Beruflich Dokumente
Kultur Dokumente
Copyright
HEYER Medical AG
VERSION: 1.1
RELEASE DATE: 01.2013
MODEL: Scalis 7, 12, 15
Declaration
HEYER Medical AG (hereinafter referred to as “HEYER”) owns the intellectual property rights
to this product and this manual. This manual may refer to information protected by copyrights
or patents and does not convey any license under the patent rights of HEYER or the rights of
others. HEYER does not assume any liability arising from any infringements of patents or
other rights of third parties.
HEYER intends to maintain the contents of this manual as confidential information. Disclosure
of the information in this manual in any manner whatsoever without the written permission of
HEYER is strictly forbidden. Release, amendment, reproduction, distribution, rent, adaption
and translation of this manual in any manner whatsoever without the written permission of
HEYER is strictly forbidden.
The contents of this manual are subject to change without prior notice.
Manufacturer’s Responsibility
All information contained in this manual is believed to be correct. HEYER shall not be liable for
errors contained herein or for incidental or consequential damages in connection with the
supply, performance or use of this manual.
HEYER is responsible for the safety, reliability and performance of this product only on the
condition that:
• all installation, additions, changes, modifications and repairs of this product are conducted
personnel authorized by HEYER; and,
• the electrical fittings of the relevant room comply with the applicable national and local
requirements; and,
• this product is operated under strict compliance with the instructions of this manual.
Warning
In order to use this device safely, you must follow the instructions listed. This manual is not
intended to replace proper medical instruction of the user.
• Do not rely on the audible alarm alone to monitor patient. During patient monitoring, selecting
a very low volume or completely muting the sound may have catastrophic results for the
patient. The most reliable way to monitor the patient is to perform manual monitoring in parallel
to the correct use of monitoring equipment.
• This multi-parameter patient monitor is intended for use only by medical professionals in
healthcare establishments.
• To avoid electrical shock, do not open any covers by yourself. Service must be carried out by
qualified personnel.
• Use of this device may cause interference on the screens of ultrasound equipment. This
monitor and any ultrasound imaging equipment must be kept as far apart as possible.
• It is dangerous to expose the electrical contacts and connectors to normal saline, other liquids
or conductive adhesives. Electrical contacts and connectors, such as cable connectors, power
supply and parameter module socket inlet and frame must be kept clean and dry. These parts
must be dried thoroughly if they come into contact with any liquid. If further eradication is
required, please contact your biomedical department or manufacturer.
Warning
If the hospital or establishment responsible for using this instrument does not establish an
appropriate maintenance schedule, this may result in abnormal equipment failure, thus
endangering human health.
For device components identified in this manual as appropriate for repair by the user, HEYER will
provide circuit diagrams, calibration methods and any other information requested by the user, in
order to facilitate repair by appropriately qualified technical personnel on behalf of the user.
Contents
Page
Chapter 1. Overview .............................................................................................. 1
1.1 Overview ........................................................................................................ 1
1.2 Safety information .......................................................................................... 1
1.3 Device Labels................................................................................................. 3
1.4 Intended Use.................................................................................................. 4
1.5 Introduction to screen layouts......................................................................... 5
1.5.1 Information section .......................................................................... 5
1.5.2 Parameters section.......................................................................... 6
1.5.3 Waveform section ............................................................................ 6
1.5.4 Menu section ................................................................................... 6
1.6 Alarms............................................................................................................ 7
1.7 Control panel.................................................................................................. 7
1.8 Menu .............................................................................................................. 7
1.8.1 Patient management........................................................................ 7
1.8.2 History review .................................................................................. 8
1.8.3 Settings............................................................................................ 8
1.8.4 Alarm limit...................................................................................... 11
1.8.5 Service .......................................................................................... 12
1.9 Sensor socket .............................................................................................. 14
1.10 External interface ......................................................................................... 15
1.11 Networks ...................................................................................................... 15
1.12 Built-in rechargeable battery......................................................................... 15
1.13 Installation.................................................................................................... 16
1.13.1 Open the box and check ................................................................ 16
1.13.2 Connect the power cables ............................................................. 16
1.14 Turning on the monitor ................................................................................. 17
1.15 Connect patient sensors............................................................................... 17
1.16 Check the recorder....................................................................................... 17
Chapter 2. Alarms ................................................................................................ 18
2.1 Alarms overview........................................................................................... 18
2.1.1 Types of alarm ............................................................................... 18
2.1.2 Alarm levels ................................................................................... 18
2.1.3 Alarm modes ................................................................................. 18
2.2 Pausing the alarms ...................................................................................... 19
2.3 Silencing the alarms ..................................................................................... 19
2.4 Alarm Setting................................................................................................ 19
2.5 Alarm limits .................................................................................................. 20
2.5.1 ECG Alarm Limit ............................................................................ 21
2.5.2 SpO2 Alarm Limit ........................................................................... 21
2.5.3 NIBP Alarm Limit ........................................................................... 22
2.5.4 Resp Alarm Limit ........................................................................... 22
2.5.5 Temp Alarm Limit........................................................................... 23
2.5.6 Load Default Alarm Limit................................................................ 23
2.6 Patient alarm information ............................................................................. 23
2.7 Technical alarm information ......................................................................... 25
Chapter 3. Recording........................................................................................... 28
3.1 Record settings ............................................................................................ 28
3.2 Record type.................................................................................................. 28
List of Figures
Fig. 1-1: demo of monitor interface................................................................... 5
Fig. 1-2: Patient management............................................................................ 7
Fig. 1-3: History review ...................................................................................... 8
Fig. 1-4: Setting .................................................................................................. 8
Fig. 1-5: Screen layouts ..................................................................................... 9
Fig. 1-6: Adjust time......................................................................................... 10
Fig. 1-7: Miscellaneous settings ..................................................................... 10
Fig. 1-8: Load default settings......................................................................... 11
Fig. 1-9: Alarm Limit......................................................................................... 12
Fig. 1-10: Service ............................................................................................... 12
Fig. 1-11: ECG calibration.................................................................................. 12
Fig. 1-12: Temp Sensor Type ............................................................................ 13
Fig. 1-13: Sensor socket .................................................................................... 14
Fig. 2-1: Alarm settings.................................................................................... 20
Fig. 2-2: Alarm Limit......................................................................................... 21
Fig. 2-3: ECG Alarm Limit ................................................................................ 21
Fig. 2-4: SpO2 Alarm Limit ............................................................................... 22
Fig. 2-5: NIBP Alarm Limit ............................................................................... 22
Fig. 2-6: Resp Alarm Limit ............................................................................... 22
Fig. 2-7: Temp Alarm Limit .............................................................................. 23
Fig. 2-8: Load Default Alarm Limit .................................................................. 23
List of Tables
Table 14-1: Guidance and HEYER declaration — electromagnetic
emissions........................................................................................... 65
Table 14-2: Guidance and HEYER declaration — electromagnetic
immunity ............................................................................................ 66
Table 14-3: Guidance and HEYER declaration — electromagnetic
immunity ............................................................................................ 67
Table 14-4: Recommended separation distances between portable and
mobile RF communication and the monitor .................................... 68
1.1 Overview
This monitor is suitable for adult, pediatric and neonatal patients and can be used to
monitor physical parameters such as electrocardiograph (ECG), noninvasive blood
pressure (NIBP), oxygen saturation (SpO2), respiration rate (Resp) and body
temperature (Temp). It can display a maximum of 8 waveforms and all information
about the parameters being monitored on the same screen.
