TOSOH AUTOMATED GLYCOHEMOGLOBIN ANALYZER

HLC-723GX
HOST CONNECTION SPECIFICATIONS

Rev.2

TOSOH CORPORATION

BIOSCIENCE DIVISION
HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

Revision record

Rev. Revision date Revision content


1 Jul 25, 2011 Created
The data format of IFCC value can be reported like 999.9 (one
2 Apr 13, 2012 decimal place form) when the PERCENT DECIMAL parameter value
is set to 8.

TOSOH CORPORATION Bioscience Division Page 2/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

Contents

1. INTRODUCTION...................................................................................................... 4

2. HARDWARE SPECIFICATIONS............................................................................. 4

3. COMMUNICATION SETTING ................................................................................. 5

4. COMMUNICATION MODE ...................................................................................... 6


4.1. GX format and old model compatible format .................................................................. 6
4.2. Hi-LEVEL and RP modes ............................................................................................... 7
5. TRANSMISSION PROTOCOL ................................................................................ 8
5.1. Handshake...................................................................................................................... 8
6. COMMUNICATION FORMAT................................................................................ 10
6.1. Query ............................................................................................................................ 10
6.1.1. Sample and container type designation.......................................................... 11
6.2. Measurement results .................................................................................................... 12
6.2.1. GX format.......................................................................................................... 12
6.2.2. G8 format .......................................................................................................... 16
6.2.3. G9 format .......................................................................................................... 19
6.2.4. G7 format .......................................................................................................... 20
6.2.5. GHbV compatible format ................................................................................. 21
6.2.6. Flag level and communication......................................................................... 24

TOSOH CORPORATION Bioscience Division Page 3/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

1.Introduction
This document describes the communication arrangements used when connecting the
HLC-723GX (hereafter called analyzer) to a host computer (hereafter called host).
2.Hardware specifications
(1) Communication specifications
Item Specification
Transmission method EIA-574 (or EIA-232, RS-232C)
Asynchronous (start-stop transmission),
half-duplex
Transmission rate 1200, 2400, 4800, 9600 bps
Transmission code ASCII
Data length 7 bit, 8 bit
Parity Even, odd, none
Stop bit 1 bit, 2 bit

(2) Connections

Communication cable
Analyzer Host

Boundary of responsibility (Free point)

(3) Connector
A D-SUB 9p (male) is used as the connector to the cable on the analyzer side.
(Connector which should be attached to the analyzer side of the communication cable
is D-SUB 9s (female).)

(4) Pin assignment

Analyzer side (9 pin) Host side

Pin No.
Signal name Pin No. 9pins 25pins Signal name
case 1 Frame GND (FG)
Receive data (RxD) 2 2 3 Receive data (RxD)
Transmit data (TxD) 3 3 2 Transmit data (TxD)
Data terminal ready (DTR) 4 4 20 Data terminal ready (DTR)
Signal GND (SG) 5 5 7 Signal GND (SG)
Data set ready (DSR) 6 6 6 Data set ready (DSR)
Request to send (RTS) 7 7 4 Request to send (RTS)
Clear to send (CTS) 8 8 5 Clear to send (CTS)

TOSOH CORPORATION Bioscience Division Page 4/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

3.Communication Setting
Communication setting for analyzer can be arranged on the RS232C screen of the
analyzer.

Figure 3-1 RS232C screen

• Number of digits of sample number


SMP 3 : Handles the sample number with the last 3 digits
SMP 5 : Uses 5 digits for the sample number
SMP 8 : Uses 8 digits for the sample number (recommended)
• Number of digits of ID number(bar code)
BC 13 : Transmits the ID number using 13 digits
BC 18 : Transmits the ID number using 18 digits
BC 20 : Transmits the ID number using 20 digits (recommended)
BC NO : Does not send the ID number
• Data length
8 BIT : Sets the data length to 8 bits
7 BIT : Sets the data length to 7 bits
• Stop bit
1 BIT : Sets the stop bit to 1
2 BIT : Sets the stop bit to 2
• Parity
NONE : Sets the parity to none
ODD : Sets the parity to odd numbers
EVEN : Sets the parity to even numbers
• Baud rate
1200 : Sets the baud rate to 1200 bps
2400 : Sets the baud rate to 2400 bps
4800 : Sets the baud rate to 4800 bps
9600 : Sets the baud rate to 9600 bps
• Query with ID
: When this key is highlighted, a query with ID is executed and only
designated samples are processed.

