ERRDUMP 3.20 - System Error Log Dump (c)1988-1998 Lunar Corporation


------------------------------------------------------------------------------

===============================================================================
Source File Mnemonic Description XInfo
===============================================================================

----[0184]----"dpxiq240 4.7e"-----------------------------07/02/07--10:03am----
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file CALABM00.F03
Low Energy corrections made: 41
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file CALABM00.F03
Low Energy corrections made: 41
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file CALABM00.F03
Low Energy corrections made: 41
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file CALABM00.F03
Low Energy corrections made: 41
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file CALABM00.F03
Low Energy corrections made: 41
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file MART7900.F03
Low Energy corrections made: 3
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file MART7900.F03
Low Energy corrections made: 3
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file MART7900.F03
Low Energy corrections made: 3
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file MART7900.F03
Low Energy corrections made: 3
High Energy corrections made: 0
dsf_read.c(1438) ERR generic error code
Corrected computed energies < 1 from file MART7900.F03
Low Energy corrections made: 3
High Energy corrections made: 0

----[0185]----"dpxiq240 4.7e"-----------------------------07/03/07--08:27am----
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 3.000 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=3.000.
SBC: +KV=38.000, -KV=38.000, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV=29.749, mA=3.805, PMT=150.9.
FEEDBACK: +KV=38.000, -KV=38.025, mA=3.060, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 3.000 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=3.000.
SBC: +KV=38.000, -KV=38.000, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV=29.358, mA=3.785, PMT=151.1.
FEEDBACK: +KV=37.793, -KV=37.769, mA=2.888, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2443) ER_DIAGNOSTIC diagnostic report 71
SUCCESS: mA RAMP to 3000 OK ON RETRY #3.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=38.000, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 5.029, mA=3.798, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.383, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=38.000, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.856, mA=3.798, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=38.013, mA=0.276, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2443) ER_DIAGNOSTIC diagnostic report 71
SUCCESS: mA RAMP to 750 OK ON RETRY #3.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 3.000 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=3.000.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV=28.296, mA=3.809, PMT=150.9.
FEEDBACK: +KV=37.915, -KV=37.927, mA=2.141, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 3.000 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=3.000.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV=28.711, mA=3.810, PMT=150.9.
FEEDBACK: +KV=38.000, -KV=37.964, mA=2.163, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 3.000 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=3.000.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV=26.855, mA=3.822, PMT=150.9.
FEEDBACK: +KV=37.927, -KV=37.805, mA=2.634, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 3.000 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=3.000.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV=26.892, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.769, -KV=37.891, mA=2.688, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_io.c(165) ER_DIAGNOSTIC diagnostic report 1
DPXReadSerial() calling ParseStatusPacket looking for NumBytes=
dsr_dpx.c(2528) ER_SERDRV serial driver call failure
dsr_dpx.c(2411) ER_DIAGNOSTIC diagnostic report 3000
Error Setting tube current
daq_acqr.c(916) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ROJA3200.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.026, mA=3.796, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.190, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2443) ER_DIAGNOSTIC diagnostic report 71
SUCCESS: mA RAMP to 750 OK ON RETRY #2.
dsr_dpx.c(1361) ER_DIAGNOSTIC diagnostic report 109
XRAY SOURCE FAILURE 'm':
