Transport X

Portable Dissolved
Gas-in-Oil Analysis
(DGA) Instrument.
Transport X
• All gases measured together as a mixture.
• Extraction of gas from oil using standard extraction
technique;
– “Dynamic headspace” extraction - based on standard “static
headspace” gas-in-oil analysis technique.
• Sample analysis takes approximately 25 mins.
• Technique uses proven detection system (PGA).
• Very portable and rugged.
• No significant consumables required.

Donal Skelly Kelman Ltd


TRANSPORT X internal schematic

Donal Skelly Kelman Ltd


How does system work?
TRANSPORT X uses manually
taken 50ml syringe samples
On-line units automate sample
process
Gases then extracted from oil
sample using headspace
extraction – very reliable
technique
Then gas sample measured
using PAS module
Simple PAS Theory:
•Gases can be separated in frequency
domain
•Uses IR source and optical filters to
select particular freq. for each gas
•Strobed IR light directed into
measurement cell containing
extracted gas sample
•IR energy absorbed and released
from gas molecules creating pressure
waves (sound waves)
•Pressure waves measured by highly
sensitive microphones
•Sound signal proportional to
concentration of that gas in cell – this
is the Photo-acoustic Affect
Key advantages and differentiators
Rugged and suitable for life in
the substation
No Carrier gases or calibration
gases
Portable unit very simple to use
with results in 20 minutes
Can go from high to low gas
samples with no carry-over
Very high accuracy and
reliability
Minimal maintenance
requirements
TRANSPORT X Oil Sampling Procedure
• Sampling procedure is as laid out in Transport X operating
manual – Very similar to International Standard relating to
sampling of insulating oil into a syringe.
• 50ml sample collected into a glass syringe.
• Should be as close to 50ml as possible.
• Samples should be stored in cool place and shaded from
direct sunlight
• Sample injected directly from syringe into closed TRANSPORT
X glass bottle assembly.
Transport X Operation
• TRANSPORT X unit is designed to be very easy to operate
(compared to other DGA systems).
• Detailed procedures for all aspects of operation contained in
user manual.
• Unit is controlled via inbuilt touch screen.
• Step by step instructions given on-screen during standard
operation of TRANSPORT X. User simply has to follow the
numbered actions.
• Quick reference guide located on inside lid of Transport X unit.
Calibration
• No calibration required.
• Unit self-checks all critical functions every time a sample
analysis is initiated.
• Series of built in self-check limits will warn user if something
has gone wrong during the analysis.
How to be Sure!
• Optional cylinder of certified gas mixture for total system
check can be supplied – used at 3 month intervals it is enough
for approx 2 years of system checks.
Communications
• PC Sync to USB interface for downloading results to Delta X
compatible database.
– Unit can store up to 20,000 sample results in its
permanent memory.
• USB to USB interface for use of USB mouse or Keyboard.
• Supplied with PERCEPTION software
TRANSPORT X Results
TRANSPORT X – Portable DGA & Moisture
• 7 diagnostic gases & moisture (ppm & RS%)
• Rugged and field ready system
• No calibration requirements
• No carrier or calibration gases required
• Embedded PC with step by step instruction
• Easy to use

•Standard sampling practices


•Ability to test main tank, LTC, OCB etc
• No cross over effect with minimal cleaning
• Results on-site in minutes
• Embedded diagnostics
• TOA compatible
Introducing Product Range
Based on cutting edge technology Kelman developed DGA product
portfolio to suit wide range of requirements
• Portable DGA (TRANSPORT X)
• Low cost composite gas trend monitor (HYDRAN)
• On-line Full 8 Gas DGA (TRANSFIX)
• On-line 3 Gas DGA (MINITRANS)
• On-line DGA for main-tank and LTC (TAPTRANS)
• On-line DGA for large single phase per tank units (MULTITRANS)
• Software to download and analyse DGA result (PERCEPTION)
Kelman MINITRANS & HYDRAN

Cost effective online DGA monitoring Hydran M2 – Single Gas on-line sensor
system
•Continuous, on-line, intelligent moisture
Direct measurement of: and gas-in-oil detector
•Hydrogen H2 •Measures H2 (0-2000ppm) and water (0-
•Carbon Monoxide CO 100% RH)
•Acetylene C2H2 •Ease of installation – mount on flooded
•Moisture H20 port with 1 ½” NPT threads
M0150
The Intellix MO150 provides
comprehensive monitoring and interactive
transformer condition
diagnostics, to provide early warning of
incipient failure conditions.

• Winding hot-spot temperature model

• Moisture model which computes the


moisture content in the
thin conductor insulation and in the barrier
insulation

• MVA model which computes


the apparent power on
primary, secondary or tertiary
winding
Kelman TRANSPORT X

• Portable 7 Gas DGA + moisture


• Very simple to use
• Reliable and accurate results in minutes
• Embedded DGA diagnostics
• Designed for use in field
• No calibration or carrier gases
Kelman MULTITRANS & TRANSFIX

Multi tank Online DGA - monitors up to 3 On-Line DGA - 8 gases + moisture + load
separate single phase tanks

• 8 gases + moisture • No carrier or calibration gases


• No Consumables • Reliable gas extraction
• Extensive Alarms and Comms • Extensive Alarms and Comms
• Hourly sampling • Hourly sampling -real-time DGA
• Targeted at large, single phase per tank • Easy Install & Minimal maintenance
transformers requirements
Kelman TAPTRANS

On-Line DGA for Transformer & LTC


in one unit
• No cross-contamination
• 8 Gases + Moisture
• No consumables
• Extensive alarms and comms
• Hourly sampling
• Targeted at transmission
transformers with LTC’s
• No competitors can achieve this
functionality due to cross-
contamination

Unique Application for Arc


Furnace Transformer
Information is the key – not just
data
• One of the keys to successful monitoring is
being able to retrieve the data remotely
• Comprehensive comms options for modern
utilities
• The next step is converting that data in to
information for decision making
PERCEPTION – Asset Management Software
Data downloaded, stored, analysed and reported on
Tabs for assets and displays
TREND CHART - Smoothing
Data Table

Alarm – Red
Caution - Yellow
Duval’s Triangle

Historical Data
shown with
time-line
Key Gases with Probability
Gas Ratios – Graphical format

Dragable 360 degrees view of ratios

Das könnte Ihnen auch gefallen