Sie sind auf Seite 1von 8

Article

pubs.acs.org/ac

Size Discrimination and Detection Capabilities of Single-Particle


ICPMS for Environmental Analysis of Silver Nanoparticles
Jani Tuoriniemi, Geert Cornelis, and Martin Hassellöv*
Department of Chemistry and Molecular Biology, University of Gothenburg, SE-412 96 Gothenburg, Sweden
*
S Supporting Information

ABSTRACT: The detection capabilities of single particle


inductively coupled plasma-mass spectrometry (spICPMS)
See https://pubs.acs.org/sharingguidelines for options on how to legitimately share published articles.

with respect to particle size and number concentrations are


Downloaded via UNIVERSIDADE DE LISBOA 00700 on January 3, 2019 at 12:40:25 (UTC).

investigated for the case of silver nanoparticles (ca. 20−80


nm). An iterative algorithm was developed where particle
measurement events were distinguished as outliers from the
more continuous dissolved ion signal if the measured intensity
was more than five times the standard deviation of the whole
data set. The optimal dwell time for 40−80 nm particles,
limiting both incomplete and multiple particle events, was 5
ms. The smallest detectable particle size (ca. 20 nm) is mainly
limited by the overlap of particle events and dissolved signal that increases with noise on both signals. The lowest measurable
number concentration is limited by the relative frequency of erroneously identified particle events, a limit that can be reduced by
acquiring more data points. Finally, the potential of spICPMS for environmental detection of nanoparticles is demonstrated for a
wastewater treatment plant effluent sample.

T he increasing application of engineered nanoparticles


(ENP) in consumer products is forecasted to have a major
impact on most segments of the future society, and ENP are
concentrations of ENP in aqueous samples. If a sufficiently
dilute suspension of metal-rich particles is nebulized into the
plasma, then a burst of ions will be generated when each
currently being used in a large variety of consumer products.1 discrete particle is vaporized, atomized, and ionized. The ions
Following indications that they are potentially harmful for of the mass of interest are subsequently counted with a very
organisms in the environment,2 the effect and exposure high frequency similar to the time scale of a particle event. The
assessment of ENP need to be supported by in situ analysis theory of spICPMS for liquid analysis was outlined by
and physicochemical characterization, which is difficult with Degueldre et al.,12−16 where they showed that the concen-
currently available methods. For instance, models predicting tration of particles can be derived from the time-resolved
the environmental concentrations of ENP in different compart- signals. They also experimentally demonstrated the feasibility of
ments such as surface waters, soils, and sediments3,4 need to be the method for a range of particles of different materials with
validated with environmental measurements,5 but the quanti- sizes larger than 80 nm.12−16 The spICPMS approach is further
fication of ENP in complex environmental matrixes is very explored in this study for the case of Ag nanoparticles (AgNP).
challenging due to the likely low environmental concen- AgNP can be found in many consumer products today,
trations3,4 (<106 particles mL−1; < pg L−1), the high abundance particularly because of their antibacterial properties, and
of natural nanoparticles,6 and the presence of interfering concerns have been raised about emissions into the environ-
elements.7 Similar challenges also apply to cosmetics, food, and
ment.17−21 spICPMS has already been applied for character-
nanomedicine protection goals. To address these challenges,
ization of the effluent of a washing machine in which AgNP are
new sensitive and selective methods working on a single
used.22 Laborda et al.23 estimated both AgNP particle number
particle level need to be developed.7−9 Moreover, given many
indications that the mode of action depends on the size of concentration and dissolved silver concentration from
ENP, these novel methods need to be able to at least quantify spICPMS measurements.
both particle numbers and size distributions. Electron A 3*σ threshold limit is usually adopted to discriminate the
microscopy methods have been shown to provide powerful ion bursts of a particle from the background signal. While 3*σ
information,10,11 but sample preparation artifacts, the high is a commonly used definition for detection limits (DL), this
particle count required, and unsuitability for high-throughput concept cannot be used for spICPMS where the intensity of
applications limit the applicability of these techniques for particle signals are outliers within a data set of which the mean
environmental samples.
The capabilities of single particle inductively coupled plasma- Received: December 5, 2011
mass spectrometry (spICPMS) are explored in this study as a Accepted: April 6, 2012
technique to simultaneously determine size and ultratrace Published: April 6, 2012

© 2012 American Chemical Society 3965 dx.doi.org/10.1021/ac203005r | Anal. Chem. 2012, 84, 3965−3972
Analytical Chemistry Article

