TABLE OF CONTENTS

THEORY OF OPERATION 1

OPERATIONAL TESTING 2

HARDWARE ACCESS 3

ADJUSTMENTS 4

PREVENTATIVE MAINTENANCE 5

TROUBLESHOOTING 6

DICOM OPTIONS 7

PARTS LIST 8

SYSTEM BLOCK DIAGRAMS Appendix A:

Conquest Imaging -3-


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging -4-


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™

Chapter 1
Theory of Operation

Conquest Imaging -5-


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging -6-


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ Theory of Operation

Operating System

The operating system used on the Host and the Signal and Image-Processing
(SIP) motherboards is Microsoft Windows XP Embedded (XP-E). The
embedded XP version allows customization of components and system
behavior. Much of the operating system is hidden from the user, as the
Philips applications use XP-E as a vehicle for implementing system behavior
through custom graphical user interfaces (GUI).

The operating system on the host is primarily used for system interaction
with the user or network and as an application platform for implementing the
iU22 operation and behavior.

The SIP operating system is configured for headless operation (no graphics
subsystem). The primary functions of the SIP operating system include:
 Process estimate data to native format data (scan converted but not
scaled to display size)
 Process estimate data flow components to sweeping display data and
digital Doppler audio
 Process estimate data to M-mode sweeping display data
 Implement the Cineloop function for save/recall/review

System Functional Architecture

The Applications layer comprises software to perform administrative


functions and miscellaneous non-imaging tasks necessary to operate the
ultrasound system. The Virtual Ultrasound System comprises all proprietary
technologies developed by Philips that are involved in ultrasound imaging
and the related physiological signals. These include hardware, firmware, and
software. The Platform layer comprises the hardware and infrastructure
required to run the applications and virtual ultrasound systems.

Conquest Imaging -7-


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Signal Flow

Ultrasound data (RF) is acquired from each transducer element and beam-
formed (summed) into a coherent data stream. This RF data stream is moved
over the RF bus between the Front-End Controller (FEC) and the Host
resident Dual Signal Conditioning (DSC) PCB.

The DSC processes the RF data to estimate space data, commonly referred to
as R-Theta data. This estimate data (2D, Doppler, and so on.) is sent to the
SIP Motherboard to be further processed into each final display data type
supported by the system (2D, 3D, 4D, Doppler, Audio, M-mode, and so on).
The SIP Motherboard stores these data types locally in the SIP hard drive
sub-system (Cineloop images).

From Cineloop images, this data is time-aligned by a streaming data manager


application and provided to Frame Composition processing for composite
image display and real-time annotation. Frame Composition uses the Host
AGP card to compose each image element into a composite display
containing image, text and graphical elements.

Conquest Imaging -8-


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU 22 Basic Block Diagram
HOST

Hard Drive


D V D

DVD drive

Modem PCI Bus - 64 Bit

PCI Bus - 32 Bit


CP/UIF

Dolphin
External
Network
AGP

PH: 209-942-2654
Dolphin

Motor Control

Rev3 070208.dms
FEC RFA
RFB

Conquest Imaging
DSC SIP IEP

FAX: 209-942-2572
AIM+

Ultrasound Sales Service Repair Training


1815 Industrial Drive, Suite 100 Stockton, CA 95206
OEMs
Hard
Drives

Scanhead Select
Channel USB
Boards
Acquisition Front Plane

Display

-9-
Acquisition Subsystem AVIO CAD AVIO RIP
Power Subsystem

Note: The power subsystem used on initially released systems is designated


for “A.x systems.” The subsequently released power subsystem is designated
for “B.0 systems.” Release A.x and B.0 systems are identified by a green or
blue LED, respectively, on the rear of the AC Tray.

Power Subsytem (A.x Systems)

The A.x power subsystem contains:


 Platform Power Supply (PPS)
 Platform Power Distribution Board (PPDB) • Acquisition Power
Supply (APS)
 AC Tray
 HV Switcher (used on A.x and B.0 systems)

Power enters the iU22 system via the AC receptacle, located on the AC Tray.
The AC Tray includes an EMI filter, circuit breaker, OEM isolation
transformer and OEM power relay. AC voltage is also routed to both custom
power supplies (Acquisition and Platform). The Power Subsystem is
controlled by the On/Standby switch, located next to the DVD drive. The
On/Standby switch, in conjunction with the Host Motherboard power control
circuitry generates a PS_ON# signal, which activates the two system power
supplies. The power supplies are interlocked, which controls voltage
application during power up.

The Acquisition and Platform supplies contain monitoring hardware, which


prevent sustained over- or under-voltage or over-current conditions. If one of
the supplies does not meet specification, the Power Subsystem shuts
down.This condition requires the AC mains to be disconnected for at least
one minute to discharge the power supply internal voltages and reset the
fault circuitry. There are no power supply adjustments. The power supplies
are factory set and sealed.

Conquest Imaging - 10 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
When the Power Subsystem is in low-power sleep mode, the power supplies
are shut down except for +5STBY, which is required to support motherboard
standby functions. When the system is in standby mode (suspend-to-RAM),
the Host software sets a battery enable register bit on the AVIO-CAD
Module and turns off both power supplies. In this state, the system is
consuming less than 7 W of energy and stays in this state indefinitely while
plugged into the AC mains. The battery enable register bit allows the battery
pack to provide energy to the Platform Power Supply +5STBY circuit if the
AC mains is disconnected.

The maximum duration of the mobile standby state is 4 hours. If the system
is unplugged for more than 4 hours and left in standby, a battery monitoring
circuit senses the battery discharge state and disconnects the battery from
the Platform Power Supply, preventing battery damage. By disconnecting
the battery, the Platform Power Supply can no longer supply +5STBY to
both motherboards, and the system enters a complete power-down state.
Subsequent iU22 power-up executes a cold boot cycle to obtain imaging
operation. Normally, if the system is still in a sus-pend state, the system will
boot much quicker as it is restoring the previous operating system state
saved in RAM. Typical boot times are greater than 2 minutes for a cold boot
and less than 35 seconds for a resume from standby.

The ACQ_DC_OK_n signal line indicates the operating condition of the


Acquisition Power Supply and allows for correct operation of the Platform
Power Supply if the Acquisition Power Supply is not inserted or is inserted
and functioning correctly. It also prevents a false-good indication if the
Acquisition Power Supply fails, including conditions where the input fuses
blow.

The DC_OK_OUT signal line is a reset signal to the system. It is dependent


on the power state of the Acquisition Power Supply and the Platform Power
Supply.

Conquest Imaging - 11 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Battery Charging

A fully-charged system battery should maintain the system in the standby


state for approximately four hours without the system being plugged in. If
the battery has been fully discharged, the system may not boot up until the
system has been plugged in for up to 20 minutes to get a sufficient charge on
the battery.

The system battery is charging when:


 The system is powered up and in use
 The system is shutdown, the circuit breaker is on, and the power cord
is connected to a power outlet.
 The system is in standby, circuit breaker is on, and the power cord is
connected to a power outlet. (During this state, standby is maintained
by current from the power outlet, not the battery.)
The system battery is being discharged when the system is in standby and if
the circuit breaker is off, or if the power cord is not connected to a power
outlet

USB Port Power

The Universal Serial Bus (USB) ports on the back of the DVD drive are
powered by the system battery if the system is off (power switch is off and
the breaker is on). The USB ports are powered by power from the
motherboard. The four USB ports on the rear of the AVIO-RIP and the
USB port for the control panel are powered by system power. If system
power is on, the ports are powered. If the system is off or in standby mode,
there is no power to the USB ports.

Conquest Imaging - 12 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Conquest Imaging - 13 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Conquest Imaging - 14 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Platform Power Supply

The Power Subsystem comprises the following hardware modules:


 Platform Power Supply (PPS)
 Platform Power Distribution Board
 Acquisition Power Supply (APS)
 AC Tray
 HV Switcher

Platform Power Supply Module

 Provides +3.3 Vdc, ±5.0 Vdc, ±12 Vdc, +12 VSB, and +5 VSB. The
+12 VSB is used as a source for the +5 VSB regulator and a source for the
battery charger, located on the AVIO-RIP. The +12 VSB and +5 VSB
are present when AC power is applied. (The iU22 system does not
need to be fully powered up. It needs only to be plugged into the
wall socket with the circuit breaker on.)
 Operates between 85 Vac and 264 Vac at 47 to 63 Hz.
 Supplies +12 Vdc to video monitor and +3.3 Vdc to touch screen
LCD and +5 Vdc to the touch screen Backlight Inverter.
 Thermal, over-voltage, and over-current protected: If temperature,
voltage, or current exceeds the normal range all voltages or currents
shut down. Cycle AC power to reset.
 Remote sense for +3.3 Vdc, +5 Vdc, and ground.
 Fused primary voltages. The fuses are located under the power supply
cover.
 Battery supplies 2.5 A for 5 ms as the current decreases during
power-down.
 Battery powers +5 VSB converter, if battery voltage is greater than
+5.88 V, and if the +12 VSB is less than 9 V. When the system is in
this state, the battery is not charged. Battery does not power +5 VSB
converter if +12 VSB is greater than 9 V, and the battery is charged
during this state.\
 Generates system reset signal using state of the Acquisition Power
Supply and Platform Power Supply.

Conquest Imaging - 15 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
 Platform Power Supply functions correctly if the Acquisition Supply
is not inserted or is inserted and functioning correctly. If something
occurs to take the Platform Power Supply down, the Acquisition
Supply also suspends operation. This may happen if the power supply
is working correctly and a voltage goes out on a system PCB.

Platform Power Distribution Board

 Mounts to the back wall of the Platform Card Cage and receives
voltages from the Platform Power Supply via bus bars and distributes
to connectors.
 Provides a low impedance connection for the Platform Power Supply
outputs.
 Contains voltage sense resistors from the Platform Power Supply for
+3.3 Vdc and +5 Vdc
 Provides test points and LEDs to indicate that the Platform Supply is
functioning correctly

Acquisition Power Supply Module

 Provides +3.5 Vdc, ±5.3 Vdc, ±15 Vdc, and +48 Vdc.
 Operates between 85 Vac and 264 Vac at 47 to 63 Hz
 Detects AC input voltages (100 to 120 Vac or 220 to 240 Vac).
 Thermal, over-voltage, and over-current protected: If temperature,
voltage, or current exceeds the normal range, all voltages or currents
shut down. Cycle AC power to reset.
 Remote sense for +3.5 Vdc, ±5.3 Vdc, and ground.
 Fused primary voltages: Fuses are under the power supply cover.
 If something occurs to take the Acquisition Power Supply down, the
Platform Power Supply also suspends operation. This may happen if the
Acquisition Power Supply is working fine and a voltage goes out on a
system PCB.
 Supplies +48 Vdc to the HV Switcher to operate the system front end.

Conquest Imaging - 16 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
AC Tray

 Filters input voltage and contains a circuit breaker


 Supplies line voltage (100, 120, 230, or 240 Vac) to the Platform
Power Supply and Acquisition Power Supply
 Supplies line voltage to the OEM transformer and OEM relay for OEM
operation.

HV Switcher

 Converts ±15 Vdc and +48 Vdc from the Acquisition Power Supply to
±HV (±30 to ±120Vdc)
 Supplies ±HV to the AIM HV regulators for low-pass filtering to reduce
power supply noise and to the Channel Boards through the Acquisition
Frontplane (for all transmit waveforms except CW)
 During CW, turns off power converters supplying ±HV bus and
switches the ±15 Vdc to the ±HV bus
 Routes HV to the transducer connector, to provide power for HV
mux devices in the transducer and to power pulsers of the matrix
array

Conquest Imaging - 17 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Conquest Imaging - 18 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Subsytems (B.0/C.0)

The B.0 power subsystem contains:


 Power supply assembly
o Platform Power Supply (PPS)
o Acquisition Power Supply (APS)
o Acquisition Power Distribution Board (APDB)
 Platform Power Distribution Board (PPDB)
 AC Tray
 HV Switcher (used on A.x, B.0 and C.0 systems)

Power Theory

Power enters the iU22 system via the AC receptacle, located on the AC
Tray. Power is routed through the circuit breaker, is filtered, and is sent to
the +12 VSB power supply. When the +12 VSB is within regulation, the
system awaits a signal from the operator to start system operation.

Pressing the On/Standby switch initiates a PS_ON_n signal at the Host/SIP


motherboards. The PS_ON_n signal is routed to the Acquisition Power
Distribution Board where it initiates the start-up logic sequence. The start-
up logic circuitry sends a signal to the AC Tray to close the AC relay,
applying power to the Acquisition Power Supply, the Platform Power Sup-
ply, and the OEM power supply. The Acquisition Power Distribution Board
also supplies +5 VSB to the AC Tray to light the blue LED. The +5 VSB is
also on when AC power is off but the battery power is on.
The OEM transformer may be configured for a 120- or 220-Vac input,
depending on whether the OEM transformer primary is wired in series or in
parallel. The OEM power supply in conjunction with the Platform Power
Distribution Board brings all output buses within regulation. Bus
monitoring occurs on the Platform Power Distribution Board and the
Acquisition Power Distribution Board.

Conquest Imaging - 19 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
When the monitoring circuits indicate that the buses are within regulation
voltage, a DC_OK_OUT signal is sent from the Acquisition Power
Distribution Board to the Host and SIP motherboards via the Platform
Power Distribution Board. When the motherboards receive this signal, they
initialize and bootup begins.

During system operation, all power buses are monitored by comparators on


the Platform Power Distribution Board. The comparators set the
DC_OK_OUT low if a fault occurs. At the same time, the circuitry on the
Acquisition Power Distribution Board identifies whether the shutdown
signal is coming from the Platform Power Distribution Board or from the
AC tray. The results are displayed with the SPS OK and PPDB OK LEDs
on the Acquisition Power Distribution Board. The LED status is latched
until AC power is removed.

Battery Charging

Same as A.x systems

USB Port Power

Same as A.x systems

Conquest Imaging - 20 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Conquest Imaging - 21 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Supply (B.0/C.0) Systems

The power supply assembly comprises the Platform Power Supply Module,
the Acquisition Power Supply Module, and the Acquisition Power
Distribution Board.

Platform Power Supply Module (PPS) (B.0/C.0 Systems)

 Supplies +12Vdc to the Platform Power Distribution Board for


conversion to +5 Vdc, +3.3 Vdc, and -12 Vdc.
 Operates between 90 and 240 Vac at 45 to 65 Hz.
 Thermal, over-voltage, and over-current protected: If
temperature, voltage, or current exceeds the normal range, all
voltages or currents shut down. Cycle AC power to reset.
 Fused primary voltages. The fuses are located under the power
supply cover.

Acquisition Power Supply Module (APS) (B.0/C.0 Systems)

 Provides +3.5 Vdc, ±5.3 Vdc, ±15 Vdc, and +48 Vdc.
 Operates between 85 and 264 Vac at 47 to 63 Hz.
 Thermal, over-voltage, and over-current protected: If temperature,
voltage, or current exceeds the normal range, all voltages or currents
shut down. Cycle AC power to reset.
 Remote sense for +3.5 Vdc, ±5.3 Vdc, and ground.
 Fused primary voltages: Fuses are under the power supply cover.
 If something occurs to take the Acquisition Power Supply down, the
Platform Power Supply also suspends operation. This may happen if
the Acquisition Power Supply is working fine and a voltage goes out
on a system PCB.
 Supplies +48 Vdc to the HV Switcher to operate the system front end.

Conquest Imaging - 22 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Acquisition Power Distribution Board (APDB) (B.0/C.0 Systems)

 Distributes power from the Acquisition Power Supply to the front end
 Uses the PS_ON_n signal from the Platform Power Distribution
Board to activate and hold the AC relay on
 Monitors +3.65 Vdc, +5.35 Vdc, -5.35 Vdc, +15 Vdc, -15 Vdc, and
+48 Vdc. Creates the APS_DC_OK signal from the monitored
voltages
 Uses the APS_DC_OK signal and the PPS_DC_OK signal (from the
Platform Power Supply) to create the SPS_OK signal.
 Uses the PLT_PWR_GOOD signal generated on the Platform Power
Distribution Board and the SPS_OK signal to generate the
DC_OK_OUT signal. The DC_OK_OUT signal prevents the Host
and the SIP motherboards from running until all buses are within the
proper voltage range.
 Provides fault detection during system operation. If a power supply
fault is detected, a latched-shutdown condition occurs, opening the
AC relay, disabling the power supplies and causing the AC tray LED
to blink. The latched-shutdown condition remains until +12 VSB is
inactive
 Provides a power-up time-out interval (approximately 2 to 3 seconds).
If the interval expires, a latched-shutdown condition occurs, opening
the AC relay, disabling the power supplies and causing the AC tray
LED to blink. The latched-shutdown condition remains as long as +12
VSB is active
 Enables the Acquisition Power Supply and Platform Power Supply at
system power-up (no delay)
 Monitors +5 VSB and illuminates the +5VSB_OK LED if the voltage
is within specifications
 Monitors system power-up, shutdown, and fault signals and generates
the appropriate system responses
 Twelve LEDs indicate power supply status. Six green LEDs indicate
normal operation. Six red LEDs indicate faults. Refer to Table 9-6 for
LED information
 +3.65 Vdc and +5.35 Vdc remote sense signals pass through the
Acquisition Power Distribution Board from the Acquisition Front
Plane to the Acquisition Power Supply

Conquest Imaging - 23 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Platform Power Distribution Board (PPDB) (B.0/C.0 Systems)

 Manages the power bus between the Platform Power Supply and
the Platform card cage • Converts +12 Vdc input to +5 Vdc,
+3.3 Vdc, and -12 Vdc
 Converts +12 VSB to SW_BATT or to + 5 VSB, depending on
whether the system is in normal operation or in standby
 Monitors status of buses and converters and combines the outputs
into one signal (PLT_POWER_GOOD) if the individual outputs
are within specifications
 Power-on command provided by PS_ON_n_HOST and
PS_ON_n_SIP signals to become PS_ON_n. This signal is then
sent to the APDB where it is used to activate the AC relay and
hold it on

AC Tray (B.0/C.0 Systems)

 Controls AC power to the Acquisition and Platform Power Supplies


by using a board-mounted AC relay. The relay is energized by logic
on the Acquisition Power Distribution Board. The relay controls AC
power to the Acquisition Power Supply, the Platform Power Supply,
and OEM input power.
 Configures winding of OEM AC transformer for 90 to 120 Vac, or
220 Vac input power (via two connectors). The AC Trays are sent
correctly configured from the factory.
 Provides EMI filtering on AC power lines
 Displays power status with a blue LED on the rear of the tray
 Generates and routes +12 VSB (standby power) to the Platform
Power Distribution Board
 Provides ground lugs for input power and customer equipment
connections
 Receives +5 VSB to light the LED from the Acquisition Power
Distribution Board

HV Switcher
Same a A.x

Conquest Imaging - 24 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Subsystems (D.0)

The D.0 power subsystem contains:


• Power Supply Assembly
- Platform Power Supply (PPS)
- Acquisition Power Distribution Board (APDB)
- Acquisition Power Supply (APS)
• Battery Thermal Management (BTM) board
• Signal and Power Distribution Board (SPD)
• AC Tray
• HV Switcher

Power Theory (D.0 Systems)

Power enters the iE33 system via the AC receptacle, located on the
AC Tray. Power is routed through the circuit breaker, is filtered, and is sent
to the +12 VSB power supply. When the +12 VSB is within regulation, the
system awaits a signal from the operator to start system operation.
Pressing the On/Off switch activates a timer that holds the AC relay
closed for about 4 seconds.
Pressing the On/Off switch initiates a PS-ON_n signal at the Unified
Motherboard (UMB). The PS_ON_n signal is routed to the Acquisition
Power Distribution Board (APDB) where it initiates the startup logic
sequence.
The APDB also supplies +5 VSB originating from the +12 VSB on
the Battery Thermal Management (BTM) board to the AC Tray to light the
blue LED. After internal sequencing, the UMB sends a PS_ON_n signal to
the Signal and Power Distribution board (SPD). The SPD sends inhibit
signals to allow the Acquisition and Platform Power Supplies (APS and
PPS) to power-up their DC outputs. The SPD also sends a signal to hold the
AC relay closed after the initial 4-second delay.
After the AC relay is energized, the APS and PPS are turned on by a
signal from the APDB. As each power supply comes up, window
comparators monitor the individual voltages (+48 V, ±15 V, ±5.35 V, and
+3.65 V). If the voltages are within range, each power supply develops a
DC_OK signal.

Conquest Imaging - 25 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
The two signals are combined (OR’d) and produce a single
DC_OK_OUT signal. The signal is sent to the UMB. When the UMB
receives the signal, it initializes and bootup begins. The DC_OK_OUT
signal is also used to set the next On/Off switch closure as a shutdown
command.
Once the system is operational, pressing the On/Off switch opens the
AC relay and removes power from the APS and PPS. While this is
occurring, the system continues to consume DC power, which ensures that
the charges on the power supplies diminish quickly according to
specifications.
Pressing the On/Off switch also activates a circuit on the SPD that
initiates a 20-second lockout of On/Off switch. This enables the system to
power up without interference from another switch depression (during the
power-up cycle, the user is not able to force a power-off state by pressing the
switch).
At system power-up, the Battery and Thermal Management (BTM)
board sets the Platform and Acquisition fan speeds at full speed. During
system operation, the BTM board continues to monitor the system
temperature and automatically adjust fan speed.

Conquest Imaging - 26 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Conquest Imaging - 27 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Battery Charging Theory (D.0 Systems)

For systems with D.0 hardware, the system batteries have been
disconnected and the battery charging circuits are not operational.

USB Port Power Theory (D.0 Systems)

The Universal Serial Bus (USB) ports on the back of the DVD drive
and on the UAVIO are powered by system power from the UMB. If system
power is on, the ports are powered. If the system is off, there is no power to
the USB ports.

Signal and Power Distribution Board (SPD) (D.0 Systems)

• Provides On/Off switch conditioning. Disables the switch for


approximately 20 seconds after first depression, unless power is removed
(Figure 4-3)
• Provides an interface between the UMB and the rest of the system.
Cables to the SPD are designed to enable the UMB to slide out of the
system for maintenance
• Contains a remote On/Off switch for FSE use during system
maintenance
• Receives LVDS signals from the UMB, re-clocks the signals, and sends
them to the UAVIO
• Generates the platform power good signal (PLT_POWER_GOOD)
when all power supplies on the board are within regulation
• Provides a +5.01 V reference voltage to voltage comparators on the
SPD
• Regulates the +5 Vdc and +3.3 Vdc
• Uses the +5 Vdc and +3.3 Vdc to monitor the +12 Vdc.
• Provides overvoltage shutdown protection for the +5 Vdc and +3.3 Vdc
regulators. Latches the regulators off if the voltage is out of spec
• Provides -12 Vdc to track the +12 Vdc

Conquest Imaging - 28 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Battery Thermal Management (BTM) (D.0 Systems Only)

Provides Platform Card Cage and Acquisition Card Cage fan tachometer
readings to system software for fault isolation purposes. Tachometer
readings are used only for fault isolation. (They are not used for speed
control.)
• Sets platform and acquisition fans at full speed during system power-up
• Provides an I2C bus interface to send BTM board identification data to
the rest of the system

AC Tray (D.0 Systems)

• Displays power status with a blue LED on the rear of the tray.
• Controls AC power to the Acquisition and Platform Power Supplies by
using a board-mounted AC relay. The relay is energized by logic on the
Acquisition Power Distribution Board. The relay controls AC power to
the Acquisition Power Supply, the Platform Power Supply, and OEM
input power.
• Configures winding of OEM AC transformer for 90- to 127-Vac, or
220-Vac input power (via two connectors). The AC Trays are sent
correctly configured from the factory.
• Provides EMI filtering on AC power lines.
• Generates and routes +12 VSB (standby power) to the Signal and
Power Distribution Board (SPD).
• Provides ground lugs for input power and customer equipment
connections.
• Receives +5 VSB to light the LED from the Acquisition Power
Distribution Board.

HV Switcher (D.0 Systems Only)

• Converts ±15 Vdc and +48 Vdc from the Acquisition Power Supply to
±HV (±30 to ±120 Vdc)
• Supplies ±HV to the NAIM HV regulators for low-pass filtering to
reduce power supply noise and to the Channel Boards through the
Acquisition Frontplane (for all transmit waveforms except CW)
• During CW, turns off power converters supplying ±HV bus and
switches the ±15 Vdc to the ±HV bus

Conquest Imaging - 29 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
• Routes HV to the transducer connector, to provide power for HV MUX
devices in the transducer and to power pulsers of the matrix array.

