Dräger Perseus® A500 Breathing System

Disassembly/Assembly

D-41942-2012
Disassembly
91 02 842 | 16.09-4 | HQ | HO | Subject to modifications | © 2016 Drägerwerk AG & Co. KGaA

Insert the cleaning disk into the Loosen and pull out the
Unscrew the CLIC adapter. adapter. Close the adapter so expiration port.
that it engages.

Remove the cleaning disk from the


adapter. Close the adapter so that Push in the expiratory port and
Screw the CLIC adapter. tighten it with the knurled nut.
it engages.

Loosen and pull out the Turn the three levers


inspiratory port. Remove the flow sensors. counterclockwise about 120°.

Push in the inspiratory port and Insert the flow sensors. Turn the three levers clockwise
tighten it with the knurled nut. about 120°.

Lift off the upper part of the Screw off the APL valve. Remove the valves
breathing system. (red and blue).

Place the upper part of the Align the APL valve correctly and Align the valves correctly.
breathing system on the lower part. tighten it with the knurled nut. Insert the valves.

Assembly
Dräger Perseus® A500 Breathing System
Reprocessing Instructions

D-41942-2012
Manual cleaning Machine cleaning Sterilization
and disinfection and disinfection

Position components so that all interior spaces and surfaces are completely flushed and water can drain off freely.

Visual check
The following parts must be free of deposits:
* – Incident flow grills
– Non-return valves (red and blue)

If required, remove deposits on the valve


plates of the non-return valves with a soft
cloth. If the incident flow grills are damaged,
contact DrägerService®.

* For machine cleaning, position each valve


so that the valve cage is at the bottom.

This guideline does not replace the instruction for use.