RENO PLASMA STERILIZER

1
Trouble Shooting
Error 1

◆ The reason of Error 1 Occurrence

This error occurs when the vacuum value doesn’t decrease by proper value during setting time.

Vacuum & Leak test

No Leak
Leak
Vacuum & leak are normal
3. Check the area between
2. Check parts that is
1. Check sterilization items vacuum valve and pump &
connected to chamber
pressure gauge

www.renosem.com 1
RENO PLASMA STERILIZER

Vacuum & Leak test

Push the Maintenance icon 1) Push the “****” box.


2) Enter the Password “0501”, and then push
the “Enter “ icon.

www.renosem.com 2
RENO PLASMA STERILIZER

Vacuum & Leak test

1) Push the Vacuum pump 1) Close the Vacuum valve. 1) Push the Vent valve.
& Vacuum valve icons
to make a vacuum state. 2) The range of vacuum value
must be less than 200mT/1min
2) After 5 minutes, check the (if not, need to find a leak point)
vacuum value.
( normal value is less than 400mT)

www.renosem.com 3
RENO PLASMA STERILIZER

1 Trouble Shooting
Error 1

1. Check sterilization Items

Prohibited material

▪ Not completely dried Items ▪ Powder


▪ Possibly absorptive materials ▪ Oil
(Cellulosic material) ▪ Water/Wadding
▪ Wet items ▪ Cardboard
▪ Wood ▪ Organic Materials
▪ Paper ▪ Low Quality Plastic & Latex
▪ Foam ▪ Disposable suture materials
▪ Sponge, Gauze ▪ Rigid(S.S) Lumen
▪ Linen : Follow the lumen guide line
▪ Cotton ▪ Flexible (Teflon) Lumen
▪ Cloth/Towel : Follow the lumen guide line
▪ Huck towels
▪ Muslin
▪ Silk

www.renosem.com 4
RENO PLASMA STERILIZER

1 Trouble Shooting
Error 1

2. Check the parts that is connected to chamber

2-1. Check the Injection Valve 2-2. Check the SFS Nozzle

2-4. Check connecting part of 2-3. Check the parts that is connected
pressure gauge to vent valve

www.renosem.com 5
RENO PLASMA STERILIZER

Injection Valve Leak test

Prepare vacant sterilant Remove the sticker Push the red rubbers like a picture

Vent Make a vacuum for 4minutes Insert a test sterilant Assemble the sterilant case

If injection valve has a leakage, red rubber is inserted to down side like picture.

www.renosem.com 6
RENO PLASMA STERILIZER

2-1. How to change Injection Valve

Snap-ring

union

Supporting bolt

1. Remove the snap-ring.


2. Take out the nametag & solenoid
3. Remove the both unions & supporting bolts.

www.renosem.com 7
RENO PLASMA STERILIZER

2-1. How to change Injection Valve

Injection valve supporting bolt

4. Remove the both injection valve supporting bolts.


5. Change the injection valve and the re-assemble by reverse order.

www.renosem.com 8
RENO PLASMA STERILIZER

2-2. SFS Nozzle check

1) Go to maint.4 and then Push the “SV” button to decrease temperature of injection heater.
2) Change the setting value to 10degree and then push the ‘enter, button.
3) Push the ‘change’ button to confirm the setting value.
4) Check the temperature of injection heater.
If the temperature fall down less than 40 degree, try to check the SFS Nozzle.

www.renosem.com 9
RENO PLASMA STERILIZER

2-2. SFS Nozzle check

Blank Flange

1) After unfasten the Clamp, disassemble the SFS Nozzle.

2) Block the hole using blank, and then try to vacuum & leak test

3) If vacuum value is normal & no leak, need to check SFS Nozzle for finding a leakage.
(If not, you need to check others parts)

www.renosem.com 10
RENO PLASMA STERILIZER

2-2. SFS Nozzle check

Clamp High temp. centering Heat resisting O-ring

1) Unfasten the Clamp.


2) After checking High temp. centering, replace
High temp. centering if it is not good condition.
3) After removing heater supporting bolt,
disassemble heater & temp.sensor to arrow mark
direction.
4) After removing 4 wrench bolts, disassemble
injection port & SFS Nozzles body.
5) After checking heat resisting O-ring, replace it
if it is not good.

www.renosem.com 11
RENO PLASMA STERILIZER

2-3. Check the connecting part of Vent valve

1) After unfasten the clamp, disassemble the TEE pipe.