Conquest Imaging - 30 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Subsystems (E.0)

The E.0 power subsystem contains:


• Power Supply Assembly
- Platform Power Supply (PPS)
- Power System and Battery Controller (PSBC) Board
- Acquisition Power Supply (APS)
• Signal and Power Distribution Board (SPD)
• AC Tray

Power Theory (E.0 Systems)

Power enters the iE33 system via the AC receptacle, located on the
AC Tray. Power is routed through the circuit breaker, is filtered, and is sent
to the +12 VSB power supply. When the +12 VSB is within regulation, the
system awaits a signal from the operator to start system operation. Pressing
the On/Off switch activates a timer that holds the AC relay closed for about
4 seconds.
Once the Acquisition and Platform Power Supplies have charged their
internal high voltage bus, each send an AC_OK signal to the Power System
and Battery Controller (PSBC). The PSBC then sends a pulsed signal to the
UMB. (This pulsed signal replaces the momentary low-pulse signal sent
directly to the mother board from the On/Off switch).
After internal sequencing, the UMB motherboard sends a PS_ON_n
signal to the PSBC. The PSBC sends inhibit signals to allow the Acquisition
and Platform power supplies to power up their DC outputs. The PSBC also
sends a signal to hold the AC relay closed after the initial 4-second delay.
The power supplies then bring all buses up to the proper specifications
and send the DC_OK_OUT signal to the UMB. When the UMB receives
this signal, it initializes and bootup begins. The DC_OK_OUT signal is also
used to set the next On/Off switch closure as a shutdown command.
When the system starts from a standby state, AC power is applied to
the standby power supply in the AC Tray, however, the AC relay has
opened, removing power from the PPS and APS. This places the system in
the initial power-up sequence at the point when the +12 VSB powers up.

Conquest Imaging - 31 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
When the On/Off switch is pressed, the system power-up sequence
continues as described. Once the system is operational, pressing the On/Off
switch opens the AC relay and removes power from the PPS and APS.
While this is occurring, the system continues to consume DC power, which
ensures that the charges on the power supplies diminish quickly according to
specifications.

Battery Charging Theory (E.0 Systems)

For systems with E.0 hardware, the batteries have been removed. The
battery charging circuits are not functional.

USB Port Power Theory (E.0 Systems)

See “USB Port Power Theory (D.0 Systems)”.

Signal and Power Distribution Board (SPD) (E.0 Systems)

• Provides On/Off switch conditioning. Disables the switch for


approximately 20 seconds after first depression, unless power is
removed.
• Provides an interface between the UMB and the rest of the system.
Cables to the SPD are designed to enable the UMB to be slid out of the
system for maintenance.
• Contains a remote On/Off switch for FSE use during system
maintenance.
• Receives LVDS signals from the UMB, re-clocks the signals, and sends
them to the UAVIO.
• Generates the platform power good signal (PLT_POWER_GOOD)
when all power supplies on the board are within regulation.
• Provides a +5.01 V reference voltage to voltage comparators on the
SPD.
• Regulates the +5 Vdc and +3.3 Vdc.
• Uses the +5 Vdc and +3.3 Vdc to monitor the +12 Vdc.
• Provides overvoltage shutdown protection for the +5 Vdc and +3.3 Vdc
regulators. Latches the regulators off if the voltage is out of spec.
• Provides -12 Vdc to track the +12 Vdc.

Conquest Imaging - 32 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power System and Battery Controller (PSBC) (E.0 Systems)

• Combines the functions of the Acquisition Power Distribution Board


(APDB), the Battery Thermal Management (BTM) board, and the HV
Switcher.
• Routes +3.65 Vdc and ±5.3 Vdc directly to the Acquisition Front Plane
(AFP) from the Acquisition Power Supply (APS).
• Sets platform and acquisition fans at full speed during system power-up.
• Provides dual platform and Acquisition Card Cage fan tachometer
readings to system software for fault isolation purposes. Tachometer
readings are used only for fault isolation. (They are not used for speed
control.)
• Provides an I2C bus interface to send fan tachometer readings and
PSBC board identification data to the rest of the system.
• Provides out-of-specification levels for overvoltages or undervoltages
or overcurrent to cause system shutdown and latch off.
• Provides an interface to the Signal and Power Distribution Module
(SPD).
• Converts ±15 Vdc and +48 Vdc from the Acquisition Power Supply to
±HV (±30 to ±120 Vdc).
• Supplies ±HV to the NAIM HV regulators for low-pass filtering to
reduce power supply noise and to the Channel Boards through the
Acquisition Frontplane (for all transmit waveforms except CW).
• During CW, turns off power converters supplying ±HV bus and
switches the ±15 Vdc to the ±HV bus.
• Routes HV to the transducer connector, to provide power for HV MUX
devices in the transducer and to power pulsers of the matrix array.

AC Tray (E.0 Systems)

• Displays power status with a blue LED on the rear of the tray.
• Controls AC power to the Acquisition and Platform Power Supplies by
using a board-mounted AC relay. The relay is energized by logic on the
Power System and Battery Controller (PSBC). The relay controls AC
power to the Acquisition Power Supply, the Platform Power Supply, and
OEM input power.
• Configures winding of OEM AC transformer for 90- to 127-Vac, or
220-Vac input power (via two connectors). The AC Trays are sent
correctly configured from the factory.
Conquest Imaging - 33 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
• Provides EMI filtering on AC power lines.
• Generates and routes +12 VSB (standby power) to the SPD.
• Provides ground lugs for input power and customer equipment
connections.
• Receives +5 VSB to light the LED from the SPD.

HV Switcher

HV Switcher functions for E.0 or later systems are combined on the Power
System and Battery Controller (PSBC).

The control subsystem comprises:


• Control Panel Module (CPM)
• Two Control Panel Connector Board Interfaces
• Control Panel User Interface (CPUI)
• TGC/LGC Board
• Two Backlight Inverter Boards
• Two 8.4-inch LCD panels with integrated touch screens
• Disk drives (DVD and three Host hard drives: HDD0, HDD1, and
HDD2)
• Bluetooth transceiver (wireless RF control using voice recognition)

Control Panel Module (CPM)

• Supports the following user-input devices:


- USB QWERTY keyboard
- USB trackball
- Dual touch screen graphics overlays
- TGC/LGC slide pots
- Rotary encoders
- Push buttons
- Spare USB connector

• Supports the following user-output devices:


- Touch screen monitor graphics overlays
- LED status indicators
- Keyboard illumination LEDs
- Piezoelectric beeper

Conquest Imaging - 34 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
• Detects user inputs, communicates with the system CPU via a USB
port, and displays outputs
• Provides a 7-port USB hub
• Analog to digital conversion of the TGC/LGC signals
• 24-MHz clock provides reference frequency for a clock synthesizer that
drives various frequencies to the clocked devices within the control panel
(microcontroller, USB hub, and so on)
• CPM provides power and signal interface to CPUI
• Uses +5 Vdc from the Platform Power Supply
• Identifies revision levels of CPM and CPUI modules

Control Panel Connector Board Interface (Two Boards)

• Provides LVDS (low-voltage differential signaling) from the UAVIO to


the LCD display
• Connects the CPM to the Backlight Inverter
• Provides an interface for a potentiometer controlling the Backlight
Inverter (CCFL or cold-cathode fluorescent lamp)
• Provides interface to Backlight Pot
• Provides an interface from the CPM to the beeper
• Provides interface to the QWERTY control panel LEDs
• Connects touch screen sense lines to the CPM
• Receives signals from the CPM for the following:
- Four touch-screen signals to the LCD
- Four ground signals
- Two +5 Vdc signals
- Two +3.3 Vdc signals
- Two beeper signals
- Potentiometer brightness control
- Sleep control to the LCD backlight

Control Panel User Interface (CPUI)

Contains the rotary encoder and push-button switch interface to CPM

TGC/LGC Board

• Contains eight Time Gain Compensation and eight Lateral Gain


Compensation slide pots with LEDs.
Conquest Imaging - 35 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
• The 16 slide pot signals are multiplexed to eight analog signals
representing the position of each slide pot location.
• Multiplexed slide pot signals are sent to the CPM for processing.

Backlight Inverter (Two Each)

• Connected to the Control Panel Connector Board.


• 600 Vac powers the touch screen “backlight.” High voltage is present
across the lower end of the transformer.
• Voltage is bled off immediately when power is removed. There are no
components to store power even if the backlight is defective or not
connected.
• Sleep command from the system holds off backlight until a
predetermined point in the bootup cycle.

NOTE: Voltages of 600 Vac or higher are present on the lower end of
the Backlight Inverter when system
power is applied. To avoid shock hazard, do not touch the inverter.

Touch Screens (Two)

• Connected to the Control Panel Connector Board.


• Thin, pressure-resistive, transparent film overlays the 800 x 600 LCD
display mounted to the Control Panel Module. That film is scanned by a
USB chip set that sends screen coordinates to the Host operating system
when the user touches the overlay. The coordinates represent the X, Y
location of the pressure point that is registered to a displayed button.
• User selection of imaging parameters and control settings. Provides a
virtual button press, using a graphical interface.
• The touch screen lights are powered by the Backlight Inverter.

Conquest Imaging - 36 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Control System

The control subsystem comprises:

 Control Panel Module (CPM)


 Control Panel User Interface (CPUI)
 Bluetooth device (wireless RF control using voice recognition)
 Disk drives (DVD, Host Hard Drive, and two SIP Hard Drives)

Control Panel Module (CPM)

 24-MHz clock provides reference frequency for a clock synthesizer that


drives various frequencies to the clocked devices within the control panel
(microcontroller, USB hub, and so on).
 CPM provides power and signal interface to CPUI.
 Host Motherboard USB routes directly to CPM for expanding to seven-
port USB hub.
 USB interface to QWERTY keyboard, touch screen, trackball, USB port,
and CPU interface.
 Uses +5 Vdc from the Platform Power Supply.
 Identifies revision levels of CPM and CPUI modules.
 Beeper circuit provides “beeper” and “click” noises.
 Provides interface to Control Panel Connector Board and Backlight
Inverter.
 Provides LVDS (low-voltage differential signaling) from the AVIO to the
LCD display
 Connects the CPM to the Backlight Inverter
 Provides an interface for a potentiometer controlling the Backlight
Inverter (CCFL or cold-cathode fluorescent lamp).
 Provides interface to Backlight Pot
 Provides an interface from the CPM to the beeper Provides interface to
the QWERTY control panel LEDs
 Connects touch screen sense lines to the CPM

Conquest Imaging - 37 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
 Receives signals from the CPM for the following:
o Four touch-screen signals to the LCD
o Four ground signals
o Two +5 Vdc signals
o Two +3.3 Vdc signals
o Two beeper signals
o Potentiometer brightness control
o Sleep control to the LCD backlight

Control Panel User Interface (CPUI)

Contains the slide pot, rotary encoder, push-button, and multi-position switch
interface to CPM
 Connected to the Control Panel Connector Board.
 600 Vac powers the touch screen “backlight.” High voltage is present
across the lower end of the transformer.
 Voltage is bled off immediately when power is removed. There are no
components to store power even if the backlight is defective or not
connected.
 Sleep command from the system holds off backlight until a
predetermined point in the bootup cycle.

Caution: Shock Hazard – Voltages greater the 600 volt present.

 Connected to the Control Panel Connector Board.


 Thin, pressure-resistive, transparent film overlays the 800 x 600 LCD
display mounted to the Control Panel Module. That film is scanned by
a USB chip set that sends screen coordinates to the Host operating
system when the user touches the overlay. The coordinates represent
the X, Y location of the pressure point that is registered to a displayed
button.
 User selection of imaging parameters and control settings. Provides a
virtual button press using a graphical interface.
 The touch screen light is powered by the backlight inverter.

Conquest Imaging - 38 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Disk Drives

 The DVD drive is used to upgrade system software and transducer


files (transducer-specific AP&I data) and may also be used to store
images.
 Two hard drives for the SIP Motherboard: Each drive has one high-
speed partition (on the outer edge of the platter) and one or more
operating system partitions. One of the two drives contains the
operating system boot image and is used to boot the SIP application.
Both drives are involved in streaming-data operations and use the
high-speed partition to provide about 130 GB of high-bandwidth
storage. Refer to for drive partition information.
 One hard drive for the Host Motherboard: Contains the ultrasound
system operating system, embedded operating system, system
software, and transducer information.

Host Hard Drive Partitioning

Drive Partition by Function

C Windows XP Embedded operating system


D Read/write thumbnails and associated data
E Read/write other data, including Quick Save presets, logs, and so on
F Read/write installed Philips application software

Bluetooth Transceiver

The Bluetooth transceiver enables wireless remote control using RF


technology to enable voice control of the ultrasound system. The transceiver
may be used with or without a headset. Refer to the operating instructions
for more information.

Acquisition Subsystem

The Acquisition Subsystem controls the transducer, generates the electrical


impulse (pulse) that creates sound waves, and converts the returning echo
information into digital data.

Conquest Imaging - 39 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
The Acquisition Subsystem consists of:
 Scanhead Select Module (S/HSEL)
 Front End Controller Module (FEC)
 Acquisition Frontplane (AcqF)
 Four Channel Boards (CB0 through CB3)
 Analog Interface Module Plus (AIM+)
 Motor Controller PCB (daughterboard to FEC)

Scanhead Select Module - S/HSEL

 Selects between transducer connectors


 Interface between the front-end PCBs and the transducer
 Provides the signal path for TGC1 and TGC2 voltages from the
AIM+ to the Channel Boards
 Provides the signal path for CW Doppler I and Q signals from the
Channel Boards to the AIM+
 Connects transducer personality information from the transducer to
the AIM+
 Provides relay status signals to the AIM+, indicating the status of the
multiplexers
 3X or 4X multiplexers for the 128 Channel Board transceivers to the
transducer connectors
 CW Doppler multiplexer for connecting transmit signal from AIM+ to
CW/PW connector
 4X multiplexer for array data, clock, control signals, and power
signals from the AIM+ to transducer connectors or to no connection if
no transducer connector is selected
 3X multiplexer for transducer temperature sensing lines to AIM+ and
sensors for transducer 3D position data and plane selection data.
 3X multiplexer for Motor Controller signals from the AIM+ to the
transducer
 Signal path from CW/PW connector to the AIM+ for CW Doppler
personality signals
 Decoupling filters for all power supply input buses
 Transducer-disconnect interrupt signal to the AIM+ from each
transducer connector
 I2C interface and flash EEPROM for board-level identification
 Provides a path from the AIM+ for VX1, VX2, VX3+, VX3-, HV+,
Conquest Imaging - 40 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
HV-, and DC_BIAS1 voltages to the transducers. Uses VRELAY (+12
V), and SS_+5V power buses.

Front End Controller (FEC)

 Uses 40-MHz/32-MHz differential clocks and 8-MHz sync clock


(all derived from AIM+)
 Contains a slot for the Galil Motor Controller PCB for 3D
acquisition
 Programs Motor Controller with a motion profile for speed and
position control
 AP&I power monitor control through FEP bus and utilized by
FEC processor
 Monitors system temperature from sensors located on several
modules in the card cage
 Physio functions
o Contains jacks for internal ECG input, external or high-
level ECG input, Pulse/Aux1 input, Phono/Aux2 input, and
ECG analog out or ECG trigger out
o Processes respiratory, pulse, phono, and auxiliary (from
third-party equipment)
o Simultaneous acquisition and processing of four channels
of physio data
o Data managed by FEC CPU
o Data sent to AVIO for storage in the Host CPU memory
 FEC processor is Motorolla 603a CPU processor
 Responsible for event timing signal set to control the beamformer
signal path for each PRI
 Sends control data from FEC to the Channel Board, AIM+, and
S/HSEL
 Sends TGC data from FEC to the TGC ASIC on the AIM+
 Accepts real-time RF data from Channel Boards on SUM bus A
and SUM bus B
 FEC is I2C bus master for entire Acquisition Subsystem
 RF pre-processing: RF headers, low-pass filter, and line
composition buffer to reorder output lines to the DSC
 Contains serial LVDS host interface to the AVIO

Conquest Imaging - 41 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
 Configures the AIM+ FPGA using the FEP bus (FPGA is a Field
Programmable Gate Array: ageneral purpose chip which can be
programmed to carry out a specific hardware function.)
 Sequencing in Krusty FPGA and controls beamformer operation
 RF sequencer responsible for controlling RF processor (Tigris
FPGA)
 Drives the RF header bus and control signals to Tigris FPGA
 Signal path mode switching

Acquisition Frontplane (AqF)

 Provides connections and signal paths for the AIM+, FEC, and the
Channel Boards
 Distributes power supply buses and clock signals to the AIM+, FEC,
and Channel Boards • Provides data bus, address bus, and clock
termination networks
 Contains the Acquisition Control Bus, which is the interface to the
Platform Card Cage

Channel Boards (CB) iU22

 4 Channel Boards (A0 through A3), 4 Nile ASICs per board with
16 DAC transmit each. Only 8 DAC are used for each board.
 Nile ASICs contain a transmit burst controller, waveform
generator, waveform weighting, and waveform pre-distortion.
 Analog signal path D/A conversion and external power
amplification.
 32 receive channels per board connect to the FEC bus, TIM bus,
and the clocks.
 Bus interfaces: Sum A, Sum B, TIM, FEP, FEC.

Conquest Imaging - 42 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Channel Boards (CB) iE33

• Four Channel Boards (A0 through A3), four Nile ASICs per board with
16 DAC transmit each. Only eight DAC are used for each board.
• Channel Board transmit functions are not used for Matrix transducers
• Nile ASICs contain a transmit burst controller, waveform generator,
waveform weighting, and waveform pre-distortion.
• Analog signal path D/A conversion and external power amplification.
• 32 receive channels per board connect to the FEC bus, TIM bus, and the
clocks.
• Bus interfaces: Sum A, Sum B, TIM, FEP, FEC.
• CW Doppler processing (array current summed and pencil probe)
• CW Pulser waveform generator and pulser for static CW

Analog Interface Module Plus (AIM+) iU22

 Monitors system temperature, voltage, and beamformer pulser power


consumption Controls Scanhead Select Module interface (transducer
select, personality retrieval, and connectivity status)
 Generates TGC1 and TGC2 waveforms
 CW Doppler acquisition, including pencil probe
 Acquisition Clock sources
 Regulates HV Pulser
 HV MUX current limited source voltages
 Regulates acquisition analog supply
 Pulser power monitor source
 High Speed Serial Interface to Scanhead Select Module
 TGC and HV MUX data demultiplexing and signal conversion
 Contains 320 MHz master oscillator and generates clock signals for
all Acquisition hardware (64 MHz, 40 MHz, 32 MHz and Link_Clk)
distributed as differential PECL. Clock_Sync is distributed as
LVCMOS
 A/D data acquisition block for measuring transducer mounted sensors,
acquisition system voltages, pulser supply voltage and current and
system temperature transducers
 Power supply regulators and current limits for all voltages supplied to
the transducer array connectors (±12 Vdc, +5 Vdc, +REF, and ±HV
MUX bias)
Conquest Imaging - 43 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
 Transducer personality data identifies which transducer(s) are
connected to the system Pulser regulators (±20-80 Vdc, ±60-120
Vdc) and LV CW regulator with disable support
 CW Doppler Pulser waveform generator
 CW Doppler processing (array current summed and pencil probe)
 Local regulation of ±15 V to ±12 V for transducer use
 Local generation of +10 V reference for TGC waveform D/As and
transducer use
 Passive motor control path for future 3D position motor control
 Interface between FEC and the Scanhead Select Module
 Supplies the TGC voltages to the Channel Boards
 Regulates the front-end pulser voltages
 Serves as the pulser for static CW Doppler
 Performs the analog process for CW Doppler

Neo Analog Interface Module (NAIM) iE33

• Monitors system temperature, voltage, and beamformer pulser power


consumption
• Controls Scanhead Select interface (transducer select, personality
retrieval, and connectivity status)
• Transducer personality data identifies which transducer(s) are
connected to the system
• Generates TGC1 and TGC2 waveforms
• Supplies TGC voltages to the Channel Boards for non-Matrix
transducers or to the beamformer resident in Matrix transducers
• Pulser regulators (±20-80 Vdc, ±60-120 Vdc) and LV CW regulator
with disable support
• HV MUX current limited source voltages
• Regulates acquisition analog supply
• Pulser power monitor source
• High Speed Serial Interface to Scanhead Select
• TGC and HVMUX data demultiplexing and signal conversion
• Contains 320 MHz master oscillator and generates clock signals for all
Acquisition hardware (64 MHz, 40 MHz, 32 MHz and Link_Clk)
distributed as differential PECL. Clock_Sync is distributed as
LVCMOS

Conquest Imaging - 44 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
• A/D data acquisition block for measuring transducer mounted sensors,
acquisition system voltages, pulser supply voltage and current and
system temperature transducers
• Power supply regulators and current limits for all voltages supplied to
the transducer array connectors (±12 Vdc, +5 Vdc, +REF, and ±HV
MUX bias)
• Local regulation of ±15 V to ±12 V for transducer use
• Local generation of +10 V reference for TGC waveform D/As and
transducer use
• Interface between FEC and the Scanhead Select
• Regulates the front-end pulser voltages
• Contains Matrix transducer functions including data interface for
microbeamformer, HV MUX (transmit power) control, and TGC
control

Galil Motor Controller PCB

 Motor Controller PCB located on the FEC


 Interfaces with the FEC CPU to set up motion profiles and to pass
parameters. Initiates motion and passes location results via the
Krusty FPGA on FEC
 Controls the speed and location of motorized probes and reports
information to the system in real-time
 Calibrates after transducer connection
 Monitors and limits the power to the motor for safety reasons
 Constant velocity or stepped acquisition
o During constant velocity mode, the FEC programs the Motor
Controller with a motion profile consisting of a speed and a
range over which this speed should be maintained.
o During stepped acquisition, the Motor Controller sends the
transducer to a position, fires a few frames, and waits to send
the transducer to a different location. The Motor Controller
holds the position until told to move elsewhere.
 System level identification of the module (firmware and software
version) provided through the RS-232 interface

Conquest Imaging - 45 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Motor Controller Circuit Description

The Galil Motor Controller is an optional daughterboard installed in the


FEC PCB. The board controls the speed and location of a variety of
motorized transducers and reports position information to the system in
real-time. The board is also responsible for monitoring and limiting the
amount of power delivered to the motor to comply with safety issues. The
Omni Motor Controller is integrated onto the FEC PCB. The Omni Motor
Controller is an optional feature. Not all FEC PCBs have an integrated
Omni Motor Controller.
Transducers are calibrated when connected to the system. This verifies to the
system the transducer range of motion and sets the position counter to the
home state or that of an end-point. Once calibrated, the Motor Controller is
available for position or velocity control.
The Motor Controller can be used for constant velocity acquisition or
stepped acquisition. During constant velocity moves, the Motor Controller
is programmed via the FEC with a motion profile consisting of a speed and
range over which the speed should be maintained. While this is happening,
the Krusty FPGA on the FEC is set up by the system with look-up tables to
fire a sequence. Once set up, the Krusty notifies the Motor Controller to
start the transducer motor. When the transducer motor comes up to speed,
the Krusty is notified that it may shoot frames and retrieve valid data from
the Motor Controller.
During stepped acquisition, the desired sweep rate is too slow for a constant
velocity control. At predefined thresholds, when the motor is stalled, the
system calculates and initiates a series of stepped movements. The FEC sets
up the motion profile as with the constant velocity control (only with a
string of positions) and Krusty initiates the motion. Once the desired
position is achieved, the Krusty shoots a frame and signal for the Motor
Controller to move to the next position.
A second type of stepped acquisition occurs when the transducer is moved
to a location and fires a set of 2D frames. (Essentially, the transducer is
being used as a directional 2D probe.) Motor Controller operation is the
same for both types of stepped acquisition. The Motor Controller holds the
transducer position until instructed to move elsewhere.

Conquest Imaging - 46 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
When not in use, the Motor Controller is shut down to reduce the chance
of motor voltages appearing on the transducer connector and to reduce the
possibility of noise originating from the PCB.

Signal Processing Subsystem

The Signal Processing Subsystem comprises functional blocks that operate


on and modify the received signals after beamforming and before image
processing. The functional blocks of the signal processing subsystem are
partitioned over the following modules.
 Dual Signal Conditioning PCB (DSC)
 Signal and Image Processing (SIP) Motherboard

Dual Signal Conditioning PCB (DSC)

• Processes dual real-time streams of information from the front-end


including echo, Color, CPA imaging, Power Motion Imaging, Tissue
Doppler Imaging, PW Doppler, CW Doppler, 2D/M-mode, and
Color M-mode) before sending the data to the motherboards
• Buffers RF data and passes information to the Yukon ASIC where the
front-end sample rate is decoupled from the DSC clock
• Writes data from each RF bus to the PCI bus for storage in an RF
capture buffer (in system memory)
• Writes RF headers, RF data, or both to the RF capture buffer
• Filters lines for capture using the data-type header
• Yukon ASIC, dual data streams
- RF A/B Bus summed and swapped
- Normalization (applies depth versus gain curves)
- RF multizone fine delay
- RF multizone blend
- Lateral RF interpolation filter
- Synthetic aperture

Conquest Imaging - 47 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
• Styx ASICs
- Quadrature Bandpass Filtering (QBP)
- Lateral gain compensation
- Automatic Gain Control
- Band equalization
- Focal zone matching
- Digital time gain compensation
- Detection
- Compression
- Contrast resolution enhancement (frequency compounding
summation)
- Detected multizone blending
- Axial interpolation
- Multi-rate filtering (MRF)
• Flow processing
• PCI interface
• Receives data from RF Buses (RF A and RF B)
• Passes CW data without processing

Signal and Image Processing Motherboard (SIP)

• Contains signal and image processing for all imaging modes.


