Beruflich Dokumente
Kultur Dokumente
1
Reagent Chemicals, 5th Edition
The Fifth Edition of REAGENT pan is attached. Place a cover on the some samples, particularly those having
CHEMICALSwas published in 1974. As coarser sieve, and shake the stack in a darker colors, a better color match may
stated a t the end of the Preface, two rotary horizontal direction and verti- be obtained by using 100-mL short-form
supplements will be issued in ANA- cally by tapping on a hard surface for Nessler tubes.
LYTICAL CHEMISTRY prior to publi- not less than 20 min, or until sifting is For the most accurate estimation of
cation of the next full edition, and re- practically complete. Weigh accurately color, it is desirable to use a comparator
prints sized to fit REAGENTCHEMI- the amount of material remaining on constructed so that white light is re-
CALS will be made available. The re- each sieve. flected from a white glass plate with
prints will incorporate any modifica- An alternate procedure may be used equal intensity through the longitudinal
tions in specifications and tests for in which the screening through the axis of the tubes being compared. The
which the American Chemical Society standard sieves is carried out in a me- tubes are shielded so that no light enters
Committee on Analytical Reagents has chanical sieve shaker. This shaker the tubes from the side. In most cases,
given early notification following its reproduces the circular and tapping satisfactory estimates of the color can be
meetings. Since the publication of the motion given to the testing sieves in obtained by holding the sample and
Fifth Edition, three such announce- hand sifting, but with a mechanical ac- standards close to each other over a
ments have appeared in ANALYTICAL tion, following the directions provided white plate.
CHEMISTRY,as follows: 47, No. 7,1149 by the manufacturer of the shaker. The best source of light is generally
(1975);48, No. 4,792 (1976);and 48,No. considered to be diffuse daylight.
9, 1432 (1976). General Procedures However, for general routine analyses,
the use of a titrating lamp equipped with
a “daylight” type fluorescent tube is
Interpretation of Measurement of Physical satisfactory.
Preparation of the APHA No. 500
Requirements Properties Platinum-Cobalt Standard. Measure
500 mL of reagent water into a 1000-mL
At the end of page 14 insert the volumetric flask, add 100 mL of hydro-
O n page 5 delete the paragraph o n chloric acid, and mix well. Weigh 1.245
Particle Size for Granular Materials. statement on Color ( A P H A ) [American
Public Health Association] cited g of potassium chloroplatinate
Substitute the statement cited below. (K2PtCld and 1.000 g of cobaltous
When a mesh or a mesh range is stat- below.
The color intensity of liquids may be chloride (CoClz.6HzO) to the nearest
ed on a reagent label, the label shall in- mg, and transfer to the flask. Swirl until
clude reference to a coarse sieve and to rapidly estimated by the use of plati-
num-cobalt standards, as described in complete dissolution is effected, dilute
a finer sieve. The sieve number (mesh) to the mark with reagent water, and mix
is related to the sieve opening as indi- ASTM standard method D 1209-69.
This method is particularly applicable thoroughly. The spectral absorbance of
cated in the accompanying table. this APHA No. 500 standard must fall
to those materials in which the color-
Sieve No., Sieve opening, producing bodies have light absorption within the limits given below, when
Mesh mm characteristics nearly identical with measured in a suitable spectrophotom-
those in the standards. Colors having eter, using a 1-cm light path and reagent
2 9.52 water as the reference liquid in a
4 4.76 hues other than light yellow or reddish
yellow cannot be determined by these matched cell.
8 2.38
10 2.00 standards.
20 0.84 The platinum-cobalt color standards Wavelength, nm Absorbance
30 0.59 contain carefully controlled amounts of
potassium chloroplatinate and cobaltous 430 0.110 to 0.120
40 0.42 455 0.130 to 0.145
50 0.297 chloride. Each platinum-cobalt color
unit is equivalent to 1 mg of platinum 480 0.105 to 0.120
60 0.250 510 0.055 to 0.065
80 0.177 per liter of solution (1 ppm), and the
standards are named accordingly. For
When tested according to the Proce- example, the No. 20 platinum-cobalt Preparation of the APHA Plati-
dure given below, at least 95% of the standard contains 20 ppm platinum. num-Cobalt Standards. Prepare the
material shall pass through the coarse These platinum-cobalt standards are required color standards by diluting the
sieve and at least 70% shall be retained also called APHA and Hazen stan- volume of APHA No. 500 platinum-
on the finer sieve. This requirement dards. cobalt standard given in the following
applies only to materials that are 60 Apparatus. The apparatus for this table with reagent water to a total vol-
mesh or coarser. measurement consists of a series of ume of 100 mL. The use of a buret is
Procedure. The sieves used in this Nessler tubes, a color comparator, and recommended in measuring the No. 500
procedure shall be those known as the a light source. These tubes must match standard. This series of standards is
U.S. Standard Sieve Series. For the de- each other with respect to the color of usually sufficient to permit an experi-
tails of the standardization of such the glass and height of the graduation enced analyst to make color compari-
sieves, reference may be made to ASTM mark. They must be fitted with suitable sons with the necessary precision and
E 11-70. closures to prevent loss of liquid by accuracy. If a more exact estimate of
Place 25 to 100 g of the material to be evaporation and contamination of the color is desired, additional standards
tested upon the appropriate coarser standards by dust or dirt. The most may be prepared to supplement those
standard sieve which is mounted above commonly used type of Nessler tube is given by using proportional amounts of
the finer sieve t o which a close-fitting the 100-mL tall-form tube. However, for the No. 500 platinum-cobalt standard.