Sie sind auf Seite 1von 37

ULTRAWAVE

APPLICATION
BOOK

Application guidelines for


your UltraWAVE
Milestone Srl UltraWAVE – Application book

Milestone Srl
Via Fatebenefratelli, 1/5
24010 Sorisole (BG) Italy
Tel +39 035 573857 – Fax +39 035 575498
www.milestonesrl.com

THE ULTRAWAVE APPLICATION BOOK


1- Introduction and Applications for UltraWAVE ECR

MICROWAVE PROGRAM GUIDELINES


2- Sample weight
3- Acid mixture
4- Microwave program

GENERAL DIGESTION METHODS


5- General Digestion method for Food samples
6- General digestion method of Environmental samples
7- General digestion method for Polymers samples
8- General digestion method for Petrochemicals samples
9- General digestion method for Pharmaceutical samples
10- General digestion method for Biological samples

LIST OF TESTED SAMPLES IN MILESTONE ANALYTICAL LAB

APPLICATIONS FOR ALL ULTRAWAVE VERSIONS

UW-01 Mineral Oil


UW-02 Polyethylene loaded with Si (Total recover of Si)
UW-03 Terephtalic acid (PTA)
1
Milestone Srl UltraWAVE – Application book

UW-04 Toys samples (Paper and polymer)


UW-05 Boron Carbide
UW-06 Silicon Carbide
UW-07 Water based markers (Ink)
UW-08 EPA 3051 General digestion method for Soil
UW-09 EPA 3052 General digestion method of soil with Si
UW-10 SRM samples B5-B6 (Lake Sediment)
UW-11 Carbon Fibre
UW-12 Silica Sand
UW-13 Coal
UW-14 Textile sample (Cotton)
UW-15 SRM Sample NIST 1575 (Pine Needles)
UW-16 SRM Sample USGS G-2 (Granite)
UW-17 SRM Sample USGS BCR-2 (Basalt)
UW-18 SRM Sample NIST 2702 (Marine sediment)
UW-19 SRM sample NIST 2711 (Montana soil)
UW-20 SRM Sample NIST 2709a (San Joaquim soil)
UW-21 Fly ash
UW-22 Aluminium Titanate (Al2TiO5)
UW-23 High amount of Vegetable oil
UW-24 Sulphur
UW-25 Titanium Oxide (TiO2)

APPLICATIONS ONLY FOR ULTRAWAVE ECR VERSION

UW-ECR-01 Alumina
UW-ECR-02 Ferberite (Iron Wolframite mineral, FeWO4)
UW-ECR-03 Inorganic Fertilizers
UW-ECR-04 Manganese doped Zirconia
UW-ECR-05 Nickel-Chromium (NiCr)
UW-ECR-06 Calcium Tungstate Mineral, CaWO4 (Scheelite)
UW-ECR-07 Tantalite
UW-ECR-08 Iron Manganese Tungstate Mineral, FeMnO4 (Wolframite)
UW-ECR-09 Yttria Zirconia
UW-ECR-10 Zirconia with Rare Earth Oxide
UW-ECR-11 Zirconium Oxide or Zirconium Hydroxide
UW-ECR-12 Platinum and Rhodium Alloy
UW-ECR-13 Carbon Black with high Si content

2
Milestone Srl UltraaWAVE – Appliccation book

Milestone Srl
Via Fatebennefratelli, 1/5
24010 Sorissole (BG) Italy
Tel +3
39 035 5738557 – Fax +39 035 575498
www.mileestonesrl.com

THE ULTTRAWAVVE APPLICATIO


ON BOO
OK
Introducttion

The purpoose of the application


a book is too support you,
y in ordeer to find thhe proper method
m forr
your samples.
s

The lissted appliccation notees are a goood point too start wheen developping a new
w
methood for yourr samples.. These guuidelines are a the moost relevant methodss
develooped by Milestone Laaboratoriess, based onn real sam mples from Milestonee
worldwwide users.

Anywaay pleasee consider that some applications might neeed somee


modifiications, in order to fi t with yourr specific sample.

Applicatioons for UlttraWAVE ECR


E

The Appliccation book is divideed into twoo parts, whhere the seecond part is only deedicated too
UltraWAVEE systems with
w ECR (EEnhanced CChemical Resistance)
R ) kit.

The ECR vversion consists of pressure


p linnes made of a speciial alloy haaving an outstanding
o g
chemical rresistance to a broad range of ccorrosive media,
m suchh as:

• Applications that reqquire moree than 15 mL of hyddrochloric aacid, consiidering thee


complete rack (i.e. more
m than 1 mL in eaach 15-possitions rackk or 3 mL in each 5--
positions rack,
r etc…
…)
• Routine appplications with high volume of hydrochlorric acid.
• Applications that reqquire the usse of aqua regia (3:1 HCl:HNO3))

Alll UltraWAV
VE system
ms can be upgraded with the EECR kit

Alll applications in thee first partt of this ap


pplication book are of course
suuitable forr UltraWAVVE ECR as well.

3
Milestone Srl UltraaWAVE – Appliccation book

M
MICROW
WAVE PPROGRA
AM GUID
DELINESS
UltraWAVEE allows tto digest a wide rrange of samples such s as eenvironmenntal, food,,
pharmaceutical, pettrochemicaal, polymerr, metals, ggeologicals, etc.

This guiddelines, willl help you to find the suitable microwavee digestionn
f all your samples.
method for
These aree only guiddelines, in case you need
n furtheer informatiion, pleasee
contact your
y local M Milestone distributor
d

Sample wweight
Sample wweight is an important parameterr especiallyy for digesttion of Orgaanic samplles.
Here below
w are somee guidelinees based onn sample tyypes.

Samples rreactivity Samp


ples type 15 positions
p ra
ack 5 positions rack

Soils, aqueeous solution s,


Not reaactive
environmeental sampless, Up to 1 g Up to 3 g
sampples
fertilizerss, feed, etc..