• Scan conversion.
• Contains image (Cineloop) memory.
• Wall filter and gain.
• Connected to the Host Motherboard by the Inter-Platform Link (IPL), a
64-bit PCI-PCI custom bridge. (IPL handles just data. Control is
accomplished via FireWire cable.)
• Sends image data (native images) to Host for graphics overlay.
• Monitors +3.3 Vdc, +5.0 Vdc, ±12.0 Vdc, CPU core voltage, CPU
temperature, and the motherboard ambient temperature.
• Stores images to two hard drives.

Frame Composition

Software on the Host Motherboard accomplishes frame composition. Scan-


converted image data is routed from the SIP to the Host, where it is
processed into frames and overlaid with the image graphics.

Conquest Imaging - 48 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Host Motherboard

• Receives native images from the SIP via the IPL.


• Performs frame composition on native images.
• Adds overlay graphics to native images.
• System-mounted microphone input for recording to the VCR.
• Provides the PCI bus interface to the AVIO, DSC, IEP, and SIP
Motherboard.
• Connected to SIP Motherboard by the Inter-Platform Link (IPL), a
64-bit PCI-PCI custom bridge.
• SIP and Host communicate commands via embedded FireWire. Image
data moves across IPL.
• System memory.
• Monitors +3.3 Vdc, +5.0 Vdc, ±12 Vdc, CPU core voltage, CPU
temperature, and the motherboard ambient temperature.
• System software is stored on Host hard drive.

Display Subsystem

The Display Subsystem formats the echo information for display, and
transfers it to the video monitor and peripheral devices.
The Display Subsystem consists of:
• AVIO-RIP PCB
• AVIO-CAD PCB
• Image Export Processor Board (IEP)
• AGP Graphics Card

Audio Video Input/Output Rear Interface Panel (AVIO-RIP PCB)

• Video distribution for SXGA (RGBVH) composite (CVBS) and


component (Y-C) video to and from the AGP graphics card,
supports NTSC and PAL.
• Provides mute controls for VCR line inputs (line out and microphone)
and speakers.
• Y-C video multiplexing for two sources of component video. (One port
is for internal SVHS VCR; the other is user-defined. The second
source may be endoscopic transducer picture-in-picture views.)
• Independent two-speed control for Acquisition and Platform Card Cage
fans.
Conquest Imaging - 49 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
• Print control and status connection for an external printer.
• Battery charger and terminal discharge monitor.
• USB hub interface expansion from one USB on the Host Motherboard
to four on the AVIO. Each has independent power control and
provides individual port current limiting and port status to the hub
controller.
• Provides a battery backup source to the Platform Power Supply for +5
VSB generation.
• Provides pass-through routing to the serial number device interface
(iButton). Obtains
64 bits of data from the iButton using read ROM command. The
iButton stores the system serial number and identifies the system
during software or feature upgrades.
• Contains a 32-bit PCI bus to the Host Motherboard bridged to the
AV_PCI bus.
• AV_PCI bus provides interface to Acquisition Card Cage, UI
graphics controller (SVGA touch screen LCD), and local bus
interface adapter.
• AV_LB bus (local bus) is the primary control for volume control D/A,
audio routing, fan speed, print/print status, ancillary control signals for
the AVIO data path, Doppler audio generator data and control
interface, and voltage and temperature monitoring functions.
• Buffers and performs ESD/EMI functions on component and
composite video from the AGP graphics card. Allows user access to
video and routes video back to AGP graphics.
• Buffers composite capture input and the S-Video Luma input so the
outputs can be directed to the same pin on the AGP graphics card.
(AGP does not support dual video inputs.)
• Boosts low-frequency for power amp.
• Monitors ±12 VDC, +12V_STBY, battery voltage, battery charge
current, and AVIO ambient temperature.
• Board-level identification. Contains an EEPROM for device
configuration information and storage. System software can write
to the VPD data storage area
• Supplies AVIO-CAD interface.

Conquest Imaging - 50 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Audio Video Input/Output (AVIO-CAD PCB)

• Generates analog Doppler audio at a configurable sample rate locked


to the acquisition system clock (PRF rate data)
• Extends the platform PCI bus to the Acquisition Card Cage as a 32-bit
segment (Acquisition control bus interface to the FEC)
• PCI bus based dual-head SVGA controller and LVDS transmission to
the touch screen LCD
• Control interface to the AVIO-RIP
• Provides an interface to the iButton

Image Export Processor – IEP

• Accelerated JPEG compression and export of frame composed


images.
• Contains a GPU (graphics processor unit) to optimize frame sizes and
frame rates.
• RGB from GPU is color-space converted, buffered for raster
(line) to block conversion (RBC), and presented to the JPEG
encoder for compression (up to 30:1).
• Compressed image frame data is then high-speed transferred
to the Host memory.
• Interfaces with the Host Motherboard to setup compression
parameters.
• No local processor to execute diagnostics. Diagnostics execute on the
Host Motherboard and consist of verification of write/read access
to selected FSEs, composition of test patterns on the GPU, and
capture of GPU output and comparison against references, both
compressed and non-compressed.
• Operates independent of the system display path.
• Standard PCI card and resides in a 64-bit/66-MHz slot on a Host PCI
bus.

Conquest Imaging - 51 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
IEP Circuit Description

The Image Export Processor provides accelerated JPEG compression and


export of frame composed images. It is a standard PCI card located in a 64-
bit/66-MHz slot on the Host Motherboard.
Line image data is obtained from the Host Motherboard via the PCI bus and
is passed through the PCI-PCI bridge to the GPU. The frame composition
capabilities of the GPU are similar to and independent of those occurring in
the system display path. The parallel processing enables optimized frame
sizes and frame composition of exported video data.
Frame composition software on the Host Motherboard alternates on a frame-
by-frame basis between the system display path and the background image
export path. The alternation rate is 75 Hz driven by the AGP vertical sync
interrupt. Thus, frame composition of the IEP is accomplished at a frame rate
of 37.5 Hz. Frame size is fixed at 1024 x 768 pixels (compared to display
resolution of 1280 x 1024).
The 24-bit RGB output from the GPU is color space converted to YCbCr
color space and is buffered and raster-to-block converted. Lossy JPEG
compression (between 5:1 and 30:1) is set up by the Host Motherboard.
The Data Flow Controller is programmed with frame size, region of interest,
JPEG quantization tables, and miscellaneous control signals from the Host
Motherboard via the PCI bus.
Compressed images are transferred through the local bus to the PCI-local
bus bridge which controls data output between the Image Export Processor
and the image memory on the Host Motherboard.

Graphics Card -AGP (ASUS V8420)

• Provides RGB lookup tables to map Doppler spectral column values to


the RGB values
• Formats spectral data to fit the display size

Conquest Imaging - 52 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Dolphin Card

• Provides the Inter-Platform Link for data transfer between the SIP and
Host Motherboards
• 64-bit PCI-PCI bridge
• 100 MB/s bandwidth in each direction at approximately 667 MB/s

FireWire Interface (IEEE 1394)

• Provides TCP/IP connection and control path between Host and


SIP Motherboards
• Provides wireless LAN connection between Host and SIP
Motherboards

USB
• Provides control and access to QWERTY keyboard, trackball, touch
screen, Bluetooth transceiver, peripherals, and CD/DVD drive (via a
USB-IDE adapter)

Output Power Monitor Theory

Output power monitor provides patient safety by ensuring the system is


operating within acoustic power and intensity (AP&I) limits. It also protects
hardware from destructive or unsafe conditions. It consists of AP&I
software, which resides on the System CPU; FEC software; pulser monitors
in each Channel Board; and a beamformer monitor.
The output power monitor is designed to detect any single failure in the
system, for example, a faulty power supply, a faulty beamformer, a software
failure, a defective transducer or transducer data files. Mode information,
(PRI, voltage, burst, and so forth) sent to the beamformer is also verified
before allowing the beamformer to start. This prevents the system from
inadvertently creating a temporary over-power condition.
The DC output level is monitored for each pulser power supply (on the
Channel Boards, on the AIM+, and within the transducer). Output load
current is measured for each pulser supply and compared to levels expected
by the Acoustic Power and Intensity Manager (APIM). The monitoring is
continuous in all operating states and shuts down the malfunctioning pulser
supply within two seconds of fault detection.

Conquest Imaging - 53 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
The current monitor, tracks the current through the AIM regulator. The
voltage monitor tracks the input and output voltages on the AIM regulator,
and the watchdog timer automatically disables the HV supplies if the current
and voltage monitors are not being read by the FEC local processor.

Acoustic Power and Intensity Manager (APIM)

Acoustic Power and Intensity Manager (APIM) software controls the


acoustic power and the surface temperature of the transducer face while
under ultrasound system control. Acoustic power and temperature limits
are set by various regulatory agencies. The APIM monitors the output
power and temperature and ensures the parameters are not exceeded. For
example, display of output power parameters (MI, TI, SPTA) on the system
monitor allows the user to compare anatomy of one patient to another at a
specific output value or to switch modes while keeping the 2D output
constant.
The APIM selects the appropriate voltage level that the FEC must use, given
the current transmit conditions of the ultrasound system and target output
level. The voltage level is calculated based on the system state, transducer
data, AP&I data, and other factors. The voltage level is set so the output
power is as close as possible to the desired setting (MI, TI, or SPTA)
without exceeding the desired level. A power monitor on the FEC verifies
that the FEC is consuming electric current at the rate expected by the
APIM.
Minor errors indicate that acoustic power and intensity may be calculated
and the errors are reported to the error log. Major errors specify conditions
where acoustic power and intensity cannot be calculated and scanning
cannot be initiated for the condition.

Conquest Imaging - 54 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Temperature Monitoring

Temperature monitoring in the Acquisition Card Cage is accomplished


using 11 temperature sensors. Two are on each Channel Board, two are on
the AIM+, and one is on the FEC. The Channel Board and AIM+ sensors
are in the input and exhaust air streams. The sensor for the FEC is located in
the exhaust air stream. The temperatures are not monitored for the first 10
minutes of system operation. After that the system fan speeds are
automatically adjusted according to the system temperatures at the various
sensor locations.If the sensor temperatures drop too low, the fan speeds drop.
Temperature monitoring in the Platform Card Cage is accomplished with a
temperature sensor on each of the motherboards and one on the AVIO-RIP.

X3-1 Matrix Transducer Theory

The X3-1 Matrix transducer is a non-motorized 3D transducer. Transducer


identification (personality) is controlled the same as for other transducers.
The transmit path for the X3-1 Matrix transducer is contained within an
ASIC in the transducer. Beamformer control lines from the NAIM are used
to control beamformer timing (for 2400 elements), TGC, and transducer
power. Some of the control information originates in the FEC PCB. The
Channel Boards are not used for the X3-1 transmit signal path. Programming
functions that are normally stored in Channel Board Ram are stored in 32
MB of memory located on the NAIM. The programming information is
transferred to the beamformer via the MX Bus.
The receive path within the system is similar to that of other transducers,
however, the signals received from the 2,400 elements are multiplexed to 128
packets of information. The packets are processed the same as signals
received from other transducers. The signals are sent to the system back-end
on the RF A and RF B data buses.

Conquest Imaging - 55 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Buss Functions

Platform Subsystems Busses

PCI Bus

The system platform PCI bus interface on the Host Motherboard provides
access to the AVIO, DSC, IPL, IEP, and the SIP Motherboards. It transfers
signal path data transactions, transducer tables, and other data. It also
transfers RF data from the DSC to the Host, estimate data from the DSC to
the SIP, Native Format and Native Format Cineloop image data (scan-
converted and post-processed) from the SIP to the Host memory, and image
import and export data from the Host.
The PCI bus is a segmented bus, meaning that separate portions of the bus
have different bus speeds.

Universal Serial Bus (USB)

The Host Motherboard accesses the DVD drive, modem, user interface,
VCR, and other digital video OEMs through several USB buses. The USB
connector on the back of the DVD drive is USB 2.0 standard. All others are
USB 1.0.

Inter-Platform Link Bus

Provides a PCI-PCI bridge that connects the Host Motherboard IPL module
to the SIP Motherboard IPL module. Connects each motherboard to a 66-
MHz PCI bus segment.

Conquest Imaging - 56 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Inter-Card Cage Buses

Acquisition Control Bus Buses

The Acquisition Control Bus provides a high-bandwidth interface between


the Acquisition Card Cage and the Platform Card Cage. It provides two
unidirectional, five-channel LVDS data and clock paths, reset and hold
lines from the Platform Card Cage to the Acquisition Card Cage, and link
clock signal lines. The physical connections are between the FEC and the
AVIO. The bus also provides a low-noise bus structure to minimize
EMI/RFI. It moves all control and ancillary data movement (non-
beamformed RF data).

RF Bus A/B

The RF Bus comprises two identical 28-bit bus structures (only 24 bits are
used). RF data is moved from the FEC to the DSC via this bus. There is no
handshake provided for data transfers.

Front-End Buses

FEP Bus

The FEP Bus is a static address, data, and control structure that provides
access to various resources on the AIM+, Channel Boards, and S/HSEL
modules. This is a 16-bit extension of the FEC local processor bus operating
at 40 MHz.

FEC Bus

The FEC Bus provides real-time access to the Nile ASICs on the Channel
Boards and the AIM+ for control and timing purposes. The bus architecture
is a static address and data with additional control signals for SDRAM
access and control. The clock rate is 40 MHz.

Conquest Imaging - 57 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
SUM Bus Structure

The SUM Bus structure is composed of two identical bus paths of four
segments, daisy-chained between the Channel Boards. Each Channel Board
sums its channels. It then sums its result with the sum of the previous
Channel Board and passes the result to the next Channel Board. Data ends
at the FEC. It can transport RF data at single or double rates. C
W RF Doppler data may be driven onto this bus by the AIM+. Clock rate is
32 or 64 MHz.

TIM Bus

Provides all PRI-PRI real-time control signals as required by the Nile ASIC
and the AIM+ CW Doppler processing module. The control signals include
transmit and receive coarse-delay control, high-voltage rail selection, PRI
reset, and skin-line identification. The clock rate is 32 MHz.

AIM Bus

This bus transports data from the FEC to the AIM+. It has a 5-bit command
bus and a 16-bit data bus. It is time-multiplexed between TGC control, HV
MUX programming data, transducer setup for certain transducers,
microbeamformer scanfile download (for some transducers). The clock rate
is 40 MHz.

Motor Bus

The Motor Bus is the transmission path for multi-phase DC motor control. It
includes four high-current motor drive signals and six encoder channels. This
bus is a subset of the interface to the Motor Controller PCB connector on the
FEC.

Conquest Imaging - 58 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU 22 Basic Block Diagram
HOST

Hard Drive


D V D

DVD drive

Modem PCI Bus - 64 Bit

PCI Bus - 32 Bit


CP/UIF

Dolphin
External
Network
AGP

PH: 209-942-2654
Dolphin

Motor Control
Data Paths

Rev3 070208.dms
FEC RFA
RFB

Conquest Imaging
DSC SIP IEP

FAX: 209-942-2572
AIM+

Ultrasound Sales Service Repair Training


1815 Industrial Drive, Suite 100 Stockton, CA 95206
OEMs
Hard
Drives

Scanhead Select
Channel USB
Boards

Acquisition Front Plane


Display

- 59 -
Acquisition Subsystem AVIO CAD AVIO RIP
This Page Intentionally Left Blank

Conquest Imaging - 60 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™

Chapter 2
Operational Testing

Conquest Imaging - 61 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 62 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ Operational Testing

Beginning an Exam

Note: Most newer Ultrasound Systems depend on the opening and closing of
exams to correctly initialize and maintain internal file structures.

To proceed the following must be true:

• Transducer ports are operational


• Transducer, Patient Data, and the QWERTY keyboard are
operational
• Touch screen is operational and that the calibration has been
successful. No further testing of the touch screen is necessary.

1. Connect and initialize any transducer. If possible, connect transducers


to all three ports.
2. Press Transducer to display the transducer touch screen.
3. Select a transducer from the touch screen.
4. Touch a clinical option and a Tissue Specific preset.
5. Press Transducer.
6. Press Patient Data and enter a patient name.
7. Click Close.

Operational Testing

2D

1. Press 2D to enter 2D imaging (if needed.)


2. Turn 2D to change the gain.
3. Adjust TGC slides. Verify changes in overall image brightness.
4. Turn Zoom. Verify the image increases from and decreases to the
original size.
5. Turn Depth. Verify the changes in depth.
6. Press Zoom to bring up the HD zoom box.

Conquest Imaging - 63 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
7. Use the trackball and Select control to move and size the HD zoom
box.
8. Press Zoom to enlarge the boxed area.
9. Press Zoom to return to 2D imaging.
10. Adjust Res/Spd. Verify the icon updates on the touch screen.
11. Adjust Gray Map. Verify the map changes on the display and the
map number changes on the touch screen.
12. If a curved array or sector transducer is connected, adjust Sector
Width to reduce the field of view, and restore it to the default size.
13. Adjust Compress to change image compression. Verify the
compression value on the touch screen and the value (C#) on the
upper left corner of the main display changes.
14. Touch Next.
15. Adjust Output Power. Verify the MI and TI values change on the
main display.
16. Press iSCAN. Verify the iSCAN icon changes to an amber color for
about 3 seconds.
17. Press Freeze. Verify the image is frozen and that the Cineloop
cursors and image number are displayed below the image.
18. Press Print. Verify a beep is heard and a thumbnail shows upon the
right side of the main display. Successful execution of the above steps
verifies operation of the 2D, Depth, Focus, Zoom, iSCAN, Freeze,
and Print controls and all touch screen knobs except those on the far
right and far left.

Color/CPA

1. Press Color. Verify that the color box is displayed.


2. Turn Color to change the gain. Verify that the color box is filled with
color noise.
3. Adjust Scale. Verify that the scale displayed on the color bar changes.
4. Adjust Baseline. Verify the baseline changes on the color bar.
5. Press CPA. Verify that the color bar changes to a an orange hued
color.

Conquest Imaging - 64 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
6. Turn CPA. Verify that the color in the color box changes to the orange
colors.
Successful execution of the above steps verifies operation of the Color
and CPA knobs and the remainder of the touch screen knobs not used
in 2D imaging.

PW Doppler, CW Doppler, and M-mode

1. While still in CPA, press PW. Verify the sample volume marker is
displayed.
2. Press Update to toggle between the 2D/CPA image and the scrolling
Doppler display.
3. Touch Simult. Verify that the 2D/CPA and scrolling Doppler displays
are live.
4. Turn PW to change the Doppler gain. Verify there is Doppler noise on
the display and that you hear Doppler audio.
5. Turn Volume. Verify that the volume level changes.
6. Verify your system and transducer support CW Doppler.
7. If so, press CW to begin CW imaging. Verify the proper display.
8. Turn CW to change the Doppler gain.
9. Press 2D to return to 2D imaging.
10. Press M-Mode and Update to obtain a scrolling M-mode display.
11. Turn M-Mode to change the M-mode gain. Successful execution of
the above steps indicates:
• PW, CW, M-mode, Update and Volume are operational.
• Doppler audio signal path and speakers are operational.

Real Time 3D

1. Verify the system contains a motor controller PCB and that the real-
time 3D licensed option is enabled.
2. Connect a motorized transducer.
3. Select a clinical option and a Tissue Specific preset. If there are no
Tissue Specific presets displayed, install or flash the motor controller,
or install the licensed option.
4. Touch 3D/4D.

Conquest Imaging - 65 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
5. Acquire an image using Single Sweep 3D. Verify that a volume has
been acquired.
6. Acquire an image using 4D and note the presence of a live volume.
(Move the region of interest box to the image data if necessary.)
Successful execution of the above steps indicates:
• the 3D motor controller has been successfully installed and flashed
• the appropriate licensed options are installed

Connectivity and Review

To proceed the following must be true:

• End Exam and Review are working


• The patient database is functional
• DVD drive is operational
• Transfer of JPEG loops to the DVD and back means that the IEP
board is functional after it had been re-flashed during the SW
installation

1. Press 2D and press Color.


2. Turn Color to adjust color gain until there is color noise in the color
box.
3. Press Print. Verify that you hear a beep and the acquired image
appears in the thumbnail on the right side of the display.
4. Press Capture (while still in live imaging). Verify an hourglass
appears and a delayed beep is heard. Also verify that another
thumbnail is displayed with a loop icon on the right side of the
display.
5. Press End Exam.
6. Press Review to display the Patient Directory.
7. Insert a new DVD into the DVD drive.
8. Click Format to format the disk.
9. Select the patient exam that was most recently ended and click Send
To.
10. In the Send To dialog box, select DVD and click OK.
11. Verify that the green dot on the DVD queue icon comes on briefly.

Conquest Imaging - 66 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
12. Delete the patient that was copied to the DVD from the hard drive.
13. Select the DVD directory, highlight the exam that was copied, and
transfer the exam to the hard drive.
14. Select the hard drive directory and the exam that was copied from the
DVD. Click on View Selected. Verify that if in a 4:1 image
presentation, the first loop appears as a JPEG loop icon, while the
print appears as it was taken.
15. Double-click on the JPEG loop icon to obtain a full-screen playback
of that loop.
16. Touch Cine Play.
17. Click Close.

Annotations and Body Markers

1. Press 2D and obtain a 2D image.


2. Press Annotate. Verify that the Annotation touch screen is displayed
with a blinking green cursor on the main display.
3. Touch any annotation term on the touch screen. Verify that the
annotation appears on the main display at the cursor location.
4. Press Erase Text. Verify that annotation disappears from the display.
5. Press Body Mark. Verify a body mark graphic appears in the upper
right of the display.
6. Press Body Mark to erase the body mark.
Successful execution of the above steps implies that Annotation, Erase
Text, and Body Mark are operational.

Analysis

1. Connect and select a transducer used for OB imaging.


2. Touch a clinical option and a Tissue Specific preset used for OB
imaging.
3. Select
4. Press 2D and obtain a 2D image.
5. Press Freeze.
6. Press Caliper. Verify that a single caliper is displayed.
7. Press Select and verify a second caliper is displayed.
8. Press the left and right Select to toggle between the two calipers. (The
active caliper is green.)

Conquest Imaging - 67 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
9. Press Calc to display the measurement labels on the left side of the
display.
10. Press Select to assign the measurement to the highlight label.
11. Press Report. Verify that the report is displayed with the patient
name and the measurement just made.
12. Press Report to return to 2D imaging.
13. Press Pointer a couple of times. Verify an arrow appears and
disappears from the display. Successful execution of the above steps
indicates that Caliper, Calc, Report, Pointer, and the Select buttons
work.

Voice Recognition

1. Verify that the Voice Recognition licensed option is enabled.


2. Power up the wireless headset.
3. Press Setups on the keyboard.
4. Click Voice Control and Headset Config. Select the headset and click
Pair.
5. Click Done to exit Setups.
6. Double-click on the Voice icon (headset) to establish voice control.
7. Speak into the headset. Verify the headset icon appears amber or
green.

Peripherals VCR

1. Set up the VCR.


2. Connect and initialize any transducer.
3. Acquire a PW Doppler image.
4. Record several seconds of video.
5. Rewind the tape to the beginning of the current recording.
6. Play back the video and verify that the graphics and gray bar on the
display are in sharp focus and that the gray bar displays the full range
of gray shades as seen on the monitor.
7. Verify that the displayed image is spatially correct, horizontally and
vertically.

Conquest Imaging - 68 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Printer

1. Press Setups.
2. Click Print/Network.
3. Click the Device Selection tab.
4. Select the installed printer.
5. Connect and initialize any transducer.
6. Acquire a 2D image.
7. Press Freeze.
8. Press Print. Verify that the print contains region of interest and full-
screen images.
9. Touch Alt Print. Verify that the print contains region of interest and
full-screen images.
10. Verify that the graphics and gray bar on the print are in sharp focus
and that the gray bar displays the full range of gray shades as seen on
the monitor.
11. Verify that the print image is spatially correct, horizontally and
vertically.

Conquest Imaging - 69 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 70 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™

Chapter 3
Hardware Access

Conquest Imaging - 71 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 72 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ Hardware Access

Accessing the Acquisition Card Cage and Frontplane

1. Remove the two screws securing the upper front panel enclosure.
2. Slide the upper front panel enclosure up to remove it and set it aside.

Conquest Imaging - 73 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Lower Front Panel Enclosure

3. Remove the two screws securing the lower front panel enclosure.
4. Slide the lower front panel enclosure up to remove it and set it aside.

Conquest Imaging - 74 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Top Control Panel Articulation Enclosure

5. Slide forward and remove the top enclosure that surrounds the control
panel articulation and set it aside. There are no screws securing this
cover to the frame. It is held in place when the front panel enclosure is
installed.