• Automatically transmitted
: When highlighted, results are automatically transmitted.

TOSOH CORPORATION Bioscience Division Page 5/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

4.Communication mode
There are two communication modes, query mode, on which analyzer queries test
order to the host using sample ID after reading the bar-code on the sample, and
result transmission mode on which analyzer transmits result to the host every time
result is obtained. Various formats compatible with the old models are available on
the result transmission mode.

4.1. GX format and old model compatible format


Five formats, GX format (standard), G9 format, G8 format, G7 format and GHbV
A1c2.2 compatible format are available as communication formats. Select the
desired format using TRANS MODE on the PARAMETER screen of the analyzer.
These formats are compatible with HI-LEVEL mode.

Figure 4-1 PARAMETER screen(TRANS MODE)

GX FORM: GX format (default setting)


G9 FORM: G9 format (same as G8 FORM)
G8 FORM: G8 format
G7 FORM: G7 format
GHb-V FORM: GHbV A1c2.2 compatible format (Standard use)
GHb-V(SP): GHbV A1c2.2 compatible format (When “5 digits”
is selected for the number of digits of sample
number, if the sample is not CALIB, the sample
number is extended by adding 0 to the top digit
of the sample number. Refer to [6.2.5.GHbV
compatible format] for details.)

TOSOH CORPORATION Bioscience Division Page 6/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

4.2.Hi-LEVEL and RP modes


The HI-LEVEL mode is available in GX format and the old model compatible formats.
The HI-LEVEL mode is processed using Handshake protocol.
RP mode is a dedicated communication mode for Tosoh Data Management Program
501RP(hereafter 501RP). For models connectable to 501RP and settings for 501RP,
please refer to 501RP’s manual.
The HI-LEVEL or RP mode can be selected by setting RP MODE on the
PARAMETER screen of the analyzer as below. Please make sure to set RP MODE
as “NO” except when the analyzer is connected with 501RP.

Set RP MODE as “YES”: RP mode(only when connected with 501RP)


Set RP MODE as “NO”: HI-LEVEL mode(except when connected with
501RP)

TOSOH CORPORATION Bioscience Division Page 7/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

5.Transmission protocol
Handshake protocol is used as transmission protocol.
The control codes used in communications are shown below.
Code Name Meaning
(Hexadecimal)
02 STX Text start
03 ETX Text end
04 EOT Transmission end
06 ACK Normal reception response
15 NAK Abnormal reception response

Note:
If the host computer receives dust data, such as those caused by noise, from the
analyzer, please be sure to process them properly by the host computer.

5.1.Handshake
This protocol is used on the analyzer.
Communication content
<STX> text <ETX><BCC> →
← <ACK> or <NAK>
<EOT> →

Note:
1. <BCC> is the exclusion logic sum (exclusive OR), in character units, from
the start of the text to <ETX> and consists of a 1 byte binary. Moreover,
since <BCC> becomes the value of 00~7F by hexadecimal, please
process not to mistake for the control code used in transmission
procedure.
2. When <NAK> is returned transmission will be tried again up to 5 times.
After 5 times <EOT> will be transmitted and the protocol will be
reinitialized.
3. When multiple data sets exist, <EOT> will be transmitted at the end of
each data set.
4. The string from the beginning of the text to <BCC> must be transmitted
within 20 seconds. If this time limit is exceeded the protocol will be
reinitialized.
5. If there is no <ACK> or <NAK> response within 20 seconds, <EOT> will
be transmitted and the protocol will be reinitialized.