2) Block the hole using blank, and the try to do
Vacuum & leak test
3) If vacuum value is normal & no leakage, need to replace
3 parts that is connected to clamp or TEE pipe part.

Caution
In this case you can’t vent at maint mode because vent valve
TEE pipe Clamp is not connected to chamber, so need to vent slowly with
hand

www.renosem.com 12
RENO PLASMA STERILIZER

2-4. Check the connecting part of pressure gauge

1) Unfasten the clamp that is connected to pressure gauge, and then disassemble the Elbow.
2) Connect the pressure gauge directly to the hole
3) Try to do Vacuum & leak test.
4) If vacuum value is normal & no leak, check the clamp & elbow and replace it if it is not
good condition.

13
www.renosem.com 13
RENO PLASMA STERILIZER

1 Trouble Shooting
Error 1

3. Check all parts between vacuum valve and vacuum pump


& pressure gauge

3-1. Check the pressure gauge 3-2. Check the vacuum valve

3-3. Check the Elbow


3-4. Check the bellows
& Corona Plasma

3-5. Check the Vacuum Pump

www.renosem.com 14
RENO PLASMA STERILIZER

3-1. Check the pressure gauge

◆ In case of using Calibration kit

A
A
B

B
“U” pipe

1) After disassemble the used pressure gauge, connect “U type pipe” like picture.
2) Connect reference gauge to A line & used gauge to B line.
3) Make a vacuum for 5 minutes.
4) After press the DSP button , check the monitor → “A” is reference gauge value, “B” is used gauge value,
“C” is “ A-B”
5) If “A-B” value is less than 200mT, it is normal
1) If “A-B” value is over than 200mT, adjust “C” value to about “0” by adjusting zero point of used gauge.
www.renosem.com 15
RENO PLASMA STERILIZER

3-1. Check the pressure gauge

◆ In case of doesn’t have Calibration kit

1) Use a new pressure gauge as reference gauge.

2) After make a vacuum for 5 minutes with used gauge and then make a note the vacuum value
not to forget it.

3) After disassemble the pressure gauge, assemble the new pressure gauge in there, and then make a
Vacuum for 5minutes and also make a note it.

4) Re-assemble the used gauge and then make a vacuum.


After vacuum, change present vacuum value to reference value that made a noted by adjusting
zero point.

www.renosem.com 16
RENO PLASMA STERILIZER

3-2. Check the Vacuum Valve

◆ If Vacuum Valve is not opened, Error 1 can be occurred.

1) Go to Maint mode, and then push the vacuum valve.


And check the solenoid working sound

2) If you cannot hear solenoid sound, check the input voltage(+24v)


of Vacuum Valve.

www.renosem.com 17
RENO PLASMA STERILIZER

3-3. Check the Elbow & Corona Plasma

◆ Elbow

elbow

⚫ Check the clamp & centering of both side of Elbow

⚫ After changing Elbow and check the leakage.

www.renosem.com 18
RENO PLASMA STERILIZER

3-3. Check the Elbow & Corona Plasma


◆ Corona Plasma
Rev.3

Corona Plasma leak test


⚫ After removing Corona
Plasma , connect vacuum
valve directly to vacuum pump
using bellows and the check
the leakage

www.renosem.com 19
RENO PLASMA STERILIZER

3-4. Check the Bellows

Bellows

⚫ Change the Bellows and then try to do leakage test.

www.renosem.com 20
RENO PLASMA STERILIZER

3-5. Check the Vacuum Pump

◆ Poor performance of Vacuum Pump


☞ if the vacuum value doesn’t fall down less than 400mmT /5 min

1) Check the working Ampere of Vacuum pump without load (standard value is 4.5A ~ 6A)
After making a vacuum for 3 minutes, measure the value with hook meter

2) If the measured value is over than standard value, check the pump condenser.
Check point

www.renosem.com 21
RENO PLASMA STERILIZER

3-5. Check the Vacuum Pump

3) Check the oil quantity & oil replacement

→After changing pump oil, check the vacuum value.


