Beruflich Dokumente
Kultur Dokumente
6.- Applications
2.- The Physics Surface Breaking Cracks
Properties of Electricity SBC using Sliding Probes
Current Flow & Ohm's Law Metal Thinning (Corrosion)
Induction & Inductance Tube Inspection
Self Inductance Conductivity
Mutual Inductance Heat Treat Verification
Circuits & Phase Thickness of Thin Mat'ls
Impedance Thickness of Coatings
Depth & Current Density
Phase Lag
7.- Advanced Techniques
Scanning
3.- Instrumentation Multi-Frequency Tech.
Eddy Current Instruments Swept Frequency Tech.
Resonant Circuits Pulsed ET Tech.
Bridges Background Pulsed ET
Impedance Plane Remote Field Tech.
Display - Analog Meter Impedance Matching
Eddy current inspection is one of several NDT methods that use the principal of
“electromagnetism” as the basis for conducting examinations. Several other
methods such as Remote Field Testing (RFT), Flux Leakage and Barkhausen
Noise also use this principle.
One of the major advantages of eddy current as an NDT tool is the variety of
inspections and measurements that can be performed. In the proper
circumstances, eddy currents can be used for:
Crack detection
Material thickness measurements
Coating thickness measurements
Conductivity measurements for:
o Material identification
o Heat damage detection
o Case depth determination
o Heat treatment monitoring
In thin materials such as tubing and sheet stock, eddy currents can be used to
measure the thickness of the material. This makes eddy current a useful tool for
detecting corrosion damage and other damage that causes a thinning of the
material. The technique is used to make corrosion thinning measurements on
aircraft skins and in the walls of tubing used in assemblies such as heat
exchangers. Eddy current testing is also used to measure the thickness of paints
and other coatings.
Eddy currents are also affected by the electrical conductivity and magnetic
permeability of materials. Therefore, eddy current measurements can be used to
sort materials and to tell if a material has seen high temperatures or been heat
treated, which changes the conductivity of some materials.
Measurements have been carried out both in the laboratory and in the field.
Corrosion trials have demonstrated how material loss can be detected and
quantified in multi-layer aluminum structures. More recently, studies carried out
on three and four layer structures show the ability to locate cracks emerging from
fasteners. Pulsed eddy-current measurements have also been applied to
ferromagnetic materials. Recent work has been involved with measuring the case
depth in hardened steel samples.
2 THE PHYSICS
Properties of Electricity
Electricity
Amperage
The flow of electrons is measured in units called amperes or amps for short. An
amp is the amount of electrical current that exists when a number of electrons,
having one coulomb of charge, move past a given point in one second.
A coulomb is the charge carried by 6.25 x 1018electrons or
6,250,000,000,000,000,000 electrons.
Electromotive Force
The unit of measure for EMF is the volt. One volt is defined as the electrostatic
difference between two points when one joule of energy is used to move one
coulomb of charge from one point to the other. A joule is the amount of energy
that is being consumed when one watt of power works for one second. This is
also known as a watt-second. For our purposes, just accept the fact that one joule
of energy is a very, very small amount of energy. For example, a typical 60-watt
light bulb consumes about 60 joules of energy each second it is on.
Resistance
Ohm's law is the most important, basic law of electricity. It defines the
relationship between the three fundamental electrical quantities: current, voltage,
and resistance. When a voltage is applied to a circuit containing only resistive
elements (i.e. no coils), current flows according to Ohm's Law, which is shown
below.
I=V/R
Where:
Electrical Current
I=
(Amperes)
V = Voltage (Voltage)
R = Resistance (Ohms)
Ohm's law states that the electrical current (I) flowing in an circuit is
proportional to the voltage(V) and inversely proportional to the resistance (R).
Therefore, if the voltage is increased, the current will increase provided the
resistance of the circuit does not change. Similarly, increasing the resistance of
the circuit will lower the current flow if the voltage is not changed. The formula
can be reorganized so that the relationship can easily be seen for all of the three
variables.
The Java applet below allows the user to vary each of these three parameters in
Ohm's Law and see the effect on the other two parameters. Values may be input
into the dialog boxes, or the resistance and voltage may also be varied by moving
the arrows in the applet. Current and voltage are shown as they would be
displayed on an oscilloscope with the X-axis being time and the Y-axis being the
amplitude of the current or voltage. Ohm's Law is valid for both direct current
(DC) and alternating current (AC). Note that in AC circuits consisting of purely
resistive elements, the current and voltage are always in phase with each other.
Exercise: Use the interactive applet below to investigate the relationship of the
variables in Ohm's law. Vary the voltage in the circuit by clicking and dragging
the head of the arrow, which is marked with the V. The resistance in the circuit
can be increased by dragging the arrow head under the variable resister, which is
marked R. Please note that the vertical scale of the oscilloscope screen
automatically adjusts to reflect the value of the current.
See what happens to the voltage and current as the resistance in the circuit is
increased. What happens if there is not enough resistance in a circuit? If the
resistance is increased, what must happen in order to maintain the same level of
current flow?
Induction and Inductance
Induction
Faraday also noticed that the rate at which the magnetic field changed also had an
effect on the amount of current or voltage that was induced. Faraday's Law for
an uncoiled conductor states that the amount of induced voltage is proportional to
the rate of change of flux lines cutting the conductor. Faraday's Law for a straight
wire is shown below.
Where:
It should be noted that since it is the changing magnetic field that is responsible
for inductance, it is only present in AC circuits. High frequency AC will result in
greater inductive reactance since the magnetic field is changing more rapidly.
By studying this image of a coil, it can be seen that the number of turns in the
coil will have an effect on the amount of voltage that is induced into the circuit.
Increasing the number of turns or the rate of change of magnetic flux increases
the amount of induced voltage. Therefore,Faraday's Law must be modified for a
coil of wire and becomes the following.
Where:
The equation simply states that the amount of induced voltage (VL) is
proportional to the number of turns in the coil and the rate of change of the
magnetic flux (dø/dt). In other words, when the frequency of the flux is increased
or the number of turns in the coil is increased, the amount of induced voltage will
also increase.
Where:
Soon after Faraday proposed his law of induction, Heinrich Lenz developed a
rule for determining the direction of the induced current in a loop.
Basically, Lenz's law states that an induced current has a direction such that
its magnetic field opposes the change in magnetic field that induced the
current. This means that the current induced in a conductor will oppose the
change in current that is causing the flux to change. Lenz's law is important in
understanding the property of inductive reactance, which is one of the properties
measured in eddy current testing.
Inductive Reactance
It should be noted that the inductive reactance will increase if the number of
winds in the coil is increased since the magnetic field from one coil will have
more coils to interact with.
