SERVICE MANUAL

HCP
Humidity chambers
INCO
CO2 incubators
Contents
1. About this manual 4 10. Sterile filter 24
1.1 Purpose and target group 4
1.2 Related documents 4 11. Doors 25
1.3 Tools needed 4 11.1 Installing, removing and adjusting
1.4 Manufacturer address and custo- door 25
mer service 4 11.2 Adjusting the door 26
11.3 Removing and installing the door
2. Safety regulations 5 lock 27
11.4 Replacing seals 29
3. Maintenance plan 5 11.5 Door heating 30
4. Finding Errors and Error Mes- 11.6 Adjusting and removing inner
sages 6 door 31

5. Overview/Appliance assembly 12. Humidity control 32


8 12.1 Humidity cannot be increased 32
12.2 Humidity cannot be reduced 35
6. Removing covers 9 12.3 Replacing the control module 37
6.1 Removing the top panel 9 12.4 Replacing the humidity sensor 38
Re-attach the top panel 9 13.1 CO2 sensor 39
Put the rear panel back on 10 13.2 Auto-zero measuring gas pump 41
6.2 Removing the rear panel 10 13.3 Replacing solenoid valve 42
7. Control unit 11 13.4 CO2 tube connections 43
7.1 Overview 11 13.5 Replacing CO2 sterile filter 43
Replacement 43
7.2 Mainboard 13
7.3 Replacing main switch module 13. Replacing the INCO
with the push/turn control 15 steam generator pump 44
7.4 Replacing the controller 16 1.1 Required spare parts 44
7.5 Replacing the switched mode po- 1.2 Required tools 44
wer supply 17 1.3 Replacement 44
7.6 Replacing the interface module 18 1.4 Circuit diagram details 49
1.5 Positions of components in the
8. Chamber fan 19
control unit 50
8.1 Function test 19
8.2 Replacing fan power supply unit19 14. Circuit diagrams 52
8.3 Replacing fan 20
9. Temperature control 22
9.1 Pt100 temperature sensor 22
9.2 Temperature limiter 23

Service HCP/INCO | 03/2016 3


1. About this manual
1.1 Purpose and target group
These instructions describe service and repair work on humidity
chambers of the type HCP and CO2 incubators of the type INCO.
They are exclusively for use by electricians who are familiar with
the safety regulations of electrical engineering.

1.2 Related documents


Apart from this service guide, the information and regulations in
the user manual for the respective appliance must also be ob-
served for all service and repair work.
Please refer to the circuit diagrams from page 44 of these instruc-
tions for all work on the electrical system.

1.3 Tools needed


The following tools and equipment are required to perform the
work described in this guide:
► Torx screwdriver sizes 8 and 10
► Phillips screwdriver size 1
► Wire-cutting pliers
► Flat pliers to remove the connection cable
► Allen key 2 mm for door hinge,
1.5 mm for humidity sensor
► Multimeter with voltage and resistance measuring range
► Cable straps
► Adhesive tape to label cables and components

1.4 Manufacturer address and customer service


MEMMERT GmbH + Co. KG
PO Box 17 20
91107 Schwabach
Germany
Äußere Rittersbacherstr. 38
91126 Schwabach
Germany
Phone: +49 (0) 09122 / 925-0
Fax: +49 (0) 09122 / 14585
E-mail: sales@memmert.com
Internet: www.memmert.com
Customer Service:
Phone: +49 (0) 09122/925-128
and: +49 (0) 09122/925-126
E-mail: service@memmert.com
For service enquiries, please always specify the appliance number
on the nameplate.
© 2016 Memmert GmbH + Co. KG
Date 03/2016
We reserve the right to make changes

4 Service HCP/INCO | 03/2016


2. Safety regulations
Danger!
After removing covers, voltage-carrying parts
may be exposed if the mains plug has not
been disconnected. Any work that involves
opening the appliance may only be performed
by trained electricians who are familiar with
the safety regulations for electrical engineer-
ing. If you want to change parts you must
always first disconnect the mains plug.

Caution!
When dismantling the appliance you could be
injured on sharp (unprocessed) metal edges.
You should wear gloves.

Please also observe the safety regulations in


the operating instructions for the respective
appliance.

After all maintenance or repair work, the appliance must be sub-


ject to a function test. The appliance must not be left unattended
until it has reached the steady state.
If the interference or malfunction cannot be eliminated:
Contact the manufacturer.

3. Maintenance plan
Annually:
► Check and replace sterile filter if it is dirty
(see page 24)
Every two years:
► Grease door hinge with silicone grease
► Replace all sterile filters (see page 24)
► Replace humidity restriction pump (see page 36)

Service HCP/INCO | 03/2016 5


4. Finding Errors and Error Messages
Error descrip- Cause of error Rectifying errors see
tion
General faults
Controller dis- Miniature fuse 100 mA (200 Check fuse, page 13
play does not mA) faulty replace if necessary
light up Appliance fuse 15 A faulty Check fuse, page 13
replace if necessary
Mainboard faulty Replace mainboard page 14
Temperature Appliance "locked" by UserID Undo lock with UserID card Instruc-
cannot be card tions
adjusted/ for use
appliance Push/turn control faulty Replace main switch mod- page 15
can generally ule with push/turn control
not be operated
by the push/
turn control
Chamber fan Miniature fuse 50 mA on fan Replace the fuse page 19
does not turn power supply unit faulty
Fan power supply unit faulty Replace fan power supply page 19
unit
Error messages in the monitoring display
Caution icon Mechanical temperature lim- ► Switch off appliance
is permanently iter has switched off heating and leave to cool down page 23
lit up ► Check temperature
limiter
The appliance is only op-
erational again after it has
cooled down and the error
has been eliminated
Caution icon Monitoring controller has Increase temperature differ-
flashes switched off heating, since ence between monitoring page 22
alarm limit has been reached and working temperature.
Replace Pt100 temperature
sensor of monitoring con-
troller if necessary
Error message Pt100 temperature sensor of Replace Pt100 temperature page 22
E-3 monitoring controller faulty sensor of the monitoring
controller on slot J3

6 Service HCP/INCO | 03/2016


Error descrip- Cause of error Rectifying errors see
tion
Error message Pt100 temperature sensor of Replace Pt100 temperature page 22
E-3 alternates monitoring controller faulty. sensor of the monitoring
with display of Monitoring controller is work- controller on plug J3
temperature ing in emergency mode with
the temperature value of the
Pt100 temperature sensor
from the operating controller
Error messages in timer display
Error message Serious configuration error Replace controller page 16
E-0 (e.g. incorrect chamber type
or wrong temperature range)
Error message Checksum error (error when The error can be rectified
CONF saving setpoint values) by the controller after a
(is only dis- setpoint parameter has
played for again been saved. If the er-
approximately ror should continue occur-
10 sec. after ring, or cannot be rectified:
switching on) Replace controller
page
16
Error message Communication with main- Check connection lead; page 14
E-L1 board 1 interrupted replace motherboard
if necessary
Error messages in temperature display
Error message Mainboard actuator Triac Replace mainboard page 14
E-1 faulty
Error message Mainboard faulty Replace mainboard page 14
E-2
Error message Pt100 temperature sensor of Replace Pt100 temperature page 22
E-3 operating controller faulty sensor of the operating
controller on plug J4
Error message Pt100 temperature sensor of Replace Pt100 temperature page 22
E-3 alternates operating controller is faulty. sensor of the operating
with display of Operating controller work- controller on plug J4
temperature ing in emergency mode with
the temperature value of the
Pt100 temperature sensor
from the monitoring control-
ler
Error messages in the CO2 display (only for incubators of the INCO type)
Error message CO2 sensor faulty Replace CO2 measuring page 39
E-5 board and CO2 sensor

Error message Checksum error (error when Replace CO2 measuring page 39
CONF in CO2 saving setpoint values) board and CO2 sensor
display
Error messages in humidity display (only for corresponding model)
Error message Humidity sensor faulty Replace humidity sensor page 38
E-6

Service HCP/INCO | 03/2016 7


5. Overview/Appliance assembly
1

4 3

Fig. 1 Construction (shown in figure: Maximum fittings for INCO incubator; the actual fittings may vary,
depending on the appliance type)
1 Top panel (see page 9).
2 Rear panel (see page 10)
3 Right-side panel
4 Door (see page 25)
5 Inner door (see page 31)
6 Fan shaft (see page 19)
7 Left-side panel
8 Control unit (see page 11)

8 Service HCP/INCO | 03/2016


6. Removing covers
For some service work it is necessary to remove case panels (top
panel, rear panel). How these panels are removed is described
below. What panel(s) need(s) to be removed for the respective
service work is described in the relevant chapter.

6.1 Removing the top panel


Danger!
After removing the top panel, voltage-carrying
parts in the control unit may be exposed if the
mains plug has not been disconnected. Do not
reach inside the control unit. If you want to examine or
replace the components inside, disconnect the mains
plug first.
1. Undo the 4 screws to the left and right on the rear side of
the appliance holding down the top panel (1).
2. Slightly raise the panel and pull it out backwards (2).

Fig. 2 Removing the top panel

Re-attach the top Put on the top panel and push forwards into the slot. Insert the
panel screws with retaining washers into the openings in the rear panel
and screw down the top panel (1).
The retaining washers are necessary to create a safe PE conductor
connection with the top panel.

Service HCP/INCO | 03/2016 9


6.2 Removing the rear panel
1. Disconnect the mains plug.
2. Remove the top panel (see previous page).
3. Undo connection to PE conductor (see page 11/12).
4. Pull rear panel upwards out of the appliance. For the INCO
CO2 incubator, pull away the rear panel slightly from the ap-
pliance so that it is not caught by the ventilation pipe (Fig. 3).
Make sure that the rear panel does not jam when pulling out.

Fig. 3 Removing the rear panel


1 Ventilation pipe (only for CO2 INCO incubators)
2 Opening for ventilation pipe in the rear panel

Put the rear panel 1. Insert rear panel carefully from above into the guide rails and
back on push downwards until they slot into place. Make sure that
the rear panel does not jam and is not scratched by the
ventilation pipe that sticks out of the rear of the appliance
(Fig. 3).
2. Reconnect the PE conductor.

10 Service HCP/INCO | 03/2016


7. Control unit
7.1 Overview
(see also circuit diagrams from page 44).

The control unit can be reached after the top panel has been removed (see page 9).