Conquest Imaging - 75 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Control Panel Articulation Arm Beard Enclosure Assembly

6. Remove the four screws securing the articulation arm beard enclosure
assembly.
7. Rotate the front side of the enclosure assembly downward.
8. Then slide the enclosure assembly forward to remove it. Set it aside.

Conquest Imaging - 76 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Rotating the Front Panel Hanger Brackets

9. Rotate all four front panel hanger brackets. Rotate the brackets up 90
degrees until they are flush with the inside of both side panels. There
are two brackets on each side of the panel. This is necessary to allow
sufficient clearance for the removal of the Frontplane assembly.

Conquest Imaging - 77 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Remove the Front Plane Shield

Frontplane Assembly Screws

10. Remove the 17 screws securing the Frontplane assembly to the front
of the Acquisition Card Cage.

Conquest Imaging - 78 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Disengaging the Frontplane assemply

11. Gently pull (do not force) the mechanical arm up to disengage the
Frontplane assembly from the card cage PCBs. Do not use more than
25 pounds (11.4 Kilograms) of force.

Conquest Imaging - 79 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Caution:

Conquest Imaging - 80 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Uncabling the Frontplane Assembly

12. Gently lift the Frontplane assembly up and off the four hanger
brackets.
13. Tilt the top of the Frontplane assembly toward you and lower it until
it rests on the static mat.
14. Disconnect the four power cable connections. The Frontplane
assembly is now completely disconnected.

Conquest Imaging - 81 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Acquisition Card Cage PCBs and Scanhead Select Assembly

Removing the Front Column Disk Bay Enclosure

NOTE To remove the Scanhead Select Assembly, you must first start with
the removal of the Frontplane assembly

1. Remove the 10 screws securing the front disk bay enclosure.

Conquest Imaging - 82 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Rotating the PCB Retainer Brackets

2. Rotate all four PCB-retainer brackets away from the card cage before
attempting to remove any PCBs. Two brackets are at the top of the card
cage and two are at the bottom. They swing outward from the middle.

Conquest Imaging - 83 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Remove the Acquisition Card Cage PCBs

3. Slide the Front-End Controller (FEC) assembly out of the card cage until
you can reach the ribbon cable connected at the bottom of the assembly.
4. Disconnect the ribbon cable and set the FEC assembly aside on an ESD
mat.
5. Remove all the remaining PCBs from the card cage by lifting the
individual board retainers to disengage them from the Scanhead Select
Assembly. Set the boards aside on ESD mats.

Conquest Imaging - 84 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Releasing the Captive Screws

6. Unscrew the 10 captive screws inside the back of the Acquisition Card
Cage.

Conquest Imaging - 85 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Remove the Scanhead Select Assembly

7. Gently lift the Scanhead Select Assembly, tilt it forward, and remove the
assembly from between the Acquisition Card Cage and the system frame.

Conquest Imaging - 86 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
AC Tray

Removing the AC tray

1. Remove the six screws securing the rear panel enclosure assembly. Set the
assembly aside.

Conquest Imaging - 87 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
2. Remove the power cord.)
3. Remove the two screws securing the AC tray to the cart frame.

For A.x systems, proceed to step 4.


For B.0 systems, proceed to step 7.

Conquest Imaging - 88 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Removing the AC Tray (A.x Systems)

4. Slide the AC Tray out toward you until the tray stops.
5. Disconnect the red/black, yellow/ black, OEM power, and main power
cables.
6. Remove the tray from the system by sliding it the rest of the way out of
the support shelf. This completes the A.x system version of the AC tray
disassembly procedure.

Conquest Imaging - 89 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Removing the AC Tray (B.0 Systems)

7. Slide the AC Tray out toward you until the tray stops.
8. Disconnect the green/yellow, brown, blue, and main power cables.
9. Remove the tray from the system by sliding it the rest of the way out of
the support shelf. This completes the B.0 system version of the AC tray
disassembly procedure.

Conquest Imaging - 90 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Software Change History (iU22)

1.0.0.95 Released on March 19, 2004

453561163821 Software collection


8000-2630-01 Software BOM part number
453561162541 System part number
453561191281 Programmed Host Drive, 1.0.0.95

New Features
• Broadband Sector (Phased) Array Transducers
- S4-1
• Broadband Linear Array Transducers
- L8-4
- L12-5 50 mm
- L17-5 High Frequency Linear Array
• Broadband Curved Array Transducers
- C5-2
- C8-4v
- C8-5
- C9-4
- C9-5ec
- 3D9-3v
- 3D6-2 Trans-Abdominal Broadband Curved Array
• Voice commands
• iSCAN Intelligent Optimization
• SonoCT Real-Time Compound Imaging
• XRES Visualization Technology
• Freehand 3D imaging and 4D imaging
• Contrast imaging
• Panoramic Imaging
• Connectivity

Conquest Imaging - 91 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
1.0.1.3 Released on June 15, 2004
This software is a corrective build. It is available on new-build
systems and for upgrades on an as-needed basis. There is no
upgrade program associated with this release. Fixes associated
with this release will be incorporated into a subsequent baseline
software version.

453561178561 Software collection


453561191291 Programmed Host Drive, 1.0.1.3

New Features
There are no new features associated with this release.

1.1.0.138 Released on June 22, 2004


Upgrade name: System Update 2.0

453561181411 Software collection


453561191301 Programmed Host Drive, 1.1.0.138

New Features
• S3-1 Broadband Phased Array (Sector) Transducers
• Adult Cardiology clinical option
• Cardiology (Adult Echo) Calculations
• Support for DICOM color printers
- Agfa DryStar 4500
- Agfa DryStar 5500
- Fuji DryPix 7000
- Fuji DryPix FM-DP L
- Kodak DryView 8900
- Kodak Keli 160
- Konica DryPro 722
- Konica DryPro 752

Conquest Imaging - 92 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
1.1.1.147 Released on August 24, 2004
Software version 1.1.1.147 is a maintenance build to allow
replacement of the ASUS video card with the MSI video card
when the ASUS is end-of-life. There is no upgrade associated
with this release. Fixes associated with this release will be
incorporated into a subsequent baseline software version.

453561186421 Software collection


453561191311 Programmed Host Drive 1.1.1.147

New Features
There are no new features associated with this release.

1.1.2.197 Released on October 8, 2004


Software version 1.1.2.197 is a baseline upgrade that supports
several feature enhancements and the corrections to system
operation. There are no new hardware requirements for the
software. Included with the upgrade kit is a Certificate of
Authenticity label. The label represents the Philips purchase of
Microsoft-licensed software. Install the label as indicated in the
upgrade instructions.

453561192621 Software 1.1.2.197


453561193441 Programmed Host Drive, 1.1.2.197

New Features

• Allows use of retired ultrasound and retired ultrasound


mainframe SOP classes for improved interoperability with older
PACS.
• Adds the Advanced control to the Archive area of the Device
Selection tab in the Print/Network setups to select between
ILE and ELE transfer syntaxes to archive servers. The iU22
system supports both syntaxes. Some PACS crash when
attempting to use a transfer syntax that they do not support. For
1.1.2 software, the setting Use implicit transfer syntax only is
disabled after it has been selected.

Conquest Imaging - 93 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
• Adds the ability to optimize image quality on exported images
by allowing the user to apply one of four lookup tables to the
exported images.
• Adds the ability for the user to modify these printer settings:
image display format, minimum density, maximum density,
film destination, and configuration information.
• Adds the ability for the user to configure whether the system
automatically selects the appropriate DICOM printer for color
or black-and-white images.
• Adds a Study Description field to the Patient Data display.

NOTE There are no image quality changes to the C5-2 Abdomen, S4-1
Abdomen, L8-4 Vascular, or 3D6-2 OB presets. Changes were made for the
C5-2 OB and L12-5 OB Tissue Specific presets for Harmonic SonoCT
imaging.

1.1.3.283 Released on February 2, 2005


Software version 1.1.3.283 is a baseline upgrade that supports
the release of the Hewlett Packard LaserJet 1320 (HP LJ1320)
and corrects various operational issues. There are no changes or
enhancements to image quality associated with this release.

453561203711 Software 1.1.3.283


453561209821 Programmed Host Drive, 1.1.3.283

New Features
There are no new features associated with this release.

1.1.4.311 Released May 16, 2005


Software version 1.1.4.311 is a corrective build only. There are
no changes or enhancements to image quality associated with
this release.

453561215211 Software 1.1.4.311

Conquest Imaging - 94 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
New Features
There are no new features associated with this release.
2.0.0.320 Expected to be released September 2005 Upgrade
name System Update 3.0 This software supports the new
features listed and makes several corrections to system
software.

453561238121 Software 2.0.0.320

New Features
• L9-3 Linear Array Transducer
• X3-1 Matrix Transducer
• S7-2 Sector Array Transducer
• S5-1 Sector Array Transducer
• D2cwc CW Transducer (Pedoff)
• D5cwc CW Transducer (Pedoff)

Software Change History (iE33)

1.0.0.399 Released on November 30, 2004

453561195251 Software collection


453561175241 System part number

New Features
• Transducers listed in Table 15-1
• Live xPlane Imaging
• Live 3D Echo
• Panoramic Imaging
• Pulse phono
• Voice commands
• Connectivity

Conquest Imaging - 95 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Acquisition Board Layout

Conquest Imaging - 96 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Acquisition Card Cage

iE33 - Neo Analog Input


Module
(NAIM)

Conquest Imaging - 97 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Supported Media iU22

• If you want to read files stored on the DVD on a PC, a UDF reader must be
installed on the PC if the PC does not use Windows XP. The UDF reader
may be downloaded free from the Internet.
• On initially released systems, Philips recommends the use of Verbatim or
Sony DVDs. Other DVD brands have not performed as well on the system.
On systems with 2.0 software, compatibility issues when using certain
brands of DVDs in the system DVD drive can result in DVD write errors
and system lockups. Philips recommends using only TDK DVD +RW and
Maxell DVD + RW media in the system DVD drive.

Types of Media Supported in the DVD Drive (S/W Below 2.0)

Media File System Notes


DVD+R UDF Use Verbatim or Sony DVDs (not
Imation)
DVD-R UDF Use Verbatim or Sony DVDs (not
Imation)
DVD+RW UDF Use Verbatim or Sony DVDs (not
Imation)
DVD-RW UDF Use Verbatim or Sony DVDs (not
Imation)
CD-RW UDF Requires 20 minutes to format, only
supports to 10X
CD-R ISO No need for UDF reader, can be read
anywhere. ISO can be formatted on the
CD using the DVD format utility

Conquest Imaging - 98 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Types of Media Supported in the DVD Drive (2.0 S/W and
Above)

Media File System Notes


DVD+R UDF Files require UDF reader to be read on a
PC.
DVD+RW UDF Files require UDF reader to be read on a
PC.
CD-R ISO No need for UDF reader, can be read
anywhere. ISO can be formatted on the
CD using the DVD format utility

Only the preferred types of media supported in the DVD drive are listed.
Drives are also compatible with CD-RW, DVD-RW, and DVD-RW.
However, CD-RW and DVD-RW drives have slow write speeds.

Supported Media iE33

• If you want to read files stored on the DVD on a PC, a UDF reader must be
installed on the PC if the PC does not use Windows XP. The UDF reader
may be downloaded free from the Internet.
• Philips recommends the use of Verbatim or Sony DVDs. Other DVD
brands have not performed as well on the system.

Types of Media Supported in the DVD Drive

Media File System Notes


DVD+R UDF Files require UDF reader to be read on a
PC.
DVD+RW UDF Files require UDF reader to be read on a
PC.
CD-R ISO No need for UDF reader, can be read
anywhere. ISO can be formatted on the
CD using the DVD format utility

This Page Intentionally Left Blank


Conquest Imaging - 99 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Conquest Imaging - 100 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™

Chapter 4
Adjustments

Conquest Imaging - 101 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 102 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ Adjustments

System Adjustments

Power Supply

There are no power supply voltage adjustments. The voltages can be


measured directly on the Acquisition or Platform Power Supply modules or
remotely using the UpLink application. Refer to “Troubleshooting”
section. If a voltage is out of tolerance, the system automatically shuts down
and cannot be turned on until the faulty PCB or assembly has been
removed.

Video Monitor Articulation, Elevation, Tilt and Swivel

Adjust video monitor articulation (elbow and knuckle) during system


installation, preventative maintenance, and after monitor replacement.
Monitor articulation requires adjustment when replacing cabling in the
monitor articulation arm.

Video Monitor elevation, tilt, and swivel

1. Release the travel lock.


2. Lift the video monitor. When the monitor is in a normal viewing
position, verify that it does not move up or down without an external force
(that it is neutrally buoyant).
3. If it is not neutrally buoyant, use a 3/16-inch allen wrench to
adjust the set screw on the articulation joint. Adjust the screw
clockwise to increase tension.
4. Grab the monitor from the bottom edge and adjust the tilt. Verify
that the monitor can be moved without difficulty and does not tilt on its
own weight.
5. If necessary, use a 5/32-inch Allen wrench to adjust the two set
screws on the tilt joint. Adjust the screw clockwise to increase
tension.
6. Grab the monitor on the left and right sides and swivel the monitor
from side to side. Verify that the monitor moves with the proper amount of
force.
7. If needed, adjust the swivel nut. Adjust clockwise to increase force.
Conquest Imaging - 103 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
3/16 inch allen head set screw adjustment

Video Monitor Articulation Arm Buoyancy Adjustment

5/32 inch allen head set screw adjustments

Video Monitor Tilt Adjustment


Conquest Imaging - 104 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
7/16 inch hex head nut adjustment.

Video Monitor Swivel Adjustment

Video Monitor Adjustments

There are no adjustments on the video monitor. However, the monitor


brightness default setting can be adjusted in the setups.
The Monitor control on the 2D touch screen allows you to adjust monitor
brightness to compensate for changes in ambient light. Use this control to
temporarily change monitor brightness. When you cycle power, the system
resets monitor brightness to the default value.

Setting video monitor brightness (power-up default value)

1. Press the Setup key.


2. Click System Settings.
3. Click the Display tab.
4. Select a setting for Monitor Brightness from 1 (darkest) to 5
(lightest).
5. Click Done.

Conquest Imaging - 105 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Adjusting the monitor for ambient lighting

1. Touch Next on the 2D touch screen.


2. Turn Monitor to select a setting from 1 (darkest) to 5 (lightest).

NOTE The Display tab may appear different on different software versions.
However, the location of the Brightness Setting remains the same.

Video Monitor to System Video Adjustment

Video Monitor This procedure calibrates the video monitor pixel clock
and A/D converters (R/G/B) to the Host AGP graphics board. Perform the
calibration only if the main display or AGP graphics board have been
replaced. The monitor must be functional (have power applied and display
video) before running the calibration. Allow the monitor to warm up for 5
minutes before performing the calibration.
When the calibration is executed, the system supplies a video pattern
consisting of alternating black and white pixels (gray screen) and a USB
command to synchronize this input. The system sends a 16-gradation

Conquest Imaging - 106 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
grayscale for display, with a USB command to set the A/D converters in the
video module.
The monitor blinks once during the end of the calibration process. Observe
the display for the blink. The calibration is considered successful if the
blink is verified.

To adjust the monitor to the system video card output

1. Remove the lower rear cover from the video monitor.


2. Connect a USB cable from the USB port on the back of the drive bay
to the USB connector on the video monitor. Some systems may have a
hublet connected to this connector. You can either disconnect the cable to
the hublet and connect to the connector on the back of the drive bay or
connect to the empty connector on the hublet.

NOTE Avoid using the USB connectors on the AVIO-RIP. Some systems
may have AVIO-RIP USB connections with versions earlier than 2.0.

3. Verify there is no external keyboard or mouse plugged in.

NOTE The procedure does not work if an external keyboard or mouse are
plugged in.

4. Power up the system. Allow the system to warm up for 5 minutes


before continuing.
5. Press the Setup key.
6. Click Config/Test.
7. Click the Tech/Admin tab.
8. Enter the user name and password.
9. Click Test/Utilities.
10. Click Platform. Click the arrow to expand.
11. Click Main Display. Click the arrow to expand.
12. Click Calibration.
13. Click Execute.
14. Click Screen Adjust.

Do Not Make any other selections

Conquest Imaging - 107 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
15. Click Auto Adjust. Two test patterns briefly appear. The calibration
is complete after the second pattern is displayed. (The first pattern is a fine
grid image; the display sets the pixel clock to match the system graphics
board. The second pattern is a grayscale pattern; the display sets the A/D
converters for optimum dynamic range.)
16. Disconnect the USB cable.
17. Reboot the system.

Touch Screen Calibration

Calibrating the touch screen

1. Press the Setup key.


2. Click Config/Test.
3. Click the Tech Admin tab.
4. Enter the user name and password.
5. Click the Test & Utilities tab.
6. Click Platform. Click the arrow to expand.
7. Click Touch Panel. Click the arrow to expand.
8. Click Calibration.
9. Click Execute. The touch screen calibration applet is displayed.
10. Touch the appropriate touch panel when prompted.
11. Touch in the center of each of the five targets that are displayed
sequentially on the touch screen. A message is displayed indicating the
number of pixels off for the calibration procedure.
12. Verify errors are less than 10 pixels. If errors are more than 10
pixels, repeat step 10 and step 11 until the touch screen is calibrated.
13. Click OK.
14. Click Quit.
15. Verify the calibration results:
a. Press Transducer.
b. Select a transducer and a preset.
c. If you are unable to select a transducer and a preset, reboot the system
and repeat step 1 through step 15.

Control Panel Lateral Translation Force Adjustment

The control panel lateral translation force and rotational torque settings are
adjusted simultaneously.
Conquest Imaging - 108 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Adjusting the lateral translation force and rotational torque setting

1. Remove the 10 screws securing the front pivot (orange) cover from
under the control panel.
2. Check the force necessary to rotate the control panel and adjust if
necessary.
3. Using an Allen wrench, adjust the screw shown in. Adjust for
customer satisfaction. There is no torque specification for this adjustment.
4. Replace cover.

Allen head adjustment screw.

Control Panel Lateral Translation Force Adjustment

Key Slide Catch Adjustment

The keyboard slide catch can be adjusted without removing any of the
system enclosures.

Conquest Imaging - 109 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Adjusting the keyboard slide catch

1. Pull the keyboard out from below the control panel.


2. Adjust the screw at the back of the keyboard drawer from
underneath the control panel.
3. Slide the keyboard in until it catches. It should catch and release
easily.

Slotted adjustment screw

Keyboard Slide Catch Adjustment

Control Panel Articulation Adjustment

Conquest Imaging - 110 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Adjusting the control panel articulation-cable tension

1. Adjust the control panel articulation trigger-release tension.


a. Raise the control panel articulation arm using the trigger release.
b. Remove the four screws securing the lower handle half to the
upper and remove the handle half.
c. Looking at the cable-adjustment nut from the non-trigger side of
the cable, adjust the nut clockwise to increase tension on the
cable or counterclockwise to decrease tension.
d. Adjust tension to the customer’s preference.
e. Replace the lower handle half.
2. Adjust the control panel articulation tension at the base of the
articulation arm.
a. Lower the control panel articulation arm using the trigger release.
b. Remove the top lid enclosure from over the base.
c. Using a wrench, adjust the nut clockwise to increase tension or
counterclockwise to decrease tension.
d. Adjust tension for operator preference.
e. Replace the top lid enclosure.
3. Using the articulation trigger release, check the articulation
adjustment by moving the control panel up and down.

Adjustment Nut

Control Panel Articulation Trigger Adjustment

Conquest Imaging - 111 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Hex adjustment nut

Control Panel Articulation Base Adjustment

System Brake Adjustments

Adjusing the system brake

1. Verify system power is off.


2. Chock the rear casters to prevent rolling.
3. Tilt the system to gain access to the brake mechanism.
4. Loosen the adjustment nut on the left brake arm. Adjust the arm in or
out so there are ten threads showing when the adjustment nut is
tightened.
5. Loosen the adjustment nut on the right brake arm. Adjust the arm in or
out so there are six threads showing when the adjustment nut is
tightened.

Conquest Imaging - 112 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
6. Lower the system to the floor.
7. Verify brake operation.
8. Repeat step 4 and step 5 until the operation is sufficient.

System Brake Adjustment

Conquest Imaging - 113 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 114 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™

Chapter 5
Preventative Maintenance

Conquest Imaging - 115 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 116 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ Preventative Maintenance

Introduction

Planned maintenance is recurring and must be performed at least every 6


months. Planned maintenance includes visual inspections, cleaning,
functional performance checks, and safety testing of the system and its
equipment.
Prior to visiting the customer site for planned maintenance, ask the customer
to archive the patient database and perform a disk defragmentation (where
applicable).
After performing planned maintenance, a Service Report and a Preventative
Maintenance Certificate must be completed. A copy of the completed
Service Report and a Preventative Maintenance Certificate must be provided
to the customer and to the Home Office.

Conquest Imaging - 117 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Some facilities have regulations that require planned maintenance more
frequently. It is the responsibility of the Customer Service Engineer (CSE),
to be aware of and adhere to all local regulations regarding planned
maintenance intervals for this system and its equipment.

WARNING - Some ultrasound parts may have had patient contact,


exposing them to blood borne pathogens. When handling parts that may
have had patient contact, take appropriate precautions against exposure to
blood-borne pathogens.

Conquest Imaging - 118 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Tools
 Standard field service tools
 Electrical Safety Analyzer Dale 601
 Ultrasound Phantom RMI 403GS, ATS 539 or similar
 ESD Vacuum
 Molygraph SL3144 Bearing Grease
 Micro-fiber or lint-free soft cloth.
 Liquid screen cleaner specifically designed for LCDs

Discuss Customer Concerns

1. Perform the following:


a) Discuss with the customer how the system has performed.
b) Find out about the concerns that the customer might have
regarding the system.
c) Fill out the appropriate section of the Preventative Maintenance
Certificate to document customer concerns and the actions taken
to resolve the issues.
d) Discuss with the customer whether there are unused studies
present on the system. If there are, work with the customer to either
send the unused studies to a storage device or delete the unused
studies.

Note: If the customer has concerns in regards to system performance, steps


may be required to address this. If there are no concerns, skip this step and
proceed to the next step.

System Maintenance

Fan/Air Filters The recommended interval for cleaning the filters is every 6
months, but periodic checks and cleaning should be done more often, if the
system is in a dusty environment. The customer can clean the filters.

Conquest Imaging - 119 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
To clean and replace the filters

1. Remove the left-side fan filter (power supply filter):


a. Remove the filter tray enclosure from the lower left side of the
system. The filter tray enclosure snaps into rubber grommets
and can be removed without tools.
b. Remove the filter from the filter tray.
2. Open the rear cover door and pull out the rear air filter.
3. Vacuum both filters and then clean them with soap and water.
4. If there are spare filters, install the spare filters into the system. If
there are no spares, reinstall dry and cleaned filters.

Note: Do not replace filters until thoroughly dry.

5. Reinstall the system covers.

NOTE On the power supply filter, the smaller end of the trapezoidal
shape goes toward the front of the system.

Casters The recommended interval for caster maintenance is every 12


months.

1. Check casters for cut or damaged tires.


2. Check for missing or damaged hubcaps.
3. Move the system and check caster operation.
4. Replace the casters if necessary.

Break and Swivel-lock The recommended interval for brake/swivel-lock


assembly maintenance is every 12 months. Both front casters can be brake-
locked. The right rear caster can be swivel-locked. The left rear caster
cannot be either. All four casters swivel when the swivel lock is off.

1. Check the brake pedal for smooth operation, and confirm that the
mechanism is functional
2. Check the caster swivel-lock pedal for smooth operation, and confirm
that the mechanism is functional.
3. Remove the mud plate enclosure.

Conquest Imaging - 120 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
4. Check the brake and swivel-lock cables. Verify there are no kinks or
other cable damage. Replace the cables if necessary.
5. Remove the left and right brake-pedal extension covers.
6. Apply Molygraph SL3144 Bearing Grease to the center brake linkage,
if needed. Move the system and check that the brake and swivel-lock
mechanisms are functional.

Control Panel The recommended interval for control panel articulation


maintenance is every 12 months.

1. Check the control panel articulation for smooth operation (up-down


and left-right).
2. Check the control panel side-to-side release mechanism for proper
operation.
a. Check control panel side-to-side movement. Adjust the tension
if necessary.
b. Check the control panel swivel movement.
3. Check the control panel vertical-release lever for proper operation.
Adjust the cable tension on the release lever if necessary.
4. Check the control panel articulation-arm movement. Adjust the
tension at the base of the arm if necessary.
5. Check that the keyboard slide mechanism is fully functional. Adjust
the slide catch if necessary.
6. Verify that the keyboard light turns on when the keyboard is pulled out
and turns off when the keyboard is pushed in. Adjust the keyboard if
necessary.

Trackball The recommended interval for cleaning the control panel trackball is
every 12 months, but periodic checks and cleaning should be done more often, if
the system is in a dusty environment. Check the control panel trackball for smooth
operation. Clean the trackball and the contacts in the trackball reservoir if needed.