Food (fattty samples),


Samplees with
pharma, ingredients, Up to 0,5 g Up to 1,,5 g
medium rreactivity
feed
f

Highly reeactive Polymer, petrochemicaal,


Up to 0,3 g Up to 0,,8 g
sampples oils, solvents
s

Acid mixttures
The choosse of the acid
a mixtuure is a cruucial part of sample preparatioon especiaally for thee
inorganic samples.
w some inddications about acids , based on sample types.
Here below

Sam
mples Type Reag
gents

Orgaanic samplee HNO3, H2O2

Organic sam
mple with Silicates
S HNO3, H2O2,
H HF

4
Milestone Srl UltraaWAVE – Appliccation book

HNO33, HCl.
Mettal samples (Aqua regiaa or reverse aquaregia.
a Inn some casess
HF is neecessary for a complete ddigestion)
HNO3, HF, H2SO4, etc...
Inorganic samplees (See the appplication bookk or contact the Milestonne
application
n specialist)

Saamples mu
ust be com
mpletely wet
w by the acid mixtuure.

Digestionn program
Most of thhe samples can be used simultaaneously wiith a standaard microw
wave prograam:
Steep 1 20 min 240°C
Steep 2 10 min 240°C

In some ccases, due to the reacctivity of thhe sampless or to the high tempperature reqquired, thee
microwavee program should bee slightly m modified. Below
B is reeported a ttable that gives
g clearr
indicationss of the appproach too be used for the diffferent sam mples. The table considers twoo
main paraameters:
- Reaactivity of the
t samplee
- Tarrget tempeerature

Samples rreactivity Samp ples type Targeet temperatu


ure Staandard ramping time
Soils, aqueeous solutionns, 10 min up too 180°C
Not reaactive
environmeental sampless, 180-220°C
1 or
sampples
fertilizerss, feed, etc.. 15 min up too 220°C
Food (fattty samples), 20 min up too 220°C
Samplees with
pharma ingredients, 200-240°C
2 or
medium rreactivity
feed
f 25 min up too 240°C
5 min to 100°C
Polymer, petrochemica
p al,
Highly rreactive 15 min to 180°C
1
large amoount of food,, 220-250°C
2
sampples 10 min to 220°C*
2
oils, solvents
*deppending on the
t target T
25 min up too 250°C
Geologiccal, metals,
High temperature 250-270°C
2 Or
ceramicss, inorganics
30 min up too 270°C

5
 

 
Application Note

General digestion method for food samples


Summary
With Oil or high fat samples, reduce the amount to
This method provides the acid digestion of the sample 1 g in the 5 positions rack and 0,3 g in the 15
in a Single Reactor Chamber (SRC) using temperature positions rack.
control and microwave heating for the metal
determination by spectroscopic methods. This method
is suitable for samples like: Infant formula, Milk
powder, Meat, vegetables, rice, etc…
Start Pressure
40 bar of Nitrogen
Instrumentation
Milestone UltraWAVE with internal temperature and
Procedure
pressure control in all vessels, 640 or 660 terminal 1. Switch ON the chiller
with easyCONTROL software installed. 2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
Nitrogen tank 3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
Basic Load 5. Tight the clamps
6. Run the microwave program.
Place the following basic load into the main 1 L 7. Wait the cooling and the pressure release
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67% 8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.
Sample amount and Reagent
Microwave program
Add 2 - 3 drops of pure water into the empty vials
before weight, this will help to completely wet the Step Time T1 T2 P Power
powder sample. 1 00:20:00 220°C 60°C 110 1500W
2 00:10:00 220°C 60°C 110 1500W

Notes
*Milestone suggest Glass or Quartz vials.

THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.
Rack type Sample Amount Reagents

5 positions* 2g 10 ml of HNO3
This procedure is only a guideline and it may need to be modified
15 positions* 0,5 g 5 ml of HNO3 or changed to obtain the required results on your sample.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

General digestion method for environmental samples


Summary Start Pressure
This method provides the acid digestion of the sample 40 bar of Nitrogen
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal Procedure
determination by spectroscopic methods. This method
1. Switch ON the chiller
is suitable for samples like: Soils, Plant materials,
2. Place the main TFM vessel with the basic load, into
Waste water, etc… the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
Instrumentation 4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
Milestone UltraWAVE with internal temperature and 6. Run the microwave program.
pressure control in all vessels, 640 or 660 terminal 7. Wait the cooling and the pressure release
with easyCONTROL software installed. 8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.
Nitrogen tank
Microwave program
Basic Load
Step Time T1 T2 P Power
Place the following basic load into the main 1 L
1 00:15:00 220°C 60°C 110 1500W
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%
2 00:10:00 220°C 60°C 110 1500W
Sample amount and Reagent Notes
Rack type Sample Amount Reagents *Milestone suggest Glass or Quartz vials.

5 positions* 2g 10 ml of HNO3 THE CLEANING /MAINTENANCE PROCEDURE


MUST BE EXECUTED FOLLOWING THE USER
15 positions* 0,5 g 5 ml of HNO3
MANUAL AND INSTALLER INSTRUCTIONS.

In case of “foaming” of samples: This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
- ‘Pre-wet’ these samples with 1ml H2O in
these vials, then add the acids for your
digestions

- Add acid drop-by-drop

- Wait for the exothermic reaction to subside


to down

- Use Vortex system, before the digestion


run.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


Application Note

General digestion method for polymers


Summary 5. Tight the clamps
6. Run the microwave program.
This method provides the acid digestion of the sample 7. Wait the cooling and the pressure release
in a Single Reactor Chamber (SRC) using temperature 8. Untight the clamps and open the main chamber.
control and microwave heating for the metal 9. Collect the solutions in a marked flask and dilute it.
determination by spectroscopic methods. This method
Microwave program
is suitable for samples like: Polyethylene, Polyester,
Polypropylene, etc... Step Time T1 T2 P Power
1 00:10:00 110°C 60°C 90 1500W
Instrumentation
2 00:15:00 180°C 60°C 90 1500W
Milestone UltraWAVE with internal temperature and 3 00:10:00 230°C 60°C 120 1500W
pressure control in all vessels, 640 or 660 terminal 4 00:05:00 230°C 60°C 120 1500W
with easyCONTROL software installed.

