Beruflich Dokumente
Kultur Dokumente
IT1000-AC-W-Ex2/22
IT1000-DC-W-Ex2/22
IT1000-BATT-W-Ex2/22
IT1000-AC-E-Ex2/22
IT1000-DC-E-Ex2/22
IT1000-BATT-E-Ex2/22
II 3G Ex ic nA IIC T4 Gc
December 2017
ST.2309.1312
Rev. 9
File: IT1000_EX222_THE.DOC
Published By:
© SysTec Systemtechnik und Industrieautomation GmbH, Bergheim, Germany
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted
in any form or by any means, mechanical, photocopying, recording, or otherwise, without the prior written
permission of SysTec GmbH.
Terms and product names mentioned in this publication are trademarks, registered trademarks or
service marks of their respective owners. Use of a term should not be regarded as affecting the validity
of any trademark, registered trademark or service mark.
EPSON ESC/P® is a registered trademark of the SEIKO EPSON Corporation.
TOLEDO® is a registered trademark of Mettler-Toledo Inc.
Please Note:
While every precaution has been taken in the preparation of this manual, SysTec GmbH assumes no
responsibility for errors or omissions. Neither is any liability assumed for damages resulting from the use of
the information contained herein.
The publisher is grateful for any information and/or advice that may contribute to correct errors or omissions
in following editions.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 5
Contents
1 Introduction ........................................................................................................................... 9
1.1 About This Manual ......................................................................................................... 9
1.2 Safety Symbols Used In This Manual ............................................................................... 9
1.3 Safety Instructions ......................................................................................................... 9
1.4 Marking ...................................................................................................................... 11
2 System Description ............................................................................................................... 12
2.1 General....................................................................................................................... 12
2.2 Basic Units And Options ............................................................................................... 13
2.3 Housing ...................................................................................................................... 14
2.4 Description Of Components .......................................................................................... 15
3 Intended Use ........................................................................................................................ 17
4 Assembly ............................................................................................................................ 18
5 Installation ........................................................................................................................... 19
5.1 General....................................................................................................................... 19
5.2 Equipotential Bonding ................................................................................................... 19
5.3 Shielding..................................................................................................................... 19
5.4 Connection Of Cables................................................................................................... 20
6 Installation Of External Connections With Type Of Protection 'Ex-nA' ......................................... 21
6.1 Connection Of Power Supply To IT1000-AC-*-Ex2/22 ..................................................... 21
6.2 Connection Of Power Supply To IT1000-DC-*-Ex2/22 And IT1000-BATT-*-Ex2/22 ............ 22
6.3 Operation Of IT1000-BATT-*-Ex2/22 ............................................................................. 23
6.4 Connection Of Analog Scale To ADM ............................................................................. 26
6.5 Connection Of Serial Interface SIM ................................................................................ 29
6.6 Connection Of Parallel I/Os PIM ..................................................................................... 34
6.7 Connection Of Parallel I/Os IOM .................................................................................... 36
6.8 Connection Of 15-Bit Analog Output DAU15 .................................................................. 38
7 Commissioning ..................................................................................................................... 40
8 Installation Examples ............................................................................................................. 41
8.1 Installation Example IT1000-AC-*-Ex2/22....................................................................... 41
8.2 Installation Example IT1000-DC-*-Ex2/22....................................................................... 42
8.3 Installation Example IT1000-BATT-*-Ex2/22 With Akku-Box ............................................. 43
9 Service Mode ....................................................................................................................... 44
9.1 General....................................................................................................................... 44
9.2 Display And Keyboard .................................................................................................. 45
9.3 Operator Prompting ...................................................................................................... 46
9.4 Weighing Functions ...................................................................................................... 48
9.5 Power Up ................................................................................................................... 49
9.6 Overview Service Mode ................................................................................................ 50
10 Calibrate Scale ................................................................................................................... 52
10.1 Entering Calibration Mode ........................................................................................... 52
10.2 Capacity & Interval .................................................................................................... 53
10.3 Span Adjust .............................................................................................................. 56
10.4 Linearization .............................................................................................................. 58
10.5 Zero Adjust ............................................................................................................... 59
10.6 Adaptation ................................................................................................................ 59
6 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
1 Introduction
1.1 About This Manual
This manual contains information and Technical Data for the use, the installation and the operation of
the industrial weighing terminals IT1000-AC-W-Ex2/22, IT1000-AC-E-Ex2/22, IT1000-DC-W-Ex2/22
and IT1000-DC-E-Ex2/22, IT1000-BATT-W-Ex2/22 and IT1000-BATT-E-Ex2/22.
Information applicable to all versions refers to IT1000-Ex2/22. Specific information is marked with the
full designation of the respective version, e.g. 'IT1000-AC-W-Ex2/22'.
The weighing terminal must only be operated by trained personnel. For simple weighing transactions to
be carried out by the enduser the Operating Instructions IT1000 (ST.2309.0603) are available.
WARNING
Failure to observe this precaution could result in serious injuries or fatal accidents due to
Ex ignition of an explosive atmosphere (gas and/or dust). Please make absolutely sure that these
precautions are observed in order to ensure safe operation of the equipment.
WARNING
Failure to observe this precaution could result in serious injuries or fatal accidents. Please make
absolutely sure that these precautions are observed in order to ensure safe operation of the
equipment.
CAUTION
Failure to observe this precaution could result in damage to or destruction of the equipment or
bodily harm! Please make absolutely sure that these precautions are observed in order to ensure
safe operation of the equipment.
Note: This indicates an advice for the designated use of the equipment and/or additional information to
avoid inappropriate handling.
WARNING
Exercise utmost care when making checks, tests and adjustments that can actuate movable
parts such as feeding devices, gates, flaps, conveyors etc. Make absolutely sure that nobody is
within reach of movable parts.
WARNING
For installation, service and operation of the unit, the ATEX directive as well as all locally
Ex applicable regulations for safety and the prevention of accidents must be observed!
10 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
WARNING
This weighing terminal must not be used in Ex-Zone 0, 1, 20 and 21.
Ex It is the sole responsibility of the employer / operating company to classify the area of
installation (zone, group, temperature class). To that effect, the assistance of the Technical
Supervisory Service or any other technical authority may be utilized!
WARNING
When this unit is included as a component part of a system, the resulting system design must
Ex be reviewed by qualified personnel who are familiar with the construction and operation of all
individual components in the system and the potential hazards involved.
WARNING
Input voltage of the instrument must comply with local mains supply!
Ex
WARNING
This module and its associated equipment must only be installed, adjusted and maintained by
Ex qualified personnel authorized by SysTec GmbH!
WARNING
Connection of peripheral devices to IT1000A-Ex2/22 is made in compliance with Ex type of
Ex protection 'Ex-nA'. The connection values must be observed. Connected peripheral devices
installed in hazardous area must have appropriate Ex protection.
WARNING
When the instrument is permanently connected to the power supply, an easily accessible
Ex separator must be included in the supply circuit! It must either be located in safe area or -if
installed in Ex area- have appropriate Ex protection for the place of installation.
WARNING
The weighing terminal must not be installed in areas where electrostatic charging is possible
Ex which may produce spark discharges, brush discharges or propagating brush discharges at the
front panel.
Comment: According to common knowledge, the use of the equipment and the cleaning with a
damp cloth do not produce such a high surface charge density. However, the front panel must
not be wiped with a dry cloth!
WARNING
Risk of electrical shock! When operating electrical devices, parts of this devices are connected
to dangerously high voltages. Inappropriate use of such devices may lead to serious bodily
injuries or substantial damages to property.
WARNING
The power supply unit of the weighing terminal provides SELV voltages in accordance with EN
60950. Make sure that any peripheral device connected to the weighing terminal containing its
own power supply also uses SELV voltages!
