Sie sind auf Seite 1von 2

INSTRUCTION MANUAL 1.

Features
● Designed to meet international safety standards.
● Marks listed in the table below are used on this
instrument.
# CAUTION
● Set the Function Switch to an appropriate position
 IEC61010-1.IEC61010-031 & IEC61010-2-032
DIGITAL CLAMP METER  Measurement Category (CAT.) IV 600V # User must refer to the manual. before starting measurement.
● Firmly insert the test leads.
 Pollution Degree 2
 Instrument with double or reinforced insulation ● Disconnect the test leads from the instrument for

KEW SNAP SERIES


● Double molded main body provides comfortable
Indicates that this instrument can clamp on current measurement.
single handed grip ● Do not expose the instrument to the direct sun,
● Data Hold Function
bare conductors when measuring a voltage
corresponding to the applicable measurement high temperature and humidity or dewfall.
KEW2046R 600A TRMS Type ● LCD Backlight function to facilitate working at dimly
category, which is marked next to this symbol. ● Altitude 2000m or less. Appropriate operating
lit situations.
KEW2056R 1000A TRMS Type ● REL function to indicate measurement variation  AC temperature is within 0℃ and 40℃.
● This instrument isn't dust & water proofed. Keep
(Current, voltage, Resistance measurement)  DC away from dust and water.
● MIN/MAX function enables easy reading of min & AC & DC ● Be sure to power off the instrument after use.
max value during measurement. When the instrument will not be in use for a long
● PEAK Hold Function enables Peak value measure-
This instrument satisfies the marking
requirement defined in the WEEE Directive. period, place it in storage after removing the
ment of starting current. (only at ACA Range) This symbol indicates separate collection for batteries.
● With Continuity & Diode Check Function electrical and electronic equipment. ● Use a cloth dipped in water or neutral detergent
● Capacity measurement of capacitors
for cleaning the instrument. Do not use abrasives
● Temperature measurement, switchable between ℃ # DANGER or solvents.
and ℉ ● Never make measurement on a circuit in which
● NCV (Non Contact Voltage) Function for wiring check voltage over AC600V exists. Measurement categories (Over-voltage categories)
● 600V input protection ● Do not attempt to make measurement in the
To ensure safe operation of measuring instruments,
● Sleep Function to extend battery life presence of flammable gasses. Otherwise, the IEC61010 establishes safety standards for various
● With Bar Graph, 6039 count display use of the instrument may cause sparking, which electrical environments, categorized as CAT.Ⅰ to
can lead to an explosion. CAT.Ⅳ, and called measurement categories.
2. Safety Warnings ● Transformer jaw tips are designed not to short the
Higher-numbered categories correspond to electrical
This instrument has been designed,manufactured and circuit under test. If equipment under test has environments with greater momentary energy, so a
tested according to IEC 61010: Safety requirements e x p o s e d c o n d u c t i v e p a r t s , h o w e v e r, e x t r a m e a s u r i n g i n s t r u m e n t d e s i g n e d f o r C AT. Ⅲ
for Electronic Measuring apparatus, and delivered in precaution should be taken to minimize the environments can endure greater momentary energy
the best condition after passed the inspection. possibility of shorting.
This instruction manual contains warnings and safety than one designed for CAT.Ⅱ.
● Never attempt to use the instrument if its surface
rules which must be observed by the user to ensure CAT.Ⅰ: Secondary electrical circuits connected to an
or your hand is wet. AC electrical outlet through a transformer or
safe operation of the instrument and retain it in safe ● Do not exceed the maximum allowable input of
condition. similar device.
any measuring range. CAT.Ⅱ: Primary electrical circuits of equipment
Therefore, read through these operating instructions ● Never open the Battery cover during a
before using the instrument. connected to an AC electrical outlet by a
measurement. power cord.
