Sie sind auf Seite 1von 4

Basic Research—Technology

Removing Fractured Endodontic Instruments with a Modified


Tube Technique Using a Light-curing Composite
Michael Wefelmeier, DMD,* Maria Eveslage, Dipl-Stat,† Sebastian B€
urklein, DMD,†‡
Klaus Ott, DMD,* and Markus Kaup, DMD*

Abstract
Introduction: The aim of this in vitro study was to
assess an alternative method using light-curing compos-
ite for removing fractured endodontic instruments with
F racturing of endodontic instruments is a rare but annoying complication during
root canal treatment with a reported prevalence between 1.83% (1) and 3.3%
(2). In retreatment cases, this incident occurs more often (1). Machtou and Reit
a tube technique. Methods: Two different stainless steel (3) point out that removal of the separated instruments would be the best treatment
endodontic instruments (ISO 20: Hedstrom files, K-files; option.
VDW, Munich, Germany) were cut at the diameter of Even though modern techniques and advances in vision have improved clinicians’
0.4 mm. These fragments were fixed in a vise leaving a ability to remove fractured endodontic instruments, removal may not always be
free end of 1 or 2 mm. Cyanoacrylate (Instant Fix; Henry possible or desirable. There is no sufficient information based on high-level evidence
Schein Dental, Melville, NY), dual-curing Rebilda DC about the management of separated instruments, which complicates the decision-
(VOCO, Cuxhaven, Germany), and light-curing SureFil making process (4–6). All efforts in managing this complication should be based
SDR (Dentsply, York, PA) were placed into microtubes on thorough knowledge of each treatment option, considering the success rates
(N’Durance Syringe Tips; Septodont, Saint-Maur, France) well balanced against the potential risks of leaving or removing the fragment (7). Frac-
and shifted over the instruments (n = 20 in each group). tured endodontic instruments might not directly affect the prognosis (2, 4, 6, 8) of the
After polymerization, pull-out tests were performed with a tooth because the fractured instrument itself may not directly lead to infection.
constant speed of 2 mm/min; failure load was measured However, the fractured instrument may hinder chemomechanical disinfection of the
digitally. Data were analyzed using the Kruskal-Wallis entire root canal system and thus can limit the prognosis (3, 5, 6, 9–11)
test followed by the Dunn test for pairwise comparison. depending on the stage in the root canal treatment procedure when the separation
Results: The median failure load was up to 62.5 N for occurred (6, 12, 13). Because of the different situations after instrument fracture
SDR, 35.8 N for Rebilda, and 14.7 N for cyanoacrylate, (eg, presence or absence of apical disease [6], type of tooth [14], location/length/
respectively. Both tested composites yielded significantly type of the instrument [7, 14, 15], root canal curvature [14, 16], and time of
higher values in pull-out tests than cyanoacrylate. The dis- fracture [17]), there is no clinical evidence on the force required for its removal.
connecting force was highest when light-cured composite Even the technique for the removal of fractured instruments has to be evaluated indi-
SDR was used for fixation. Removing Hedstrom files re- vidually for each different situation (5).
sulted in higher values than removing K-files. The median If removing is necessary, attempts to remove fractured instruments can lead to
force when using SDR was 79.7 N (interquartile range, ledge formation, overenlargement, canal transportation, or perforation (18). The chal-
66.0–86.8 N) in Hedstrom files and 53.3 N (interquartile lenging steps in removing fractured instruments are the minimally invasive approach
range, 47.1–58.5 N) in K-files. Conclusions: Within the and exposure (5). For clinicians, several nonsurgical treatment options are available.
limitations of this study, the use of light-curing composite Besides the ‘‘braiding technique’’ (19) in which small files are used to remove or at least
inside of the microtube was superior compared with the bypass the instrument, the use of ultrasonic devices is an effective way to expose and
use of cyanoacrylate or chemically cured composite, eventually remove fragments (14, 20). If ultrasonic procedures fail, the tube
which are being used presently. (J Endod 2015;-:1–4) technique is the next best chance to remove fractured instruments (20). In these cases,
it is helpful to be able to release as much force as possible with the minimally invasive
Key Words approach.
Endodontic instrument, fractured, removal, tube tech- For the successful use of commercial mechanical tube systems like the Masserann
nique kit (Micro-Mega, Besançon, France) and the IRS Instrument Removal System (Jadent,
Aalen, Germany), a straight-line access to the fractured instrument is necessary (21).
Even the smallest diameter of the Masserann-kit (1.2 mm) is pretty wide compared with
the average root diameter (22, 23).
From the *Department of Operative Dentistry, Universit€ats- To approach the fractured instrument, the IRS Instrument Removal System only
klinikum M€unster; †Institute of Biostatistics and Clinical
Research; and ‡Central Interdisciplinary Ambulance in the needs 0.6 mm; however, the instrument needs to be exposed at least up to 2–3 mm
School of Dentistry, Universit€atsklinikum M€unster, M€unster, (20). Alternatively, a microtube filled with cyanoacrylate or with dual-curing composite
Germany. can be shifted over the exposed end of the fractured instrument (24, 25). However, using
Address requests for reprints to Dr Michael Wefelmeier, microtubes filled with adhesive materials is associated with disadvantages when
Department of Operative Dentistry, Albert-Schweitzer-Campus
1, Building W30, Universit€atsklinikum M€unster, 48149,
compared with mechanical systems (eg, the extended cyanoacrylate may set inside the
M€unster, Germany. E-mail address: mwefel@uni-muenster.de root canal) (20). Additionally, only relatively low tensile forces are achieved (20). The
0099-2399/$ - see front matter aim of this in vitro pilot study was to compare the different well-established microtube
Copyright ª 2015 American Association of Endodontists. techniques with a new approach for instrument fixation.
http://dx.doi.org/10.1016/j.joen.2015.01.018

