project
• A new dimension in Anaesthesia
• Adult / Pediatrics / Neonates
• Closed … ?
• To continue … later on more?

???
Juli 16 12
Product Range

6. New Project ???


A Step beyond the future

Juli 16 13
Safety Concept

Leon plus:
1. Safety Flow
• Emergency Start opportunity
• Backup Electronic Mixer
2. Systemtest
• Automatic Running
• Under 4 minutes
• Gastype Identification
• Compliance compensation
3. Ratio Function
• Guarrantee to 25% in O2 delivery
• Switch over function pneumatically / electronicallly
Juli 16 14
Safety Concept

4. Gas supply
• Emergency bottle backup system
• Drive gas switch over function
• Capability of additional Reserve gas cylinders
• Emergency Air valve
5. Power Supply
• Battery backup up to 110 minutes
• Safety „shut down“ concept
6. User Interface
• „Power off“ - unconsciously impossible
• PSV with backup ventilation
• HLM within CPAP control
Juli 16 15
Concept Ventilation Modes

1. Man. / Spont. (Manuell / Spontaneous Mode)


2. IMV (Intermittend mandatory Ventilation)
3. SIMV (Synchronized intermittend mandatory Ventilation)
4. PCV (Pressure controlled Ventilation)
5. PCV / VTG (Pressure controlled Volume targeted Ventilation)
6. SPCV (Synchronized Pressure controlled Ventilation)
7. PSV (Pressure supported Ventilation)
8. HLM (Heart – Lung – Machine Mode)
9. MON (Monitoring Mode)

Juli 16 16
Concept Ventilation Modes

1. Man. / Spont. (Manuel / Spontaneous Mode)

The Manual / Spontaneous Mode is devided in two functions, the


Manual Function, as the patient is ventilated by using the APL –
valve for adjusting the ventilation pressure and within the
handbag comprimation, the patient is manually ventilated.

The second function is Spontaneous, the patient is able to use


the flow (coming from freshgas settings) to breath
spontaneously on an adjustable CPAP level. Also the Mode can
be used for induction to preoxygenate the patient with mask.

Juli 16 17
Concept Ventilation Modes

1. Man. / Spont. (Manuel / Spontaneous Mode)

Juli 16 18
Concept Ventilation Modes

1. Man. / Spont. (Manuel / Spontaneous Mode)

Juli 16 19
Concept Ventilation Modes

2. IMV (Intermittend mandatory Ventilation)

The IMV Mode is the Volume controlled Mode. The selected VT is


delivered by a constant flow in a defined frequency and time.
The pressure is limited with the function of Pmax, to prevent
from too high pressures.
The inspriation increasing time is driven by the Plateau
adjustement in percentage. The reached plateau is creating the
zero flow phase.
Range: VT (3*) 20 – 1600 ml *Neonatal Option

Frequency 4 – 80 (100*) bpm

Juli 16 20
Concept Ventilation Modes

2. IMV (Intermittend mandatory Ventilation)

Juli 16 21
Concept Ventilation Modes

2. IMV (Intermittend mandatory Ventilation)

Juli 16 22
Concept Ventilation Modes

3. SIMV (Synchronized intermittend mandatory Ventilation)

The SIMV Mode is working as the IMV mode within the


opportunity of a free adjustable Flowtrigger.

The range of the flow trigger can be selected from:


0,1 – 10,0 L/min

Juli 16 23
Concept Ventilation Modes

3. SIMV (Synchronized intermittend mandatory Ventilation)

Juli 16 24
Concept Ventilation Modes

4. PCV (Pressure controlled Ventilation)

The PCV mode is working as a time and pressure controlled


ventilation. Also here the adjustement of the inspiration
increasing time is done with the selection of the plateau in
percentage.

Range: Pinsp. 5 – 60 cm H2O


PEEP 0 – 20 cm H2O
Plateau 10 – 90%

Juli 16 25
Concept Ventilation Modes

4. PCV (Pressure controlled Ventilation)

Juli 16 26
Concept Ventilation Modes

4. PCV (Pressure controlled Ventilation)

Juli 16 27
Concept Ventilation Modes

5. PCV / VTG (Pressure controlled Volume targeted Ventilation)

This Mode is a combination between Pressure controle and


Volume Targeting. This function combines the advantage of
pressure controle as a higher volume on a lower pressure level,
with the advantage of a volume constant ventilation. In
activating this function by setting a tidalvolume, the parameter
Pinsp is switching to the parameter PMax automatically.

Juli 16 28
Concept Ventilation Modes

5. PCV / VTG (Pressure controlled Volume targeted Ventilation)

Juli 16 29
Concept Ventilation Modes

6. SPCV (Synchronized Pressure controlled Ventilation)

The SPCV Mode is working as the PCV mode within the


opportunity of a free adjustable Flowtrigger.