WARNING
The device uses the short-circuit / overcurrent protection of the on-site mains supply.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 11
1.4 Marking
II 3G Ex ic nA IIC T4 Gc
Ex classification II 3D Ex ic tc IIIB T50°C Dc IP54 or
II 3D Ex ic tc IIIC T50°C Dc IP65
CE marking
2 System Description
2.1 General
IT1000-Ex2/22 is a general purpose weighing terminal for use in a variety of applications such as data
logging, data capturing, parts counting, setpoint control and filling, suitable for installation in Ex zones
2 and 22.
The weighing module ADM is used to connect one scale base with strain gauge loadcells with an
overall impedance of 21 Ω to 4500 Ω. In addition a battery-backed realtime clock can be installed.
Power supply through the built-in wide-range power supply unit is either for 110–240 VAC or 12–
30 VDC or 10–24 VDC (battery-powered version). Housings are available for both, desk-top/wall-
mount and panel-mount installation. Cables of all external components are connected at screw
terminals.
Weight and additional information is indicated on a back-lit 13-character LCD, height of characters
14mm. Operation is made via a sealed membrane keyboard with 5 keys.
Instead of operating the terminal locally via keyboard and display, full remote control is possible from a
PC. In applications legal for trade data can be stored in an approved data archive on the PC's harddisk.
This option requires the installation of the realtime clock in the weighing terminal and the PC program
PC COM+.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 13
2.3 Housing
IT1000-Ex2/22 is incorporated in a stainless steel housing with cable glands to connect cables to
peripheral devices. It is available in two versions:
The housing for desk-top / wall-mount installation (IT1000-**-W-Ex2/22) is protected to IP65.
The housing for panel-mount installation (IT1000-**-E-Ex2/22) is inserted in a suitable cutout of a
switch cabinet and fixed from the rear with 6 mounting brackets. When installed accordingly, the front
of the terminal -in combination with the circumferential gasket between terminal and switch cabinet- is
protected to IP65.
When installed in Ex zone 2 (as per II 3G Ex ic nA IIC T4 Gc) and Ex zone 22 with non-conductive dust
(as per II 3D Ex ic tc IIIB T50°C Dc IP54) the surrounding housing must be protected to IP54 or better.
When installed in Ex zone 22 with conductive dust (as per II 3D Ex ic tc IIIC T50°C Dc IP65) the
surrounding housing must be protected to IP65 or better.
The surrounding housing (e.g. switch cabinet) must conform to the requirements of EN60079-0 section
26.4.
Ex
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 15
X3 / X4
Socket for
ADM weighing
module
Notes:
• For the storage of volatile data the module of the optional realtime clock contains a lithium battery.
Dispose of used batteries as indicated by manufacturer.
• If no realtime clock is installed, entry of date and time is required on power up!
• For replacing the module observe the instructions in section 'Replacing the battery'!
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 17
3 Intended Use
In compliance with RL 2014/34/EU appendix I, the weighing terminal IT1000-Ex2/22 is an
apparatus of group II category 3G, that following RL 99/92/EG (ATEX 137), can be used in
Ex zone 2, as well as in gas groups IIA, IIB and IIC, which are at risk of explosion through
combustible substances of temperature classes T1 to T4. The following section 'Installation of
'IT1000-**-E-Ex2/22 in switch cabinet' must be observed.
IT1000-Ex2/22 is also an apparatus of group II category 3D, that following RL 99/92/EG (ATEX
137) can be used in zone 22 (dust), the maximum surface temperature is 50°C. The following
instructions on the installation of the IT1000-**-E-Ex2/22 in a surrounding housing must be
observed.
Installation of 'IT1000-**-E-Ex2/22 in switch cabinet: The housing for panel-mount installation
is inserted in a suitable cutout of a switch cabinet and fixed from the rear with 6 mounting
brackets. If installed accordingly, the front of the terminal -in combination with the
circumferential gasket between terminal and switch cabinet- is protected to IP65. When
installed in Ex zone 2 (as per II 3G Ex ic nA IIC T4 Gc) and Ex zone 22 with non-conductive
dust (as per II 3D Ex ic tc IIIB T50°C Dc IP54) the surrounding housing must be protected to
IP54 or better. When installed in Ex zone 22 with conductive dust (as per II 3D Ex ic tc IIIC
T50°C Dc IP65) the surrounding housing must be protected to IP65 or better. The surrounding
housing (e.g. switch cabinet) must conform to the requirements of EN60079-0 section 26.4.
The weighing terminal must not be installed in areas where electrostatic charging is possible
which may produce spark discharges, brush discharges or propagating brush discharges at the
front panel.
All external connections are designed for type of protection Ex 'nA'. The connection values
shown in chapter 'Installation of external connections with type of protection Ex-nA' must be
observed.
When installed in hazardous area, all peripheral devices connected to IT1000-Ex2/22 must have
appropriate Ex protection.
The metal housing of the IT1000-Ex2/22 terminal must be integrated at the PA stud into the
equipotential bonding system of the installation. Use suitable serrated washers.
Ex
The permissible range of ambient temperature is –10 °C to +40 °C.
When installing IT1000-**-E-Ex2/22 in a switch cabinet (panel-mount version), this
temperature range also applies to the inside the switch cabinet.
Furthermore, for use and installation all requirements as per EN 60079-14 must be observed.
Any other but the intended use of the equipment, modifications and extensions must not be
made without the explicit approval of the manufacturer and is deemed not to be intended. Part
of the intended use is observance of this Technical Manual and the adherence to inspection and
maintenance instructions and intervals. The manufacturer does not assume any liability
whatsoever for damages resulting from non-intended use. Any risk resulting from non-intended
use is solely borne by the user.
The weighing terminal and all connected components must be integrated into a common
system of equipotential bonding.
Protective covers or other parts covering the housing completely or in part must not be used.
The housing must be protected against permanent exposure to ultraviolet radiation.
18 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
4 Assembly
It must be made sure that installation of the weighing terminal is only carried out when all
power to the instrument is disconnected and no potentially explosive atmosphere is present.
Ex The weighing terminal must not be installed in areas where electrostatic charging is possible
which may produce spark discharges, brush discharges or propagating brush discharges at the
front panel.
Installation must be made in compliance with the generally accepted rules of technology. In particular
for all work carried out on electrical installations the specific safety regulations must be observed.
The weighing terminal may be installed in Ex zone 2 or 22. The terminal must not be damaged. The
place of installation must be clean.
The IT1000-**-W-Ex2/22 (wall-mount / desk-top version) has mounting brackets for wall or table-top
installation. They must be fixed with suitable screws which are not part of the supply. All screws must
be tightened securely.
The IT1000-**-E-Ex2/22 (panel-mount version) has screw clamps to fix the terminal in the cutout of
the switch cabinet. To connect the bonding conductor, an M5 stud is provided at the rear or bottom of
the housing. Use serrated washers to prevent the nuts from becoming loose.
The permissible temperature range is –10 °C to +40 °C at a max. of 95% relative humidity (non
condensing). Direct exposure to ultraviolet radiation must be avoided.
For wall-mount installation the unit can be fixed at the wall first and then the cables can be installed
with the front lid open.
Before the device is put into operation, the housing must be closed and tightened securely with the
hexagonal nuts provided. The nuts must be tightened with a torque of 1.1 Nm.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 19
5 Installation
5.1 General
This apparatus must only be installed by qualified personnel who are familiar with all regulations
applicable to the different types of protection and the methods of installation, as well as the
Ex general principles for the classification of zones. This level of competence must correspond to
the work that is to be carried out and appropriate training should be provided on a regular basis.
When the weighing terminal is to be installed in Ex zone 2 / 22 the regulations as per
EN 60079-14 must be adhered to.
The weighing terminal must not be installed in areas where electrostatic charging is possible
which may produce spark discharges, brush discharges or propagating brush discharges at the
front panel.
It must be made sure during installation of the weighing terminal that all power to the
instrument is disconnected and that no potentially explosive atmosphere can be present at any
time.
Installation must be carried out in compliance with applicable DIN/VDE regulations. Also, all
country-specific regulations must be observed. Connection of supply voltage must be made in
compliance with VDE 0100 and VDE 0160.