● The instrument is to be used only in its intended
# WARNING CAT.Ⅲ: Primary electrical circuits of the equipment
● Read
applications or conditions. Otherwise, safety connected directly to the distribution panel, and
through and understand the instructions
c o n t a i n e d i n t h i s m a n u a l b e f o re u s i n g t h e functions equipped with the instrument doesn't feeders from the distribution panel to outlets.
instrument. work, and instrument damage or serious personal CAT.Ⅳ: The circuit from the service drop to the
● Keep the manual at hand to enable quick reference injury may be caused. service entrance, and to the power meter
whenever necessary. and primary over current protection device
● The instrument is to be used only in its intended # WARNING
applications. (distribution panel).
● Never attempt to make measurement if any
● Understand and follow all the safety instructions
abnormal conditions, such as broken case and
contained in the manual. exposed metal parts are found on the instrument.
● It is essential that the above instructions are
● Do not rotate the Function Switch while the test
adhered to. Incoming wire
● Failure to follow the above instructions may cause leads are being connected. Interior wiring
injury, instrument damage and/or damage to ● Do not install substitute par ts or make any
equipment under test. modification to the instrument. For repair or re- CAT.IV CAT.III
calibration, return the instrument to your local
The symbol # indicated on the instrument means distributor from where it was purchased. CAT.II
Transformer
that the user must refer to the related parts in the ● Do not try to replace the batteries if the surface of
CAT.I
manual for safe operation of the instrument. It is the instrument is wet. Socket
essential to read the instructions wherever the # ● Disconnect all the cords and cables from the
symbol appears in the manual. object under test and power off the instrument
before opening the Battery Cover for Battery
# DANGER is reserved for conditions and actions replacement.
that are likely to cause serious or fatal injury. ● Verify proper operation on a known source before
# WARNING is reserved for conditions and actions use or taking action as a result of the indication of
that can cause serious or fatal injury. the instrument.
# CAUTION is reserved for conditions and actions
that can cause injury or instrument damage.

3. Specification Frequency/DUTY Function(Auto-ranging for Frequency) ● Applicable Standards ● CF (Crest Factor) is found by dividing the peak
3-1. Measuring range & accuracy Measuring Accuracy IEC 61010-1 : 2001 value by the effective value.
(accuracy guaranteed at 23℃ ±5℃, humidity Range Measurement CAT. IV 600V Pollution degree 2 Examples: Sine wave: CF=1.414
Range KEW2046R KEW2056R
45~85%) AC Current 600A, 1000A Function IEC 61010-031:2002, IEC 61010-2-032 Square wave with a 1: 9 duty ratio: CF=3
ACA 40Hz-400Hz
Measuring Accuracy ±0.5%rdg±5dgt EMC:EN 61326 Effective value Average value Conversion Reading errors for Crest factor
Function ACV 1Hz〜10kHz Waveform factor
Range KEW2046R KEW2056R ・ EN 55022 Vrms Vavg Vrms/ Vavg
average sensing
instrument CF

±2.0%rdg±5dgt(50/60Hz)
0.1-99.9%
±2.5%rdg±5dgt ・ EN 61000-4-2 (performance criterion B) A 1
A
2
π A
π
2
0-600.0A (Pulse width/Pulse period) 0 2 2 2 0%
Peak 1500A ±3.5%rdg±5dgt(40~500Hz) ・ EN 61000-4-3 (performance criterion B) ≒0.707 ≒0.637 ≒1.111 ≒1.414