1. Press down on the black rubber portion of the retainer ring with your
thumb and index finger.
2. Turn the retainer ring counterclockwise to loosen the ring.
3. Unscrew the ring and remove it.
4. Remove the trackball from the reservoir.

Conquest Imaging - 121 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
5. Clean dust and debris from the trackball reservoir with a cotton swab
moistened with alcohol.
6. Clean the shaft and rollers with a cotton swab moistened with alcohol.
7. Place the trackball in the trackball reservoir.
8. Carefully align the retaining ring to prevent cross-threading.
9. Turn the retaining ring clockwise until it is snug.

Monitor The recommended interval for monitor cleaning and articulation


maintenance is every 12 months.

Articulation
1. Check the monitor articulation mechanism for smooth operation.
2. Check that the monitor does not droop (tilt). Adjust the allen-head
screws at the tilt point.
3. Check that the monitor holds a stable vertical position (buoyancy)
during normal use. Adjust the tension on the jack screw at the
articulation joint.
4. Check the monitor swivel mechanism. Adjust the swivel nut at
the articulation arm swivel point.

Monitor and Touch screen cleaning

1. The video monitor and the touch screen have plastic front surfaces.
The surface is resistant to fluids normally found in clinical
environments, such as ultrasound gel, alcohol, and disinfectants, but
fluids spilled on the screens should be wiped off with a soft cloth to
avoid forming spots on the display surface.
2. The video monitor and touch screen should be cleaned as necessary.
For routine cleaning, use only cleaners specifically designated for
LCD screens. Disinfectants may be used when necessary, but they
are, are not recommended for routine screen cleaning.
3. Clean dust from the monitor and touch screen surfaces with a micro-
fiber or lint-free soft cloth. (Micro-fiber is recommended.)

Conquest Imaging - 122 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
4. Remove fingerprints or other marks on the screen using a liquid
screen cleaner specifically designed for LCDs. Spray directly on the
cleaning cloth or apply sparingly to the screen. Pre-moistened
screen wipes may also be used.
5. Dry the surfaces using a micro-fiber or lint-free cloth.

Note: • To avoid damage to the monitor screen and touch screen, do not
touch them with any sharp object such as pencils or calipers.
• Repeated use of common cleaners such as glass cleaner or
products containing alcohol may damage the monitor and touch-
screen surfaces and should not be used. Use cleaners specifically
designed for cleaning LCDs.
• Do not use paper towels to clean LCD surfaces; they may scratch
the surfaces. Use only a micro-fiber or other lint-free cloth.
Monitor Video Adjustments The Monitor control on the touch screen
allows you to adjust monitor brightness to compensate for changes in
ambient light. Use this control to temporarily change monitor brightness.
When you cycle power, the system resets monitor brightness to the default
value.

1.Press Setups.
2.Click System Settings.
3.Click the Display tab.
4.Select a setting for Monitor Brightness from 1 (darkest) to 5
(lightest).
5. Click Done.
To adjust the monitor for ambient lighting

1. Touch Next on the 2D touch screen.


2. Turn Monitor to select a setting from 1 (darkest) to 5 (lightest).

Clearing the Patient Directory During normal system operation, orphaned


patient files are left in the patient database (on the SIP drives). The number
of orphaned files grows with system use. The patient database must be
cleared during each preventive maintenance procedure to ensure that the
number of orphaned files never gets large enough to interfere with system
operation.

Conquest Imaging - 123 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Exporting Image to DVD

Patient images and image loops are stored on the SIP drives. The Review
Exam display allows you to export images in different formats so they can
be viewed on a personal computer. Single-frame images are exported in
JPEG format, and Cineloop sequences are exported in AVI format. When
images are exported in a compressed format, the system adds a label to
each image to indicate that the image has been compressed. The label
“JPEG” is added to the left of the image number on all images exported
from the current exam. The compression factor for exported images is
selected in the setups.

In JPG or AVI Format

1. Press Review. The patient directory is displayed.


2. Select the images you want to export.
3. Click To DVD as JPG/AVI in the Review Exam display.
4. If the Format DVD dialog box is displayed, click OK, click Format
in the Review Exam display, and follow the prompts. Then click To
DVD as JPG/AVI again.
5. In the Copy to DVD dialog box, select or create a directory in which
to store the images on the DVD.
6. Click OK to start copying. A message indicates when the images have
been copied.

In Dicom Format

1. Press Review. The patient directory is displayed.


2. Click Select All. The images are highlighted with reverse video.
3. Click Send To. The Send To dialog box is displayed.
4. Select DVD.
5. Insert a DVD into the drive.
6. Click OK.
7. Remove the DVD when copying has completed.

Conquest Imaging - 124 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Method 1

1. Power up the system.


2. Check with the customer and save selected exams to DVD.
3. Press Setup on the QWERTY top row.
4. Select Print/Network button on the left side of the monitor display.
5. Select the Global Config tab.
6. Select HD Maint. on the Global Config page
7. Click on Reset Patient Database or Reset HD.

Method 2

1. Power up the system.


2. Save all images that the customer wants to retain to the DVD.
3. Power down the system and press Ctrl+Alt+R as the system splash
screen is displayed.
4. Using the key generator in UpLink software, obtain a password and
log onto the MTE.
5. Log onto the command shell.
6. Type f and press Enter.
7. Type cd host and press Enter.
8. Type cd bin and press Enter.
9. Type dmmScript and press Enter. It takes several minutes to run
the script.
10. Type cleanLsr.bat and press Enter.
11.Power down the system using the circuit breaker and wait several
seconds.
12. Reboot the system.

Conquest Imaging - 125 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Backup System

Formatting a DVD

DVD and CD media are available in a number of types. Not all media types
are fully compatible with the system DVD drive. You can use the following
media types with the system:

 DVD+RW (re-writable, 4.7 GB capacity)


 DVD+R (write once, 4.7 GB capacity)
 CD-R (write once, 650 MB capacity)

The system supports multi-session writing to both DVD+RW and CD-R


media. This means that you can record information onto these discs in
multiple sessions, provided each session is closed before a new session begins.
The system automatically closes an open session when you end an exam and when
you click Eject in the Patient Directory. If you always use Eject in the Patient
Directory when ejecting discs, a CD-R recorded on the system will be compatible
with most CD-ROM drives found on PCs. Likewise, a DVD+R recorded on the
system will be compatible with most DVD drives.

1. Insert a DVD into the drive.


2. In the Review Exam display or the Patient Directory, click Format. If
the DVD is unformatted, an icon is displayed to indicate this.
3. In the Format DVD dialog box, click OK.

NOTE In some situations, (for example, if an unformatted disk is in the DVD drive
and the user elects to “Send a copy to DVD” in the Network Setups menu), the
DVD LED continues to flash after a disk is formatted and a study has been copied
to it. This operating scenario, although seemingly not normal, results in no harmful
effect. You may still store new exams to the disk.

Conquest Imaging - 126 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Copying Settings to a DVD If you want to back up or load only certain
settings, select and copy only those settings to a DVD. When you import settings,
all of the settings on the DVD will be loaded into the system.

1. Press Setups.
2. Click Config/Test.
3. Click the Manage Setups tab
4. Insert a DVD into the DVD drive.
5. If you want to copy all settings in all categories, click Select All and
continue with step 9.
6. Select any setups categories of settings you want to copy.
7. To select Quick Save presets, do one of the following:
- To select all Quick Save presets, select All for Transducer and click
Select Displayed.
- To select specific Quick Save presets, select a transducer and then
select the Quick Save presets you want to copy. Repeat this step
with other transducers.
8. To select voice profiles, do one of the following:
- To select all voice profiles, click Select All.
- To select specific voice profiles, select the ones you want to copy.
9. Click Copy to DVD.
10. If a dialog box indicates that the DVD needs to be formatted, click OK
in the dialog box, and then click Format and follow the prompts.
11.When a prompt indicates that the transfer is complete, press the eject
button on the drive, remove the DVD, and click Done.

Savings User Settings

This procedure saves all Tissue Specific presets, system settings, analysis
settings, annotation set tings, body markers, printer and network settings,
audio/video settings, and voice profiles. The settings must be saved and
restored prior to and after re-imaging or upgrading.

1. Power up the system.


2. Press Setups.
3. Click Config/Test.
4. Click the Manage Setups tab.
5. Click Select All or click the desired individual settings.
6. Insert a formatted DVD into the drive.
Conquest Imaging - 127 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
7. Click Copy to DVD.
8. Save the DVD for future use.

Importing Setting from DVD The system allows you to import all of the
settings on a DVD into your system.

1. Press Setups.
2. Click Config/Test.
3. Click the Manage Setups tab.
4. Insert a DVD containing settings into the drive.
5. Clik Import from DVD.
6. To select settings to import in the Import Setups dialog box, do one
of the following:
- If you want to import all settings in all categories, click Voice
Profiles, Quick Saves, and each of the setups categories, and
then continue with step 9.
- To select specific Quick Save presets or voice profiles, click the +
icon next to the category to expand the list and then select the
settings you want to import.
7. To import the selected settings, click OK.
8. To import the settings on the DVD, click OK.
9. When a prompt indicates that the transfer is complete, press the eject
button on the drive, remove the disc, and click Done.

Restoring User Setups

1. Perform step 1 through step 5 of “Savings User Settings”.


2. Insert the DVD with the saved user settings into the drive.
3. Click Import from DVD. An Import Setups dialog box is displayed.
4. Click OK.
5. Remove the DVD from the drive.

System Performance Tests

Begin a Patient Exam


1. Connect and initialize any transducer.
2. Press Patient Data and enter a patient name.
3. Click Close and return to 2D.
2D

Conquest Imaging - 128 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
1. Connect and initialize a curved transducer.
2. Enable AGC.
3. Adjust 2D gain to maximum.
4. Turn Depth to minimum.
5. Capture an image.
6. Turn Depth to maximum.
7. Adjust speed/IQ to minimum.
8. Adjust the focal zone position and range.
9. Enable Harmonic imaging.
10.Invert the image top to bottom.
11.Invert the image left to right.
12.Restore the image to the normal orientation.
13. Press Freeze.
14. Scroll through each frame in the Cineloop buffer.
15. Return to live imaging.
16.Reduce the sector width.
17.Steer the reduced sector left, right, and center.
18.Print an image to the hard disk.
19. Press Freeze.
20. Scroll through each frame in the Cineloop buffer.
21. Return to live imaging.
22.Expand the sector width to maximum.
23.Zoom the image.
24.Pan to display all areas of the image.
25.Change the zoom factor.
26.Pan to display all areas of the image.
27. Press Freeze.
28.Invert the image left to right.
29.Print an image to the hard disk.
30. Scroll through each frame in the Cineloop buffer.
31.Play the Cineloop images.
32.Adjust the Cineloop playback speed.
33.Pause Cineloop playback.
34. Scroll through each frame in the Cineloop buffer.
35. Return to live imaging.
36. Restore the image to the normal orientation.

Conquest Imaging - 129 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Sono CT

1. Touch SonoCT.
2. Verify that the button is active. The transducer orientation indicator
changes from a dot to a pair of triangles.
3. Turn Depth both directions.
4. Verify that the image and depth markers change accordingly.

CPA

1. Press 2D.
2. Press CPA. Verify that the CPA box appears in the image area.
3. Turn CPA clockwise to increase CPA gain until color noise is visible
within the box.
4. Use the trackball to move the CPA box to the four corners of the
image area.
5. Resize the CPA box. Vary the box size to its minimum and maximum.
6 . Press CPA to exit CPA mode.

2 D Color

1. Connect and initialize a linear transducer.


2. Enable Color mode.
3. Adjust the Color gain to fill the color box with color.
4. Adjust the color box position and size.
5. Print an image to the hard disk.
6. Press Freeze.
7. Scroll through each frame in the Cineloop buffer.
8. Return to live imaging.
9. Exit Color mode.

H D Zo o m

1. Enable the HD Zoom box.


2. Adjust the size and position of the HD Zoom box.
3. Zoom the HD Zoom box to fill the image area.
4. Print an image to the hard disk.
5. Adjust the zoom factor.
6. Un-zoom the image.

Conquest Imaging - 130 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Dua l

1. Enable Dual mode.


2. Enable the right image.
3. Press Freeze.
4. Scroll through each frame in the Cineloop buffer.
5. Return to live imaging.
6. Enable the left image.
7. Exit Dual mode.

PW Do ppl er

1. Connect and initialize a sector transducer.


2. Enable PW Doppler mode.
3. Adjust the position the M-line.
4. Press Update to activate scrolling Doppler.
5. Adjust PW Doppler gain to display Doppler data.
6. Print an image to the hard disk.
7. Enable Simultaneous mode.
8. Press Color.
9. Change the display format.
10.Enable HighQ automatic Doppler.
11.Enable Mean Trace.
12. Press Freeze.
13. Scroll through each frame in the Cineloop buffer.
14. Return to live imaging.
15. Return to 2D mode.

M- Mo de

1. Connect and initialize any transducer.


2. Press M-Mode.
3. Adjust the position of the M-Line.
4. Enable scrolling M-mode.
5. Adjust the position of the M-Line.
6. Print an image to the hard disk.

Conquest Imaging - 131 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
7. Press Freeze.
8. Scroll through each frame in the Cineloop buffer.
9. Return to live imaging.
10.Adjust the M-mode scrolling speed.
11.Return to 2D Mode.

Pan ora mic I magi ng

1. Connect and initialize any curved transducer.


2. Touch Panoramic.
3. Touch Start.
4. Acquire an image, moving the transducer as the image is being
acquired.
5. Touch Stop.
6. Touch Cine Review.
7. Use the trackball to scroll through frames of data.
8. Touch Cine Play.
9. Verify that each frame is played sequentially.
10. Exit Panoramic Imaging.

3D

1. Connect and initialize any transducer.


2. Touch 3D/4D.
3. Touch Start 3D.
4. Acquire an image, moving the transducer as the image is being
acquired.
5. Allow the buffer to fill with data.
6. Touch Stop 3D.
7. Rotate the image on each of the three axes.
8. Verify that the image rotates correctly.
9. Exit 3D.

Conquest Imaging - 132 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
I mage Ca pt ur e an d Con ne ct ivity

1. Press Patient Data.


2. Enter a patient name.
3. Click Close.
4. Press Print. Within a few seconds, you hear a very faint, high-
pitched beep. Listen carefully, it can be hard to hear.
5. Press Patient Data. The Pata Data formappears with the same
data you previously entered. If you did not enter an ID, one is
generated.
6. Press End Exam. All data on the Patient Data form is erased.
7. Click Close. The display returns to the same state as before you
displayed the Patient Data form. The patient name is not displayed in
the image header.
8. Press Review. The Patient Directory appears.
9. Select HD from the Drive menu. A list of saved exams appears in the
directory. It is finished when the status box at the bottom of the
display says, “Query Completed.” Verify that the patient you entered
is in the list, and that one image is associated with it (the Images col-
umn has a “1” in it).
10.Click Close.

Revi ew Exa ms

1. End any open (or paused) exams.


2. Press Review to display the Patient Directory.
3. Select the current exam.
4. Review each image in the directory.
5. Delete the exam
6. Return to 2D mode.

ECG If Present
1. Connect an S3-1 transducer
2. Connect the ECG cable to the system.
3. Set up the ECG simulator, 2.0 mV and 180 bpm.
4. Press Physio, then enable ECG.

Conquest Imaging - 133 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
5. Adjust the ECG gain, ECG position, and speed as necessary.
6. Verify that the ECG waveform is displayed correctly and that the
measured heart rate stabilizes at 180 ±5 bpm.
7. Disable ECG and press 2D.

Use r In te rfa ce Te st

1. In the setups, click Config/Test.


2. Click the Tech Admin tab.
3. Enter your user name and password.
4. Click Test and Utilities.
5. Run the QWERTY test.
6. Verify that a response is indicated for each control you press.
7. Verify that a response is indicated for each touch screen control.
NOTE Do not press the keys noted on the display.

S yst e m Co nfi de nc e T est The System Confidence Test or System Test


is a comprehensive test of the system operational status. This test comprises
numerous sub-tests: the System Functional Test, all of the Field Replaceable
Unit (FRU) Comprehensive Tests, and all of the Interface Tests. If a sub-test
fails, the system completes the remaining sub-tests. System Test execution
from this screen is available to The System Confidence Test is also accessible
from within the Tech Admin functions. Test execution from within Tech
Admin is limited to service personnel. If a sub-test fails, a failure code is
displayed.
The System Confidence Test should be run any time a system error is
displayed, or if the user suspects problems with the system. If an error
message is displayed during the test, reboot the system with the On/Standby
switch. This test can take up to 30 minutes.

1. Power up the system.


2. Verify there are no transducers connected.
3. Press Setups.
4. Click Config/Test.
5. Click the System Test tab.
6. Click Run. The system displays a message when the test is complete,
telling you whether the test passed or failed.
7. Click Cancel or Done to exit the setups.

Conquest Imaging - 134 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Pe rip he rals - VCR’s , Prin te rs et c .

1 . Tes t for no r ma l ope ra tion.

S af ety Ch ec ks

Complete Safety Checks as mandated by the facility.

Conquest Imaging - 135 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 136 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™

Chapter 6
Troubleshooting

Conquest Imaging - 137 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 138 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ Troubleshooting

Troubleshooting

The system contains a number of on-board diagnostic tools. Description and


use of these tools are provided in the Tech Admin Help. However, if the
system does not boot up, the on-board diagnostic tools may not be
accessible. The information in this section may be used to troubleshoot non-
booting systems to allow access to the diagnostics.
Some system faults that cause nonboot conditions may be corrected using
procedures detailed in System Administration.

Troubleshooting Procedure

Note: Never attach or disconnect probes during power-up.

Power Supplies

The Platform Power Supply requires the ACQ_DC_OK signal from the
Acquisition Power Supply to be 0 volts or less than 2.2 Vdc (logic low)
before it powers up. The Platform Power Supply remains off when
ACQ_DC_OK is between 2.2 Vdc and 6 Vdc. The signal is approximately
4 Vdc if the Acquisition Power Supply is connected and not in a
DC_OK_OUT state. This logic allows the Platform supply to power up with
or without the Acquisition Power Supply connected and yet stay shut down
if the Acquisition supply is connected and faulty.
Once powered up with valid output levels, the platform supply provides a
DC_OK_OUT signal to both motherboards, allowing each motherboard
BIOS to run and configure motherboard hardware.
Either motherboard can energize the supply, but either motherboard can
present an over-current load. A jumper (JP2) on the Power Distribution
Board (Platform Card Cage) during troubleshooting that can force the
PS_ON_n signal active. This is useful if one of the motherboards is
suspected of preventing the Platform Power Supply from powering up.
Any condition causing the power supplies to shut down (over-voltage or
over-current) requires the AC breaker to be turned off (or the power cord
disconnected) for a minimum of 1 minute. That allows capacitors within the
supply to drain, removing the internal fault latch voltage.

Conquest Imaging - 139 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
When troubleshooting the power supplies, disconnect P1 (24-pin) and P3 (8-
pin) connectors of cable 453561182601 at the motherboards instead of
removing the connection at the Power Distribution Board. Removing this
cable end allows remote sense signal lines to remain connected to the power
supply output and allows the Platform supply to provide proper voltage
regulation. Power cables may be disconnected from one or both
motherboards for troubleshooting purposes.

FEC Configuration Bypass

Perform the FEC configuration bypass procedure if you suspect that the
system did not boot up because the FEC local processor did not configure the
AVIO-CAD bridge for I2O messaging (no beeps from the Host
Motherboard during the bootup process).
Performing the FEC configuration bypass procedure
1. Power down the system by turning off the main breaker and wait
several seconds.
2. Turn the main breaker back on.
3. Using a nonconductive implement, press the push-button switch
located behind the small hole at the bottom of the AVIO-RIP
faceplate. This activates a hardware latch to disable the I2O
messaging unit and ensures the AVIO-CAD FEC PCI bridge responds
to PCI configuration cycles on the next power up, regardless of the
operational state of the FEC.

NOTE: Nonconductive implements include a stiff wire with insulation still


on, a paper clip wrapped with tape, or a plastic tool designed specifically for
the job.

4. Power up the system.


5. Observe the power-up sequence to determine if the procedure
affected the BIOS beep behavior.

NOTE: Each power cycle resets the FEC configuration bypass latch, so the
above sequence must be repeated each time you want a power-up cycle
without FEC I2O messaging.

If the BIOS beep behavior changes, or if the boot sequence goes further,
refer to Use the FEC debug port to determine if the front end is faulty.
Conquest Imaging - 140 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Use the FEC debug port to determine if the front end is faulty

The FEC debug port may be used to determine the configuration of the front
end PCBs (including part numbers), the version number of front end ASICs,
and if there are faults in the Acquisition Subsystem. The detectable faults
include power monitor faults and hardware configuration errors. The port
also reports the checksums for the Channel Board flash. The system should
remain off until the proper setup has occurred.

1. Refer to drawing and connect the system to the laptop.


2. On the laptop, click Start, Programs, Accessories, Communications,
HyperTerminal.
3. If the new connection dialog box appears, enter a name and click OK.
4. Configure the HyperTerminal settings as indicated in table.

Parameter Setting
Connect To tab
Connect using COM1
COM1 Properties
Bits per second 19200
Data bits 8
Parity None
Stop bits 1
Flow control Xon/Xoff

5. Click OK to launch HyperTerminal.


6. Power up the system. The front end status is displayed in the
HyperTerminal window.

Conquest Imaging - 141 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Connecting your Laptop

Laptop Serial Port Cable

P1 P2 Signal
P1-1 P2-1 (GND)
P1-2 P2-2 (Tx)
P1-3 P2-3 (Rx)
P1-4
P1-5 P2-4 (GND)
P1-6
P1-7
P1-8
P1-9

Conquest Imaging - 142 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Remote DOS Commands

Command Description
ASSOC Allows you to view the file association
ATTRIB Displays and changes file attributes
CD Changes directories
CHDIR Changes directories
Checks the hard drive running FAT for
CHKDISK errors
Checks the hard drive running NTFS for
CHKNTFS errors
CLS Clears the screen
COMP Compares files
Copies one or more files to an alternate
COPY location
DATE Views or changes the system date
Consolidates information on a hard drive
DEFRAG to help with data/program storage
DEL Deletes one or more files
Deletes a file (recovery console
DELETE command)
Lists the contents of one or more
DIR directories
FIND Searches for text within a file
FINDSTR Searches for a string of text within a file
Displays or modifies file types used in
FTYPE file extension associations
Displays a list of commands along with
HELP a brief explanation
Allows you to view network adapter
settings and assigned values (network
IPCONFIG command)
MD Creates a new directory
MEM Displays memory on the system
MKDIR Creates a new directory

Conquest Imaging - 143 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Command Description
Moves one or more files from one
MOVE directory to another directory
Updates, fixes, or allows you to view the
NET network or network settings
Tests/sends information to another
PING network computer or network device
Starts a separate window in the
Windows operating system from the
START MS-DOS prompt
Allows you to use Telnet to access
another computer/device from the
TELNET prompt
Allows you to view or modify system
TIME time
TRACERT Allows you to view network packets
routed across a network
TYPE Displays the contents of a file
VER Displays the version information
VOL Displays volume information about the
designated drive

Power Input Requirements

Determining Electrical Power Requirements

1. Explain to the facilities representative that the system requires a


dedicated and isolated line meeting the power requirements as listed
in tables below for A.x systems or for B.0 systems.
2. Verify the outlet to be used is a dedicated and isolated line:
a. Have the facility representative open the circuit breaker.
b. With a DVM or three-prong test plug, confirm that power is still
applied to all other outlets and hardwired devices in the immediate
area and that power is removed only from the dedicated and
isolated outlet box.
3. Verify the proper wiring of the outlet using a three-prong test plug.
This plug also tests for the presence of a ground, but it cannot detect
neutral/ground reversals.
Conquest Imaging - 144 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
4. Inform the facility representative of the current draw of the system
and its acceptable voltage range. It is the responsibility of the facility
representative to determine the proper size of the wire based upon the
length of the feed.
5. Using your DVM at the outlet, measure the hot/neutral, hot/ground,
and neutral/ground voltages with a comparable load (for example,
another ultrasound system) on the line.
6. To determine line loss, if a facility representative is available:
a. Have the facility representative measure the voltage at the line’s
circuit breaker primary in the distribution sub-panel.
b. Measure the voltage at the wall outlet.
c. Calculate the percentage of voltage loss across the
line using this formula: Percent line loss = [(A -
B)/A] x 100 where A and B are the voltages at the
circuit breaker and wall outlet with the system on,
respectively.
d. Continue with step 9.
7. If a facility representative is not available:
a. Measure the voltage at the outlet with the load on and off.
b. Calculate the percentage of voltage loss across the line using the
formula: Percent line loss = [(A - B)/A] x 100, where A is the
voltage at the wall outlet with the system off and B is the voltage
at the wall outlet with the system on.
8. Use an ONEAC line viewer and oscilloscope to measure power-line
noise. Repeat with a load applied, for example, another ultrasound
system.
9. Record your power quality observations and measurements.
10.Review the electrical site evaluation with the facility representative.