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.750, PMT=523.4.
ACTUAL - PROGRAM: KV= 9.546, mA=3.807, PMT=150.9.
FEEDBACK: +KV=38.049, -KV=38.049, mA=0.957, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(766) ER_DIAGNOSTIC diagnostic report
daq_loop.c(792) ERR generic error code
daq_acqr.c(967) ERR generic error code
Error occurred during acquisition.
daq_acqr.c(1313) ERR generic error code
Bad scan file saved to LSTBAD01.F03
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ROJA3200.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 4.932, mA=3.809, PMT=150.9.
FEEDBACK: +KV=38.062, -KV=38.049, mA=0.659, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2443) ER_DIAGNOSTIC diagnostic report 71
SUCCESS: mA RAMP to 750 OK ON RETRY #2.
dsr_dpx.c(1361) ER_DIAGNOSTIC diagnostic report 109
XRAY SOURCE FAILURE 'm':
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.750, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.189, mA=3.799, PMT=151.1.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.620, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(766) ER_DIAGNOSTIC diagnostic report
daq_loop.c(792) ERR generic error code
daq_acqr.c(967) ERR generic error code
Error occurred during acquisition.
daq_acqr.c(1313) ERR generic error code
Bad scan file saved to LSTBAD01.F03
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ROJA3200.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.343, mA=3.794, PMT=149.7.
FEEDBACK: +KV=37.927, -KV=37.939, mA=0.134, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.660, mA=3.795, PMT=149.7.
FEEDBACK: +KV=37.964, -KV=37.988, mA=0.170, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.979, mA=3.785, PMT=149.7.
FEEDBACK: +KV=37.830, -KV=37.866, mA=0.648, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 5.444, mA=3.796, PMT=149.9.
FEEDBACK: +KV=37.964, -KV=37.988, mA=0.428, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #5-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 7.324, mA=3.810, PMT=149.7.
FEEDBACK: +KV=38.062, -KV=38.049, mA=0.807, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 109
FAILURE: UNABLE TO RAMP CURRENT TO 750, MAX RETRIES (#5) EXCEEDED.
daq_acqr.c(642) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ARANGA00.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.514, mA=3.795, PMT=150.6.
FEEDBACK: +KV=37.964, -KV=37.976, mA=0.150, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.836, mA=3.799, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.679, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2443) ER_DIAGNOSTIC diagnostic report 71
SUCCESS: mA RAMP to 750 OK ON RETRY #3.
dsr_dpx.c(1361) ER_DIAGNOSTIC diagnostic report 109
XRAY SOURCE FAILURE 'm':
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.750, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.478, mA=3.783, PMT=150.9.
FEEDBACK: +KV=38.110, -KV=38.110, mA=0.673, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(766) ER_DIAGNOSTIC diagnostic report
daq_loop.c(792) ERR generic error code
daq_acqr.c(967) ERR generic error code
Error occurred during acquisition.
daq_acqr.c(1313) ERR generic error code
Bad scan file saved to LSTBAD01.F03
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ARANGA00.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.575, mA=3.795, PMT=150.6.
FEEDBACK: +KV=37.964, -KV=37.976, mA=0.160, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 0.903, mA=3.795, PMT=150.6.
FEEDBACK: +KV=37.964, -KV=37.964, mA=0.071, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.148, mA=3.796, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=37.988, mA=0.210, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 7.471, mA=3.799, PMT=150.6.
FEEDBACK: +KV=38.000, -KV=38.049, mA=0.828, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2443) ER_DIAGNOSTIC diagnostic report 71
SUCCESS: mA RAMP to 750 OK ON RETRY #5.
dsr_dpx.c(1361) ER_DIAGNOSTIC diagnostic report 109
XRAY SOURCE FAILURE 'm':
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.750, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.284, mA=3.796, PMT=151.1.
FEEDBACK: +KV=37.964, -KV=37.964, mA=0.283, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(766) ER_DIAGNOSTIC diagnostic report
daq_loop.c(792) ERR generic error code
daq_acqr.c(967) ERR generic error code
Error occurred during acquisition.
daq_acqr.c(1313) ERR generic error code
Bad scan file saved to LSTBAD01.F03
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ARANGA00.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur

----[0186]----"dpxiq240 4.7e"-----------------------------07/03/07--09:30am----
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.722, mA=3.796, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=37.988, mA=0.175, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.002, mA=3.794, PMT=150.6.
FEEDBACK: +KV=37.927, -KV=37.927, mA=0.200, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.040, mA=3.810, PMT=150.6.
FEEDBACK: +KV=38.062, -KV=38.062, mA=0.610, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.978, mA=3.795, PMT=150.6.
FEEDBACK: +KV=37.964, -KV=37.988, mA=0.162, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #5-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.223, mA=3.796, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=37.988, mA=0.319, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 109
FAILURE: UNABLE TO RAMP CURRENT TO 750, MAX RETRIES (#5) EXCEEDED.
daq_acqr.c(642) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ARANGA00.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur

----[0187]----"dpxiq240 4.7e"-----------------------------07/03/07--09:37am----
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 4.797, mA=3.804, PMT=150.6.
FEEDBACK: +KV=38.025, -KV=38.037, mA=0.514, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2443) ER_DIAGNOSTIC diagnostic report 71
SUCCESS: mA RAMP to 750 OK ON RETRY #2.
dsr_dpx.c(1361) ER_DIAGNOSTIC diagnostic report 109
XRAY SOURCE FAILURE 'm':
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.750, PMT=523.4.
ACTUAL - PROGRAM: KV= 4.834, mA=3.796, PMT=150.9.
FEEDBACK: +KV=37.976, -KV=37.988, mA=0.482, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(766) ER_DIAGNOSTIC diagnostic report
daq_loop.c(792) ERR generic error code
daq_acqr.c(967) ERR generic error code
Error occurred during acquisition.
daq_acqr.c(1313) ERR generic error code
Bad scan file saved to LSTBAD01.F03
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_io.c(165) ER_DIAGNOSTIC diagnostic report 1
DPXReadSerial() calling ParseStatusPacket looking for NumBytes=
dsr_dpx.c(2003) ER_DIAGNOSTIC diagnostic report
daq_acqr.c(1392) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ARANGA00.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.124, mA=3.796, PMT=150.6.
FEEDBACK: +KV=37.964, -KV=37.988, mA=0.126, PMT=754.4.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.223, mA=3.796, PMT=150.6.
FEEDBACK: +KV=37.964, -KV=37.976, mA=0.229, PMT=754.4.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.002, mA=3.796, PMT=150.6.
FEEDBACK: +KV=37.964, -KV=37.988, mA=0.199, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.201, mA=3.798, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.486, PMT=754.4.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #5-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.113, mA=3.796, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=37.988, mA=0.305, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 109
FAILURE: UNABLE TO RAMP CURRENT TO 750, MAX RETRIES (#5) EXCEEDED.
daq_acqr.c(642) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ARANGA00.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.550, mA=3.795, PMT=150.6.
FEEDBACK: +KV=37.976, -KV=37.976, mA=0.156, PMT=755.6.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2443) ER_DIAGNOSTIC diagnostic report 71
SUCCESS: mA RAMP to 750 OK ON RETRY #2.
dsr_dpx.c(1361) ER_DIAGNOSTIC diagnostic report 109
XRAY SOURCE FAILURE 'm':
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.750, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.091, mA=3.796, PMT=150.9.
FEEDBACK: +KV=37.964, -KV=37.976, mA=0.603, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(766) ER_DIAGNOSTIC diagnostic report
daq_loop.c(792) ERR generic error code
daq_acqr.c(967) ERR generic error code
Error occurred during acquisition.
daq_acqr.c(1313) ERR generic error code
Bad scan file saved to LSTBAD01.F03
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ARAQUO01.F03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring Femur
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 0.940, mA=3.800, PMT=149.9.
FEEDBACK: +KV=38.074, -KV=38.086, mA=0.582, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 4.907, mA=3.799, PMT=149.9.
FEEDBACK: +KV=37.988, -KV=38.013, mA=0.461, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.672, mA=3.795, PMT=149.9.
FEEDBACK: +KV=37.964, -KV=37.976, mA=0.171, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.624, mA=3.795, PMT=149.9.
FEEDBACK: +KV=37.976, -KV=37.976, mA=0.159, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #5-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.799, mA=3.805, PMT=149.9.
FEEDBACK: +KV=38.013, -KV=38.025, mA=0.613, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 109
FAILURE: UNABLE TO RAMP CURRENT TO 750, MAX RETRIES (#5) EXCEEDED.
daq_acqr.c(642) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\ARAQUO01.S03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring AP Spine
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.111, mA=3.790, PMT=149.9.
FEEDBACK: +KV=37.866, -KV=37.854, mA=0.115, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 1.184, mA=3.795, PMT=149.9.
FEEDBACK: +KV=37.964, -KV=37.976, mA=0.122, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.783, mA=3.796, PMT=149.9.
FEEDBACK: +KV=37.964, -KV=38.000, mA=0.266, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 4.395, mA=3.796, PMT=149.9.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.441, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #5-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.002, mA=3.795, PMT=149.9.
FEEDBACK: +KV=37.976, -KV=37.988, mA=0.200, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 109
FAILURE: UNABLE TO RAMP CURRENT TO 750, MAX RETRIES (#5) EXCEEDED.
daq_acqr.c(642) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\MART8001.S03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring AP Spine
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 5.530, mA=3.798, PMT=149.9.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.549, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.894, mA=3.809, PMT=149.9.
FEEDBACK: +KV=38.098, -KV=38.098, mA=0.850, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 4.846, mA=3.805, PMT=149.9.
FEEDBACK: +KV=38.025, -KV=38.037, mA=0.497, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.210, mA=3.795, PMT=149.9.
FEEDBACK: +KV=37.952, -KV=37.952, mA=0.259, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #5-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.284, mA=3.796, PMT=149.9.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.343, PMT=750.7.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 109
FAILURE: UNABLE TO RAMP CURRENT TO 750, MAX RETRIES (#5) EXCEEDED.
daq_acqr.c(642) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\MART8001.S03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring AP Spine

----[0188]----"dpxiq240 4.7e"-----------------------------07/03/07--02:06pm----
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #1-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 2.319, mA=3.790, PMT=150.9.
FEEDBACK: +KV=37.854, -KV=37.854, mA=0.254, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #2-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.738, mA=3.796, PMT=150.9.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.360, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #3-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 5.139, mA=3.796, PMT=150.9.
FEEDBACK: +KV=37.964, -KV=38.000, mA=0.302, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #4-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 5.432, mA=3.798, PMT=150.9.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.443, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 109
FAILURE 'm': 0.750 mA RAMP ATTEMPT #5-
mA feedback out of acceptable range
EXPECTED - HOST: KV=76.000, mA=0.750.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 3.320, mA=3.796, PMT=150.9.
FEEDBACK: +KV=37.976, -KV=37.988, mA=0.250, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.050.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 109
FAILURE: UNABLE TO RAMP CURRENT TO 750, MAX RETRIES (#5) EXCEEDED.
daq_acqr.c(642) ERR generic error code
daq_main.c(521) ERR generic error code
Scan aborted due to error.
ddb_acq.c(490) ERR generic error code
Error scanning patient: c:\db_data2\MART8001.S03
ddb_main.c(547) ERR generic error code
[F1] Error Acquiring AP Spine