→ If the vacuum value doesn’t decrease by base value, it need to be replaced Vacuum pump.
Oil Inlet

Max Level
Reference Level

Minimum Level

Oil Drain

www.renosem.com 22
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

◆ The reasons of Error 2

☞ It causes when sterilant solution is not diffused to the chamber at sterilizing 1&2 stage.

User Mistake Unit Fault

6.Pressure gauge
1. Sterilant 2. Sterilization 3.Needle & 4. Injection 5. SFS Nozzle
& Vacuum Valve
reuse Item Mini Valve fault valve fault fault
fault

www.renosem.com 23
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

1. Sterilant reuse
◆ After complete sterilizing, if user doesn’t remove used sterilant cassette,
and then reuse it , the error 2 occurs.

2. Sterilization item

1) Non woven sheet reuse


2) Pouch reuse
3) Poor silicon pad use
4) Possibly absorptive materials use
5) Many Lumens use
6) Sterilization items load without tray( it is possible to interrupt sterilant injection)

www.renosem.com 24
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

3.Needle

1) Needle block (Refer to Injection flow time check )


2) Needle bending

www.renosem.com 25
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

3. How to change needle


union

1. Hold the backside union with spanner and then unfasten the front union
with other spanner.
2. Remove the used needle.
3. Insert the new needle and fasten both union again.

www.renosem.com 26
RENO PLASMA STERILIZER

Injection Valve Leak test

Prepare vacant sterilant Remove the sticker Push the red rubbers like picture

Vent Make a vacuum for 4minutes Insert a test sterilant Assemble the sterilant case

If injection valve has a leakage, red rubber is inserted to down side like picture.

www.renosem.com 27
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

4. Check the Injection Valve

1)How to change Injection Valve


Snap-ring

union

Supporting bolt

1. Remove the snap-ring.


2. Take out the nametag & solenoid
3. Remove the both unions & supporting bolts.

www.renosem.com 28
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

4. Check the Injection Valve

1)How to change Injection Valve

Injection valve supporting bolt

4. Remove the both injection valve supporting bolts.


5. Change the injection valve and the re-assemble by reverse order.

www.renosem.com 29
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

5. Check the SFS Nozzle

1) Go to maint.4 and then Push the “SV” button to decrease temperature of injection heater.
2) Change the setting value to 10degree and then push the ‘enter, button.
3) Push the ‘change’ button to confirm the setting value.
4) Check the temperature of injection heater.
If the temperature fall down less than 40 degree, try to check the SFS Nozzle.

www.renosem.com 30
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

5. Check the SFS Nozzle

Clamp High Temp. centering

1) Unfasten the Clamp.

2) Bore the injection port holes of SFS Nozzle with a proper tool.

www.renosem.com 31
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

6. Pressure gauge & Vacuum Valve

1) If the vacuum value change rapidly, pressure gauge need to be changed.

2) If as soon as sterilizing cycle starts, the vacuum values is “0”, , pressure gauge need to be changed

3) After hydrogen Peroxide is injected to chamber , if the vacuum value falls down from 20T to 0 slowly,
vacuum valve need to be changed because vacuum valve has a leakage.

www.renosem.com 32
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

◆ Injection flow time speed check

Blue : Inactivation state


Red : Activation state

1) Make a vacuum for 5minutes.

2) Close Vacuum Valve.

www.renosem.com 33
RENO PLASMA STERILIZER

2
Trouble Shooting
Error 2

◆ Injection flow time speed check

3) Open the Injection Valve1.


6) Check the Injection Valve2 by same way.
.
4) Count a second to reach 20460mT.
5) If it can be reached within 80seconds, injection
valve is normal.

www.renosem.com 34
RENO PLASMA STERILIZER

3
Trouble Shooting
Error 3

◆ Cause
- This error occurs when user notices problem and presses Cancel button.