When the instrument is permanently connected to the power supply, an easily accessible
separator must be included in the supply circuit! It must either be located in safe area or -if
installed in Ex area- have appropriate Ex protection for the place of installation.
All cables are led into the housing through Ex cable glands. Strip cable jacket only as long as
required for the wires to reach the screw terminals. Use wire end ferrules with plastic collar on
Ex stranded cable and avoid protruding wires. When assembling the cable glands make sure that
shield of cable is connected inside the cable gland. Observe instructions in section 'Connection
of cables'.
Before the device is put into operation, make sure that the housing is closed and tightened
securely with all hexagonal nuts provided (torque: 1.1 Nm).
In compliance with EN 60079-14 the apparatus must be integrated into the equipotential
bonding system of the installation. The bonding conductor must have a cross section of min.
Ex 4 mm². To connect the bonding conductor, an M5 stud with serrated washer is provided at the
rear or bottom of the housing.
5.3 Shielding
Use only shielded connection cables. The shield must be connected at both sides in the cable
glands. It is mandatory to additionally install equipotential bonding as per EN60079-14
Ex paragraph 'additional requirements for the type of protection i-intrinsic safety' / section
'grounding of conductive shields' / exception b). The bonding conductor must have a cross
section of min. 4 mm².
20 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Cable jacket Shield Compression nut Plastic insert Bushing Counter nut
Use crimped wire end ferrules with plastic collar on stranded cable and avoid protruding wires.
Ex
All connecting cables must be firmly installed to avoid tension load at the cable glands and to
comply with the requirements of EN60079-14 paragraph 'cables and lines for permanent
Ex installation'.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 21
L1 N PE
Terminal KL1 Terminal KL2
110 V (–15%) to 240 V (+10%),
50/60Hz
The connection of the supply voltage at the free cable end of the weighing terminal IT1000-AC-
*-Ex2/22 is designed for type of protection Ex-nA. The listed nominal connection values must
Ex be adhered to.
For the connection of the line cord in hazardous area a suitable method must be chosen in
compliance with EN60079-0 (e.g. Ex-e, Ex-d). A suitable protection against overload and short
circuit in compliance with VDE regulations must be provided.
DC/DC converter
+V 0 V PE
Terminal KL1 Terminal KL2
12–30 VDC
The connection of the supply voltage to the weighing terminal IT1000-DC-*-Ex2/22 is designed
for type of protection Ex-nA. The listed nominal connection values must be adhered to.
Ex For the connection of the supply cable in hazardous area a suitable method must be chosen in
compliance with EN60079-0 (e.g. Ex-e, Ex-d). A suitable protection against overload and short
circuit in compliance with VDE regulations must be provided.
Note: When the IT1000-BATT terminal is supplied from an external battery, under '\ Service Menu \
General \ Batt' the respective type must be chosen, this choice affects the automatic power off
switching when battery is low:
When battery is low, this is indicated with a flashing symbol at the right hand side of the display. The
operator should then recharge the battery.
The charging of the rechargeable battery is indicated by a flashing triangle (version 2.29 or higher).
To avoid deep discharge of the battery, the IT1000-BATT is automatically switched off when the
voltage falls below the min. value. Before this happens, the message 'Low Battery' is shown for
approx. 2 min. If the unit is switched on again, 'Low Battery' is shown for approx. 2 minutes and then
the terminal is automatically switched off again to avoid deep discharge damage to the battery.
24 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
The following rechargeable batteries can be used for the supply of the weighing terminal:
1) AkkuBox Ex; 12 VDC for use in Ex zone 2 and 22
Article-No. E3AKK002 / E3AKK003
2) The use of alternative 12 V batteries is possible. However, when used in hazardous
area they must have appropriate ex protection. The requirements of EN 60079-14 must
be observed.
The operating instructions of the rechargeable battery must be observed.
1. Open protective caps before plugging and check connector/socket for damages.
2. Align male/female connectors along red arrows and insert into groove.
1) 2)
3) 4)
Terminal Assignment
1 +Excitation
2 – Excitation
3 +Sense
4 – Sense
5 +Signal
6 – Signal
Terminal Assignment
1 +Excitation
2 – Excitation
3
4
5 +Signal
6 – Signal
28 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
W&M approved applications require that the calibration parameters be protected against unauthorized
modifications. To that effect the jumper can be sealed with thread and lead seal or adhesive label.
Peripheral devices are connected at CPU1000 screw terminal KL3. The connection is designed
compliant to type of protection 'Ex-nA'. The following connection values must be observed:
Ex
SIM-RS232:
CPU1000 Description Nominal value
Terminal KL3
1 GND ±12 VDC
2 — not used
3 — not used
4 — not used
5 TxD ±12 VDC
6 RTS ±12 VDC
7 RxD ±12 VDC
8 CTS ±12 VDC
30 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
SIM-RS485-2-wire:
CPU1000 Description Nominal value
Terminal KL3
1 — not used
2 — not used
3 — not used
4 — not used
5 A (TxD+/ RxD+) 5 VDC
6 B (TxD– / RxD–) 5 VDC
7 A (TxD+/ RxD+) 5 VDC
8 B (TxD– / RxD–) 5 VDC
SIM-RS485-4-wire:
SIM-RS485-OPTO:
CPU1000 Description Nominal value
Terminal KL3
1 — not used
2 — not used
3 — not used
4 — not used
6 TxD+ 5 VDC
7 TxD– 5 VDC
8 RxD+ 5 VDC
9 RxD– 5 VDC
SIM-20mA:
CPU1000 Description Nominal value
Terminal KL3
1 — not used
2 — not used
3 — not used
4 — not used
6 TXIN 12 VDC / 20 mA
7 TXOUT 12 VDC / 20 mA
8 RXIN 12 VDC / 20 mA
9 RXOUT 12 VDC / 20 mA
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 31
If termination and pull-up / pull-down resistors are to be used in hazardous area, specific
requirements apply. Contact SysTec service for details.
Ex
For the installation of connection cables for serial interfaces please follow the recommendations listed
below:
• Install data cables to prevent capacitive or inductive interference from other cables, machines and/or
electrical devices that could interrupt data transmission and lead to loss of data.
• Non-factory made cables must comply with the following specification::
Shielded twisted pair, e.g. LIYCY 3 x 2 x 0.25 mm², shield to be grounded on both
sides;
Resistance ≤ 125 Ω/km
Cross section 0.25 mm2–0.5 mm2
Capacitance ≤ 130 nF/km
Cable length 20mA max. 100 m
Cable length RS232 max. 15 m
Cable length RS485 max. 1200 m
Impedance RS485 approx. 150 Ω
Nominal voltage ≥ 250 V
34 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Note: If an I/O module is inserted in the PIM socket, the SIM socket must remain free!
Peripheral devices are connected at CPU1000 screw terminal KL3. The connection is designed
compliant to type of protection 'Ex-nA'. The following connection values must be observed:
Ex
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 35
CPU1000
Signal Function Nominal value
Terminal KL3
1 OV 0 V auxiliary voltage 10 VDC / 15 mA
2 +10 V +10 V auxiliary voltage 10 VDC / 15 mA
3 IN0 Input 0 24 VDC / 7 mA
4 IN1 Input 1 24 VDC / 7 mA
5 In– Gnd for inputs 24 VDC / 14 mA
6 OUT0 Output 0 24 VDC / 100 mA
7 OUT1 Output 1 24 VDC / 100 mA
8 OUT+ External supply for outputs 24 VDC / 200 mA
Note: The internal 10 VDC supply (KL3, terminal 2) may be used to connect switches and push
buttons to the digital inputs (15 mA max.). External devices connected to the digital outputs must
always be supplied from an external 24 VDC power supply. If connected components are installed in
hazardous area, they must comply with appropriate type of protection, e.g. 'Ex-d'.
36 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Peripheral devices are connected at CPU1000 screw terminal KL3. The connection is designed
compliant to type of protection 'Ex-nA'. The following connection values must be observed:
Ex
Note: The internal 10 VDC supply (KL3, terminal 2) may be used to connect switches and push
buttons to the digital inputs (15 mA max.). External devices connected to the digital outputs must
always be supplied from an external 24 VDC power supply. If connected components are installed in
hazardous area, they must comply with appropriate type of protection, e.g. 'Ex-d'.