is mainly determined by the continuous signal originating from nondigested Au and Ag nanoparticles smaller than 100 nm
dissolved analytes. Moreover, it can be calculated that 0.13% of diameter.26−28 However, it is possible that the f neb for particles
a Gaussian distributed dissolved signal is erroneously counted still differs slightly from that of the dissolved analyte in the
as particles if a 3*σ detection limit is assumed. A Poisson lower concentrations used in this study due to increased
distribution is probably more appropriate for mean dissolved adherence on, e.g., tubing and spray chamber walls. If this is the
signals approaching zero,23 and in this case, ca. 0.5% of the data case, it would influence number concentration measurements
points are erroneously counted as particles. Previous work on based on nebulization efficiency (see below), but the analyte
spICPMS12−16,23 did not elucidate how many particles need to mass per particle reaching the plasma is not affected by f neb.
be counted in order to reach a certain level of confidence in the Ipart, the signal intensity caused by one nanoparticle, can thus be
results. In addition, no method for determining the particle size calculated replacing the first part of eq 1 by the number of
without the use of standard particles of known sizes, which are atoms in one particle npart. It should thus also be possible to
currently not available for most nanoparticles, has hitherto been measure the mass of analyte in the particles using calibration
presented. In addition, the previous studies by Degueldre et with dissolved standards.
al.12−16 only showed detection of particles of 80 nm and above. Distinguishing Particle Events and Size Detection
More recent work showed that 40 nm AgNP can be Limits. Counting particle events in this study is based on
discriminated from the dissolved signal,23 but enhanced detecting outliers in the more continuous dissolved signal. In
reactivities and toxicity have been demonstrated for sizes the ideal case, the signal intensities of dissolved ions do not
below approximately 30 nm.24,25 Consequently, there is a need overlap with nanoparticle signal intensities in a frequency−
to further develop and optimize spICPMS to reduce the lowest intensity diagram. However, reducing particle size may lead to
detectable size limit and to improve the methods for overlap with the dissolved signal, due to the cubic relation of
discriminating between the particle events and the background particle size and mass. Additionally, all noise factors of the
signal. dissolved and/or particulate signal increase overlap, e.g.,
The current study investigates the detection capabilities for increasing number of incompletely measured particle events
AgNP using a sector-field ICPMS (Thermo-Finnigan Element or low stability in the dissolved signal. Concentration detection
2). The software of this instrument allows one to reduce dwell limits in conventional ICPMS are usually stated in terms of
times (also called sampling time), i.e., the detector acquisition standard deviation of the background signal, σb. The size
time for each data point below 5 ms and the minimum dwell detection limit in spICPMS may similarly be expressed in terms
time used in previous work.12−16,23 It is discussed how this of standard deviation of the dissolved signal σdiss, but it is not
affects the size detection limits of spICPMS. A method for possible to calculate σdiss unless one can distinguish the
determining particle size using calibration with dissolved dissolved signal from the particle events. Mitrano et al.26
standards is also presented, and data analysis is improved. In proposed an iterative algorithm based on a detection limit 3*σ,
addition, the number concentration detection limit of spICPMS where σ is the standard deviation of the whole data set. The
and how it may be improved is discussed. Finally, as a feasibility data points exceeding μ + 3*σ, where μ is the mean of the data
check, an environmental sample from a wastewater treatment set, are considered nanoparticle signals and are removed.
plant (WWTP) effluent is analyzed. Calculation of μ + 3*σ is then repeated multiple times, and


nanoparticle signals are removed from the remaining data
THEORY points until no further data points are removed. More details
can be found in Supporting Information. The resulting
Signals of Dissolved and Particle Bound Analytes. The
detection threshold signal intensity for particle detection set
number of ions of dissolved analytes, Idiss, arriving at the
by this algorithm is for a low frequency of particle events
detector in each data point (dwell) during the dwell time, tdwell
approximately Idiss + n*σdiss. This sets the requirement Ipart >
(s), is given by:23
n*σdiss for a particle event to be recognized as such (derivation
Idiss = tdwellfneb cdissAq × fis fion fplasma ftrans in Supporting Information). The choice of detection threshold,
(1)
n*σ, in the algorithm is thus critical for the analysis, and n will
Equation 1 contains two parts that account for dilution and be varied in this study.
efficiency factors in the sample introduction system, and plasma Counting Particle Events and the Particle Number
and mass spectrometer, respectively. In the first part giving the Concentration Detection Limit. The average number of
number of analyte atoms introduced to the plasma, f neb is the particles λ entering the plasma during a dwell, based on the first
fraction of the analyte that passes through the nebulizer and part of eq 1, is given by:
spray chamber, the nebulization efficiency, cdiss, is the λ = qfneb c ptdwell (2)
concentration of ions in the sample in mol L−1, A is Avogadro’s
number, and q is the flow rate of sample into the nebulizer in where cp is the particle number concentration in the sample.
Ls−1. The second part is the counting yield that determines the The probability, pk, for obtaining single (k = 1) and multiple (k
fraction of analyte atoms entering the plasma that are detected. > 1) particle events and the fraction of dwells containing
f is is the abundance of the monitored isotope, f ion is the fraction particle events, pp, can be calculated using Poisson statistics
of analyte that becomes singly charged positive ions in the (further discussion in Supporting Information). The total
plasma, f plasma is the fraction of analyte exiting the plasma number of particle events counted in a sequence of D dwells is
through the sampler cone, and f trans is the transport efficiency or D*pp. At low λ, the chance for multiple particle events, pk>1, is
the fraction of ions that is transported from the sampler cone to negligible relative to pk=1 and the approximation pp ∼ λ is valid.
the detector. The terms in the second part of eq 1 are most The particle count at the detector increases linearly with cp in
likely equal for dissolved or particulate signals because several this case but, at higher λ, i.e., higher cp, nebulization efficiency,
studies have shown that there is no significant difference flow rate, and/or dwell time, a lower particle count than
between conventional ICPMS analysis of acid digested or predicted from the initial linear trend is obtained. The particle
3966 dx.doi.org/10.1021/ac203005r | Anal. Chem. 2012, 84, 3965−3972
Analytical Chemistry Article