WARNING Points to be noted to avoid injury to the patient and the operator.
Warnings
• Do not rely on the audible alarm alone to monitor patient. During patient monitoring,
selecting a very low volume or completely muting the sound may have catastrophic
results for the patient. The most reliable way to monitor the patient is to perform
manual monitoring in parallel to the correct use of monitoring equipment.
• This multi-parameter patient monitor is intended for use only by medical professionals
in healthcare establishments.
• To avoid electrical shock, do not open any covers by yourself. Service must be carried
out by qualified personnel.
• Use of this device may cause interference on the screens of ultrasound equipment.
This monitor and any ultrasound imaging equipment must be kept as far apart as
possible.
• It is dangerous to expose the electrical contacts and connectors to normal saline, other
liquids or conductive adhesives. Electrical contacts and connectors, such as cable
connectors, power supply and parameter module socket inlet and frame must be kept
clean and dry. These parts must be dried thoroughly if they come into contact with any
liquid. If further eradication is required, please contact your biomedical department or
manufacturer.
• There is the risk of electrical shock if the monitor casing is opened. All servicing and
future upgrading to this equipment must be carried out by personnel trained and
authorized by the manufacturer.
• Always ensure that the device and accessories can function safely and normally before
use.
• You must customize the alarm setups according to the individual patient situation and
make sure that the alarm sound can be activated when an alarm is triggered.
• Do not touch the patient, operating table, bed or the device during defibrillation.
• Do not use cellular phones in the vicinity of this device. High-level electromagnetic
radiation emitted from such devices may greatly affect the performance of the monitor.
• When used with electrosurgery equipment, the doctor or nurse must give top priority to
patient safety.
• Do not place the monitor or external power supply in any position that might cause it to
fall onto the patient. Do not lift the monitor by the power supply cord or patient cable;
use only the handle on the monitor.
• Dispose of the packaging material in accordance with the applicable waste control
regulations, and keep out of reach of children.
• Grounding:
Connect the monitor only to a three-wire, grounded, hospital-grade socket. The three-
pin plug must be inserted into a properly wired three-wire socket; if a three-wire socket
is not available, a qualified electrician must install one in accordance with the
applicable electrical code.
Do not under any circumstances remove the grounding conductor from the power plug.
Do not use extension cords or adapters of any type. The power cord and plug must be
intact and undamaged.
Caution
If you have any doubt about the earthing layout and its performance, you must use the
built-in battery to power the monitor.
Caution: alerts users that failure to follow the instructions may cause damage to
the equipment or affect the test results.
Warning: warns users of potential dangers. Failure to follow the instructions may
result in personal injury.
This symbol indicates that the instrument is IEC 60601-1 Type CF equipment.
The unit displaying this symbol contains an F-Type isolated (floating) patient
applied part providing a high degree of protection against shock and is suitable
for use during defibrillation.
Power On/Off
AC power
This symbol indicates that the instrument is IEC 60601-1 Type CF equipment.
The unit displaying this symbol contains an F-Type isolated (floating) patient
applied part providing a high degree of protection against shock and is suitable
for use during defibrillation.
This symbol indicates that the instrument is IEC 60601-1 Type BF equipment.
SCALIS is designed as a bedside portable monitor that can operate in all professional
medical facilities, including but not restricted to: emergency department, operating
room, post-anesthesia recovery, critical care, surgical intensive care, respiratory
intensive care, medical intensive care, pediatric intensive care and neonatal intensive
care. SCALIS cannot be used on helicopters. SCALIS is intended strictly to monitor a
single patient at any one time.
Warning
This is not a therapeutic device
1) Patient information:
Bed number (refers to the hospital bed number of patient monitored);
Type of patient (“Adult”, “Pediatric” or “Neonate”);
Name of patient (if operator does not input a patient name, this position will display
“NO NAME”);
2) Technical alarm information:
Reports the current condition of the monitor or sensors; this section will display
alarm information;
3) Physiological alarm information:
If the patient's physiological parameters exceed the alarm limit, this section will
display alarm information;
6) Battery status:
Current battery capacity or its condition.
The user may change the settings of the monitored parameters mentioned above. This
will be described in detail in later chapters.
