Sie sind auf Seite 1von 9

A Preclinical Study of an Implanted Device

in the Pulmonary Veins, Intended for the Treatment of


Atrial Fibrillation in an Ovine Model
TIM VANDECASTEELE, D.V.M.,* TIM BOUSSY, M.D.,† MATTHEW PHILPOTT, M.SC. ENG.,‡
ELI CLEMENT, M.SC. ENG.,‡ STIJN SCHAUVLIEGE, D.V.M., PH.D.,§ WIM VAN DEN
BROECK, D.V.M., PH.D.,* GUNTHER VAN LOON, D.V.M., PH.D.,¶
PIETER CORNILLIE, D.V.M., PH.D.,* and GLENN VAN LANGENHOVE, M.D., PH.D.‡
From the *Department of Morphology, Faculty of Veterinary Medicine, Ghent University, Merelbeke, Belgium; †AZ
Groeninge Medical Hospital, Department of Cardiology, Kortrijk, Belgium; ‡Fulgur Medical, Merelbeke, Belgium;
§Department of Surgery and Anaesthesia of Domestic Animals, Faculty of Veterinary Medicine, Ghent University,
Merelbeke, Belgium; and ¶Department of Large Animal Internal Medicine, Faculty of Veterinary Medicine, Ghent
University, Merelbeke, Belgium

Background: Atrial fibrillation is the most frequent arrhythmia in adults of which the interventional
cure is hampered by high recurrence rates. Recurrence after ablation is due to an incomplete isolation of
the pulmonary veins. A new ablation technique was performed, in the antra of ovine pulmonary veins, by
device implantation, which was heated through a wireless heat-generating system.
Methods and Results: Implants were placed transatrially in the pulmonary veins of sheep. Using
a wireless heating system, the energy was afterward transferred through wires to the implanted device
according to a defined protocol. The position of the implant and the applied lesions were macroscopically
evaluated. Samples of the ablated tissue of the atrio-pulmonary vein junction were histologically and
immunohistochemically examined.
Conclusions: Six ablation procedures in four sheep were successfully performed without adverse
cardiac reactions. Implantation of the device and the wireless heat generation was feasible. Sufficient
heat was produced at the level of the antra of the pulmonary veins to create ablation lesions, which were
histologically and immunohistochemically confirmed. (PACE 2016; 39:822–829)

electromagnetic heating, sheep, implant, wireless, ablation

Introduction Nevertheless, ablation of the pulmonary veins is,


Atrial fibrillation (AF) is the most common still to date, a technically demanding technique
cardiac arrhythmia in adults associated with and recurrence of AF is seen due to incomplete
substantial mortality and morbidity. It is generally isolation of the pulmonary veins. During long-
accepted that the main triggers of abnormal term follow-up, a recurrence rate of more than
pulses, which induce AF, are localized at the 50% was mentioned after a single procedure
level of the pulmonary veins base.1–3 In the past and without the use of antiarrhythmic drugs.9
decade, several catheter-based ablation techniques Consequently, the need for a better treatment
were tested and improved to isolate electrically strategy which ensures a high success rate along
the pulmonary veins from the left atrium.4–8 with a relatively quick and simple procedure
remains, ideally offering the possibility to repeat
the procedure noninvasively.
Conflict of Interest: Glenn Van Langenhove, Matthew Philpott, In this context, an implantable device was
and Eli Clement are employed by Fulgur Medical, responsible developed, after positioning in an electromagnetic
for the development of the implantable device and heat-
generating system. All authors have read the journal’s field, which can be wirelessly heated, creating the
authorship agreement, and the manuscript has been reviewed desired transmural ablation lesions at the level
and approved by all the named authors. of the pulmonary vein-left atrial junction. In the
Address for reprints: Tim Vandecasteele, D.V.M., Department first step, an in situ study examines wireless heat
of Morphology, Faculty of Veterinary Medicine, Ghent generation after which the heat is transferred to
University, Salisburylaan 133, 9820 Merelbeke, Belgium. the implanted ablation device. This implant has a
Fax: 3292647790; e-mail: Tim.Vandecasteele@UGent.be tubular shape through which a complete circular
Received March 17, 2016; revised May 10, 2016; accepted May lesion can be applied. This procedure was per-
23, 2016. formed in an ovine model by open heart surgery
doi: 10.1111/pace.12899 due to access of the pulmonary veins. The aim of

©2016 Wiley Periodicals, Inc.