Example in the case of 1 data set


<STX> text1 <ETX><BCC> →
← <ACK>
<EOT> →

Example in the case of multiple data sets


<STX> text1 <ETX><BCC> →
← <ACK>
<STX> text2 <ETX><BCC> →
← <ACK>
:
:
<STX> textn <ETX><BCC> →

TOSOH CORPORATION Bioscience Division Page 8/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

← <ACK>
<EOT> →

Example in the case of retransmission succeeded (1 retransmission)


<STX> text1 <ETX><BCC> →
← <NAK>
<STX> text1 <ETX><BCC> →
← <ACK>
<EOT> →

Example in the case of retransmission failed (up to 5 retransmissions)


<STX> text1 <ETX><BCC> →
← <NAK>
<STX> text1 <ETX><BCC> →
← <NAK>
<STX> text1 <ETX><BCC> →
← <NAK>
<STX> text1 <ETX><BCC> →
← <NAK>
<STX> text1 <ETX><BCC> →
← <NAK>
<EOT> →

Example in the case of ACK reception timeout


<STX> text1 <ETX><BCC> →
(20 seconds passed)
<EOT> →

TOSOH CORPORATION Bioscience Division Page 9/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

6.Communication format
6.1.Query
First, an order request for the sample ID is transmitted to the host from the
analyzer, then the analyzer waits for a response from the host. If no responses
are given within 30 seconds, the analyzer will consider it that there is no orders for
the sample ID.
To activate query mode, turn the QUERY key on (highlighted) on the RS232C
screen.

Analyzer → host
Number of
Information Content
characters
G 1 Query code (Fixed as the character G)
, 1 Comma
[Sample ID] 20 Aligned left; characters to the right are filled with spaces when 20
characters are not used. No zero suppression

Host → analyzer
Number of
Information Content
characters
A 1 Order code (Fixed as the character A)
, 1 Comma
[Sample ID] 20 Aligned left; characters to the right are filled with spaces when 20
characters are not used. No zero suppression
Set the sample ID for which the analyzer has requested
, 1 Comma
[Sample type] 2 00 = Whole blood, 01 = Diluted (Refer to note)
[Container 2 00 = Primary tube, 01 = Sample cup (Refer to note)
type]
[Reserved] 2 (Fixed as spaces)
, 1 Comma
[A1c item] 2 00 = No request 01 = Request
[Reserved] 6 (Fixed as spaces)
, 1 Comma
[Reserved] 8 (Fixed as spaces)
Note: Sample and container type designations are subject to the SMP MODE on the
PARAMETER screen of the analyzer. Refer to "5.1.1 Sample and container
type designation" for details.

Communication example
(Analyzer) (Host)
<STX>G,ID-0123456789 <ETX><BCC> →
← <ACK>
<EOT> →
← <STX>A,ID-0123456789 ,0001 ,01 , <ETX><BCC>
<ACK> →
← <EOT>

TOSOH CORPORATION Bioscience Division Page 10/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

6.1.1.Sample and container type designation


The sample type (diluted, whole blood) and container type (sample cup, primary
tube) designated by the query will not be applied if they are not the same as those
defined by SMP MODE on the PARAMETER screen of the analyzer as below. In
such a case, those defined by SMP MODE will be used.

Figure 6-1 PARAMETER screen (SMP MODE)

Samples
Setting Sample type Container type Query
STANDARD Judged by the sample Judged by the sample Yes
sensors (SAMP1, SAMP2) if sensors.
it is a sample cup or a
primary tube. Sample in a
sample cup is considered as
a diluted sample and that in
a primary tube is considered
as a whole blood sample.
WHOLE BLD Fixed as a whole blood Judged by the sample Yes
sample sensors.
DILUTED Fixed as a diluted sample Judged by the sample Yes
sensors.
HOST Designated by the host Judged by the sample Yes
sensors.

Calibrator(see Note-1 below)


Setting Sample type Container type Query
Will be ignored Fixed as a diluted sample Fixed as a sample cup No (see
Note-2 below)
Note:
1. Samples, which are set on calibrator holders, are processed as
calibrators. Calibrators have to be set in sample cups.
2. Even if query is designated, no query will be made for calibrators.