1 2 3 4 5 6 7 8 9 10 11 12 13

Fig. 4 Control unit humidity chamber HCP


1 Connection for PE conductor
2 Pump for humidification (see page 35)
3 Mainboard (see page 13)
4 Fan
5 Controller (see page 16)
6 Control module for humidity (see page 37)
7 Chipcard reader
8 Power supply unit for chamber fan (see page 19)
9 Main switch (see page 15)
10 Sterile filter (see page 24)
11 Switched mode power supply (see page 17)
12 Humidity restriction pump (see page 36)
13 Interface module (USB, Ethernet) (see page 18)

Service HCP/INCO | 03/2016 11


1 2 34 5 6 7 8 9 10 11 12 13 14 15 16 17

Fig. 5 Control unit CO2 INCO incubator


1 Connection for PE conductor
2 Sterile filter for auto-zero measuring gas pump
3 Pump for humidification (only for incubators with humidification, see page 35)
4 Auto-zero measuring gas pump
5 Mainboard (see page 13)
6 Fan
7 Control module for humidity (see page 37)
8 PCB for CO2 sensor (see page 39)
9 Controller (see page 16)
10 Chipcard reader
11 Power supply unit for chamber fan (see page 19)
12 Main switch (see page 15)
13 Sterile filter (see page 24)
14 Switched mode power supply (see page 17)
15 Humidity restriction pump (see page 36)
16 Solenoid valve (see page 42)
17 Interface module (USB, Ethernet) (see page 18)

12 Service HCP/INCO | 03/2016


7.2 Mainboard
4

2 3

Fig. 6 Mainboard (see also control unit overview on pages 11 and 12)
1 Mainboard
2 Fuse for controller/display
3 Appliance fuse
4 Screws (4 units)

7.2.1 Check/replace fuses


The fuses for controller/display and appliance protection are lo-
cated on the mainboard (Fig. 6).
1. Disconnect the mains plug.
2. Remove the top panel (see page 9).
3. Measure the fuse resistance with a continuity tester. If the
fuse is in order the resistance should only be a few Ohms.
4. To replace the fuse, push an insulated flat screwdriver be-
neath the side of the fuse and lever it out of the mounting
bracket. Press the new fuse from above into the mounting
bracket.
► Appliance fuse: 15 A quick-blow 32 x 6.3 mm
(American)
► Fuse for controller/display: 100 mA slow-blowing 20 x 5 mm
200 mA slow-blowing 20 x 5 mm
at 115V
Spare part Variants Spare part no.
Appliance fuse 15 A E02387
100 mA/230 V E02497
Fuse for controller/display
200 mA/115 V E01007

Service HCP/INCO | 03/2016 13


7.2.2 Replace mainboard
1. Disconnect the mains plug.
2. Remove the top panel (see page 9).
3. Pull off all cables from the mainboard (Fig. 6). Label un-
marked cables with adhesive tape to avoid confusion when
reconnecting.
4. Undo 4 screws from the mainboard. Make sure that the
spacer bolts are not lost when doing this.
5. Remove the mainboard.
6. Insert a new mainboard and screw it tight with the spacer
bolts.
7. Reconnect the cable.
8. Put the top panel back on; plug in the mains plug.
9. Check that the appliance is working. If the fault persists: Con-
tact the manufacturer.
Spare part Variants Spare part number
230 V B04093
Mainboard
115 V B04101

14 Service HCP/INCO | 03/2016


7.3 Replacing main switch module with the push/turn control
(See also overview pictures on pages 11, 12 and 16)
1. Disconnect the mains plug.
2. Remove the top panel (see page 9).
3. Pull off the push/turn control (1) on the operating side for-
wards.
4. Remove the plugs (4-pin) (4) in the control unit from the con-
troller slot J2 (see Fig. 8).
5. Undo the mounting bolts on left and right (2).
6. Slightly raise the main switch module and pull backwards.
7. Place new main switch module into the control panel.
8. Attach the module with two screws (2).
9. Connect the plug (4) to the controller slot J2.
10. Attach the push/turn control (1) at the front.
11. Put the top panel back on; connect the mains plug.
12. Switch on appliance and check the function of the new main
switch module and the push/turn control. If the error has not
been rectified: Contact the manufacturer.

1 2

Fig. 7 Main switch module (see also control unit overview


on pages 11 and 12)
1 Push/turn control
2 Mounting bolts (left and right)
3 Main switch
4 Plug (controller slot J2)

Spare part Spare part number


Main switch module B04425

Service HCP/INCO | 03/2016 15


7.4 Replacing the controller
1. Disconnect the mains plug.
2. Remove the top panel (see page 9).
3. Pull off the push/turn control to the front (see page 15).
4. Undo three screws (Fig. 8, 3) and remove the main switch
module (4).
5. Cut through three (HCP) or four (INCO) cable ties (2) with
wire-cutting pliers and discard.
6. Pull off and label the plugs of the respective controller slots
(Jx).
7. Remove controller (1) to the rear.
8. Insert new controller and fasten controller and main switch
module with three screws (3).
9. Reconnect plug to the respective controller slots.
10. Push the push/turn control back on.
11. Put the top panel back on; connect the mains plug.
12. Subject the appliance to a function test. If it should still work
incorrectly, contact the manufacturer.

2 2 J13 2 J4 J3 J11 3 3 4 3 J2 J9 2 J8

Fig. 8 Replace controller (the example shows the CO2 INCO incubator controller, see also control unit over-
view on pages 11 and 12)
1 Controller
2 Cable straps
3 Screws
J(x) Controller slots

Spare part Variants Spare part number


RS232/USB/Ethernet B04373
Controller INCO
RS485 B04374
RS232/USB/Ethernet B04378
Controller HCP
RS485 B04379

16 Service HCP/INCO | 03/2016


7.5 Replacing the switched mode power supply
1. Disconnect the mains plug.
2. Remove the top panel (see page 9).
3. Disconnect all the plugs on the switched mode power supply
and label them if necessary (1).
4. Undo the screws (4 screws) at the corners (2) and remove the
switched mode power supply.
5. Insert a new switched mode power supply and screw tight.
6. Reconnect the plug to the switched mode power supply.
7. Re-attach the top panel.
8. Connect the mains plug and subject the appliance to a func-
tion test. If the fault should persist or if it works incorrectly,
contact the manufacturer.

2
2

Fig. 9 Switched mode power supply


1 Connections
2 Screw connection

Spare part Spare part number


Switched mode power supply B04372

Service HCP/INCO | 03/2016 17


7.6 Replacing the interface module
1. Disconnect the mains plug.
2. Remove the top panel (see page 9).
3. Pull off the plug on the USB interface module or Ethernet
module (depending on the appliance fittings) on the rear cas-
ing panel (1).
4. Undo 2 screws at top and bottom with which the module is
attached to the rear wall (2); remove module.
5. Insert a new module and screw tight.
6. Connect the plug.
7. Re-attach the top panel.

1 2

Fig. 10 Replacing the interface module


1 Plug
2 Screw connection

Spare part Spare part number


USB interface module B04118
Ethernet module B04121

18 Service HCP/INCO | 03/2016


8. Chamber fan
8.1 Function test
Visual test to see if the fan wheel
is turning. It is visible inside the
chamber through the ventilation
2
slits in the fan shaft on the rear
(Fig. 12).
Check the fan power supply 3
unit in the control unit (see also
the control unit overviews from 4
page 11):
1. Remove the top panel
(see page 9). 1
2. Check the miniature fuse on
the fan power supply unit
(50 mA slow-burning) (Fig. Fig. 11 Fan power supply unit in
11). control unit
1 Miniature fuse
3. Check the output voltages: 2 Screws
terminal 1: L 3 Terminal 6
terminal 2: N 4 Terminal 1
terminal 3: N
terminal 4: approx. 18 Volt/
DC
terminal 5: GND
If one of the voltages is missing, the fan power supply unit must
be replaced.

8.2 Replacing fan power supply unit


1. Disconnect the mains plug.
2. Pull off connection leads from the terminals on the fan
power supply unit (Fig. 11).
3. Undo three mounting bolts (2) and remove the power supply
unit.
4. Insert a new power supply unit and screw tight.
5. Connect connection leads to the power supply unit terminals.
Plug in mains plug and repeat voltage test (see above). In case
this does not rectify the problem: Contact the manufacturer.

Service HCP/INCO | 03/2016 19


8.3 Replacing fan
The chamber fan unit is always replaced in full because of the
shaft seal, i.e. with mounting plate, terminal and shaft feed-
through.
1. Disconnect the mains plug.
2. Remove the top panel (see page 9) and pull out the rear
panel (see page 10).
3. Inside the chamber (Fig. 12), undo the screws (2) of the fan
shaft (1) and remove the fan shaft (3); the fan wheel is now
accessible.

1 2

Fig. 12 Removing the fan shaft


1 Fan shaft
2 Screws
3 Taking off fan shaft

4. Undo the headless screw on the fan wheel (Fig. 13, 4) and
remove the fan wheel towards the front (5).
5. On the rear of the appliance, unplug the two connection
leads on the terminal block (10).
6. Undo four screws (9) with washer (7) and serrated washer (8)
from the fan mounting plate (6) and pull out fan module to
the rear (11).
7. Insert a new fan module and screw tight; connect leads.
8. Inside the chamber, attach the fan wheel to the shaft with
the headless screw.
9. Screw the fan shaft back on.
10. Plug in mains plug and check fan function. In case this does
not rectify the problem: Contact the manufacturer.

20 Service HCP/INCO | 03/2016


5

9
10

11

Fig. 13 Replacing the chamber fan


4 Headless screw
5 Remove fan wheel
6 Fan mounting plate
7 Disc
8 Toothed disc
9 Screw
10 Terminal block
11 Remove fan wheel

Spare part Variants Spare part number


Chamber fan IP54 24 V B03283
230 V B03284
Fan power supply unit
115 V B03286
108 E03816
Fan shaft
153/246 E04364

Service HCP/INCO | 03/2016 21


9. Temperature control
9.1 Pt100 temperature sensor
The appliance is equipped with two temperature sensors which
measure the temperature inside the chamber: one sensor for the
working controller and one sensor for the monitoring controller
(Fig. 14).
Replacement 1. Disconnect the mains plug.
2. Remove the top panel (see page 9) and pull out the rear
panel (see page 10).
3. Remove insulation material (1) at the top right of the rear
panel.
4. Sensor for working controller: Disconnect connection lead
(4) of the Pt100 sensor (2) in the control unit from the 4-pin
controller slot J4.
Sensor for monitoring controller:
Disconnect connection lead (4) of the Pt100 sensor (3) in the
control unit from the 4-pin controller slot J3.
5. Cut through cable tie (5) with wire-cutting pliers. Pull sensor
back out of rear panel (6) and thread upwards through the
floor of the control unit (7).
6. Thread new temperature sensor downwards from the control
unit and push into rear panel. Connect connection lead to
the corresponding controller slot (see above).
7. Re-attach the insulation material; push the rear panel back in
and mount the top panel. Connect the mains plug.
8. Subject the appliance to a function test. If the fault should
persist/if it works incorrectly, contact the manufacturer.
4
5

3
6

Fig. 14 Replacing a Pt100 temperature sensor


1 Insulation material
2 Sensor
3 Sensor
4 Cable connections to the controller in the control unit
5 Cable straps
6 Pull out Pt100 temperature sensor
7 Pull out Pt100 temperature sensor upwards into the control unit

Spare part Spare part number


Pt100 temperature sensor E04956

22 Service HCP/INCO | 03/2016


9.2 Temperature limiter
The temperature limiter in the rear panel switches the heating off
automatically in case of overtemperature.
Replacement 1. Disconnect the mains plug.
2. Remove the top panel (see page 9) and pull out the rear
panel (see page 10).
3. Remove insulation material in the vicinity of the temperature
limiter (top left of the rear panel, see Fig. 15).
4. Pull off connection leads (1) of the temperature limiter from
the mainboard (slots S25 and S26) in the control unit.
5. Pull out connection leads downwards through the raised
floor (2).
6. Undo the mounting bolts (3) and remove the retaining brack-
et of the temperature limiter (4); remove temperature limiter.
7. Place on a new temperature limiter and attach with bracket.
8. Pull connection lead up through the control unit floor into
the control unit and connect to the mainboard.
9. Re-attach the insulation material; push the rear panel back in
and mount the top panel. Connect the mains plug.
10. Subject the appliance to a function test. If the fault should
persist/if it works incorrectly, contact the manufacturer.