Similarly to resistance, inductive reactance reduces the flow of current in a
circuit. However, it is possible to distinguish between resistance and inductive
reactance in a circuit by looking at the timing between the sine waves of the
voltage and current of the alternating current. In an AC circuit that contains only
resistive components, the voltage and the current will be in-phase, meaning that
the peaks and valleys of their sine waves will occur at the same time. When there
is inductive reactance present in the circuit, the phase of the current will be
shifted so that its peaks and valleys do not occur at the same time as those of the
voltage. This will be discussed in more detail in the section on circuits.
Mutual Inductance
(The Basis for Eddy Current Inspection)
The magnetic flux through a circuit can be related to the current in that circuit
and the currents in other nearby circuits, assuming that there are no nearby
permanent magnets. Consider the following two circuits.
The magnetic field produced by circuit 1 will intersect the wire in circuit 2 and
create current flow. The induced current flow in circuit 2 will have its own
magnetic field which will interact with the magnetic field of circuit 1. At some
point P, the magnetic field consists of a part due toi and a part due to i . These
1 2
The coils in the circuits are labeled L1 and L2 and this term represents the self
inductance of each of the coils. The values of L1 and L2 depend on the
geometrical arrangement of the circuit (i.e. number of turns in the coil) and the
conductivity of the material. The constant M, called themutual inductance of
the two circuits, is dependent on the geometrical arrangement of both circuits. In
particular, if the circuits are far apart, the magnetic flux through circuit 2 due to
the current i will be small and the mutual inductance will be small. L and M are
1 2
constants.
We can write the flux, B through circuit 2 as the sum of two parts.
B2 = Li + iM
2 2 1
An equation similar to the one above can be written for the flux through circuit 1.
B1 = Li + iM
1 1 2
Though it is certainly not obvious, it can be shown that the mutual inductance is
the same for both circuits. Therefore, it can be written as follows:
M =M 1,2 2,1
How is mutual induction used in eddy current
inspection?
In the applet below, the probe and the sample are shown in cross-section. The
boxes represent the cross-sectional area of a group of turns in the coil. The liftoff
distance and the drive current of the probe can be varied to see the effects of the
shared magnetic field. The liftoff value can be set to 0.1 or less and the current
value can be varied from 0.01 to 1.0. The strength of the magnetic field is shown
by the darkness of the lines.
A circuit can be thought of as a closed path in which current flows through the
components that make up the circuit. The current (i) obeys Ohm's Law, which is
discussed on the page oncurrent flow. The simple circuit below consists of a
voltage source (in this case an alternating current voltage source) and a resistor.
The graph below the circuit diagram shows the value of the voltage and the
current for this circuit over a period of time.
This graph shows one complete cycle of an alternating current source. From the
graph, it can be seen that as the voltage increases, the current does the same. The
voltage and the current are said to be "in-phase" since their zero, peak, and valley
points occur at the same time. They are also directly proportional to each other.
In the circuit below, the resistive component has been replaced with an inductor.
When inductance is introduced into a circuit, the voltage and the current will be
"out-of-phase," meaning that the voltage and current do not cross zero, or reach
their peaks and valleys at the same time. When a circuit has an inductive
component, the current (i ) will lag the voltage by one quarter of a cycle. One
L
cycle is often referred to as 360o, so it can be said that the current lags the voltage
by 90o.
This phase shift occurs because the inductive reactance changes with changing
current. Recall that it is the changing magnetic field caused by a changing
current that produces inductive reactance. When the change in current is
greatest, inductive reactance will be the greatest, and the voltage across the
inductor will be the highest. When the change in current is zero, the inductive
reactance will be zero and the voltage across the inductor will be zero. Be
careful not to confuse the amount of current with the amount of change in the
current. Consider the points where the current reaches it peak amplitude and
changes direction in the graph below (0o, 180o, and 360o). As the current is
changing directions, there is a split second when the change in current is
zero. Since the change in current is zero, no magnetic field is generated to
produce the inductive reactance. When the inductive reactance is zero, the
voltage across the inductor is zero.
The resistive and inductive components are of primary interest in eddy current
testing since the test probe is basically a coil of wire, which will have both
resistance and inductive reactance. However, there is a small amount of
capacitance in the circuits so a mention is appropriate. This simple circuit below
consists of an alternating current voltage source and a capacitor. Capacitance in a
circuit caused the current (i ) to lead the voltage by one quarter of a cycle
c
In the image below, a circuit diagram is shown that represents an eddy current
inspection system. The eddy current probe is a coil of wire so it contains
resistance and inductive reactance when driven by alternating current. The
capacitive reactance can be dropped as most eddy current probes have little
capacitive reactance. The solid line in the graph below shows the circuit's total
current, which is affected by the total impedance of the circuit. The two dashed
lines represent the portion of the current that is affected by the resistance and the
inductive reactance components individually. It can be seen that the resistance
and the inductive reactance lines are 90o out of phase, so when combined to
produce the impedance line, the phase shift is somewhere between zero and 90o.
The phase shift is always relative to the resistance line since the resistance line is
always in-phase with the voltage. If more resistance than inductive reactance is
present in the circuit, the impedance line will move toward the resistance line and
the phase shift will decrease. If more inductive reactance is present in the circuit,
the impedance line will shift toward the inductive reactance line and the phase
shift will increase.
The relationship between impedance and its individual components (resistance
and inductive reactance) can be represented using a vector as shown below. The
amplitude of the resistance component is shown by a vector along the x-axis and
the amplitude of the inductive reactance is shown by a vector along the y-axis.
The amplitude of the the impedance is shown by a vector that stretches from zero
to a point that represents both the resistance value in the x-direction and the
inductive reactance in the y-direction. Eddy current instruments with impedance
plane displays present information in this format.
The impedance in a circuit with resistance and inductive reactance can be
calculated using the following equation. If capacitive reactance was present in the
circuit, its value would be added to the inductance term before squaring.
The phase angle of the circuit can also be calculated using some trigonometry.
The phase angle is equal to the ratio between the inductance and the resistance in
the circuit. With the probes and circuits used in nondestructive testing,
capacitance can usually be ignored so only inductive reactance needs to be
accounted for in the calculation. The phase angle can be calculated using the
equation below. If capacitive reactance was present in the circuit, its value would
simply be subtracted from the inductive reactance term.
or
The applet below can be used to see how the variables in the above equation are
related on the the vector diagram (or the impedance plane display). Values can be
entered into the dialog boxes or the arrow head on the vector diagram can be
dragged to a point representing the desired values. Note that the capacitive
reactance term has been included in the applet but as mentioned before, in eddy
current testing this value is small and can be ignored.
In previous pages, Ohm's Law was discussed for a purely resistive circuit. When
there is inductive reactance or capacitive reactance also present in the circuit,
Ohm's Law must be written to include the total impedance in the circuit.
Therefore, Ohm's law becomes:
I=V/Z
Ohm's law now simply states that the current (I), in amperes, is proportional to
the voltage (V), in volts, divided by the impedance (Z), in ohms.