◆ Action
- If you push ‘STOP ’ or “Cancel” button during the process,
it will be carried out as FINISH processing.
Don’t Open the door during FINISH processing
to protect user from vapor hydrogen peroxide in the chamber.
If you push ‘stop’ button before sterilizing 1 process,
It will take 1min for FINISH processing.
But if you push ‘stop’ button after sterilizing 1 process,
It will take 13min for FINISH processing.
And you can open the door after FINISH processing.
.

www.renosem.com 35
RENO PLASMA STERILIZER

4
Trouble Shooting
Error 4(Check the vacuum pump)

◆ Cause
- This error occurs due to abnormality of pump in any section during the process.
.
◆ Action
Oil inlet
1. Check result on printer paper.
2. Check Pump oil level(not Min) and color
(Transparent→Turbid).
3. Check the ampere with current measuring tool.
If on the current measuring tool it shows 4.5 ~ 6 ampere,
pump is damaged. Oil outlet
(How to check)
1) Hook tool to blown color electric line on EOCR.
2) Activate Pump in maintenance mode by manual.
3) Check ampere value on current measuring tool.

EOCR

www.renosem.com 36
RENO PLASMA STERILIZER

4
Trouble Shooting
Error 4(Check the vacuum pump)

4. When activates pump if it is not working and show error4 on the monitor,
check pump condenser inside of pump circuit box.
As opens pump circuit box if condenser is broken like picture, pump must be changed.

www.renosem.com 37
RENO PLASMA STERILIZER

5
Trouble Shooting
Error 5(Sterilization error)

◆ Cause
- This error occurs when the vacuum is not reached the setting value the
sterilization 2 cycle.
.
◆ Action
- same to error1

www.renosem.com 38
RENO PLASMA STERILIZER

6
Trouble Shooting
Error 6(Self test failed)

◆ Cause
- This error occurs when processing in not carried out smoothly during the self
test process .
.
◆ Action
- same to error1

www.renosem.com 39
RENO PLASMA STERILIZER

7
Trouble Shooting
Error 7,9,11,13(Chamber, Injection, Doors - Overheating)

◆ Cause
- This error occurs when the chamber temperature or front/rear door temperature
reaches over 58 degree and the injection heater temperature reaches over 220
degree

.◆ Action
- 1. Check printer paper result.
2. Check temperature on the Maint.4 screen
to check temperature on setting value (SV)
and Process Value (PV).
* Don’t turn off the machine when you check.
3. Check the SSR status. If the SSR is “turn ON”- short,
The heater is always working. Exchange the SSR Board.

www.renosem.com 40
RENO PLASMA STERILIZER

7
Trouble Shooting
Error 7,9,11,13(Chamber, Heater, Doors - Overheating)
4. Check the Heater Resistance working well or not.
ⓐ How to check chamber heater resistance (60Ω/ ±10%)

If it is shown infinity on multimeter it means that electric line is gone or short.


Change or find short position.
ⓑ How to check injection heater resistance (400Ω/ ±10%)

ⓒ How to check injection heater resistance (160Ω/ ±10%)

www.renosem.com 41
RENO PLASMA STERILIZER

8
Trouble Shooting
Error 8,10,12,14(Chamber, Injection, Doors – Heater Low )

◆ Cause
- This error occurs when the chamber temperature does not reach over 45 degree
and vaporizer temperature does not reach over 145 degree(in case of front/rear door
temperature setting value is 40degree)

◆ Action
- 1. Check printer paper result.
2. Check controller to check temperature on controller
after taking out right side outer cover.
3. Check how many time machine has been warming up.
At least 2~3hours must be heated up.
* Don’t turn off the machine when you check.

www.renosem.com 42
RENO PLASMA STERILIZER

8
Trouble Shooting
Error 8,10,12,14(Chamber, Heater, Doors – Heater Low )
4. Check the Heater Resistance working well or not.
ⓐ How to check chamber heater resistance (60Ω/ ±10%)

If it is shown infinity on multimeter it means that electric line is gone or short.


Change or find short position.
ⓑ How to check injection heater resistance (400Ω/ ±10%)

ⓒ How to check injection heater resistance (160Ω/ ±10%)

www.renosem.com 43
RENO PLASMA STERILIZER

8
Trouble Shooting
Error 8,10,12,14(Chamber, Heater, Doors – Heater Low )

5. Check the Heater SSR . If the SSR is “broken” or disconnection, The heater is not
working. Exchange the SSR Board or tighten the wires.
1) Check 24V 2) LED of SSR is always blinking when main power is supplied.