2
3

4 3

Fig. 15 Replacing the temperature limiter


1 Undo the connection leads in the control unit
2 Pull connection leads downwards and out through the control panel floor
(2).
3 Remove screws
4 Remove mounting bracket and take out temperature limiter

Spare part Spare part number


Temperature limiter E05123

Service HCP/INCO | 03/2016 23


10. Sterile filter
Depending on the model (humidification and/or CO2 fumigation),
the appliance is fitted with up to four sterile filters which clean
the CO2 or the fresh air before it is brought into the chamber (Fig.
16):
Filter 1: CO2 IN1 + IN2
Filter 2: Humidity limit control
Filter 3: Humidity limit control
Filter 4: Auto-zero measuring gas pump (see also page 41)
Replacement 1. Disconnect the mains plug.
2. Remove the top panel (see page 9).
3. Pull off tubes from both sides of the sterile filter (1).
4. Take sterile filter out of holder (3).
5. Place new sterile filter into holder.
6. Connect tubes to both sides of the new filter.
7. Re-attach the top panel.

Fig. 16 Replace sterile filters


1 Pull off tubes from both sides of sterile filter
2 Sterile filter
3 Pull sterile filter upwards out of holder

Spare part Spare part number


Sterile filter E03401

24 Service HCP/INCO | 03/2016


11. Doors
11.1 Installing, removing and adjusting door
11.1.1 Removing the door
1. Swivel door by 180 degrees and undo screws (4) at the two
eccentric sockets (1) at the top and bottom.
2. Pull out door (5) horizontally in the direction of the arrow.

11.1.2 Installing the door


1. Grease the moving parts of the door hinges with thin silicone
grease (see also operating instructions "Maintenance").
2. Inset a hard paper disk on the lower door hinge between the
eccentric tappet (3) and the door (5).
3. Insert the door. When doing this, make sure that the adjuster
cam of the eccentric tappet (3) is pointing forwards (1a).
4. Put screws back in (4) and tighten up.

1b
1a
1

Fig. 17 Installing/removing door


1 Eccentric socket
1a Eccentric tappet moved forwards: factory settings for door
1b Eccentric tappet moved to the right: Doors are also adjusted to the right
2 Headless screw
3 Eccentric tappet
4 Screws
5 Door

Service HCP/INCO | 03/2016 25


11.2 Adjusting the door
1. Undo mounting bolts (2) at the top and bottom of door.
2. The top section of the door hinge (1) can now be moved
slightly in the direction of the arrow.
3. After undoing the headless screw (5), the door can be adjust-
ed by turning the eccentric tappet (3) with a screwdriver.
The headless screw (5) is fixed with lock washer varnish. It can
be undone with a jolting motion using a 2 mm Allen key.
4. Apply lock washer varnish to the headless screw (5) and
screw the headless screw back tight.
5. Tighten the mounting bolts (2) again.

1
2

3
4
5

Fig. 18 Adjusting the door


1 Upper section of the door hinge
2 Mounting bolts
3 Eccentric tappet
4 Eccentric socket
5 Headless screw

The locking plate (7) can also be adjusted in


the direction of the arrows after undoing the
6 screws. Afterwards, screw the locking plate
7 back firmly.

Fig. 19 Adjusting the locking plate


6 Screw
7 Locking plate

26 Service HCP/INCO | 03/2016


11.3 Removing and installing the door lock
1. Undo screws (6).
2. Remove the lock cover (5).
3. Screw off the door panel (2) and lift up.
4. Pull off the door-handle lid (1).
5. Pull off the circlip (2) sideways and remove the door-handle
(3).
6. Undo the guide screw (4) remove the circlips (10) and coil
springs (12).
7. Pull out the axles (11) sideways and
push the locking pins (7) outwards.
8. Remove the door lock mechanism (6, 8, 9).

6 5

Fig. 20 Door lock


1 Door
2 Door panel
3 Door seal
4 Locking mechanism
5 Lock cover
6 Screws

Service HCP/INCO | 03/2016 27


1 2 3 4 5 6 7

8 9 10 11 12 13 14 15

Fig. 21 Locking mechanism


1 Door-handle cover
2 Circlip
3 Door-handle
4 Guide screw
5 Sleeve
6 Locking bracket
7 Locking pins
8 Locking bolt
9 Transmission pin
10 Disc for door handle
11 Flat-head screw
12 Locking arm (straight)
13 Locking arm (angled)
14 Axles
15 Circlips (small)

Spare part Model Spare part number


108 B03299
Door 153 B04052
246 B03643
Door eccentric tappet all B03749
Door lock, complete, without locking pins all B04037
108 E02993
Locking pins
153/246 E03287
Locking plate all B02720

28 Service HCP/INCO | 03/2016


11.4 Replacing seals
11.4.1 Replacing door seal
1. Remove the locking cover (2) (see page 27).
2. Lift up door panel (4) and remove old seal (3). On no account
should the screw mounting of the door panel be undone,
since parts inside the door are held together by this.
3. Cut out the new door seal at the ends using the old one as a
template and push onto the door panel.
4. Insert the door panel and press on.
5. Re-attach the lock cover.

1 2 3 4 3

Fig. 22 Replacing the door seal (cross-section)


1 Screw
2 Lock cover
3 Door seal
4 Door panel

11.4.2 Replacing the chamber seal and the cage seal


The appliance has two seals:
► Chamber seal
► Cage seal
The seal can be pulled out after the door is opened and the new
seal be pushed into place.
Push in the pre-cut length of seal into the space, without stretch-
ing it, and cut off the end if necessary.

1
3
2

Fig. 23 Replacing the chamber and cage seals


1 Chamber seal
2 Cage seal
3 Seal cross-section

Service HCP/INCO | 03/2016 29


Spare part Model Spare part number
108 Z00702
Chamber seal 153 Z00722
246 Z00723
108 Z00716
Cage seal 153 Z00800
246 Z00801
108 Z00707
Door seal 153 Z00721
246 Z00708

11.5 Door heating


The door heating is operated with safety extra-low voltage.

11.5.1 Function test through resistance measurement


The correct function of the door heating can be tested with a
resistance measurement. The measurement should be taken at
the upper and lower spring contact (Fig. 24). The resistance must
be approx. 12 Ω.

11.5.2 Contact
Adjust door (see page 26), in case the spring contact (Fig. 24) in
the door does not touch the door contact surface (Fig. 25) when
the door is closed.
The spring contacts can be pulled out of the mounting with flat
pliers and replaced.

Fig. 24 Spring contact in door Fig. 25 Door contact surface

Spare part Spare part number


Spring contact E02743

30 Service HCP/INCO | 03/2016


11.6 Adjusting and removing inner door
11.6.1 Adjusting inner door
1. Loosen screws (1); do not remove screws completely, as oth-
erwise the mounting block (inside) will fall out.
2. Check the screw (2) in the hinge and tighten if necessary.
3. Adjust the inner door (4) and tighten screws (1) again.
If a mounting block has fallen out, you can reach it by pulling out
the side section and then mount it again.

11.6.2 Removing the inner door


Undo the screw (2) (counter with open-end wrench), tilt glass
door forwards and lift out from lower hinge.

1
2
3

Fig. 26 Adjusting/removing inner door


1 Screws
2 Screw
3 Inner door hinge
4 Inner door
5 Outer door (open)

Spare part Variants Spare part number


108 B03300
Glass inner door (incl. hinge) 153 B03670
246 B03644
top B02902
Hinge
bottom B02903

Service HCP/INCO | 03/2016 31


12. Humidity control
(only for corresponding model)
If the humidity control of the chamber does not work properly
(see also operating instructions), this may have several different
causes. It is described below how the fault can be determined
and what needs to be done to remedy this.
First check whether an error message of the control module for
humidification is displayed at the front of the appliance (see op-
erating instructions). If this is the case, the control module must
be replaced (see page 37).
If this is not the case, check whether an error message from the
humidity sensor is displayed. In this case the humidity sensor
must be replaced (see page 38).
If no error message is displayed, proceed as follows:

12.1 Humidity cannot be increased


If humidification (increase of humidity in the chamber) is not
working properly, this can be seen by the fact that the humidity
setpoint display on the control panel does not change if a higher
humidity is set (see operating instructions). Reducing the humid-
ity works, however.
Proceed as follows to determine the cause:
1. Check whether there is water in the external tank/canister (1);
if not, fill up with distilled water.
2. Disconnect the mains plug.
3. Remove the top panel (see page 9) and rear panel (see page
10).
4. Check whether the water tubes on the appliance and on the
steam generator (3) on the rear side of the appliance are con-
nected properly (2). If not, make sure that they are fastened
tight.

2
3

Fig. 27 Checking the tube connections


1 Water tank/canister
2 Tube connections
3 Steam generator

32 Service HCP/INCO | 03/2016


If the fault is not down to the water supply, continue as follows:
1. Check that all plugs in the control unit are connected to the
control module for humidification (1) and for the pump (2).
2. Check that the fuse for the control module is in order. If not:
Remove the mains plug and replace the fuse.
3. Check that the pump is running. If not: Remove the mains
plug and replace the pump (see page 35).
4. If the error has still not been found, then the steam genera-
tor at the back of the appliance is faulty. In this case it must
be replaced (see next chapter).

