How defibrillators work explained simply
We will discuss the following aspects. Please scroll down and start reading (and enjoy !):

 Basic explanation of fibrillation and defibrillation.


 What are capacitors and why we need to use them in defibrillators.
 Charging and discharging capacitors.
 How the defibrillator creates the “biphasic truncated exponential waveform”.
 Brief introduction to “automated external defibrillator” (AED).
 Brief introduction to “implantable cardioverter defibrillator” (ICD)

Introduction
We all know that electrical current can kill a person. However, current given using defibrillators also saves
lives. The human heart can develop an extremely serious condition called “ventricular fibrillation”, which if
not treated, leads to death in minutes. A defibrillator is a device which, as its name suggests, can stop
fibrillation. In this page of the website, you will learn the basics of how a defibrillator works. At the end,
you will be able to understand and draw the basic circuit of a defibrillator as shown below. Please do not
get a heart attack seeing this diagram ! I will explain it step by step and you will understand it all.
Basic explanation of fibrillation and defibrillation

Note: If you are a medical professional, hopefully you already know the basics of fibrillation and
defibrillation is. If not, do learn about it urgently from elsewhere ! However, in addition to medical
professionals, this website also gets a lot of non medical visitors who are interested in learning things to
satisfy their curiosity. For these visitors, before explaining details about how defibrillators work, the next
section will explain, in very basic terms, what fibrillation and defibrillation is. If you already know all this,
you may wish to just skim over it, or if you wish, skip to the next section by clicking HERE.
What is “ventricular fibrillation”?

In this section, I will explain to you in a simplified way, what “ventricular fibrillation” is. However, before we
begin, I need to warn you that I am not a good artist and therefore the “hearts” that I will draw in our
discussion will not look anything like the real heart in your body !

I am sure that you are aware that the function of the heart is to pump blood. You can think of the heart as
a pump consisting of a flexible “bag” surrounded by heart muscle cells. I have drawn the heart muscle
cells as yellow squiggly lines.
The heart muscle cells have some unique features. Firstly, and luckily for us, these muscle cells are very
strong ! To pump blood out, the heart muscle cells contract and become shorter. This “squeezes” the
heart and pushes the blood out (red arrow).

To effectively pump blood out of the heart, the heart muscle cells contract in a “coordinated way”. In a
very basic way, I have shown this in the images below. You will notice that first the lower heart muscle
cells contract (B), emptying blood from that area to the area in the middle. Then the muscle cells in the
middle also contract (C), emptying blood to the top. Finally the muscle cells at the top also contract (D).
You an see that as the muscle cells contract in a coordinated way, blood gets pumped out of the heart.

It is important that the heart muscle cells contract in a coordinated manner. If the heart muscle cells
contract in a disorganised way, blood will not be pumped out of the heart effectively. In the example
below, the heart muscle cells in the middle have contracted before the heart muscle cells at the bottom
have contracted. The blood therefore gets “trapped” at the lower half of the heart and the heart doesn’t
pump blood effectively.

To coordinate the muscle cell contractions, the heart has a “control system”. This control system uses
electrical signals to communicate with the muscle cells. I have shown this “control system” in dark blue
below.
Using medical terminology, the “control system” can be called the “pacemaker and conduction system”.
However, to keep things simple in our discussions, I will continue to call it the “control system”.

At regular intervals (i.e. for each heart beat), the control system sends electrical signals (shown as green
dots) to the heart muscles.

The heart muscle cells “listen” for the presence of the electrical signal sent from the control system. When
heart muscle cells “hear” the electrical signal, they contract. In the diagram below, the heart muscle cells
at the lower part of the heart have got the electrical signal, and are therefore contracting. The other parts
of the heart have not got this signal yet and are therefore in a relaxed state.

As the electrical signal travels further, more of the heart muscle cells contract. In the image below, you
can see that now the heart muscle cells in the middle have received the electrical signal and are
contracting.
Finally the electrical signal from the control system reaches the heart muscle cells at the top, which then
contract, pushing out the last bit of blood.

You can see that in this way, the control system, using electrical signals, is able to make the heart muscle
cells contract in a coordinated way and effectively pump blood out of the heart.

A very serious situation called “ventricular fibrillation” can happen to the heart where the electrical signals
are completely disorganised, making the heart unable to pump blood. Needless to say, if nothing is done
quickly about this, both the heart and the person it belongs to, would not survive beyond a few minutes. I
will now explain to you what ventricular fibrillation is.
Let me first explain to you what the word “ventricular” of “ventricular fibrillation” means. The images of the
heart I have drawn so far have looked a little like a “single bag”. In real life, the heart is composed of
four individual “bags”, or to say it more elegantly, “chambers”. The lower two chambers are very strong
and do most of the pumping work of the heart. These two lower chambers are called “ventricles”.

We will soon discuss an electrical disturbance called “fibrillation”. When fibrillation affects the “ventricles”,
it is called “ventricular fibrillation”.
Now that you know what “ventricles” are, let me explain what “fibrillation” is. To explain this, I will be
taking you back to school! Shown below is a small school where you can see children quietly sitting in a
classroom. At the front is the teacher who is teaching the alphabet by making the children repeat after
her. She starts by saying the letter “A”.

These lovely children repeat after her. As you can see, the children are learning well.
The teacher continues with the letter “B”.

I am sure that you will agree that the children are learning in a very “organised” manner. The teacher first
says something and the children listen. Then the children repeat what the teacher just said. These
children are full of good discipline!
The heart muscle cells, like the children you saw in the classroom, are able to “listen” and respond. The
heart muscle cells “listen” to their equivalent of the “teacher”, the “control system”.

You will recall that it is the control system that “tells” the heart muscle cells to contract in a particular
sequence. Because all the heart muscle cells listen and respond to the messages from the control
system, they contract in a coordinated way, effectively pumping blood out of the heart.
In other words, the heart muscle cells are like the good children we just saw in the school. Like the good
children, they too, “listen” and obediently follow what is told to them.

Let us go back to the school to see how the children are getting on. Things seem to be going well and the
teacher has reached the letter “ M ”.
But the relative calm in the classroom is about to change quickly. You see, there is one child (shown in
grey) who is getting bored of listening to the teacher. He is getting more and more “irritable”.

