4128
INORGANIC NONMETALS (4000)
4500-NO,~ C. Second-Derivative Ultraviolet Spectrophotometric Method
1. General Discussion
4, Principle: Even though NO,~ strongly absorbs UV light,
determining NO,~ by measuring the absorbance st one wave-
Jength may not be feasible because natural organic matter (NOM)
and other solutes also absorb UV light, The spectra of NOM ftom
different water sources usually ae different, so if NOM is present
the UV NO,~ screening procedure (4500-NO,”B) will not give
reliable results. The UV spectrum of NO.” is quite differen from
that of NOM. In the NO,~ spectrum, the absorbance increases
rapidly from 230 to 210 nm, whereas in NOM spectra, the absor-
bance increases gradually in the same wavelength range. Comput-
ing the second derivative of a sample spectrum effectively
eliminates the background NOM contribution
>. Interferences: The NO;~ and NO,~ UV spectra sre similar,
‘bat NO.” concentrations usually are much lower than NO3” con
centrations. Bicarbonate absorbs weakly at wavelengths <210 nm
bt docs not affect the second -derivative signal of NO”. Bromide
(Br) interferes at seawater concentrations (68 mg Br~/L, salinity
35 Se), so this method cannot be used to determine NO,~ in
seawater or samples containing high concentrations of Br. Neither
iron (Fe) nor copper (Cu) interferes at 2 mg/L, but both metals
setiously interfere at 20 mg/L." For NO,”-N concentrations up to
3 mplL, the UV absorbance spectrum has a peak at 203 um. For
higher concentrations the peak gradually shifts to 207 nm and the
second-dervative maximum also shifts from 224 am to higher
wavelengths. The shape of the UV and second-derivaive spectra
‘may be affected by the spectophotometer slit width. The method
has been tested for potable water but not for wastewater
2. Apparatus
1 Spectrophotometer capable of scanning from 250 to
200 nm. Some spectrophotometers can compute the second
10 or >,
respectively. Ions that are potential interferences but do not nor-
rally occur at significant levels in potable waters ae Br, eblorate
(CiOs>, cyanide (CN), hydrosufde (HS), iodide (7), NO”,
and perchlorate (ClO,~). Although the electrodes function satisfac-
teil in buffers ranging from pH 309, erate responses bave been
noted when pHs not held constant, Because the electrode responds
to NO,” sctiviy ter than concentration, oni strength must be
constant in all amples and sandals, Minimize these problems by
using a buffer solution containing silver sulfate (Ag,SO,) to remove
Br-,Cl",CN*,T" and HS" interferences, sulfamic acid to remove
NO,” alates at pl 3 to liminale HCO,” and maintain a constant
ET aod ionic reap, and aluminum sulfa [Al($O,)s] # com
plex orgatic aid, For example, 10 ml. of buffer will compensate
for about 781 mg/L of chloride interference from the sample mati.
2. Apparatus
4. pHaVASE meter (concentration mode) or pH meter, ex-
ppanded-scale or digitel, capable of 0.1 mV resolution (no con-
centration mod.
b. Nitrate ion-selective electrode, combination* or hal-cellt
with double-junction reference electrode: Fill outer chamber
With solution containing ammonium sulfate [(NH,),S0q) Care-
fully follow manufacturer's instructions on care and storage.
"Thema Ssienife Ori Model 97-OTBN, or equiv
{Teme Seientife Onan Mel SSOTBNWE. cr equlet
+ Menno Seine ri Medel 90-0200 or uae,
5r motorized stirrer: Tetrafiuoroethylene (TFE)-
coated sttring bar.
4. Pipets: If adjustable pipets are used, calibrate according to
manufacturer's directions.
3. Reagents
4a. Reagent water: Prepare as described in 4500-NO,” B3a.
Use forall solutions and dilutions
Sock nitrate solution A, 100 mg/L: Purchase from a com-
‘mercial source (shel life as stated by manufacturer) or prepare
by drying KNO, in a 103-105°C oven for 24 h. Dissolve
(0.7218 g + 0.0005 in water and dilute o 1000 mL; 1.00 mL. =
100 4g NO3"N, Solution is stable for at last 6 months
Stock nitrate solution B, 1 mL = 100 ug NO,” -N: Purchase
from a commercial source different from stock nitrate solution A.
or prepare as directed in J @ above using KNOs from a com-
mercial source different from that used to prepare stock nitrate
solution A.
4. Standard nitrate solutions: Dilute 1.0, 10.0, and 50.0 ml.
stock nitrate solution A to 100 mi. with reagent water to obtain
standard solutions of 1.0, 100, and 50.0 mg NO, -NIL, respec
tively. I calibration curve is nonlinear, consult the manual that
came with the meter. Additional standards may be recom-
mended. Some pH/ISE meters automatically perform nonlinear
calibration.
. Calibration verification standard: Dilute 10.0 ml. stock
nitrate solution B to 100 mL. with reagent water. Nitrate concen
tration is 10 mg NO,"-N/L.
‘Buffer solution: Dissolve 17.32 g Al,(S0)y~ 18H,0,
343g AgsSO,, 1.28 g HsBOs, and 2.52 g sulfamic acid
(G.NSO,H) in about 800 ml. reagent water, Adjust to pH1 3.0 by
slovy adding 0.102V NaOH. Dilute to 1000 mL and store in an
amber glass botle, Altematively, obtain from 2 commercial
source.§
{Sodium hydroxide (NaOH), 0.1M.
4 Reference electrode filing solution: Dissolve 053. §
(NH,),S0, in reagent water and dilute to 100 ml, or obiain a
‘Fitenwo Site Orin 920710, or equivalent