JOE — Volume -, Number -, - 2015 Removing Fractured Endodontic Instruments 1


Basic Research—Technology
Materials and Methods 1. Fixation material: cyanoacrylate, instrument: Hedstrom, and fixa-
In pull-out tests, the disconnecting force between 3 different tion length: 1 mm
fixation materials and 2 different stainless steel endodontic instru- 2. Fixation material: cyanoacrylate, instrument: Hedstrom, and fixa-
ments was determined. Twenty specimens were investigated in each tion length: 2 mm
group. 3. Fixation material: cyanoacrylate, instrument: K-file, and fixation
Two different endodontic instruments (ISO 20: Hedstrom files, length: 1 mm
Kerr files; VDW Dental, Munich, Germany) were cut exactly at the 4. Fixation material: cyanoacrylate, instrument: K-file, and fixation
same diameter of 0.4 mm. These fragments were fixed in a vise with length: 2 mm
an overlap of either 1 or 2 mm. Microtubes (N’Durance syringe tips; 5. Fixation material: Rebilda DC, instrument: Hedstrom, and fixation
Septodont, Saint-Maur, France) with an outer diameter of 0.85 mm length: 1 mm
and an inner diameter of 0.64 mm (22-G) were shifted over these in- 6. Fixation material: Rebilda DC, instrument: Hedstrom, and fixation
struments and fixed as shown (Fig. 1). length: 2 mm
In group 1, cyanoacrylate-based adhesive (Instant Fix; Henry 7. Fixation material: Rebilda DC, instrument: K-file, and fixation
Schein Dental, Melville, NY) was aspirated into the tubes before putting length: 1 mm
them over the endodontic instrument. For faster setting of the cyanoac- 8. Fixation material: Rebilda DC, instrument: K-file, and fixation
rylate, the tubes were stored in water for 30 minutes to guarantee a ho- length: 2 mm
mogenous setting and maximal adhesion. 9. Fixation material: SDR, instrument: Hedstrom, and fixation length:
In group 2, a dual-curing composite resin (Rebilda DC; VOCO, 1 mm
Cuxhaven, Germany) was used to fix the endodontic instrument in 10. Fixation material: SDR, instrument: Hedstrom, and fixation length:
the microtube. The setting time was 30 minutes to guarantee complete 2 mm
polymerization. 11. Fixation material: SDR, instrument: K-file, and fixation length:
In group 3, a light-curing composite resin (Surefil SDR; Dentsply, 1 mm
York, PA) was used to fix the endodontic instruments inside of the 12. Fixation material: SDR, instrument: K-file, and fixation length: 2 mm
tube. An optical fiber (Conrad Electronic SE, Hirschau, Germany) The tube on the one side and the endodontic instruments on the
with a diameter of 0.5 mm was inserted into the microtube and pushed other side were fixed in a mount. The tubes were pulled with a constant
forward until the fiber got in contact with the endodontic instrument speed of 2 mm/min, and the resulting force was measured digitally (LF
(Fig. 1). Then, the SDR was light polymerized by Smartlite PS (Dents- Plus; Lloyd Instruments, Bognor Regis, England).
ply) through the optical fiber for 1.5 minutes. The light source was
applied in contact to the fiber (Fig. 2).
After polymerization, the compound between the tubes and the Statistics
endodontic instruments was used for pull-out tests. A total of 240 sam- To compare the different instruments, instrument lengths, and
ples were prepared as follows: fixation materials in regard to the force necessary to break the

Figure 1. A schematic drawing showing the fixed instrument, metallic tube, and 2 different methods of adhesion (lower right: cyanoacrylate; Rebilda DC, dual-
curing composite; upper right: SDR, smart dentin replacement, light-curing composite).