The name appears in upper left side of each waveform. The ECG waveform gain and
filter mode will be also displayed beside the ECG wave name. On the right side of the
ECG waveform is a mark demonstrating the unit of 1 mV. The gain setting of the
respiration waveform is displayed on the right side of the name of respiration wave.
When the user pushes the keys on the patient monitor, a window may pop up in the
waveform section. The waveform section will return to normal after the window is
minimized.
1.8 Menu
1.8.3 Settings
By selecting the “Setting” tab, the user may choose to enter the “Alarm Setting”,
“Record Setting”, “Screen Layout”, “Adjust Time”, “Miscellaneous”, “ECG Setting”, “ST
Setting”, “SpO2 Setting”, “NIBP Setting”, “Resp Setting”, “Temp Setting”, or “Load
Default” windows.
If the screen layout is set to “Trend Table”, the user can change time intervals for the
trend analysis from “1 min” to “60 min” by adjusting the “Trend Time”.
1.8.3.6 Miscellaneous
By entering the “Miscellaneous” window, the user can change the key volume and the
screen brightness. Key volume can be set at any level from 0 to 10 (0 represents
mute). Screen brightness settings range from 1 to 10 (10 representing the maximum
brightness). If “Wave Smooth” is set to “On”, the wave will be displayed in smooth
mode.
If “Yes” is chosen, then the current settings will be replaced with the default settings.
1.8.5 Service
By selecting the “Service” tab, the user can choose to enter the “ECG Calibrate”,
“Temp Sensor Type”, “NIBP Pneumatic Test”, “NIBP Calibration”, “NIBP reset”, “Demo
Mode”, “Version Info”, “User Setting” and “Factory Service” windows, and so on.
When the blood pressure cuff is connected, the user can start the air leakage test by
selecting this option. The test will determine whether condition of the gas tubes and
apparatus is airtight or not. The test results will be displayed as follows:
If the air leakage test is passed, the system will not make any prompts.
If the test result is a fail, the corresponding failure prompts will be displayed in the
screen segment providing information about noninvasive blood pressure.
For detailed instructions, please refer to Chapter 8.5 "Checking for leaks".
If the blood pressure measurement is abnormal but the monitor does not signal any
prompts to suggest the cause of the problem, we suggest using this option. If the BP
module functions abnormally, resetting it may resolve the error.
Warning
Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC standards (e.g. IEC 60950 for data processing equipment
and IEC 60601-1 for medical equipment). Furthermore all configurations must comply with
the valid version of the system standard IEC 60601-1-1. Everybody who connects
additional equipment to the signal input part or signal output part is configuring a medical
system and is therefore responsible for ensuring that the system complies with the
requirements of the valid version of the system standard IEC 60601-1-1. If in doubt,
please consult the technical service department or your local representative.
1.11 Networks
The network port on the monitor is a standard RJ45 network interface and can be used
to communicate with the central station through an Ethernet cable to enable remote
monitoring. In the top right corner of the screen is a network icon which depicts the
current network status. If the network cable is disconnected, the following network
status icon is shown “ ”. Once the monitor has established connection with the
central station, the following icon is shown “ ”. If the monitor is communicating
normally with the central monitoring system, the following icon is shown “ ”.
When operating on battery power, the monitor measures the battery level and triggers
an alarm when the battery level is insufficient; the prompt “BAT LOW” appears in the
information section. If this happens, the AC power cable should be plugged in; this will
immediately start to charge the battery before it can deplete. If battery continues to be
used as the power supply, the monitor will power off automatically when the battery is
exhausted.
Warning
If you have any doubts about the earthing arrangement or its performance, you must use
the built-in battery to power the monitor.
1.13 Installation
• Make sure the AC power supply complies with following specifications: 100-240
VAC, 50/60 Hz.
Connect the power cable provided to the monitor. Plug the power cable to the
INPUT socket on the monitor. Connect the other end of the power cable to an
earthed 3-phase power output.
Note
• Connect to the ground line if necessary. Refer to the Chapter 13 "Patient Safety"
for details.
Notes
If the monitor finds any fatal errors during the self-test, it will trigger an alarm.
Check all functions that are to be used during monitoring, and make sure that the monitor
is operating normally.
The battery must be recharged to full capacity after each use to ensure adequate electrical
back-up.
Warning
If any signs of damage are detected, or if the monitor displays any error messages, do not
use it on any patients. Contact the biomedical engineer in the hospital or HEYER
Customer Service Center immediately.
Note
The interval between two presses of the POWER button must be more than 1 minute.
Note
Patient alarms: triggered if the patient's physiological parameters exceed the set limits,
e.g. patient’s body temperature exceeds the temperature alarm limit.
The monitor has set levels for technical alarms and patient alarms.
The audible alarms are emitted by the speaker, the visible alarms occur at the LEDs,
and the alarm descriptions are displayed on the screen. The patient alarm is displayed
in the patient alarm information section, while technical alarms are displayed in the
monitor alarm information section.
If there is “*” in the technical or physiological information section, this means a low-level
alarm. “**” means a medium-level alarm, and the background color of the relevant
section will turn yellow. “***” means a high-level alarm, and the background color of the
relevant section will turn red. For example: “** HR TOO HIGH” indicates a medium-
level alarm.
Patient alarms have 2 kinds of alarm mode: LATCH or Non Latch. LATCH means that
once an alarm has been triggered, the system will continue to indicate the alarm until
manual intervention (such as pushing the “SILENCE” button on the panel). Non Latch
means that the system will stop signaling an alarm once the condition that triggered the
alarm is no longer applicable.
Three options can be set for the alarm suspension time: 1 minute, 2 minutes and
3 minutes. To select the desired time, the user must enter the “Alarm Setting” window
and select the suspension time accordingly. If the “PAUSE” key is pressed again, the
system will restore to the normal state.
If an alarm state still exists while the alarm is silenced, the information segment on the
screen will display this alarm information.
If there is no alarm state triggered while the alarm is silenced, all alarms will be
cancelled.