Eventually he can’t resist irritating the child next to him. Instead of listening to the teacher, he decides to
talk to the child next to him.
The child next to him gets distracted and replies.

Unfortunately, the comments made between the initial children starts to affect the other children nearby.
Soon the whole class is completely disorganised and all the children are erratically talking to and replying
to the children near them. Even though the teacher is bravely continuing to teach, none of the children are
listening to her. This is a disaster !

As I will explain to you soon, a similar thing can happen in the heart, making it go into ventricular
fibrillation. As explained before, normally the heart muscle cells “listen and obey” the “control system” of
the heart, so that they all work in a coordinated way to pump blood.
However, just like the naughty child in the classroom, a faulty heart muscle cell may become irritated and,
ignoring the control system, start on its own to electrically “talk” with cells nearby (blue arrows).

And just like in the classroom, this may eventually lead to all the heart muscle cells to erratically talk
(electrically) to nearby cells. Instead of listening to the control system, they are all “chatting” among
themselves!
The heart muscle cells are so busy talking (“electrically”) among themselves, that they ignore the
electrical messages coming from the control system. They now contract in such a disorganised way that
the heart appears to just “shiver” instead of contracting in an organised way. When this happens, the
heart is said to be in “ventricular fibrillation” and in this condition, it cannot pump blood. This is a serious
medical emergency and if nothing is done, the patient will die within minutes.

You may know what an electrocardiogram (ECG or EKG) is. Your doctor or other health care
professional can use it to see the electrical activity (voltage) of your heart over a period of time. As we
have discussed, the electrical signals in the heart are normally very organised. A normal ECG shows the
organised electric signals as a waveform that repeats itself in a regular way. The red arrows show the
beginning of each new heart beat. The waveform itself is somewhat complex, but hopefully you will see
that it has a regular pattern, indicating that the electrical activity in the heart is occurring in a nice
organised manner.

Now look at what happens to the ECG when the heart is in ventricular fibrillation. The disorganised
electrical activity causes the ECG pattern to be completely disorganised as well.
What does a defibrillator do?

Now let us see how we can sort both, the noisy classroom and the fibrillating heart. Let us start with the
classroom. Below is the disorganised classroom that you recently saw. However, this time there is an
extra person, shown in pink. Who is this extra person ? Well, that is you! You have been sent to sort out
the noisy disorderly children who are not learning anything at present.

The solution to the problem seems simple. You quietly and sweetly say; “silence”. Why quietly ? Well you
are a nice sweet person and don’t believe in raising your voice.
To your surprise (and perhaps anger ?) you find that nothing happens. You are being ignored and the
disordered chatter among the children is continuing.

Now you decide to show your true strength. You take in the biggest breath of your life and shout as loud
as you can, “SILENCE !”
To your relief, something seems to be happening. There is complete silence ! Both, the teacher and the
children are completely silent.

And then the teacher tries to teach again.


The children, who can now hear the teacher, start to learn ! Everyone, especially you, is happy that this
situation has been solved.

Now let us return to the fibrillating heart. As you will recall, in the fibrillating heart, the individual heart cells
are crazily talking (electrically) to each other instead of “listening and obeying” the control system. This,
as you know, leads to disorganised contractions and the heart cannot pump blood.
So, just like you did with the children in the classroom, you need to tell all the heart muscle cells to be
“silent ”. Since the heart cells communicate with each other using electricity, it is possible to “electrically”
tell all of them to be “silent”. This is can be achieved by passing a large electrical current through the
heart. Let us do it now ! We place two large metallic pads (grey in diagram) on either side of the fibrillating
heart. We send a large current (blue circles) between the pads. The large electrical current goes from
one pad, through the heart, to the other pad. Sending a large current in this manner is the “electrical” way
of shouting “silence !” to the chatting heart muscle cells.
The exact mechanism of how a large current silences the electrical activity of the heart is a bit too
complicated to explain on this equipment website. However, soon after the large current goes through
the heart, the heart muscle cells stop sending electrical signals to each other and become electrically
“silent”. Just like the silence that was there in the classroom after you shouted “silence”.

Now that everything is quiet, soon the heart muscle cells “hear” the coordinated electrical signals coming
from the “control system” and start to contract in a organised way. The heart is therefore now able to
pump blood effectively.
The diagram below summaries what we just talked about. We start with a heart in fibrillation (A). We now
say the equivalent of “silence !” by passing a large current through the heart (B). All the heart muscle
cells become electrically silent (C). Now the heart muscle cells can “hear” the signals coming from the
control system and they start to contract in a organised manner (D and E).

The use of a current to stop “fibrillation” is called “defibrillation”. And what is the device that passes the
current through the heart to make defibrillation happen called ? You guessed it ! It is called a “defibrillator”
!
I would like to take you briefly back to the noisy class room. You will remember that when you “quietly”
said “silence” that nobody listened to you and the classroom remained disorganised.

But when you loudly said “silence”, everyone listened and obeyed you. In other words, you had to shout
loudly to make everyone hear you.

With defibrillation, a similar concept applies. To defibrillate an heart, the electrical current passed through
it needs to be large enough to silence all the heart muscle cells. A small current would not work. A major
design requirement of defibrillators is to have the ability to provide the necessary large current. How it
does this will be explained to you soon.

The need for a “capacitor”

A crucial electrical component in defibrillators is a device called a “capacitor”. To explain what a capacitor
is, I have decided, with my magical powers, to make you a fireman (or fire woman!). Shown below is you,
proudly wearing your red fireman’s uniform. Next to you is a water pipe, which of course will be your main
weapon to combat any fires you will face. Like all visitors to this website, you are a dedicated worker and
are happily waiting for your first fire.
While waiting for the first fire to happen, you decide to test your equipment. You turn on the water supply
and to your surprise, the water flow is very low. Thinking that there must be some fault, you immediately
tell your boss about it. She tells you that, to save money, that she will only provide a low flow of water.
You plead with her for a higher water flow, and all she says in response is; “No !”