2 Wefelmeier et al. JOE — Volume -, Number -, - 2015


Basic Research—Technology
Results
Regardless of the type of instrument or instrument length, the use
of light-curing SDR reached the highest median amount of force, which
was necessary to break the connection between the microtube and the
instrument (Table 1). For all instruments and instrument lengths, sig-
nificant differences between SDR and Rebilda as well as cyanoacrylate
were achieved (P < .0001).
In all pull-out tests with SDR and 2-mm Hedstrom files, the
connection between composite resin and the instrument did not fail.
However, this was primarily because of previous fracturing of the end-
odontic instrument itself.
Two different mechanisms of failure of the adhesive joint were
observed when using Rebilda in 2-mm Hedstrom files or
2-mm K-files. The connection between composite and the inner sur-
face of the tube failed and led to total disconnection, which was
observed in 20% (K-files) or 40% (Hedstrom) of the samples. The
increased variance resulting from this phenomenon can clearly be
seen in Table 1.
The glue or composite resin reacts differently with the 2 types of
instruments. The adhesive joint seems to be more durable in Hedstrom
files for Rebilda and SDR (eg, the median force when using SDR was
79.7 N [IQR = 66.0–86.8 N] in Hedstrom files and 53.3 N [IQR =
47.1–58.5 N] in K-files). The connection is more durable in any com-
bination of fixation materials and instruments with instrument lengths of
2 mm compared with 1 mm (Table 1).

Discussion
All tested parameters had a relevant influence on the durability of
the adhesive joint. The disconnecting force was highest when light-
cured composite SDR was used for fixation. For the dual-curing com-
Figure 2. Microtube and optical fiber to show the way of the light, which is
necessary for polymerization after shifting both over the tip of the endodontic
posite resin Rebilda, the biggest variances of values were observed.
instrument. Fixation with cyanoacrylate was the weakest (Table 1).
Fixation with cyanoacrylate led to slightly higher values for instru-
ments with a larger core diameter (K-file > Hedstrom). The strength of
adhesive joint, descriptive statistics were calculated. Values are pre- the adhesive bond seems to be higher when the layer of cyanoacrylate in
sented as median and interquartile range (IQR) throughout the text. the gap between the instrument and the tube is spread out relatively uni-
Because normal distribution could not be assumed, the 3 groups formly. Cyanoacrylate adhesives are not designed to bridge a gap
were compared using the Kruskal-Wallis (26) test followed by the >.1 mm and thus cannot create a secure adhesive connection.
Dunn test (27) for pairwise comparison applying the closed testing In general, significantly higher values in pull-out tests were
principle (28). These comparisons were performed for the 2 instru- achieved with both tested composites than with cyanoacrylate. Using
ments and 2 instruments lengths separately, and all P values were ground-twisted K-files resulted in lower values compared with
therefore adjusted by the Bonferroni method to account for multiple machined Hedstrom files with a smaller core diameter and a more pos-
testing. The multiple significance level was set to a = 0.05. Statistical itive rake angle (29), resulting in more room for the fixation material.
analyses were conducted using IBM SPSS Statistics 22 (IBM Corp, Furthermore, the angulation of the instruments’ cutting edges may affect
Somers, NY) and R Version 3.1.0 (SAS Institute Inc, Cary, NC). the resulting data. If the angle is more parallel to the direction of force

TABLE 1. Mean Force, Standard Deviation, and Range of All Pull-out Tests
Endodontic instrument Fixation length Fixation material Mean Standard deviation Range (minimum–maximum)
Hedstrom 1 mm cyanoacrylate 11.24 N 3.83 N 4.93–18.50 N
Rebilda DC 32.42 N 11.30 N 11.69–51.67 N
SDR 64.66 N 9.13 N 49.13–81.47 N
2 mm cyanoacrylate 17.69 N 7.42 N 6.16–30.88 N
Rebilda DC 55.82 N 25.51 N 14.17–96.59 N
SDR 86.15 N 4.33 N 78.91–93.60 N
K-file 1 mm cyanoacrylate 11.56 N 4.44 N 2.85–18.86 N
Rebilda DC 29.83 N 7.35 N 18.20–42.12 N
SDR 47.67 N 7.07 N 33.42–58.83 N
2 mm cyanoacrylate 27.59 N 5.55 N 18.54–36.31 N
Rebilda DC 43.20 N 17.28 N 10.34–64.36 N
SDR 59.79 N 9.45 N 45.62–76.15 N
Rebilda DC, dual-curing composite; SDR, smart dentin replacement light-curing composite.