Fig. 28 Control unit


1 Control module for humidification
2 Pump

Service HCP/INCO | 03/2016 33


12.1.1 Replacing steam generator
1. Disconnect the mains plug.
2. Remove the top panel (see page 9) and pull out the rear
panel (see page 10).
3. Remove the fan shaft in the chamber of the appliance (see
page 20).
4. Loosen tube (1) on the steam generator (2) (Fig. 29).
5. Undo the plug connector of the steam generator (3) in the
control unit on the top of the appliance.
6. Unscrew the steam generator (4 screws) (4).
7. Remove the steam generator.
8. Put a new steam generator into place. When doing this,
make sure that the copper pipe with the attached tube on
the rear of the steam generator fits through the opening into
the interior of the appliance (7). Screw the steam generator
tight.
9. In the interior of the appliance, cut back the tube of the
steam generator coming out of the rear to about 3 mm in
length (7). Re-attach the fan shaft.
10. Connect the plug connector of the steam generator in the
control unit on the top of the appliance.
11. Connect the water tube on the right of the steam generator
(1).
12. Reinsert the rear panel and attach the top panel.
13. Reconnect the mains plug.
14. Check that the humidification is now working properly. If this
is not the case, contact the manufacturer.
1 7
2

4
3 mm

5 6
Fig. 29 Replacing steam generator
1 Tube connection 5 Insulation material/rear panel
2 Steam generator 6 Interior of appliance
3 Connection cable to the control unit 7 Tube on steam generator
4 Screw connections (four screws)

Spare part Model Spare part number


108/153 B04436
Steam generator
246 B04437

34 Service HCP/INCO | 03/2016


12.1.2 Replacing the steam generator pump
Description see page 44.

12.2 Humidity cannot be reduced


12.2.1 Finding errors
If the decrease of humidity in the chamber is not working proper-
ly, this can be seen by the fact that the humidity setpoint display
on the control panel does not change, even if a lower humidity is
set (see operating instructions). Increasing the humidity (humidi-
fication) works, however.
Proceed as follows to determine the cause:
1. Remove the top panel (see page 9).
2. Check that all tubes are connected to the humidity restriction
pump and the two accompanying sterile filters, and that they
are undamaged (Fig. 31).
3. Check the two sterile filters (2) for contamination (permeabil-
ity). If they are dirty, replace the sterile filters (see also page
24).
4. Check that the connection lead for the humidity restriction
pump (4) is connected properly.

2
4
6

5 1
Fig. 30 Humidity restriction in the control unit
1 Humidity restriction pump
2 Sterile filter
3 Tube connections
4 Connection lead
5 Screw connection
6 Removing the humidity restriction pump

Service HCP/INCO | 03/2016 35


If this does not result in the cause of the error being found, the
humidity restriction pump must be replaced:

12.2.2 Replacing the humidity restriction pump


1. Disconnect the mains plug.
2. Pull of the tubes from the pump (Fig. 31, 3).
3. Pull off the flat connector of the connection lead (4) on the
humidity/CO2 module.
4. Undo both screws on the mounting bracket (5) and remove
the pump (6).
5. Insert a new pump and screw it tight to the mounting brack-
ets.
6. Connect the flat connector of the pump to the humidity/CO2
module.
7. Push on the tube connections on the pump.
8. Check that the humidity restriction is now working properly.
If this is not the case, contact the manufacturer.
Spare part Spare part number
Humidity restriction pump E03660

36 Service HCP/INCO | 03/2016


12.3 Replacing the control module
1. Disconnect the mains plug.
2. Remove the top panel (see page 9).
3. Undo all plug connectors on the control module in the con-
trol unit (1).
4. Undo the screws of the control module (2) (4 screws) and
remove the control module (see also control unit overview on
page 11).
5. Screw in a new control module.
6. Reconnect the plug connections. The number of the respec-
tive slot is marked on the cables.
7. Re-attach the top panel.
8. Reconnect the mains plug.
9. Check that the humidification is now working properly. If this
is not the case, contact the manufacturer.

Fig. 31 Replacing control module for humidification


1 Control module
2 Screw connection (four screws)

Spare part Design Spare part number


active humidification B04376
Control unit
humidification passive humidification B04375
(INCO only)

Service HCP/INCO | 03/2016 37


12.4 Replacing the humidity sensor
1. Disconnect the mains plug.
2. Remove the top panel (see page 9) and pull out the rear
panel (see page 10).
3. Remove the insulation material surrounding the humidity
sensor on the rear side of the appliance (3).
4. Undo the connection cable (2) of the humidity sensor (4) on
the control module in the control unit.
5. Undo the set screw on the humidity sensor (5) and pull out
the humidity sensor (6).
6. Pull the humidity sensor out and upwards through the open-
ing in the floor of the control unit (1).
7. Push the new humidity sensor down through the opening
in the floor of the control unit and insert into the rear of the
appliance.
8. Screw tight the humidity sensor with the set screw and put
the insulation material back in.
9. Connect the plug connection of the humidity sensor to the
control module in the control unit.
10. Re-attach the rear and top panels.
11. Reconnect the mains plug.
12. Check that the humidification and dehumidification is now
working properly. If this is not the case, contact the manufac-
turer.

4
1

3
4
6

Fig. 32 Replacing the humidity sensor


1 Pull out the humidity sensor upwards
2 Connection lead of humidity sensor
3 Remove insulation material
4 Humidity sensor
5 Set screw
6 Pull humidity sensor out of rear panel of appliance

Spare part Spare part number


Humidity sensor E05249

38 Service HCP/INCO | 03/2016


13. CO2-supply
(only for incubators of the INCO type)

13.1 CO2 sensor


When replacing the CO2 sensor in the rear panel of the appliance
the accompanying PCB in the control unit must always be re-
placed as well, as the two components are tuned to each other
(see control unit overview on page 12, Fig. 5, item 8).
1. Disconnect the mains plug.
2. Remove the top panel (see page 9) and pull out the rear
panel (see page 10).
3. Replace the PCB of the CO2 sensor (Fig. 34, 1) in the control
unit, which is screwed to the control module of the humid-
ity unit (2). To do this, undo the two plastic screws (3 and 4)
and pull the PCB upwards (5). Carefully insert the new PCB
(included in the spare parts delivery) and attach with the two
plastic screws.

3 1 2

Fig. 33 Replacing the PCB of the CO2 sensor


1 PCB of the CO2 sensor
2 Control module for humidification
3 Plastic screw
4 Plastic screw
5 Remove PCB

Service HCP/INCO | 03/2016 39


4. Remove insulation material on the CO2 sensor in the rear
panel of appliance (Fig. 35).
5. Pull off 25-pin signal lead (2) and both control cables (3).
6. Pull off silicone tubes at the top and left on the CO2 sensor
(Fig. 36).
7. Undo both screws (4) of the mainboard.
8. Carefully undo sensor from the silicone tube on the rear pan-
el (Fig. 37). Hold the silicone tube tightly so that it remains
positioned on the connecting piece on the rear panel.
9. Connect the new sensor first to the silicone tube.
10. Attach the sensor with both screws.
11. Put the silicone tubes on the sensor on the top and to the
left.
12. Re-attach the insulation material.
13. Connect the 25-pin signal lead and the two control cables.
14. Push the rear panel back in and mount the top panel.

2
4

4
3 5

Fig. 34 Fig. 35
1 Remove insulation material 4 Pull off tubes at the top and on the left of the
2 Remove 25-pin signal lead sensor
3 Remove control cable 5 Undo the screws on the left and right of the
PCB (2 screws)

Fig. 36 Pull off the sensor with the PCB from the tube coming out the rear. Pull the tube from the sensor,
and not from the appliance, when doing this.

40 Service HCP/INCO | 03/2016


13.2 Auto-zero measuring gas pump
The auto-zero measuring gas pump is used to calibrate the CO2
sensor to zero. It briefly sprays the sensor with fresh air to do
this.
Function test 1. Remove the top panel (see page 9).
2. Switch on appliance; the auto-zero measuring gas pump
must run for about 1 second. Then the pump must run for
the duration of the zero point calibration. In this period, AUTO
ZERO can be seen in the display.
In case the pump does not run, it is defective and must be re-
placed:
Replacement 1. Disconnect the mains plug.
2. Undo the tube fuse (1) and pull off tube from the measuring
gas pump between the measuring gas pump and the sterile
filter (2).
3. Pull off the connection leads of the auto-zero measuring gas
pump (3) from the humidity/CO2module (mainboard).
4. Undo the side mounting bolts (4) and remove the pump (5).
5. Screw tight new pump to the mounting bracket.
6. Connect the connection leads of the auto-zero measuring gas
pump to the humidity/CO2 module.
7. Push on the tubes.
8. Connect the mains plug and subject the pump to a function
test; switch on the appliance for this purpose (see above). If
the error has not been rectified: Contact the manufacturer.

1
2

5
3
4

Fig. 37 Replacing the auto-zero measuring gas pump


1 Undo the tube safety catch
2 Pull off tube
3 Pull off connection leads from the humidity/CO2 module
4 Undo the mounting bolts on the mounting bracket
5 Remove the auto-zero measuring pump upwawrds

Spare part Spare part number


Auto-zero measuring gas pump E03604

Service HCP/INCO | 03/2016 41


13.3 Replacing solenoid valve
The two solenoid valves (2) regulate the CO2 supply via the two
external connections.
1. Disconnect the mains plug.
2. Interrupt CO2 supply at the gas bottle.
3. Remove the top panel (see page 9).
4. Undo the tube connections (1) on the solenoid valve (2) and
pull off tubes (3).
5. Unplug the connection leads (4).
6. Undo the two side screws (5) on the mounting bracket and
remove the pump (6).
7. Screw a new solenoid valve onto the mounting bracket.
8. Connect the leads.
9. Push on the tubes and screw tight.
10. Re-assemble the top panel; connect the mains plug and turn
on the CO2 supply with primary pressure of approx.1 bar.

4
3

2
1
6 3
5

Fig. 38 Replacing solenoid valve


1 Undo the tube connections
2 Solenoid valves
3 Pull off tubes
4 Unplug connection leads
5 Undo the screws on the mounting bracket
6 Remove the solenoid valve upwards
Spare part Spare part number
Solenoid valve 24 V E05232

42 Service HCP/INCO | 03/2016


13.4 CO2 tube connections
Caution:
The tubes (4) between the gas bottle connections on the rear of
the appliance (3) and the solenoid valves (1) must withstand the
maximum CO2 gas pressure. Only a special pressure tube may be
used here. All tube connections must be secured with a safety
catch.
Replacement 1. Disconnect the mains plug.
2. Switch off the gas supply.
3. Remove the top panel (see page 9).
4. Undo the tube connections (2) and remove the CO2 tube.
5. Put on a new tube and screw it tight.
6. Re-attach the top panel.