TOSOH CORPORATION Bioscience Division Page 11/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

6.2.Measurement results
GX format or G9 format, G8 format, G7 format or GHbV compatible format may be
selected as the data format.
GX format or G8 format is recommended for this analyzer.
6.2.1.GX format
6.2.1.1.HI-LEVEL mode
The data format is shown below.

Information Number of characters Content


[Analyzer name] 12 “HLC-723GX ”
[Operation mode] 1 1(=VAR mode), Fixed
[Sampling position] 1 2(=table), Fixed
[Specimen type] 1 0=specimen, 1=calibrator
[Percent decimal] 1 1 or 2 (number of decimal places)
[Sample number] 3,5,8 Sample number (See Note-1 below)
(Recommended : 8)
[1st component] 6
[2nd component] 6
[3rd component] 6
[4th component] 6
[5th component] 6
[6th component] 6
[7th component] 6
[8th component] 6
[9th component] 6
[10th component] 6 Spaces or 0.0 ~ 100.0 or 0.00 ~ 100.00
[11th component] 6 (aligned right) (See Note-3 and -4 below)
[12th component] 6
[Total A1%] 6
SA1c(JDS)% 6
SA1c(NGSP)% 6
SA1c(Mono S)% 6
[Reserved] 6
[Reserved] 6
[Reserved] 6
[Reserved] 6
SA1c(IFCC) 5 Spaces or 0 ~ 99999 or 0.0 ~ 999.9
mmol/mol (aligned right)
[CALIB type] 2 00=JDS, 01=NGSP, 02=IFCC(JDS),
03=IFCC(NGSP), 04=Mono S,
05=IFCC(Mono S), 06=NGSP(JDS),
07=JDS(NGSP)
[Reserved] 2 (Spaces, fixed)
[Flag judgment] 2 00~99 (00=normal) (See Note-7 below)
[Reserved] 4 (Spaces, fixed)
[Sample ID] 0,13,18,20 Aligned left; characters to the right are
filled with spaces when 20 characters
are not used. No zero suppression (See
Note-5 below) (Recommended : 20)
Note:

1. The number of digits used for the sample number is set by the
SMPx key on the RS232C screen of the analyzer.
Various digit conversion rules are shown below.

TOSOH CORPORATION Bioscience Division Page 12/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

In the case of 5
Actual sample In the case In the case of
Specimen type digits (put 0 in
number of 3 digits 8 digits
second digit)
Normal samples 0001 ~ 7999 001 ~ 999 00001 ~ 70999 MMDD0001 ~
(last 3 digits) MMDD7999
CALIB 9001 ~ 9999 001 ~ 999 90001 ~ 90999 MMDD9001 ~
(last 3 digits) MMDD9999
*MMDD is the date of measurement. (MM: month, DD: day)

2. The number of digits used for each percent value and IFCC
value is subject to PERCENT DECIMAL of the PARAMETER
screen as shown below.
PERCENT DECIMAL Percent value IFCC value

1 0.0-100.0 0-99999

2, 9 0.00-100.00 0-99999

8 0.00-100.00 0.0-999.9

Figure 6-2 PARAMETER screen (PERCENT DECIMAL)

3. Transmitted components (1st to 12 th) are set by TRANS


COMPONENT of the PARAMETER screen, as shown in the list
below. Only two patterns, “23456789A---” or “1234567-----”, are
available on GX format.

Figure 6-3 PARAMETER screen (TRANS COMPONENT)

TOSOH CORPORATION Bioscience Division Page 13/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

TRANS COMPONENT: TRANS COMPONENT:


1234567----- 23456789A---
1st component FP% A1a%
2nd component A1a% A1b%
3rd component A1b% F%
4th component F% LA1c%
5th component LA1c% SA1c%
6th component SA1c% A0%
7th component A0% H-V0%
8th component Reserved H-V1%
9th component Reserved H-V2%
10th component Reserved Reserved
11th component Reserved Reserved
12th component Reserved Reserved

4. Each percent value is replaced with spaces of the number of


characters if the measurement result satisfied the condition
formula for level 1 flag. (Replacement with zero (0.0) from the
spaces is available. Refer to [6.2.6 Flag level and
communication] for details).