Conquest Imaging - 145 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Specifications (A.x Systems)

Measurement
100 Vac 120 Vac 230 Vac
(System Only)

Hot/Neutral 100 Vac 120 Vac 230 Vac

Current Draw 7.44/7.67 6.15/6.34 3.24/3.36


(normal/peak)1 A A A
Power Draw 741/764 737/760 746/773
(normal/peak)1 VA VA VA
734/754 727/750 721/743
W W W
< 5-Vac < 5-Vac < 5-Vac
Neutral-to-Ground
RMS RMS RMS

Line Loss < 3% < 3% < 3%

Distortion < 20% < 20% < 20%

Common Mode < 1.5 Vp- < 1.5 Vp- < 1.5 Vp-
Noise p p p
Normal Mode
< 10 Vp-p < 10 Vp-p < 10 Vp-p
Noise

Current draw and power draw specifications are for the rated voltage ranges. (The
“normal” specification includes system operation with the VCR recording, all
other OEMs at the idle state, and with the system scanning. The “peak”
specification includes system operation with OEMs.)

Conquest Imaging - 146 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Specifications (B.0 Systems)

Measurement
100 Vac 120 Vac 230 Vac
(System Only)
Hot/Neutral 100 Vac 120 Vac 230 Vac
Current Draw
8.5/9.6 A 7.2/7.7 A 3.64/4.0 A
(normal/peak)1
Power Draw 850/950 840/940 840/935
(normal/peak)1 VA VA VA
850/950 830/930 800/900
W W W
< 5-Vac < 5-Vac < 5-Vac
Neutral-to-Ground
RMS RMS RMS
Line Loss < 3% < 3% < 3%

Distortion < 20% < 20% < 20%


Common Mode < 1.5 Vp- < 1.5 Vp- < 1.5 Vp-
Noise p p p
Normal Mode
< 10 Vp-p < 10 Vp-p < 10 Vp-p
Noise

Current draw and power draw specifications are for the rated voltage ranges. (The
“normal” specification includes system operation with the VCR recording, all
other OEMs at the idle state, and with the system scanning. The “peak”
specification includes system operation with OEMs.)

Conquest Imaging - 147 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Supply Data

The system software monitors the power-supply enable voltages to ensure


they are within tolerance. If the voltages are not within the tolerances, the
system shuts down. If the system shuts down, none of the voltages are
enabled (all voltages are 0 V). System power LED status is indicated in
System Power LED Status table. If the power supply is fully operational
and a cable is bad, you measure voltages on the individual PCBs.
On systems with B.0 hardware additional troubleshooting information is
available via LEDs on the Acquisition Power Distribution Board. table for
system power status indicated by the APDB LEDs and for fault indications
and corrective actions.
The software monitors and logs the voltages but does not shut down the
system. The power supply hardware shuts down the system but does not
monitor or log voltages. That is, the voltage log may record voltages that are
out of specification before the system shuts down, or the hardware may shut
the system down before an entry has been made in the log.

Conquest Imaging - 148 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
System Power LED Status

A.x Systems B.0 Systems


System Power AC Tray On/Standby AC Tray On/Standby
State Green Green LED2 Blue Green LED2
LED1 LED1
Off and no Off Off Off Off
AC power
applied
Off and AC On Off On Off
is applied

Standby3 with On Blinking Off Blinking


no AC applied

Standby3 with On Blinking Off Blinking


AC applied

Operating On On On On

AC applied, On Off Blinking Off


power supply

1. On the rear of the system where the power cord connects.


2. On the front of the system next to the switch.
3. Standby mode (suspend to RAM). Processor is not processing. +5 VSB from the
battery supplies power to memory.

Conquest Imaging - 149 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Acquisition Card Cage Voltages

Voltage Voltage Range Monitored


(Vdc) By
+5.35 5.0 to 5.7 AIM+
-5.35 -5.0 to -5.7 AIM+
+3.65 3.4 to 3.9 AIM+
+15 14.0 to 16.1 AIM+
-15 -14.0 to -16.1 AIM+
+48 44.6 to 51.4 AIM+
+HV_IN 0 to 125 AIM+
-HV_IN 0 to -125 AIM+
+HV_OUT (pulser 0 to 120 AIM+
-HV_OUT out) (pulser 0 to -120 AIM+
out) out)
+LV (pulser 0 to 12 AIM+
-LV (pulser out) 0 to -12 AIM+
TGC1, TGC2 -1.4 to 1.4 AIM+
HV Bias 0 to -HV_OUT AIM+
VX1 1.5 to 5 AIM+
VX2 1.5 to 5 AIM+
VX3+ 5 to 12 AIM+
VX3- -5 to -12 AIM+
Matrix TGC 0 to 12 AIM+
Matrix VREF 0 to 12 AIM+
+48 44.6 to 51.4 FEC

Conquest Imaging - 150 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Subsystem LED Status (B.0 Systems)

Board or FRU LED1 ON/Standby Description


Acquisition
Power
DS11 Green/Green +5 VSB from AC tray is okay
Distribution
Board
Power-on signal from
ON/Standby switch and
DS10 Green/Off
motherboards is okay
(PS_ON_n)
AC is applied to APS and PPS
DS1 Green/Off and is within specification. The
relay drive signal is sent to the
AC tray to close the relay
APS and PPS are okay. All
DS4 Green/Off system bus (SPS OK) within
ranges
Platform Power Distribution
Board and all platform power are
DS12 Green/Off
okay. (PLT_POWER_GOOD
signal, PPDB_OK, is okay.)

DC_OK_OUT signal is okay.


DS7 Green/Off
Allows motherboards to boot
DS2 Off/Red Not for field use
DS3 Off/Red Not for field use
DS5 Off/Red Not for field use
DS6 Off/Red Not for field use
DS8 Off/Red Not for field use
DS9 Off/Red Not for field use
Platform Power
Distribution All Red/Red Not for field use
Board
AC AC, +12 VSB, and +5 VSB are
AC Tray Blue/Blue
Tray present. Batteries are charging

1. LEDs are listed in the order they are lit on a normally operating system.

Conquest Imaging - 151 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Acquisition Power Distribution Board LED Locations (B.0 Systems)

Acquisition Power Distribution Board LED Locations (B.0 Systems)

Conquest Imaging - 152 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Power Subsystem LED Fault Indications and Corrective Actions (B.0
Systems)

AC Possible
DS4 - DS12 - DS7 - Tray - Location Corrective
SPS_OK PPDB_OK DC_OK_OUT Blue of Action
LED Failure
Replace
Power
On On Off On APDB
Supply
Assy1
Replace
Power
Power
Off On Off On Supply
Supply
Assy
Assy
Replace
On Off Off On PPDB
PPDB
Replace
Power
Off Off On On APDB
Supply
Assy
Replace
Power
Off On On On APDB
Supply
Assy
Replace
Power
On Off On On APDB
Supply
Assy
1. Close
1. No
circuit
Off Off Off Off AC
breaker
power
and
2. AC check
Tray power
2. Replace
AC Tray

Conquest Imaging - 153 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Possible AC
DS4 - DS12 - DS7 - Tray -
Location Corrective
SPS_OK PPDB_OK DC_OK_OUT of Blue
Action
Failure LED
Replace
Power
On On On Blinking APDB
Supply
Assy
Replace
Power
Any Power
Blinking Supply
combination Supply
Assy
Assy

1. The Power Supply Assy comprises the Acquisition Power Supply, the
Platform Power Supply, and the Acquisition Power Distribution Board.
2. Refer to LED locations on previous illustrations.

Hub Plus PCB LEDs

The Hub Plus PCB LEDs provide information valuable to troubleshooting


the board. The LED locations are shown in Figure 9-3 and the LED
indications are provided in Table 9-8

Conquest Imaging - 154 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Hub Plus PCB LED Indications (B.0 Systems)

LED Indication Notes


DS1 When on, indicates +5 Located in the DVD bay
VSB power is present on near the rear panel, the
the board. LED is status can be seen from the
momentarily off after rear of the system through
depression of the the fifth vent hole from the
On/Standby switch left in the bottom row.
DS1 and DS2 are both
visible through the same
vent hole. DS1 is on the
edge of the board closest
to the rear of the system
DS2 When on, indicates that Located in the DVD bay
+3.3 Vdc is present on the near the rear of the DVD
board. The presence of drive, status can be seen
+3.3 Vdc indicates DVD from the rear of the system
drive activity. through the fifth vent hole
from the left in the bottom
row. DS1 and DS2 are
both visible through the
same vent hole. DS2 is
located on the edge of the
board closest to the front
of the system
DS3 Indicates USB bus Check LED status by
activity. Momentarily removing the left DVD
comes on upon activation drive cover screw and
of the On/Standby switch, observing through the
then goes off, then comes screw hole
on once the system is fully
booted up. Stays on until
power down. Indicates that
the On/Standby switch has
been pressed and the signal
has been sent to the
motherboards.
Temperature Monitoring
Conquest Imaging - 155 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Temperature monitoring occurs on PCBs in the Acquisition and Platform
card cages (Table 9-9). Temperatures are logged in the SystemMonitorLog
during system powerup. The system warms up for approximately 10 minutes
before the temperatures are logged again. If the temperature at each sensor
rises above a set point, the temperature is logged to the SystemMonitorLog
and the system fan speed increases. If the temperature continues to rise above
the set point and reaches the alarm setting, the temperature is logged into the
ErrorLog instead. Once this happens, an error message notifies the user of
the over-temperature condition and that the system will shut down in 30
minutes. If the user has not shut the system down after the warning, the
system is shut down by the error handler.

NOTE Systems with 1.1.0.138 software intermittently fail the Fan


Controller Test in the System Confidence Test. If you suspect a temperature
problem and the Fan Controller Test fails, do not replace parts until you
reboot the system, and re-run the Confidence test (the test consistently passes
on a rebooted system). Review the SystemMonitorLog and ErrorLog to
verify that there are no temperature errors. If the Fan Controller Test fails for
this issue, there is no thermal danger to the patient or to the system.

On PCBs that have input and output airflow sensors (Channel Boards and
AIM+), if the difference between the two sensor readings is greater than
25°C (77°F), the temperatures are logged to the SystemMonitorLog every 10
seconds.
The rate of change from the previously logged temperature is also
monitored. If the last logged value is 50, and the rate of change is set at 5, the
temperature is logged in the SystemMonitorLog when the temperature is
45°C (113°F) or 55°C (131°F). The amount of time between logged values
depends only on how long it takes for the temperature to vary by the set
amount. It may be 2 minutes, 2 days, or longer.

Conquest Imaging - 156 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
PCB Temperature Sensor Information

PCB Temperature Fan Temp Alarm Temp Rate of


Sensor Change
AVIO RIP 43°C 67°C 5°C

AVIO CAD 47°C 72°C 5°C

Host Motherboard 1 1 1

SIP Motherboard 1 1 1

FEC 54°C 75°C 5°C

AIM+ top sensors 47°C 70°C 5°C

AIM+ bottom 54°C 75°C 5°C


sensors
Channel Boards 0–3, 56°C 75°C 5°C
top
Channel Boards 0–3, 40°C 60°C 5°C
bottom
Channel Board - - 25°C
temperature
difference

Conquest Imaging - 157 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Abnormal On/Standby LED States (A.x Systems)

Pressing and holding the On/Standby switch for more than 5 seconds,
anytime the system is operating or booting, forces an abnormal shutdown
sequence on the power sub-system and leaves the On/Standby LED in an
indeterminate state (usually ON). To preclude this from happening, use the
On/Standby switch as a momentary control (pressing it for no more than 1 or
2 seconds). If the system does not respond to this input, turn off the AC
breaker on the lower rear of the system for 1 minute and then back on again.
This ensures that the Host and SIP Motherboards are placed in the same
internal power state for proper power sequencing.
Also, wait at least 2 seconds after using the On/Standby switch to turn the
system off before turning it on again. Not doing so may result in the SIP
failing to initialize on bootup.

NOTE This issue is not a problem on systems with B.0 hardware and above.

Conquest Imaging - 158 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Diagnostic Codes and Descriptions

Description / Service
Code Condition
Consequence Responsibility
EHA_UNIT_TEST EHA Unit Test System fault and No action
recovery are part of a
unit test of the Error
Handling Application.
The unit test that uses
this is a TestRunner
application named
ehaTest.tfe.
TBD Diagnostic A fault has occurred Report this fault as
Code to Be that has not been a build
Determined assigned a diagnostic discrepancy.
code. This code
should occur only in
engineeringsupervised
and developmental
systems.
000 Information An informational No action
Only dialog box has been
displayed as part of
normal operation.
001 System Over One of several Check the platform
Temperature internal temperature log for temperature
sensors indicates that sensor readings.
the system is too hot Check room
to continue operation. temperature and
The default response for air-fan
is to alert the user and blockage. Check
shut down the system the preventative
in 30 minutes if the maintenance
user does not do so schedule.
manually.

Conquest Imaging - 159 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
003 Logging The logging service is Use the UpLink
Obstructed no longer fully Remote Service
functioning. The Application to
default log may be in transfer the logs to
use. The user is not Bothell.
informed of this
missing service and
the workflow is not
interrupted. The user
may proceed without
logging.
004 Power The EHA cannot turn There may be a
Dissipation down the front-end problem with the
Fault pulser voltage enough users’
to allow scanning understanding of
while leaving other the controls or
controls alone. The choice of
user must reselect the transducer or
transducer. application. User
may have noticed a
loss in signal
before the dialog
box was displayed.
Educate user as
toproper use of the
system.

Conquest Imaging - 160 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
005 Acoustic Power The system has There may be a
Monitor Fault measured that the hardware problem
transducer and pulsers with the transducer
are consuming too or front end. Run
much power. The the on-board self
default response is to tests. On 1.1.1.147
discontinue scanning systems, this code
and alert the user that sometimes appears
the transducer might when
be defective. disconnecting a
selected
multiplexed
transducer (3D6-2,
C9-4, L12-5 50, or
L17-5). Review
the error log to
determine if the
error occurs during
other instances.
006 Transducer The system has There may be a
Over measured that an hardware problem
Temperature intracavity transducer with the transducer
is too hot to continue or front end. Run
operation. The the on-board self
automated response is tests.
to pause scanning and
alert the user of the
situation. If the
condition persists,
further scanning is not
allowed.
007 Thermal Circuit The system has There may be a
Error detected a fault in the hardware problem
transducer with the transducer
temperature sensing or the front end.
circuit. The default Move the
response is to mark transducer to
the transducer as not another connector.
selectable on the If the problem
touch screen and alert follows, replace
the user that the the transducer.
transducer may be
damaged.

Conquest Imaging - 161 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
008 License An expired option Determine if the
Expired was detected on boot user wants to
up. The system will purchase the
be rebooted with this expired option.
option disabled. An
expiration dialog box
was displayed
followed by a reboot
dialog box.
009 CRC Check While booting, a Follow installation
Failed checksum did not procedure. Reboot
match in the system or use the
configuration files. setups to identify
the offending file
from the Error
Log. Then
navigate to
Install/Upgrade
and restore the file.
010 CRC Check While booting, a Follow installation
Failed checksum did not procedure. Reboot
match in the system or use the
configuration files. setups to identify
the offending file
from the Error
Log. Then
navigate to
Install/Upgrade
and restore the file.
011 SN Mismatch The system serial Follow installation
number does not procedure.
match that contained
in the licensed options
file.
BOM File The system bill of Follow installation
Open Error materials file could procedure. Reboot
not be opened. the system or use
the setups and
navigate to
Install/Upgrade.

Conquest Imaging - 162 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
013 AIC File Open The allowed internal Follow installation
Error capabilities file could procedure. Reboot
not be opened. the system or use
the setups and
navigate to
Install/Upgrade.
014 LO File Open The licensed options Follow installation
Error file could not be procedure. Reboot
opened. the system or use
the setups and
navigate to
Install/Upgrade
and Manage
Licensed Options
File. Restore the
licensed options
file.
015 LO File Open Unspecified error Follow installation
Error occurred while procedure. Reboot
opening licensed the system or use
options file. the setups and
navigate to
Install/Upgrade
and Manage
Licensed Options
File. Restore the
licensed options
file.
016 3D Calibration Motor controller Could be
Failed failure during transducer cap
calibration of deformation,
mechanical 3D motor or encoder
transducer. circuit or motor
controller. See
Error Log for
originating error.

Conquest Imaging - 163 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
017 3D Motion A motion failure was Could be
Fault While reported by the front transducer cap
Scanning end during operation deformation,
of mechanical 3D motor/encoder
transducer. circuit connections
or motor
controller. See
Error Log for
originating error.
018 SIP Halted SIP halted volume Restart system and
Volume reconstruction rescan.
Reconstruction because the elevation
position data was
inaccurate.
019 MC/FEC FEC could not None.
Communication communicate with
Failure motor controller
during operation of a
mechanical 3D
transducer.
020 Hardware The installed Press Setup to
Mismatch hardware review hardware
configuration is not mismatch in Error
an allowed Log or to
configuration for the reconfigure
currently installed software or
software. hardware, or
reboot with the
installation disk.
021 Motor The motor controller Replace motor
Controller indicated that it did controller.
Timeout not keep up with or
complete a motion
profile in the time
allotted.
022 Front End FEC sensed a HV May be a failed
Current over-current when Channel Board or
Calibration applying a test load other front-end
Failure located on AIM+ at field-replaceable
power up or unit. Replace
transducer select. suspect unit.
Scanning was
terminated.

Conquest Imaging - 164 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
023 MUX Over FEC sensed that the The transducer
Current MUX chips on the may be damaged.
transducer Replace
multiplexer FRU are transducer.
drawing too much
current from the HV
supply. Scanning was
terminated.
024 HQ Out of The measurement tool Reselect the
Sync could not calculate transducer and
HQ indices on the acquire new data.
Doppler data.
025 Transducer The transducer in the The user may have
Mismatch - not port about to be changed
selected activated has not been transducers while
or is no longer the system was
selected. selecting. Could
also be the
transducer
EEPROM or
S/HSEL relay.
026 Transducer There is no longer a The user may have
Mismatch - not transducer in the port changed
connected about to be activated. transducers while
the system was
selecting. Could
also be the
transducer
EEPROM or
S/HSEL relay.
027 Transducer The DAT file for the The user may have
Mismatch - transducer in the port changed
DAT file about to be activated transducers while
does not match what the system was
the system has set up. selecting. Could
also be the
transducer
EEPROM or
S/HSEL relay.

Conquest Imaging - 165 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
028 Transducer The CTL file for the The user may have
Mismatch - transducer in the port changed
CTL file about to be activated transducers while
does not match what the system was
the system has set up. selecting. Could
also be the
transducer
EEPROM or
S/HSEL relay.
029 Transducer The relays for the port The user may have
Mismatch - about to be activated changed
Relay not are not selected. transducers while
selected the system was
selecting. Could
also be the
transducer
EEPROM or
S/HSEL relay.
30 Motor Motor controller boot Disconnect and
Controller Boot error was reported at reconnect
Error S/HSEL module. The transducer
transducer is not
selectable.
31 FCC Timeout Frame composition Reselect
domain timed out transducer to clear
waiting for signal the problem.
path to start after a
control change.
32 FCC Frame composition Reselect
Consistency domain failed its transducer to clear
Check Failed consistency check. the problem.
33 DSC Low DSC detected an Reselect
Level fault error. transducer to clear
the problem.

Conquest Imaging - 166 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
200 Unanticipated The system has Filter Error Log to
Serialized Fault detected a fault for "DiagCode" and
- Restart which there is no find 200. See prior
Required recovery. It has been errors in Error
passed to EHA as a Log. Use UpLink
serialized exception. software to
transfer the logs to
Bothell. Restart the
system with the
On/Standby
switch.
201 FEC failed with The Front End Filter Error Log to
error code 1 - Controller has "DiagCode" and
Restart reported an error for find 201. See prior
Required which there is no FEC error
recovery. Scanning description in
has stopped. Error Log. Use
UpLink software
to transfer the logs
to Bothell. Restart
the system with the
On/Standby
switch.
202 FEC failed with The Front End Filter Error Log to
error code 2 - Controller has "DiagCode" and
Restart reported an error for find 202. See prior
Required which there is no FEC error
recovery. Scanning description in
has stopped. Error Log. Use
UpLink software
to transfer the logs
to Bothell. Restart
the system with the
On/Standby
switch.

Conquest Imaging - 167 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
203 FEC failed with The Front End Filter Error Log to
some other Controller has "DiagCode" and
error code - reported an error for find 203. See prior
Restart which there is no FEC error
Required recovery. Scanning description in
has stopped. Error Log. Use
UpLink software
to transfer the logs
to Bothell. Restart
the system with the
On/Standby
switch.
204 ACQC The Acquisition Filter Error Log to
unanticipated Control domain has "DiagCode" and
error - Restart reported an error for find 204. See prior
Required which there is no errors in Error
recovery. Scanning Log. Use UpLink
has stopped. software to
transfer the logs to
Bothell. Restart the
system with the
On/Standby
switch.
205 ACQC bridge The Acquisition Filter Error Log to
Server request Control domain has "DiagCode" and
bound overflow reported an overflow find 205. See prior
- Restart of messages and can errors in Error
Required not recover. Scanning Log. Use UpLink
has stopped. software to
transfer the logs to
Bothell. Restart the
system with the
On/Standby
switch.

Conquest Imaging - 168 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
206 SIP exception The SIP has reported Filter Error Log to
with no an error for which "DiagCode" and
recovery - there is no recovery. find 206. See prior
Restart Scanning has stopped. errors in Error
Required Log. Use UpLink
software to
transfer the logs to
Bothell. Restart the
system with the
On/Standby
switch.

Conquest Imaging - 169 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
207 APPUI The Application User Filter Error Log to
unanticipated Interface domain has "DiagCode" and
error - Restart reported an error for find 207. See prior
Required which there is no errors in Error
recovery. Scanning Log. Use UpLink
has stopped. software to
transfer the logs to
Bothell. Restart the
system with the
On/Standby
switch.About 40
software
exceptions are
responsible for all
207 errors.There is
no hardware fault
associated with
this error code.
When a 207 error
is reported, review
the error log for
hardware errors
immediately
preceding the error
and address that
hardware. If no
error is reported,
do not replace
hardware; inform
the customer that
this error will
likely be addressed
by a future
software build.
Download error
logs using the
UpLink
application and
send to Bothell. If
necessary, contact
RTAC to have logs
reviewed for
confirmation.There
is no hardware
Conquest Imaging fault associated- 170 -
Ultrasound Sales Service Repair Training with this error
1815 Industrial Drive, Suite 100 Stockton, CA 95206code. Do not
PH: 209-942-2654 FAX: 209-942-2572 replace hardware.
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
208 FCC The Frame Filter Error Log to
unanticipated Composition domain "DiagCode" and
error - Restart has reported an error find 208. See prior
Required for which there is no errors in Error
recovery. Scanning Log. Use UpLink
has stopped. software to
transfer the logs to
Bothell.
209 DMM The DMM domain Filter Error Log to
unanticipated has reported an error "DiagCode" and
error - Restart while saving patient find 209. See prior
Required data into the database errors in Error
for which there is no Log. Use UpLink
recovery. Restart the software to
system with the transfer the logs to
On/Standby switch. Bothell. Restart the
system with the
On/Standby
switch.
210 DMM The DMM domain Filter Error Log to
unanticipated has reported an error "DiagCode" and
error - Restart while updating patient find 210. See prior
Required data into the database errors in Error
for which there is no Log. Use UpLink
recovery. software to
transfer the logs to
Bothell. Restart the
system with the
On/Standby
switch.
211 Unhandled A Boris process has Filter Error Log to
Exception reported an unhandled "DiagCode" and
exception from the find 211. See prior
top of the process. errors in Error
EHA gets the last Log. Use UpLink
chance to recover. software to
transfer the logs to
Bothell.

Conquest Imaging - 171 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
212 DMM Image The DMM domain Filter Error Log to
directory delete has reported an error "DiagCode" and
error while deleting image find 212. See prior
directory. errors in Error
Log. Use UpLink
software to
transfer the logs to
Bothell. Restart
system and delete
image directory.
214 DMM -- task Cannot start up Filter Error log to
scheduler Windows task "DiagCode" and
cannot start up. scheduler service. find 214. See prior
Reboot errors. Use UpLink
required. software to
transfer the logs to
Bothell.
215 With 1.1.3.283 Advise customers
systems (Update 4.0), to avoid using
the code is noticed those tools and
mainly during instead use the
vascular and OB keyboard
exams while using the annotation keys. If
touch screen the customer does
annotation tools. a lot of annotation,
retain their system
at 1.1.2.197
(Update 3.0) until
the issue is
resolved.
218 ANNOTSP The Annotation Filter Error log to
unanticipated Signal Path domain "DiagCode" and
error - Restart has reported an error find 218. See prior
Required for which there is no errors. Use UpLink
recovery. Scanning software to
has stopped. User transfer the logs to
must restart the Bothell.
system with the
On/Standby switch.