38 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Peripheral devices are connected at CPU1000 screw terminal KL3. The connection is designed
compliant to type of protection 'Ex-nA'. The following connection values must be observed:
Ex
7 Commissioning
Prior to start up follow the check list below:
Check to make sure that during commissioning no potentially explosive atmosphere can be
present.
Ex
Check whether the conditions concerning the intended use are fulfilled (see section 'Intended
use').
Check assembly as described in 'Assembly'.
Check of installation (equipotential bonding, connection of external components) as described in
chapter 'Installation'.
Check whether installation has been carried out in compliance with EN 60079-14.
Check whether housing is thoroughly closed with all hexagonal srews tightly fastened (torque:
1.1 Nm).
Check to make sure that the components connected to the outputs and interfaces (valves,
movable parts) cannot cause any damage, and whether they comply with applicable Ex
regulations when they are installed in hazardous area.
Switch on supply voltage.
Configure weighing terminal (interface parameters...) in the Service Mode as described in
section 'Service Mode'.
Calibrate scale as described in Calibration Instructions.
Test inputs/outputs, serial interface and scale function in Service Mode as described in section
'Hardware Test'.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 41
8 Installation Examples
Shown below are three typical system configurations with the weighing terminal IT1000-Ex2/22.
These are principal schematics depicting the options to connect peripheral devices and the connections
in hazardous area as well as in safe area. All devices installed in hazardous area must be designed
compliant to an appropriate type of protection. All sections of this manual and the installation standard
EN 60079-14 must be complied with in full detail.
9 Service Mode
9.1 General
The Service Mode is a program for configuration, calibration and hardware test of the IT1000 weighing
terminal. The following sections give an introduction on how to operate the terminal via keyboard and
display and describe the individual functions of the Service Mode.
Notes:
• IT1000 and its associated equipment must be installed, adjusted and maintained by qualified
personnel only!
• Before accessing the Service Mode all peripheral devices must be installed and configured!
• Access to the Service Mode is protected by the Service Password.
• Inappropriate changes of Service Mode settings may cause malfunction of the instrument!
and Press simultaneously in the initial step to reset the terminal (warm start).
~ = Motion
Key Description
F key Show tare weight, enter target value
↑ key Back to previous program step
Total key Show totals
Total
Clr Clr key Delete parameters (Clr)
Tare key Autotare scale, or clear tare when scale is tared (repetitive tare is also
possible)
+1
+1 key Choose menu item or parameter (+1)
Zero key Set gross weight to zero (only within zero setting range)
0
0 0 key Append '0' in parameter entry (0)
Prompts or entries that apply only under certain conditions are shown in an extra frame. The condition
is shown in bold face in the upper left hand corner of the frame:
Wrong password entered:
Confirm entry
Confirm entry
Press to shift decimal point to the left, press repeatedly until desired position
is reached
Confirm choice
Confirm choice
Confirm choice
48 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
11.9kg Press Tare key once more to clear tare and return to
display of gross weight (function depending on
Service Mode settings)
9.5 Power Up
When the terminal is switched on the power up messages are displayed and the program proceeds to
its initial step.
15.03.04 Date
15:06 Time
11.9kg Initial step of operating sequence (in the operating modes CHECK and FILL
first the target value must be entered).
and Press simultaneously in the initial step to reset the terminal (warm start).
Servicemode
Invalid Entry Press Print key twice to return to password entry and try again.
Press F key to exit Service Mode and store settings, return to normal
Saving...
operation.
50 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Part 2
52 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
10 Calibrate Scale
10.1 Entering Calibration Mode
Prior to power up the jumper W1 on the ADM board must be removed. Only in this position can the
changed parameters be saved in memory after the calibration.
Calibration released
Press Zero and Tare key simultaneously to enter Adapt Mode (version V1.44 or higher) to optimize
performance of digital filtering.
Linearizat. Linearization
On exiting the calibration, the entered or modified parameters are either stored or ignored:
or
Save Cal.? 0 Exit calibration, ignore all changes, do not save data (o).
or:
'Two Intervals' selected:
or:
'Three Intervals' selected:
*) Intervals for multiple-range or multi-interval scales must be specified with identical number of trailing
decimals for all intervals. If -for instance- a scale is to be calibrated with 3 ranges and increment
sizes of 2 kg, 1 kg and 0.5 kg, these increments must be entered as 2.0 kg, 1.0 kg and 0.5 kg.
56 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Fixp. active Warning: Linearization points are already set that may effect the following
calibration. Display for approx. 2 sec.
Geo Value 20 Entry of Geo Value (0 to 31). Enter the applicable Geo Value of the calibration
location (see also chapter Geo Values). If Geo Value of place of installation is
different, that can be changed later.
Cal.Zero? 1 If calibration at zero is to be carried out (1), unload scale before proceeding; if
zero is already calibrated, this step can be skipped (0).
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 57
Cal.Load? 1 If calibration at load is to be carried out (1), load test weight; if load is already
calibrated, this step can be skipped (0).
For best results use the highest possible calibration load.
To store calibration parameters exit group 'Span Adjust' and answer question 'Save Cal.?' with '1'.
Note: If changes of calibration parameters are to be ignored, setup must be exited and question
answered with '0' before any other group of the calibration may be called up.
58 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
10.4 Linearization
Certain weighing inaccuracies are the result of the inherent non-linearity of the load sensor. Up to 6
linearization points can be set freely to compensate this imperfection. Linearization points should be
entered at the points of the weighing curve where the greatest deviations (i.e. max. non-linearity) are
experienced.
Linearization points must be smaller than the max. weighing range. The linearization algorithm interpret
the sections between two neighboring linearization points as a straight line.
X10: 150.00 Display of actual weight with tenfold resolution (for verification).
New Lpt.? 1 Continue with next linearization point (1) or exit linearization (0).
or
Linearization points already stored:
New Lpt.? 1 Enter new linearization point with (1) or exit linearization (0).
Lpt.1 1500.0 Choose linearization point that is to be deleted and press Clr key.
New Lpt.? 1 Enter new linearization point with (1) or exit linearization (0).
Note: A new fixpoint can be inserted between existing ones, after that all fixpoints are renumbered in
rising order.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 59
Fixp. active Warning: Linearization points are already set that may effect the following
calibration. Display for approx. 2 sec.
Zero: 0.00 Display of the new zero point with tenfold resolution (for verification).
Z-mV 0.05701 Display of rated signal of the new zero point in mV/V
10.6 Adaptation
In this group parameters are entered to adapt the scale to its environment.
Incl.Swi.? 0 Connection of incline sensor for mobile weighing on parallel input IN1,
options:
0 = no incline sensor
1 = incline sensor connected
• If an incline sensor is connected and the permissible inclination exceeded, the display of the
indicator is blanked. Instead of the weight only a slash '/' is shown.
• When an incline sensor is connected, external taring via input IN1 is not possible.
Underld.20d 1 Display blanking for gross weights smaller than –20d (under zero):
0 Disabled
1 Enabled (default)
For applications compliant to OIML R76 underload blanking must be enabled
X10: 7.77 Displays the actual weight with tenfold resolution (for information only).
Set zero
62 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Cal.Zero? 1 If calibration at zero is to be carried out (1), unload scale before proceeding; if
zero is already calibrated, this step can be skipped (0).
S1-mV 9.99999 Enter rated signal of first loadcell (see test certificate of loadcell).
Continue with next loadcell, after the last loadcell continue in the next step.
Note: The captured values for Zero and Load are automatically entered as calibration values and can be
verified in group 'Span Adjust'.
64 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
In this group the setting of parameters can be checked that are relevant for W&M approved scales.
Any deviation is indicated.
(E=Error, (W=Warning)
*) If setting for Power Up Zero is unequal ±2 % or ±10 %, only a warning is shown, because
depending on the specific application also the setting 'Off' maybe acceptable, e.g. for hopper
scales.