While you were arguing with your boss, a real fire needs your urgent attention. You run with the water
pipe and let the water flow into the fire.
Well, you are not surprised that the low water flow is not putting the fire out. Just as you told your boss,
the water flow is simply not enough to put the fire out. In fact it appears that the fire is laughing at you !

You are not the type who gives up and you start to think about a solution. The water that is being supplied
to you is coming at a continuous low flow. However, what you need is a brief high flow of water that will
once and for all put the fire out.
You suddenly think of a great solution. With your own money (you know your boss won’t give you any)
you buy a bucket! You then do the following things. I will show the things you do in some detail as it will
be relevant when we discuss the role of the capacitor in defibrillators. You first, using the “continuous low
flow” supply of water, start filling the bucket.

After a little while, you have a bucket that has been filled with enough water to put the fire out.
Now you tip the bucket over the fire. A brief but high flow of water falls onto the fire, putting it out. You
have won the battle ! Your boss is happy and “promises” to pay for the bucket someday.

Now that you have put the fire out, let us return to defibrillators. As you may have guessed, there are
some “similarities” between putting the fire out and defibrillation . As a fireman, to successfully put the fire
out, you needed a brief large flow of water. In a similar way, as you learnt earlier in our discussion, you
need a brief large current to defibrillate a heart.
You will recall that when you were a fireman, the big problem you had was that the water supply provided
to you had a very low flow. Without the bucket, this low water flow was simply not enough to put the fire
out. The bucket was your savior !

The designers of defibrillators face a similar problem to the low flow water supply you faced as a fireman.
All electronic devices, including defibrillators, need a source of power (current) to work. For an example,
because defibrillators have to be easy to transport, they often get their power from batteries.
Unfortunately, batteries can only produce a continuous low flow of current. The defibrillator has to
“somehow” use the continuous low flow current from the batteries to provide the brief high flow current
needed for defibrillation.

Defibrillators can also get power from the wall sockets in your hospital. However, even the maximum
current available from such wall plugs is not enough to directly produce the high current flow required for
defibrillation. I will further explain this.
As a very approximate example, the current needed for defibrillation can be in the region of 20 A
(Amperes). So how big is a current of 20 A ? Shown below is a diagram of a table lamp. A modern lamp
such as this consumes about 0.02 Amperes.

As shown below, when you do the calculations, you will see that the 20 A defibrillation current equals the
current consumed by approximately 1000 table lamps ! So even when the defibrillator is plugged into a
wall power socket, the current from the socket cannot be used directly to provide the huge current needed
for defibrillation. How the defibrillator is able to produce the large currents needed for defibrillation from
batteries or wall power sockets involves some magic which I will reveal to you soon ! By the way, please
do not learn the numbers shown below as they are very approximate.
At this point, I would like to “warn” you that the discussion ahead will involve us dealing with some basic
electrical concepts. Hopefully this is not too surprising to you as defibrillation is all about electricity!
However, do not worry too much if electricity frightens you as I will try and keep things as simple as
possible. Also, this website has a separate section explaining the “basics of electricity”. If words like
current, potential difference, alternating current and direct current make little sense to you, it might be a
good idea to first visit the “basics of electricity” section of this website by clicking HERE (please remember
to come back!).

When you were a fireman, the bucket was the savior. The defibrillator needs the electrical equivalent of
the bucket. It needs something that can collect the continuous low flow of current from batteries or wall
power and then release it as a brief large flow of current.
Fortunately there is such a device and it is called a “capacitor”. Capacitors play an extremely important
role in defibrillators and I will of course explain to you how they work. The capacitor in a defibrillator is
able to collect the continuous low flow of current, store what it has collected, and then release it as the
brief large flow of current needed for defibrillation. The basic design of a capacitor is extremely simple. It
consists of two parallel flat metal plates, shown below as blue and pink squares. Wires (shown in green)
are connected to each of the plates.
The metal plates lie very closely next to each other. In real life, the plates are so close to each other that
you would need a magnifying glass or microscope to see the separation. I however, cannot drawn them
so close, so do remember to use your imagination when you see my diagrams.

In the world of electronics, when a circuit is drawn, symbols are often used to represent various devices.
Fortunately for us, the symbol representing a “capacitor” is very easy to remember as it looks so much
like the object it represents.
How capacitors work

As explained before, a capacitor consists of two metal plates lying next to each other. To know how
capacitors work, we need to first understand how a metal can become “electrically charged”. The actual
physics of how this happens is somewhat complex and it is not necessary for our purposes to understand
it in great detail. If you struggle to understand what I explain in this section, please just skip it and move
onto the next section. Let us start ! Metals have “positive charges” and “negative charges” inside them.

I would like you to, only for this discussion, think of the charges inside the metal as being very romantic
boys and girls! The negative charges and positive charges are strongly attracted to each other. Given
any opportunity, the opposite charges would love to form a relationship and be together as a happy
couple !
However, the positive charges are very “lazy”. Let me explain! Say that in the metal, there are positive
charges at one end and there are negative charges at the other end. As explained before, the positive
charges and negative charges are attracted to each other and they would like to be together to form a
happy couple. In this situation, you would expect both charges to move towards each other.

However, what happens is that only the negative charges move to be with the positive charges. The
positive charges stay fixed in one place, which is why I call them “lazy”.
The same situation exists when a positive and negative charge couple split up. Say a negative charge is
interested in another positive charge (negative charges are not very faithful !). When the negative charge
leaves the positive charge to be with its new lover, the positive charge it was with will not “follow” it. It just
lazily remains fixed in one place.

So basically, when attracted to each other or when leaving each other, it is only the negative charges that
move.
From a physics point of view, we are not interested in the happy couples (i.e. where a negative charge is
together with a positive charge). These “happy couples” (shown in grey below) are for the moment stable.
What interests us are only the charges that do not have a partner (i.e. the ones who are lonely and are
desperate for a partner, shown below with hearts). It is these single “available” charges that are active in
electrical circuits.