Figure 1. Time resolved signals of spICPMS and the corresponding signal intensity frequency histograms for samples of ultrapure water, a 1 μg L−1
dissolved Ag standard, a fresh 40 nm AgNP suspension, and an aged 40 nm AgNP suspension. The dashed lines indicate the 5*σ detection threshold.

number concentration can be determined either on the basis of for this instrument is found in Supporting Information.
measured nebulization efficiency and using eq 2 or from Dissolved standards were included in each analysis sequence,
calibration curves prepared from particle standards (see below). and the nebulization efficiency was measured in triplicate using
The particle number detection limit can be related to the the waste collection method.30,31 A detailed description of the
relative error in particle counting Er (derivation in Supporting measurement of nebulization efficiency and calibration of
Information): particle size using dissolved standards is found in Supporting
Information. The nanoparticle stock suspensions were diluted
1
Er = 2z in series up to 109 times in ultrapure water. The concentrations
N (3) are listed in Table SI-2, Supporting Information. The nominal
where z is the z-score of the normal distribution which assumes dwell time was set at 1, 5, or 10 ms, and 10 000 or 40 000
the value of 1.96 for a 95% confidence interval.29 Given a dwells of 107Ag measurements were collected for each dilution.
desired accuracy, the particle number detection limit thus Calibration curves, frequency histograms, and particle counts
depends only on the number of particles counted, which were calculated from the exported raw data using spICPMS
increases with D. theory in MATLAB (version 6). For each dilution series, the

■ EXPERIMENTAL SECTION
Chemicals. AgNP suspensions were obtained from British
particle concentrations were determined using eq 2 and the
measured nebulization efficiency on a sample with a high
enough concentration to obtain sufficient counting statistics
biocell international (BBI, UK) (20, 40, 60, and 80 nm in (details in Table SI-2, Supporting Information). These
nominal diameter). The particles were citrate-coated and measured values were used to calculate the concentrations of
consisted of elemental silver. Unless otherwise stated, the the other samples in the dilution series using the dilution factor.
samples were measured briefly after the first opening of the WWTP Samples. WWTP effluents were sampled at the
container. Reference Au nanoparticle suspensions (30 and 60 GRYAAB WWTP (Gothenburg, Sweden) the 19th of May
nm nominal diameter) were obtained from NIST. Dissolved 2009 and analyzed the same day. The sample was a daily flow
silver standards (ultra scientific) for ICPMS calibration were composite sample from the final treated water from which three
diluted in ultrapure water from a 1000 ppm solution in 2% subsamples were filtered through an acid rinsed and neutralized
HNO3. 5.0 μm Millipore Durapore Millex (PVDF syringe filter with 3.9
AgNP spICPMS Measurements. A Thermo Element 2 cm2 filter area) followed by a 0.45 μm Millipore Durapore
ICPMS (settings given in Table SI-2, Supporting Information) Sterivex-HV (PVDF filter with 10 cm2 filter area). The first 2
was used in all experiments after sensitivity and short-term mL were discarded, and 10 mL of filtrate was collected into acid
stability optimization using 1 μg L−1 In in 2% HNO3. A washed and neutralized polycarbonate vials. The dwell time was
discussion regarding issues with the dwell times that are specific not optimized for these samples, and the minimum possible
3967 dx.doi.org/10.1021/ac203005r | Anal. Chem. 2012, 84, 3965−3972
Analytical Chemistry Article

dwell time, 0.1 ms, was used together with a relatively large
number of dwells (D = 65 500) to maximize the chance of
detecting a significant number of nanoparticle events in this
first feasibility study that did not involve an elaborate sample
characterization. To calibrate for particle number concentration
dilution series of the NIST, Au standard particles were
measured (calibration curves in Figure SI-8, Supporting
Information). Concentrations in the WWTP samples were
obtained from the calibration curves containing the frequency
of particle events plotted against the number concentration
provided by the manufacturer. To ensure that sensitivity was
not size-dependent to any significant extent, the calibration
curves were measured for both 30 and 60 nm Au nanoparticle
sizes.