5. The number of digits used for the sample ID is set by the BCxx
key on the RS232C screen of the analyzer.

6. When multiple data sets are to be transmitted, EOT is


transmitted after the last data set.

7. If the measurement result satisfied two or more condition


formula for flag defined on FLAG screen, only one flag code is
transmitted to the host according to the priorities of the flags.
The level 1 flag is given priority over the level 0 flag. If two or
more condition formula of the same level of flag were satisfied, a
higher priority is given to the flag set lowest in the table on the
FLAG screen. (Refer to [6.2.6 Flag level and communication] for
details)

Communication examples are the case that TRANS COMPONENT of the PARAMETER
screen is set as “23456789A---”.
Communication example: 1 data set
(Analyzer) (Host)
<STX>HLC-723GX 12011026000111A0.911A1.011A0.91111.71110.5AA86.
6 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0AA12.4A1A0.0 A10.5AAA0.0AAA
0.0AAA0.0AAA0.0AAA0.0AAA9200AA00AAAABAR-CODE-ID111111111<ETX>
<BCC> →
← <ACK>
<EOT> →

TOSOH CORPORATION Bioscience Division Page 14/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

Communication example: 3 data sets


(Analyzer) (Host)
<STX>HLC-723GX 12011026000111A0.911A1.011A0.91111.71110.5AA86.
6 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0AA12.4A1A0.0 A10.5AAA0.0AAA
0.0AAA0.0AAA0.0AAA0.0AAA9200AA00AAAABAR-CODE-ID111111111<ETX>
<BCC> →
← <ACK>
<STX>HLC-723GX 12011026000211A0.911A1.011A0.91111.71110.5AA86.
6 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0AA12.4A1A0.0 A10.5AAA0.0AAA
0.0AAA0.0AAA0.0AAA0.0AAA9200AA00AAAABAR-CODE-ID111111111<ETX>
<BCC> →
← <ACK>
<STX>HLC-723GX 12011026000311A0.911A1.011A0.91111.71110.5AA86.
6 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0AA12.4A1A0.0 A10.5AAA0.0AAA
0.0AAA0.0AAA0.0AAA0.0AAA9200AA00AAAABAR-CODE-ID111111111<ETX>
<BCC> →
← <ACK>
<EOT> →

Communication example: decimal place of percent value is two


(Analyzer) (Host)
<STX>HLC-723GXAAA120210260001AA0.86110.95110.93111.67110.54186.6
4AA0.00AA0.00AA0.00AA0.00AA0.00AA0.00A12.36A A0.00A10.54AA0.00AA0
.00AA0.00AA0.00AA0.00AAA9200AA00AAAABAR-CODE-ID111111111<ETX>
<BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: Satisfied condition formula for level 1 flag with #ERR TRANS
FORM set as SPACE
(Analyzer) (Host)
<STX>HLC-723GXAAA120210260001AA0.00110.86110.95110.93111.67110.5
4A86.64AA0.00AA0.00AA0.00AA0.00AA0.00A12.36A10.54AA0.00AA0.00AA0
.00AA0.00AA0.00AA0.00AAA9201AA00AAAABAR-CODE-ID111111111<ETX>
<BCC> Æ
Å <ACK>
<EOT> Æ

TOSOH CORPORATION Bioscience Division Page 15/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

6.2.2.G8 format
6.2.2.1.HI-LEVEL mode
The data format is shown below.

Information Number of characters Content


[Operation mode] 1 1(= VAR mode), Fixed
[Sampling position] 1 2(= table), Fixed
[Sample number] 3, 5, 8 Sample number (See Note-1 below)
(Recommended : 8)
[1st component] 5, 6
[2nd component] 5, 6
[3rd component] 5, 6
Spaces or 0.0 ~ 100.0 or 0.00 ~ 100.00
[4th component] 5, 6
(aligned right) (See Note-3 below)
[5th component] 5, 6
when IFCC value is transmitted, spaces
[6th component] 5, 6
or 0 ~ 99999 or 0 ~ 999999
[7th component] 5, 6 or 0.0 ~ 9999.9
[8th component] 5, 6
[9th component] 5, 6
[10th component] 5, 6
[Flag judgment] 2 00 ~ 99 (00 = normal)
[Sample ID] 0, 13, 18, 20 Aligned left; characters to the right are
filled with spaces when 20 characters
are not used. No zero suppression (See
Note-5 below) (Recommended : 20)
Note:

1. The number of digits used for the sample number is set by the
SMPx key on the RS232C screen of the analyzer.
Various digit conversion rules are shown below.
In the case of 5
Actual sample In the case In the case of 8
Specimen type digits (put 0 in
number of 3 digits digits
second digit)
Normal samples 0001 ~ 7999 001 ~ 999 00001 ~ 70999 MMDD0001 ~
(last 3 digits) MMDD7999
CALIB 9001 ~ 9999 001 ~ 999 90001 ~ 90999 MMDD9001 ~
(last 3 digits) MMDD9999
*MMDD is the date of measurement. (MM: month, DD: day)
2. The number of digits used for each component value is subject
to PERCENT DECIMAL of the PARAMETER screen as shown
below.
PERCENT DECIMAL Percent value IFCC value

1 0.0-100.0 0-99999

2, 9 0.00-100.00 0-999999

8 0.00-100.00 0.0-9999.9

3. Transmitted components (1st to 10 th) are set by TRANS


COMPONENT of the PARAMETER screen, as shown in the list
below. Only two patterns, “23456789AT--” or “1234567--T--”, are
available on G8, G9 and G7 format.

TOSOH CORPORATION Bioscience Division Page 16/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

TRANS COMPONENT: TRANS COMPONENT:


1234567--T-- 23456789AT--
1st component FP% A1a%
2nd component A1a% A1b%
3rd component A1b% F%
4th component F% LA1c%
5th component LA1c% SA1c% or IFCC (*1)
6th component SA1c% or IFCC (*1) A0%
7th component A0% H-V0%
8th component Reserved H-V1%
9th component Reserved or IFCC (*2) H-V2%
10th component Total A1% or IFCC (*3) Total A1% or IFCC (*3)

(*1) When TRANS CONFIG of the PARAMETER screen is set to


“SA1C->IFCC”, this value is in IFCC as 0-99999 or 0-999999 or
0.0-9999.9.

(*2) When TRANS CONFIG of the PARAMETER screen is set to


“BLK->IFCC”, this value is in IFCC as 0-99999 or 0-999999 or
0.0-9999.9.

(*3) When TRANS CONFIG of the PARAMETER screen is set to


“TA1->IFCC”, this value is in IFCC as 0-99999 or 0-999999 or
0.0-9999.9.

Figure 6-4 PARAMETER screen (TRANS CONFIG)

4. Each percent value is replaced with spaces of the number of


characters, if the measurement result satisfied the condition
formula for level 1 flag. (Replacement with zero (0.0) from the
spaces is available. Refer to [6.2.6 Flag level and
communication] for details).

5. The number of digits used for the sample ID is set by the BCxx
key on the RS232C screen of the analyzer.

6. When multiple data sets are to be transmitted, EOT is


transmitted after the last data set.

7. If the measurement result satisfied two or more condition


formula for flag defined on FLAG screen, only one flag code is

TOSOH CORPORATION Bioscience Division Page 17/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

transmitted to the host according to the priorities of the flags.


The level 1 flag is given priority over the level 0 flag. If two or
more condition formula of the same level of flag were satisfied, a
higher priority is given to the flag set lowest in the table on the
FLAG screen. (Refer to [6.2.6 Flag level and communication] for
details)

Communication examples are the case that TRANS COMPONENT of the PARAMETER
screen is set as “23456789AT--”.
Communication example: 1 data set
(Analyzer) (Host)
<STX>1210260001110.6110.8110.8112.2115.8190.3110.0
110.0110.0117.300BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: 3 data sets


(Analyzer) (Host)
<STX>1210260001110.6110.8110.8112.2115.8190.3110.0
110.0110.0117.300BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<STX>1210260002110.6110.8110.8112.2115.8190.3110.0
110.0110.0117.300BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<STX>1210260003110.6110.8110.8112.2115.8190.3110.0
110.0110.0117.300BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: decimal place of percent value is two