Now let us understand what we mean when we refer to the “charge” of a metal plate of a capacitor.
Shown below is one metal plate of a capacitor. I have drawn this plate as a very thick one as this will help
me to explain things better. A metal plate of a capacitor can have no charge, be positively charged, or be
negatively charged. Let us first start by seeing a metal plate with “no charge”. Such a plate has an equal
number of positive and negative charges inside it. Being “romantic”, these charges form “couples” and
therefore there are no extra positive or negative charges left. As mentioned before, in physics, we are
only interested in extra positive or negative charges. When we refer to the “charge ” of a metal plate, we
are actually referring to how many extra “free” positive or negative charges are there. Since the metal
capacitor plate in example below has no extra charges, we can say that this plate has “no charge”.

Let us now make a capacitor by putting another metal plate next to the plate we saw before. Since this
new metal plate also has no extra charges, we can also say that it has “no charge”.
A capacitor with “no charge” in it has no energy for defibrillation. It is like an empty water bucket that
needs to be filled before it can be used to put a fire out.

A capacitor needs to be “charged” before it can be used to provide current for defibrillation. Charging a
capacitor is the equivalent of filling the bucket with water. During charging, electrons (i.e. current) goes
into the capacitor and collect there (i.e. get stored).
Shown below is a capacitor that has been “charged”. In this “charged” capacitor, plate A is “negatively
charged” (because it has excess free negative charges) and plate B is “positively charged” (because it
has excess free positive charges). In this charged state, the capacitor is capable of giving the necessary
current required for defibrillation. In the sections that follow, I will explain to you how a capacitor is
charged, and then, how it can release that charge (discharge).

To understand how a capacitor is charged, let us first recall some basics about electricity. You will
remember that electrons flow only when there is a potential difference across the wire that carries them.
Shown below is a battery, which as you know, is something that provides a potential difference. Here the
potential difference provided by the battery is making the electrons (current) flow in the wire.

It is quite messy to draw a battery looking like the one above. So instead, I will use the
international symbol that represents a battery. Please note that the battery symbol looks somewhat
similar to the capacitor symbol, which I have shown in grey for you to compare.
We will soon connect our capacitor to a battery to see what happens. However, before that, let us see
that happens when we connect the capacitor to a wire that has no potential difference (i.e. there is no
battery). As shown below, both plates of the capacitor have no extra charges. In other words, both plates
have “no charge”.

Now let us connect the capacitor to a battery. The battery creates a potential difference across the
capacitor. As I will explain to you soon, interesting things will happen at both plates of the capacitor.
To make things easy to explain, I have labelled the plates of the capacitor as “A” and “B”. Let us first see
what happens at plate B. Because of the potential difference, the negative charges (electrons) in plate B
are attracted to the battery. These electrons (which are not very loyal to their partners !) leave their
positive charges and go to the battery. You will recall that the positive charges, being lazy, don’t run after
electrons that leave them.

As you can see below, because electrons have left plate B, there are now free positive charges in this
plate. Therefore we can say that plate B is “positively charged”.
Now let us see what happens in the other plate (i.e. plate A). The electrons in the battery are “attracted”
to the positive charges on plate B. These electrons therefore go from the battery towards plate A,
“hoping” to meet the free positive charges on plate B.

However, the electrons from the battery cannot cross from plate A to plate B as there is a gap between
the two plates. So while the electrons in plate A are attracted by the positive charges in plate B (attraction
shown as hearts in diagram below), they can only “admire” them from across the gap. This attraction
between the charges is only possible because the plates in real life, unlike what is shown in my drawings,
are placed very close to each other.
You can see that now there are free negative charges (electrons) that have collected in plate A because
they cannot cross the gap between the plates. We therefore can say that plate A is “negatively charged”.

In summary, when a potential difference is applied across a capacitor, excess positive charges will form
on one plate and excess negative charges will collect on the other plate. This process of making one
plate positively charged and making the other plate negatively charged is called “charging the capacitor”.
After a while, the capacitor has quite a lot of positive charges on one plate and negative charges
(electrons) on the other plate. It is the electrical equivalent of filling the water bucket with water when you
were attempting to put that fire out.

After a while, the capacitor cannot collect any more charges. We can now say that the capacitor is
“charged”.
You have seen how a capacitor “collects” the current that flows into it. Now let us see how a capacitor
“stores” the charges that it has collected. Let us disconnect our capacitor from the battery and see what
happens. When we do this, the collected charges do not just fall out of the capacitor !

Instead the the charges just remain there. So what holds them in place?
You will remember the free negative and positive charges are attracted to each other. So even though the
capacitor is disconnected, this attraction across the gap between the plates keeps the charges in place.
This is how the capacitor is able to “store” charges.

So far we have seen how a capacitor “stores” charge. I will now explain how we can “use” the stored
charge. Shown below is a charged capacitor and a light bulb next to it. At the moment the light bulb is not
connected to the capacitor. As it is not connected to anything, the light bulb is not light up.
Let us connect the light bulb to the charged capacitor and see what happens. When we connect the light
bulb to the capacitor, we now provide a easy pathway (shown in green) for the charges to meet each
other.

The electrons (negative charges) rush through the wires and bulb to meet the positive charges. The
movement of electrons (i.e. current) through the bulb makes it light up. The process whereby the
capacitor gives up its charges and makes a current flow can be called “discharging the capacitor”. When
the negative charges meet the positive charges, they form couples (shown in grey).
Capacitors can very quickly discharge the charges they store. This property of capacitors makes them
very useful for defibrillators, where a large amount of electrons (current) needs to be sent through the
heart quickly. It is like rapidly emptying the bucket to put out the fire.

Eventually all the free negative and positive charges meet up to form couples. There are no more “free”
negative charges left to move. i.e. the current stops flowing and the bulb is no longer lights up. The
capacitor is now said to be “discharged”.
So you have seen how a capacitor (like the water bucket you used to put the fire out) can be used to to
collect electrons from a low flow current, store them as a charge, and then discharge them rapidly. The
capacitor in a defibrillator is used in this way and we will soon discuss these steps in more detail.