JOE — Volume -, Number -, - 2015 Removing Fractured Endodontic Instruments 3


Basic Research—Technology
(K-files), the resistance of the connection between composite resin and Acknowledgments
fractured instrument seems to be lower.
The application of chemically polymerized Rebilda DC showed The authors deny any conflicts of interest related to this study.
significantly lower values and greater standard deviations than light-
cured SDR. This could be explained by different shrinkage of the tested References
composites (30) and the content of filler but not a lack of polymeriza- 1. Tzanetakis GN, Kontakiotis EG, Maurikou DV, et al. Prevalence and management of
instrument fracture in the postgraduate endodontic program at the Dental School of
tion or other physical properties. In their studies, axial shrinkage of SDR Athens: a five-year retrospective clinical study. J Endod 2008;34:675–8.
was 2.26% and Rebilda DC reached 2.96%. The shrinkage force 2. Spili P, Parashos P, Messer HH. The impact of instrument fracture on outcome of
amounted to 20 N for SDR in average and about 37 N for Rebilda endodontic treatment. J Endod 2005;31:845–50.
DC (30). 3. Machtou P, Reit C. Non-surgical retreatment. In: Bergenholtz G, Horsted-Bindslev P,
During polymerization, the composite seems to shrink toward Reit C, eds. Textbook of Endodontology, 1st ed. Oxford: Blackwell Munksgaard; 2003.
4. Murad M, Murray C. Impact of retained separated endodontic instruments during
the structured surface of the endodontic instrument. As a result of root canal treatment on clinical outcomes remains uncertain. J Evid Based Dent
this, the connection between the composite and the inner surface Pract 2011;11:87–8.
of the tube failed using Rebilda DC for fixation and led to total 5. Madarati AA, Hunter MJ, Dummer PM. Management of intracanal separated instru-
disconnection. These observations may elucidate the standard devia- ments. J Endod 2013;39:569–81.
6. McGuigan MB, Louca C, Duncan HF. The impact of fractured endodontic instru-
tions in Table 1. ments on treatment outcome. Br Dent J 2013;214:285–9.
Using this modified microtube technique may offer some advan- 7. McGuigan MB, Louca C, Duncan HF. Clinical decision-making after endodontic in-
tages compared with other tube techniques. Within the limitations of strument fracture. Br Dent J 2013;214:395–400.
the results of this experiment, the following aspects about the clinical 8. Panitvisai P, Parunnit P, Sathorn C, et al. Impact of a retained instrument on treat-
relevancy might be considered: ment outcome: a systematic review and meta-analysis. J Endod 2010;36:775–80.
9. Sj€ogren U, Hagglund B, Sundqvist G, et al. Factors affecting the long-term results of
1. The microtubes can be bent in any desirable direction or a Cancel- endodontic treatment. J Endod 1990;16:498–504.
10. Di Fiore PM. A dozen ways to prevent nickel-titanium rotary instrument fracture.
lier instrument (SybronEndo, Orange, CA) might be used for placing J Am Dent Assoc 2007;138:196–201.
the tubes over the instrument so that nothing will interfere with the 11. Ng YL, Mann V, Gulabivala K. A prospective study of the factors affecting outcomes of
straight line of sight a microscope requires. non-surgical root canal treatment: part 2: tooth survival. Int Endod J 2011;44:610–25.
2. Both microtubes and optical fibers are available in a wide range of 12. Simon S, Machtou P, Tomson P, et al. Influence of fractured instruments on the suc-
diameters down to 0.25 mm. Because of this fact, the size of the tube cess rate of endodontic treatment. Dent Update 2008;35:172–9.
13. Roda R, Gettleman BH. Nonsurgical retreatment. In: Cohen S, Hargreaves KM, eds.
can be adapted individually, and additional reduction of radicular Pathways of the Pulp, 9th ed. St Louis: Mosby Elsevier; 2006.
dentine is minimized. 14. Cuje J, Bargholz C, H€ulsmann M. The outcome of retained instrument removal in a
3. A circumferential staging platform facilitates the removal of frac- specialist practice. Int Endod J 2010;43:545–54.