822 August 2016 PACE, Vol. 39
IMPLANTATION IN THE PULMONARY VEINS

Procedure
The procedure was performed at IMM
Recherche (Paris, France) under the Ethics com-
mittee approval number “RYTHMOLOGIE-12-12-
FLUX MEDICAL 14–34.” Studies were performed
on four sheep (1–1.5 years old).
Using 6 mg/kg propofol (Fresofol, Pharmatel
Fresenius Kabi, Hornsby, Australia) induction
and 2% isoflurane (Abbott Australasia, Kurnell,
Australia) maintenance, sheep were anesthetized
and placed in dorsal recumbency. Arterial and
venous cannulae were inserted into the carotid
artery and jugular vein using a cut-down technique
for pressure monitoring. The chest was opened
using conventional techniques. A 40/36-F two-
staged venous cannula (Edwards Life Sciences,
Figure 1. Left image = Verhaert implant device with
Irvine, CA, USA) was inserted into the right
heater coil (1); right image = pick-up coil (2) and
atrium and secured with a pursestring suture.
electromagnetic or applicator coil (3).
A 22-F aortic cannula (Edwards Life Sciences)
was inserted into the proximal part of the
descending aorta. The extracorporeal circulation
system consisted of a venous reservoir, roller
this study was to evaluate this technique, directly pump (Cobe Cardiovascular, Arvada, CO, USA),
after the ablation, by macroscopic, histological, and a Capiox RX 25 membrane oxygenator
and immunohistochemical investigation of the (Terumo Europe, Leuven, Belgium) connected by
induced lesions in sheep. noncoated tubing. Cardiopulmonary bypass was
established by priming with 1,000 mL crystal-
Materials and Methods loid prime solution (lactated Ringer’s solution
Device 750 mL, 20% mannitol 100 mL, aprotinin 100 mL,
The implants (Fig. 1) contain a heating coil 8.4% sodium bicarbonate 50 mL, and heparin
(Verhaert, Kruibeke, Belgium), which includes a 5,000 IU), which was allowed to circulate in
temperature feedback system. Implant size ranged the extracorporeal circulation system for 2 hours.
from 12 mm to 20 mm. An electromagnetic coil Flow was maintained at >3 L/min with a perfusion
generating an electromagnetic field, in which a pressure of 50–70 mm Hg. After opening of the left
pick-up coil was placed according the same plane, atrium, suction cannulas were placed inside the
constitutes the heat-generating system. The heat pulmonary veins. Afterward, the most suitable-
generated in the pick-up coil was transferred sized implant, according to the diameter of the
through wires to the heating coil. pulmonary vein, was inserted into the antrum,

Figure 2. Ablated atrio-pulmonary vein junction tissue samples of two sheep. AS = atrial side of
the sample; PS = pulmonary side of the sample; the oval indicates the ablation region; red line =
trimming line indicating the orientation of the histological sections.

PACE, Vol. 39 August 2016 823


VANDECASTEELE, ET AL.