Attention
When the system is operating in “SILENCE” mode, any newly triggered alarms will cause
the monitor to exit silence mode, and the system will return to normal alarm conditions.
The adjustment range for temperature alarm limits is: 0 – 50°C (32 – 122°F).
Selecting “Yes” will cancel all current alarm limit settings and restore the default alarm
limit settings.
Attention
1. When different levels of alarm are triggered simultaneously, the audible alarm emitted
corresponds to the highest level alarm.
2. While the alarm is paused, monitoring will not process any alarm information.
1) Record wave 1, Record wave 2: there are five wave options to choose from: Off,
ECG1, ECG2, Pulse wave, Respiratory wave. The user may choose to record two
waves simultaneously or select the recording of one wave while recording of the
other wave is set to “Off”.
2) Record length: there are two options: “continuous” and “8 seconds”. “Continuous”
means that wave output is recorded continuously until the “RECORD” key is
pressed down again.
3) Record period: the time interval between two record outputs. There are 10 options:
Off, 10 minutes, 20 minutes, 30 minutes, 40 minutes, 50 minutes, 1 hour, 2 hours, 3
hours and 4 hours. The recording length is 8 seconds.
4) Record speed: there are two options: “25.0mm/s” and “50.0mm/s”.
5) Record grid: “Off” means non- grid output. “On” means grid output.
Enter the “Patient Info” window by choosing the “Admit new patient” menu, and input
the patient information (see graphic below).
Select the “Ok” button to quit; the patient’s information will be accepted.
Note
If do not discharge a patient first before admitting a new patient, the new patient’s
measurement data will be saved with the preceding patient's data. The monitor cannot
distinguish the new patient’s data from the data of the previous patient.
The ECG measures the electrical activity of the heart and depicts this electrical activity
in wave-form on the monitor, together with the heart rate.
The RA (right arm) electrode — place in the subclavian area near the right shoulder.
The LA (left arm) electrode — place in the subclavian area near the left shoulder.
The LL (left leg) electrode — place under the left abdomen.
The RL (right leg) electrode — place under the right abdomen.
The V (chest) electrode — place on the chest.
• When connecting the cables and electrodes, make sure no conductive parts are in
contact with the ground. Verify that all ECG electrodes, including neutral electrodes,
are securely attached to the patient.
• Verify lead fault detection before start of monitoring. Unplug the ECG cable from the
socket, the screen will display the error message “ECG LEAD OFF” and the audible
alarm is activated.
• Use five lead wires. Use only silver/silver chloride (Ag/AgCl) ECG electrodes and
cables that meet AAMI standards.
• The following table shows the standard lead names and colors for Europe and the
United States.
The R wave height is big and located completely above the baseline or under.
The amplitude of the P wave and the T wave is smaller than 0.2mV.
Warnings
• Do not touch the patient, table, bed or the equipment during defibrillation.
• Use only the recommended manufacturer ECG cable for monitoring. Other ECG cables
and leads may cause improper performance and/or provide inadequate protection
during defibrillation.
Note
Interference from a non-earthed device near the patient and ESU interference can cause
inaccuracy of the waveform.
Warning
When using ESU equipment, leads should be placed in a position at equal distance from
ESU electrotome and the grounding plate to avoid cautery. ESU equipment wire and ECG
cable must not be tangled up.
1) Pacemaker: When it is turned on, the pacing signal is shown as a vertical line
above ECG wave l; when it is turned off, the pacing signal will not be detected.
2) Channel 1 lead, Channel 2 lead: There are 7 leads: I, II, III, AVR, AVL, AVF, V.
3) Channel 1 size, Channel 2 size: There are four sizes: “×0.25”, “×0.5”, “×1”, “×2”.
A 1 millivolt scale is displayed to the right of the ECG wave, the height of which is
in direct ratio with the wave amplitude.
4) Notch: work frequency suppression switch; when it is “On”, it will filter the AC
interference of the ECG signal.
5) Filter: There are 3 filter modes: diagnostic, monitor and surgery.
In “Diagnostic” mode, the ECG wave without filtering is displayed;
In “Monitor” mode, the artifact which causes the false alarm is filtered out;
In “Surgery” mode, artifacts and the interference caused by electrosurgical
equipment can be reduced.
The filter modes are displayed above the ECG wave.
6) Heart volume: the range is from 0 to 10, “0” means that the sound of the
heartbeat is muted, “10” means it is on the maximum volume.
7) Wave speed: There are three levels of the ECG waveform tracing speed to be
chosen: 12.5, 25.0 and 50.0 mm/s.
8) HR source the options are “Auto”, “ECG”, “SpO2”. When “ECG” is selected, HR
and heart sound are from ECG; when “SpO2”is selected, HR and heart sound are
from SpO2; when “Auto” is selected, the patient monitor will automatically detect
the ECG and SpO2 signal. HR will come from ECG when and ECG signal exists,
otherwise it will come from SpO2.
9) ST switch: When it is “On”, ST analysis will be performed; otherwise it will not.
Note
When the “Pace” is turned on, arrhythmia related to PVC Premature Ventricular
Contractions (including the PVCs computation), will not be detected; simultaneously, ST
section analysis will not be performed
Warning
1) Apnea time: This is an alarm to indicate when the patient is not breathing. The
apnea time (time between breaths) can be set between 10 seconds and 40
seconds. If the setting is set to ”Off”, this means that the apnea alarm is off.
2) Wave speed: you can set the wave speed to 6.25mm/s, 12.5mm/s or 25.0 mm/s.
3) Wave size: The user may set the wave amplitude enlargement factor. There are a
total of 5 levels: ×0.25, ×0.5, ×1, ×2, ×4.
4) RR Source: when “ECG” is selected, RR is from ECG leads; when “CO2” is
selected, RR is from the CO2 module and AwRR is displayed in the parameter
area.
• Respiration monitoring is not recommended on a patient who moves a lot, as this may
trigger a false alarm.
