Beruflich Dokumente
Kultur Dokumente
IMMOBILIZATION AND
LOCALIZATION
JAMES BALTER, PHD
The steep dose gradients and small margins associated with of external markers to room lasers remains the standard
modern conformal radiation therapy (RT) and intensity- practice for patient positioning in the majority of centers
modulated radiation therapy (IMRT) demand introspec- performing conformal RT and IMRT.
tion into the concepts of patient positioning. This chapter The patient coordinate system may also be referred to
attempts to aid in the quest to understand what is to be by the position and orientation of the regional skeletal
accomplished by the tools and procedures used to position anatomy of the patient visible on radiographic film or on
a patient for treatment and what paradigms for positioning, electronic portal and diagnostic images. Verification of
including variations on methods of image guidance, may be bony landmark position is routinely performed at the ini-
needed. The reader is challenged to consider each anatom- tiation of treatment and subsequently throughout the treat-
ic target uniquely and to consider how a specific patient may ment course on an infrequent basis (typically weekly).
be unique compared with others with the same disease site. Although it is possible that some targets (eg, brain tumors
and upper neck lesions) may maintain a fixed location with
respect to regional skeletal anatomy, targets in the tho-
Immobilization Systems rax, abdomen, and pelvis may vary significantly in posi-
Quite often, a vendor will describe the “reproducibility” of tion relative to the skeleton.
an immobilization system. Herein lies the first opportuni- The third coordinate system is the target coordinate sys-
ty for consideration. It is necessary to both make a target tem. It is reasonable to define the target in this situation as
immobile (ie, not moving during irradiation) and position the clinical target volume. This coordinate system is the
the target correctly. It is a very complex task to achieve both most difficult to establish routinely over the course of treat-
goals simultaneously via a mechanical system. The most ment; however, target location is the dominant concern of
successful applications of this joint paradigm are found in immobilization and localization. Examples of technolo-
the use of stereotactic head frames, which are invasive and gy for target localization include ultrasonography, radi-
rely on minimal organ motion within the skull. ography of implanted fiducial markers, and in-room
computed tomography (CT) scanners.
Coordinate Systems Related to Patient By these definitions, it becomes clear that the primary goal
of treatment verification is to register the target and treatment
Positioning room coordinate systems. Although the patient coordinate
It is useful to consider at least three coordinate systems for system is more conveniently established, its relationship to
patient positioning. The first set of axes defines the treatment the target coordinate system may be highly variable. Persistent
room and is defined by the axes of gantry and collimator rota- use of skin marks or skeletal anatomy to position a patient
tion. This system is known most precisely and is usually main- requires establishment of this variation and incorporation of
tained (through positioning lasers) to an accuracy of 2 mm the residual uncertainty into the treatment planning process
or better with routine physics quality assurance. to ensure sufficient dose to the target.
A second coordinate system is somewhat arbitrary but
has been the standard tool for treatment setup verification
for decades. The patient coordinate system defines the gross Systems for Patient Positioning
position of the anatomic section of the patient to be treat- The variety of systems developed for patient positioning
ed. This system is routinely assessed by marks placed on is quite extensive, and a detailed list is beyond the scope of
the patient’s skin and/or immobilization device. Alignment this chapter. For some details, the reader is referred to an
91
92 / Intensity-Modulated Radiation Therapy
random variation in position for the same immobilization during which images are acquired or on the subsequent
technique may vary significantly between patients in the treatment fraction.
same population. Furthermore, these may vary based on This strategy is certainly acceptable for reduction of
the skill and experience of the therapist performing the gross (eg, over 1 cm) systematic errors. It has no applica-
setup. These variations can be ameliorated somewhat tion in the reduction of random variation (as is described
through extensive training and development of standard further). The potential for this strategy to reduce the mag-
procedures for use of positioning systems. It is a critical, nitude of smaller systematic errors, for which it is gener-
although typically neglected, part of the process of using a ally believed to be acceptable, is discussed further here.
new immobilization system for an institution to establish Figure 6-1 shows the actual data of daily position vari-
its own population statistics of setup variation. In addi- ation of a patient for whom daily imaging was performed.
tion, these measurements should be used for rational deci- The patient had daily position adjustments. The images
sions in localization strategy, as well as margins and acquired following setup (to skin marks) but prior to adjust-
treatment planning directives. The steep dose gradients ment of position were compared with the reference digi-
seen in IMRT plans make such a process crucial to safe and tally reconstructed radiograph to establish position error.
effective treatment. The data shown represent position variation in the cranial-
caudal axis. A simulation is shown of the impact of week-
ly setup adjustment with “perfect” correction. Perfect
Localization Strategies correction is simulated by taking the error seen on the days
Table 6-1 shows examples of population setup variations of imaging (every fifth fraction) and applying that correc-
for different body sites. These data were acquired via fre- tion on all subsequent fractions until the next imaging ses-
quent portal imaging during a planned observation peri- sion. It is clearly conceivable that no significant improvement
od. Although images were acquired daily, only a subset of is made to patient position by weekly adjustment. In fact,
weekly images was used for setup verification. Orthogonal given the selection of the “day,” the potential remains to
portal images were acquired daily, and setup variations were increase setup error over doing nothing at all (in this exam-
measured retrospectively. The mean random variation is ple, both the average and the random variation increased
clearly a poor predictor of the range of random setup errors slightly with ideal weekly adjustment).
for the population.