----[0189]----"dpxiq240 4.7e"-----------------------------07/03/07--02:08pm----

----[0190]----"dpxiq240 4.7e"-----------------------------07/03/07--02:08pm----

----[0191]----"DIAGS 4.7e"--------------------------------07/03/07--02:08pm----
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #1-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.152, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=37.988, mA=0.616, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.098.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #2-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.519, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=38.000, mA=0.654, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #3-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.787, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=37.988, mA=0.681, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #4-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 7.031, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=38.000, mA=0.706, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #5-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.592, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=37.988, mA=0.660, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 85
FAILURE: UNABLE TO RAMP CURRENT TO 150, MAX RETRIES (#5) EXCEEDED.
dst_menu.c(471) ER_DIAGNOSTIC diagnostic report
dsr_dpx.c(1361) ER_DIAGNOSTIC diagnostic report 109
XRAY SOURCE FAILURE 'm':
EXPECTED - HOST: KV=76.000, mA=-0.001.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.592, mA=3.796, PMT=150.9.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.660, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(766) ER_DIAGNOSTIC diagnostic report
dst_cntr.c(984) ER_DIAGNOSTIC diagnostic report
dst_cntr.c(871) ER_DIAGNOSTIC diagnostic report
dst_menu.c(861) ER_DIAGNOSTIC diagnostic report
serdrv.c(602) ERR generic error code -1
ser_write:ProtProcess
csr_port.c(440) ER_SERDRV serial driver call failure -1
WriteSerialPort:-1 of 1 bytes written
dsr_err.c(286) ER_DIAGNOSTIC diagnostic report 1
SCANNER ERROR: Scanner Detected emergency stop button or interlock engaged.
dsr_io.c(165) ER_DIAGNOSTIC diagnostic report 4
DPXReadSerial() calling ParseStatusPacket looking for NumBytes=
dsr_io.c(255) ER_SERDRV serial driver call failure 1
dsr_com.c(1308) ER_SERDRV serial driver call failure
Scanner refuses 'SetMotorRelay' command W. Enable='E', Disable='K'
dsr_dpx.c(656) ER_DIAGNOSTIC diagnostic report

----[0192]----"DIAGS 4.7e"--------------------------------07/03/07--02:09pm----
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #1-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.592, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=37.988, mA=0.660, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.098.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #2-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.836, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=38.000, mA=0.686, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #3-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.982, mA=3.796, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=38.000, mA=0.699, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #4-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.860, mA=3.796, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=38.000, mA=0.688, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(2671) ER_DIAGNOSTIC diagnostic report 85
DAC UNDERFLOW 'U': 0.150 mA RAMP ATTEMPT #5-
Underflow of mA pgm voltage w/o acceptable mA feedback
EXPECTED - HOST: KV=76.000, mA=0.150.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.714, mA=3.795, PMT=150.9.
FEEDBACK: +KV=37.952, -KV=38.000, mA=0.674, PMT=758.1.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(2453) ER_DIAGNOSTIC diagnostic report 85
FAILURE: UNABLE TO RAMP CURRENT TO 150, MAX RETRIES (#5) EXCEEDED.
dst_menu.c(471) ER_DIAGNOSTIC diagnostic report
dsr_dpx.c(1361) ER_DIAGNOSTIC diagnostic report 109
XRAY SOURCE FAILURE 'm':
EXPECTED - HOST: KV=76.000, mA=-0.001.
SBC: +KV=38.000, -KV=37.988, mA=0.100, PMT=523.4.
ACTUAL - PROGRAM: KV= 6.714, mA=3.798, PMT=150.9.
FEEDBACK: +KV=37.976, -KV=38.000, mA=0.673, PMT=756.8.
TOLERANCE - SBC: +KV= 0.244, -KV= 0.244, mA=0.024.
dsr_dpx.c(766) ER_DIAGNOSTIC diagnostic report
dst_cntr.c(984) ER_DIAGNOSTIC diagnostic report
dst_cntr.c(871) ER_DIAGNOSTIC diagnostic report
dst_menu.c(861) ER_DIAGNOSTIC diagnostic report

----[0193]----"dpxiq240 4.7e"-----------------------------07/03/07--02:10pm----
---------------------------------------------------------------(end of log)----