• Place the RA and LL electrodes on opposite sides of the patient’s body in order to
obtain the best respiration wave. Avoid the liver area and the ventricles of the heart in
the line between the RESP electrodes to prevent cardiac overlay or artifacts from
pulsating blood flow. This is particularly important in neonates.
Note
Place the SpO2 sensor cable along the back of the patient’s hand. Make sure the
fingernail is just opposite to the light emitted from the sensor.
Note
If the sensor cannot be positioned accurately, this may result in an inaccurate SpO2
reading, or it may even mean that the SpO2 cannot be measured because no pulse is
detected. If this is true, you must position the sensor again.
Excessive patient movement may result in an inaccurate reading. In this situation, you
must keep the patient still or change the site for monitoring in order to reduce the adverse
influence of excessive movement.
Warning
In the event of prolonged and continuous monitoring, you should check the peripheral
circulation and the skin every 2 hours. If any adverse changes take place, you should
change the measurement position in time.
In the event of prolonged and continuous monitoring, you should periodically check the
positioning of the sensor. If the positioning of the sensor moves during monitoring, the
measurement accuracy may be affected.
1) Pulse volume: the volume range is 0 to 10. 0 shows that pulse sound is muted; 10
is the maximum volume.
2) Sensitivity: the sensitivity for measuring the oxygen saturation value has three
option: “high”, “medium” and “low”.
3) Wave Speed: the wave scanning velocity has two options to choose between: 12.5
and 25mm/s.
4) Pulse Rate: when se to “On”, the parameter area will show the pulse rate;
otherwise the pulse rate will not be displayed.
5) Wave Mode: when “Line” is selected, the monitor will use line mode to draw the
pleth wave; when “Fill” is selected, the monitor will use fill mode to draw the pleth
wave.
Warning
and move it when the skin deteriorates. More frequent examinations may be required
for different patients.
7) In the event of prolonged continuous monitoring, inspect the SpO2 measurement, the
peripheral circulation and the condition of the skin about every 2 hours. If the
discovered changes are not good, you should change the SpO2 measurement
promptly. At the same time, check periodically to ensure that the sensor is secure; this
is to avoid changes in the sensor fastness situation caused by the moving and so on,
as such factors can affect the accuracy of the measurement.
8) Inaccurate measurement can be caused by incorrect application or use of an SPO2
sensor; in case of lost measurement sensor has to be relocated.
9) Excessive movement during measurement of SpO2 may result in an inaccurate
measurement. In this case, the patient must be calmed down, and the measurement
sensor must be replaced.
Warnings
• Use accessories specified by HEYER only, otherwise the device may not function •
normally.
• Before starting a measurement, verify that you have selected a setting appropriate for
your patient (adult, pediatric or neonatal)
• Do not apply the cuff to a limb that has an intravenous infusion or catheter in place.
This could cause tissue damage around the catheter when the infusion is slowed or
blocked during cuff inflation.
• Make sure that the air tube connecting the blood pressure cuff and the monitor is
neither blocked nor tangled.
Notes
• If blood pressure measurement is to be used as the basis for establishing therapy, this
must be obtained using a different blood pressure measurement method, such as the
cuff/stethoscope auscultation method. Accordingly, the doctor must note that the
values obtained using other methods may differ from those obtained sing the SCALIS.
1) Insert the gas tube into the blood pressure socket of the monitor;
2) Place the blood pressure cuff around the patient’s upper arm or the thigh;
3) Use a suitable size cuff for the patient. Make sure that the symbol Ф is located over
a suitable artery. Make sure that the cuff is not too tight; otherwise this may cause
the limb periphery to change color, or it can even cut off blood supply;
4) Check that the edges of the cuff fall in the range marked <->.If not, swap for a more
appropriate cuff;
5) Make sure that the cuff is deflated completely;
6) Attach the cuff to the gas tube. The body part which is to be measured should
placed at the same horizontal level as the patient’s heart. If this is not possible, use
the following adjustment method to revision the measurement result:
If the cuff is at a higher horizontal level than the heart, add 0.75mmHg (0.10kPa) for
each centimeter difference.
Notes
• The width of the cuff should be either 40% of the limb circumference (50% for
neonates) or 2/3 of the upper arm length. The inflatable part of the cuff should be long
enough to encircle 50-80% of the limb. The wrong size of cuff can cause erroneous
readings. If there are any doubts about the cuff size, then use a larger cuff.
• For neonatal measurements, the maximum pressure of the cuff is 147mmHg. The
maximum pressure is 140mmHg when the cuff is used normally. The initial inflating
pressure of the cuff is 70mmHg during measurement.
1) Unit: select the unit of pressure from the options “mmHg” or “kPa”.
2) Mode: there are 3 kinds of measurement mode: manual, automatic, STAT.
In the manual measurement mode, pressing down the blood pressure
measurement button on the control panel will result in a single manual
measurement of blood pressure;
In the automatic measurement mode, pressing down the blood pressure
measurement button on the control panel will result in a single immediate blood
pressure measurement, followed by automatic measurements initiated by the
monitor after a defined period;
In the STAT measurement mode, pressing down the blood pressure measurement
button on the control panel will start continuous measurement for 5 minutes.
Connect the pressure gauge and the cuff via a 3-way tube to the blood pressure
trachea jack on the monitor. Set the monitor to the “Calibration” mode, then inflate the
cuff using an air pump. First increase the pressure to 250 mmHg, then slowly deflate.
When the monitor display 200, 150 and 50 mmHg, the disparity between the standard
pressure gauge value and the monitor pressure value should be less than 3 mmHg. If
the value exceeds 3 mmHg, please contact our company’s service technician.
Note
Warning
You should calibrate NIBP measurements once every two years (or as required in your
hospital’s maintenance regulations). You should monitor performance according to the
following information.
Warning
1) Do not perform noninvasive blood pressure measurement on patients who have the
sickle cell anemia or damaged skin or this will result in injury.