Conquest Imaging - 172 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
219 UAA The Ultrasound Filter Error log to
unanticipated Analysis Application "DiagCode" and
error - Restart domain has reported find 219. See prior
Required an error for which errors. Use UpLink
there is no recovery. software to
Scanning has stopped. transfer the logs to
User must restart the Bothell.
system with the
On/Standby switch.
220 UIMONSTER The UIMonster Filter Error log to
unanticipated domain has reported "DiagCode" and
error - Restart an error for which find 220. See prior
Required there is no recovery. errors. Use UpLink
Scanning has stopped. software to
User must restart the transfer the logs to
system with the Bothell.
On/Standby switch.
221 UIOGRE The UIOgre domain Filter Error log to
unanticipated has reported an error "DiagCode" and
error - Restart for which there is no find 220. See prior
Required recovery. Scanning errors. Use UpLink
has stopped. User software to
must restart the transfer the logs to
system with the Bothell.
On/Standby switch.
222 UITK The Annotation Filter Error log to
unanticipated Signal Path domain "DiagCode" and
error - Restart has reported an error find 220. See prior
Required for which there is no errors. Use UpLink
recovery. Scanning software to
has stopped. User transfer the logs to
must restart the Bothell.
system with the
On/Standby switch.
223 VOX The Annotation Filter Error log to
unanticipated Signal Path domain "DiagCode" and
error - Restart has reported an error find 220. See prior
Required for which there is no errors. Use UpLink
recovery. Scanning software to
has stopped. User transfer the logs to
must restart the Bothell.
system with the
On/Standby switch.

Conquest Imaging - 173 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
224 Machine Caught unknown Filter Error log to
Monitor crash exception in Machine "DiagCode" and
with unknown monitor. find 220. See prior
exception. errors. Use UpLink
Reboot software to
required. transfer the logs to
Bothell.
225 DMM -- Caught unknown Filter Error log to
JobManager exception in main "DiagCode" and
crash with thread of find 225. See prior
unknown JobManager. Job errors. Use UpLink
exception. manager terminated. software to
Reboot User must restart the transfer the logs to
required. system with the Bothell.
On/Standby switch.
226 DMM -- JobManager has Filter Error log to
JobManager reported an error for "DiagCode" and
unanticipated which there is no find 226. See prior
error. Reboot recovery. User must errors. Use UpLink
required. restart the system software to
with the On/Standby transfer the logs to
switch. Bothell.
501 An Invalid Cannot start up an Filter Error log to
Domain ID was unidentifiable "DiagCode" and
received for a Domain. find 501. See prior
domain that errors in error log.
failed Use UpLink
initialization. software to
transfer the logs to
Bothell.
502 Analysis Cannot start up Filter Error log to
Application Analysis Application "DiagCode" and
Domain failed Domain. find 502. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 174 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
503 Acquisition Cannot start up Filter Error log to
Control Acquisition Control "DiagCode" and
Domain failed Domain. find 503. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
504 Annotation Cannot start up Filter Error log to
Signal Path Annotation Signal "DiagCode" and
Domain failed Path Domain. find 504. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
505 Applications Cannot start up Filter Error log to
User Interface Applications User "DiagCode" and
Domain failed Interface Domain. find 505. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
506 Acquisition Cannot start up Filter Error log to
Protocols and Acquisition Protocols "DiagCode" and
Replay Domain and Replay Domain. find 506. See prior
failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
507 BUMPR Cannot start up Filter Error log to
Control BUMPR Control "DiagCode" and
Software Software Domain. find 507. See prior
Domain failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 175 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
508 Deadline Cannot start up Filter Error log to
Detection and Deadline Detection "DiagCode" and
Measurement and Measurement find 508. See prior
Domain failed Domain. errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
509 Development Cannot start up Filter Error log to
Applications Development "DiagCode" and
Domain failed Applications Domain. find 509. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
510 Display Layout Cannot start up Filter Error log to
Services Display Layout "DiagCode" and
Domain failed Services Domain. find 510. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
511 Digital Media Cannot start up Filter Error log to
Management Digital Media "DiagCode" and
Domain failed Management Domain. find 511. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
512 Data Cannot start up Data Filter Error log to
Management Management Services "DiagCode" and
Services Domain. find 512. See prior
Domain failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 176 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
513 Debug Print Cannot start up Filter Error log to
Services Debug Print Services "DiagCode" and
Domain failed Domain. find 513. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
514 Error Handling Cannot start up Error Filter Error log to
Services Handling Services "DiagCode" and
Domain failed Domain. find 514. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
515 Error Handling Cannot start up Error Filter Error log to
Application Handling Application "DiagCode" and
Domain failed Domain. find 515. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
516 Frame Cannot start up Frame Filter Error log to
Composition Composition Control "DiagCode" and
Control Domain. find 516. See prior
Domain failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
517 File Cannot start up File Filter Error log to
Management Management Services "DiagCode" and
Services Domain. find 517. See prior
Domain failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 177 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
518 Signal and Cannot start up Signal Filter Error log to
Image and Image Processing "DiagCode" and
Processing Server Domain. find 518. See prior
Server Domain errors in error log.
failed Use UpLink
initialization. software to
transfer the logs to
Bothell.
519 Image Analysis Cannot start up Image Filter Error log to
Domain failed Analysis Domain. "DiagCode" and
initialization. find 519. See prior
errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
520 Host Image Cannot start up Host Filter Error log to
Memory Image Memory "DiagCode" and
Domain failed Domain. find 520. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
521 System Admin Cannot start up Filter Error log to
Install Upgrade System Admin Install "DiagCode" and
Domain failed Upgrade Domain. find 521. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
522 System Admin Cannot start up Filter Error log to
JTAG Manager System Admin JTAG "DiagCode" and
Domain failed Manager Domain. find 522. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 178 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
523 System Admin Cannot start up Filter Error log to
Logging System Admin "DiagCode" and
Application Logging Application find 523. See prior
Domain failed Domain. errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
524 Logging Cannot start up Filter Error log to
Services Logging Services "DiagCode" and
Domain failed Domain. find 524. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
525 System Cannot start up Filter Error log to
Monitoring System Monitoring "DiagCode" and
Domain failed Domain. find 525. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
526 Measurement Cannot start up Filter Error log to
Tools Domain Measurement Tools "DiagCode" and
failed Domain. find 526. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
527 Platform Cannot start up Filter Error log to
Independence Platform "DiagCode" and
Toolkit Domain Independence Toolkit find 527. See prior
failed Domain. errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 179 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
528 Plug-in Cannot start up Plug- Filter Error log to
Theories in Theories Domain. "DiagCode" and
Domain failed find 528. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
529 Platform Cannot start up Filter Error log to
Services Platform Services "DiagCode" and
Domain failed Domain. find 529. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
530 Peripheral Cannot start up Filter Error log to
Platform Peripheral Platform "DiagCode" and
Services Services Domain. find 530. See prior
Domain failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
531 QLAB Domain Cannot start up Filter Error log to
failed QLAB Domain. "DiagCode" and
initialization. find 531. See prior
errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
532 System Admin Cannot start up Filter Error log to
System Config System Admin "DiagCode" and
Domain failed System Config find 532. See prior
initialization. Domain. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 180 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
533 System Admin Cannot start up Filter Error log to
System System Admin "DiagCode" and
Component System Component find 533. See prior
Config Services Config Services errors in error log.
Domain failed Domain. Use UpLink
initialization. software to
transfer the logs to
Bothell.
534 Streaming Data Cannot start up Filter Error log to
Buffer Streaming Data "DiagCode" and
Management Buffer Management find 534. See prior
Domain failed Domain. errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
535 Signal and Cannot start up Signal Filter Error log to
Image and Image Processing "DiagCode" and
Processing Control Domain. find 535. See prior
Control errors in error log.
Domain failed Use UpLink
initialization. software to
transfer the logs to
Bothell.
536 Study Cannot start up Study Filter Error log to
Management Management Domain. "DiagCode" and
Domain failed find 536. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
537 System Admin Cannot start up Filter Error log to
Application System Admin "DiagCode" and
Domain failed Application Domain. find 537. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 181 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
538 System Admin Cannot start up Filter Error log to
System Tools System Admin "DiagCode" and
Domain failed System Tools find 538. See prior
initialization. Domain. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
539 System Admin Cannot start up Filter Error log to
TestUtil System Admin "DiagCode" and
Domain failed TestUtil Domain. find 539. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
540 User Cannot start up User Filter Error log to
Annotations Annotations "DiagCode" and
Application Application Domain. find 540. See prior
Domain failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
541 Ultrasound Cannot start up Filter Error log to
Data Ultrasound Data "DiagCode" and
Management Management Domain. find 541. See prior
Domain failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
542 User Interface Cannot start up User Filter Error log to
Playback Interface Playback "DiagCode" and
Domain failed Domain. find 542. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 182 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
543 User Interface Cannot start up User Filter Error log to
Toolkit Domain Interface Toolkit "DiagCode" and
failed Domain. find 543. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
544 Ultrasound Cannot start up Filter Error log to
Control Ultrasound Control "DiagCode" and
Domain failed Domain. find 543. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
545 Ultrasound Cannot start up Filter Error log to
Imaging Ultrasound Imaging "DiagCode" and
Domain failed Domain. find 543. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
546 Virtual Cannot start up Filter Error log to
Database Virtual Database "DiagCode" and
Domain failed Domain. find 543. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
547 XUML Cannot start up Filter Error log to
Architecture XUML Architecture "DiagCode" and
Domain failed Domain. find 543. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.

Conquest Imaging - 183 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
548 Applications Cannot start up UI Filter Error log to
User Interface Monster Domain. "DiagCode" and
sub-domain find 543. See prior
failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
549 Applications Cannot start up UI Filter Error log to
User Interface Ogre Domain. "DiagCode" and
sub-domain find 543. See prior
failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
550 Applications Cannot start up UI Filter Error log to
User Interface Ogre Domain. "DiagCode" and
sub-domain find 543. See prior
failed errors in error log.
initialization. Use UpLink
software to
transfer the logs to
Bothell.
601 Some Domain Cannot start up a Filter Error log to
failed Domain. This could "DiagCode" and
initialization. be one of the find 543. See prior
following domains: errors in error log.
job manager, SQL Use UpLink
server, web services, software to
DVDController, drive transfer the logs to
mapping, sip power, Bothell.
gppc stop, task
scheduler start, task
scheduler stop or an
operating system
fault.

Conquest Imaging - 184 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
602 Physical Cannot start up Filter Error log to
Memory Physical Memory "DiagCode" and
Domain failed Domain. find 543. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
603 Distributed Cannot start up Filter Error log to
Domain failed Distributed Domain. "DiagCode" and
initialization. find 543. See prior
errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
604 GPPC Start Cannot start up GPPC Filter Error log to
Domain failed Start Domain. "DiagCode" and
initialization. find 543. See prior
errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
605 System Cannot start up Filter Error log to
Monitor System Monitor "DiagCode" and
Domain failed Domain. find 543. See prior
initialization. errors in error log.
Use UpLink
software to
transfer the logs to
Bothell.
900 Message The EHA has Restart system.
Overflow experienced an
During overflow of requests
Recovery to handle system
faults, dialog box
responses, or control
commands. The
default response is to
stop scanning and
display the "Restart
required" dialog box.

Conquest Imaging - 185 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Description / Service
Code Condition
Consequence Responsibility
901 Fault During The EHA has Restart system.
Recovery experienced an
exception while
processing a system
fault recovery. The
default response is to
stop the system and
display the "Restart
required" dialog box.
If that fails, the EHA
will give up and abort
recovery
902 Fault In The fault map that Restart system.
Recovery Map EHA uses to choose
recovery actions is in
error or has
omissions. Any fault
that is not correctly
mapped in the fault
map file gets a default
mapping that displays
the "Restart required"
dialog. Box
903 Fault Ignored This fault has been Restart system.
By EHA ignored because it
was reported while
EHA had an open
dialog box for a prior
active fault.

System Software Tools

The system contains a number of software tools used to run tests, record and
monitor system behavior, and conduct analysis. Tool use is documented in
several places including: the operation manual, operations manual help, and
in Tech Admin Help. This section provides an overview of all of the tools
and more comprehensive information regarding specific tools that is not
covered in the other documentation.

Conquest Imaging - 186 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Manage Setups Function

The Manage Setups function allows Setups data to be copied from the
ultrasound system to a removable disk (DVD). The setups data may then be
copied to another ultrasound system or reloaded onto the current ultrasound
system after an upgrade procedure. The Manage Setups capabilities are
available to all users. Refer to the operations manual (on the User
Information CD) for general information and refer to Section 14, “System
Administration” for use of Manage Setups when servicing equipment.

Licensed Options

Licensed options enable the use of purchasable options available on the


ultrasound system. The options may be customer-purchased or temporarily
enabled for customer use. Up to five licensed options may be temporarily
enabled at one time for each ultrasound system through the use of license
keys. Temporary licensed options are automatically disabled after 1 month.
When purchased, the options are listed as “enabled” in the Licensed Option
list.

NOTE Do not enable more than four temporary licensed options on the
same ultrasound system. The fifth temporary licensed option must be
reserved for access to Tech Admin.

The Lic Options function displays the licensed options and whether they are
permanently enabled or on a trial-basis. Refer to the user documentation for
general information and refer to Section 14, “System Administration” for
use of Lic Options when servicing equipment.

System Confidence Test

The System Confidence Test or System Test is a comprehensive test of the


system operational status. This test comprises numerous sub-tests: the
System Functional Test, all of the Field

Replaceable Unit (FRU) Comprehensive Tests, and all of the Interface Tests.
If a sub-test fails, the system completes the remaining sub-tests. System Test
execution from this screen is available to all users and displays only a pass-
fail result on the system monitor. Sub-test results are annotated in system
Conquest Imaging - 187 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
logs. System users should notify Philips Ultrasound if the confidence test
fails. The System Confidence Test is also accessible from within the Tech
Admin functions. Test execution from within Tech Admin is limited to
service personnel. If a sub-test fails, a failure code is displayed.
The System Confidence Test should be run any time a system error is
displayed, or if the user suspects problems with the system. If an error
message is displayed during the test, reboot the system with the On/Standby
switch.

Conquest Imaging - 188 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Invoking the System Confidence Test

1. Power up the system.


2. Verify there are no transducers connected.
3. Press the Setup key.
4. Click Config/Test.
5. Click the System Test tab.
6. Click Run. The system displays a message when the test is complete,
telling you whether the test passed or failed.
7. Click Cancel or Done to exit the setups.

NOTE The test takes approximately 30 minutes.

Conquest Imaging - 189 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
NOTE The Service Export tab shown on this figure is not displayed on
systems below 2.0 software.

Service Export (2.0 Software)

Service Export is used to transfer encrypted error logs to a DVD disk. If the
data is not saved, an alert is displayed, advising you to check the media in
the drive and try again.

Service Export Tab

NOTE System software builds below 2.0 did not have the Service Export
tab.

Conquest Imaging - 190 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Technical Administration

Technical Administration (Tech Admin) functions are intended for


authorized personnel only (Philips service, manufacturing, and engineering
personnel) and are documented in Tech Admin Help available on the
system.

Channel Walk Test

The channel walk test is used to determine if image drop-out is caused by


noise or by receive or transmit functions of individual channels and boards.
The channel walk test is also used to determine if a problem is caused by a
transducer or the system.

Performing the channel walk test

1. Connect and initialize an L12-5 transducer on any port.


2. Set the system parameters as indicated in Table 9-11.

Channel Walk System Settings


Parameter Setting
SonoCT
Imaging Off
XRES
Imaging Off
Output Maximum
2D Gain Maximum
TGC Slide
pots Maximum
# Focal Zones 1
Less than maximum
Focus depth

Conquest Imaging - 191 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
3. Set the Acquisition/Channel Walk utility parameters as follows
- Channel
- Transmit
- Walk ‘1’
- Automatic
- Delay 0.5 sec
- Transducer Port 1
- SH Element to 1
4. While imaging a phantom, slowly scroll the tower(s) across image and
verify the appearance of the tower for each element is consistent with
the towers for all other elements.
5. Select Receive.
6. While imaging a phantom, slowly scroll the tower(s) across image and
verify the appearance of the tower for each element is consistent with
the towers for all other elements.
7. Click Stop.
8. Connect the transducer to the next port.
9. Set SH Port to this port.
10. Click > to begin scrolling.
11. Repeat step 6 through step 10 until all the ports have been tested.
12. Click Finish.

FEC Debug Port

The FEC debug port may be used to determine the configuration of the front
end PCBs (including part numbers), the version number of front end ASICs,
and if there are faults in the Acquisitionm Subsystem. The detectable faults
include power monitor faults and hardware configuration errors. It also
checks if the Motor Controller PCB is present and reports the checksums for
the Channel Board flash.The system should remain off until the proper setup
has occurred.

NOTE The FEC debug port can be used to determine if there is a Channel
Board fault, but does not determine which board is faulty. To determine the
faulty board, log onto Tech Admin and use system diagnostics.

Conquest Imaging - 192 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Using the FEC debug port to determine if the front end is faulty

1. Connect the system to the laptop.


2. On the laptop, click Start, Programs, Accessories, Communications,
HyperTerminal.
3. If the new connection dialog box appears, enter a name and click OK.
4. Configure the HyperTerminal settings as indicated.

FEC Debug Port HyperTerminal Settings

Parameter Setting
Connect To tab
Connect using COM1
COM1 Properties
Bits per second 19200
Data bits 8
Parity None
Stop bits 1
Flow control Xon/Xoff

5. Click OK to launch HyperTerminal.


6. Power up the system. The front end status is displayed in the
HyperTerminal window.

VCR Audio

Correcting weak VCR audio

1. Determine whether the 7500-2072-02 or 453561167391 AVIO-RIP


PCB is installed in the system.
2. On systems with the 7500-2072-02 AVIO-RIP, replace it with the
453561167391.
3. Connect an external microphone (2100-0570-01) to the VCR.
4. Set AUDIO to MIX on the front of the VCR to allow audio playback
through the system.
5. Test the VCR audio performance.
6. If audio playback is sufficient, disregard the following steps.

Conquest Imaging - 193 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
7. Connect external microphone (2100-1868-01) to MIC (P18) on the
AVIO-RIP.
8. Secure the microphone above the DVD drive on the system.
9. Test the VCR audio performance.

User-Accessible Test Images

Forty-three test patterns are stored on a hard drive partition. The test patterns
must be transferred to a different drive partition before they can be viewed.
Transferring the test patterns to the drive partition must be done only once if
the test patterns are not deleted from the drive partition. Use the test patterns
to check image quality or to transfer the test patterns to a PACS for image
quality tests.

Transfering and view the test images

1. Power up the ultrasound system.


2. Press Review.
3. Select Test Imgs from the Drive pull down menu.
4. Select Test Patterns from the exam selection list.
5. Click Send to. A pop-up window is displayed.
6. Click the Hard Drive box and click OK. The images are transferred to
the hard drive partition.

NOTE This may take several seconds. If necessary check the transfer status,
by clicking the DVD Drive status icon (the second icon from the left) at the
bottom of the display.

7. Select Hard Drive from the Drive pull down menu.


8. Select Test Patterns from the exam selection list.
9. Click Display Exams. The test patterns are displayed.

Conquest Imaging - 194 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™

Chapter 7
Dicom

Conquest Imaging - 195 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 196 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ System Dicom

Configuring the System for Network Operation

1. Power up the ultrasound system.


2. Press the Setup key.
3. Click Print/Network.
4. Click the Global Config tab.
5. Click System.
6. On systems with software versions below 2.0, enter the following:
• AE Title
• Port Number (a change in this field will cause the system to
reboot)
• IP Address
• Subnet Mask
• Gateway
• Station Name
• System Location

NOTE Station Name and System Location are not required fields. The
system must be connected to a live network for the settings to be visible and
for the changes to take effect.

7. On systems with software version 2.0 or higher:


a. Enter the following:
• AE Title
• Port Number (a change in this field will cause the system to
reboot)
• Station Name
• System Location
b. Click Dynamic (DHCP) or Static IP Configuration.

NOTE The values entered for Station Name and System Location must
match the corresponding Modality Worklist field values. These fields can
be left blank if the station name or system location are not being used in the
worklist or if no worklist server is configured. Consult with the site
administrator for more information.

Conquest Imaging - 197 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Adding a Network Device

1. Perform step 1 through step 4 of “Configuring the System for


Network Operation” section.
2. Click Devices.
3. Click New Device.
4. Use the pull-down menu to select the device type or type in the device
type.
5. Configure the new device by entering the following:

NOTE * indicates required field.

 Device Name *
 AE Title *
 Port # *
 IP Address *

6. Click OK.
7. Check that the new device appears in the Device Configuration
window.

Verifying Network Devices

1. Perform step 1 through step 4 of “Configuring the System for


Network Operation” section.
2. Click Devices.
3. In the Device Configuration window, highlight the device that you
want to verify.
4. Click Verify. Verification confirms multi-frame and single-frame
storage. A message is displayed with the verification status.

NOTE When archiving, verify that the system negotiates multi-frame


ultrasound (Cineloop) storage with JPEG compressed transfer syntax.
Compression and multi-frame cannot be turned off.

Conquest Imaging - 198 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Adding Dicom Printers

1. Perform step 1 through step 4 of “Configuring the System for


Network Operation” section.
2. Click Devices.
3. Click New Devices.
4. Select the device type from the menu, either DICOM B&W Printer
or DICOM Color Printer.
5. Select the printer type from the menu. Do not set up a printer if it is
not listed. (Do not type the printer type in the printer type field.)
6. Enter the device name, AE title, port number, and IP address.
7. Click OK.

Selecting Your Printer

To select a printer and edit the printer settings

1. Press the Setup key.


2. Click Print/Network.
3. Click the Device Selection tab.
4. Click Printer.
5. Enable the printer or printers by selecting the checkbox to the left of
each printer name.

NOTE A maximum of two DICOM printers, two local printers and one
local report printer, may be selected from the association lists.

6. If you want to change the number of copies, film size, medium type,
display format or orientation, Click Edit.
7. Type the desired settings or use the menus.
8. Click OK.

Archive Devices, Storage Commit Devices, and Structured Report


Devices

Conquest Imaging - 199 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Selecting archive devices, storage commits, or structured report devices

1. Press the Setup key.


2. Click Print/Network.
3. Click the Device Selection tab.
4. Click Archive.
5. Enable the archive server or servers by selecting the check-box to the
left of each server name.

NOTE A maximum of three archive servers may be selected from the list.

6. Select a storage commit server (if configured) from Commit Server.


If a commit server is selected, also select the archive system being
monitored from Target Archive Server.
7. Select a report server (if configured) from Structured Report Server.

Modality Worklist and Performed Procedure Step Servers

Selecting a worklist or performed procedure step server

1. Press the Setup key.


2. Click Print/Network.
3. Click the Device Selection tab.
4. Click Worklist.
5. Select the Worklist Server and Performed Procedure Step (PPS)
Server from the menus.
6. Select the appropriate Define Query options.