Nitrogen tank
Note
Basic Load
Milestone suggest to use glass or quartz vials
Place the following basic load into the main 1 L
In case of high NOx production, it’s suggested
TFM vessel: 150 ml of H2O, 5 ml of HNO3 65%
to use the High temperature PTFE disc (p/n
Sample amount and Reagent MCL0699A). See user manual for more details
or contact the Milestone specialist
Rack type Sample Amount Reagents
THE CLEANING /MAINTENANCE PROCEDURE MUST
8ml of HNO3, BE EXECUTED FOLLOWING THE USER MANUAL AND
5 positions Up to 600 mg*
1ml H2O de. INSTALLER INSTRUCTIONS.

4ml of HNO3,
15 positions Up to 250 mg*
1ml H2O de.
This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.

Pressure loading
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


Application Note

General digestion method for petrochemical products


Summary Pressure loading

This method provides the acid digestion of the sample 40 bar of Nitrogen
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal Procedure
determination by spectroscopic methods. This method 1. Switch ON the chiller
is suitable for samples like: Fuel oil, heavy oil, Crude 2. Place the main TFM vessel with the basic load, into
oil, etc… the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
Instrumentation 4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
Milestone UltraWAVE with internal temperature and 6. Run the microwave program.
pressure control in all vessels, 640 or 660 terminal 7. Wait the cooling and the pressure release
with easyCONTROL software installed. 8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.
Nitrogen tank
Microwave program
Basic Load
Step Time T1 T2 P Power
Place the following basic load into the main 1 L 1 00:10:00 110°C 60°C 90 1500W
TFM vessel: 150 ml of H2O, 5 ml of HNO3 65% 2 00:15:00 180°C 60°C 90 1500W
3 00:10:00 230°C 60°C 120 1500W
Sample amount and Reagent
4 00:05:00 230°C 60°C 120 1500W
Rack type Sample Amount Reagents

8ml of HNO3,
5 positions Up to 600 mg*
1ml H2O de. Note
5ml of HNO3, Milestone suggest to use glass or quartz vials
15 positions Up to 300 mg*
1ml H2O de.
In case of high NOx generation, it’s suggested
to use the High temperature PTFA disc (p/n
*For solvent samples (i.e. Naphtha) reduce MCL0699A). See user manual for more details
the sample amount to 300 mg in 5 positions or contact the Milestone specialist
rack and 150 mg in 15 positions rack
THE CLEANING /MAINTENANCE PROCEDURE MUST
MILESTONE SUGGEST TO USE GLASS OR BE EXECUTED FOLLOWING THE USER MANUAL AND
QUARTZ VIALS. INSTALLER INSTRUCTIONS.

This procedure is only a guideline and it may need to be modified


or changed to obtain the required results on your sample.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

General digestion method for pharmaceutical samples


Summary Pressure loading

This method provides the acid digestion of the sample 40 bar of Nitrogen
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal
Procedure
determination by spectroscopic methods. 1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
Instrumentation the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
Milestone UltraWAVE with internal temperature and 4. Close the UltraWAVE chamber trough the software
pressure control in all vessels, 640 or 660 terminal 5. Tight the clamps
with easyCONTROL software installed. 6. Run the microwave program.
7. Wait the cooling and the pressure release
Nitrogen tank 8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.
Basic Load
Microwave program
Place the following basic load into the main 1 L
TFM vessel: 130 ml of H2O, 5 ml of HNO3 65% Step Time T1 T2 P Power
1 00:20:00 240°C 60°C 110 1500W
Sample amount and Reagent
2 00:10:00 240°C 60°C 110 1500W
Rack type Sample Amount Reagents

5 positions Up to 800 mg* 8ml of HNO3


Note
15 positions Up to 500 mg* 5ml of HNO3
Milestone suggest to use glass or quartz vials

THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.

This procedure is only a guideline and it may need to be modified


or changed to obtain the required results on your sample.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

General method for biological samples


Summary 7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.
This method provides the acid digestion of the sample 9. Collect the solutions in a marked flask and dilute it.
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal
Microwave program
determination by spectroscopic methods. This method
Step Time T1 T2 P Power
is suitable for samples like: Blood, human or animal
1 00:15:00 220°C 60°C 110 1500W
tissue, hairs, etc…
2 00:10:00 220°C 60°C 110 1500W
Instrumentation
Milestone UltraWAVE with internal temperature and Note
pressure control in all vessels, 640 or 660 terminal
with easyCONTROL software installed. *Milestone suggest Glass or quartz vials.

Nitrogen tank THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Basic Load INSTALLER INSTRUCTIONS.

Place the following basic load into the main 1 L


TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%. This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
Sample amount and Reagent
Rack type Sample Amount Reagents

15 positions* 0,5 g 5 ml of HNO3

22 positions* 0,2 g 3 ml of HNO3

Sample amounts refers to dry matter

Pressure loading
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Mineral  Oil
 
Summary 6. Run the microwave program.
7. Wait the cooling and the pressure release
This method provides the acid digestion of the sample 8. Untight the clamps and open the main chamber.
in a Single Reactor Chamber (SRC) using temperature 9. Collect the solutions in a marked flask and dilute it.
control and microwave heating for the metal
determination by spectroscopic methods.
Microwave program

Step Time T1 T2 P Power


Instrumentation
1 00:10:00 120°C 60°C 110 1500W
Milestone UltraWAVE with internal temperature and 2 00:10:00 160°C 60°C 130 1500W
pressure control in all vessels, 640 or 660 terminal
3 00:20:00 220°C 60°C 130 1500W
with easyCONTROL software installed.
4 00:10:00 220°C 60°C 130 1500W
Nitrogen tank

Basic Load *Milestone suggest Quartz or Glass vials

Place the following basic load into the main 1 L THE CLEANING /MAINTENANCE PROCEDURE MUST
TFM vessel: 150 ml of H2O, 5 ml of HNO3 65%. BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.
Sample amount and Reagent
This procedure is only a guideline and it may need to be modified
Rack type Sample Amount Reagents
or changed to obtain the required results on your sample.