W&M Setup ok If all settings comply with W&M regulations a corresponding message is
displayed.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 65
When a Multiple Range Scale is loaded, the scale display changes automatically from a lower to a
higher range. During unloading of the scale, the scale display remains in the higher range. If the scale
was tared (weighing in net mode), on return to gross zero the tare is only cleared when the Tare key is
pressed. Only after clearing the tare, the display returns to gross mode and switches back to the
smallest range.
Automatic switching back to the smallest range on reaching zero is only carried out if the scale was
not tared (weighing in gross mode).
Example of a weight curve for loading and unloading a Multiple Range Scale showing the weight and
the displayed scale interval:
When the scale is loaded further after it was tared, the tare weight is rounded automatically to the
next scale interval as soon as the display is switched from one range to the next. The max. tare value
corresponds to the capacity in the highest range.
The Multiple Range Scale provides weighing of light and heavy loads on the same scale with a high
degree of accuracy. Since resolution -and loadcell signal- in each range (e.g. 3,000 increments) is
considered separately, use of inexpensive loadcells is possible.
As a restriction for the configuration of scale intervals, the decimal point location must be identical for
all ranges. Example: Use 0.5 kg and 1.0 kg instead of 0.5 kg and 1 kg.
The max. permissible resolution of the loadcells (e.g. 3000d) must not be exceeded in any of the scale
ranges, and for W&M approved applications the signal must not be smaller than 0.33 µV/e.
68 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
The scale display changes automatically from one interval range to the next during loading and
unloading of the scale.
Example weight curve for loading and unloading a Multi Interval Scale showing the weight and the
scale interval displayed:
After taring the scale, the scale display shows the net weight with the smallest scale interval. Small
weights are shown with the highest accuracy, even when the scale is already loaded to a higher
weighing range. Autotare is possible up to full capacity, however, manual tare must not exceed the
capacity of the smallest range.
For the configuration of a multi-interval scale all capacities and intervals can be chosen freely.
*) Depending on the characteristics of a specific application, also the setting 'Off' may be acceptable
(e.g. for hopper scales).
Under menu item 'W&M Info' the setting of parameters and the compliance with W&M requirements
can be checked.
The scale settings are stored in a powerfail safe EEPROM and protected by inserting jumper W1 which
can be sealed by W&M to prevent unauthorized modification.
70 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
11 Adapt Mode
The Adapt Mode is called up in the Service Mode from the step 'Calibration' by pressing
simultaneously the Zero and Tare keys (version V1.44 or higher).
The Adapt Mode provides options to optimize the digital filtering of the captured weight:
Calibrate
Damping 0 Damping
With this parameter the 'Filter Size' as set in the calibration can be changed
by up to +5 or –5 steps. Default setting is 0.
–5 to +5: Adaptation of filter size
12 Interface Configuration
In the group 'Interfaces' of the Service Mode, settings are entered for transmission speed, format and
protocol of the serial interface, which can be configured either for connection of a PC or a printer.
Settings must comply with the configuration of the corresponding peripheral device. For applications
with a different hardware configuration (e.g. remote display), the assignment of the interface can be
changed.
End Char __3 Enter end character as decimal value, (e.g. 3 = ETX)
For data transmission to the PC the ACK/NAK protocol is used. For a description of this protocol refer
to chapter 'Data Transmission'.
72 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
S1-0 Length 0 Control sequence for print attribute Special 1 On (to be output before the field
contents), for use with the print format generator.
Enter length of sequence (0–4, 0=Off)
Char 1 __0 ASCII code for control character, if length of sequence in the previous step
was entered greater than 1, enter up to four control characters, one after the
other.
S1-1 Length 0 Control sequence for print attribute Special 1 (output after the field)
S2-0 Length 0 Control sequence for print attribute Special 2 On (output before the field)
S2-1 Length 0 Control sequence for print attribute Special 2 (output after the field)
S3-0 Length 0 Control sequence for print attribute Special 3 On (output before block 'Cycle')
S3-1 Length 0 Control sequence for print attribute Special 3 Off (output after block 'Cycle')
Note: To configure a formfeed after the cyclic part, specify 'Length 1' and
'Char 1 = 12' (Epson compatible printers only).
S4-0 Length 0 Control sequence for print attribute Special 4 On (output before block 'Totals')
S4-1 Length 0 Control sequence for print attribute Special 4 Off (output after block 'Totals')
Note: To configure a formfeed after printing totals, specify 'Length 1' and
'Char 1 = 12' (Epson compatible printers only).
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 73
Field-No. _1 To exit print format generator, press ↑-key in the first step of the sequence.
To configure a form feed after the cyclic part or after printing totals, set parameter S3-1 and S4-1
appropriately.
14 General
In this group parameters are entered to select the language for prompts, the operating mode and the
way the outputs are controlled (analog and digital).
Unit: g Choose unit sign that is to appear on the remote display, options: g, kg, t
Not for operating mode Online, Online MP85, Online BDI, Online MT-SICS or Online MTDeman:
CO_DLY 999.99 Enter delay in sec between the transmission of two consecutive data strings
over the continuous output, (default = 0.2sec).
Light Off 999 Backlighting is switched off after time when scale is not in use, enter time in
minutes;
0 = backlighting is not switched off
This function is not supported by the current program version; ignore and
Power Off 999
continue.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 77
15 Hardware Test
WARNING
Exercise utmost care when making checks, tests and adjustments that can actuate movable
parts such as feeding devices, gates, flaps, conveyors, etc. Make absolutely sure that nobody
is within reach of movable parts.
Failure to observe this precaution could result in bodily injury!
PI: 10 PO: 01 Display state of the two optional digital inputs (PI) and outputs (PO),
1 = input/output set.
With the Tare key the output 1 and with the 0 key the output 0 can be set /
reset.
The picture on the left shows the state:
input 0 = Off input 1 = On
output 0 = On output 1 = Off
SIM ok Test of serial interface 'ok' or 'not ok' (RS232 and RS485-4-wire);
RS232: jump lead from terminal 1 to 3 and from terminal 2 to 4 (connect RTS
with CTS and TxD with RxD).
RS485.4: jump lead from terminal 1 to 3 and from terminal 2 to 4 (connect
TxD+ with RxD+ and TxD– with RxD–).
Note: Test of RS485 2-wire and 20 mA interfaces is not possible.
mV/V 0.0499 Display of analog input signal for checks and plausibility test.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 79
Reset OK? 1
Note: With (1) all parameters are overwritten and default values are set! It is strongly recommended
that you print your current print format settings before carrying out a reset!
18 Protocol
Print all current parameter settings.
Printing...
Example:
Service Parameters
------------------
Interface
9600 Baud
8 Databits
No Parity
No Control
XOR All Char
StartChar 2
End Char 3
Prt: Tab=Spaces
S1-0 Length 0
S1-1 Length 0
S2-0 Length 0
S2-1 Length 0
S3-0 Length 0
S3-1 Length 0
S4-0 Length 0
S4-1 Length 0
General
German
DD.MM.YY
Online
Tar.Free
O0: LT S1
O1: LT S2
TareKey: G/N
Term.No. 1
Light Off 0
Power Off 0
Format
19 Supervisor Mode
In the Supervisor Mode parameters are entered during normal operation. The Supervisor Mode can be
called up from the first step of the application program (weight display).
Supervisor Note: If no realtime clock module is installed, date and time must be entered
after every power up!
Cons.-No.9999 Enter start value of consecutive-No. for printout (is increased automatically for
totalizing)
Not for operating modes Online, Online MP85 or Online BDI, not if continuous output is enabled:
Not if printer is enabled, not for operating mode Online, Online MP85 or Online BDI and not if
continuous output is enabled:
W&M Setup 0
W&M Setup 1
Entered or changed parameters are stored when the Supervisor Mode is exited.
Saving... Exit Supervisor Mode, store changes and return to normal operation.