Figure 3. Time/temperature table of the ablation procedure of the pulmonary veins during 250 seconds (panels A, B,
and G), 400 seconds (panel E), 500 seconds (panel F), 600 seconds (panels C and D), 800 seconds (panel H). Tstent and
Tenviro indicate implant temperature and body temperature, respectively, I refers to the used current, Tset indicates
the set temperature.

which is the common venous space, of both ostia The electromagnetic field, created by the
through which the pulmonary veins drain into applicator coil, generates an alternating current
the left atrium (for terminology see Vandecasteele in the pick-up coil, when placed in the same
et al.10 ). Subsequently, the implanted device was plane of the applicator coil, which is consequently
connected through wires to a pick-up coil to transferred into heat. This heat causes the ablation
ensure the heat transfer. Also, a temperature process of the connected implant’s heater coil.
control device was put in place. Before the ablation procedure was started, the

824 August 2016 PACE, Vol. 39


PACE, Vol. 39

Table I.
Description of Temperature Data, Visualized in the Graphs of Figure 3, of the Devices Implanted in Different Sheep According to the Set Temperature

Implants Sheep 1 Sheep 2 Sheep 3 Sheep 4

Implant 1 Temperature Failure: as a current increase To a higher temperature Gradually but faster Failure: temperature
increase around 100 seconds compared to the test in surpassed the 60°C limit
induced no temperature sheep 1 (demonstrated by a large

IMPLANTATION IN THE PULMONARY VEINS


increase (Fig. 3A) temperature peak),
associated with the
production of a small
amount of smoke after
which the test was stopped.
It was found that both the
temperature and heater
August 2016

power wire were damaged


(Fig. 3G)
Set temperature 50°C 55°C for 210 seconds
Implant Followed nicely the set Reacted on the temperature
temperature temperature causing the increase (only small
implant to ablate at 50°C for variations were seen;
180 seconds (Fig. 3C) Fig. 3E)
Implant 2 Temperature Gradually Of which only the peak Was heated similarly to the Temperature raised gradually
increase temperatures reached previous test to two different set
briefly set temperature temperatures
Set temperature Increased from 43.5°C–44°C Was increased to 55°C for Was increased to 55°C for First 52°C for 120 seconds
for 180 seconds 230 seconds 300 seconds and thereafter of 53°C for
480 seconds
Implant Followed set temperature Was stable at an average Followed set temperature At each set temperature, the
temperature (Fig. 3B) temperature of 51°C almost perfectly with only implant temperature was
(Fig. 3D) some small variations fairly constant with
(Fig. 3F) approximately 2°C variation
compared to the set
temperature (Fig. 3H)
825
826

Table II.
Additional Information about the Procedures Mentioned in Figure 3 and Table I

Implants Sheep 1 Sheep 2 Sheep 3 Sheep 4

Implant 1 Failure was due to malposition of Was heated to 50°C as this The required current was relatively Was stopped due to alleged
the temperature sensor which provides the required ablation low compared to previous tests as damage of the temperature
was not in contact with the vessel lesion according to previous tests. the peaks surpassed 80 A and sensor or the heater coil during
wall (Fig. 3A) At maximum set temperature, the the average levels were around the implantation (Fig. 3G)
required current peaked at almost 70 A. The peak at the end was
120 A and averaged around due to system reset by which the

VANDECASTEELE, ET AL.
100 A (Fig. 3C) temperature dropped and
consequently the software
August 2016

increased the current output


accordingly (Fig. 3E)
Implant 2 The required current peaked Reached set temperature at certain Demonstrated a required power The power peaked just under 120 A
(almost 120 A) at a maximum set moments which is due to peak of just below 120 A and an and averaged around 60 A
temperature and was averaged insufficient contact between the average of around 80 A (Fig. 3F) (Fig. 3H)
around 60 A (Fig. 3B) temperature sensor and the
vessel wall or between the
heating coil and the vessel wall in
the vicinity of the temperature
sensor. Due to this uncertainty,
the ablation time was reduced to
230 seconds instead of the
planned 300 seconds. The
required power peaked and
averaged just below 120 A and
was increased gradually in the
last half of the test to reach the
set temperature (Fig. 3D)
PACE, Vol. 39
IMPLANTATION IN THE PULMONARY VEINS

Figure 4. Parts of pulmonary vein tissue at the level of the atrio-pulmonary vein junction. Left
image: at the top side the white colored ablation site indicated by the arrows and at the bottom
the normal pinkish-colored tissue. Right image: the removed device with the heater coil (green
wire) of which an ablation mark (indicated by the frame) is seen on a piece of the pulmonary
vein wall.