24V Rear door Heater SSR

Front door Heater SSR

Injection Heater SSR


Chamber Heater SSR

www.renosem.com 44
RENO PLASMA STERILIZER

8
Trouble Shooting
Error 8,10,12,14(Chamber, Heater, Doors – Heater Low )

3) Check 220V

www.renosem.com 45
RENO PLASMA STERILIZER

9
Trouble Shooting
Alarm 1 (Sterilant cassette is not inserted)
◆ Cause
- This alarms occurs when user push the start button without inserting sterilant cassette
at user mode.

◆ Action
1. Insert RENO-SA10 cassette into sterilant injection part.
2. HPS Sensor wire connection loosen or disconnection.
3. Check the HPS sensor damaged or position deformed. (If necessary, replace it)

HPS sensor

www.renosem.com 46
RENO PLASMA STERILIZER

10
Trouble Shooting
Alarm 2 (Remove sterilant cassette)

◆ Cause
- This alarm occurs when user push the start button with inserting sterilant cassette
at self test mode.

◆ Action
1. Remove RENO-SA10 cassette from the sterilant cart with Self-Test operation.
2. HPS Sensor wire connection short.
3. Check the HPS sensor damaged or position deformed(If necessary, replace it)

www.renosem.com 47
RENO PLASMA STERILIZER

11
Trouble Shooting
Alarm 3 & Alarm 4(front/rear door is not closed)

◆ Cause
- these alarms occurs when door up sensor didn’t detect the signal or door lock
sensor didn’t detect the signal.
◆ Action(Alarm 3)
1. The front door is not closed.
Close the front door.
2. Check the maint. 3 screen
and close/open front door by manual.
If the Alarm 12 happens,
the problem is front door up (FDU) Sensor.
① Check the FDU Sensor wire connection (FUD or RDU Sensor)
② FDU sensor damaged or position deformed.
(If necessary, replace it)
3. If the Alarm 14 happens,
the problem is Front Door lock sensor(DLS1).
① Check the DLS1 sensor wire connection.
② DLS1 Sensor wire connection is deformed or is damaged
(If necessary, adjust the sensor position or replace the sensor)
※ in case of Alarm 4, you can check the RDU sensor or
DLS2 sensor (DLS1 or DLS2)

www.renosem.com 48
RENO PLASMA STERILIZER

12
Trouble Shooting
Alarm 5 & Alarm 6(rear/front door is opened)

◆ Cause
- These alarms occurs when user tried to open the door with the opposite door
is opened or Door plate detection sensor didn’t detect the signal.

◆ Action(Alarm 5)
1. Close the Rear door and then open the front door.
2. If the rear door is closed but door indicator is opened,
① check the Rear Door Lock sensor (DLS II ) sensor wire connection.
② Check the DLS II sensor is damaged or it’s position deformed.
(If necessary, replace it)
※ in case of Alarm 6, you can check the DLS I sensor.

DLS I or DLS II

www.renosem.com 49
RENO PLASMA STERILIZER

13
Trouble Shooting
Alarm 7(Check the chamber pressure)

◆ Cause
- This alarm occurs when user try to open the door at vacuum state,

Go to Maint 2 screen, the PV value must be higher than SV value


1. Change the SV value(the difference of both value must be higher than 30mbar.
2. Check the VTV(exhaust valve) connection.
3. Check the Hepa filter is damaged or is blocked

※ Atmospheric pressure is different by the weather


and installation location (elevation).
If the PV(Process value) is lower than SV(Setting value),
doors are not opened and go to alarm 7.
So Atmospheric pressure could calibration
by SV(Setting Value) touching. Recommend value

www.renosem.com 50
RENO PLASMA STERILIZER

14
Trouble Shooting
Alarm 8(Air pressure low)
◆ Cause
- This alarm occurs when air tank pressure value is lower than 3.5 bar.

1. Check the Comp. pressure is lower than 3.5bar on the monitor mode(Y)
-> Check compressor or output air pressure.
2. Check the Air pressure Gauge problems.
3. Check the Air pressure leakage. (by using sound or water bubble)
4. Check the compressor and if necessary, replace it

(Main regulator) (Air Pressure gauge)


www.renosem.com 51
RENO PLASMA STERILIZER

15
Trouble Shooting
Alarm 9(Power Outage)
◆ Cause
- This alarm occurs when main power is suddenly cut off or user turn off the main s/w

1. Check the power outage occurrence.

1) if the power outage occurs during sterilizing process


① Go to ‘home’ screen and then select ‘Non lumen cycle’
② Select ‘start’ button.
③ If the process is completed or error2 occurs, remove the sterilant cassette.
④ Insert the new sterilant cassette.
⑤ Continue the sterilizing process.
2) If the power outage occurs during user doesn’t using it.
☞ you can use the unit as normal state.