■ RESULTS AND DISCUSSION


Dissolved and Particulate Signals. Typical spICPMS raw
data and frequency histograms using 5 ms dwell time are shown
in Figure 1. The low intensity dissolved signal and the particle
related spike in ultrapure water is due to both impurities and
carry-over from previous samples. The signal of the 1 μg L−1
Figure 2. Frequency−intensity histograms of 80 nm AgNP measured
dissolved Ag standard has a RSD of 34% and is relatively with dwell times of 1, 5, and 10 ms. The dashed lines indicate the
symmetrically distributed around a mean close to 100 ion particle detection threshold.
counts. Even in the “fresh” 40 nm AgNP sample (analyzed
within one week after delivery), one can recognize a nonzero
dissolved signal resembling that of ultrapure water and particle
events that appear as intense spikes. The intensities of these time because increasing tdwell to 10 ms does not increase
particle spikes are lower for the “aged” particle suspension resolution of the particle signal from the dissolved signal. The
(analyzed 3 months after opening) compared to the fresh broad tail in this sample can be explained by multiple particle
suspension, suggesting a particle size decrease while the events. The overlap in the data with tdwell = 1 ms is high because
dissolved signal intensity is increased, probably due to partial of a large proportion of incompletely measured particle events.
dissolution upon exposure of the fresh standard to oxygen. The Similar trends were observed for 40 and 60 nm particles
overlap of nanoparticle and dissolved signal is more (Figures SI-4 and SI-5, Supporting Information).
pronounced in the aged standard. Figure 1 also shows that a Measurement of Particle Size. The particle sizes were
5*σ threshold value indeed identifies most measured intensities determined by calibrating the ion count with respect to mass of
of ultrapure water and 1 μg L−1 dissolved Ag as below the Ag introduced into the plasma using dissolved standards.
particle detection threshold, whereas the nanoparticle events of Provided that the chemical composition is known, this allows
the fresh 40 nm AgNP suspension are identified as outliers and estimation of the particle size distribution by assuming a
are thus retained as nanoparticle events. This approach also spherical shape for all particles. The calculated diameters are
demonstrates the difficulty in distinguishing nanoparticles from thus corresponding volumetric spherical diameters. Figure 3
dissolved signals in the case of overlapping signals as is the case shows the calculated size distributions of 20, 40, 60, and 80 nm
for the aged 40 nm AgNP suspension. Both the fresh and aged particles calculated from data obtained using a 5 ms dwell time.
40 nm standard may contain false positives, i.e., signal The calculated diameters of the 40, 60, and 80 nm AgNP
intensities of dissolved Ag that are above the 5*σ threshold standards are in relatively close agreement with the nominal
and that are thus erroneously regarded as particle signals, but diameters. Diameter calculation requires the nebulization
the proportion of false positives to total detected particle events efficiency, a value that was determined in this study using an
is likely higher in the aged sample than in the fresh sample. indirect determination of which the accuracy is disputed,
Analogous data for 1 ms dwell time is shown in Figure SI-3, especially when the nebulization efficiency is relatively low (i.e.,
Supporting Information. < 5%).30 However, it can be shown (Supporting Information)
Effect of Dwell Time. Figure 2 shows frequency−intensity that any error in nebulization efficiency is reduced to its third
histograms for 80 nm Ag nanoparticles measured using 1, 5, root when propagated to an error in diameter. Figure 3 shows
and 10 ms dwell times. While the intensity of the dissolved that a 50% variation of the nebulization efficiency results in ca.
signal increases with tdwell, the particle signal does not. The ± 10 nm variation of the peak maximum of 80 nm particles.
overlap between dissolved and nanoparticle signals thus reduces The indirectly measured nebulization efficiencies were 15−20%
as tdwell is reduced (equations found in Supporting Informa- where the discrepancy with, e.g., directly measured nebulization
tion). However, the standard deviation of the dissolved signal efficiency, a value that is generally regarded as more accurate, is
and the probability that a particle event is only partially likely lower than 50%.30 Moreover, direct methods cannot be
measured increase with reducing tdwell, especially when the run at the same time as the spICPMS determination, whereas
dwell time is comparable to or lower than the duration of the nebulization efficiency is highly dependent on sample
particle event, (ca. 0.1−0.5 ms).32 Shortening tdwell can thus also introduction settings33 that may differ from the conditions
increase the overlap of dissolved and nanoparticle signals, and during a direct nebulization efficiency determination where the
an optimum tdwell with a minimal signal overlap therefore exists. plasma is not ignited. However, the accuracy of the nominal
Figure 2 shows that 5 ms is in most cases an optimum dwell particle sizes was not investigated in this study that was not
3968 dx.doi.org/10.1021/ac203005r | Anal. Chem. 2012, 84, 3965−3972
Analytical Chemistry Article

deviation rather than the sensitivity of the ICPMS. Further


improvement of the detection capabilities with respect to size
should thus focus on reducing the dissolved signal intensity and
its standard deviation. Since the samples are diluted up to 109
times to avoid measuring multiple particle events, the dissolved
background predominantly consists of sample carry-over and
not dissolved silver coming from the sample itself. However, if a
detectable concentration of dissolved silver from the sample
remains, the magnitude of its signal can be reduced by
decreasing the sample uptake rate. Reducing the noise of the
dissolved signal can be achieved by, e.g., monodisperse droplet
delivery32,37 or a Peltier cooled desolvation system37 by which
it is possible to partially resolve 20 nm AgNP from the
dissolved background, whereas this is not possible with
conventional sample introduction.37 Developing algorithms
capable of deconvoluting overlapping signals can also
contribute to detection capabilities. The theoretical smallest
detectable equivalent spherical diameter is 6.1 nm, because this
value corresponds to the approximate minimum silver mass
needed to deliver at least one 107Ag ion to the detector; given
that knowing the mass of silver introduced to the plasma during
dwell, it can be calculated that only ca. 1/30 000 of the 107Ag
atoms reaches the detector in a typical measurement.
Choice of Detection Threshold. The number of particles
counted per 10 000 dwells are plotted as a function of the
parameter n for 20 and 40 nm AgNP and for a 2 μg L−1
dissolved standard using a 5 ms dwell time (Figure 4; data for 1

