Drop Height of 10 kg - drop weight = 72,0 cm calibration factor for deflection amplitude = 0,943
force [N] pulse force [N] pulse force [N] pulse ref. mean tester differ.
Nr. length [ms] length [ms] length [ms] Nr. ref. 1 [mm] ref. 2 [mm] ref. 3 [mm] [mm] [mm] [mm]
1 7.075 16,07 7.082 16,09 7.098 16,12 1 0,524 0,462 0,510 0,499 0,506 0,008
2 7.072 16,07 7.081 16,10 7.098 16,14 2 0,511 0,470 0,518 0,500 0,508 0,008
3 7.066 16,04 7.088 16,13 7.092 16,15 3 0,515 0,466 0,514 0,498 0,510 0,012
4 7.069 16,07 7.086 16,10 7.092 16,11 4 0,509 0,471 0,522 0,501 0,505 0,005
5 7.068 16,07 7.092 16,12 7.095 16,12 5 0,510 0,471 0,524 0,502 0,503 0,002
6 7.071 16,08 7.086 16,09 7.088 16,17 6 0,515 0,467 0,520 0,501 0,508 0,007
7 7.075 16,10 7.079 16,13 7.094 16,12 7 0,506 0,473 0,517 0,499 0,502 0,004
8 7.078 16,10 7.085 16,12 7.086 16,12 8 0,512 0,471 0,522 0,502 0,506 0,005
9 7.071 16,10 7.080 16,11 7.092 16,16 9 0,516 0,465 0,518 0,500 0,503 0,004
10 7.076 16,11 7.093 16,11 7.078 16,18 10 0,526 0,469 0,518 0,504 0,505 0,001
mean: 7.072,1 16,08 7.085,2 16,11 7.091,3 16,14 mean: 0,500 0,506 0,005
Stddev. 3,8 0,021 4,8 0,015 6,0 0,025 Stddev. 0,002 0,002 0,003
T10 1,535 1,360 1,616 1,342 1,110 1,660 T10 2,148 1,727 1,890
T1 1,587 1,923 1,285 1,342 2,203 1,174 T1 1,129 1,413 1,368
mean dev: 0,0% -0,919 0,2% -0,890 0,3% -0,861 T -9,182
Criterions Tvar 0,005
1a) T10<2,41 yes yes yes yes yes yes Criterions
1b) T1<2,41 yes yes yes yes yes yes 1a) T10<2,41 yes yes yes
2) F<20 N yes yes yes 1b) T1<2,41 yes yes yes
3) t<0,1 ms yes yes yes 2)T<2,262 yes
4) |F|<1% yes yes yes 3,4) <0,02 yes yes
5) |∆t|<1,5ms yes yes yes 5)Tvar<=0,021 yes
mean pulse force peak: 7.082,87 all criterions are meet: yes
standard deviation pulse force F.. 9,81
6) F..<10 N yes
all criterions are meet: yes
All decisions are "yes". Therefor the device is calibrated correctly and ready to use. The calibration has been repeated after one year.