SERVICE MANUAL FOR INFANT RADIANT WARMER EDITION/REVISION A/0

ALARM EXPLANATION AND


TYPE
CODE CORRECTIVE ACTION

Main RAM failure alarm H0.1 Replace main MCU

Main ROM failure alarm H0.2 Replace main MCU

1. Check the connection of circuit between


Double MCU communication failure alarm H0.3 the main MUC and auxiliary MCU
2. Replace main MUC and auxiliary MCU

E2PROM main MCU failure alarm H0.4 Replace main MCU

RAM failure alarm inside auxiliary MCU H0.5 Replace auxiliary MCU

ROM failure alarm inside auxiliary MCU H0.6 Replace auxiliary MCU

Main ADC failure alarm H0.7 Refer to Flow chart 5.1

Auxiliary ADC failure alarm H0.8 Refer to Flow chart 5.2

SRAM failure alarm H0.9 Replace U7 on mian board

Real timer RTC failure alarm H1.0 Replace U21 on mian board

Heat circuit failure alarm H1.1 Refer to Flow chart 5.3

E0.1 Refer to Flow chart 5.4

Skin temperature sensor failure alarm


E0.2 Refoer to Flow chart 5.5

Deviation failure alarm Skin


E0.3 Refer to Flow chart 5.6
temperature sensor

Over-temp alarm E0.4 Refer to Flow chart 5.7

Upper deviation E0.5 1. Check the position of sensor


2. Check the ambient temperature
Deviation alarm
3. Make the device far away the heat
Lower deviation E0.6 resource

Please set the temperature or change the


E0.7
cotrol mode correctly
Set alarm
Please check the condition of probe and the
E0.8
patient’s skin

5-8