Figure 3. The size distributions of the 20, 40, 60, and 80 nm nominal
diameter Ag particles determined by spICP-MS. The error bars
demonstrate the effect of possible 50% errors on the determination of
nebulization efficiency on the calculation of size. A tdwell of 5 ms was
used except for the 20 nm AgNP for which it was 1 ms.

aimed at demonstrating the accuracy of particle size


determination of spICPMS.
Since spICPMS measures the mass of particles that varies
with the third power of the diameter, this technique can
potentially achieve a very high resolution in diameter. However,
Figure 4. The number of particle events counted by the iterative
studies on single droplets introduced into plasma indicate that,
algorithm as a function of threshold value (n) for data obtained using a
e.g., variation in the point of vaporization and trajectories that 5 ms dwell time. The following data sets are included: 2 μg L−1
lie off the sampler cone axis may introduce variation in the dissolved Ag and 40 and 20 nm AgNP suspensions.
intensity of particle signals.34−36 Since there is no information
about the polydispersity of the samples, the size resolution
capabilities could not be assessed in this study. As for particle ms dwell time is shown in Figure SI-6, Supporting
sizing by microscopy, a large number of particles are required Information), suggesting that an n value of at least 4 must be
for the estimation of the mean diameter with sufficient used to reduce the number of false positives to less than 0.1% of
accuracy, and the sizing capabilities will ultimately be the total count in the case of the dissolved standard. The
determined by the available sample volume. particle count should thus reach a value that is independent of n
The distributions of particle event signals of the 20 nm if n > 4, in the absence of false positives. This behavior is
particles were to a large extent overlapping with the dissolved approached for the 40 nm particles measured at 5 ms dwell
signal although a dwell time of 1 ms was used. Only the high time where the particle events are completely resolved from the
intensity tail corresponding to 20−25 nm volume equivalent dissolved signal (Figure 3). In the case of 40 nm particles
spherical diameters could be discriminated. Figure 3 thus measured at tdwell = 1 ms and for all 20 nm data, the particle
confirms that 20 nm AgNP are possible to detect and to some events are not resolved from the background and it can be seen
extent possible to discriminate. The size detection limit is for that the particle count continues to decrease upon increasing n.
these samples set by the dissolved background and its standard Increasingly more true particle counts are removed together
3969 dx.doi.org/10.1021/ac203005r | Anal. Chem. 2012, 84, 3965−3972
Analytical Chemistry Article

with false positives in this case because of the overlap of particle depends on the determination of nebulization efficiency, and
and dissolved signals. It is evident that to reduce the number of the ability to discriminate particle events from the dissolved
false positives to a predefined acceptable level, e.g., 0.1% of the signal. Although the number concentrations were not
total count, n values considerably higher than 3 must be used to confirmed in this study, Figure 5 demonstrates the dynamic
reach this level over a broad concentration range. However, range and linear concentration dependent response of
when very high values for n are used, particles are omitted from spICPMS particle counting. The contribution of false positives
counting. For the remainder of the discussion, n = 5 was is significant when the particle concentration is low and the
therefore chosen as a compromise between retaining a particle number is consequently overestimated. The frequency
sufficiently high particle count over a broad concentration of false positives thus sets the limit of detection with respect to
range and ensuring that contribution of false positives is less concentration when present. At higher concentrations, the
than 0.1% of the actual particle events. (Particle count over a counting yield decreases due to the occurrence of multiple
range of concentrations are compared for n values of 3, 5, and 8 particle events that are each counted as one particle event and a
in Figure SI-7, Supporting Information.) downward slope can be observed. This behavior is observed
Concentration Calibration Curves. The logarithm of regardless of size or dwell time for the particles investigated
particle count divided by concentration is plotted as a function here. A plateau corresponding to the linear response to
of the logarithm of concentration for the 20, 40, 60, and 80 nm concentration is obtained over ca. 2 orders of magnitude of
AgNP measured using 1 and 5 ms dwell time in Figure 5. The particle number concentrations, even if particle events are not
resolved from the background, as is the case for 20 nm AgNP
and measurements using 1 ms dwell time. This suggests that
the iterative algorithm may also be successful if the nanoparticle
and dissolved signal intensities partly overlap. However, the
particle count is less accurate in this case because a fraction of
the true particle events are not counted as such. Moreover, any
error in nebulization efficiency is also propagated to the particle
count, and the effective dwell time in eq 2 is not necessarily
equal to the nominal tdwell (further discussion and illustration in
Figure SI-1, Supporting Information).38
Effect of Number of Dwells. Figure 6 shows how the
linear range of particle number determination depends on