The “ability” of a capacitor to store charge is called “capacitance”. For a given potential difference, a
capacitor with a “high capacitance” will store a high amount of charge whereas a capacitor with a “low
capacitance” will store low amount of charge. Because defibrillators need a very large charge to be
stored, the capacitors used in defibrillators need to have a high capacitance. One way to increase the
capacitance of a capacitor is to use plates with a large surface area. Shown below are two capacitors
connected to the same potential difference. The capacitor on your right has plates with a larger surface
area and therefore it has a higher capacitance.
However, a capacitor with such large plates would not fit into a defibrillator. The clever engineers however
have found a solution. They “roll” the capacitor into a cylindrical shape! In this way, the capacitor can
have a high capacitance (because the plates have a large surface area) and still be made relatively small.

How I first learnt about capacitors !


Let me now take you to a slightly different topic, i.e. that of my childhood ! When I was child, I used to
often go to local junk yards and collect various discarded electrical things to try and “repair” them.

Of course I had no real knowledge of how to repair things. And alarmingly, I used to think that since the
broken electrical items were not plugged into anything, that they must be “safe” to work with. In reality,
many of the electronic items that I used to pick up had capacitors in them, that were occasionally fully
charged. So even though they were not plugged in, I would occasionally get rather big shocks when my
little fingers met capacitors!
I don’t have many pictures of my childhood, but I did manage to find this one. It shows me playing with a
powerful electrical motor. I now realise, looking at the picture, that the motor had a large capacitor
(highlighted in red) attached to it! I suppose I should consider myself to be lucky to be alive. My parents of
course did not understand what I was up to!

Defibrillator circuit

Charging the capacitor

We will now, step by step, learn about the various parts of the defibrillator circuitry. As mentioned before,
the capacitor plays an important part in defibrillators, as it is able to be charged using a low flow current
and then is able to discharge a brief but high flow current to the heart. Let us start our discussion on the
defibrillator circuitry by describing the system that charges the capacitor. When we want to use the
defibrillator to defibrillate some unfortunate person, the first thing we need to do is to “charge” the
capacitor. This is done by the “capacitor charging system” in the defibrillator.
The actual workings of the charging system is somewhat complex and I do not think it is necessary for
you to know it in too much detail. The charging system involves a variety of voltages and it also deals with
direct current (DC) and alternating current (DC). Depending on your requirements, you may think it wise
to just skim over the next section, ignoring the bits you do not understand.

The charging system needs a source of current to charge the capacitor. This can be from batteries or
from the wall power socket in your hospital.
To have a charge large enough for defibrillation, the capacitor needs to be charged using a very high
voltage (e.g. 2000 Volts).

The main component of the charging system is the “step up transformer” (which was discussed in the
electricity basics section of this website). The step up transformer is able to raise (“step up”) the voltage
that comes into it.
The step up transformer in the charging system takes in current from either the batteries or the wall power
source. For an example, it is able to raise the supply voltage from the wall power source (e.g. AC 230
Volts) to AC 2000 Volts. However, capacitors can only be charged with a DC current, whereas step up
transformers put out current in the form of AC current. So there is a device that converts the AC current
from the transformer to DC current to charge the capacitor. Please note that all these voltages and
circuits that I describe are only examples that I am using to explain things. Please do not use them to
design “real life” defibrillators!
Defibrillators often need to be carried to the patient (i.e. need to be portable) so they often use batteries
as source of power. The step up transformer is able to raise the battery voltage to charge the capacitor
(e.g. from 12 V to 2000 V). However, step up transformers need the input current to be in the form of AC
current whereas batteries provide current in the form of DC current. Therefore, a special device converts
the DC current from the batteries to AC before the current goes into the transformer. You will happy to
know that there are no more conversions !

The operator of the defibrillator chooses the settings on the defibrillator. The settings on currently used
defibrillators do not mention things like “amount of charge” or “current” to be given to the patient. Instead,
the measurement unit used in defibrillators is the “Joule”, which is a measure of “energy”. The operator
chooses the energy level, in Joules, to be given to the patient.
The energy, in Joules, stored in a capacitor is given by the following equation. The mathematics is a bit
complicated so I have not explained the derivation of the formula. In the formula, “capacitance” is a
measure of the ability of the capacitor to store a charge. For our discussion here, we will ignore it. From
this equation, you can see that the energy stored by a capacitor is proportional to the voltage (potential
difference) across the capacitor. Higher the voltage across the capacitor, higher will be the charge stored
in the capacitor.
To charge the capacitor, the charging system will apply a potential difference (Volts) across the capacitor.
The potential difference applied to the capacitor will be adjusted to match the energy (Joules) selected
to be delivered to patient. Higher the energy to be delivered to the patient, higher will be the potential
difference applied to the capacitor to charge it.

The defibrillator circuit, including the section concerning charging of the capacitor, makes use of electrical
switches. In real life, the user doesn’t directly touch these switches with their hands as the voltages
involved are very high and the switches need to operate rapidly. Instead, the switches are controlled by
sophisticated computers. I.e. the user controls the computer, which in turn controls the various switches. I
will use the following symbols to show switches in my diagrams.
In my explanation of the defibrillation circuit, I will be using six electrical switches as shown in the diagram
below. For now, we need to only worry about the charging part of the circuit (shown in green box) .

The charging system is connected to the capacitor via two charging switches (A and B). When the
capacitor is not being charged, both these switches are in the “off” position.

To the charge the capacitor, charging switches A and B are put into the “on” position and the charging
system applies a high voltage across the capacitor. As discussed before, the magnitude of the voltage
applied is based on the energy level selected by the operator. Higher the energy level selected, higher
will be the voltage needed to be applied across the capacitor.

Current from the charging system flows into the capacitor and the capacitor starts to get excess positive
charges in one plate and excess negative charges in the other plate. The charging system monitors
the charging process.
Once the capacitor has adequate charge appropriate for the needed defibrillation energy, the charging
switches (A and B) are put to the “off” position to stop further charging. The charging process is now
complete and at this point, medical personnel will often say something like “defibrillator charged”. It is a
warning to everyone that soon the discharge will happen.

Discharge part of defibrillator circuit

The capacitor is charged and is now ready for discharge into the patient. In the diagram below, the
“discharge part” of the defibrillator circuit is shown in the green box. You will notice that there are four
switches (1,2,3,4) in this area. These switches will, later, connect the heart to the capacitor. However, at
present, these four switches are “off” and therefore the charged capacitor is not connected to the heart.
The capacitor, at this point, is simply storing the charge.
In my diagrams, the heart is of course drawn only as a symbol. The circuit connects to the heart via two
pads (shown in green).