600A CF=2.5@600A ±5.5%rdg±5dgt(500~1kHz) Note: M e a s u r a b l e i n p u t s a re : 40V r m s @ A C V o r ● Overload Protection A A×1,111-A
CF=3.0@500A 0 ×100
A A 1 A 1
* Add 2% at CF>2 50Arms@AC600A, 350A@AC1000A Range Current Range : 720A AC/ 10 sec@KEW2046R =11.1%
0-1000A 1200A AC/DC/ 10 sec@KEW2056R A 2 0.5A×1.111-
A

Peak 1500A Temperature Function Voltage Range : 720V AC/DC/ 10sec 0


1
A 0.5A 3 A
3 ×100
3
1000A CF=2.5@600A
3 3 ≒1.732
≒1.155
CF=3.0@500A Measuring Accuracy Resistance Range : 600V AC/DC/ 10sec =-3.8%
Range T
Range KEW2046R KEW2056R ● Withstand Voltage A f 1 (1.111  
D -1) A 1
0 A D A T =A・D A D = =
-50℃ ~ 0℃ ±5℃±3dgt 6880V AC(TRMS 50/60Hz)/ 5 sec f AD D ×100% A D D
DC Current 600A, 1000A Function D=f/T
℃ 0℃ ~ 150℃ ±3℃±2dgt
(between Jaws and electrical circuit/ between
Measuring Accuracy internal circuit and enclosure)
Function
Range KEW2046R KEW2056R 150℃ ~ 700℃ ±2℃±2dgt
● Insulation Resistance: 10MΩ or more/ 1000V 3-3. Function Keys
-58℉~ 32℉ ±9℉±3dgt The "●" mark shows available function at each
600A 0-600.0A ±1.5%rdg±5dgt (between electrical circuit and enclosure)
±1.5%rdg±5dgt ℉ 32℉~ 302℉ ±5℉±2dgt
Range.
1000A 0-1000A ● Conductor size
302℉~1292℉ ±2%±2dgt KEW2046R: approx. 33mm Hz/ MAX/
HOLD PEAK SELECT ZERO
DUTY MIN
AC Voltage Function Above specified accuracy is applied to Clamp meter KEW2056R: approx. 40mm
(Auto-ranging, Input impedance: approx. 10MΩ) ● Dimension ACA ● ● ● ● ● ●
itself. Accuracy of Temperature probe is excluded.
Measuring Accuracy approx. 254(L)×82(W)×36(D)mm / KEW2056R ACV ● - - ● ● ●
Range
Range KEW2046R KEW2056R 3-2. General Specification approx. 243(L)×77(W)×36(D)mm / KEW2046R
DCA ● - ● ● - ●
● Mode of operation : △Σ mode ● Weight : approx 300g @ KEW2046R
±1.5%rdg±4dgt (50/60Hz)
6/60/600V 0-600.0V 310g @ KEW2056R DCV ● - - ● - ●
±3.5%rdg±5dgt (40~400Hz) ● Display : max. 6039 counts
  (Frequency: 9999, Capacity &Temperature:4039) ● Accessories
Ω ● - ● ● - ●
DC Voltage Function & Bar graph Test Leads Model 7066 / 1 set
(Auto-ranging, Input impedance: approx. 10MΩ) Battery R03(UM-4)/ 2pcs - - ● - - -
● Over-range indication : "OL" displayed when exceeding
Measuring Accuracy the measuring range. Instruction manual English, Japanese / 1pce - - ● - - -
Range
Range KEW2046R KEW2056R   (except for AC/DCV and 1000A Function) Carrying Case Model 9094 / 1pcs
● Optional Accessories
● - ● ● - -
600mV/6/6 ● Range switching :
0-600.0V ±1.0%rdg±3dgt K-type Temperature Probe Model 8216 TEMP ● - ● ● - ●
0/600V Auto-ranging/Voltage, Resistance, Capacity Range
Single range / Continuity, Diode check, DUTY and Multi-Tran M-8008
Resistance (Diode Check/ Continuity/ Capacity) Temperature 4. Preparation for measurement
Measuring Accuracy ● Sample rate : three times per second ● Effective Value (RMS) 4-1. Checking Battery Voltage
Range Most # alternating currents and voltages are
Range KEW2046R KEW2056R ● Functional construction : Set the Function Switch to any position other than
600Ω/6k/ OFF/ ACA/ ACV/ DCA/ DCV/Ω/ ℃ / ℉ expressed in effective values, which are also "OFF". When the display is clear without "BATT" mark,
0-600.0V ±1.0%rdg±5dgt
● Keys referred to as RMS (Root-Mean-Square) values. showing battery voltage is enough. When the display is
60k/600kΩ :
The effective value is the square root of the blank or "BATT" mark is indicated, replace the batteries
6M/60MΩ 0.600-60.00MΩ ±5%rdg±8dgt SELECT(AC/DC switching &/Ω/ / / ), according to Section 7, Battery Replacement.