Given these average and patient-specific variations, it is
important to determine which strategy will most efficiently Pretreatment (On-line) Setup
achieve the precision necessary for IMRT. A few strategies Adjustment
have been studied extensively. The methods described below
One of the most labor- and time-intensive methods to
present a trade-off of effort (and treatment time) expend-
improve setup accuracy involves daily pretreatment local-
ed versus accuracy achieved.
ization of the patient. In fact, this procedure is performed
on almost all patients via the use of external marks and
Weekly Verification lasers, thus registering the patient coordinate system to some
extent with the room coordinate system. As mentioned
This strategy (or small variations thereof) is probably the
above, however, the veracity of surface localization is like-
predominant means of setup verification in use today.
ly insufficient to achieve the accuracy necessary for most
Evolved over the past decades from the advent of portal
films,13 this method involves acquiring portal images to
verify patient position at the initiation of treatment and
then infrequently through the course of treatment. 5
Corrections are generally based on a threshold for action,
and adjustment is made either on the treatment fraction 0
Error (mm)
0 5 10 15 20 25
-5
TABLE 6-1. Average and Maximum Random Variations in a
Population of Patients Setup via skin marks
IMRT treatments. site (possibly as well as user skill) and should be assessed
The advent of electronic portal imaging saw a plethora for new systems.
of studies involving daily targeting of skeletal anatomy, as Given this limit, it is reasonable to consider a slightly
well as implanted fiducial markers. These studies showed different strategy for adjustment in which measurements
a significant reduction in systematic and random varia- are acquired daily but adjustment is performed only if
tions, although with the associated cost of increased treat- the measured position error exceeds a defined threshold.
ment time. Further advances of in-room imaging, including Using this strategy, the threshold can be selected by body
the use of an ultrasound system indexed to isocenter posi- site as a trade-off of the effort expended versus the accu-
tion, as well as in-room CT scanning and registration to racy achieved. An illustration of this trade-off is shown in
reference CT data, have provided increased potential to Figure 6-2. It can be seen that as the residual error approach-
realize the target position more directly than by conven- es the limit of accuracy for measurement and correction,
tional radiographic means. These developments are still dramatic increases in effort yield minimal gain in accura-
quite dynamic, and at the time of writing, it is too early to cy until, finally, no significant gain is realized, even with
state the expected accuracy of using in-room CT. adjustment on every treatment fraction.
Within the daily verification paradigm, there are two Thus, with daily adjustment, it is important to be real-
subsets of strategies. Under one strategy, all measurements istic about expectations of effort and accuracy. One impor-
of target offset are used to correct patient position. Although tant thing to think about is whether the random variation
seemingly the most accurate strategy, this process comes in patient position is so large that it exceeds these accura-
at the highest cost in time and needs some further thought cy limits and is also unacceptably large with respect to treat-
prior to acceptance. ment planning directives. This area is still under exploration.
Consider the steps in the process of daily setup verifi- For conformal therapy of the prostate, Remeijer and col-
cation. A measurement is made of position (via imaging leagues described a margin formula that denotes planning
and some visual or computer-aided alignment technique). target volume (PTV) margin size as
This measurement is then translated into instructions to
adjust the treatment, typically by adjustment of the treat-
PTV margin ≈ 2.5Σ + 0.7σ
ment table position. Both the measurement process and where Σ represents the population standard deviation for
the couch adjustment are not perfect, and their combined systematic error and σ is the random variation standard
error leads to a limit in the final accuracy that can be deviation.14,15 It is important to note that the latest ver-
achieved (even assuming that the patient is completely sion of van Herk’s formula includes a constant 3 mm reduc-
immobile during the time of imaging, adjustment, and tion. This value was based on the equivalent uniform dose
treatment). This means that even if every measured setup formalism, including an estimate of the influence of dose
error is corrected, there is the potential to have a residual homogeneity on tumor control, and is not reproduced
error that is, in fact, introduced by the correction process.