(Analyzer) (Host)
<STX>1210260001110.64110.82110.76112.17115.79190.33110.00
110.00110.00117.2500BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: TRANS CONFIG is set to “SA1C->IFCC”


(Analyzer) (Host)
<STX>1210260001110.6110.8110.8112.211 40190.3110.0
110.0110.0117.300BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: TRANS CONFIG is set to “TA1->IFCC”


(Analyzer) (Host)
<STX>1210260001110.6110.8110.8112.2115.8190.3110.0
110.0110.011A4000BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

TOSOH CORPORATION Bioscience Division Page 18/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

6.2.3.G9 format
6.2.3.1.HI-LEVEL mode
The data format is the same as that of G8 format. (Refer to “6.2.2.1 HI-LEVEL
mode”)

TOSOH CORPORATION Bioscience Division Page 19/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

6.2.4.G7 format
6.2.4.1.HI-LEVEL mode
The data format is the same as that of G8 format. (Refer to “6.2.2.1 HI-LEVEL
mode”)

TOSOH CORPORATION Bioscience Division Page 20/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

6.2.5.GHbV compatible format


6.2.5.1.HI-LEVEL mode
The data format is shown below.

Information Number of characters Content


[Sample number] 3, 5, 8 Sample number (See Note-1 below)
(Recommended : 8)
[1st component] 5, 6
[2nd component] 5, 6
[3rd component] 5, 6 Spaces or 0.0 ~ 100.0 or 0.00 ~ 100.00
[4th component] 5, 6 (aligned right) (See Note-3 below)
[5th component] 5, 6 when IFCC value is transmitted, spaces
[6th component] 5, 6 or 0 ~ 99999 or 0 ~ 999999
[7th component] 5, 6 or 0.0 ~ 9999.9
[8th component] 5, 6
[9th component] 5, 6
[Flag judgment] 5 00 ~ 99 (00 = normal) (aligned left,
remaining characters filled with spaces)
[10th component] 5, 6 Spaces or 0.0 ~ 100.0 or 0.00 ~ 100.00
(aligned right) (See Note-3 below)
when IFCC value is transmitted, spaces
or 0 ~ 99999 or 0 ~ 999999
or 0.0 ~ 9999.9
[Sample ID] 0, 13, 18, 20 Aligned left; characters to the right are
filled with spaces when all characters are
not used. No zero suppression (See
Note-5 below)
Note:

1. The number of digits used for the sample number is set by the
SMPx key on the RS232C screen of the analyzer.
Various digit conversion rules are shown below.
In the case of 5
Actual sample In the case In the case of 8
Specimen type digits (put 0 in
number of 3 digits digits
second digit)
Normal samples 0001 ~ 7999 001 ~ 999 00001 ~ 70999 MMDD0001 ~
(last 3 digits) (See note) MMDD7999
CALIB 9001 ~ 9999 001 ~ 999 90001 ~ 90999 MMDD9001 ~
(last 3 digits) MMDD9999
*MMDD is the date of measurement. (MM: month, DD: day)

When TRANS MODE of PARAMETER screen of


analyzer is set to “GHb-V(SP)”, only for normal samples
the actual sample number is set to 00001-07999.
2. The number of digits used for each component value is subject
to PERCENT DECIMAL of the PARAMETER screen as shown
below.
PERCENT DECIMAL Percent value IFCC value

1 0.0-100.0 0-99999

2, 9 0.00-100.00 0-999999

8 0.00-100.00 0.0-9999.9

TOSOH CORPORATION Bioscience Division Page 21/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

3. Transmitted components (1st to 10 th) are set by TRANS


COMPONENT of the PARAMETER screen, as shown in the list
below. Only two patterns, “23456798AT--” or “23467----T--”, are
available on GHb-V compatible format.