PEAK HOLD/ Back Light, REL△ ,Hz/DUTY, MIN/MAX average of square of alternating current or voltage
Cont Buzzer 0-600.0Ω Buzzer sounds at 100Ω or less ● Power source : DC3V/ R03(UM-4) x 2pcs values. Many clamp meters using a conventional # CAUTION
Diode Test voltage: 0-2V ● Low battery warning : " " mark is displayed at re c t i f y i n g c i rc u i t h a v e " R M S " s c a l e s f o r A C T h e S l e e p f e a t u re a u t o m a t i c a l l y p o w e r s t h e
2.4V±0.15V or less. measurement. The scales are, however, actually instrument off in about 15 min after the last switch or
Capacity Function ● Temperature & humidity : 23℃ ±5℃ , relative humidity calibrated in terms of the effective value of a sine key operation. Therefore, the display may be blank
wave though the clamp meter is responding to the even with the Function Switch set to a position other
Measuring Accuracy accuracy guaranteed 85% or less (no condensation)
Range average value. The calibration is done with a than "OFF". To operate the instrument in this case,
Range ● Operating temperature : 0~40℃ , relative humidity 85%
KEW2046R KEW2056R turn the switch back to the "OFF" position, then to
& humidity range or less (no condensation) conversion factor of 1.111 for sine wave, which is any other position. Replace the batteries if nothing
40nF the accuracy is not guaranteed
● Storage temperature : -20~60℃ , relative humidity found by dividing the effective value by the average was displayed after above operations.
400nF value. These instruments are therefore in error if the
0.01nF & humidity range 85% or less (no condensation)
4μF ±2.5%rdg±20dgt ● Current consumption : approx. 25 mA input voltage or current has some other shape than 4-2. Checking Switch Setting & Operation
-4000μF
40μF ● Sleep Function : Automatically powered off in about sine wave. Confirm the Function Switch is set to the correct
Auto-ranging p o s i t i o n , t h e i n s t r u m e n t i s s e t to t h e c o r re c t
400μF the accuracy is not guaranteed 15 min after the last Function switch operation. measurement mode,and the Data hold function is
4000μF the accuracy is not guaranteed Rotate the Function Switch from OFF to any disabled. Otherwise,desired measurement cannot be
position to exit from the Sleep state. made.
5. Measurement (3) Press the trigger to open the transformer jaws and 5-5. Resistance/ Diode/ Cont/ Capacity Measurement Capacity
5-1. AC Current Measurement clamp them onto the one conductor under test, the (1) Set the Function Switch to "Ω/Diode/Cont/Capacity"
conductor should be at the center of the jaws, then # DANGER position. "Ω" has been selected by default; press the
● Never use the instrument onan energized circuit. SELECT key to change it to "Capacity"
# DANGER take the reading on the display. ● Do not make measurement with the Battery Cover Resistance ⇨ Diode ⇨ Cont ⇨ Capacity
● Never make measurement on a circuit in which (4) S e t t h e F u n c t i o n S w i t c h t o a n a p p ro p r i a t e removed.
voltage over AC600V exists to avoid getting position according to current under test.
