The dynamic scale loop is a tube blocking system for the evaluation of scale
inhibitors. The basic principal involves mixing two incompatible brines in a narrow
bore test coil. Scale begins to form in the coil creating a differential pressure. The
time taken for this formation gives an indication of inhibitor performance.
The control software enables the user to view all data in an easy to understand
manner. The graphs display historical pressure and temperature information,
whereas the schematic shows all current data in addition to the pumps status.
Model L1 H200
Max Pressure 1 Bar (14.5psig) 200 Bar (2900psig)
Pre-heating Unit S N
Oven O S
Cooling Bath O O
Max Temperature 90°C 200°C
Automatic Testing S S
High Pressure N O
Pump Type Peristaltic HPLC
Exchangeable Test Coil S S
High Accuracy Differential Transducer S S
Piston Flushing N S
Pressures Displayed & Logged S S
Temperature Displayed & Logged S S
Safety Pressure Relief N S
System Pressure Control S S
Untreated
Brine
Scale
Inhibitor
MIC of 2ppm
1
0.00 20.00 40.00 60.00 80.00 100.00
Test time (Minutes)
Typical result graph
The above graph show typical results. The untreated brine forms scale and starts to
block the test coil after about 12 minutes (Blank). The test is then repeated with
scale inhibitor present, firstly at 3ppm, automatically reducing to 2ppm after 45
minutes and finally reducing to 1ppm after 90 minutes. At about 110 minutes the test
coil starts to block again indicating insufficient inhibitor. At this point the unit
automatically cleans itself.
Test Coil