Of course the real human heart is very different to the ones I have drawn in my diagrams. In actual use,
the defibrillation current is applied to two areas of skin on the chest of the patient using special
“conductive pads”. Current goes from the defibrillator to the patient and returns from the patient to the
defibrillator via wires connected to the chest pads. The current goes from one pad to another, passing a
variety of structures such as skin and of course the heart. By the way, please do not use my rough
diagrams to guide where you actually place pads in a real patient !

Use of oscilloscope to see waveforms

During defibrillation, the current that goes through the heart changes over time. To study the flow of
current over a period of time, one needs a device called an “oscilloscope”. An oscilloscope has a screen
which has two axes as shown below. The horizontal axis shows time and the vertical axis shows the
quantity we are measuring, which in the case of the defibrillation, is “current”. The oscilloscope is able to
plot a graph of current over time and it will be very useful in our discussions ahead. The oscilloscope
below has been connected to the defibrillator output. At the moment, the capacitor is not connected to the
heart (i.e. switches 1,2,3,4 are all “off”) and therefore no current is flowing through the heart. The
oscilloscope shows this as a flat trace (red line shown below) since the current flow over time is zero.
To make my diagrams less cluttered, I wont be showing the wires between the heart and oscilloscope
anymore.

The tracing on an oscilloscope showing the defibrillation current over time can be called the “wave form”.
Basic discharge waveform of a capacitor

As you may recall, for the capacitor to discharge, the electrons (negative charges) need a path to flow
from one plate of the capacitor to the other plate of the capacitor. We can create this pathway by
switching “on” switches 1 and 4. The capacitor now discharges via the pathway shown by the blue
arrows. The oscilloscope waveform shows that initially the discharge current flow is high and that it then
decreases over time. This is a typical waveform seen during discharge of a capacitor.
Let us look at the waveform in more detail. When a capacitor discharges, the current flow over time
decreases in what is called an “exponential” way. When a current decreases in an exponential way, the
rate of the decrease (i.e. how much current decreases over a given time period) is proportional to current
flow at that time. The oscilloscope below demonstrates the exponential decrease of the current from the
capacitor. The pink horizontal bars represent equal time periods. You can see that for these same time
periods (pink bars), the rate of decrease (vertical blue lines) decreases over time. When the current flow
was high (i.e. at start of graph) the rate of decrease was high (tall blue line), whereas when the current
flow was low(i.e. towards end of graph) , the rate of decrease was low (short blue line). Unfortunately, it is
beyond the scope of this website to explain “exponential” decreases in more detail. If you are interested
(it is an important concept in engineering and medicine), there is plenty of material online regarding it
(type the words “exponential decay” into Google).
“Monophasic” waveform

I will soon explain to you what a “monophasic” waveform is. The word monophasic relates to the
“direction” of current flow. As you see below, at a given moment, the current can travel in one of two
directions through the heart. I will call current that goes from your left to your right as “forward current”.
And similarly, I will call current that goes the other way, “reverse current”.
The oscilloscope can, in addition to showing us the magnitude of current flow over time, also show the
“direction” of current flow over time. In the example below, you can see that the current flow is “forwards”.
In our oscilloscope, the “forward” current flow is shown by the waveform going upwards.

Let me now show you how the oscilloscope waveform looks when the current is made to go through the
heart in the reverse direction. We can make the current go in the reverse direction by discharging the
capacitor through different switches to those that we used before. This time we put “on” switches 2 and
3. As you can see below, this makes the current flow in the reverse direction. The oscilloscope shows this
reverse direction current by displaying the waveform of current in the opposite direction to before (i.e. the
waveform goes below the baseline).
Let us return to our first example, where we put switches 1 and 4 “on”, making the the current go in the
“forward” direction. As you saw before, this makes the oscilloscope show an exponential waveform,
which as the capacitor discharges, returns to the baseline (zero). Please note that the waveform in the
oscilloscope has never gone below the baseline. This tells us that, in our example, the discharge current
only went in “one direction” (i.e. forwards). A waveform showing current flowing in only one direction is
called a “monophasic waveform” (“mono” means one and “phasic” sort of means “direction”).
As you know, the capacitor discharge waveform is “exponential”. So, combining monophasic and
exponential, this kind of waveform is called a “monophasic exponential” waveform.
In the past (before around 15-20 years ago), defibrillators were designed to produce monophasic
exponential waveforms. However, it has been found that a different type of waveform is more effective at
successfully defibrillating hearts. The newer waveforms are called “biphasic waveforms”, which
essentially means that during defibrillation, the current first goes in the forward direction and then
changes direction and goes in the reverse direction. Shown below is an example of a biphasic waveform.
I will explain to you later about how biphasic waveforms are delivered.

The old defibrillators which delivered monophasic current had only two discharge switches (green
arrows). You will see later that the biphasic defibrillator circuit has two more switches (shown in grey
below).
The monophasic defibrillators also had a large component called an “inductor”, which was a large coil of
wire (shown in blue below) placed in the circuit. The inductor “slows” the current, making it have a lower
peak and “spreading” it over a longer time period (see blue waveform). These effects were considered
good for clinical reasons. I mention the inductor here only for historical reasons and will not waste your
time explaining how it works. We will soon move onto biphasic defibrillator circuit designs and there you
wont see the inductor, as it is no longer plays an important role.
I do need to give a warning to the many anaesthetists (medical personnel who put patients to sleep for
surgical operations) visiting this website. Unfortunately, there may be some anesthesia textbooks and
other learning resources which explain the monophasic defibrillator instead of the newer biphasic
defibrillators. If you come across such, please contact the author and gently suggest that it be updated.
Similarly, some examination bodies may have not updated their “question bank” to include biphasic
defibrillators. If you are able to, please try and check with them.

Biphasic defibrillator

Let us now learn about how the modern biphasic defibrillator works. The charging system of the biphasic
defibrillator is very similar to the monophasic defibrillator, so I wont repeat how it works here.
To simplify things and to make room for clearer diagrams, at times I will not show you the charging part of
the circuit. If space is very tight, I may even omit showing the capacitor. So even when not
shown, please always assume that there is a correctly charged capacitor.