3) In case of more than 1hours after power outage.


☞ you can wait at less 10minutes or at maximum 2hours.

www.renosem.com 52
RENO PLASMA STERILIZER

16
Trouble Shooting
Alarm 10 (Check front door motor)/ Alarm 11(check rear door motor)
◆ Cause
- This alarm occurs when front door motor is damaged or happens over current

◆ Action(alarm 10)

1.Push the ALMRST button at Maint 3.


2.open or close the front door.
3.If alarm occurs again, refer as below.
4.check the front door servo driver and then if any error message occurs,
contact to Renosem.
① Ano-00 Emergency stop: check the DC24V
② Ano-01 Over current : check the power, reset or replacement
③ Ano-03 Over Load : check the motor connection.
④ Ano-05 Line Fail: overload of ENCORD,, check the oveload or motor connection.
⑤ Ano-07 Follow ERR: Check the Gain, pulse of frequency, Encord, motor connection.

※ In case of Alarm 11, you can motor check rear door

www.renosem.com 53
RENO PLASMA STERILIZER

17
Trouble Shooting
Alarm 12 & Alarm 13(emergency stop(front/rear door))

◆ Cause
- These alarms occurs when Door up(down) limit sensor detected the signal because
door up/down sensor couldn’t detect the signal.

◆ Action(Alarm 12)
1. Push the ALMRST button at Maint 3.
If the alarm happens again, the problem is caused by FDU(Front Door Up) Sensor
or FDD (Front Door Down)sensor didn’t detect the signal.
① Check the FDU& FDD Sensor cable
② Remove the Side deco B cover, and then check the FDU& FDD sensor.
if the sensor is damaged, replace it.
③ to remove the alarm signal, you can remove the FEA(Front door emergency Alarm)
if you remove the this signal cable, you can move the door. And then, connect FEA
signal cable again.

※ In case of Alarm 13, you can check FDU & FDD, REA signal cable.

www.renosem.com 54
RENO PLASMA STERILIZER

17
Trouble Shooting
Alarm 12 & Alarm 13(emergency stop(front/rear door))

※ Check point

www.renosem.com 55
RENO PLASMA STERILIZER

17
Trouble Shooting
Alarm 12 & Alarm 13
※ Check point

(FDU or RDU)

(FDD or RDD)

www.renosem.com 56
RENO PLASMA STERILIZER

17
Trouble Shooting
Alarm 12 & Alarm 13(emergency stop(front/rear door))

※ Check point

www.renosem.com 57
RENO PLASMA STERILIZER

18
Trouble Shooting
Alarm 14 & Alarm 15(check front/rear door cylinder)

◆ Cause
- These alarms occurs when Door Lock Sensor (front/rear) did not detect a signal
because of low pressure of compressor or temporary missing sensor position.

◆ Action(Alarm 14)
1. Open &Close the Front door with a foot switch.
2. Check the pressure of compressor & main regulator
(it must be more than 4.0 bar)
① if not, check the air line of compressor has a leakage or not.
(If necessary, replace a compressor)
3. Check the front door lock (DLS I ) sensor connection.
4. Check the DLS I sensor position deformed or the sensor
& sensor bracket is damaged
※ in case of Alarm 15, you can check the DLS II sensor.

www.renosem.com 58
RENO PLASMA STERILIZER

18
Trouble Shooting
Alarm 14 & Alarm 15(check front/rear door cylinder)

※ Check point

(Main regulator)

DLSⅠor DLS Ⅱ

www.renosem.com 59
19 PM KIT 1 Procedure

1. Rotary Pump Oil (WooSung) : 3 liter


2. Activated Carbon Filter
3. O-ring (inlet & Outlet)
4. Oil paper & Glove
5. Oil Container
6. Funnel
7. Garbage Vinyl
RENO PLASMA STERILIZER

1. How to Change Pump Oil (Woo Sung)

✓ Preparation
- Glove, Oil Container, Funnel & Some paper,Oil
RENO PLASMA STERILIZER