84 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
20 Operating Modes
20.1 Operating Mode 'BASIC'
Note: Since IT1000 has only one socket for interface extensions, either a SIM module to connect a
printer or a PC or a PIM module for digital I/Os can be used, i.e. output on printer or PC and exchange
of signals via digital I/Os is not possible at the same time.
• Comparison of weight and control of the outputs A0 and A1 is active in the background,
independently of data entry and printing. Configuration of the outputs is made in Service Mode,
group 'General'.
• Depending on the configuration in the Service Mode, the setpoints Sp1 and Sp2 can be used as
weight thresholds or as tolerance check, (see 'Assignments of Outputs'). The values for Sp1 and
Sp2 are entered in the Supervisor Mode.
• If an incline sensor is connected, the input E1 is not available for other functions.
• The function 'Totalizing' can be disabled in Supervisor Mode in the step 'Totals?'. Print- and Total-
key are then locked.
• By pressing the F key, the current tare weight can be displayed.
The display shows ' 8.0kgT '.
• If the F key is pressed once more while tare weight is displayed, the actual weight is shown with
tenfold resolution (x10 function) and the display shows 'X10______119.03'. Press F key again to
return to initial step or display returns to initial step automatically after 5 sec.
• The inputs E0 and E1 work in parallel to the function keys as follows:
Note: Since IT1000 has only one socket for interface extensions, either a SIM module to connect a
printer or a PC or a PIM module for digital I/Os can be used, i.e. output on printer or PC and exchange
of signals via digital I/Os is not possible at the same time.
86 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
• The program automatically optimizes the average piece weight. That makes it possible to start with
a small number of counted reference parts, then add -or remove- more parts (which need not be
counted) until the display shows 'Opt. Pcs.' This ensures maximum accuracy for the subsequent
counting cycle.
• If too many parts are placed on the scale, optimization is not possible and the OPT signal on the
display goes out.
• By default the number of reference parts is 10, but it can be changed freely by the operator.
• Parts can be counted into an empty container or removed from a filled one.
• Comparison of weight and control of the outputs A0 and A1 is active in the background,
independently of data entry and printing. Configuration of the outputs is made in Service Mode,
group 'General'.
• Depending on the configuration in the Service Mode, the setpoints Sp1 and Sp2 can be used as
weight thresholds or as tolerance check, (see 'Assignments of Outputs'). The values for Sp1 and
Sp2 are entered in the Supervisor Mode.
• The inputs E0 and E1 work in parallel to the function keys as follows:
Signal Start /
rising edge input E0
Counting and Summing
• If an incline sensor is connected, the input E1 is not available for other functions.
• The function 'Totalizing' can be disabled in Supervisor Mode in the step 'Totals?'. Print- and Total-
key are then locked.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 87
Note: Since IT1000 has only one socket for interface extensions, either a SIM module to connect a
printer or a PC or a PIM module for digital I/Os can be used, i.e. output on printer or PC and exchange
of signals via digital I/Os is not possible at the same time.
With a PIM module installed inputs and outputs are assigned as follows:
• Minus threshold = target – setpoint S1
Plus threshold = target + setpoint S2
• Output A0 is used to indicate the result 'Weight ok'
• Output A1 is used to indicate the result 'Out of tolerance'
• Inputs E0 and E1 can be used in parallel to the function keys as follows:
• A checkweighing cycle is started when the scale is loaded with more than 10 % of target weight
and no-motion is detected. Thereafter the output signal corresponding to the result of the check is
set, and it remains on until the weight falls below the threshold of 10 % of target, e.g. the test
object is removed from the scale. Then the output signal is reset and a new cycle can be started.
• If an incline sensor is connected, the input E1 is not available for other functions.
• The function 'Totalizing' can be disabled in Supervisor Mode in the step 'Totals?'. Print- and Total-
key are then locked.
Note: With the setting 'PIM = 1IN / 3Out' in group 'General' of the Service Mode, the assignment of
I/Os is changed as follows:
• Output A0 = 'weight in — zone'
• Output A1 = 'weight ok'
• Output A2 = 'weight in + zone'
• Input E0 = zero setting (if gross <0) or taring (if gross > 0).
The command is only carried out when the scale is settled!
• Setpoint 1 = minus tolerance, setpoint 2 =plus tolerance.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 91
Press to display tare value for approx. 5 sec (display falls back automatically after time).
Command
Field Byte Char's Description Example
RN 1 2 Command RN
Scale-No. 3 1 optional 1
In total: 3
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 93
Response
Field Byte Char's Description Example
Error code 1 2 00 = no error, see table 'Error Codes' 00
Scale status 3 2 First digit: 00
0 = scale settled
1 = scale in motion
Second digit:
0 = gross positive
1 = gross negative
Date 5 8 Date, format as 02.05.05
per configuration
Time 13 5 Time, format 14:30
HH:MM)
Ident-No. 18 4 Ident number (non-significant digits are ___1
transmitted as space character)
Scale-No. 22 1 In IT1000 always 1 1
Gross 23 8 Formatted as per configuration (non- __430.00
significant digits are transmitted as space
character)
Tare 31 8 Formatted as per configuration (non- ___30.00
significant digits are transmitted as space
character)
Net 39 8 Formatted as per configuration (non- __400.00
significant digits are transmitted as space
character)
Unit 47 2 kg, g , t or lb, for g and t: g_
_ = second character is space
Tare code 49 2 PT = Preset Tare PT
_T = Autotare
__ = scale not tared, (_ = space character)
Weighing range 51 1 Weighing range only for multiple-range scales, 2
otherwise space characters
Terminal-No. 52 3 As specified in group 'General' of Service 001
Mode
Checksum 55 8 CRC16 checksum (non-significant digits are ___45678
transmitted as space character)
In total: 62
94 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Command
Field Byte Char's Description Example
RM 1 2 Command RM
Scale-No. 3 1 optional 1
In total: 3
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 95
Response
Field Byte Char's Description Example
Error code 1 2 00 = no error, see table 'Error Codes' 00
Scale status 3 2 First digit: 00
0 = scale settled
1 = scale in motion
Second digit:
0 = gross positive
1 = gross negative
Date 5 8 Date, format as 02.05.05
per configuration
Time 13 5 Time, format 14:30
HH:MM)
Ident-No. 18 4 Always 0 (non-significant digits are ___0
transmitted as space character)
Scale-No. 22 1 In IT1000 always 1 1
Gross 23 8 Formatted as per configuration (non- __430.00
significant digits are transmitted as space
character)
Tare 31 8 Formatted as per configuration (non- ___30.00
significant digits are transmitted as space
character)
Net 39 8 Formatted as per configuration (non- __400.00
significant digits are transmitted as space
character)
Unit 47 2 kg, g , t or lb, for g and t: g_
_ = second character is space
Tare code 49 2 PT = Preset Tare PT
_T = Autotara
__ = scale not tared, (_ = space character)
Weighing range 51 1 Weighing range only for multiple-range scales, 2
otherwise space characters
Terminal-No. 52 3 As specified in group 'General' of Service 001
Mode
Checksum 55 8 CRC16 checksum (non-significant digits are ___45678
transmitted as space character)
In total: 62
96 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
21.4 Taring
TA Automatic Tare
The TA command performs automatic taring.
Automatic taring is possible only if scale is settled. If no-motion cannot be detected within 10 seconds,
the command is aborted and error code <15> is returned to the host. If required the host must then
repeat the command.
Command
Field Byte Char's Description Example
TA 1 2 Command TA
Scale-No. 3 1 optional 1
In total: 3
Response
Field Byte Char's Description Example
Error code 1 2 00 = no error, see table 'Error Codes' 00
In total: 2
TM Manual Tare
The TM command is used to tare the scale with a value transferred from the host computer.
The tare value may include a decimal point or comma. The terminal rounds the tare value to the
increment size of the scale. If the tare value exceeds the weighing range of the scale, error code
<15> is returned to the host.
Command
Field Byte Char's Description Example
TM 1 2 Command TM
Tare Value 3 8 with decimal point or comma 000056,71
Scale-No. 11 1 optional 1
In total: 11
Response
Field Byte Char's Description Example
Error code 1 2 00 = no error, see table 'Error Codes' 00
In total: 2
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 97
TC Clear Tare
The tare is cleared and the scale is set to gross mode. The IT1000 always returns <00>.