4 ml of HNO3,
15 positions* 0,2 g
2ml H2O

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
 
 
 

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Polyethylene loaded with Si (Total Si recovery)


Summary 7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.
This method provides the acid digestion of the sample 9. Collect the solutions in a marked flask and dilute it.
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal
Microwave program
determination by spectroscopic methods. Step Time T1 T2 P Power
1 00:25:00 220°C 60°C 110 bar 1500W
Instrumentation
2 00:10:00 220°C 60°C 110 bar 1500W
Milestone UltraWAVE with internal temperature and
pressure control in all vessels, 640 or 660 terminal
with easyCONTROL software installed. Note
Nitrogen tank *Milestone suggest PTFE vials.

Basic Load Use the PTFE protection for thermowell (P/N


MCL0664A)
Place the following basic load into the main 1 L
TFM vessel: 150 ml of H2O, 5 ml of HNO3 65% THE CLEANING /MAINTENANCE PROCEDURE MUST
BE EXECUTED FOLLOWING THE USER MANUAL AND
Sample amount and Reagent INSTALLER INSTRUCTIONS.

Rack type Sample Amount Reagents


This procedure is only a guideline and it may need to be modified
4 ml of HNO3 + 4 or changed to obtain the required results on your sample.
15 positions* 100 mg
ml of HF

Pressure loading
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

Terephtalic acid (PTA)


Summary Procedure

This method provides the acid digestion of the sample 1. Switch ON the chiller
in a Single Reactor Chamber (SRC) using temperature 2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
control and microwave heating for the metal
3. Position the rack into the UltraWAVE
determination by spectroscopic methods. 4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
Instrumentation 6. Run the microwave program.
7. Wait the cooling and the pressure release
Milestone UltraWAVE with internal temperature and 8. Untight the clamps and open the main chamber.
pressure control in all vessels, 640 or 660 terminal 9. Collect the solutions in a marked flask and dilute it.
with easyCONTROL software installed.
Microwave program
Nitrogen tank
Step Time T1 T2 P Power
Basic Load
1 00:10:00 150°C 60°C 90 1500W
Place the following basic load into the main 1 L 2 00:05:00 150°C 60°C 90 1500W
TFM vessel: 130 ml of H2O, 5 ml of HNO3 3 00:10:00 250°C 60°C 120 1500W
4 00:10:00 250°C 60°C 120 1500W
Sample amount and Reagent

Rack type Sample Amount Reagents


Note
15 positions* 500 mg 8 ml of HNO3
*Milestone suggest glass or quartz vials.

THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Pressure loading INSTALLER INSTRUCTIONS.

40 bar of Nitrogen
This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Toys samples (Paper and polymer)


Summary 7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.
This method provides the acid digestion of the sample 9. Collect the solutions in a marked flask and dilute it.
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal
Microwave program
determination by spectroscopic methods. Step Time T1 T2 P Power
1 00:10:00 110°C 60°C 90 1500W
Instrumentation
2 00:10:00 180°C 60°C 90 1500W
Milestone UltraWAVE with internal temperature and 3 00:10:00 230°C 60°C 120 1500W
pressure control in all vessels, 640 or 660 terminal 4 00:05:00 230°C 60°C 120 1500W
with easyCONTROL software installed.

Nitrogen tank
Note
Basic Load
Milestone suggest to use glass or quartz
Place the following basic load into the main 1 L vials. Use PTFE vials with HF.
TFM vessel: 150 ml of H2O, 5 ml of HNO3 65%
In case of high NOx generation, it’s suggested
Sample amount and Reagent to use the High temperature PTFE disc (p/n
MCL0699A). See user manual for more details
Rack type Sample Amount Reagents or contact the Milestone specialist
8ml of HNO3, THE CLEANING /MAINTENANCE PROCEDURE MUST
5 positions Up to 500 mg*
(1mL of HF) BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.
5ml of HNO3,
15 positions Up to 300 mg*
(1mL of HF)

In case of Si presence, add 1 mL of HF


This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
Pressure loading
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Boron Carbide
 
 
Summary  
6. Run the microwave program.
This method provides the acid digestion of the sample 7. Wait the cooling and the pressure release
in a Single Reactor Chamber (SRC) using temperature 8. Untight the clamps and open the main chamber.
control and microwave heating for the metal 9. Collect the solutions in a marked flask and dilute it.
determination by spectroscopic methods.
Microwave program
Instrumentation
Step Time T1 T2 P Power
Milestone UltraWAVE with internal temperature and 1 00:30:00 280°C 70°C 110 1500W
pressure control in all vessels, 640 or 660 terminal 2 00:20:00 280°C 70°C 110 1500W
with easyCONTROL software installed.

Nitrogen tank
*Milestone suggest PTFE vials

Basic Load Use the PTFE (MCL0699A) discs for high


temperature applications
Place the following basic load into the main 1 L
TFM vessel: 130 ml of H2O, 5 ml of HNO3 65%. THE CLEANING /MAINTENANCE PROCEDURE MUST
BE EXECUTED FOLLOWING THE USER MANUAL AND
Sample amount and Reagent INSTALLER INSTRUCTIONS.