2) Patients with susceptible to blood clots must be clinically assessed before deciding
whether to carry out the automatic blood pressure measurement, because there is a
risk of haematoma at the place where the body and the cuff rub together.
3) Before starting the measurement, you must confirm that the patient type is correct
(adult, pediatric, neonate).
4) Do not attach the cuff to a limb where there is an inravenous infusion or an inserted
syringe driver, as this may cause damage around the syringe driver when the
transfusion reduces speed or stops while cuff if inflating.
5) If noninvasive blood pressure measurement is used on automatic mode for too long,
then the limb with the cuff attached may develop purpura, lack of blood flow and
neuralgia. When monitoring the patient, it is necessary to check the luster, warmth and
sensitivity of limb periphery frequently. If any abnormalities are observed, please stop
the blood pressure measurement immediately.
6) Calibration of the noninvasive blood pressure measurement is supposed to be carried
out one time every year. (Or according to the maintenance regulations of your
hospital).
7) The cuff width should be 40% of the limb circumference (50% for neonates) or the 2/3
of the upper arm length. The length of the cuff inflation part should long enough to
surround 50-80% of the limb. A cuff of inappropriate size can result in an incorrect
reading. If there are any doubt about the cuff size, a bigger cuff should be used in order
to reduce errors.
Warning
Before starting to use temperature measurement, please check whether the sensor cable
is normal. Unplug the temperature sensor cable from the socket: the screen will display
the error message “Temp sensor off” and emit an audible alarm.
Choosing “Trend Graph” in the “Review” menu will open the following window:
In the Trend Graph window, the time is shown underneath the X axis, with the most
recent time displayed on the nearest right side. The value ranges for the various
parameters is displayed on left side of the Y axis.
Choosing “Trend Table” in the “Review” menu will open the following window:
In “Trend Table” window, the time shows underneath the parameter columns: the most
recent time is displayed on the nearest right side; the parameter name and the unit are
displayed in the first column.
Alarm events can also be observed in the trend table: The time of any triggered alarms
is also saved in the trend data. If a parameter triggers an alarm, the trend data in the
corresponding alarm time period will be displayed with a yellow background color.
Selecting “Alarm Review” in the “Review” menu will display recent alarm event
information, just as the following chart shows:
• Sequence number: format is I/N whereby I means the index of alarm event and N
means the total number of alarm events in the database, as shown in Fig. 10-3. The
most recent alarm has the smallest number, e.g. No 1 means the most recent alarm.
• Alarm event time;
• Alarm event type;
• Parameters when alarm occurs;
• 2 wave channels, 16 seconds for each channel;
10.3.4 Record
The recorder will print out the current alarm event if the user presses the ”Record”
button.
Selecting “NIBP Review” in the “Review” menu will display the results and the times of
the 10 most recent pieces of noninvasive blood pressure measurement data, just as
the following window shows:
The data is arranged in order according to the time. The most recent measurement
data is displayed at the top. 10 lines of measurement data can be displayed on the
screen at a time. The buttons “ ” and “ ”can display the previous or the next
measurement data. With the buttons “ ” and “ ”, the operator can move the review
table a page forward or backward. By selecting the button “ ”, the operator may see
the earliest measurement data, and “ ”the most recent.
Selecting “Wave Review” in the “History Review” menu will display the most recent
wave data, just as the following chart shows:
The section above the wave shows the related information: wave scanning velocity,
current review time, and a table of the measured parameters.
Note
Trend data can be stored for 720 hours after turning off of the monitor. If the monitor is
turned on after more than 720 hours without power, the trend data will be erased.
Wave review data can be stored for 2 hours after turning off the monitor. If the monitor is
turned on after more than 2 hours without power, the wave review data will be deleted.
This monitor provides the function of dose calculation for 21 kinds of medicines and the
titration table.
Selecting “Dose Calculation” in the menu will open the window as shown in the
following graphic:
Concentration = Amount/volume
Inf rate = Dose / Concentration
Duration = Amount / Dose
Dose = Inf rate × Concentration
Rotate the knob and move the cursor to each calculation item in the formula
separately. Press down and rotate the knob, select the calculation values. After
selection, the value of the calculated item will be displayed in the corresponding place.
Drug name selection: move the cursor to ”drug name”, rotate the knob. It is possible to
choose from 21 kinds of medicines, AMINOPHYLLIN, DOBUTAMINE, DOPAMINE,
EPINEPHRINE, HEPARIN, ISUPREL, INOCOR, INSULIN, INSUPREL, LIDOCAINE,
NIPRIDE, NITROGLYCERIN, NOREPINEPHRINE, PITOCIN, PROCAINAMIDE,
VASOPRESIN, DRUG A, DRUG B, DRUG C, DRUG D, DRUG E. Only one type of
medicine can be calculated at a time.
If you find any damage on the monitor, stop using the monitor on the patient and
contact the biomedical engineer of the hospital or the HEYER Customer Service
Department immediately.
A complete check of the monitor, including the functional safety check, must be
performed by qualified personnel once every 6 to 12 months or each time after repair.
All checks that require the casing of the monitor to be opened must be performed by
qualified service personnel. Safety and maintenance checks can also be conducted by
HEYER personnel. You can obtain information about the customer service contract
from your local HEYER office.
HEYER provides all SCALIS circuit diagrams, parts lists and standard instructions.
Warning
If the hospital or agency that using the monitor does not follow a satisfactory maintenance
schedule, the monitor may become invalid, and human health may be endangered.
Warning
Turn off the power and disconnect the power line before cleaning the monitor or the
sensor/probe.
It is recommended that you should clean the outside surface of the monitor casing and
the display screen regularly. Only use non-caustic detergents such as soap and water
to clean the monitor casing.
Pay special attention to the following to avoid damaging the SCALIS monitor:
1. Avoid using ammonia-based or acetone-based cleaners such as acetone.
2. Most cleaning agents must be diluted before use. Dilute the cleaning agent as per the
manufacturer's instructions.