Command
Field Byte Char's Description Example
TC 1 2 Command TC
Scale-No. 3 1 optional 1
In total: 3
Response
Field Byte Char's Description Example
Error Code 1 2 always 00 00
In total: 2
Command
Field Byte Char's Description Example
SS 1 2 Command SS
Scale-No. 3 1 optional 1
In total: 3
Response
Field Byte Char's Description Example
Error code 1 2 00 = no errors, see table 'Error Codes' 00
In total: 2
98 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Command
Field Byte Char's Description Example
SZ 1 2 Command SZ
Scale-No. 3 1 optional 1
In total: 3
Response
Field Byte Char's Description Example
Error code 1 2 00 (when successful) or 00
15 (in case of error)
In total: 2
Press to view actual tare value, display falls back after 5 sec.
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
@G 2 2 Command @G
ETX 4 1 End character ETX
In total: 4
Response
Field Byte Char's Description Example
STX 1 1 Start character STX
Status byte (P) 2 1 See table 'Status bytes' P
Weight 3 6 with decimal point +21,22
CR 9 1 Carriage Return (hex 13) CR
ETX 10 1 End character ETX
ACK 11 1 Acknowledge (hex 6) ACK
In total: 11
22.5 Taring
@T Automatic Tare
In response to this command the scale is autotared
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
@T 2 2 Command @T
ETX 4 1 End character ETX
In total: 4
Response
Field Byte Char's Description Example
ACK 1 1 Acknowledge (hex 6) ACK
In total: 1
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 101
@B Clear Tare
The tare is cleared and the scale is set to gross mode.
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
@B 2 2 Command @B
ETX 4 1 End character ETX
In total: 4
Response
Field Byte Char's Description Example
ACK 1 1 Acknowledge (hex 6) ACK
In total: 1
press to view actual tare value, display falls back after 5 sec.
23.3 Initialization
Note: The space characters between start character, command and end character in the example are
only inserted to improve readability. They must not be included in the transmitted data string.
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
iD 2 2 Command iD
ETX 4 1 End character ETX
In total: 4
No Response
104 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Response
Field Byte Char's Description Example
STX 1 1 Start character STX
Error status (Err) 2 1 0 Communication activated 0
ETX 3 1 End character ETX
In total: 3
23.3.3 Synchronization
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
ETX 2 1 End character ETX
In total: 2
Response
Field Byte Char's Description Example
STX 1 1 Start character STX
Error status (Err) 2 1 0 (Slave is synchronized) 0
ETX 3 1 End character ETX
In total: 3
Command Description
STX LD ETX Read weight (displayed value)
STX LG ETX Read gross weight
STX LN ETX Read net weight
STX LT ETX Read tare weight
STX t ETX Autotare
STX g ETX Clear tare
STX z ETX Set scale to zero
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 105
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
LD 2 2 Command LD
ETX 4 1 End character ETX
In total: 4
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
LG 2 2 Command LG
ETX 4 1 End character ETX
In total: 4
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
LN 2 2 Command LN
ETX 4 1 End character ETX
In total: 4
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
LT 2 2 Command LT
ETX 4 1 End character ETX
In total: 4
Response
Field Byte Char's Description Example
STX 1 1 Start character STX
Error status (Err) 2 1 0
Error code (E_Code) 3 1 See table 'Error Codes' 0
Weight status (Stat) 4 1 See table 'Weight Status' 0
Weighing range (WR) 5 1 1–3 1
Weight (Wgt) 6 8 non-significant digits are transmitted as ___12,34
space character
ETX 14 1 End character ETX
In total: 14
Note: When weights are read (displayed value, net, cutoff point) the first digit represents the
calculating sign. For a positive weight a space character is inserted, a negative value is shown with a
minus sign.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 107
23.6 Taring
t Automatic Tare
In response to this command the scale is autotared.
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
t 2 2 Command t
ETX 4 1 End character ETX
In total: 4
Response
Field Byte Char's Description Example
STX 1 1 Start character STX
Error status (Err) 2 1 0 taring successful 0
4 taring not successful
(scale in motion)
ETX 3 1 End character ETX
In total: 3
g Clear Tare
The tare is cleared and the scale is set to gross mode.
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
g 2 2 Command g
ETX 4 1 End character ETX
In total: 4
Response
Field Byte Char's Description Example
STX 1 1 Start character STX
Error status (Err) 2 1 0 clearing tare successful 0
4 clearing tare not successful
(scale in motion)
ETX 3 1 End character ETX
In total: 3
108 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Command
Field Byte Char's Description Example
STX 1 1 Start character STX
z 2 2 Command z
ETX 4 1 End character ETX
In total: 4
Response
Field Byte Char's Description Example
STX 1 1 Start character STX
Error status (Err) 2 1 0 zero setting successful 0
4 zero setting not successful
(out of zero setting range)
ETX 3 1 End character ETX
In total: 3
24 Multidrop Connection
Instead of a point-to-point connection between weighing terminal and host computer, also a multidrop
connection (RS485 4-wire) is possible in the operating modes 'Online SysTec', 'Online BDI' and 'Online
MP85'. Up to 30 terminals can be connected to one master. The slave address of the weighing
terminal is set in the group 'General' of the Service Mode. Valid addresses range from 01 to 30.
Example with PC as master and a maximum of 30 weighing terminals as slaves:
Connection:
← $1F
$01
Transparent
communication channel
between master
and slave 1
← $1F
$02
Transparent
communication channel
between master
and slave 2
← $1F
$01
...
25 Remote Display
In the operating mode 'Remote Display' the IT1000 terminal works as a remote display. In Service
Mode of the host weighing terminal the setting 'Sys.Remote' under 'Cont.out:' (continuous output)
must be selected.
Note: Setting of interface parameters of the indicator must match those of the remote display.
CAUTION
Incorrect connections may damage the terminals.
IT1000:
Group 'Interface': 9600 Baud, 8 Data Bits, No Parity, Halfduplex (RS485-4)
Note: Interface parameters (baud rate, data bits and parity) of weighing modules and remote display
must be set to identical values.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 113
W99 11.9kg Show weight of next weighing module, after the last
return to first.
27 Configuration Example
27.1 Factory Setting
As factory setting -or after a reset-
1 2 3 4
1234567890123456789012345678901234567890
1 Date 29.03.05
2 Time 15:30
3 Cons.-No 1
4 Gross 136.0kg
5 Tare 100.0kg
6 Net 36.0kg
7
8 Total 36.0kg
All fields on the printout are positioned by specifying the line- and the column-No. of the left-most
character in the string that is to be output. For field 01 with the text 'Date' that can be easily
recognized, line 1 / column 1 points to the 'D' in 'Date'.
For field 08 (gross weight of scale) that is already a little more difficult to spot, it must be taken into
consideration that weight values are internally represented as an 8-character string and that leading
zeros are suppressed. Thus the output of the gross weight indeed starts in column 14, although the
first character appears in column 17.
A similar effect applies to the unit sign that is automatically appended to the weight value. For gross
weights the unit sign has always a length of 2 characters (e.g. 'kg'), for tare weights the length is 4
characters (e.g. 'kgPT' for manual tare), while the length for the net weight sign is 3 characters long
(e.g.'kgC' for calculated net weights). If fewer characters are required for the correct representation of
the unit sign (e.g. 't' or 'kg' for autotared tare weights), the gap is filled with a corresponding number
of space characters. When a printout is drafted, the actual length of all variables must be considered
and overlapping avoided, or -depending on the type of printer- some very strange effects may appear.
Field 15 prints the accumulated net weight (starting in column 12). After that the totals memory is
cleared, the consecutive-No. reset to 1, and a new cycle can be started.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 115
To configure a formfeed after the cyclic part or after printing totals, parameters S3-1 or S4-1 must be
set accordingly.