Rack type Sample Amount Reagents


This procedure is only a guideline and it may need to be modified
3 ml of HNO3, or changed to obtain the required results on your sample.
5 or 8 positions* 50 mg 1 ml HF,
1 ml H2SO4

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
 
 

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Silicon Carbide
 
 
Summary  
6. Run the microwave program.
This method provides the acid digestion of the sample 7. Wait the cooling and the pressure release
in a Single Reactor Chamber (SRC) using temperature 8. Untight the clamps and open the main chamber.
control and microwave heating for the metal 9. Collect the solutions in a marked flask and dilute it.
determination by spectroscopic methods.
Microwave program
Instrumentation
Step Time T1 T2 P Power
Milestone UltraWAVE with internal temperature and 1 00:30:00 280°C 70°C 110 1500W
pressure control in all vessels, 640 or 660 terminal 2 00:15:00 280°C 70°C 110 1500W
with easyCONTROL software installed.

Nitrogen tank
*Milestone suggest PTFE vials

Basic Load Use the PTFE (MCL0699A) discs for high


temperature applications
Place the following basic load into the main 1 L
TFM vessel: 130 ml of H2O, 5 ml of HNO3 65%. Use the PTFE protection for thermowell (P/N
MCL0664A)
Sample amount and Reagent
THE CLEANING /MAINTENANCE PROCEDURE MUST
BE EXECUTED FOLLOWING THE USER MANUAL AND
Rack type Sample Amount Reagents INSTALLER INSTRUCTIONS.

3 ml of HNO3,
5 positions* 50 mg 3 ml HF,
This procedure is only a guideline and it may need to be modified
6 ml H2SO4
or changed to obtain the required results on your sample.

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
 
 

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Water based marker (Ink)


Summary Microwave program

This method provides the acid digestion of the sample Step Time T1 T2 P Power
in a Single Reactor Chamber (SRC) using temperature 1 00:20:00 250°C 60°C 110 1500W
control and microwave heating for the metal 2 00:10:00 250°C 60°C 110 1500W
determination by spectroscopic methods.

Instrumentation
Note
Milestone UltraWAVE with internal temperature and
*Milestone suggest glass or quartz vials.
pressure control in all vessels, 640 or 660 terminal
with easyCONTROL software installed. THE CLEANING /MAINTENANCE PROCEDURE MUST
BE EXECUTED FOLLOWING THE USER MANUAL AND
Nitrogen tank INSTALLER INSTRUCTIONS.

Basic Load
This procedure is only a guideline and it may need to be modified
Place the following basic load into the main 1 L or changed to obtain the required results on your sample.
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%

Sample amount and Reagent


Rack type Sample Amount Reagents

4 ml of HNO3,
15 positions* 0,25 g
1 ml of H2O

Pressure loading
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

EPA 3051 General digestion method of soil samples


Summary Microwave program

This method provides the acid digestion of the sample Step Time T1 T2 P Power
in a Single Reactor Chamber (SRC) using temperature 1 00:10:00 175°C 60°C 120 1500W
control and microwave heating for the metal 2 00:10:00 175°C 60°C 120 1500W
determination by spectroscopic methods.

Instrumentation
Note
Milestone UltraWAVE with internal temperature and
*Milestone suggest Glass or quartz vials.
pressure control in all vessels, 640 or 660 terminal
with easyCONTROL software installed. THE CLEANING /MAINTENANCE PROCEDURE MUST
BE EXECUTED FOLLOWING THE USER MANUAL AND
Nitrogen tank INSTALLER INSTRUCTIONS.

Basic Load
Place the following basic load into the main 1 L This procedure is only a guideline and it may need to be modified
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67% or changed to obtain the required results on your sample.

Sample amount and Reagent


Rack type Sample Amount Reagents

15 positions* 0,5 g 10 mL of HNO3*

Pressure loading
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

EPA 3052 Siliceous based compound


Summary Microwave program
This method provides the acid digestion of the sample Step Time T1 T2 P Power
in a Single Reactor Chamber (SRC) using temperature 1 00:10:00 180°C 70°C 120 1200W
control and microwave heating for the metal
2 00:10:00 180°C 70°C 120 1200W
determination by spectroscopic methods.

Instrumentation
Note
Milestone UltraWAVE with internal temperature and
*Milestone suggest TFM vials for application with
pressure control in all vessels, 640 or 660 terminal
HF.
with easyCONTROL software installed.
THE CLEANING /MAINTENANCE PROCEDURE MUST
Nitrogen tank BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.
Basic Load
Place the following basic load into the main 1 L
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67% This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
Sample amount and Reagent
Rack type Sample Amount Reagents

4 ml of HNO3, 2
15 positions* 0,5 g
ml of HF

Pressure loading
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release
8. Un-tight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

SRM samples B5-B6 (Lake Sediment)


Summary 8. Un-tight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.
This method provides the acid digestion of the sample
in a Single Reactor Chamber (SRC) using temperature Microwave program
control and microwave heating for the metal
Step Time T1 T2 P Power
determination by spectroscopic methods.
1 00:25:00 260°C 60°C 110 1200W
Instrumentation 2 00:15:00 260°C 60°C 110 1200W

Milestone UltraWAVE with internal temperature and


pressure control in all vessels, 640 or 660 terminal
Note
with easyCONTROL software installed.
*Milestone suggest TFM vials for application with
Nitrogen tank HF.

Basic Load Use the PTFE (MCL0699A) discs for high


temperature applications
Place the following basic load into the main 1 L
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67% THE CLEANING /MAINTENANCE PROCEDURE MUST
BE EXECUTED FOLLOWING THE USER MANUAL AND
Sample amount and Reagent INSTALLER INSTRUCTIONS.

Rack type Sample Amount Reagents

4 ml of HNO3, This procedure is only a guideline and it may need to be modified


2 ml of H3PO4, or changed to obtain the required results on your sample.
15 positions* 50 mg
2 ml of HBF4,
1 ml of H2SO4

Pressure loading
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Carbon Fibre
Summary Microwave program

This method provides the acid digestion of the sample Step Time T1 T2 P Power
in a Single Reactor Chamber (SRC) using temperature 1 00:25:00 270°C 60°C 110 1500W
control and microwave heating for the metal 2 00:10:00 270°C 60°C 110 1500W
determination by spectroscopic methods.