(5) Pressing the "ZERO" key again releases "ZERO" Resistance
electrical shock.
function. (△mark at the upper right on the display (1) Set the Function Switch to "Ω/Diode/Cont/Capacity"
● Transformer jaw tips are designed not to short the (2) Connect the red test lead to V/Ω terminal and the
disappears.) position.
circuit under test. If equipment under test has (2) Connect the red test lead to V/ Ω terminal and the black test lead to COM terminal.
exposed conductive parts, however, extra pre- # CAUTION black test lead to COM terminal. Confirm "OL" is (3) Connect the test leads to the both ends of the
caution should be taken to minimize the possibility ● When the current flows from the upside (the indicated on the display, and then short-circuit the capacitor under test.
of shorting. display side) to the underside of the instrument, tips of test leads to make the indication zero. (4) Take the reading on the display.
● Do not make measurement with the Battery Cover the polarity of the reading is positive and vice (3) Connect the test leads to the both ends of the
removed. versa. resistor under test. 5-6 Temperature Measurement
● Disconnect the test leads from the instrument for (4) Take the reading on the display. (1) Set the Function Switch to "℃/℉" position.
(2) Connect the K-type Temperature Probe (Optional
current measurement. 5-3. AC Voltage Measurement
# CAUTION Accessories) to the input terminal. Positive (+) side
# DANGER ●Even if short the test lead tips, indicated value of Probe should be connected to V/Ω.
(1) Set the Function Switch to "600A" or "1000A" ● Never make measurement on a circuit in which (3) C o n t a c t t h e S e n s o r ( m e t a l p a r t ) o f K - t y p e
may not be zero. But this is because of the
position. resistance of test leads and not a failure. Temperature Probe to the object under test. Take
voltage over AC600V exists to avoid getting
(on KEW2046R, only "600A" is available) AC has electrical shock. ● When test leads are open, "OL" is indicated on the the reading on the display. Positive (+) side of
been selected by default; press the SELECT key, ● Do not make measurement with the Battery Cover display. Probe should be connected to V/Ω.
when DC has been selected, to change it to AC. removed.
AC mark is displayed at the upper left on the display. Continuity # WARNING
● Keep your fingers behind the barrier on the
(2) Press the trigger to open the transformer jaws and (1) Set the Function Switch to "Ω/Diode/Cont/Capacity" ● Never connect the Temperature Probe to an
instrument during measurement. position. "Ω" has been selected by default; press the energized circuit.
clamp them onto the one conductor under test,
(1) Set the Function Switch to "ACV" position. SELECT key to change it to "Continuity"
then take the reading on the display. Pressing the
(2) Connect the red test lead to V/ Ω terminal and the Resistance ⇨ Diode ⇨ Cont ⇨ Capacity # CAUTION
"Hz/DUTY" Key switches the indication in following ● Room temperature is indicated on the LCD when
sequence. black test lead to COM terminal.
setting the Function Switch to "℃ /℉" position. In
AC Current ⇨  Hz ⇨  DUTY (3) Connect the test leads to the circuit under test. Take case that "OL" or anything other than room
the reading on the display. Pressing the "Hz/DUTY" (2) Connect the red test lead to V/ Ω terminal and the temperature is indicated, something may wrong
key while reading is indicated on the display black test lead to COM terminal. Confirm "OL" is with the instrument. Stop the use of instrument
Hz/DUTY Function requires 50A or more at AC600A switches the indication in following sequence. indicated on the display and short circuit the tips of immediately.
Range and 350A or more at AC1000A range. AC Voltage ⇨ Hz ⇨ DUTY test leads. Indication should become zero and ● There may be a break in Probe when indication
buzzer sounds. isn't changed if Sensor (metal part) of K-type
# CAUTION (3) Connect the test leads to the both ends of the Temperature Probe is contacted with the object
● Max conductor size for KEW2046R is approx dia. conductor under test. The buzzer sounds, if the under test.
# CAUTION resistance under test is 100Ω or less.
33mm and for KEW2056R is approx dia. 40mm. ● Hz/DUTY Function requires AC40V or higher.