1. How to Change Pump Oil (Woo Sung)

To Change Oil
1. Unscrew clamp connected oil mist trap
2. Unscrew oil Inlet plug.
RENO PLASMA STERILIZER

1. How to Change Pump Oil (Woo Sung)

Drain plug Drain Valve

3. Open Drain Valve and then take oil out.


4. Turn off Drain Valve after oil out.
RENO PLASMA STERILIZER

1. How to Change Pump Oil (Woo Sung)

Oil Inlet

Max Level
Reference Level

Minimum Level

5. Put Oil through the Inlet port.


Oil Drain
6. Fasten Inlet port.
7. Assemble Oil mist trap.
RENO PLASMA STERILIZER

2. How to Change O-ring (Inlet) 2-1. How to Change O-ring (Outlet)

O-ring(Outlet)
O-ring(Inlet)
RENO PLASMA STERILIZER

3. How to replace carbon filter

1. Loosen two NW25 Clamp.


2. Replace carbon filter.

www.renosem.com 66
PM KIT 2 Procedure

1. Oil Mist Trap (Woo Sung)

2. Heat-resisting O-ring

3. NW25 Clamp
RENO PLASMA STERILIZER

1. How to replace Oil Mist Trap (Woo Sung)

1. Loosen NW40 and NW25 Clamp.


2. Replace carbon filter.

www.renosem.com 68
RENO PLASMA STERILIZER

2. How to replace Heat-resisting O-ring

Unscrew it
※ Caution
Down the injection Vaporizer Temperature before handle the vaporizer.
On the temperature controller
Press Set button. → down the temperature with arrow button.
After 10 minutes, check temperature goes down, then handle vaporizer.

www.renosem.com 69
RENO PLASMA STERILIZER

2. How to replace Heat-resisting O-ring

Heat resisting O-ring

1. Loosen 4 bolts using with wrench.


2. Replace Heat resisting O-ring.
3. Reassemble all parts in reverse order.

www.renosem.com 70
PM KIT 3 Procedure

1. Catalyst filter

2. Injection Valve (For Shintec)

3. HEPA filter

4. Door Gasket
RENO PLASMA STERILIZER

1. How to change Catalyst filter

Catalyst filter

1. Loosen NW 25 Clamp.
2. Replace Catalyst Filter.

www.renosem.com 72
RENO PLASMA STERILIZER

2. How to change injection Valve (Sintec)

◆ Sintec type

1. Unscrew nuts. 2. Separate Union. 3. On the bottom side,


unscrew fixed 4 bolt on bracket.

6. Replace injection valves


7. Reassemble in reverse order.

4. Take out the below union. 5. Unscrew 2 injection valve supporting bolts

www.renosem.com
RENO PLASMA STERILIZER

3. How to change HEPA filter

1. Grab Hepa Filter body, turn in a counter-clockwise direction, then take out Hepa Filter.
2. To replace a new HEPA Filter, turn in a clockwise.

Caution: Arrow must be in a down direction when HEPA Filter is replaced.

www.renosem.com
RENO PLASMA STERILIZER

4. How to change Door gasket

1. Enter the maint mode No. 3 screen.


2. Open a opposite side door.
Push icons by the manual. First push OUT icon
to unlock the other door. And push DOWN icon.
3. Unlock the door that fall out gasket.
Push OUT icon.
4. Turn off the main power switch for cooling Temp.
5. Half hour later, put the door gasket on the door
track through the opened other door.
(go to next page)
6. After finishing the door gasket putting, Turn on the power.
And Lock (=Press IN icon) and doing the self-test.
You must check the vacuum in the chamber.

www.renosem.com 75
4. How to change Door gasket

When you replace door gasket, removing the air layer is very important.
If there is remaining air gap, there is possibility gasket would be off.
So, put the gasket on the door track and remove the air layer like following picture.

1. Put and push the gasket on the door track. Door gasket size is different the width and length.
Check the direction.
2. Push 45 degree on the gasket one direction with head of screwdriver.
And then the air layer gathers some point.
3. Put and incline the small thin flat screwdriver on the gap between gasket and door track. Remove
the air layers on the gap. And push & put the door gasket. Do that all door gasket surface.
4. Repeat 2 and 3 several times

1.Push
3.Push

Flat-blade ScrewDriver

2.Put and Incline

Air SPACE