Figure 6. The logarithm of counted particle events divided by the


concentration as a function of the logarithm of concentration for 40
nm AgNP measured using a 5 ms dwell and 5*σ detection threshold.
Figure 5. The logarithm of counted particle events divided by the
The number of data points used was varied by randomly selecting
concentration as a function of the logarithm of concentration. The
from the 40 000 measured data points for each sample.
data obtained for 20, 40, 60, and 80 nm Ag particles using 1 ms (left)
and 5 ms (right) dwell time and 5*σ detection threshold is shown. 10
000 dwells were used for each sample. The boxes indicate the region
where the relative frequency of false positives (solid line) and multiple
randomly selecting smaller data point sets among the 40 000
particle events (dashed line) becomes significant. measured dwells. When D = 1000, only one or no particles are
counted for the first 7 data points and the true particle number
is increasingly overestimated at lower concentrations, because
particles were counted with a 5*σ detection threshold. The data the number of false positives becomes more significant relative
points in such plots are on horizontal lines in the particle to the total count. Moreover, there is more scattering in data
concentration range where true particle counts dominate over obtained from smaller data sets, because of higher statistical
false positives and a linear response of the experimental particle fluctuations. Increasing D to 5000 reduces this scatter and a
count therefore occurs. The level of these lines depends on the linear response is obtained over at least 2 orders of magnitude,
measured particle number concentration, whose accuracy but the scatter in data is still significantly higher than for D >
3970 dx.doi.org/10.1021/ac203005r | Anal. Chem. 2012, 84, 3965−3972
Analytical Chemistry Article

10 000. Figure 6 thus shows that the concentration detection applications, e.g., additive to diesel fuel to improve combustion
limit can be decreased by increasing D. However, the extent by efficiency, support in the automobile catalysts, and as chemical-
which this limit is reduced is lower at higher D values as Figure mechanical polishing in certain applications, e.g., electronics
6 shows that increasing D from 20 000 to 40 000 provides industry. The Gothenburg wastewater system receives road
much less reduction in scatter and the proportion of false runoff and industrial and domestic wastewater, so the Ce
positives. The results agree with eq 3. particles may originate from man-made sources. TiO2 nano-
Wastewater Effluent Samples. Of the screened elements particles are used in a large number of applications from
(Ag, Ce, Ti, Si, Zn, Cr, Cu, Mo, Pt, Sb, W, Y, and Zr), only the sunscreens and other cosmetics to paint.43 The predicted
Ag, Ce, and Ti analyses demonstrated particle events (Figure environmental concentration in wastewater effluents for TiO2
7). The filtration through the 450 nm nominal cutoff filters sets in Europe is 3.5 μgL−139 leading to 1 975 300, 246 920, and 73
160 particles mL−1, assuming spherical 20, 40, or 60 nm sizes,
respectively; values that are again comparable to the measured
32 656 mL−1, considering the large uncertainty on modeled
values. However, natural sources cannot be ruled out for both
Ce- and Ti-containing nanoparticles. Both brookite TiO2 and
Ce bearing allanite, for instance, have been found in river
sediments.11

■ CONCLUSIONS AND OUTLOOK


AgNP spICPMS signal intensities can be resolved from
dissolved signal intensities when the size exceeds ca. 20 nm
and the dwell time exceeds 1 ms. When these signal intensities
are not fully resolved, the particle number determination is less
accurate. Further improvement of the detection limit can be
achieved by reducing the dissolved signal intensity and its noise,
e.g., using monodisperse drop delivery.32 The choice of a
proper detection threshold is important for the quantification
of the particle number concentration. Using n values of ca. 5
compromises between a relatively low frequency of false
positives and a relatively high effective particle count. The
lowest detectable concentration is determined by the number
of acquired dwells but also by the relative frequency of false
positives. The volume equivalent spherical diameters calculated
from the signal intensities using calibration with dissolved
standards agree with the nominal diameters of the particles,
Figure 7. Time resolved signals of spICPMS for wastewater effluent suggesting that spICPMS can be used as a fast and sensitive
samples. The monitored elements were Ti, Ce, and Ag.
particle sizing method for environmental samples.
The detection of particles in an environmental sample was
an upper limit on the particle sizes. Gold particle reference demonstrated. The shortest possible tdwell of 0.1 ms was used in
materials were used to estimate the particle number order to minimize the size detection limit. Increasing the tdwell
concentrations (Figure SI-8, Supporting Information). The increases sensitivity with respect to concentration and allows
dwell time of 0.1 ms does not allow accurate sizing of the more accurate determination of the mass of analyzed element
particles. The numbers of particle events were low and in the particle and thus its size. With the instrumentation used
therefore a 8*σ detection threshold was chosen to obtain a here, it is only possible to measure a single mass within the
sufficiently low frequency of false positives. The concentrations duration of a particle event which does not allow determination
were 9568 particles mL−1 for Ag, 2312 particles mL−1 for Ce, of the composition of individual particles. This limits the
and 32 656 particles mL−1 for Ti. The predicted environmental applicability of the method for unknown samples, where it
nano-Ag concentration in wastewater effluents for nano-Ag in would be interesting to adapt the concepts developed in this
Europe is 42 ng L−1.39 Assuming a spherical particle size of 20, paper for instruments capable of simultaneous multielement
40, and 60 nm, this predicted concentration corresponds to 95 detection.44 Such a method could, for instance, detect whether
493, 11 937, 354 particles mL−1, suggesting that the measured the detected particles are in fact AgNP or whether these are
particle number concentration in WWTP is in the order of transformed into Ag2S nanoparticles.