Instrumentation *Milestone suggest Quartz or glass vials

Milestone UltraWAVE with internal temperature and Use the PTFE (MCL0699A) discs for high
pressure control in all vessels, 640 or 660 terminal temperature applications
with easyCONTROL software installed.
THE CLEANING /MAINTENANCE PROCEDURE MUST
Nitrogen tank BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.
Basic Load
Place the following basic load into the main 1 L This procedure is only a guideline and it may need to be modified
TFM vessel: 130 ml of H2O, 5 ml of HNO3 65%. or changed to obtain the required results on your sample.

Sample amount and Reagent

Rack type Sample Amount Reagents

15 positions* 100 mg 4 ml of HNO3

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps 
6. Run the microwave program.
7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Silica Sand
Summary
9. Collect the solutions in a marked flask and dilute it.
This method provides the acid digestion of the sample
in a Single Reactor Chamber (SRC) using temperature Microwave program
control and microwave heating for the metal
Step Time T1 T2 P Power
determination by spectroscopic methods.
1 00:15:00 240°C 60°C 110 1500W
Instrumentation 2 00:15:00 240°C 60°C 110 1500W

Milestone UltraWAVE with internal temperature and


pressure control in all vessels, 640 or 660 terminal *Milestone suggest PTFE vials
with easyCONTROL software installed.
Use the PTFE protection for thermowell (P/N
Nitrogen tank MCL0664A)

Basic Load THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Place the following basic load into the main 1 L INSTALLER INSTRUCTIONS.
TFM vessel: 130 ml of H2O, 5 ml of HNO3 65%.

Sample amount and Reagent This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.

Rack type Sample Amount Reagents

1 ml of HNO3,
15 positions* 100 mg
4 ml of HF

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps 
6. Run the microwave program.
7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

Coal
Summary Start Pressure
This method provides the acid digestion of the sample 40 bar of Nitrogen
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal Procedure
determination by spectroscopic methods.
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
Instrumentation
the UltraWAVE chamber.
Milestone UltraWAVE with internal temperature and 3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
pressure control in all vessels, 640 or 660 terminal 5. Tight the clamps
with easyCONTROL software installed. 6. Run the microwave program.
7. Wait the cooling and the pressure release
Nitrogen tank 8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.
Basic Load
Microwave program
Place the following basic load into the main 1 L
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67% Step Time T1 T2 P Power
1 00:20:00 240°C 60°C 110 1500W
Sample amount and Reagent
2 00:10:00 240°C 60°C 110 1500W
Add 2 - 3 drops of pure water into the empty vials Notes
before weight, this will help to completely wet the
powder sample. * Milestone suggest Glass or quartz vials. If HF is
required, use TFM vials.

In case of big amount of NOx, it’s suggested to


use the High temperature PTFE disc (p/n
MCL0699A). See user manual for more details or
contact the Milestone specialist

THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Rack type Sample Amount Reagents INSTALLER INSTRUCTIONS.

5 ml of HNO3
15 positions* 0,5 g
(1 ml of HF)
This procedure is only a guideline and it may need to be modified
In case of Si presence, add the HF. or changed to obtain the required results on your sample.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

Textile sample (Cotton)


Summary 9. Collect the solutions in a marked flask and dilute it.

This method provides the acid digestion of the sample Microwave program
in a Single Reactor Chamber (SRC) using temperature
Step Time T1 T2 P Power
control and microwave heating for the metal
1 00:15:00 220°C 60°C 110 1500W
determination by spectroscopic methods.
2 00:15:00 220°C 60°C 110 1500W
Instrumentation
Notes
Milestone UltraWAVE with internal temperature and * Milestone suggest Glass or quartz vials.
pressure control in all vessels, 640 or 660 terminal
with easyCONTROL software installed. THE CLEANING /MAINTENANCE PROCEDURE MUST
BE EXECUTED FOLLOWING THE USER MANUAL AND
Nitrogen tank INSTALLER INSTRUCTIONS.

Basic Load
This procedure is only a guideline and it may need to be modified
Place the following basic load into the main 1 L or changed to obtain the required results on your sample.
TFM vessel: 150 ml of H2O, 5 ml of HNO3 67%

Sample amount and Reagent


Rack type Sample Amount Reagents

15 positions* 0,5 g 5 ml of HNO3

5 positions* 1g 10 ml of HNO3

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

SRM Sample NIST 1575 (Pine Needles)


Summary 9. Collect the solutions in a marked flask and dilute it.

This method provides the acid digestion of the sample Microwave program
in a Single Reactor Chamber (SRC) using temperature
Step Time T1 T2 P Power
control and microwave heating for the metal
determination by spectroscopic methods. 1 00:15:00 230°C 60°C 110 1500W
2 00:10:00 230°C 60°C 110 1500W
Instrumentation Notes
Milestone UltraWAVE with internal temperature and * Milestone suggest Glass or quartz vials.
pressure control in all vessels, 640 or 660 terminal
with easyCONTROL software installed. THE CLEANING /MAINTENANCE PROCEDURE MUST
BE EXECUTED FOLLOWING THE USER MANUAL AND
Nitrogen tank INSTALLER INSTRUCTIONS.