1 2

5 4
Fig. 39
1 Solenoid valves
2 Tube connections
3 Connections for gas supply
4 Pressure tube (not temperature resistant but pressure resistant)
5 Silicon tube (max. 200 °C temperature resistant)

Spare part Spare part number


Pressure tube (4) R00258
CO2 gas tube R00201

13.5 Replacing CO2 sterile filter


See page 24.

Service HCP/INCO | 03/2016 43


13. Replacing the INCO steam generator pump
1.1 Required spare parts
Spare part Voltage [V] Part number

Steam generator pump 230 V (M6) all E07236

Coupler relay 12 V DC/230 V AC (K5) all E07237

Relay module (Celduc solid state relay) all E05343

M3 x 35 screw all E03519

2 x two-way splitter all E00096

Power module 115 V → 230 V (N32) 115 B32809

Power module distance piece 115 E29721

Retrofit wire kit all D38895

1.2 Required tools


► Flat-head screwdriver size 1
► Wire-cutting pliers
► Needle-nose pliers
► Phillips screwdriver PH 1
► Cordless screwdriver size 2.5, for 115 V HPP appliances also size 4.8
► Thread cutter M3

1.3 Replacement
NOTICE
► For the following steps, please consult the wiring diagrams which were provided
with the retrofit kit, the circuit diagrams on page 49 and the layout drawings for
the components in the control unit from page 50 onwards.

► Make sure that no residual water leaks out when working on the pump –
especially when you pull off the tubes. This could damage the electronics in the
control unit.

44 Service HCP/INCO | 03/2016


1. Disconnect the mains
plug.
2. Remove the top panel
as described on page
9.

3. Inside the control unit,


carefully pull off both
tubes from the pump
(see figure).

4. Pull off the pump’s two


connection leads from
the S13 and S14 slots in
the power module.

5. Use the Phillips screw-


driver to remove the
pump from the retain-
ing plate (loosen 2
screws).
6. Remove the pump.

7. Use two screws to


fasten the new pump to
the retaining plate.

Service HCP/INCO | 03/2016 45


NOTICE
Please note that the new
pump does not turn in
the same direction as the
previous one. Therefore, the
tube connections have to be
interchanged (see figure on
the right for flow direction).

8. Connect the tubes to


the new pump.

9. Remove the relay mod-


ule from the coupler
relay (E07237) and
replace it with relay
module E05343.
E05343

10. Pull the S3-N4 and S9-


N4 leads off the power
module, attach the
splitter and reconnect
the leads (see figure).
► For INCO 230 V appli-
ances please continue
with step 11.
► For INCO 115 V appli-
ances please continue
with step 12 on page
47.

46 Service HCP/INCO | 03/2016


11. Use the wires in the
retrofit kit to connect
the relay to the S13 and
S14 slots in the power
module, as well as to
the splitter on the S3
and S9 connections
(according to the wiring
diagram).
Continue with step 16
on page 48.

S3/S9
S13/S14

For INCO 115 V appliances,


an additional N32 power
module has to be mounted
(see figure on the right).
12. For the distance pieces,
drill 4 holes (4.8) into
the bottom of the con-
trol unit. Please consult
the layout drawings
from page 47 on-
wards.
13. Use a distance piece as
shown when attaching
the N32 power module.
14. Connect the N32
power module to the
relay and to the main
power module’s split-
ters on the S3 and S9
connections according
to the wire list and cir-
cuit diagram (Fig. 41 on
page 47).
15. Use wires from the ret-
rofit kit to connect the
relay that has already
been linked to the new
N32 power module to
the S13 and S14 slots in
the main power mod-
ule.

N32
S13/S14

Service HCP/INCO | 03/2016 47


16. Drill a hole (2.5) be-
tween the pump and
the control unit wall
and cut an M3 thread.
17. Use an M3x35 screw to
attach the relay to the
bottom of the control
unit (see figure).

18. Connect the pump’s


connection leads to the
relay (see figure).
19. Then, reattach the top
panel, as described on
page 9.
20. Carry out a safety check
and a functional test.

48 Service HCP/INCO | 03/2016


4 5 6 7 8 9 10 11

1.4 Circuit diagram details


R6

D only with
door heating

M1 rd M4 active humidity 5 6 2

M
M bn bu steam generator pump
heating

S1
R9

X11 X12 bk 21 11
CO2 auto zero pump 3 4 1

B
31 41 K5
32 42 only INCO

K1
33 43
34 44 B C 24 22 14 12
35 45
- +

A
Z3 A2 A1
suppressor

CO2-SENS
S22
S21
S20
S19
S18
S17
S16
S15
S14
S13
V2 - S22
V2 + S21
V1 - S20
door heater - S19
mv2- S18
adj. pump - S17
adj. pump + S16
mv1- S15
s.g. pump - S14
s.g. pump + S13

4 5 6 S23 S23 +24V


7 8 9 10 11 12 13
J8
J8 4 3 1
LT sw 20 20 S24 J5
S24 +24V J5
LT br 19 19
18 18
humidity sensor

ais NO S25 S25 +24V


elais C 17 17 S12
MgV+ 16 16 S26 B3 lamp - S12
S26 +24V
15 15 S11
Fig. 40 Circuit diagram detail INCO
lamp + 230 V
MgV-
N 14 14
A S11

N 13 13 4
N 12 12
N4
11 11
3

eater2
N8
J9

eater1 10 10
9 9 F only with L N PE
door heating

N
M4 active humidity 5 6 2 230V -> 24V
M

N 8 8 M1 rd 25 N5
J7

7 7 M bn bu steam generator
J6pump
heating

eater1
S1 power supply
R9

J6
2 L 6 6 bk 21 11 15V
5 5 CO2 auto zero pump 3 4 1
K5 J1 B

B
L
4 4
only INCO 200W S1 S2 S3 S4 S
N-OUT
~ ~ ~ +
3 3 N6
K1

K2
L-PTC

L-SNT

L-OUT
C E 24 22 14 12
N-IN 2 2 V+ V-
N
N
N
N
N

L
L
L

1 1 - +
A

A
L-IN
Z3
S1 S1
S2 S2
S3 S3
S4 S4
S5 S5
S6 S6
S7 S7
S8 S8
S9 S9
S10 S10

A2 A1
suppressor

only with active humidity


CO2-SENSOR
F N3
S22
S21
S20
S19
S18
S17
S16
S15
S14
S13

N32
V2 - S22
V2 + S21
V1 - S20
door heater - S19
mv2- S18
adj. pump - S17
adj. pump + S16
mv1- S15
s.g. pump - S14
s.g. pump + S13

115V --> 230V


S23 S23 +24V
J8 4 3 1
S24
J8
J5 bnS1 S2
11
S24 +24V J5 AB AB
12
M
humidity sensor

S25 S25 +24V 13


B3 lamp - S12
S12 M3 14
S26 S26 +24V bu
B S11 R10 air pump 15
lamp + S11
PTC only INCO
4
N4
only with X13
active humidity
3

N8
J9

L N PE
B1

230V -> 24V


25 N5
Alarm
J7

J6
J6 power supply 15V
B2

Fig. 41 Circuit
J1 diagram detail INCO 115 V 200W S1 S2 S3 S4 S5 S6
X2 ~ ~ ~ + - +
Temp

N6 r
L-PTC

L-SNT

5
D1

V+ V- b
N
N
N
N
N

O2 sensor
L
L
L

b
y
S1 S1
S2 S2
S3 S3
S4 S4
S5 S5
11S6 S6
3 S7 S7
4 S8 S8
S9 S9
S10

B4 y
5S10

O2 senso
10

13
8
9

1
2

6
7
8
9
Debug 10
Servo 1
Encoder 2

11
PTAL / Pres 3
PT1 4
PT2 5
PT3 6
PT4 7

Power 13

controller expansion controller expansion


RS-xxx
Printer

Card

O2-Option only INCO


humidity CO2
A 11
bn 11 21
K3 12 22
M 13 23
12 14 M3
N21

+ - PTC Controller
R11 N1
R10
bu
EXT.DISP
air pump
14
15
24
25
PTC
USB
N22

only with A1 A2
active humidity X13 X14

bu
N1a N1b
B1

55161 55163 55163


Alarm

only INCO M M2
49
B2

Service HCP/INCO | 03/2016 rd


chamber fan
X2
Temp
D1
1.5 Positions of components in the control unit
INCO 108 L

100,2
132
Steam generator pump E07236
Schlauchpumpe E07236

Coupler relayE07237
Koppelrelais E07237
22 Celduc E05343
Celduc E05343
80

135
9

Power module
E38251 (230E38251
Trafonetzteil V)
115 V Geräte:
B32809 (115B32809
V)

Fig. 42 Control unit INCO 108L

INCO 153 L
100,2
128

Steam generator pump E07236


Schlauchpumpe E07236

Koppelrelais
Coupler E07237
relay E07237
22 Celduc E05343
Celduc E05343
85

229
49

Trafonetzteil
Power module E38251
E38251 (230 V)/B32809 (115 V)
115 V Geräte: B32809

Fig. 43 Control unit INCO 153L

50 Service HCP/INCO | 03/2016


Power module
Trafonetzteil E38251
INCO 246 L
E38251
115 (230
V Geräte: V)
B32809
B32809 (115 V) 145

10

100
138
Steam generator pump E07236
Schlauchpumpe E07236

Koppelrelais
Coupler E07237
relay E07237
Celduc E05343
Celduc E05343
22
85

Fig. 44 Control unit INCO 246L

Service HCP/INCO | 03/2016 51


52
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

X1.1
A PE A
N
L heating

heating R4
R2

R5
R6

heating
heating
B heating heating
B
D only with

M
R1 R3 M1 rd M4 active humidity 5 6 2
steam generator pump heaters
M bn bu

R9
X5 X6 X7 X8 X9 X10 X11 X12 bk 21 11
S1

heating
CO2 auto zero pump 3 4 1 X6 X5 X5 X7

door heating
B
B
11 21 31 41 11 21 31 41 K5
12 22 32 42 12 22 32 42 only INCO
R1 R2
13 23 33 43 13 23 33 43

K1
K2
C 14 24 34 44 14 24 34 44 B C 24 22 14 12 C
bn bn bk bk
15 25 35 45 15 25 35 45