Figure 5. Histological section of the ablated atrio-pulmonary vein junction (H&E staining). Left
image: AS = atrial side; 1 = myocardium; 2 = fibrin deposits; 3 = interruption of the tissue.
Scale bar is 250 µm. Right image: PS = pulmonary side; 2 = fibrin deposits; 4 = lung tissue; 5 =
hemorrhagic necrosis; arrow = pulmonary vein wall; Scale bar is 1 mm.

temperature sensor was checked to ensure that the histological sections. The orientation in which
indicated temperature correlated with the body the histological sections were cut is indicated in
temperature. Figure 2.
At the start of the heating procedure, the
temperature control software was set to 1°C Immunohistochemical Analysis
above the body temperature after which the The myocardial sleeve of the ablated pul-
temperature was gradually increased from the monary vein was stained immunohistochemically
initial temperature in 1°C steps every few seconds with the myosin marker MYBPC3 (polyclonal,
while trying to reach the required ablation K-16: sc-50115, Santa Cruz Biotechnology, Santa
temperature of 58°C during different time periods. Cruz, CA, USA) to detect the myosin-binding
This procedure was repeated for each ostium, one protein C (cardiac type), which is reactive
implant at a time. with ovine tissue. Myosin-binding protein C is
present in the myofibrils of cardiac tissue.11
Pathological Analysis
The 5-µm-thick slides were immunostained using
All sheep were euthanized directly after the the Dako automated Autostainer Plus (Dako,
ablation procedure. Immediately after euthanasia, Glostrup, Denmark). No antigen retrieval was
the heart-lung packages were excised after which performed. The sections were incubated for
the pulmonary veins, containing the ablation 5 minutes with 3% hydrogen peroxide and 30
device, were removed and examined macroscopi- minutes with rabbit serum. First, the primary
cally. antibody (1:100) was incubated for 60 minutes,
afterward secondary antibody (rabbit/anti-goat,
Histological Analysis biotinylated, polyclonal, 1:500, Dako) was applied
Four percent formaldehyde-fixed paraffin- for 30 minutes followed by streptavidin and
embedded scarred tissue of the atrio-pulmonary horseradish peroxidase (streptavidin-HRP, 1:500,
vein junction of two sheep was examined on Dako) for 30 minutes. The visualization was
5-µm-thick hematoxylin and eosin (H&E)-stained achieved with DAB (Dako) for 5 minutes.

PACE, Vol. 39 August 2016 827


VANDECASTEELE, ET AL.

Figure 6. Immunohistochemical staining with myosin marker MYBPC3 of myocardial sleeve tissue inside pulmonary
vein wall. The double arrow indicates the stent implantation site. AS = atrial side; PS = pulmonary side. Scale bar:
top left and top right image is 200 µm; top middle image is 100 µm; bottom image is 2 mm.