The applet below can be used to see how the current and voltage of a circuit are
affected by impedance. The applet allows the user to vary the inductance (L),
resistance (R), voltage (V)and current (I). Voltage and current are shown as they
would be displayed on an oscilloscope. Note that the resistance and/or the
inductive reactance values must be changed to change the impedance in the
circuit.
Also note that when there is inductance in the circuit, the voltage and current are
out of phase. This is because the voltage across the inductor will be a maximum
when the rate of change of the current is greatest. For a sinusoidal wave form like
AC, this is at the point where the actual current is zero. Thus the voltage applied
to an inductor reaches its maximum value a quarter-cycle before the current does,
and the voltage is said to lead the current by 90o.
The depth that eddy currents penetrate into a material is affected by the
frequency of the excitation current and the electrical conductivity and magnetic
permeability of the specimen. The depth of penetration decreases with increasing
frequency and increasing conductivity and magnetic permeability. The depth at
which eddy current density has decreased to 1/e, or about 37% of the surface
density, is called the standard depth of penetration ( ). The word 'standard'
denotes plane wave electromagnetic field excitation within the test sample
(conditions which are rarely achieved in practice). Although eddy currents
penetrate deeper than one standard depth of penetration, they decrease rapidly
with depth. At two standard depths of penetration (2 ), eddy current density has
decreased to 1/e squared or 13.5% of the surface density. At three depths (3 ),
the eddy current density is down to only 5% of the surface density.
Since the sensitivity of an eddy current inspection depends on the eddy current
density at the defect location, it is important to know the strength of the eddy
currents at this location. When attempting to locate flaws, a frequency is often
selected which places the expected flaw depth within one standard depth of
penetration. This helps to assure that the strength of the eddy currents will be
sufficient to produce a flaw indication. Alternately, when using eddy currents to
measure the electrical conductivity of a material, the frequency is often set so that
it produces three standard depths of penetration within the material. This helps to
assure that the eddy currents will be so weak at the back side of the material that
changes in the material thickness will not affect the eddy current measurements.
The applet below illustrates how eddy current density changes in a semi-infinite
conductor. The applet can be used to calculate the standard depth of penetration.
The equation for this calculation is:
Where:
= Standard Depth of Penetration (mm)
= 3.14
f = Test Frequency (Hz)
= Magnetic Permeability (H/mm)
= Electrical Conductivity (% IACS)
(Note: The applet has an input box for relative permeability since this is often the
more readily available value. The applet multiplies the relative permeability of
the material by the permeability of free space to get to H/mm units.)
The applet also indicates graphically the phase lag at one and two standard
depths of penetration. Phase lag will be discussed on the following page.
Phase Lag
Phase lag is a parameter of the eddy current signal that makes it possible to
obtain information about the depth of a defect within a material. Phase lag is the
shift in time between the eddy current response from a disruption on the surface
and a disruption at some distance below the surface. The generation of eddy
currents can be thought of as a time dependent process, meaning that the eddy
currents below the surface take a little longer to form than those at the
surface. Disruptions in the eddy currents away from the surface will produce
more phase lag than disruptions near the surface. Both the signal voltage and
current will have this phase shift or lag with depth, which is different from the
phase angle discussed earlier. (With the phase angle, the current shifted with
respect to the voltage.)
In
Radians
In
Degrees
Where:
=Phase Lag (Rad or Degrees)
x=Distance Below Surface (in or mm)
=Standard Depth of Penetration (in or mm)
At one standard depth of penetration, the phase lag is one radian or 57o. This
means that the eddy currents flowing at one standard depth of penetration ( )
below the surface, lag the surface currents by 57o. At two standard depths of
penetration (2 ), they lag the surface currents by 114o. Therefore, by measuring
the phase lag of a signal the depth of a defect can be estimated.
On the impedance plane, the liftoff signal serves as the reference phase
direction. The angle between the liftoff and defect signals is about twice the
phase lag calculated with the above equation. As mentioned above,
discontinuities that have a significant dimension normal to the surface, will
produce an angle that is based on the weighted average of the disruption to the
eddy currents at the various depths along its length.
In the applet below, the relationship between the depth and dimensions of a
discontinuity and the rotation produced on the impedance plane is explored. The
red lines represent the relative strength of the magnetic field from the coil and the
dashed lines indicate the phase lag of the eddy currents induced at a particular
depth.
3 EINSTRUMENTATION
Eddy Current Instruments
The most basic eddy current testing instrument consists of an alternating current
source, a coil of wire connected to this source, and a voltmeter to measure the
voltage change across the coil. An ammeter could also be used to measure the
current change in the circuit instead of using the voltmeter.
While it might actually be possible to detect some types of defects with this type
of equipment, most eddy current instruments are a bit more sophisticated. In the
following pages, a few of the more important aspects of eddy current
instrumentation will be discussed.
Resonant Circuits
The applet below represents an eddy current probe with a default resonant
frequency of about 1.0 kHz. An ideal probe might contain just the inductance, but
a realistic probe has some resistance and some capacitance. The applet initially
shows a single cycle of the 1.0 kHz current passing through the inductor.
Exercise 1: Using your mouse, adjust the resistance by sliding the slide bar.
Does the frequency change?
Exercise 2: Note that changing the inductance and/or the capacitance changes
the resonant frequency of this resonant circuit. Can you find several
combinations of capacitance and inductance that resonate at 1.0 kHz?
Bridges
The bridge circuit shown in the applet below is known as the Maxwell-Wien
bridge (often called the Maxwell bridge), and is used to measure unknown
inductances in terms of calibrated resistance and capacitance. Calibration-grade
inductors are more difficult to manufacture than capacitors of similar precision,
and so the use of a simple "symmetrical" inductance bridge is not always
practical. Because the phase shifts of inductors and capacitors are exactly
opposite each other, a capacitive impedance can balance out an inductive
impedance if they are located in opposite legs of a bridge, as they are here.
Unlike this straight Wien bridge, the balance of the Maxwell-Wien bridge is
independent of the source frequency. In some cases, this bridge can be made to
balance in the presence of mixed frequencies from the AC voltage source, the
limiting factor being the inductor's stability over a wide frequency range.
Exercise: Using the equations within the applet, calculate appropriate values for
C and R2 for a set of probe values. Then, using your calculated values, balance
the bridge. The oscilloscope trace representing current (brightest green) across
the top and bottom of the bridge should be minimized (straight line).
In the simplest implementation, the standard capacitor (C) and the resistor in
parallel with it are made variable, and both must be adjusted to achieve balance.
However, the bridge can be made to work if the capacitor is fixed (non-variable)
and more than one resistor is made variable (at least the resistor in parallel with
the capacitor, and one of the other two). However, in the latter configuration it
takes more trial-and-error adjustment to achieve balance as the different variable
resistors interact in balancing magnitude and phase.