A
A
- +
Z3 X9 X8 X8 X10
A2 A1
suppressor R3 R4
ye ye bu bu

D X9 X10 X11 X12 D


CO2-SENSOR
R5 R6
N3

S22
S21
S20
S19
S18
S17
S16
S15
S14
S13
10 gn gn rw-heater

S22
S21
S20
S19
S18
S17
S16
S15
S14
S23 S23 +24V

V2 -
V1 -

24
23
22
21
V2 +
mv2-
mv1-
J8
J8 4 3 1
LT sw 20 20 S24 J5

adj. pump -
s.g. pump -
S24 +24V J5

adj. pump +
s.g. pump + S13

door heater -
LT br 19 19

zus. GND
IPP: SNT- 24
29

NO / PEL- 23
29 expansion Relais NO 18 18

IPP: SNT+ 21

NC / PEL+ 22
14. Circuit diagrams

E S25 S25 +24V


E
Relais C 17 17 S12
B3 lamp - S12
MgV+ 16 16 S26 S26 +24V
28 28 controller MgV- 15 15 S11
A lamp + S11
N 14 14

humidity sensor
N 13 13 4 only INCO

variant:
adress:
27A 27a Relais N 12 12
N4

3
L-heater2 11 11
10 10 N8

J9
L-heater1

N2
F 27B 27b Relais N 9 9
L N PE F
230V -> 24V
N 8 8 25 N5

J7
N-heater1 7 7 J6 4
J6 power supply
J1-N4

L 6 6 15V
25 25 TB L 5 5 J1 O2
200W S1 S2 S3 S4 S5 S6
N-OUT 4 4
~ ~ ~ + - + red
26 26 TB L-OUT 3 3 N6

LEISTUNGSTEILE_55162
N-IN 2 2 V+ V-
5 black

N
N
N
N
N
L-PTC
L
L
L
L-SNT
L-IN 1 1 O2 sensor blue

SL 30
yellow

55162
G G
plug O2

B4 yellow

S1 S1
S2 S2
S3 S3
S4 S4
S5 S5
S6 S6
S7 S7
S8 S8
S9 S9
TB S2

S10 S10
O2 sensor cable
N7
V-/N6
V+/N6

bk
rd

O2-Option only INCO


H H
bn 11 21
12 22
M 13 23
M3 14 24
bu
air pump 15 25
PTC R10
8 only with
active humidity X13 X14
I colour-info I
bu bk = black

B1
M

Alarm
bn = brown
rd M2

B2
chamber fan rd = red
X2

Temp
bu = blue

D1
J ye = yellow J
gn = green

8
9
10
1
2
11
3
4
5
6
7
13

8
9
1
2
6
7
Optional controller expansion controller expansion date name

PT1 4
PT2 5
PT3
PT4

Card 11

Servo
Power 13

Debug 10

Printer
S2

RS-xxx
humidity CO2

Encoder
edited by 11.12.15 fm

PTAL / Pres 3

3,6V

230V
K S1 checked by 11.12.15 jl designation: K
N20
unit or size:
Controller N1 EXT.DISP INCO2/HCP

N21
K5 eingefügt wegen

ETHERNET
D Änderung M4 auf 230V 09.12.15 fm 108

USB
N22
C RC-Glied bezeichnet 23.06.15 fm nominal voltage: drawing number:
55161
N1a 55163 N1b 55163 B RC-Glied eingefügt 29.09.14 fm
L only INCO 230V D32939 L
A Umbezeichnung der
Klemmen N4/S23 -> S26 07.07.14 fm
rev. modification date name INCO2_HCP108 18.02.2016 09:38:44
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Service HCP/INCO | 03/2016


1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

X1.1
A PE A
N
L heating

heating R4
R2

R5
R6

heating
heating
B heating heating
B
F only with

M
R1 R3 M1 rd M4 active humidity 5 6 2
steam generator pump heaters
M bn bu

R9
X5 X6 X7 X8 X9 X10 X11 X12 bk 21 11
S1

heating
CO2 auto zero pump 3 4 1 X6 X5 X5 X7

door heating
B
B
11 21 31 41 11 21 31 41 K5
12 22 32 42 12 22 32 42 only INCO
R1 R2
13 23 33 43 13 23 33 43

K1
K2
C 14 24 34 44 14 24 34 44 C E 24 22 14 12 C
bn bn bk bk
15 25 35 45 15 25 35 45

A
A
- +
Z3 X9 X8 X8 X10
A2 A1
suppressor R3 R4
ye ye bu bu

Service HCP/INCO | 03/2016


only with active humidity
D X9 X10 X11 X12 D
CO2-SENSOR
F R5 R6
N3

S22
S21
S20
S19
S18
S17
S16
S15
S14
S13
10 N32 gn gn rw-heater

S22
S21
S20
S19
S18
115V --> 230V
S23 S23 +24V

V2 -
V1 -

24
23
22
21
V2 +
mv2-
mv1- S15
J8
J8 S1 S2 4 3 1
LT sw 20 20 S24 J5

adj. pump - S17


s.g. pump - S14
S24 +24V J5

adj. pump + S16


s.g. pump + S13

door heater -
LT br 19 19 AB AB

zus. GND
IPP: SNT- 24
29

NO / PEL- 23
29 expansion Relais NO 18 18

IPP: SNT+ 21

NC / PEL+ 22
E S25 S25 +24V
E
Relais C 17 17 S12
B3 lamp - S12
MgV+ 16 16 S26 S26 +24V
28 28 controller MgV- 15 15 S11
B lamp + S11
N 14 14

humidity sensor
N 13 13 4 only INCO

variant:
adress:
27A 27a Relais N 12 12
N4

3
L-heater2 11 11
10 10 N8

J9
L-heater1

N2
F 27B 27b Relais N 9 9
L N PE F
230V -> 24V
N 8 8 25 N5

J7
N-heater1 7 7 J6 4
J6 power supply
J1-N4

L 6 6 15V
25 25 TB L 5 5 J1 O2
200W S1 S2 S3 S4 S5 S6
N-OUT 4 4
~ ~ ~ + - + red
26 26 TB L-OUT 3 3 N6

LEISTUNGSTEILE_55162
N-IN 2 2 V+ V-
5 black

N
N
N
N
N
L-PTC
L
L
L
L-SNT
L-IN 1 1 O2 sensor blue

SL 30
S1
S2
S3
S4
S5
S6
S7
S8
S9
yellow

55162
S10
G G
plug O2

B4 yellow

S1
S2
S3
S4
S5
S6
S7
S8
S9
TB S2 D

S10
O2 sensor cable
N7
V-/N6
V+/N6

bk
rd

O2-Option only INCO


H A 11 H
bn 11 21
K3 12 22
M 13 23
12 14 M3 14 24
PTC R11 bu
+ - R10 air pump 15 25
only with PTC
8 A1 A2 X13 X14
active humidity colour-info
I I
bu bk = black

B1
M

Alarm
bn = brown
rd M2

B2
chamber fan rd = red
X2

Temp
bu = blue

D1
J ye = yellow J
gn = green

8
9
10
1
2
11
3
4
5
6
7
13

8
9
1
2
6
7
Optional controller expansion controller expansion date name

PT1 4
PT2 5
PT3
PT4

Card 11

Servo
Power 13

Debug 10

Printer
S2

RS-xxx
humidity CO2

Encoder
edited by 11.12.15 fm

PTAL / Pres 3

3,6V

230V
K F K5 + N32 eingefügt wegen designation: K
S1 Änderung M4 auf 230V 09.12.15 fm checked by 11.12.15 jl
N20
E RC-Glied bezeichnet 23.06.15 fm unit or size:
Controller N1 EXT.DISP INCO2/HCP

N21
Bzg. Leistungsteil falsch

ETHERNET
D -> wurde geändert 23.01.15 fm 108

USB
N22
C RC-Glied eingefügt 29.09.14 fm nominal voltage: drawing number:
55161
N1a 55163 N1b 55163 Umbezeichnung der
B Klemmen N4/S23 -> S26 07.07.14 fm
L only INCO 115V D32940 L
A Relais K3 und
PTC R11 eingefügt 06.06.14 fm
rev. modification date name INCO2_HCP108 18.02.2016 09:38:44
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

53
54
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

A A
components list
- B1: PT100 temperature sensor
B - B2: PT100 alarm sensor B

- B3: humidity sensor


- B4: O2 sensor
C - K1: magnetic valve gas inlet 1 C
- K2: magnetic valve gas inlet 2
A
- K3: relais (only 115V) B
- K5: relais for steam generator pump
D D
- M1: CO2 auto zero pump
- M2: chamber fan
- M3: air pump
E - M4: steam generator pump E
- N1: controller
- N1a: controller expansion humidity
F
- N1b: controller expansion CO2 F
- N2: power module
- N3: CO2 sensor
- N4: CO2-rh-module
G - N5: power supply 15V G
- N6: power supply 24V/200W
- N7: O2 module
- N8: rh converter 3pol. → 4pol.
H - N20: ethernet power supply H

- N21: ethernet converter


- N22: USB converter B

I - N32: 115V->230V power board I


- R1 → R6: heating element
- R9: door heating
- R10: PTC sense
J - R11: PTC power A J

- S1: main switch


date name
- S2: temperature limitor edited by 11.12.15 fm
K - X1.1: main plug checked by 11.12.15 ho designation: K
- X2: printer cable unit or size:
INCO2/HCP
- X5 → X12: terminal heating 108
nominal voltage: drawing number:
- X13 → X14: terminal chamber fan B Bezeichnung K5 + N32
L eingefügt 09.12.15 fm 230V/115V L
A Bezeichnung R11
und K3 eingefügt 06.06.14 fm
rev. modification date name INCO2_HCP108 18.02.2016 09:38:44
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Service HCP/INCO | 03/2016


1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

X1.1
A PE A
N
L heating heating heating

R2 R4 R5
heating heating heating

R6a
B R1 R3 R6 B

heating
heating
D only with

M
R7 M1 rd M4 active humidity 5 6 2
steam generator pump heaters
M bn bu

R9
X5 X6 X7 X8 X9 X10 X11 X12 bk 21 11
S1

heating
CO2 auto zero pump 3 4 1 X6 X5 X5 X7

door heating
B
B
11 21 31 41 11 21 31 41 K5
12 22 32 42 12 22 32 42 only INCO
R1 R2
13 23 33 43 13 23 33 43