We are now ready to learn how the biphasic waveform is created in modern defibrillators. In the image
below, I have enlarged the biphasic waveform. To simply my explanations to you, I have divided it into
four parts and, as shown below, have given each part of the waveform a name. The waveform may look
complicated to some of you, but please do not worry. Once you understand it, you will be able to draw it
easily. In the next sections, I will explain to you how each segment of the waveform is created by the
defibrillator.

Forward current part of biphasic waveform

The biphasic defibrillator starts by sending current through the heart in the “forward” direction. This is
done in a similar way to how the monophasic defibrillator delivers current to the heart. i.e. switches 1 and
4 are put “on”. However there is crucial difference. In the case of the biphasic defibrillator, these two
switches are switched “off” before the capacitor has finished discharging. When these switches are
switched “off”, the current flow to the heart stops and the oscilloscope tracing returns to the baseline
(zero). However, the capacitor still has some charge left in it, as was prevented from emptying
completely.
“Switching interval”

As you saw previously, the forward current was stopped by switching “off” switches 1 and 4. The biphasic
defibrillator now needs to “reverse” the direction of the current flow to the heart. It does this by switching
“on” switches 2 and 3, which as you can see below, makes the current go in the reverse direction in the
heart. The oscilloscope shows the reverse current by making the graph go below the baseline.
However, if you look carefully, you will see that there is a brief period between the forward current and
reverse current, where the oscilloscope graph shows no current flow. In the magnified the oscilloscope
graph shown below, you can see this period as a short red line on the baseline (zero). I call this time
period the “switching interval” and will explain to you why it is there.

The switching interval starts when the “forward current” switches (1 and 4) are switched “off” and ends
when the “reverse current” switches (2 and 3) are switched “on”. During the switching interval”, all four
switches are in the “off” position. The “switching interval” is something that has been deliberately added
by the designers of defibrillators. The switching interval provides adequate time for the “forward” current
switches (1 & 4) to first “properly” switch “off” before the “reverse” current switches (2 & 3) are put “on”.
Let us imagine for a moment that the defibrillator has no switching interval, so that we can understand its
role. Without a switching interval, there is a chance that the forward current switches 1 & 4 don’t have
“enough time” to switch “off” before the reverse current switches 2 & 3 switch “on”. In this situation, all 4
switches could be briefly “on” at the same time.

This situation causes a “short circuit” to develop across switches 1 and 3. The current from the capacitor
returns directly back to it via a pathway through switches 1 and 3 (see red wires and arrows) and does
not go to through the heart. The energy in the short circuit current gets dissipated in the form of heat.
One big problem of the short circuit situation that we just described is that it “wastes” the capacitor charge
in the form of heat. The capacitor could end up losing most of its charge in this way.

Now in our example, even though the short circuit condition no longer exists, there is no charge left in the
capacitor to give the reverse current. I hope our discussion has shown you the importance of having
a “switching interval” between the forward current and reverse current parts of the biphasic waveform.
Let us see how the oscilloscope shows the “switching interval”. As you know, during the switching
interval, all the four switches are “off”and therefore no current flows through the heart during it.
Therefore, the oscilloscope shows the switching interval as a straight line on the baseline (indicating no
current flow through the heart). As shown in the oscilloscope view below, the switching interval line is
found between the forward current and reverse current parts of the biphasic waveform.

In real defibrillators, the switches are electronically controlled and operate at very high speeds. Therefore
they need only a short “switching interval” between the end of the forward current and the starting of the
reverse current. That is why you only see a short “switching interval” on the oscilloscope (green arrow and
short red line).

“Reverse current”

Let us now turn our attention to the “reverse current. As you saw before, to give the heart the “reverse
current”, switches 2 and 3 are switched “on”. The reverse current makes the oscilloscope graph go below
the baseline (red arrow).
In the diagram below, I have named the end of the reverse current as “reverse current natural end”. What
I mean is, this is how the waveform would look like if we let the capacitor completely discharge into the
heart, without us doing anything more to the switches. In this situation, as demonstrated below, the
waveform shows that the reverse current gradually decreases and eventually becomes zero. As I will
explain later, in real life, we manipulate the switches and prevent the reverse current from ending in a
“natural” way.
“Truncation”

As mentioned before, if we don’t change any switches, the “reverse current” returns to the baseline
“gradually” (see red arrows). Unfortunately, as I will explain to you, this “gradual” decrease at the end of
the waveform is problematic.

You will remember from our discussion on the basics of defibrillation that the defibrillation current needs
to be large to be effective.
Conversely, if a small current is used in defibrillation, a rather dangerous effect may occur. A small
current may actually cause fibrillation in the heart! The mechanism of how this happens is unfortunately
beyond the scope of this website.

So, from a defibrillation point view, we need to use reasonably large currents to have effective
defibrillation and we also need to avoid small currents that can cause fibrillation. Keeping these in mind,
let us look at the biphasic waveform. You can see in the example below, that the “forward current” and
most of the “reverse current” is large and therefore good for defibrillation. However, the tail end (shown in
red) of the “reverse current” is small and as discussed, has the potential to fibrillate the heart. i.e. the low
current at the tail end of the waveform has the potential to cause fibrillation in the heart that has just been
defibrillated by the earlier high current part of the waveform. Therefore, the defibrillator needs to avoid
giving the low current tail to the patient.
To avoid the problem we discussed, the defibrillator “cuts off” the harmful tail end of the biphasic
waveform. The “official” term used to describe this “cutting” is called “truncation”. Once the waveform tail
end is cut, it is called a “truncated waveform”. I will next explain how the waveform truncation done.

Let me show you how the defibrillator truncates the waveform. Let us commence by starting the “reverse
current” part of the waveform. As you know, this is achieved by switching “on” switches 2 and 3.
The defibrillator truncates the waveform in an interesting way. As the “reverse current” flows through
switches 2 and 3, the defibrillator truncates the current by also switching “on” switch 1. This immediately
creates a “short circuit” where the current from the capacitor returns back to it without going to the heart.
Because there is no further current going through the heart, the waveform returns to the baseline instantly
(see red line on waveform). I.e. the waveform has been truncated and the harmful tail end of the current
does not go to the heart.
During the “short circuit”, the remaining energy of the capacitor is dissipated in the form of heat in a
special part of the circuit, shown in red below. I will not discuss this heat dissipating part as it is somewhat
complicated.