To better understand the magnitude of these residual
errors, it is important to do a self-assessment. Any RT
department that wishes to set margins based on the expect- 1
ed improvement from daily setup correction should first 0.9
determine the accuracy of setup measurement and then
attempt to understand the accuracy of adjustment. Setup 0.8
Frequency of repositioning
here because it requires assumptions that are beyond the and provide an average or predicted correction. Such tools
scope of this chapter. The key point of this finding is that are generally not available in imaging systems or record-and-
systematic error plays a far more important role in dose verify software but are not too difficult to produce (a simple
coverage and margins than random variations and that paper worksheet can suffice).
attempts to reduce the magnitude of systematic varia-
tion achieve most of the benefit possible in positioning for
the majority of patients. Intratreatment Movement
The subject of intratreatment movement is also critical to
IMRT. Various means have been developed to account for
Off-line Setup Correction Strategies movement owing to breathing. The most widespread of
Although a reduction in systematic offset can be achieved these technologies follows the concept of immobilization.
through daily on-line setup measurement and adjustment, Both active breath-hold19–24 and gating25–27 systems are
more efficient means of reducing systematic error have designed to restrict the range of positions of internal anato-
been realized. Given that the systematic error is the aver- my to a fraction of that seen during normal breathing. Note
age difference between the reference and treated patient that these systems do not eliminate the need for margins
position, the combination of patient setup error measure- for movement. Given that the reproducibility of active
ments from multiple days will yield a better estimate of the breathing control is limited and the use of phase or ampli-
average error than any single daily setup measurement. tude windows for gating permits some intratreatment move-
Such measurements do not need to take place completely ment and some variation in position at the gated state,
while the patient is waiting on the treatment table and thus caution needs to be taken to understand the benefit of these
can decrease treatment time. The strategies that involve systems prior to margin reduction. It is strongly recom-
decisions about treatment made while the patient is not on mended that these systems be evaluated locally via fluo-
the table are considered off-line methods. roscopy or similar means prior to clinical use.28–30
There are a number of different off-line strategies of vary-
ing complexity, but the overall issue involves the question
of how many measurements are needed to establish the sys- Summary
tematic error.16–18 Although the results of studies vary, it Our understanding of position variation has evolved sig-
has generally been accepted that three to five measurements nificantly over the past few years. Parallel developments in
are sufficient for estimating the systematic component of localization technology, computer control, immobilization
setup variation with sufficient accuracy. There is some ques- technology, and theory of positioning have led to more ratio-
tion as to whether the first treatment fraction should be nal and efficient means of achieving reasonable margins
considered one of these measurements because many fac- than previously thought possible. Adoption of these tech-
tors, including patient discomfort, the length of the initial nology and concepts, however, requires significant intro-
treatment session, and resolution of potentially complex spection into the RT process and may require extensive
gross errors (eg, improper communication of offsets to the training and infrastructure development. Nonetheless, it
isocenter from the simulated position of the patient), prior is likely that both technology and paradigms are necessary
to initial treatment may come into play. for efficient margin reduction on the order necessary to per-
It is still somewhat difficult to implement off-line strate- ceive a benefit from IMRT in most regions of the body.
gies in most clinics. Although the use of pretreatment local-
ization involves training and use of software that is now
becoming available on most commercial imaging systems, References
and the impact on clinic flow is easily seen, the most efficient 1. Bentel GC. Patient positioning and immobilization in radiation
use of off-line strategies is still open to discussion. Two lead- oncology. New York: McGraw-Hill; 1998.
ing strategies involve either (1) a designated therapist spend- 2. Bentel GC, Marks LB, Sherouse GW, et al. The effectiveness of
ing a fraction of his or her day analyzing all portal images immobilization during prostate irradiation. Int J Radiat
Oncol Biol Phys 1995;31:143–8.
acquired in preparation for off-line measurement (with expe-
3. Sherouse GW, Bourland JD, Reynolds K, et al. Virtual
rience, a single therapist can align all of the images for a 60 simulation in the clinical setting: some pr ac tical
patient/day clinic in less than 2 hours per day for a four- cons ider at ions . Int J R a d iat O nco l Bio l P hys
measurement off-line protocol) or (2) having the therapist 1990;19:1059–65.
at the treatment unit analyze the images for a particular patient 4. Dickens CW. Personalized fixat ion using a vacuum
during or after the delivery of each fraction in the off-line consolidation technique. Br J Radiol 1981;54:257–8.
protocol or analyzing all images immediately before the frac- 5. Jakobsen A, Iversen P, Gadeberg C, et al. A new system for patient
tion on which action is to be taken. Either of these strategies fixation in radiotherapy. Radiother Oncol 1987;8:145–51.
can be implemented efficiently with sufficient planning. It 6. Verhey LJ, Goitein M, McNulty P, et al. Precise positioning of
is very useful to develop a database to record alignment results patients for radiation therapy. Int J Radiat Oncol Biol Phys
96 / Intensity-Modulated Radiation Therapy