Conquest Imaging - 200 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™

Chapter 8
Parts List

Conquest Imaging - 201 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
This Page Intentionally Left Blank

Conquest Imaging - 202 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ Parts List

iU22/iE33 System
Signal Interconnect
Cables
Part Number Cable Description Notes
2100-1919-01 Cable Assy, Adapter
USB 2.0 to IDE
453561171931 Cable Assy, Adapter
USB 2.0 to IDE
without DC Connector
3500-3615-02 Cable Assy, AGP to
AVIO
3500-3625-01 Cable Assy, AVIO-
CAD to AVIO-RIP
453561155361 Cable Assy, AVIO to
Display, 1st Segment
3500-3617-03 Cable Assy, AVIO to
Speaker
453561153501 Cable Assy, NOTE There is no
On/Standby Switch connector key on
LED connector P2.
If the light comes
on with the system
powered on, then
the cable
connection to the
LED is connected
correctly.
3500-3682-01 Cable Assy,
On/Standby Switch
Assy
3500-3644-02 Cable Assy,
On/Standby Switch,
Host
3500-3621-01 Cable Assy, PCI -
CD/DVD Audio
3500-3589-03 Cable Assy, QWERTY
Conquest Imaging - 203 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33 System
Signal Interconnect
Cables
Part Number Cable Description Notes
LED
3500-3606-01 Cable Assy, SIP to
IDE
3500-3645-01 Cable Assy, SIP Power
On/Standby to
Bulkhead
453561153801 Cable Assy, Stereo,
Mic Out, Line In, Line
Out
453561153792 Cable Assy, Trackball
to CPM USB, iU22

iU22/iE33 System
OEM Signal
Interconnect Cables
Part Number Cable Description Notes
3500-3595-01 Cable Assy, AVIO
Board to VCR Audio
Line In/Line Out
3500-3596-01 Cable Assy, AVIO
Board to VCR S-Video
3500-3591-01 Cable Assy, AVIO -
Modem USB
2100-1868-01 Cable Assy, AVIO to
Cart Microphone
2275-0176 Cable Assy, BNC,
M/M, 6-ft, RG 59B/U,
75Ohm
453561153451 Cable Assy, USB,
AVIO - OEM

Conquest Imaging - 204 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33 System
Test Cables
Part Number Cable Description Notes
453561170191 Cable Assy,
Acquisition Frontplane
to Laptop Serial
Connector

iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
3500-3365-01 Acquisition Frontplane
Assy
3500-3391-01 AC Tray Assy 120V/240V
3500-3745-01 AC Tray Assy 100V
453561155191 Arm, Lower, Display Monitor
Articulation articulation
453561155471 Articulation Arm
Assy, Flat Screen
Display
1563-0179 Ball Stud, 5/8 Length, Rear enclosure
6.32x.25 Stud access door
453561152171 Ball Stud, Gas Spring, Monitor
10 mm articulation
453561155131 Bar, Lower, Display Monitor
Articulation articulation
453561150072 Bar, Upper, Display Monitor
Articulation articulation
453561151571 Base, Pivot, Display
Articulation
1700-0138-01 Battery, 6 V, 8.0 AH,
Sealed, Lead Acid
3500-3744-01 Battery Tray Assy
2950-1619-01 Bezel, LED, 5 mm,
0.25 Panel Hole, 2PC
1519-0022 Bolt, 1/4 20x3/4, Hex,
St/G5, Zn
Conquest Imaging - 205 -
Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6589-01 Bracket, Access Door Rear enclosure
door latch
3000-0864-01 Bracket, Access Door Rear enclosure
Latch door latch
1065-6623-01 Bracket Assembly,
Filter Tray, Aft
1065-6622-01 Bracket Assembly,
Filter Tray, Forward
1065-6239-01 Bracket, Catch,
Acquisition, Rear
1065-6249-01 Bracket, Disk Drive,
Left
1065-6248-01 Bracket, Disk Drive,
Right
1065-6675-01 Bracket, EMI, DVD
1065-6546-01 Bracket, L, Grounding Acquisition card
cage
1065-6513-01 Bracket, Mount,
Monitor, Horiz,
EIZ0567
1065-6511-01 Bracket, Mount,
Monitor, Right,
EIZ0567
1065-6516-01 Bracket, Mount, Vesa,
Eizo, L500, Monitor
1065-6559-02 Bracket, Platform, Rear cooling fans
Card Cage Fan
1065-6595-01 Bracket, Power Cord Provides retention
for the power cord
to AC tray
receptacle
453561160051 Bracket, Shipping For shipping
Strap purposes only

Conquest Imaging - 206 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6663-01 Bracket, Speaker
Mounting
453561157701 Bracket, Swivel, Brake System brake
Assy assembly
453561155161 Bracket, Tilt, Monitor
3500-3743-01 Brake Assembly, Core
Cart
1065-6649-01 Brake Assembly, Left System brake
Pedal assembly
1065-6648-01 Brake Assembly, System brake
Right Pedal assembly
2950-0586-01 Bumper, 0.5-in x 0.14- Video monitor
in H, Clear Vinyl,
Adhesive Back
1065-6356-01
3000-0877-01
1065-6349-01
2950-1594-01
2950-1570-01
453561163761
453561160311
3500-3684-01 Caster Linkage, Brake System brake
Assy assembly Caster
linkage includes
1065-6552-01 (2
each) and 1065-
6553-01
1065-6553-01 Caster Link, Center, System brake
Brake Assy assembly
1065-6552-01 Caster Link, Lever, System brake
Brake Assy assembly
1065-6674-01 Catch, Transducer
Holder

Conquest Imaging - 207 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6500-01 Chimney, EMI, Left,
Platform Card Cage
1065-6501-01 Chimney, EMI, Right,
Platform Card Cage
2700-0270-01 Circuit Breaker, 15 A,
250 V, with Line
Filter, UL
2950-1179-01 Clamp, CA, 3/8 D, Used on AC Tray
1/2W, 1/4 Screw, Ins, to retain power
Al cord
1065-6566-01 Clamp, Cable, Single Used on the flat
panel video
monitor
1065-6205-01 Clip, OEM Strap Used with strap
1065-2386-01.
Strap clips are used
only for OEMs
positioned on the
lower OEM deck
shelf.
3500-3522-04 Control Panel Assy, Includes: Control
iU22 panel PCB, Control
panel connector
PCB
1065-6430-02 Cover, Access, Cable, Monitor
Display Articulation articulation
1065-5753-02 Cover, Acquisition
Frontplane
1065-6445-01 Cover, Arm, Display Monitor
Articulation articulation
1065-6556-01 Cover, Cable Access
Tower, Left
1065-6557-01 Cover, Cable Access
Tower, Right

Conquest Imaging - 208 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6242-01 Cover, Coffin, Rat
Hole
1065-6567-01 Cover, Disk Bay, Rear
1065-6227-01 Cover, HV Assy, Left
1065-6226-01 Cover, HV, Right
1065-6350-02 Cover, Rear, Platform
Card Cage
1065-6454-01 Cup, Reservoir, One removable gel
Transducer Holder drip cup per system
2100-1940-01 Disk Drive, DVD/CD,
Rewritable, ATAPI
INTFC
453561170151 Disk Drive, DVD/CD,
Rewritable, ATAPI
INTFC
1065-6653-01 Elbow, Swivel, Monitor
Articulation Display articulation
3000-0860-01 Enclosure, Access Rear main system
Door, Rear card cage enclosure
3000-0858-01 Enclosure Assy, Front Video monitor
3500-3658-02 Enclosure, Cover,
Filter Door
3500-3703-01 Enclosure, Cover,
Main, Rear
1065-6204-02 Enclosure, Cover, Used on systems
OEM Cable with one and two
1065-6377-01 Enclosure, Cover,
OEM Cable Drop
3500-3675-02 Enclosure, Cover,
OEM Deck
1065-6203-02 Enclosure, Cover,
OEM Strap
3500-3657-02 Enclosure, Cover,
Power Supply

Conquest Imaging - 209 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
3500-3628-01 Enclosure, Disk Bay, Also encloses the
Column Front Assy scanhead select
connectors.
Includes ON/STBY
switch button,
1065-6356-01.
1065-6185-02 Enclosure, Disk Bay,
Column Insert
1065-6186-02 Enclosure, Disk Bay,
Column Rear
3500-3653-02 Enclosure, Disk Bay,
Column Side, Left
3500-3652-02 Enclosure, Disk Bay,
Column Side, Right
1065-6184-01 Enclosure, Disk Bay,
Rear Display
3500-3546-02 Enclosure, Front
Bumper, Left
3500-3545-02 Enclosure, Front
Bumper, Right
3500-3656-02 Enclosure, Front
Panel, Lower
3500-3544-02 Enclosure, Front
Panel, Upper
1065-6464-01 Enclosure, Keyboard, QWERTY
Housing
1065-6188-01 Enclosure, Main Rear main system
Cover, Rear card cage enclosure
1065-6192-02 Enclosure, Mud Plate
1065-6376-02 Enclosure, OEM
Shelf, Lower
453561155221 Enclosure, OEM
Shelf, Upper
1065-6510-01 Enclosure, Rear Video monitor

Conquest Imaging - 210 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6191-02 Enclosure, Transducer
Holder Base, Left
1065-6190-02 Enclosure, Transducer
Holder Base, Right
1065-6520-01 Enclosure, Shroud,
Monitor, Lower
1065-6521-01 Enclosure, Shroud,
Monitor, Upper
3500-3660-02 Enclosure, Side Cover,
Rear, Left
3500-3659-02 Enclosure, Side Cover,
Rear, Right
3500-3548-02 Enclosure, Side Cover,
Left
3500-3547-02 Enclosure, Side Cover,
Right
1065-6180-02 Enclosure, Top,
Control Panel
Articulation
1065-6523-01 Extension, Brake System brake
Pedal, Left assembly
1065-6522-01 Extension, Brake System brake
Pedal, Right assembly
2100-1910-01 Fan, 12 V, 360 MA, Two fans located
105 CFM, 119 mm, on the filter tray
with Connector and one on each
side of the AC tray
3500-3701-01 Fan Mounting Plate Platform power
Assy, Platform supply
453561150251 Filter, Acquisition
Card Cage
453561150191 Filter, Platform Card Rear enclosure
Cage access door

Conquest Imaging - 211 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
3500-3700-01 Filter Tray/Fan Assy Acquisition card
cage cooling fans
2700-0151 Fuse, 4 A, 250 V, Fast,
1/4x1-1/4, UL/CSA
1065-6389-01 Gasket, Foam Speaker Boxes
Enclosure Seal
1065-6708-01 Gasket, Retainer, Acquisition card
Bottom cage
1065-6709-01 Gasket, Retainer, Top Acquisition card
cage
3600-0246-01 Gasket, RFI, Finger,
Clip, BE-CU, 2.5
453561165761 Grease, Molygraph, For brake pedal
SL3144, 3-oz linkage Part
Cartridge includes a 3-pack
of 3-oz cartridges
3600-0047 Grommet, 0.375ID, Monitor
0.660D, 0.06PNL, articulation
RBR
1065-6398-01 Grommet, Door, Acquisition power
Removable supply and filter
tray
2700-0240-01 Holder, Fuse Part of battery
assembly
1065-6458-01 Holder, Transducer, Part of system
Enclosure control panel Holds
transducer or gel
bottle
2950-1155-01 Holder, Microphone,
Rubber
3500-3738-01 Holder, Transducer, Located on system,
Assy, Left left side column
3500-3737-01 Holder, Transducer, Located on system,
Assy, Right right side column

Conquest Imaging - 212 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
3500-xxxx-xx HV Switcher Assy Includes mounting
plate and hardware
1065-6652-01 Insert, Arm, Lower, Monitor
Display Articulation articulation
3000-0879-01 Insert, Transducer
Holder, Left
3000-0878-01 Insert, Transducer
Holder, Right
2950-1611-01 Isolator, Vibration, Used on filter tray
Fan fans and AC tray
fans
1065-5752-02 Insulator, Frontplane
1065-6435-01 Knob, Locking, Monitor
Display Articulation articulation
1065-6462-01 Knob, Small, Push
Rotate
1065-6655-01 Knuckle, Display Monitor
Articulation articulation
453561163221 Label, 3x1, Met Poly, Serial number
Sil, Therm Xfr
453561157471 Label, High Voltage Backlight inverter
Safety 600 Vac warning
453561161311 Label, Lock/Unlock, Lock/Unlock
Display Articulation symbols
Arm, Left
453561161301 Label, Lock/Unlock, articulation lower
Display Articulation arm)
Arm, Right
453561156511 Label, Main System Model number, CE
453561154171 Label, Panel, AVIO- Connector
RIP, PCB designation
symbols
453561161261 Label, Pedal, Brake Brake lock symbol
Lock

Conquest Imaging - 213 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
453561161151 Label, Pedal, Swivel Swivel lock symbol
Lock
453561156521 Label, Power Switch Power switch
location label
453561156501 Label, System Badge iU22, ultrasound
product name
453561161321 Label, USB Symbol (Located on back
of disk bay)
453561161791 Label, Warning, PCB controls not
Monitor field adjustable
453561160341 Latch Assy, Display Monitor
Articulation articulation
1065-6439-01 Link, Tension Adjust, Monitor
Articulation Display articulation
2100-1926-01 Modem, V.90,
Portable, USB
2100-1906-01 Monitor, 17-in, Color, Video monitor
LCD Display
3500-3677-01 Monitor Assy, Flat- Video monitor
Panel
3500-3701-01 Mounting Plate Assy, Rear cooling fans
Fan
1544-0012 Nut, 1/4-20, Keps, St, System brake
Zn assembly
1555-0037-01 Nut, 1/4-28, Keps, System brake
Ext, St, Zn assembly
1555-0036-01 Nut, 5/16-18, Hex, Monitor
Nylon Insert, St, Zn articulation
1655-0020-01 Nut, M6, Hex, Keps, Used on AC tray
St, Zn ground stud
1065-6157-03 Partition, Acquisition
Card Cage
PCB Assy,
Acquisition Frontplane

Conquest Imaging - 214 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
PCB Assy, Analog
Input Module
PCB Assy, AVIO AVIO CAD
Control and Display
PCB Assy, AVIO Rear AVIO RIP
Interface Panel AVIO
RIP
7500-1914-xx PCB Assy, Breakout
Board, Platform Power
Supply
PCB Assy, Channel
Board CB0-CB3
PCB Assy, Controller, Promise IDE
Ultra ATA 133, PCI Controller (SIP
66 Mhz module board)
PCB Assy, Dual DSC
Signal Conditioning
PCB Assy, Front End FEC
Controller (iU22)
PCB Assy, Front End FEC
Controller with 3D
Motor Controller
PCB Assy, Image IEP
Export Processor
PCB Assy, IDE
Motherboard, Xeon,
Dual Processor
PCB Assy, PCA,
Motor Controller
PCB Assy, Scanhead SHSEL
Select
PCB Assy, SCI Dolphin D-340
Adapter, PCI, 64B, 66
Mhz

Conquest Imaging - 215 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
453561155171 Pin, 0.1875, Diameter
453561171591 Pin, Hinge, Display Monitor
Articulation articulation
1065-6696-01 Pin, Knurl Monitor
articulation
3500-3551-01 Plate, Adapter, Touch
Screen
453561161401 Plate, Mount, Brake System brake
Assy assembly
1065-6551-01 Plate, Plunger, Brake System brake
Assy assembly
1065-6468-01 Platform, OEM Deck
1065-6499-01 Plenum, Platform Card
Cage
1065-6155-02 Plenum, Top,
Acquisition Card Cage
2950-1603-01 Plunger, Spring, 1/4- Keyboard assembly
20, 2-4 Lbs, Ss
453561151171 Power Supply,
Acquisition
3500-3699-01 Power Supply
Assembly, Acquisition
1700-0134-01 Power Supply,
Platform
3800-0082-01 Relay, Ss, 12 Vdc, 5 A
at 100 to 240 Vac
453561168951 Retainer, Cable, Arm, Monitor
Lower, Articulation articulation
Display
1065-6574-01 Retainer, Insert, Monitor
Display Articulation articulation
1065-6529-01 Retainer, Jack Screw, Monitor
Display Articulation articulation

Conquest Imaging - 216 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6153-05 Retainer, PCB, Acquisition card
Bottom, Left cage
1065-6272-04 Retainer, PCB, Acquisition card
Bottom, Right cage
1065-6152-04 Retainer, PCB, Top, Acquisition card
Left cage
1065-6271-04 Retainer, PCB, Top, Acquisition card
Right cage
1065-6694-01 Ring Brg, Elbow, Monitor
Swivel, Display articulation
Articulation
1065-6693-01 Ring Brg, Lower Arm, Monitor
Display Articulation articulation
1065-6699-01 Ring Brg, Yoke, Monitor
Display Articulation articulation
1065-6440-01 Ring, Retainer, Elbow, Monitor
Display Articulation articulation
1065-6695-01 Ring, Slip, Elbow, Monitor
Display Articulation articulation
1065-6549-01 Rod, Pivot, Brake System brake
Assy assembly
453561153071 Screw, 3/8-16,
Shoulder 1.0L, Skh,
Hd, Nylok, S, Z
1514-0060 Screw, 4-40x3/8, Pnh,
Sem, Pd, St, Zn
1516-0112 Screw, 6-32x1/4, Pnh,
Sems, Pd, St, Zn
1530-0344-01 Screw, 6-32x3/8, Flh,
Pd, 100D, St, Zn
1516-0074 Screw, 6-32x3/8, Pnh,
Int Sem, Pd, St, Zn
1530-0088-01 Screw, 6-32x3/8, Set, System brake
Flt Pt, Hd, Ss, Pas assembly

Conquest Imaging - 217 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1530-0428-01 Screw, 6-32x3/8, Skh, Monitor
Cap, Nlk, Nyltip, S articulation
1517-0060 Screw, 8-32x3/4, Pnd,
Sems, Pd, St, Zn
1518-0042 Screw, 10-32x1/4,
Pnd, Sems, Pd, St, Zn
1518-0041 Screw, 10-32x3/8,
Pnd, Int Sem, Pd, St,
Zn
1530-0103-01 Screw, 10-32x1/2, Flh,
Pd, 100D, Ss, Pas
1518-0051 Screw, 10-32x1/2,
Pnh, Sem, Pd, St, Zn
1530-0410-01 Screw, 10-32x.5, Flh,
Pd, 82D, Nylok, S, Z
1530-0157-01 Screw, 10-32x5/8,
Skh, Cap, Hd, Nylok,
Zn
1518-0043 Screw, 10-32x3/4,
Pnh, Sem, Pd, St, Zn
1065-6555-01 Screw, Brake Lever, System brake
Brake Assy assembly
1065-6438-01 Screw, Jack, Display Monitor
Articulation articulation
1630-0113-01 Screw, M3x6, Pnh, Pd,
Int Sems, St, Bz
1630-0128-01 Screw, M4.5x16, Phd, Video monitor
Td, Tf, St, Zn
1630-0025-01 Screw, M6x15, Din AC tray ground
42801, Hex, Spcl, Br, stud
Ni
1065-5751-03 Shield, Acquisition
Frontplane

Conquest Imaging - 218 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6707-01 Shield, Cable, Video, Monitor
Display Articulation articulation
3500-3739-01 Shield, EMI, DVD
453561157281 Shield, Strip, Clip-On, Acquisition card
Perp, 16-in Length cage fingers
1065-6528-01 Sleeve, Elbow, Monitor
Display Articulation articulation
2601-0145-01 Sleeve, Ferrite Used on V.90 USB
modem cable
453561161691 Sleeve, Mounting, Monitor
Display Articulation articulation
2950-1497-01 Slide Rail, Keyboard
3500-3627-01 Speaker Box Assy, Speaker box
Left assemblies
3500-3626-01 Speaker Box Assy, Speaker box
Right assemblies
1065-6361-01 Speaker Box, Left
1065-6362-01 Speaker Box, Lid, Left
1065-6359-01 Speaker Box, Lid,
Right
1065-6358-01 Speaker Box, Right
1065-6355-02 Speaker Grill, Left
1065-6354-02 Speaker Grill, Right
3500-3731-01 Speaker Module Assy
453561174311 Spreader, Disk Drive Housing for
Assembly CD/DVD drive,
power ON/STBY
switch and system
microphone
2950-1630-01 Spring, Disc, Monitor
5/16ROD, 0.49OD, articulation
0.0075T, Ss
2950-1636-01 Spring, Gas, 650N Monitor
articulation

Conquest Imaging - 219 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6391-01 Spring, Stand-by, Disk
Bay
4600-0312-01 Standoff, 6-32, 1.25L,
1/4 Hex, M/F, AL, IR
4600-0310-01 Standoff, 6-32,
0.375L, 1/4 Hex, M/F,
AL, IR
1065-6525-01 Stop, Pivot, Display Monitor
Articulation articulation
1065-2386-01 Strap, Universal OEM Used with strap
clips 1065-6205-
01. Strap clips are
used only for
OEMs positioned
on the lower OEM
deck shelf.
2210-0126 Tape, Velcro, 1W, For mounting the
Black, Type 170 modem to the
AVIO
2210-0125 Tape, Velcro, 1W, rear interface panel
Black, Type 400
3100-3161-01 Terminal Block, 9
Position
3100-3162-01 Terminal, Tab, 0.187,
#6 Stud MT, BRS, Tin
3500-3689-01 Toggle Block, Brake System brake
Assy assembly
2601-0100-01 Transformer, Power,
360 Vac, 5 Tap Input,
50-60 Hz
1065-6545-01 Tray, AC
1065-6233-01 Tray, Acquisition
Power Supply
1065-6560-01 Tray, Battery

Conquest Imaging - 220 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1065-6228-01 Tray, Filter
1065-5966-01 Tray, OEM Used on systems
with one or no
OEMs installed.
1565-0090-01 Washer, Equalize
Pressure, 0.28ID,
0.63OD, 0.19T, S
1560-0065 Washer, Flat, #4,
0.122ID, 0.280OD,
0.03T, St, Zn
1560-0046 Washer, Flat, #6,
0.156ID, 0.375OD,
0.049T, St, Zn
1560-0061 Washer, Flat, #6,
0.312OD, 0.032T, St,
Zn
1560-0051 Washer, Flat, #10,
0.203ID, 0.500D,
0.04T, St, Zn
1565-0088-01 Washer, Flat, #10,
0.406OD, 0.04Thk, St,
Zn
1560-0052 Washer, Flat, 1/4,
0.28ID, 0.50D,
0.065T, St, Zn
1565-0092-01 Washer, Flat, 2.5ID,
3.2OD, 0.04Thk, St,
Zn
1565-0091-01 Washer, Flat, 5/16, Monitor
0.875OD, 0.064Thk, articulation
St, Zn
1561-0036 Washer, S/R Lock,
1/4, 0.489OD, 0.062T,
St, Zn

Conquest Imaging - 221 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
1565-0087-01 Washer, Thrust, Monitor
0.375ID, 0.75OD, articulation
0.0625T, Brz
1065-6465-01 Wrist Rest, Keyboard, iU22 system
Enclosure
453561155151 Yoke, Mounting, Monitor
Display Articulation articulation
453561165301 Adapter, USB System option, The
Bluetooth Assembly, Bluetooth adapter
with 1.20 Software plugs into a USB,
(Bluetooth receiver) port located on the
AVIO-Rear
Interface Panel
453561153401 Cable Assy, Ethernet Part of accessory
Filter kit Not illustrated
453561151461 Headset, Bluetooth, System option
Sony HBH-35

Conquest Imaging - 222 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
Peripherals and
Applicable Cables
OEM
Part NumbersOEM Power Cable Signal Cable
Description
Printer, Color,
Sony, UPD- 453561153451
2100-1941-01 3500-3598-01
23MD, (USB)
Digital, USB
Printer, B&W,
Sony, UP- 453561153451
453561157291 3500-3598-01
D895MD, (USB)
Digital, USB
2100-1892-XX (Alt)
VCR,
Mitsubishi,
HSMD3000U, 3500-3595-01
3500-3542-01 3500-3598-01
with RS-232 (RCA)
I/F, 120V
(NTSC)
3500-3596-01
(RCA)
2100-1883-XX 453561153451
(Alt) (USB)
VCR,
Mitsubishi,
3500-3595-01
HSMD3000E, 3500-3543-01 3500-3598-01
(RCA)
with RS-232
I/F,
3500-3596-01
(RCA)
2100-1882-XX 453561153451
(Alt) (USB)

Conquest Imaging - 223 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/iE33™ Parts List (D.0/E.0)

iU22/1E33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
453561195283 Dual Power Supply
Assy.
PCB Assy,
Acquisition Power
Distribution Board
453561253751 Battery Tray Assy
PCB Assy, Battery and
Thermal Management
1700-0138-01 Battery, 6V, 8.0AH,
Sealed, Lead Acid
PCB Assy, HV
Switcher
453561185592 AC Tray Assy.
2700-0270-01 Circuit Breaker, 15A, 90–120 Vac
250V, w/Line Filter
2100-1910-01 Fan, 12 V, 360 MA,
105 CFM, 119 mm,
with Connector
453561160312 Caster, 5-in, Swivel Front
without Lock, with
Brake
453561163762 Caster, 5-in, SH ABS, Rear
Swivel Lock, without
Brake
453561259481 Power Supply Assy,
Mobility
453561270171 Fan Assembly,
Acquisition Card Cage
453561259291 Fan Assy, Platform
453561259951 AC Tray Assy, 90-
120V

Conquest Imaging - 224 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/1E33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
2100-1910-01 Fan, 12 V, 360 MA, Platform
105 CFM, 119 mm,
with
Connector

Conquest Imaging - 225 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms
iU22/1E33
Ultrasound System
Parts List
Part Number Part Description Notes/Reference
21715A X3-1 Transducer
(453561176611)
453561192601 X7-2 Transducer
21314A S5-1 Sector Array
(453561176741) Transducer
21778A S7-2omni
(453561181031) Transducer
21781A S7-3t
(453561181211) Transesophageal
Transducer
21750A S8-3 Sector Array
(453561181011) Transducer
21780A S12-4 Sector Array
(453561184501) Transducer
453561224871 C5-2 Curved Array
Transducer
453561190821 C8-5 Curved Linear
Transducer
4000-0944-02 L8-4 Linear Array
(453561167101) Transducer
453561177101 L9-3 Linear Array
Transducer
21756A L11-3 Linear Array
(453561181021) Transducer
453561153312 L15-7io Linear
Array Transducer
453561204991 D2tcd Static PW
Transducer
453561172792 D2cwc Transducer
4000-0990-01 D5cwc Transducer

Conquest Imaging - 226 -


Ultrasound Sales Service Repair Training
1815 Industrial Drive, Suite 100 Stockton, CA 95206
PH: 209-942-2654 FAX: 209-942-2572
Rev3 070208.dms