You will recall that this can be graphically displayed using the impedance plane
diagram as seen above. Impedance also has an associated angle, called the phase
angle of the circuit, which can be calculated by the following equation.
The impedance plane diagram is a very useful way of displaying eddy current
data. As shown in the figure below, the strength of the eddy currents and the
magnetic permeability of the test material cause the eddy current signal on the
impedance plane to react in a variety of different ways.
If the eddy current circuit is balanced in air and then placed on a piece of
aluminum, the resistance component will increase (eddy currents are being
generated in the aluminum and this takes energy away from the coil, which
shows up as resistance) and the inductive reactance of the coil decreases (the
magnetic field created by the eddy currents opposes the coil's magnetic field and
the net effect is a weaker magnetic field to produce inductance). If a crack is
present in the material, fewer eddy currents will be able to form and the
resistance will go back down and the inductive reactance will go back up.
Changes in conductivity will cause the eddy current signal to change in a
different way.
When a probe is placed on a magnetic material such as steel, something different
happens. Just like with aluminum (conductive but not magnetic), eddy currents
form, taking energy away from the coil, which shows up as an increase in the
coils resistance. And, just like with the aluminum, the eddy currents generate
their own magnetic field that opposes the coils magnetic field. However, you will
note for the diagram that the reactance increases. This is because the magnetic
permeability of the steel concentrates the coil's magnetic field. This increase in
the magnetic field strength completely overshadows the magnetic field of the
eddy currents. The presence of a crack or a change in the conductivity will
produce a change in the eddy current signal similar to that seen with aluminum.
In the applet below, liftoff curves can be generated for several nonconductive
materials with various electrical conductivities. With the probe held away from
the metal surface, zero and clear the graph. Then slowly move the probe to the
surface of the material. Lift the probe back up, select a different material and
touch it back to the sample surface.
Experiment
Generate a family of liftoff curves for the different materials available in the
applet using a frequency of 10kHz. Note the relative position of each of the
curves. Repeat at 500kHz and 2MHz. (Note: it might be helpful to capture an
image of the complete set of curves for each frequency for comparison.)
1) Which frequency would be best if you needed to distinguish between two high
conductivity materials?
2) Which frequency would be best if you needed to distinguish between two low
conductivity materials?
The impedance calculations in the above applet are based on codes by Jack Blitz
from "Electrical and Magnetic Methods of Nondestructive Testing," 2nd ed.,
Chapman and Hill.
Analog meters such as the D'Arsonval design pictured in the applet below, must
"rectify" the AC into DC. This is most easily accomplished through the use of
devices called diodes. Without going into elaborate detail over how and why
diodes work as they do, remember that they each act like a one-way valve for
electrons to flow. They act as a conductor for one polarity and an insulator for
another. Arranged in a bridge, four diodes will serve to steer AC through the
meter movement in a constant direction.
An analog meter can easily measure just a few microamperes of current and is
well suited for use in balancing bridges.
Exercise: Using the equations within the applet, calculate appropriate values for
C and R2 for a set of probe values. Then balance the bridge using your calculated
values. The analog meter should swing close to the left end if its scale indicates
little or no current across the bridge. Across the bridge should be minimized
(straight line).
4 Probes / Coil Desing
Probes - Mode of Operation
Absolute Probes
Absolute coils can be used for flaw detection, conductivity measurements, liftoff
measurements and thickness measurements. They are widely used due to their
versatility. Since absolute probes are sensitive to things such as conductivity,
permeability liftoff and temperature, steps must be taken to minimize these
variables when they are not important to the inspection being performed. It is
very common for commercially available absolute probes to have a fixed "air
loaded" reference coil that compensates for ambient temperature variations.
Differential Probes
Reflection probes have two coils similar to a differential probe, but one coil is
used to excite the eddy currents and the other is used to sense changes in the test
material. Probes of this arrangement are often referred to as driver/pickup probes.
The advantage of reflection probes is that the driver and pickup coils can be
separately optimized for their intended purpose. The driver coil can be made so
as to produce a strong and uniform flux field in the vicinity of the pickup coil,
while the pickup coil can be made very small so that it will be sensitive to very
small defects.
Hybrid Probes
Probes - Configurations
As mentioned on the previous page, eddy current probes are classified by the
configuration and mode of operation of the test coils. The configuration of the
probe generally refers to the way the coil or coils are packaged to best "couple"
to the test area of interest. Some of the common classifications of probes based
on their configuration include surface probes, bolt hole probes, inside diameter
(ID) probes, and outside diameter (OD) probes.
Surface Probes
Pencil probes have a small surface coil that is encased in a long slender housing
to permit inspection in restricted spaces. They are available with a straight shaft
or with a bent shaft, which facilitates easier handling and use in applications such
as the inspection of small diameter bores. Pencil probes are prone to wobble due
to their small base and sleeves are sometimes used to provide a wider base.
Bolt hole probes are a special type of surface probe that is designed to be used
with a bolt hole scanner. They have a surface coil that is mounted inside a
housing that matches the diameter of the hole being inspected. The probe is
inserted in the hole and the scanner rotates the probe within the hole.
ID or Bobbin Probes
Probe Shielding
The most important feature in eddy current testing is the way in which the eddy
currents are induced and detected in the material under test. This depends on the
design of the probe. As discussed in the previous pages, probes can contain one
or more coils, a core and shielding. All have an important effect on the probe,
but the coil requires the most design consideration.
A coil consists of a length of wire wound in a helical manner around the length of
a former. The main purpose of the former is to provide a sufficient amount of
rigidity in the coil to prevent distortion. Formers used for coils with diameters
greater than a few millimeters (i.e. encircling and pancake coils), generally take
the form of tubes or rings made from dielectric materials. Small-diameter coils
are usually wound directly onto a solid former.
The region inside the former is called the core, which can consist of either a solid
material or just air. When the core is air or a nonconductive material, the probe
is often referred to as an air-core probe. Some coils are wound around a ferrite
core which concentrates the the coil's magnetic field into a smaller area. These
coils are referred to as "loaded" coils.
The wire used in an eddy current probe is typically made from copper or other
nonferrous metal to avoid magnetic hysteresis effects. The winding usually has
more than one layer so as to increase the value of inductance for a given length of
coil. The higher the inductance (L) of a coil, at a given frequency, the greater the
sensitivity of eddy current testing.
It is essential that the current through the coil is as low as possible. Too high a
current may produce:
The simplest type of probe is the single-coil probe, which is in widespread use.
The following applet may be used to calculate the effect of the inner and outer
diameters, length, number of turns and wire diameter of a simple probe design on
the probe's self inductance. Dimensional units are in millimeters.