The range of the flow trigger can be selected from:


0,1 – 10,0 L/min

Juli 16 30
Concept Ventilation Modes

7. PSV (Pressure supported Ventilation)

Psv provides pressure support during insufficient


spontaneous breathing.
If the Leon plus is not triggered by the patient after
the adjustable apnoe time (backup), the Leon plus
initiates the inspiration.
In addition, you can use a button to start a manual
breath, not activated by the patient.
Every breath (triggered, backup, manual) is flow
cycled

Juli 16 31
Concept Ventilation Modes

8. HLM (Heart – Lung – Machine Mode)

This Mode is working on a CPAP Level.


It is displaying a direct measurement of CPAP.
It is without „standard“ patient alarms,
Because this mode is for dedicated use on
„Bypass“, as the patient is on Heart lung
machine oxygenated.

The adjustement of CPAP is on APL – Valve


and this CPAP Level is also observated with
an alarm.

Juli 16 32
Concept Ventilation Modes

8. HLM (Heart – Lung – Machine Mode)

Juli 16 33
Concept Ventilation Modes

9. MON (Monitoring Mode)

Juli 16 34
Concept User Interface Batteries charging
Running time on
Mode Status General Patient Category Batteries Date and Time
Warning
Indication

Gas Monitoring
measurement Values

MAC Value,
adjustable
Age

5 functional
Gas supply Tabs:
pressures Curves
Trend Curves
Trend Tab
Electronic Alarm Log
Rotameter Option

Carrier gas
selection
2 pages for
Numerics
O2 Effective
Value

Ventilation Ventilation
Selection of Flow Additional Parameter Mode Stopwatch
O2 mixture Adjustement Parameters Adjustement selection

Juli 16 35
Concept User Interface Selection of Selection of Alarm
Brightness sound level

Results of
last
Systemtest

Ecometer

Juli 16 36
Alarm window, displaying 4
Concept User Interface alarms simultaneously and
scroll function for more Alarms and
than 4 alarms on right
side, actual
readings

Autoset
function

Gas supply
pressures

Juli 16 37
Concept User Interface
Detailed single loop
Zoom function
Freezing function

Three loops
displayed
simultaneously

Juli 16 38
Hardware / Configurations

1. Patient Modul
2. Options panel
3. Gas measurement
4. Vaporizer Receptacle
5. Bronchial aspiration system
6. Writing Surface DIN A3
7. Central foot rest
8. Mobil Cart / Wall Mounting / Pendant
9. Accessories

Juli 16 39
Hardware / Configurations
Flowsensors
1. Patient Modul
• Fresh gas decoupled
• Temperature controlled:
preventing condensation
warming respiratory gas
• APL decoupled during controlled ventilation
• Visual observation of the „Hanging Bellow“
• Optional Infrared Sensor „Freshgas Shortage“
• Inspiratory and expiratory flowsensors
• Absorber can be exchanged during operation
• Can be fully sterilized
• Compliance compensated
• Compatible to our other products:
Leon and Leon (Basic Version)
and also Leon Mri
Detailed view:
Infrared Sensor
„Freshgas shortage“ alarm

Juli 16 40
Hardware / Configurations

1. Patient Modul CO2 Absorber Reusable

CO2 Absorber Disposible

Juli 16 41
Hardware / Configurations O2 Safety Flow

2. Options panel Sidestream


Gasmeasurement

External O2 Outlet

External Freshgas Outlet

Juli 16 42
Hardware / Configurations

3. Gas measurement
• Sidestream Monitor with real time measurement and real time
graphs
• Paramagnetical O² measurement
• Alternativ Configurations available
• Different watertrap for exactly measures in patient category „child“
and „neonates“

Neonatal Watertrap
Oxima: fast inspiratory Blue color coding
and expiratory O2 Different sample rate
analyzing

KE 25: slow
inspiratory O2
analyzing

Juli 16 43
Hardware / Configurations

4. Vaporizer Receptacle
• Selectatec or Dräger compatible solutions
• Integrated one way valve
• Receptacle for two Vaporizers

Juli 16 44
Hardware / Configurations Integrated aspiration system

5. Bronchial aspiration system


• Integrated system
• Attachable system
• Filter protection
• Adapter solutions

Filter protection

Suction bundle
Heinen + Löwenstein

Juli 16 45
Hardware / Configurations

6. Writing Surface DIN A3

Juli 16 46
Hardware / Configurations

7. Central foot rest


Leon MRI standard, Leon plus optional

Juli 16 47
Hardware / Configurations

8. Mobil Cart / Wall Mounting / Pendant

Juli 16 48
Hardware / Configurations

8. Mobil Cart / Wall Mounting / Pendant

Juli 16 49
Hardware / Configurations

9. Accessories
Bag Arm
(PN: 0209111)

Juli 16 50
Hardware / Configurations

9. Accessories

New Adapter (PN: 0209754) for


External Freshgas outlet or
External O2 outlet

Juli 16 51
Hardware / Configurations

9. Accessories

Locking for the backdoors (PN: 0150315)

Juli 16 52
Hardware / Configurations

9. Accessories
Direct VGA Exit (PN: 0209890bg)
New option only available on Leon plus

Juli 16 53
Hardware / Configurations

9. Accessories
TIVA solution (PN: 0140060bg)
„Infusion Rod“

Juli 16 54
Thank you very much, for your open interest …

Juli 16 55

Das könnte Ihnen auch gefallen