tured endodontic instruments with ultrasonic devices or microtubes 15. H€ulsmann M, Schinkel I. Influence of several factors on the success or failure of
removal of fractured instruments from the root canal. Endod Dent Traumatol
(13). The more radicular dentin can be saved; the lower is at risk of 1999;15:252–8.
perforation (23). For this modified tube technique, high forces can 16. Suter B, Lussi A, Sequeira P. Probability of removing fractured instruments from
be transferred to the fractured instrument with an exposure of root canals. Int Endod J 2005;38:112–23.
1–2 mm. 17. Torabinejad M, Lemon RR. Procedual accidents. In: Walton RE, Torabinejad M, eds.
4. Furthermore, there are huge differences in application time. Principles and Practice of Endodontics, 3rd ed. Philadelphia: WB Saunders; 2002.
18. Choksi D, Idnani B, Kalaria D, et al. Management of an intracanal separated instru-
Although the polymerization of SDR can be controlled by the clini- ment: a case report. Iran Endod J 2013;8:205–7.
cian and is induced by light for 1.5 minutes, longer setting times for 19. Gilbert BO, Rice T. Retreatment in endodontics. Oral Surg Oral Med Oral Pathol
the other materials were necessary. Preliminary tests showed a con- 1987;64:333–8.
stant level of maximal fixation after 20 minutes for cyanoacrylate and 20. Ruddle CJ. Broken instrument removal. The endodontic challenge. Dent Today
2002;21:70–6.
Rebilda. 21. Vivekananda Pai AR, Mir S, Jain R. Retrieval of a metallic obstruction from the root
5. In addition, the polymerization of SDR only depends on the in- canal of a premolar using Masserann technique. Contemp Clin Dent 2013;4:543–6.
tensity of light, which is inside and in front of the tube. Material 22. H€ulsmann M. Methods for removing metal obstructions from the root canal. Endod
outside of the tube will not polymerize and can be removed Dent Traumatol 1993;9:223–37.
easily. 23. Yoldas O, Oztunc H, Tinaz C, et al. Perforation risks associated with the use of Mas-
serann endodontic kit drills in mandibular molars. Oral Surg Oral Med Oral Pathol
The investigation of additional rotational forces and different types Oral Radiol Endod 2004;97:513–7.
24. Andrabi SM, Kumar A, Iftekhar H, et al. Retrieval of a separated nickel-titanium in-
of endodontic instruments will have to show whether this technique is a strument using a modified 18-guage needle and cyanoacrylate glue: a case report.
meaningful rewarding addition to the standard techniques frequently Restor Dent Endod 2013;38:93–7.
used by clinicians. Further studies concerning rotary nickel-titanium in- 25. Coutinho Filho T, Krebs RL, Berlinck TC, et al. Retrieval of a broken endodontic in-
struments are necessary to elucidate if the results can be extrapolated to strument using cyanoacrylate adhesive. Case report. Braz Dent J 1998;9:57–60.
instruments with other metallurgical properties and cross-sectional de- 26. Kruskal W, Wallis WA. Use of ranks in one-criterion variance analysis. J Am Stat
Assoc 1952;47:583–621.
signs. 27. Dunn OJ. Multiple comparisons using rank sums. Technometrics 1964;6:241–52.
28. Marcus R, Peritz E, Gabriel KR. On closed testing procedures with special reference
Conclusion to ordered analysis of variance. Biometrika 1976;63:655–60.
Within the limitations of this in vitro pilot study, the use of light- 29. Himel VT, McSpadden JT, Goodis HE. Instruments, Materials and Devices. In:
curing composite resin inside of the microtube was superior compared Cohen S, Hargreaves KM, eds. Pathways of the Pulp, 9th ed. St. Louis: Mosby Elsev-
ier; 2006.
with the use of cyanoacrylate or chemically cured composite resin. 30. Taub€ock TT, Feilzer AJ, Buchalla W, et al. Effect of modulated photo-activation on
The applicable forces differed significantly (SDR > Rebilda polymerization shrinkage behavior of dental restorative resin composites. Eur J Oral
DC > cyanoacrylate, Hedstrom file > K-file). Sci 2014;122:293–302.

4 Wefelmeier et al. JOE — Volume -, Number -, - 2015

Das könnte Ihnen auch gefallen