A more precise value of L is given by:
In eddy current testing, the use of reference standards in setting up the equipment
is particularly important since signals are affected by many different variables
and slight changes in equipment setup can drastically alter the appearance of a
signal. As with most other NDT methods, the most useful information is obtained
when comparing the results from an unknown object to results from a similar
object with well characterized features and defects. In almost all cases, eddy
current inspection procedures require the equipment to be configured using
reference standards.
For crack detection, corrosion thinning and other material damage, reference
standards are used to setup the equipment to produce a recognizable signal or set
of signals from a defect or set of defects. In many cases, the appearance of a test
signal can be related to the appearance of a signal from a known defect on the
reference standard to estimate the size of a defect in the test component. Signals
that vary significantly from the responses produced by the reference standard
must be further investigated to the determine the source of the signal.
The reference standard should be of the same material as the test article. If this is
not possible or practical, it should be of material that has the same electrical
conductivity and magnetic permeability. Component features (material thickness,
geometry, etc.) should be the same in the reference standard as those in the test
region of interest. If the reference standard is the type with intentional defects,
these defects should be as representative of actual defects in the test component
as possible. The closer the reference standard is to the actual test component, the
better. However, since cracks and corrosion damage are often difficult and costly
to produce, artificial defects are commonly used. Narrow notches produced with
electron discharge machining (EDM) and saw cuts are commonly used to
represent cracks, and drilled holes are often used to simulate corrosion pitting.
Conductivity standards.
Flat plate discontinuity standards.
Flat plate metal thinning standards (step or tapered wedges).
Tube discontinuity standards.
Tube metal thinning standards.
Hole (with and without fastener) discontinuity standards
Signal Filtering
Filtering is applied to the received signal and, therefore, is not directly related to
the probe drive frequency. This is most easily understood when picturing a time
versus signal amplitude display. With this display mode, it is easy to see that the
signal shape is dependent on the time or duration that the probe coil is sensing
something. For example, if a surface probe is placed on the surface of conductor
and rocked back and forth, it will produce a wave like signal. When the probe is
rocked fast, the signal will have a higher frequency than when the probe is
rocked slowly back and forth.
The signal does not need a wavelike appearance to have frequency content and
most eddy current signals will be composed of a large number of frequencies.
Consider a probe that senses a notch for 1/60th of a second. In a period of one
second the probe could (in theory) go over the notch 60 times, resulting in the
notch signal having a frequency of 60 Hz. But, imposed on this same signal,
could be the signal resulting from probe wobble, electronic noise, a conductivity
shift and other factors which occur at different frequencies.
Filters Effects
Filter Settings
The main function of the LPF is to remove high frequency interference noise.
This noise can come from a variety of sources including the instrumentation
and/or the probe itself. The noise appears as an unstable dot that produces jagged
lines on the display as seen in the signal from a surface notch shown in the left
image below. Lowering the LPF frequency will remove more of the higher
frequencies from the signal and produce a cleaner signal as shown in the center
image below. When using a LPF, it should be set to the highest frequency that
produces a usable signal. To reduce noise in large surface or ring probes, it may
be necessary to use a very low LPF setting (down to 10Hz). The lower the LPF
setting, the slower the scanning speed must be and the more closely it must be
controlled. The image on the right below shows a signal that has been clipped
due to using a scan speed too fast for the selected HPF setting.
The HPF is used to eliminate low frequencies which are produced by slow
changes, such as conductivity shift within a material, varying distance to an edge
while scanning parallel to it, or out-of-round holes in fastener hole inspection.
The HPF is useful when performing automated or semiautomatic scans to keep
the signal from wandering too far from the null (balance) point. The most
common application for the HPF is the inspection of fastener holes using a
rotating scanner. As the scanner rotates at a constant RPM, the HPF can be
adjusted to achieve the desired effect.
Use of the HPF when scanning manually is not recommended, as keeping a
constant scanning speed is difficult, and the signal deforms and amplitude
decreases. The size of a signal decreases as the scan speed decreases and a flaw
indication can be eliminated completely if the scan is not done with sufficient
speed. In the images below, it can be seen that a typical response from a surface
notch in aluminum without HPF (left image) looks considerably different when
the HPF is activated (right image). With the HPF, looping signals with a positive
and similar negative deflection are produced on the impedance plane.
The use of a minimal HPF setting (1 or 2 Hz) may be used when manually
scanning, provided the operator can largely control the scan speed and becomes
familiar with the indication signal changes as scan speed is varied slightly. An
good example of such an application would be the manual scan of the radius of a
wheel that is rotated by hand, but the speed of rotation can be kept relatively
constant.
6 Applications
Surface Breaking Cracks
Of course, factors such as the type of material, surface finish and condition of the
material, the design of the probe, and many other factors can affect the sensitivity
of the inspection. Successful detection of surface breaking and near surface
cracks requires:
The applet below depicts a simple eddy current probe near the surface of a
calibration specimen. Move the probe over the surface of the specimen and
compare the signal responses from a surface breaking crack with the signals from
the calibration notches. The inspection can be made at a couple of different
frequencies to get a feel for the effect that frequency has on sensitivity in this
application.
Surface Crack Detection Using Sliding Probes
Sliding probes have been named so because they move over fasteners in a sliding
motion. There are two types of sliding probes, fixed and adjustable, which are
usually operated in the reflection mode. This means that the eddy currents are
induced by the driver coil and detected by a separate receiving coil.
Sliding probes are one of the fastest methods to inspect large numbers of fastener
holes. They are capable of detecting surface and subsurface discontinuities, but
they can only detect defects in one direction. The probes are marked with a
detection line to indicate the direction of inspection. In order to make a complete
inspection there must be two scans that are orthogonal (90 degrees) to each other.
Probe Types
Adjustable probes, as the name implies, are adjustable with the use of spacers,
which will change the penetration capabilities. The spacer thickness between the
coils is normally adjusted for the best detection. For tangential scans or 90 degree
scanning with an offset from the center, a thinner spacer is often used.
The spacer thickness range can vary from 0 (no
spacer) for inspections close to the surface and
small fastener heads to a maximum of about 0.3
inch for deep penetration with large heads in the
bigger probe types. A wider spacer will give more
tolerance to probe deviation as the sensitive area
becomes wider but the instrument will require more
gain. Sliding probes usually penetrate thicker
materials compared to the donut probes.
Reference Standards
Reference standards used should be manufactured from the same material type,
alloy, material thickness, and chemical composition that will be found on the
aircraft component to be inspected. Sizes and tolerances of flaws introduced in
the standards are usually regulated by inspection specifications.
Inspection Variables
Scan Patterns
A typical scan is centralized over the fastener head and moves along the axis of
the fastener holes. This scan is generally used to detect cracks positioned along
the axis of the fastener holes. For detecting cracks located transverse or 90
degrees from the axis of the fastener holes, a scan that is 90 degrees from the axis
of the fastener holes is recommended.