During current measurement, keep the ● To measure a frequency, measure the voltage
6. Other functions
on Diode
transformer jaws fully closed. Other wise, accurate 6-1. Sleep Function
the electrical circuit in advance. Then press the (1) Set the Function Switch to "Ω/Diode/Cont/Capacity"
measurements cannot be taken. Hz/DUTY key to enter into frequency measurement. (1) This is a function to prevent the instrument
position. "Ω" has been selected by default; press the from being left powered on in order to conserve
● Readings of frequency may fluctuate or be SELECT key to change it to "Diode" battery life. This function causes the instrument
5-2. DC Current Measurement influenced under noisy environment. Resistance ⇨ Diode ⇨ Cont ⇨ Capacity to enter Sleep mode about 15 minutes after the
last key operation. To exit the Sleep mode, turn
# DANGER 5-4. DC Voltage Measurement the Function switch to "OFF", then to any other
● Never make measurement on a circuit in which # DANGER position.
(2) Connect the red test lead to V/Ω terminal and the (2) Sleep Function is disabled when;
voltage over DC600V exists to avoid getting ● Never make measurement on a circuit in which black test lead to COM terminal. MIN/MAX or PEAK Function is selected.
electrical shock. voltage over DC600V exists to avoid getting
● Do not make measurement with the Battery Cover
Continuous measurement is made with the Sleep
electrical shock. Function being disabled. To activate Sleep
removed. ● Do not make measurement with the Battery Cover
Function again, disable the MIN/MAX or PEAK
removed. Function.
(1) Set the Function Switch to "600A" or "1000A" ● Keep your fingers behind the barrier on the
position. AC has been selected by default; press instrument during measurement. (3) Connect the red and black test leads to the Anode
the SELECT key, when AC has been selected, to and Cathode of the diode under test respectively. # CAUTION
change it to DC. (only 600A is available on (1) Set the Function Switch to "DCV" position. Take the reading on the display. If the connection ● The instrument consumes small amount of battery
KEW2046R) DC mark is displayed at the upper left (2) Connect the red test lead to V/ Ω terminal and the is reversed, the display indicates "OL". power in the Sleep mode. Set the Function Switch
on the display. black test lead to COM terminal. to the OFF position after use.
(2) With the transformer jaws closed and without (3) Connect the red and black test leads to the # CAUTION
● Some of diodes cannot be tested. Indication on 6-2. HOLD Key
clamping them onto the conductor, press the positive (+) and negative (-) sides of the circuit
the display will be "OL". (1) Data Hold Function
"ZERO" key to zero adjust the display. (△ mark is under test respectively. Take the reading on the
(Zener diode, LED and so on) This is a function to freeze the measured value on
displayed at the upper right on the display.) display. If the connection is reversed, the display
indicates the "-" mark. the display. Press the "HOLD" key to freeze the

reading. Press the MIN/MAX Key to select MAX or MIN. The 8. Maintenance
The reading will be held regardless of subsequent max or min value within measuring range is being ● Cleaning
variation in input. "H" is indicated on the upper left held until this function is disabled. "MIN" or "MAX" is Use a cloth dipped in water or neutral detergent for
corner of the display while the instrument is in the indicated on the display while this function is being cleaning the instrument.
Data Hold mode. To exit Data Hold mode, press activated. To disable this function, press down the Do not use abrasives or solvents. Otherwise,
the "HOLD" key again. MIN/MAX Key at least 2 sec or change functions. instrument get damaged, deformed or discolored.

(2) AC/DC Voltage Range

  # CAUTION
Pressing the MIN/MAX Key without applying voltage
disables the Auto-ranging function and fixes the
Range to 6V. Connect the test leads to the circuit
under test and press the MIN/MAX Key after an (4) Press the PEAK Key at least 2 sec to reset the
# CAUTION appropriate range is selected by Auto-ranging function. indication or release PEAK Function.
● Held readings are released when Sleep Function Pressing the MIN/MAX Key enables min or max Buzzer sounds twice, and the Function is
is activated while the instrument is in the Data Hold value measurement. Press the MIN/MAX Key to released.
mode. select MAX or MIN. The max or min value within
measuring range is being held until this function is # CAUTION
(2) Backlight ON/OFF disabled. "MIN" or "MAX" is indicated on the display ● PEAK indication for Crest value is up to 1500A.