Basic Load
This procedure is only a guideline and it may need to be modified
Place the following basic load into the main 1 L or changed to obtain the required results on your sample.
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%

Sample amount and Reagent


Rack type Sample Amount Reagents

15 positions* 0,5 g 4 ml of HNO3

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

SRM sample USGS G-2 (Granite)


Summary 8. Untight the clamps and open the main chamber
9. Collect the solutions in a marked flask and dilute it.
This method provides the acid digestion of the sample
in a Single Reactor Chamber (SRC) using temperature Microwave program
control and microwave heating for the metal
Step Time T1 T2 P Power
determination by spectroscopic methods.
1 00:20:00 260°C 60°C 110 1500W
Instrumentation 2 00:20:00 260°C 60°C 110 1500W

Milestone UltraWAVE with internal temperature and Notes


pressure control in all vessels, 640 or 660 terminal * Milestone suggest PTFE vials.
with easyCONTROL software installed.
Use the PTFE (MCL0699A) discs for high
Nitrogen tank temperature applications

Basic Load THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Place the following basic load into the main 1 L INSTALLER INSTRUCTIONS.
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%

Sample amount and Reagent This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
Rack type Sample Amount Reagents

2 ml of HNO3,
1 ml of HCl,
15 positions* 50 mg
2 ml of H3PO4,
2 ml of HBF4

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

SRM sample USGS BCR-2 (Basalt)


Summary 8. Untight the clamps and open the main chamber
9. Collect the solutions in a marked flask and dilute it.
This method provides the acid digestion of the sample
in a Single Reactor Chamber (SRC) using temperature Microwave program
control and microwave heating for the metal
Step Time T1 T2 P Power
determination by spectroscopic methods.
1 00:20:00 260°C 60°C 110 1500W
Instrumentation 2 00:20:00 260°C 60°C 110 1500W

Milestone UltraWAVE with internal temperature and Notes


pressure control in all vessels, 640 or 660 terminal * Milestone suggest PTFE vials.
with easyCONTROL software installed.
Use the PTFE (MCL0699A) discs for high
Nitrogen tank temperature applications

Basic Load THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Place the following basic load into the main 1 L INSTALLER INSTRUCTIONS.
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%

Sample amount and Reagent This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
Rack type Sample Amount Reagents

2 ml of HNO3,
1 ml of HCl,
15 positions* 50 mg
2 ml of H3PO4,
2 ml of HBF4

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

SRM sample NIST 2702 (Marine sediment)


Summary 8. Untight the clamps and open the main chamber
9. Collect the solutions in a marked flask and dilute it.
This method provides the acid digestion of the sample
in a Single Reactor Chamber (SRC) using temperature Microwave program
control and microwave heating for the metal
Step Time T1 T2 P Power
determination by spectroscopic methods.
1 00:20:00 260°C 60°C 110 1500W
Instrumentation 2 00:20:00 260°C 60°C 110 1500W

Milestone UltraWAVE with internal temperature and Notes


pressure control in all vessels, 640 or 660 terminal * Milestone suggest PTFE vials.
with easyCONTROL software installed.
Use the PTFE (MCL0699A) discs for high
Nitrogen tank temperature applications

Basic Load THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Place the following basic load into the main 1 L INSTALLER INSTRUCTIONS.
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%

Sample amount and Reagent This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
Rack type Sample Amount Reagents

3 ml of HNO3,
15 positions* 50 mg
2 ml of HBF4

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

SRM sample NIST 2711 (Montana soil)


Summary 8. Untight the clamps and open the main chamber
9. Collect the solutions in a marked flask and dilute it.
This method provides the acid digestion of the sample
in a Single Reactor Chamber (SRC) using temperature Microwave program
control and microwave heating for the metal
Step Time T1 T2 P Power
determination by spectroscopic methods.
1 00:20:00 260°C 60°C 110 1500W
Instrumentation 2 00:20:00 260°C 60°C 110 1500W

Milestone UltraWAVE with internal temperature and Notes


pressure control in all vessels, 640 or 660 terminal * Milestone suggest PTFE vials.
with easyCONTROL software installed.
Use the PTFE (MCL0699A) discs for high
Nitrogen tank temperature applications

Basic Load THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Place the following basic load into the main 1 L INSTALLER INSTRUCTIONS.
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%

Sample amount and Reagent This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
Rack type Sample Amount Reagents

3 ml of HNO3,
15 positions* 50 mg 2 ml of HBF4,
1 ml of HCl

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

SRM sample NIST 2709a (San Joaquin soil)


Summary 8. Untight the clamps and open the main chamber
9. Collect the solutions in a marked flask and dilute it.
This method provides the acid digestion of the sample
in a Single Reactor Chamber (SRC) using temperature Microwave program
control and microwave heating for the metal
Step Time T1 T2 P Power
determination by spectroscopic methods.
1 00:20:00 260°C 60°C 110 1500W
Instrumentation 2 00:20:00 260°C 60°C 110 1500W

Milestone UltraWAVE with internal temperature and Notes


pressure control in all vessels, 640 or 660 terminal * Milestone suggest PTFE vials.
with easyCONTROL software installed.
Use the PTFE (MCL0699A) discs for high
Nitrogen tank temperature applications

Basic Load THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
Place the following basic load into the main 1 L INSTALLER INSTRUCTIONS.
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67%

Sample amount and Reagent This procedure is only a guideline and it may need to be modified
or changed to obtain the required results on your sample.
Rack type Sample Amount Reagents

3 ml of HNO3,
15 positions* 50 mg 2 ml of HBF4,
1 ml of HCl

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

Fly Ash
Summary Start Pressure
This method provides the acid digestion of the sample 40 bar of Nitrogen
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal Procedure
determination by spectroscopic methods.
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
Instrumentation
the UltraWAVE chamber.
Milestone UltraWAVE with internal temperature and 3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
pressure control in all vessels, 640 or 660 terminal 5. Tight the clamps
with easyCONTROL software installed. 6. Run the microwave program.
7. Wait the cooling and the pressure release
Nitrogen tank 8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.
Basic Load
Microwave program
Place the following basic load into the main 1 L
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67% Step Time T1 T2 P Power
1 00:25:00 250°C 60°C 110 1500W
Sample amount and Reagent
2 00:10:00 250°C 60°C 110 1500W
Add 2 - 3 drops of pure water into the empty vials Notes
before weight, this will help to completely wet the
powder sample. * Milestone suggest Glass or quartz vials. If HF is
required, use TFM vials.

THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.