Signal Interpretation
When the probe moves over a fastener hole with a crack, the indication changes
and typically will create a larger vertical movement. The vertical amplitude of
the loop depends on the crack length, with longer cracks giving higher
indications.
If the crack is in the far side of the fastener, as the probe moves over it, the dot
will follow the fastener line first but will move upwards (clockwise) as it goes
over the crack. If the crack is in the near side, it will be found first and the dot
will move along the crack level before coming down to the fastener level.
If two cracks on opposite sides of the fastener hole are present, the dot will move
upwards to the height by the first crack length and then come back to the fastener
line and balance point. If the second crack is longer than the first one, the dot will
move even higher and complete the loop (clockwise) before going down to the
balance point. See Figures 3 and 4.
If the probe deviates from the center line, the crack indication will move along
the loop that we saw in Figure 5 and is now present in Figure 6. The crack
indication is at "a" when the probe is centralized and moves toward "b" as it
deviates in one direction, or "c" as it deviates in the opposite direction. Point "b"
gives an important indication even if it loses a small amount of amplitude it has
gained in phase, giving a better separation angle. This is because we deviated to
the side where the crack is located.
Crack Angle Deviation
A reduction in the crack indication occurs when the crack is at an angle to the
probe scan direction. This happens if the crack is not completely at 90 degrees to
the normal probe scan or changes direction as it grows. Both the fixed and
adjustable sliding probes are capable of detecting cracks up to about 30 degrees
off angle. See Figures 7 and 8.
Electrical Contact
When inspecting fasteners that have just been installed or reference standards
that have intimate contact with the aluminum skin plate, it is not unusual to
obtain a smaller than normal indication. In some extreme cases, the fastener
indication may disappear almost completely. This is due to the good electrical
contact between the fastener and the skin. This condition allows the eddy
currents to circulate without encountering a boundary, and therefore, no obstacle
or barrier. Because of this effect, it is recommended to paint the holes before
fastener installation.
Tube Inspection
A tube inspection using a bobbin probe is simulated below. Click the "null"
button and then drag either the absolute or the differential probe through the tube.
Note the different signal responses provided by the two probes. Also note that the
absolute probe is much more sensitive to dings and the build up of magnetite on
the outside of the tube than the differential probe is.
Conductivity Measurements
The technique usually involves nulling an absolute probe in air and placing the
probe in contact with the sample surface. For nonmagnetic materials, the change
in impedance of the coil can be correlated directly to the conductivity of the
material. The technique can be used to easily sort magnetic materials from
nonmagnetic materials but it is difficult to separate the conductivity effects from
the magnetic permeability effects, so conductivity measurements are limited to
nonmagnetic materials. It is important to control factors that can affect the results
such as the inspection temperature and the part geometry. Conductivity changes
with temperature so measurements should be made at a constant temperature and
adjustments made for temperature variations when necessary. The thickness of
the specimen should generally be greater than three standard depths of
penetration. This is so the eddy currents at the back surface of the sample are
sufficiently weaker than the variations in the specimen thickness that are not seen
in the measurements.
When the probe is brought near a conductive but nonmagnetic material, the coil's
inductive reactance goes down since the magnetic field from the eddy currents
opposes the magnetic field of the coil. The resistance in the coil increases since it
takes some of the coil's energy to generate the eddy currents and this appears as
additional resistance in the circuit. As the conductivity of the materials being
tested increases, the resistance losses will be less and the inductive reactance
changes will be greater. Therefore, the signals will be come more vertical as the
conductivity increases, as shown in the image above.
To sort materials using an impedance plane device, the signal from the unknown
sample must be compared to a signal from a variety of reference
standards. However, there are devices available that can be calibrated to produce
a value for electrical conductivity which can then be compared to published
values of electrical conductivity in MS/m or percent IACS (International
Annealed Copper Standard). Please be aware that the conductivity of a particular
material can vary significantly with slight variations in the chemical composition
and, thus, a conductivity range is generally provided for a material. The
conductivity range for one material may overlap with the range of a second
material of interest, so conductivity alone can not always be used to sort
materials. The electrical conductivity values for a variety of materials can be
found in the material properties reference tables.
The following applet is based on codes for nonferrous materials written by Back
Blitz from his book, "Electrical and Magnetic Methods of Nondestructive
Testing", 2nd ed., Chapman & Hill (1997). The applet demonstrates how an
impedance plane eddy current instrument can be used for the sorting of materials.
Conductivity Measurements
for the Verification of Heat
Treatment
Since solution heat-treated and aged materials are stronger, components can be
made using less material. A lighter or more compact design is often of great
importance to the designer and well worth the cost of the heat treating process.
However, think of the consequences that could arise if a component that was
supposed to be solution heat-treated and aged somehow left the manufacturing
facility and was put into service unheat-treated or annealed. This is a real
possibility since heat-treated aluminum parts look exactly like unheat-treated
parts. Consider 2024 aluminum as an example. Select tensile properties and its
electrical conductivity for various heat treatment conditions are given in the
following table.
Electrical
Heat Treatment Condition Ultimate Strength Yield Strength
Conductivity
Annealed (O) 26 ksi (180 MPa) 11 ksi (75 MPa) 50 % IACS
Solution Heat Treated and
64 ksi (440 MPa) 42 ksi (290 MPa) 30 % IACS
Naturally Aged (T42)
Solution Heat Treated,
Coldworked and 70 ksi (485 MPa) 66 ksi (455 MPa) 38 % IACS
Artificially Aged (T861)
It can be seen that the yield strength for the material is 42 kilopounds/square inch
(ksi) (290 MPa) in the solution heat-treated and naturally aged condition (T42
condition). The yield strength can be increased to 66 ksi (455 MPa) when
coldworked and artificially aged (T861 condition). But in the annealed condition,
the yield strength is reduced to 11 ksi (75 MPa). If an annealed part were
accidentally used where a part in the T42 or T861 was intended, it would likely
fail prematurely. However, a quick check of the conductivity using an eddy
current instrument of all parts prior to shipping would prevent this from
occurring.
Thickness Measurements of
Thin Material
Eddy current techniques are used to measure the thickness of hot sheet, strip and
foil in rolling mills, and to measure the amount of metal thinning that has
occurred over time due to corrosion on fuselage skins of aircraft. On the
impedance plane, thickness variations exhibit the same type of eddy current
signal response as a subsurface defect, except that the signal represents a void of
infinite size and depth. The phase rotation pattern is the same, but the signal
amplitude is greater. In the applet, the lift-off curves for different areas of the
taper wedge can be produced by nulling the probe in air and touching it to the
surface at various locations of the tapered wedge.
If a line is drawn between the end points of the lift-off curves, a comma shaped
curve is produced. As illustrated in the second applet, this comma shaped curve
is the path that is traced on the screen when the probe is scanned down the length
of the tapered wedge so that the entire range of thickness values are measured.