Results Discussion
Ablation Data This study was performed in sheep in an open
The data of the ablation procedures performed heart surgery procedure in order to have a good
with eight devices in four sheep (two devices per accessibility to the pulmonary veins. In numerous
animal) are indicated in Figure 3 and Tables I and cardiovascular studies, pigs are used as an animal
II. Not one of the following tests was aborted due model in which the localization of the draining
to any cardiovascular disorder or abnormalities in openings or ostia of the pulmonary veins into
rhythmicity. the left atrium was studied in detail.10 However,
anatomically, the number of ostia in pigs shows
Pathological Analysis similarity with sheep or goats as at the level of the
pulmonary vein-left atrial junction two vestibules
Macroscopic Evaluation
of the left atrium were described in a clinical study
Observation of the heart-lung packages at the in goats.12
level of the pulmonary veins revealed no damage
to the surrounding structures. The endothelium Ablation Data
of all pulmonary veins showed clear ablation
lesions (Fig. 4) along the circumference of the We demonstrated that ablation of the pul-
device, although by comparing the samples monary veins through wireless energy transfer
some differences in degree of discoloration were is feasible. Eight stents were implanted and
noticed. At the level of the device, no thrombus tested in four living sheep of which two ablation
formation was noticed in any of the sheep. procedures failed. One failure occurred during the
initialization phase and one during the ablation
phase itself. Cause of this failure was damage to
Microscopic Evaluation the temperature sensor during implantation which
The ablation lesion, investigated on two gave wrong temperature data. Thus during system
tissue samples, was recognized on the histolog- check of the implants, six of eight implants met
ical sections by fibrin deposition, visualized in the technical requirements.
Figure 5. Figure 6 demonstrates immunohisto- The difference in extensiveness of the ablation
chemically the myosin staining of myocardial lesions was due to the fact that no clear tempera-
sleeve tissue of the ablated pulmonary vein, ture protocol was predefined, and that different
including the ablation site. At this latter spot, temperatures and different ablation times were
a lighter immunohistochemical reaction can be applied. Histologically, on H&E-stained sections,
observed in comparison to the tissue proximal and the effects of heating at the level of the pulmonary
distal to the site of the implanted stent. veins could only be noticed by visualization

828 August 2016 PACE, Vol. 39


IMPLANTATION IN THE PULMONARY VEINS

of fibrin deposits in acute euthanized animals. through remote energy transfer. All sheep survived
Immunohistochemically, denaturation of myosin the implantation and heating of the ablation
proteins of the myocardial sleeve at the level of devices without any rhythm disturbances. Macro-
the heated implanted stent proves the induction scopically, in all successfully treated animals
of a lesion in the pulmonary veins wall caused by ablation lesions were noticed in the pulmonary
the generated heat at the ablation site. After long- veins and histological and immunohistochemical
term studies, greater effects may be seen as fibrin confirmation was provided of the applied lesion.
completely replaces the necrotic tissue. Immunohistochemical staining was used to indi-
The degree of contact between the implant cate the disappearance of myosin fibers during the
and the vessel wall is also of great importance ablation process as this is difficult to visualize
to create a perfect circular lesion. However, on H&E-stained tissue. Myosin destruction was
overstretching of the pulmonary vein wall is examined as Tornberg13 indicated that during
possible but within its safe range in terms of a heating procedure, this feature is the first
elasticity. In this study, the implant diameter transition occurring in muscle tissue.14,15
ranged from 12 mm to 20 mm. The information The macroscopical difference between nor-
deduced from these procedures will further mal and ablated tissue was clearly demarcated.
improve stent design intended for future studies. Heat production at the level of the pulmonary
The proof of the efficacy of this system, veins was proven histologically by the presence
by which the heat is picked up wirelessly of fibrin deposits. No thrombus formation was
and transferred from the pick-up coil to the seen on the device and no damage to the
pulmonary veins through a direct connection, surrounding tissue was noted. Further research
will allow for development of an implantable is needed to clear out which temperature and
device that can be heated by placing the animal ablation time result in an optimal ablation.
in an electromagnetic field. The fact that the Moreover, self-expanding devices would provide
implanted device can remain in place and may an additional advantage in terms of implant-
be endothelialized constitutes a major advantage pulmonary vein contact. Future trials have to
as a second treatment can be performed without confirm the feasibility of the procedure through
any surgery. a transfemoral approach, the application of a
transmural lesion, the long-term persistence of
Conclusions pulmonary vein isolation, the endothelialization
In this ablation study in a sheep model, all of the implant, the absence of excessive neointima
devices could be placed inside in one of both formation and the feasibility of noninvasively
antra of the pulmonary veins and were heated repeating the procedure.