magnitude of the predicted concentration. It is unknown to
what extent AgNP are being emitted as particles or whether ASSOCIATED CONTENT
they in fact occur as dissolved ions in applications such as food
packaging plastics, kitchen utensils and surfaces, textiles,
*
S Supporting Information

washing processes, etc., but recent investigations suggest that Additional information as noted in the text. This material is
available free of charge via the Internet at http://pubs.acs.org.


both silver forms are reprecipitated as nanoscale silver sulfide
(Ag2S) particles in wastewater treatment plants.40−42 spICPMS
analysis does not allow one to verify whether the detected silver AUTHOR INFORMATION
particles are elemental silver nanoparticles or silver sulfide, Corresponding Author
which in this case would also impede size estimation. *E-mail: martin.hassellov@chem.gu.se. Tel.: +46-31-786 9050.
Manufactured CeO2 nanoparticles are used in a number of Fax: +46-31-77227856.
3971 dx.doi.org/10.1021/ac203005r | Anal. Chem. 2012, 84, 3965−3972
Analytical Chemistry Article

Notes (26) Mitrano, D. M.; Lesher, E. K.; Bednar, A.; Monserud, J.;
The authors declare no competing financial interest. Higgins, C. P.; Ranville, J. F. Environ. Toxicol. Chem. 2011, 31, 115−


121.
ACKNOWLEDGMENTS (27) Allabashi, R.; Stach, W.; de la Escosura-Muñiz, A.; Liste-Calleja,
L.; Merkoçi, A. J. Nanopart. Res. 2009, 11, 2003−2011.
We thank the following funding agencies for support: CEFIC (28) Krystek, P.; Ulrich, A.; Garcia, C. C.; Manohar, S.; Ritsema, R. J.
Long range research initiative, European Commission FP7 Anal. At. Spectrom. 2011, 26, 1701−1721.
projects NANOFATE and MARINA, the Swedish Environ- (29) Levine, D. M.; Ramsey, P. P.; Smidt, R. K. Applied Statistics for
mental Research Council FORMAS, and the Gothenburg Engineers and Scientists; Prentice Hall: Upper Saddle River, NJ, 2001.
graduate school Environment and Health. We thank Dr. (30) Smith, D.; Browner, R. Anal. Chem. 1982, 54, 533−537.
Nicklas Paxeus at GRYAAB for help with the wastewater (31) Gustavsson, A. Spectrochim. Acta 1986, 41B, 291−294.
sampling. (32) Gschwind, S.; Flamigni, L.; Koch, J.; Borovinskaya, O.; Groh, S.;