K1
C

K2
C 14 24 34 44 14 24 34 44 B 24 22 14 12 C
bn bn bk bk
15 25 35 45 15 25 35 45

A
A
- +
Z3 X6 X8 X8 X7
suppressor A2 A1
R3 R4
ye ye bu bu

Service HCP/INCO | 03/2016


D X9 X10 X9 X11 D
CO2-SENSOR
R5 R6
N3

S22
S21
S20
S19
S18
S17
S16
S15
S14
S13
10 gn gn maul-heater

S22
S21
S20
S19
S18
S17
S16
S15
S14
S23 S23 +24V X11 X10 X9 X10

V2 -
V1 -

24
23
22
21
V2 +
mv2-
mv1-
J8
J8 4 3 1
LT sw 20 20 S24 J5 R6a R7

adj. pump -
s.g. pump -
S24 +24V J5

adj. pump +
s.g. pump + S13

door heater -
LT br 19 19

zus. GND
IPP: SNT- 24
29 18

NO / PEL- 23
29 expansion Relais NO 18

IPP: SNT+ 21

NC / PEL+ 22
E S25 S25 +24V maul-heater rw-heater E
Relais C 17 17 S12
B3 lamp - S12
MgV+ 16 16 S26 S26 +24V
28 28 controller MgV- 15 15 S11
A lamp + S11
N 14 14

humidity sensor
N 13 13 4 only INCO

variant:
adress:
27A 27a Relais N 12 12
N4

3
L-heater2 11 11

N1
10 10 N8

J9
L-heater1
F 27B 27b Relais N 9 9
L N PE F
230V -> 24V
N 8 8 25 N5

J7
N-heater1 7 7 J6 4
J6 power supply
J1-N4

L 6 6 15V
25 25 TB L 5 5 J1 O2
200W S1 S2 S3 S4 S5 S6
N-OUT 4 4
~ ~ ~ + - + red
26 26 TB L-OUT 3 3 N6

LEISTUNGSTEILE_55162
N-IN 2 2 V+ V-
5 black

N
N
N
N
N
L-PTC
L
L
L
L-SNT
L-IN 1 1 O2 sensor blue

SL 30
yellow

55162
G G
plug O2

B4 yellow

S1 S1
S2 S2
S3 S3
S4 S4
S5 S5
S6 S6
S7 S7
S8 S8
S9 S9
TB S2

S10 S10
O2 sensor cable
N7
V-/N6
V+/N6

bk
rd

O2-Option only INCO


H H
bn 11 21
12 22
M 13 23
M3 14 24
bu
air pump 15 25
PTC R10
8 only with
active humidity X13 X14
I colour-info I
bu bk = black

B1
M

Alarm
bn = brown
rd M2

B2
chamber fan rd = red
X2

Temp
bu = blue

D1
J ye = yellow J
gn = green

8
9
10
1
2
11
3
4
5
6
7
13

8
9
1
2
6
7
Optional controller expansion controller expansion date name

PT1 4
PT2 5
PT3
PT4

Card 11

Servo
Power 13

Debug 10

Printer
S2

RS-xxx
humidity CO2

Encoder
edited by 11.12.15 fm

PTAL / Pres 3

3,6V

230V
K S1 checked by 11.12.15 jl designation: K
N20
unit or size:

N21
Controller N1 EXT.DISP INCO2/HCP
K5 eingefügt wegen

ETHERNET
D Änderung M4 auf 230V 10.12.15 fm 153

USB
N22
C RC-Glied bezeichnet 23.06.15 fm nominal voltage: drawing number:
55161
N1a 55163 N1b 55163 B RC-Glied eingefügt 29.09.14 fm
L only INCO 230V D32941 L
A Umbezeichnung der
Klemmen N4/S23 -> S26 07.07.14 fm
rev. modification date name INCO2_HCP153 18.02.2016 09:27:05
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

55
56
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

X1.1
A PE A
N
L heating heating heating heating

R2 R4 R6 R7
heating heating heating heating

B R1 R3 R5 R6a B
E only with

M
M1 rd M4 active humidity 5 6 2
steam generator pump heaters
M bn bu

R9
X5 X6 X7 X8 X9 X10 X11 X12 bk 21 11
S1

heating
CO2 auto zero pump 3 4 1 X6 X5 X6 X5

door heating
B
B
11 21 31 41 11 21 31 41 K5
12 22 32 42 12 22 32 42 only INCO
R1 R2
13 23 33 43 13 23 33 43

K1
D

K2
C 14 24 34 44 14 24 34 44 C 24 22 14 12 C
bn bn bk bk
15 25 35 45 15 25 35 45

A
A
- +
Z3 X8 X7 X8 X7
suppressor A2 A1
R3 R4
ye ye bu bu
only with active humidity
D X10 X9 X10X9 D
CO2-SENSOR
E R5 R6
N3

S22
S21
S20
S19
S18
S17
S16
S15
S14
S13
10 N32 gn gn maul-heater

S22
S21
S20
S19
S18
S17
S16
S15
S14
115V --> 230V
S23 S23 +24V X12X11 X12X11

V2 -
V1 -

24
23
22
21
V2 +
mv2-
mv1-
J8
J8 S1 S2 4 3 1
LT sw 20 20 S24 J5 R6a R7

adj. pump -
s.g. pump -
S24 +24V J5

adj. pump +
s.g. pump + S13

door heater -
LT br 19 19 AB AB

zus. GND
IPP: SNT- 24
29

NO / PEL- 23
29 expansion Relais NO 18 18

IPP: SNT+ 21

NC / PEL+ 22
E S25 S25 +24V maul-heater rw-heater E
Relais C 17 17 S12
B3 lamp - S12
MgV+ 16 16 S26 S26 +24V
28 28 controller MgV- 15 15 S11
B lamp + S11
N 14 14

humidity sensor
N 13 13 4 only INCO

variant:
adress:
27A 27a Relais N 12 12
N4

3
L-heater2 11 11

N2
10 10 N8

J9
L-heater1
F 27B 27b Relais N 9 9
L N PE F
230V -> 24V
N 8 8 25 N5

J7
N-heater1 7 7 J6 4
J6 power supply
J1-N4

L 6 6 15V
25 25 TB L 5 5 J1 O2
200W S1 S2 S3 S4 S5 S6
N-OUT 4 4
~ ~ ~ + - + red
26 26 TB L-OUT 3 3 N6

LEISTUNGSTEILE_55162
N-IN 2 2 V+ V-
5 black

N
N
N
N
N
L-PTC
L
L
L
L-SNT
L-IN 1 1 O2 sensor blue

SL 30
yellow

55162
G G
plug O2

B4 yellow

S1 S1
S2 S2
S3 S3
S4 S4
S5 S5
S6 S6
S7 S7
S8 S8
S9 S9
TB S2

S10 S10
O2 sensor cable
N7
V-/N6
V+/N6

bk
rd

O2-Option only INCO


H A 11 H
bn 11 21
K3 12 22
M 13 23
12 14 M3 14 24
PTC R11 bu
+ - R10 air pump 15 25
only with PTC
8 A1 A2 X13 X14
active humidity colour-info
I I
bu bk = black

B1
M

Alarm
bn = brown
rd M2

B2
chamber fan rd = red
X2

Temp
bu = blue

D1
J ye = yellow J
gn = green

8
9
10
1
2
11
3
4
5
6
7
13

8
9
1
2
6
7
Optional controller expansion controller expansion date name

PT1 4
PT2 5
PT3
PT4

Card 11

Servo
Power 13

Debug 10

Printer
S2

RS-xxx
humidity CO2

Encoder
edited by 11.12.15 fm

PTAL / Pres 3

3,6V

230V
K S1 checked by 11.12.15 jl designation: K
N20
E K5 + N32 eingefügt wegen unit or size:
Änderung M4 auf 230V 10.12.15 fm

N21
Controller N1 EXT.DISP INCO2/HCP

ETHERNET
D RC-Glied bezeichnet 23.06.15 fm 153

USB
N22
C RC-Glied eingefügt nominal voltage: drawing number:
29.09.14 fm
55161
N1a 55163 N1b 55163 der
B Umbezeichnung
Klemmen N4/S23 -> S26 07.07.14 fm
L only INCO 115V D32942 L
A Relais K3 und
PTC R11 eingefügt 06.06.14 fm
rev. modification date name INCO2_HCP153 18.02.2016 09:27:05
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Service HCP/INCO | 03/2016


1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

A A
components list
B
- B1: PT100 temperature sensor B
- B2: PT100 alarm sensor
- B3: humidity sensor
- B4: O2 sensor
C - K1: magnetic valve gas inlet 1 C
- K2: magnetic valve gas inlet 2
A
- K3: relais (only 115V) C

Service HCP/INCO | 03/2016


D
- K5: relais for steam generator pump D
- M1: CO2 auto zero pump
- M2: chamber fan
- M3: air pump
E - M4: steam generator pump E
- N1: controller
- N1a: controller expansion humidity
F
- N1b: controller expansion CO2 F
- N2: power module
- N3: CO2 sensor
- N4: CO2-rh-module
G - N5: power supply 15V G
- N6: power supply 24V/200W
- N7: O2 module
H
- N8: rh converter 3pol. → 4pol. H
- N20: ethernet power supply
- N21: ethernet converter
- N22: USB converter C

I - N32: 115V->230V power board I


- R1 → R7: heating element
- R9: door heating
J
- R10: PTC sense J
- R11: PTC power A

- S1: main switch


date name
- S2: temperature limitor edited by 11.12.15 fm
K - X1.1: main plug checked by 11.12.15 ho designation: K
unit or size:
- X2: printer cable 153
INCO2/HCP
- X5 → X12: terminal heating C Bezeichnung
eingefügt
K5 + N32
10.12.15 fm nominal voltage: drawing number:
- X13 → X14: terminal chamber fan B Bezeichnung Z3
L eingefügt 23.06.15 fm 230V/115V L
B A Bezeichnung R11
- Z3: suppressor und K3 eingefügt 06.06.14 fm
rev. modification date name INCO2_HCP153 18.02.2016 09:27:05
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

57
58
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

X1.1
A PE A
N
L heating heating heating heating

R2 R4 R5 R7
heating heating heating heating

B R1 R3 R6 R8 B
D only with

M
M1 rd M4 active humidity 5 6 2
steam generator pump heaters
M bn bu

R9
X5 X6 X7 X8 X9 X10 X11 X12 bk 21 11
S1

heating
CO2 auto zero pump 3 4 1 X6 X5 X5 X7

door heating
B
B
11 21 31 41 11 21 31 41 K5
12 22 32 42 12 22 32 42 only INCO
R1 R2
13 23 33 43 13 23 33 43

K1
B

K2
C 14 24 34 44 14 24 34 44 C 24 22 14 12 C
bn bn bk bk
15 25 35 45 15 25 35 45

A
A
- +
Z3 X6 X8 X8 X7
suppressor A2 A1
R3 R4
ye ye bu bu

D X9 X10 X9 X10 D
CO2-SENSOR
R5 R6
N3

S22
S21
S20
S19
S18
S17
S16
S15
S14
S13
10 gn gn maul-heater
S23 S23 +24V X11 X12 X11 X12