This procedure is only a guideline and it may need to be modified


or changed to obtain the required results on your sample.
Rack type Sample Amount Reagents

0,4g H3BO3
15 positions* 0,5 g powder, 4 ml
HNO3, 2ml HF

Add the reagents sequentially, in the order how they are


written.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

Aluminium Titanate
  (Al2TiO5)
 
Summary 6. Run the microwave program.
7. Wait the cooling and the pressure release
This method provides the acid digestion of the sample 8. Untight the clamps and open the main chamber.
in a Single Reactor Chamber (SRC) using temperature 9. Collect the solutions in a marked flask and dilute it.
control and microwave heating for the metal
determination by spectroscopic methods.
Microwave program

Step Time T1 T2 P Power


Instrumentation
1 00:30:00 280°C 60°C 110 1500W
Milestone UltraWAVE with internal temperature and 2 00:15:00 280°C 60°C 110 1500W
pressure control in all vessels, 640 or 660 terminal
with easyCONTROL software installed.
*Milestone suggest PTFE vials
Nitrogen tank
Use the PTFE (MCL0699A) discs for high
Basic Load temperature applications

Place the following basic load into the main 1 L Use the PTFE protection for thermowell (P/N
TFM vessel: 130 ml of H2O, 5 ml of HNO3 65%. MCL0664A)

THE CLEANING /MAINTENANCE PROCEDURE MUST


Sample amount and Reagent BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.
Rack type Sample Amount Reagents

5 ml HF, This procedure is only a guideline and it may need to be modified


5 positions* 50 mg
6 ml H3PO4
or changed to obtain the required results on your sample.

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
 
 
 

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note
 

High amount of Vegetable


  oil
 
Summary 6. Run the microwave program.
7. Wait the cooling and the pressure release
This method provides the acid digestion of the sample 8. Untight the clamps and open the main chamber.
in a Single Reactor Chamber (SRC) using temperature 9. Collect the solutions in a marked flask and dilute it.
control and microwave heating for the metal
determination by spectroscopic methods.
Microwave program

Step Time T1 T2 P Power


Instrumentation
1 00:10:00 120°C 60°C 110 1500W
Milestone UltraWAVE with internal temperature and 2 00:03:00 120°C 60°C 110 1500W
pressure control in all vessels, 640 or 660 terminal
3 00:08:00 160°C 60°C 150 1500W
with easyCONTROL software installed.
4 00:05:00 160°C 60°C 150 1500W
Nitrogen tank 5 00:10:00 250°C 60°C 150 1500W
6 00:10:00 250°C 60°C 150 1500W
Basic Load
Place the following basic load into the main 1 L
*Milestone suggest Quartz or Glass vials
TFM vessel: 170 ml of H2O, 5 ml of HNO3 65%.
THE CLEANING /MAINTENANCE PROCEDURE MUST
Sample amount and Reagent BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.
Rack type Sample Amount Reagents

6 ml of HNO3, This procedure is only a guideline and it may need to be modified


5 positions* 1g
3 ml H2O or changed to obtain the required results on your sample.

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
 
 
 

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

Sulphur
Summary Start Pressure
This method provides the acid digestion of the sample 40 bar of Nitrogen
in a Single Reactor Chamber (SRC) using temperature
control and microwave heating for the metal Procedure
determination by spectroscopic methods.
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
Instrumentation
the UltraWAVE chamber.
Milestone UltraWAVE with internal temperature and 3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
pressure control in all vessels, 640 or 660 terminal 5. Tight the clamps
with easyCONTROL software installed. 6. Run the microwave program.
7. Wait the cooling and the pressure release
Nitrogen tank 8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.
Basic Load
Microwave program
Place the following basic load into the main 1 L
TFM vessel: 150 ml of H2O, 5 ml of HNO3 67% Step Time T1 T2 P Power
1 00:20:00 250°C 60°C 130 1500W
Sample amount and Reagent
2 00:10:00 250°C 60°C 130 1500W
Add 2 - 3 drops of pure water into the empty vials Notes
before weight, this will help to completely wet the
powder sample. * Milestone suggest Glass or quartz vials.

THE CLEANING /MAINTENANCE PROCEDURE MUST


BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.

This procedure is only a guideline and it may need to be modified


or changed to obtain the required results on your sample.

Rack type Sample Amount Reagents

4 ml of HNO3
15 positions* 0,2 g
1 ml of H2O

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com


 

 
Application Note

Titanium Oxide (TiO2)


Summary Microwave program

This method provides the acid digestion of the sample Step Time T1 T2 P Power
in a Single Reactor Chamber (SRC) using temperature 1 00:25:00 260°C 60°C 130 1500W
control and microwave heating for the metal 2 00:15:00 260°C 60°C 130 1500W
determination by spectroscopic methods.
Notes
Instrumentation *Milestone suggest PTFE vials

Milestone UltraWAVE with internal temperature and Use the PTFE (MCL0699A) discs for high
pressure control in all vessels, 640 or 660 terminal temperature applications
with easyCONTROL software installed.
THE CLEANING /MAINTENANCE PROCEDURE MUST
Nitrogen tank BE EXECUTED FOLLOWING THE USER MANUAL AND
INSTALLER INSTRUCTIONS.
Basic Load
Place the following basic load into the main 1 L This procedure is only a guideline and it may need to be modified
TFM vessel: 130 ml of H2O, 5 ml of HNO3 67% or changed to obtain the required results on your sample.

Sample amount and Reagent


Rack type Sample Amount Reagents

4 ml of HNO3
15 positions* 0,1 g
1 ml of HF

Start Pressure
40 bar of Nitrogen

Procedure
1. Switch ON the chiller
2. Place the main TFM vessel with the basic load, into
the UltraWAVE chamber.
3. Position the rack into the UltraWAVE
4. Close the UltraWAVE chamber trough the software
5. Tight the clamps
6. Run the microwave program.
7. Wait the cooling and the pressure release
8. Untight the clamps and open the main chamber.
9. Collect the solutions in a marked flask and dilute it.

Milestone Srl – Helping Chemists www.milestonesrl.com application@milestonesrl.com

Das könnte Ihnen auch gefallen