3. Do not use abrasive material, such as steel wool etc.
4. Do not let the cleaning agent enter the monitor. Do not immerse any part of the system
into liquid.
5. Do not leave cleaning agents on any part of the equipment.
Note
Diluted sodium hypochlorite from 500ppm (1:100 diluted bleach) to 5000ppm (1:10 bleach)
is very effective. The concentration of the diluted sodium hypochlorite depends on how
many organisms (blood, mucus) are left on the surface of the casing.
Notes
• You can use hospital-grade ethanol to clean the SCALIS monitor and its sensor/probe.
Leave it to dry naturally or use a clean cloth to dry it.
12.4 Sterilization
To avoid extensive damage to the equipment, sterilization is only recommended when
stipulated as necessary in the Hospital Maintenance Schedule. Sterilization facilities
must be cleaned first.
Appropriate sterilization materials for ECG leads and blood pressure cuffs are
mentioned in the chapters “ECG Monitoring”, “Respiration Monitoring” and “NIBP
Monitoring” respectively.
Caution
• Follow the manufacturer’s instructions to dilute the solution, or use the lowest possible
concentration.
• Do not let liquid enter the monitor.
• Do not immerse any part of the monitor in liquid.
• Do not pour liquid onto the monitor during sterilization.
• Use a moistened cloth to wipe off any agent remaining on the monitor
12.5 Disinfection
To avoid extensive damage to the equipment, disinfection is only recommended when
stipulated as necessary in the Hospital Maintenance Schedule. Disinfection facilities
should be cleaned first.
Appropriate disinfection materials for the ECG lead, SpO2 sensor, blood pressure cuff
and TEMP probe are introduced in the relevant chapters.
Caution
Warning
Do not come into contact with patients, beds or the monitor during defibrillation.
13.1 Environment
Follow the instructions below to ensure complete and safe electrical installation. The
environment where the PM-5000 Multi-Parameter Patient Monitor will be used should
be reasonably free from vibration, dust, corrosive or explosive gases, extremes of
temperature, humidity, and so on. For cabinet-mounted installation, allow sufficient
room at the front for operation and sufficient room at the rear for servicing with the
cabinet access door open.
13.3 Grounding
To protect the patient and hospital personnel, the casing of the Patient Monitor must be
grounded. Accordingly, the Patient Monitor is equipped with a detachable 3-pin cable
which grounds the instrument to the power line ground (protective earth) when plugged
into an appropriate 3-pin socket. If a 3-pin socket is not available, consult the hospital
electrician.
To protect the patient and hospital personnel, the case of patient monitor must be
grounded. If after consulting the instrument specifications the user is confused as to
whether a particular combination of equipment is dangerous, the user should consult
the manufacturer or other experts in this field to ensure that all instruments which
require security will not be damaged by the proposed combination.
The equipotential grounding system is for the safety function of the protective
grounding conductor if ever there is a break in the protective grounding system.
Examinations in or on the heart (or brain) should only be carried out in medically used
rooms incorporating an equipotential grounding system. Check each time before use
that the instrument is in perfect working order.
13.5 Condensation
When the equipment is moved from one room to another, condensation may form. This
is because the equipment may be exposed to humid air and temperature differences,
among other reasons.
Warning
If there are places where flammable anesthetics are used, there will be a risk of explosion.
Caution
Notes
The monitor should not be used adjacent to or stacked upon other equipment. If adjacent
or stacked use is necessary, the monitor should be observed to verify normal operation in
the configuration in which it will be used.
The monitor needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided below.
Portable and mobile RF communications equipment can affect this monitor. See tables
14-1, 14-2, 14-3, and 14-4 below.
The monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitor should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment — guidance
RF emissions Group1 The monitor uses RF energy only for its internal
CISPR 11 function. Therefore, its RF emissions are very low and
are not likely to cause any interference with nearby
electronic equipment.
RF emissions Class A The monitor is suitable for use in all establishments
CISPR 11 other than domestic and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes
Harmonic Class A
Emissions
IEC61000-3-2
Voltage Compliance
Fluctuations/
Flicker
Emissions IEC
61000-3-3
The monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of the monitor should assure that it is used in such an environment.
Immunity test IEC 60601 Compliance level Electromagnetic environment —
Test level guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete
Discharge ±8 kV air ±8 kV air or ceramic tile. If floors are
(ESD) covered with synthetic material,
IEC 61000-4-2 the relative humidity should be at
least 30%.
Electrical Fast ±2 kV for power ±2 kV for power Mains power quality should be
Transient/Burst supply lines ±1 kV supply lines ±1 kV that of a typical commercial or
IEC 61000-4-4 for input/output for input/output hospital environment.
lines (>3m). lines (>3m).
Surge ±1 kV differential ±1 kV different Mains power quality should be
IEC 61000-4-5 mode ±2 kV mode ±2 kV that of a typical commercial or
common mode common mode hospital environment.
Voltage dips, <5% UT <5% UT Mains power quality should be
Short (>95% dip in UT) (>95% dip in UT) that of a typical commercial or
interruptions and for 0.5 cycle for 0.5 cycle hospital environment. If the user
voltage variation of our product requires continued
on power supply 40% UT 40% UT operation during power mains
input lines (60% dip in UT) for (60% dip in UT) for interruptions, it is recommended
IEC 61000-4-11 5 cycle 5 cycle that our product be powered from
an uninterruptible power supply or
70% UT 70% UT a battery.
(30% dip in UT) for (30% dip in UT) for
25 cycle 25 cycle
<5% UT <5% UT
(>95% dip in UT) (>95% dip in UT)
for 5 sec for 5 sec
Power 3 A/m 3 A/m Power frequency magnetic fields
frequency should be at levels characteristic
(50/60 HZ) of a typical location in a typical
magnetic field commercial or hospital
IEC 61000-4-8 environment.
The monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the monitor should assure that it is used in such an environment
symbol:
Note — At 80 MHz and 800 MHz, the higher frequency range applies.
Note — These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the monitor is used
exceeds the applicable RF compliance level above, the monitor should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the monitor.
b
Over the frequency ranges 150kHz to 80MHz, field strengths should be less than 3V/m.
For transmitters at a maximum output power not listed above, the recommended separation
distance in meters (m) can be determined using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note — At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note — These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Warning
The patient monitor may not meet its performance specifications if stored or used outside
the manufacturer’s specified temperature and humidity range.
15.1 Classification
Item Specification
MDD classification Class IIb
Anti-electroshock degree Class I equipment with internal power supply
TEMP/SpO2/NIBP: BF
Anti-electroshock degree
ECG/RESP CF
Explosion proof level Ordinary equipment. without explosion proof
Harmful liquid proof degree Ordinary equipment, without liquid proof
Working system Continuous running equipment
15.5 Battery
2.3 Ah 12V rechargeable battery
Operating time after full charge is more than 1 hour
Operating time after the first alarm of low battery will be about 5 minutes
Maximum charging time is less than 6 hours.
15.7 Storage
Trend 720 hours
NIBP review 1000 NIBP events
Wave review 2 hours
Alarm review 200 alarm events
All storage data are non volatile.
15.8 Environment
Temperature
Working 0 - 40°C
Storage -20 - 50°C
Humidity
Working 15% - 90 %
Storage 15% - 90 % (no coagulation)
Atmospheric pressure
Working 86.0 kPa - 106.0kPa;
Storage 86.0 kPa - 106.0kPa
15.9 ECG
a) Couplet rhythm — two waves of duration is 1500ms; if calculate all the QRS
complex, heart rate is 80bpm, if only calculate large R wave or S-wave, the heart
rate is 40bpm.
b) Slow change couple rhythm — if calculate all the QRS complex, heart rate is
60bpm, if only calculate large waves, heart rate is 30bpm.
c) Fast couple rhythm — if the calculate all the QRS complex, heart rate is 120bpm.
d) bi-directional contraction — if calculate all the QRS complex, heart rate is 90bpm, if
only large waves, heart rate is 45bpm.
15.9.7 Sensitivity
> 200 µV P-P
15.9.9 Bandwidth
Surgery 1 - 20 Hz
Monitor 0.5 - 40 Hz
Diagnostic 0.05 - 130 Hz
15.9.10 CMRR
Diagnostic Mode >90 dB
Monitor Mode >110 dB
Surgery Mode >110 dB
When the pace switch is “On”, the patient monitor can inhibit pace pulse without affect
heart rate calculation: ± 2 mV - ± 5mV, width: 0.5ms - 2ms, rise time: 10µs - 100µs
single pulse overshoot normal QRS waves of pacing pulses.
Overshoot (a0) range should 0.025ap to 0.25ap range, independent of the choice of
time constant, but not more than 2mV;
When the pace switch is “On”, pace pulse inhibition of rapid ECG signal to 1V / s RTI
minimum input slew rate.
Baseline Stability: After boot, 10s baseline drift in the output rate should not exceed
10µV / s RTI;
The maximum response time is less than 10 s , for step change of heart rate from
80bpm to 40bpm;
15.10 Respiration
15.10.1 Method
Impedance between RA-LL
15.10.4 Bandwidth
0.3 - 2.5 Hz
15.10.5 Gain
x0.25, x0.500, x1, x2, x4
15.11.1 Method
Oscillometry
15.11.8 Resolution
1mmHg
15.12 SpO2
15.12.2 Resolution
1%
15.12.3 Accuracy
70% - 100% ±2 %
<69% unspecified
15.13 Temperature
Channel 2
Measure and Alarm Range 0 - 50°C
Resolution 0.1°C
Accuracy (no sensor) ± 0.1°C (0°C – 50°C)
1. Within the warranty period we will remedy free of charge any defects or damages on the device
that are shown to be caused by a manufacturing or material error. The warranty does not include
easily breakable parts, e.g. glass or consumable parts.
2. Warranty services can only be claimed upon submission of a delivery note (bill of delivery or
invoice); the type and method of damage remedy (repair or replacement) shall be at the
discretion of HEYER MEDICAL AG. Warranty services do not result in an extension of the
warranty period, nor do they entail a new warranty being granted. There is no independent
warranty period for installed spare parts.
3. Excluded from the warranty are: Damages caused by improper use, operating errors, mechanical
stress or non-observance of the operating instructions, as well as damages caused by force
majeure or by extraordinary environmental conditions.
4. Warranty services may only be claimed if proof is submitted to confirm that all service and
maintenance work has been carried out by authorized staff.
5. The warranty includes all faults that impair a faultless functioning of the device on the basis of
technical defects of individual components. The warranty obligation can only be recognized by us
if the device has been used properly and according to its intended use and no repair attempts
have been undertaken by the client himself or by third parties. The warranty claim does not
include faults caused by mechanical damages or if the device is being operated with accessories
originating from third parties.
6. The warranty is also void if changes, alterations or repairs are made to the device by persons not
authorized to do this.
7. The warranty claim only applies to customers of HEYER MEDICAL AG; it cannot be transferred to
third parties.
8. The rejected device is to be shipped back to our plant postage free. In case of a request by our
customer service department, the costs for shipping to the plant are in general to be initially borne
by the customer. After successful repair, we will send the device back freight collect. If HEYER
MEDICAL AG confirms the existence of a warranty claim, the customer will receive
reimbursement for the costs of delivery and/or transport of the apparatus. Repair parts that do not
fall under the warranty claim will be billed by us. The shipping of the device to us always counts
as a complete assignment to eliminate all faults and/or replace missing parts, unless the
customer expressly excludes partial services. Additional claims to transfer or reduce and replace
damages of any kind in particular also of damages not incurred on the object of delivery itself are
excluded.