Soon all the remaining energy of the capacitor is dissipated as heat. The capacitor will then have no
charge left in it and if another attempt at defibrillation is needed, the capacitor needs to be charged again
as described in the beginning of our discussion.
By the way, earlier in our discussions, I did mention the need for having the “switching interval” to prevent
short circuits. In that period of the biphasic waveform, a short circuit would be very much undesirable as it
would empty the capacitor prematurely leaving no energy left for the “reverse current” (see graph A
below). However, as you have seen, a short circuit is purposefully created by the defibrillator to truncate
the waveform at the end (graph B). So, at the correct place in the biphasic waveform, a “short circuit” is
useful !

Summary of biphasic defibrillator sequence

Let us at this point summarise the sequence of events that happen when we use the biphasic defibrillator:

 To provide the necessary large current for defibrillation, we need to charge the capacitor. The
charging system can use battery current or current from a wall socket.
 Charging switches A and B are switched “on” and the capacitor is charged using an high voltage
DC current. Once charged to the required level, switches A and B are switched “off”.
 For the “forward current”, switches 1 and 4 are switched “on”, and then switched off.
 After a brief “switching interval” where all switches are “off”, the “reverse current” is started by
switching “on” switches 2 and 3.
 The “reverse current” is then stopped (truncated) by also switching switch 1 “on”. Current now
flows directly back to the capacitor via switches 1 and 3 (short circuit) without going to heart. The
remaining charge in the capacitor is lost in the form of heat and the capacitor eventually has no
charge left.
 Delivery of the biphasic waveform is complete and all switches are put “off”.
 The capacitor is ready to be charged again for a repeat defibrillation attempt.
I have also summarised the switching sequence using the table shown below. You may find it easier to
understand.

However, please do not simply memorise the above table. Under exam stress, you may easily label the
switches in a different order and then the table will not make sense! So if you choose to use the table,
after writing it, please check to see if it makes sense.
“H bridge”

As you have seen, we used four switches to control the direction of current flow. In the world of
electronics, using four switches in this way to change the direction of current is very common and the
arrangement is called an “ H bridge”. The “H” in “H bridge” refers to the switches being drawn in circuit
diagrams in the form of an “H” (see green area in diagram). I am not sure why it is called a “bridge”, but I
like to think of it as a “bridge” that the current has to pass over.
In real life, the charging system and the internal switches of the defibrillator are electronically controlled
by a computer. The user selects the required energy and the computer controls the charging system to
charge the capacitor. When appropriate, the user presses the button to deliver the current to the patient.
The computer then controls the high speed electronic switches of the H bridge to deliver the biphasic
defibrillation current.
Throughout our discussion, we have “seen” the current flow as a biphasic waveform on an oscilloscope.
However, keep in mind that in real life, the defibrillation current flows via pads placed on the patients
chest. The biphasic current goes through the heart first in one direction and then changes direction to go
the other way. (Also in real life, the patient wouldn’t be smiling as shown in image below!)

By the way, the biphasic waveform that we have so far studied is “officially” called “biphasic truncated
exponential” waveform, often the name being shortened to “BTE” waveform.
The biphasic truncated exponential (BTE) waveform is not the only waveform that is currently in use.
There are other biphasic waveforms, such as one called “Rectilinear Biphasic Waveform” (RBW), which I
have shown below. Unfortunately, it is beyond the scope of this website to discuss which waveform is
medically more effective at defibrillating the heart. You can get information about various waveforms
elsewhere on the internet.

Automated External Defibrillator (AED)

The success of defibrillation is highly dependent on how soon after fibrillation that the defibrillation current
is given. Literally, every second counts.
In the past, if someone had ventricular fibrillation in a non hospital setting, there was a considerable delay
before an ambulance with trained medical personnel and a defibrillator would arrive. To reduce the delay
before defibrillation, there is a move to place defibrillators in public places, work places, schools, etc.
Such defibrillators are specially designed to be easy to use by the public and are called “public access
defibrillators”. They are also called “automatic external defibrillators” (AED).

If someone is suspected of having ventricular fibrillation (i.e. cardiac arrest), a member of the public can
take a nearby public access defibrillator and follow its instructions. The defibrillator comes with two pads
attached to wires. The instructions will show where to stick the two pads onto the patient. The AED, using
an electronic voice, will also guide the member of the pubic about what to do.

Once connected to the patient, the AED analyses the heart rhythm (it is a very clever piece of equipment)
and delivers a defibrillation current if it detects that the patient has ventricular fibrillation. I cannot go into
details of the various AED devices, but you can see plenty of videos on the internet showing how they are
used. Also, please find out, in places where you visit often, where the AEDs are located. Someday, that
information may save someone’s life !
As shown below, the basic design of the AED is very similar to the circuits we have discussed so far.
However, there are design differences between AEDs and hospital based defibrillators and it is beyond
the scope of this website to discuss them in detail. For example, AEDs have computer systems to analyse
and diagnose ventricular fibrillation. They also need circuits to produce the “voice” that guides the
member of public helping the patient. As the AED may be placed at all sorts of places, they often rely on
long lasting batteries and need to be made quite physically strong.

Implantable Cardioverter Device (ICD)

Finally I would like to briefly mention that there are devices called “implantable cardioverter devices”, or
more commonly known as “ICD”s. These are essentially like the AEDs we just talked about, except that
the device is made extremely compact. Some people have heart conditions that make them more likely to
get ventricular fibrillation and other life threatening rhythm problems in their heart. In these patients, one
option is to place an ICD inside the body and permanently connect it to the heart. The ICD continuously
analyses the heart rhythm and delivers a defibrillation current when necessary. Needless to say, putting
all that circuitry into something so tiny is an engineering miracle.
We have now reached the end of our discussion on how defibrillators work. I really hope you have