Our recommendation is to always start the layout with paper and pencil. The typical grid pattern is 10
characters per inch in width, and 6 lines per inch in length. Appropriate forms can be found further
down in this manual.
It is not mandatory to number the fields in rising order without gaps (although this usually improves
readability). Fields may remain free by assigning the 'not used' attribute.
116 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
1 2 3 4
1234567890123456789012345678901234567890
The blank line between 'Cycle' und 'Totals' is inserted by simply starting the printout of the totals in
line 2.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 117
Field length
System variable Operating mode
(characters)
Date 8 All
Time 5 All
Cons.No. 5 All
Gross,tare, net 8 All
Unit sign gross 2 All
Unit sign tare 4 All
Unit sign net 3 All
Total gross, total tare, total net 10 All
Unit sign totals 2 All
Target 8 FILL
118 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
28 Data Transmission
If data transmission is enabled, a data string is sent to the host system after each completed weighing
cycle. This is the case, for instance, after a completed filling cycle (operating mode 'FILL') or after
confirming the result of the classification (operating mode 'CHECK').
The principal structure of a data string looks as follows:
PC → IT1000
Control Character / Data Comment
ACK Positive confirmation for correctly received data string
or
PC → IT1000
Control Character / Data Comment
NAK Negative confirmation for not correctly received data string
The IT1000 timeout delay for the reception of ACK or NAK is 6 sec. If a response cannot be received
within 6 sec or if a negative response is received (NAK), the transmission of the data string is repeated
up to 4 times. If after a total of 5 transmission attempts a response cannot be received or if only
negative responses are received, an error message is indicated on the display of the IT1000 terminal
which must be acknowledged by the operator. The operator has the choice either to start the
transmission anew by pressing the Print key (e.g. after rectifying the problem) or switch off the data
transmission by pressing the F key.
122 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
28.2.2 Ranger-A-Protocol
Sign 2 1 – or + –
Weight 3 6 with decimal point and __123.4
_ = leading underline char's
Sync 9 1 CR (0d hex) for synchronization
In total: 9
124 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
Transmission parameters are: 4800 baud, 7 data bits, even parity, no hardware handshake. Checksum
is defined as the 2's complement of the 7 low order bits of the binary sum of the 7 low order bits of all
characters preceding the checksum including STX and CR. Bit 8 of the checksum is parity over the 7
low order bits of checksum.
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 125
29.2 Maintenance
Maintenance at regular intervals must be carried out by qualified personnel (see chapter
'Installation') authorized by SysTec GmbH. At these inspections first of all it must be made sure
Ex that the housing is tight, all cables are undamaged and all screws are securely fastened.
Maintenance of scale platforms is required at regular intervals depending on use and environment. The
accuracy of scales can be affected by dirt, foreign objects, etc. and appropriate maintenance is
strongly recommended. Also recommended is the calibration with certified test weights at regular
intervals.
A function check is possible with the program Service Mode.
The battery must only be replaced by qualified personnel (see chapter 'Installation') authorized
by SysTec GmbH. Replacement must only be carried out in a de-energized state and in the
Ex absence of any potentially explosive atmosphere.
Replace only with type DS1687-5 of Dallas/Maxim.
29.4 Cleaning
• Only permit qualified personnel to operate this unit! Disconnect all power to this unit before
cleaning!
The keyboard overlay is resistant against acetone, trichlorine, alcohol, ether, nitric acid (20 %),
hexane, sulfuric acid (20 %) and all-purpose cleaning agents.
Clean the keyboard with a soft clean cloth that has been dampened with a mild window type
cleaner. Do not spray cleaner directly on the unit. Concentrated leaches or acids or pure alcohol
Ex must not be used. In order to avoid electrostatic charge, do not rub or wipe the keyboard with
a dry cloth.
If cleaning agents are used that contain leach, acid or alcohol, pure water must be used to
wash off any residue.
29.7 Repair
Ex
Only qualified service personnel (see chapter 'Installation'), authorized by SysTec GmbH, may carry out
repair work using genuine factory supplied spare parts.
29.8 De-Installation
1. Make sure that all power to the weighing terminal and all connected devices is
disconnected and that no potentially explosive atmosphere can be present.
Ex
2. Remove cabling.
3. Use suitable tool to de-install weighing terminal.
29.9 Disposal
For the disposal of the weighing terminal all country-specific and locally applicable regulations must be
observed!
128 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
30 Trouble Shooting
IT1000-Ex2/22 does not contain any customer serviceable parts! It may only be installed,
adjusted and serviced by qualified personnel!
Ex
Immediately disconnect defective units from all power sources. Only qualified service personnel
(see chapter 'Installation'), authorized by SysTec GmbH, may carry out repair work with
Ex genuine factory supplied spare parts, or serious risks for the user may evolve.
Fault finding must not be carried out as long as any potentially explosive atmosphere can be
present.
Ex If possible, try first to determine the conditions under which the problem occurs. Try to find out
whether the appearance of the difficulties can be reproduced under the same conditions.
For the systematic analysis of an unknown problem the information as listed below is required:
• Serial-No. of the unit and its peripheral components.
• Exact wording of any error message displayed.
• Type and model of peripheral devices related to the problem (e.g. scale, printer etc.)
To obtain professional assistance contact your service station stating the information listed above.
During Calibration:
Error Transm. • Host switched off or off-line, data • Check cable and connectors
cable not connected or damaged • Press Print key to repeat
transmission
• Press F key to abort transmission
31 Technical Data
31.1 General Data
Housing Stainless steel wall-mount / desk-top housing,
ingress protection IP65, weight: approx. 1.5 kg
Stainless steel panel-mount housing,
ingress protection see chapter 'Intended use', weight: approx. 1.5 kg
Temperature range Storage: –10 °C to +50 °C at 95 % max. relative humidity without
condensation,
operation: –10 °C to +40 °C at 95 % max. relative humidity without
condensation;
Safety Separation between primary and secondary circuits SELV, compliant to
EN60950, overvoltage category II;
Ex classification II 3G Ex ic nA IIC T4 Gc
II 3D Ex ic tc IIIB T50°C Dc IP54 or
II 3D Ex ic tc IIIC T50°C Dc IP65
2 — —
3 — —
4 — —
SIM-RS232
5 TxD ±12 VDC
1 — —
2 — —
3 — —
4 — —
SIM-RS485-2-wire
5 A (TxD+ / RxD+) 5 VDC
1 — —
2 — —
3 — —
SIM-RS485-4-wire 4 — —
SIM-RS485-OPTO 5 TxD+ 5 VDC
6 TxD– 5 VDC
7 RxD+ 5 VDC
8 RxD– 5 VDC
1 — —
2 — —
3 — —
4 — —
SIM-20mA
5 TXIN 12 VDC / 20 mA
6 TXOUT 12 VDC / 20 mA
7 RXIN 12 VDC / 20 mA
8 RXOUT 12 VDC / 20 mA
132 Technical Manual IT1000-**-*-Ex2/22 Rev. 9
1 OV 10 VDC / 15 mA
2 +10 V 10 VDC / 15 mA
3 IN0 24 VDC / 7 mA
4 IN1 24 VDC / 7 mA
PIM
5 In– 24 VDC / 14 mA
1 OV 10 VDC / 15 mA
2 +10 V 10 VDC / 15 mA
3 IN0 24 VDC / 7 mA
IOM 4 OUT2 24 VDC / 100 mA
(socket
SIM+PIM) 5 In– 24 VDC / 7 mA
1 — —
2 — —
3 I+ 10 V / 20 mA
4 I– 10 V / 20 mA
DAU15
5 — —
6 U+ 10 V / 20 mA
7 U– 10 V / 20 mA
8 — —
1 +Excitation 5 V / 240 mA
2 – Excitation 5 V / 240 mA
3 +Sense 5 V / 1 mA
ADM
4 – Sense 5 V / 1 mA
5 +Signal 5 V / 1 mA
6 – Signal 5 V / 1 mA
Technical Manual IT1000-**-*-Ex2/22 Rev. 9 133
32 Dimensions