Pressing the HOLD key 2 sec or more lights up the while this function is being activated. To disable this Error indication is given when exceeding this
Backlight. Pressing the HOLD key 2 sec or more function, press down the MIN/MAX Key at least 2 range value.
again turns off the Backlight. sec or change functions. ● Sleep Function is disabled when PEAK Function is
selected. Care should be taken when performing DISTRIBUTOR
6-3. NCV Function continuous measurement.
Red LED on the upper area on the Panel lights up at 6-5. ZERO Function
All functions except for OFF when electric field 6-7. Over-flow indication
exceeding 100V is detected by the sensor installed # CAUTION When the input exceeds the measuring range at
in the Jaws. MIN/MAX, PEAK keys are disabled while ZERO each Function other than Voltage ,1000A and
It indicates a presence of voltage in an electrical Function is being activated. Temperature Range "OL" or "-OL" is indicated on the
c i rc u i t o r e q u i p m e n t w i t h o u t display.
touching them. Zero Adjustment Function at Current Range "△"
NCV Sensor can detect electrical mark is to be indicated at the upper right on the 7. Battery Replacement
f i e l d o n l y f ro m t h e d i re c t i o n display while ZERO function is being operated.
indicated in the right figure. Indication of relative value at current, voltage, # WARNING
Put the fixed element (left side) resistance: ● To avoid electrical hazard, set the Function Switch
closer to the conductor under test. Pressing the ZERO Key indicates REL (relative to "OFF" and remove the test leads from the
Detection against in-wall outlet is value) Press the ZERO Key to save the initial value instrument before trying to replace batteries.
impossible. at the start of measurement as a reference value.
Then the difference between the later measured # CAUTION
# DANGER values and the reference value is indicated on the ● Do not mix old and new batteries.
● The LED may not light up due to installation display. The Auto-ranging function is disabled, while ● Install batteries in
correct polarity as indicated in
condition of electrical circuit or equipment. Never this function is being activated, and the Range is the Battery Compartment.
touch the circuit under test to avoid possible fixed to the Range selected at the start of
danger even if the LED for NCV doesn't light up. measurement. Relative value is indicated within
● Check the functionality of LED on a well-known following ranges.
(Measuring range) = Kyoritsu reserves the rights to change
power supply prior to measurement. When the specifications or designs described in this manual
LED doesn't light up, do not make measurement. (Full-scale value at the fixed Range) - (Initial value) without notice and without obligations.
● NCV indication is affected by external voltage,
how to hold or place the instrument. To disable this function, press down the MIN/MAX
Key at least 2 sec or change functions.
6-4. MIN/MAX Function
6-6. PEAK Function (600A only on KEW2046R) Replace the batteries when a Low Battery Voltage
# CAUTION (1) Set the Function Switch to "AC Current" position warning "BATT" mark is indicated on the display.
● Held readings are released when Sleep Function and clamp onto a conductor under test. Note that when the battery is completely exhausted,
is activated while the instrument is in the Data (2) Pressing the PEAK Key indicates " P MAX" on the the display blanks without "BATT" mark shown.
Hold mode. display and initiates measurement. (1) Set the Function Switch to "OFF" position.
● SELECT, ZERO, Hz/DUTY keys are disabled (3) Readings indicates the PEAK of current crest (2) Unscrew and remove the Battery Compartment
while MIN/MAX Function is being activated. value. When measuring sine wave, reading is Cover on the bottom of the instrument.
about √2 times of RMS value. (3) Replace the batteries observing correct polarity.
(1) AC/DC Current Range (600A only on KEW2046R) Use new R03 (AAA) or LR03 / 1.5V batteries.
Pressing the MIN/MAX Key at 600A & 1000A (4) Install the Battery Compartment and tighten the
Function enables min or max value measurement. screws. 6-10 92-1803B