When making this measurement, it is important to keep in mind that the depth of
penetration of the eddy currents must cover the entire range of thicknesses being
measured. Typically, a frequency is selected that produces about one standard
depth of penetration at the maximum thickness. Unfortunately, at lower
frequencies, which are often needed to get the necessary penetration, the probe
impedance is more sensitive to changes in electrical conductivity. Thus, the
effects of electrical conductivity cannot be phased out and it is important to
verify that any variations of conductivity over the region of interest are at a
sufficiently low level.
Dimensions of cylindrical tubes and rods can be measured with either OD coils
or internal axial coils, whichever is appropriate. The relationship between change
in impedance and change in diameter is fairly constant, except at very low
frequencies. However, the advantages of operating at a higher normalized
frequency are twofold. First, the contribution of any conductivity change to the
impedance of the coil becomes less important and it can easily be phased out.
Second, there is an increase in measurement sensitivity resulting from the higher
value of the inductive component of the impedance. Because of the large phase
difference between the impedance vectors corresponding to changes in fill-factor
and conductivity (and defect size), simultaneous testing for dimensions,
conductivity, and defects can be carried out.
Fairly precise measurements can be made with a standard eddy current flaw
detector and a calibration specimen. The probe is nulled in air and the direction
of the lift-off signal is established. The location of the signal is marked on the
screen as the probe is placed on the calibration specimen in areas of decreasing
coating thickness. When the probe is placed on the test surface, the position of
the signal will move from the air null position to a point that can be correlated to
the calibration markings.
Specialized eddy current coating thickness detectors are also available and are
often pocket-sized with the probe resembling a small pencil. They are usually
operated by a small battery and provide a digital read-out in the appropriate units.
Calibration adjustments, some of which are laid down by standards such as BS
EN 2360 (1995) and ASTM B 244 and E 376, may be assisted by the use of an
inbuilt microprocessor.
7 Advanced Techniques
Scanning
Eddy current data can be collected using automated scanning systems to improve
the quality of the measurements and to construct images of scanned areas. The
most common type of scanning is line scanning where an automated system is
used to push the probe at a fixed speed. Line scan systems are often used when
performing tube inspections or aircraft engine blade slot inspections, where
scanning in one dimension is needed. The data is usually presented as a strip
chart recording. The advantage of using a linear scanning system is that the probe
is moved at a constant speed, so indications on the strip chart can be correlated to
a position on the part being scanned. As with all automated scanning systems,
operator variables, such as wobble of the probe, are reduced.
Several of these factors are often present simultaneously. In the simple case
where interest is confined to detecting defects or other abrupt changes in
geometry, a differential probe can be used to eliminate unwanted factors,
providing they vary in a gradual manner. For example, variations in electrical
conductivity and tube thinning affect both coils of a differential probe
simultaneously. However, if unwanted parameters that occur abruptly are
affecting the measurements, they can sometimes be negated by mixing signals
collected at several frequencies.
Swept Frequency
Swept frequency eddy current techniques involve collecting eddy current data at
a wide range of frequencies. This usually involves the use of a specialized piece
of equipment such as an impedance analyzer, which can be configured to
automatically make measurements over a range of frequencies. The swept-
frequency technique can be implemented with commercial equipment but it is a
difficult and time-consuming measurement. The advantage of a swept frequency
measurement is that depth information can be obtained since eddy current depth
of penetration varies as a function of frequency.
Swept frequency measurements are useful in applications such as measuring the
thickness of conductive coatings on conductive base metal, differentiating
between flaws in surface coatings and flaws in the base metal and differentiating
between flaws in various layers of built-up structure. An example application
would be the lap splice of a commercial aircraft. Swept frequency measurements
would make it possible to tell if cracking was occurring on the outer skin, the
inner skin or a double layer. Below is an example of the type of data that can be
obtained from swept-frequency measurements.
It can be seen that in the etched condition, the material labeled "good" exhibits a
much different signal response than the material labeled "bad." It can also be
seen that a frequency of around 2.2 MHz provides the largest separation in the
curves. Therefore, this frequency should be used if a single frequency is used to
sort the parts made from the two metals.
Pulsed Eddy Current Inspection
To improve the strength and ease interpretation of the signal, a reference signal is
usually collected, to which all other signals are compared (just like nulling the
probe in conventional eddy current inspection). Flaws, conductivity, and
dimensional changes produce a change in the signal and a difference between the
reference signal and the measurement signal that is displayed. The distance of the
flaw and other features relative to the probe will cause the signal to shift in time.
Therefore, time gating techniques (like in ultrasonic inspection) can be used to
gain information about the depth of a feature of interest.
The use of pulsed eddy currents has long been considered for testing metals
(Libby, 1971) and it has been applied to operations in specialized areas, such as
in the nuclear energy industry, where testing equipment is often constructed to
order. However, significant progress in this direction has taken place only
recently after appropriate advances in technology (Krzwosz et al. 1985; Sather,
1981; Waidelich, 1981; Wittig and Thomas 1981), but at the time of writing,
commercial equipment was not yet available. The method has the potential
advantages of greater penetration, the ability to locate discontinuities from time-
of-flight determinations, and a ready means of multi-frequency measurement. At
present, it does not generally have the precision of the conventional methods. The
apparatus is somewhat complicated in design and not readily usable by the
average operator who is experienced with the conventional eddy current
equipment. Its main successes are in the testing of thin metal tubes and sheets, as
well as metal cladding for measuring thickness and for the location and sizing of
internal defects.
When comparing the pulsed method with the conventional eddy current
technique, the conventional technique must be regarded as a continuous wave
method for which propagation takes place at a single frequency or, more
correctly, over a very narrow frequency bandwidth. With pulse methods, the
frequencies are excited over a wide band, the extent of which varies inversely
with the pulse length; this allows multi-frequency operation. As found with
ultrasonic testing, the total amount of energy dissipated within a given period of
time is considerably less for pulsed waves than for continuous waves having the
same intensity. For example, with pulses containing only one or two wavelengths
and generated 1000 times per second, the energy produced is only about 0.002 of
that for continuous waves having the same amplitude. Thus, considerably higher
input voltages can be applied to the exciting coil for pulsed operation than for
continuous wave operation.
The currents are detected by a receiving probe located either adjacent to or on the
opposite side of the metal sample from the exciting probe when access is
possible. The range of propagated frequencies depends on the logarithmic
decrement of the exciting circuit, and because the speed of the waves is a
function of frequency, dispersion takes place and the pulse changes in shape as it
progresses through the metal. As one would expect, the height of the peak and its
time delay can be related to the thickness of the metal. Waidelich reports a
maximum penetration of 90 mm for aluminum sheet and 10 mm for steel. For 6
mm thick sheets, the peak value of the received pulse voltage was 13 V for
aluminum but only 20 mV for steel. Krzwosz et al. (1985) has shown how pulses
that result from the presence of internal simulated defects produce broadening
with an increase in depth.