References
1. Jais P, Haissaguerre M, Shah DC, Chouairi S, Gencel L, fibrillation therapy in the pig. J Interv Card Electrophysiol 2008;
Hocini M, Clementy J. A focal source of atrial fibrillation 21:187–193.
treated by discrete radiofrequency ablation. Circulation 1997; 95: 8. Gerstenfeld EP, Michele J. Pulmonary vein isolation using a
572–576. compliant endoscopic laser balloon ablation system in a swine
2. Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, model. J Interv Card Electrophysiol 2010; 29:1–9.
Garrigue S, et al. Spontaneous initiation of atrial fibrillation by 9. Medi C, Sparks PB, Morton JB, Kistler PM, Halloran K, Rosso R,
ectopic beats originating in the pulmonary veins. N Engl J Med Vohra JK, et al. Pulmonary vein antral isolation for paroxysmal
1998; 339:659–666. atrial fibrillation: Results from long-term follow-up. J Cardiovasc
3. Chen SA, Hsieh MH, Tai CT, Tsai CF, Prakash VS, Yu WC, Electrophysiol 2011; 22:137–141.
Hsu TL, et al. Initiation of atrial fibrillation by ectopic beats originat- 10. Vandecasteele T, Vandevelde K, Doom M, Van Mulken E, Simoens
ing from the pulmonary veins: Electrophysiological characteristics, P, Cornillie P. The pulmonary veins of the pig as an anatomical
pharmacological responses, and effects of radiofrequency ablation. model for the development of a new treatment for atrial fibrillation.
Circulation 1999; 100:1879–1886. Anat Histol Embryol 2015; 44:1–12.
4. Haissaguerre M, Shah DC, Jais P, Hocini M, Yamane T, 11. Winegrad S. Cardiac myosin binding protein C. Circ Res 1999;
Deisenhofer I, Chauvin M, et al. Electrophysiological breakthroughs 84:1117–1126.
from the left atrium to the pulmonary veins. Circulation 2000; 12. Reddy VY, Houghtaling C, Fallon J, Fischer G, Farr N, Clarke J,
102:2463–2465. Mcintyre J, et al. Use of a diode laser balloon ablation catheter
5. Pappone C, Rosanio S, Oreto G, Tocchi M, Gugliotta F, Vicedomini to generate circumferential pulmonary venous lesions in an open-
G, Salvati A, et al. Circumferential radiofrequency ablation of thoracotomy caprine model. Pacing Clin Electrophysiol 2004;
pulmonary vein ostia: A new anatomic approach for curing atrial 27:52–57.
fibrillation. Circulation 2000; 102:2619–2628. 13. Tornberg E. Effects of heat on meat proteins—Implications on
6. Shah DC, Haissaguerre M, Jais P, Hocini M, Yamane T, Deisenhofer structure and quality of meat products. Meat Sci 2005; 70:493–508.
I, Garrigue S, et al. Curative catheter ablation of paroxysmal 14. Martens H, Vold E. DSC studies of muscle protein denaturation.
atrial fibrillation in 200 patients: Strategy for presentations ranging In: Proceedings of the 22nd European Meeting of Meat Research
from sustained atrial fibrillation to no arrhythmias. Pacing Clin Workers, 1976, Malmö, Sweden (p. J 9.3).
Electrophysiol 2001; 24:1541–1558. 15. Wright DJ, Leach IB, Wilding P. Differential scanning calorimetric
7. Jones DL, Guiraudon GM, Skanes AC, Guiraudon CM. Anatomical studies of muscle and its constituents. J Sci Food Agric 1977;
pitfalls during encircling cryoablation of the left atrium for atrial 28:557–564.

PACE, Vol. 39 August 2016 829

Das könnte Ihnen auch gefallen