V2 - S22
V1 - S20

24
23
22
21
V2 + S21
mv2- S18
mv1- S15
J8
J8 4 3 1
LT sw 20 20 S24 J5 R7 R8

adj. pump - S17


s.g. pump - S14
S24 +24V J5

adj. pump + S16


s.g. pump + S13

door heater - S19


LT br 19 19

zus. GND
IPP: SNT- 24
29

NO / PEL- 23
29 expansion Relais NO 18 18

IPP: SNT+ 21

NC / PEL+ 22
E S25 S25 +24V rw-heater rw-heater E
Relais C 17 17 S12
B3 lamp - S12
MgV+ 16 16 S26 S26 +24V
28 28 controller MgV- 15 15 S11
A lamp + S11
N 14 14

humidity sensor
N 13 13 4 only INCO

variant:
adress:
27A 27a Relais N 12 12
N4

3
L-heater2 11 11

N3
10 10 N8

J9
L-heater1
F 27B 27b Relais N 9 9
L N PE F
230V -> 24V
N 8 8 25 N5

J7
N-heater1 7 7 J6 4
J6 power supply
J1-N4

L 6 6 15V
25 25 TB L 5 5 J1 O2
200W S1 S2 S3 S4 S5 S6
N-OUT 4 4
~ ~ ~ + - + red
26 26 TB L-OUT 3 3 N6

LEISTUNGSTEILE_55162
N-IN 2 2 V+ V-
5 black

N
N
N
N
N
L-PTC
L
L
L
L-SNT
L-IN 1 1 O2 sensor blue

SL 30
yellow

55162
G G
plug O2

B4 yellow

S1 S1
S2 S2
S3 S3
S4 S4
S5 S5
S6 S6
S7 S7
S8 S8
S9 S9
TB S2

S10 S10
O2 sensor cable
N7
V-/N6
V+/N6

bk
rd

O2-Option only INCO


H H
bn 11 21
12 22
M 13 23
M3 14 24
bu
air pump 15 25
PTC R10
8 only with
active humidity X13 X14
I colour-info I
bu bk = black

B1
M

Alarm
bn = brown
rd M2

B2
chamber fan rd = red
X2

Temp
bu = blue

D1
J ye = yellow J
gn = green

8
9
10
1
2
11
3
4
5
6
7
13

2
5
6
7
Optional controller expansion controller expansion date name

PT1 4
PT2
PT3
PT4

Card 11

Servo 1
Power 13

Debug 10

Printer 8
S2

RS-xxx 9
humidity CO2

Encoder
edited by 11.12.15 fm

PTAL / Pres 3

3,6V

230V
K S1 checked by 11.12.15 jl designation: K
N20
unit or size:

N21
Controller N1 EXT.DISP INCO2/HCP
K5 eingefügt wegen

ETHERNET
D Änderung M4 auf 230V 10.12.15 fm 246

N22
USB
C RC-Glied bezeichnet 23.06.15 fm nominal voltage: drawing number:
55161
N1a 55163 N1b 55163 B RC-Glied eingefügt 29.09.14 fm
L only INCO 230V D32943 L
A Umbezeichnung der
Klemmen N4/S23 -> S26 07.07.14 fm
rev. modification date name INCO2_HCP246 18.02.2016 09:29:13
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Service HCP/INCO | 03/2016


1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

X1.1
A PE A
N
L heating heating heating heating

R2 R4 R6 R8
heating heating heating heating
E only with

M
B R1 R3 R5 R7 M4 active humidity B
D C
bn bu steam generator pump
M1 rd 5 6 2
21 11 S1 heaters
Z3 M K5

13
5
3
1
A1
suppressor

R9
X5 X6 X7 X8 X9 X10 X11 X12 bk

heating
X5 X6 X5 X6

5
3
1
CO2 auto zero pump 3 4 1

door heating
B
B

11 21 31 41 11 21 31 41

13
A1
K4 only INCO 24 22 14 12
12 22 32 42 12 22 32 42 R1 R2
13 23 33 43 13 23 33 43 - + K1
K2

C 14 24 34 44 14 24 34 44
C

14
6
4
2
A2
A2 A1 bn bn bk bk
15 25 35 45 15 25 35 45
A
A

6
4
2
X7 X8 X7 X8

14
A2
R3 R4
ye ye bu bu

Service HCP/INCO | 03/2016


only with active humidity
D X9 X10 X9 X10 D
CO2-SENSOR
E R5 R6
N32 N3

S22
S21
S20
S19
S18
S17
S16
S15
S14
S13
10 115V --> 230V gn gn maul-heater
S23 S23 +24V X11 X12 X11 X12

V2 - S22
V1 - S20

24
23
22
21
24
23
22
21
V2 + S21
mv2- S18
mv1- S15
J8
J8 S1 S2 4 3 1
LT sw 20 20 LT sw 20 20 S24 J5 R7 R8

adj. pump - S17


s.g. pump - S14
S24 +24V J5

adj. pump + S16


AB AB

s.g. pump + S13

door heater - S19


LT br 19 19 LT br 19 19

zus. GND
zus. GND

IPP: SNT- 24
IPP: SNT- 24
29 29

NO / PEL- 23
NO / PEL- 23
29 expansion Relais NO 18 18 29 expansion Relais NO 18 18

IPP: SNT+ 21
IPP: SNT+ 21

NC / PEL+ 22
NC / PEL+ 22
E S25 S25 +24V 25 rw-heater rw-heater E
Relais C 17 17 Relais C 17 17 S12
B3 lamp - S12
MgV+ 16 16 MgV+ 16 16 S26 S26 +24V
28 28 controller MgV- 15 15 28 28 controller MgV- 15 15 S11
B lamp + S11
N 14 14 N 14 14

humidity sensor
N 13 13 N 13 13 4 only INCO

variant:
variant:

adress:
adress:
27A 27a Relais N 12 12 27A 27a Relais N 12 12
N4

3
L-heater2 11 11 L-heater2 11 11

N2
10 10 10 10 N8

J9
L-heater1 L-heater1

N2/2
F 27B 27b Relais N 9 9 27B 27b Relais N 9 9
L N PE F
230V -> 24V
N 8 8 N 8 8 N5

J7
N-heater1 7 7 N-heater1 7 7 J6 4
J6 power supply 15V
J1-N4

L 6 6 L 6 6
25 25 TB L 5 5 25 25 TB L 5 5 J1 S1 S2 S3 S4 S5 S6 O2
200W
N-OUT 4 4 N-OUT 4 4 ~ ~ ~ + - +
26 26 TB L-OUT 3 3 26 26 TB L-OUT 3 3 N6 red

LEISTUNGSTEILE_55162
LEISTUNGSTEILE_55162
N-IN 2 2 N-IN 2 2 V+ V-
5 black

N
N
N
N
N
L-PTC
L
L
L
L-SNT
L-IN 1 1 L-IN 1 1 O2 sensor blue

SL 30
SL 30
yellow

55162
55162
G G
plug O2

B4 yellow

S1 S1
S2 S2
S3 S3
S4 S4
S5 S5
S6 S6
S7 S7
S8 S8
S9 S9
TB S2
S10 S10
O2 sensor cable
N7
V-/N6
V+/N6

bk
rd

O2-Option only INCO


H A 11 H
bn 11 21
K3 12 22
M 13 23
12 14 M3 14 24
PTC R11 bu
+ - R10 air pump 15 25
only with PTC
8 A1 A2 X13 X14
active humidity colour-info
I I
bu bk = black

B1
M

Alarm
bn = brown
rd M2

B2
chamber fan rd = red
X2

Temp
bu = blue

D1
J ye = yellow J
gn = green

8
9
10
1
2
11
3
4
5
6
7
13

1
2
6
7
Optional controller expansion controller expansion date name

PT1 4
PT2 5
PT3
PT4

Card 11

Servo
Power 13

Debug 10

Printer 8
S2

RS-xxx 9
humidity CO2

Encoder
edited by 11.12.15 fm

PTAL / Pres 3

3,6V

230V
K S1 checked by 11.12.15 jl designation: K
N20
E K5 + N32 eingefügt wegen unit or size:
Änderung M4 auf 230V 10.12.15 fm

N21
Controller N1 EXT.DISP INCO2/HCP

ETHERNET
D RC-Glied bezeichnet 23.06.15 fm 246

N22
USB
C RC-Glied eingefügt nominal voltage: drawing number:
29.09.14 fm
55161
N1a 55163 N1b 55163 Umbezeichnung der
B Klemmen N4/S23 -> S26 07.07.14 fm
L only INCO 115V D32944 L
A Relais K3 und
PTC R11 eingefügt 06.06.14 fm
rev. modification date name INCO2_HCP246 18.02.2016 09:29:13
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

59
60
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

A components list A
- B1: PT100 temperature sensor
- B2: PT100 alarm sensor
B
- B3: humidity sensor B
- B4: O2 sensor
- K1: magnetic valve gas inlet 1
- K2: magnetic valve gas inlet 2
C - K3: relais (only 115V) A C
- K4: contactor (only 115V) C

- K5: relais for steam generator pump


D - M1: CO2 auto zero pump D
- M2: chamber fan
- M3: air pump
- M4: steam generator pump
E E
- N1: controller
- N1a: controller expansion humidity
- N1b: controller expansion CO2
F - N2: power module F
- N2/2: power module (only 115V)
- N3: CO2 sensor
- N4: CO2-rh-module
G - N5: power supply 15V G
- N6: power supply 24V/200W
- N7: O2 module
H - N8: rh converter 3pol. → 4pol. H
- N20: ethernet power supply
- N21: ethernet converter
- N22: USB converter C
I - N32: 115V->230V power board I

- R1 → R8: heating element


- R9: door heating
J - R10: PTC sense J
- R11: PTC power A
- S1: main switch date name
- S2: temperature limitor edited by 11.12.15 fm
K checked by 11.12.15 ho designation: K
- X1.1: main plug unit or size:
- X2: printer cable 246 INCO2/HCP
- X5 → X12: terminal heating C Bezeichnung K5 + N32 nominal voltage: drawing number:
eingefügt 10.12.15 fm
B Bezeichnung Z3
L - X13 → X14: terminal chamber fan eingefügt 23.06.15 fm L
A Bezeichnung R11
230V/115V
- Z3: suppressor B und K3 eingefügt 06.06.14 fm
rev. modification date name INCO2_HCP246 18.02.2016 09:29:13
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

Service HCP/INCO | 03/2016


Memmert GmbH + Co. KG | P.O. Box 1720 | D-91107 Schwabach | Tel. +49 (0) 9122 / 925 - 0 | Fax +49 (0) 9122 / 145 85 | E-Mail: service@memmert.com | www.memmert.com

04.03.2016
HCP/INCO Service englisch

D10840