The frequency content of the pulses depends on their lengths, and in the extreme,
contains continuous spectra ranging from less than 100 Hz to 1 or 2 kHz. By
performing a Fourier transformation, the pulse obtained by the receiving probe
can be displayed in the form of the variation of amplitude (or phase) with
frequency. By sampling different delay times within a pulse, different parts of the
spectrum can be evaluated (Sather, 1981). If both amplitude and phase are
measured, two parameters (i.e. presence of defects, variations in tube thickness,
and changes in fill-factor or liftoff) can be evaluated for each frequency selected
in the same way as with the multi-frequency method, although, at present, with a
lower degree of precision.
Dodd et al.(1988) has designed and developed a pulsed magnetic saturation
method for the eddy current testing of ferromagnetic metals. The DC field pulses
are generated by passing a high-current pulse through an electromagnet so as to
produce saturation in the metal object; the pulse length is made equal to the
thickness of the object, thus ensuring complete eddy current penetration where
feasible. The DC pulse, on the order of 1 ms duration, simultaneously produces
an eddy current pulse, which is detected by a probe; the output of the probe is
characteristic of the material being tested.
This technique has the advantage of producing high magnetic peak powers with
low average powers, thus keeping any heating of the test sample down to an
acceptable level. It has been applied successfully to the internal testing of the
walls of steel steam generator tubes, and tubes of diameter 10.9 mm and wall
thickness 5 mm have been examined with peak powers of 500 kW. Small defects
close to the external surfaces can be detected, and by taking advantage of the
multi-frequency properties of pulsed eddy currents, their indications can be
resolved from those that originate from other characteristics of the tubes.
More recent work on the use of pulsed eddy currents has been reported by Gibbs
and Campbell (1991), who inspected cracks under fasteners in aluminum aircraft
structures. Here, a Hall element was used as a receiver. Radial position,
approximate depth, and relative size of defects hidden under fastener heads could
be determined in countersunk areas for defect depths of up to 7 mm for
nonferrous fasteners and 14 mm for ferrous fasteners.
Eddy current testing for external defects in tubes where external access is not
possible (e.g. buried pipelines), is conducted using internal probes. When testing
thick-walled ferromagnetic metal pipes with conventional internal probes, very
low frequencies (e.g. 30 Hz for a steel pipe 10 mm thick) are necessary to
achieve the through-penetration of the eddy currents.
This situation produces a very low sensitivity of flaw detection. The degree of
penetration can, in principle, be increased by the application of a saturation
magnetic field. However, because of the large volume of metal present, a large
saturation unit carrying a heavy direct current may be required to produce an
adequate saturating field.
For more information on Remote Field Testing can be found in the Specialized
NDT Methodssection of this site.
Impedance Matching
Although the Maxwell inductance bridge forms the basis of most eddy current
instruments, there are several reasons why it cannot be used in its simplest form
(i.e. Hague, 1934), including the creation of stray capacitances, such as those
formed by the leads and leakages to earth. These unwanted impedances can be
eliminated by earthing devices and the addition of suitable impedances to
produce one or more wide-band frequency (i.e. low Q) resonance circuits.
Instruments having a wide frequency range (i.e. from 1 kHz to 2 MHz) may
possess around five of these bands to cover the range. The value of the
impedance of the probe is therefore an important consideration in achieving
proper matching and, as a result, it may be necessary to change the probe when
switching from one frequency band to another.
8 QUIZZES
Eddy Current Inspection Quizzes
These quizzes draw from the same database of questions and differ only in the
number of questions presented. Each time a quiz is opened, a new set of random
questions will be produced from the database. The Collaboration for NDE
Education does not record the names of individuals taking a quiz or the results of
a quiz.
20 Question ET Quiz
Conductivity standards
Tube discontinuity standards
Hole discontinuity standards
All of the above
High frequency
Variable frequency
Low frequency
None of the above
Roentgen's formulas
Maxwell's inductance bridge
Reciprocity
The Inverse Square Law
A absolute probe
A differential probe
A reflection probe
Both B and C
Material thickness
Material geometry
Material conductivity
All of the above
Both B and C
Remain unchanged
Increase because the material will concentrate the magnetic
field in and around the coil
Decrease because the material will concetrate the magnetic
field in and around the coil
Decrease because the magnetic field of the coil will be
weakened by the material
Conductivity standards
Tube discontinuity standards
Hole discontinuity standards
All of the above
35 Question ET Quiz
Probe shielding is used to:
1
Shape the eddy currents to the curvature of the part
Reduce the inductive coupling of the probe and part
Increase the probe impedance
Reduce the effects of nonrelevant features in close proximity
to the probe
Conductivity standards
Tube discontinuity standards
Hole discontinuity standards
All of the above
Pancake probes
Pencil probes
Encircling probes
Sliding probes
Both B and C
Increase
Decrease
Remain the same
Remain the same until the probe touches the material
1
2
3
4
Phase lag:
22
Increase with discontinuity depth
Decrease with discontinuity depth
Is the same as phase angle
Is the same as the angle separating the liftoff and flaw
signals on an impedance plane
Pencil probes
Surface probes
Bolt hole probes
Bobbin probes
Bobbin probes
Surface coils
Encircling coils
Pencil probes
Remain unchanged
Increase because the material will concentrate the magnetic
field in and around the coil
Decrease because the material will concetrate the magnetic
field in and around the coil
Decrease because the magnetic field of the coil will be
weakened by the material
Material thickness
Material geometry
Material conductivity
All of the above
Roentgen's formulas
Maxwell's inductance bridge
Reciprocity
The Inverse Square Law
50 Question ET Quiz
Phase lag:
8
Increase with discontinuity depth
Decrease with discontinuity depth
Is the same as phase angle
Is the same as the angle separating the liftoff and flaw
signals on an impedance plane
Roentgen's formulas
Maxwell's inductance bridge
Reciprocity
The Inverse Square Law
1
2
3
4
Pancake probes
Pencil probes
Encircling probes
Sliding probes
A absolute probe
A differential probe
A reflection probe
Both B and C
Couplant
Fluorescent particles
Reference standards
Non abrasive cleaners
Pencil probes
Surface probes
Bolt hole probes
Bobbin probes
Bobbin probes
Surface coils
Encircling coils
Pencil probes
Ferrite
Aluminum
Lead
Both A and B
Conductivity standards
Tube discontinuity standards
Hole discontinuity standards
All of the above
Both B and C
A reference standard
An identified defect
A crack which is � the rejection criteria
A crack twice the rejection criteria
High frequency
Variable frequency
Low frequency
None of the above
STANDARDS
http://www.ndt-ed.org/EducationResources/CommunityCollege/EddyCurrents/cc_ec_index.htm