Niemax, K.; Gunther, D. J. Anal. At. Spectrom. 2011, 26, 1166−1174.
REFERENCES (33) Sharp, B. L. J. Anal. At. Spectrom. 1988, 3, 939−963.
(34) Dziewatkoski, M. P.; Daniels, L. B.; Olesik, J. W. Anal. Chem.
(1) Roco, M. C. J. Nanopart. Res. 2011, 13, 427−445. 1996, 68, 1101−1109.
(2) Klaine, S. J.; Alvarez, P. J. J.; Batley, G. E.; Fernandes, T. F.; (35) Olesik, J. W.; Kinzer, J. A.; McGowan, G. J. Appl. Spectrosc.
Handy, R. D.; Lyon, D. Y.; Mahendra, S.; McLaughlin, M. J.; Lead, J. 1997, 51, 607−616.
R. Environ. Toxicol. Chem. 2008, 27, 1825−1851. (36) Stewart, I. I.; Hensman, C. E.; Olesik., J. W. Appl. Spectrosc.
(3) Gottschalk, F.; Sonderer, T.; Scholz, R. W.; Nowack, B. Environ. 2000, 54, 164−174.
Sci. Technol. 2010, 43, 9216−9222. (37) Franze, B.; Stenge, I.; Engelhard, C. J. Anal. At. Spectrom. 2012,
(4) Arvidsson, R.; Molander, S.; Sanden, B. A.; Hassellov, M. Hum. DOI: 10.1039/C2JA00003B.
Ecol. Risk Assess. 2011, 17, 245−262. (38) Heithmar, E. M. Abstracts of papers of the American Chemical
(5) Paterson, G.; Macken, A.; Thomas, K. V. Anal. Methods 2011, 3, Society, 237th ACS National Meeting, Salt Lake City, UT USA, 2009;
1461−1467. Vol. 237, pp 58−COLL.
(6) Gallego-Urrea, J. A.; Tuoriniemi, J.; Pallander, T.; Hassellov, M. (39) Gottschalk, F.; Sonderer, T.; Scholz, R. W.; Nowack, B. Environ.
Environ. Chem. 2010, 7, 67−81. Sci. Technol. 2009, 43, 9216−9222.
(7) Howard, A. G. J. Environ. Monit. 2010, 12, 135−142. (40) Choi, O.; Deng, K. K.; Kim, N. J.; Ross, L.; Surampalli, R. Y.;
(8) Hassellöv, H.; Kaegi, R. In Environmental and Human Health Hu, Z. Q. Water Res. 2008, 42, 3066−3074.
Impacts of Nanotechnology; Lead, J. R., Smith, E., Eds.; John Wiley & (41) Kaegi, R.; Voegelin, A.; Sinnet, B.; Zuleeg, S.; Hagendorfer, H.;
Sons, Ltd: Chichester, U.K., 2009. Burkhardt, M.; Siegrist, H. Environ. Sci. Technol. 2011, 45, 3902−3908.
(9) Von der Kammer, F.; Ferguson, P. L.; Holden, P. A.; Masion, A.; (42) Kim, B.; Park, C. S.; Murayama, M.; Hochella, M. F. Environ. Sci.
Roger, K. R.; Klaine, S. J.; Koelmans, A. A.; Horne, N.; Unrine, J. M. Technol. 2010, 44, 7509−7514.
Environ. Toxicol. Chem. 2011, 31, 32−49. (43) Kaegi, R.; Ulrich, A.; Sinnet, B.; Vonbank, R.; Wichser, A.;
(10) Kaegi, R.; Wagner, T.; Hetzer, B.; Sinnet, B.; Tzuetkov, G.; Zuleeg, S.; Simmler, H.; Brunner, S.; Vonmont, H.; Burkhardt, M.;
Boller, M. Water Res. 2008, 42, 2778−2786. Boller, M. Environ. Pollut. 2008, 156, 233−239.
(11) Plathe, K. L.; von der Kammer, F.; Hassellov, M.; Moore, J.; (44) Schilling, G. D; Andrade, F. J.; Barnes, J. H.; Sperline, R. P.;
Murayama, M.; Hofmann, T.; Hochella, M. F. Environ. Chem. 2010, 7, Denton, M. B.; Barinaga, C. J.; koppennaal, D. W.; Hieftje, G. M. Anal.
82−93. Chem. 2007, 79, 7662−7668.
(12) Degueldre, C.; Favarger, P. Y. Colloids Surf., A 2003, 217, 137−
142.
(13) Degueldre, C.; Favarger, P. Y. Talanta 2004, 62, 1051−1054.
(14) Degueldre, C.; Favarger, P. Y.; Bitea, C. Anal. Chim. Acta 2004,
518, 137−142.
(15) Degueldre, C.; Favarger, P. Y.; Wold, S. Anal. Chim. Acta 2006,
555, 263−268.
(16) Degueldre, C.; Favarger, P. Y.; Wold, S. Talanta 2006, 68, 623−
628.
(17) Luoma, S. In Project on emerging nanotechnologies; Woodrow
Wilson international centre for scholars, The PEW charitable trusts,
2008.
(18) Benn, T. M.; Westerhoff, P. Environ. Sci. Technol. 2008, 42,
4133−4139.
(19) Geranio, L.; Heuberger, M.; Nowack, B. Environ. Sci. Technol.
2009, 43, 8113−8118.
(20) Kaegi, R.; Sinnet, B.; Zuleeg, S.; Hagendorfer, H.; Mueller, E.;
Vonbank, R.; Boller, M.; Burkhardt, M. Environ. Pollut. 2010, 158,
2900−2905.
(21) Benn, T.; Cavanagh, B.; Hristovski, K.; Posner, J. D.;
Westerhoff, P. J. Environ. Qual. 2010, 39, 1875−1882.
(22) Farkas, J.; Peter, H.; Christian, P.; Gallego-Urrea, J. A.;
Hassellov, M.; Tuoriniemi, J.; Gustafsson, S.; Olsson, E.; Hylland, K.
Environ. Int. 2011, 37, 1057−1062.
(23) Laborda, F.; Jimenez-Lamana, J.; Bolea, E.; Castillo, J. R. J. Anal.
At. Spectrom. 2011, 26, 1362−1371.
(24) Aruguete, D. M.; Liu, J.; Hochella, M. F., Jr. In Environmental
and Human Health Impacts of Nanotechnology; Blackwell Publishing,
Ltd.: West Sussex, United Kingdom, 2009; pp 79−108.
(25) Auffan, M.; Rose, J.; Bottero, J.-Y.; Lowry, G. V.; Jolivet, J.-P.;
Wiesner, M. R. Nat. Nanotechnol. 2009, 3, 634−641.

3972 dx.doi.org/10.1021/ac203005r | Anal. Chem. 2012, 84, 3965−3972

Das könnte Ihnen auch gefallen