TRANS COMPONENT: TRANS COMPONENT:


23467----T-- 23456798AT--
1st component A1a% A1a%
2nd component A1b% A1b%
3rd component F% F%
4th component SA1c% or IFCC (*1) LA1c%
5th component A0% SA1c% or IFCC (*1)
6th component Reserved A0%
7th component Reserved H-V1%
8th component Reserved H-V0%
9th component Reserved or IFCC (*2) H-V2%
10th component Total A1% or IFCC (*3) Total A1% or IFCC (*3)

(*1) When TRANS CONFIG of the PARAMETER screen is set to


“SA1C->IFCC”, this value is in IFCC as 0-99999 or 0-999999 or
0.0-9999.9.

(*2) When TRANS CONFIG of the PARAMETER screen is set to


“BLK->IFCC”, this value is in IFCC as 0-99999 or 0-999999 or
0.0-9999.9.

(*3) When TRANS CONFIG of the PARAMETER screen is set to


“TA1->IFCC”, this value is in IFCC as 0-99999 or 0-999999 or
0.0-9999.9.

4. Each percent value is replaced with spaces of the number of


characters, if the measurement result satisfied the condition
formula for level 1 flag. (Replacement with zero (0.0) from the
spaces is available. Refer to [6.2.6 Flag level and
communication] for details).

5. The number of digits used for the sample ID is set by the BCxx
key on the RS232C screen of the analyzer.

6. When multiple data sets are to be transmitted, EOT is


transmitted after the last data set.

7. If the measurement result satisfied two or more condition


formula for flag defined on FLAG screen, only one flag code is
transmitted to the host according to the priorities of flags. The
level 1 flag is given priority over the level 0 flag. If two or more
condition formula of the same level of flag were satisfied, a
higher priority is given to the flag set lowest in the table of the
FLAG screen. (Refer to [6.2.6 Flag level and communication] for
details)

TOSOH CORPORATION Bioscience Division Page 22/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

Communication examples are the case that TRANS COMPONENT of the PARAMETER
screen is set as “23456798AT--”.
Communication example: 1 data set
(Analyzer) (Host)
<STX>10260001110.6110.8110.8112.2115.8190.3110.0110.0
110.000111117.3BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: 3 data sets


(Analyzer) (Host)
<STX>10260001110.6110.8110.8112.2115.8190.3110.0110.0
110.000111117.3BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<STX>10260002110.6110.8110.8112.2115.8190.3110.0110.0
110.000111117.3BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<STX>10260003110.6110.8110.8112.2115.8190.3110.0110.0
110.000111117.3BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: decimal place of percent value is two


(Analyzer) (Host)
<STX>10260001110.64110.82110.76112.17115.79190.33110.00110.00
110.0000111117.25BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: TRANS CONFIG is set to “SA1C->IFCC”


(Analyzer) (Host)
<STX>10260001110.6110.8110.8112.211140190.3110.0110.0
110.000111117.3BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

Communication example: TRANS CONFIG is set to “TA1->IFCC”


(Analyzer) (Host)
<STX>10260001110.6110.8110.8112.2115.8190.3110.0110.0
110.00011111140BAR-CODE-ID111111111<ETX><BCC> Æ
Å <ACK>
<EOT> Æ

TOSOH CORPORATION Bioscience Division Page 23/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

6.2.6.Flag level and communication


6.2.6.1.Flag level
Each flag can be set as level 0 flag or level 1 flag on FLAG screen.

Figure 6-5 FLAG screen

Flag level 0

Flag level 1

If the measurement result satisfied two or more condition formula for flag defined
on the FLAG screen, only one flag code is transmitted to the host according to the
priorities of flags. The level 1 flag is given priority over the level 0 flag. If two or
more condition formula of the same level of flag were satisfied, a higher priority is
given to the flag set lowest in the table of the FLAG screen.
For example, when the measurement result satisfied the conditions defined in the
1st row and 3rd row of the table in Figure 6-5, “40” is transmitted to the host as the
flag code.
If you need to change the priority of flag to be transmitted, please arrange the
order of rows in the table on the FLAG screen carefully studying the cases.

6.2.6.2.Flag level and #ERR TRANS FORM


If the measurement result satisfied the condition formula for the level 0 flag, the
numeric value is transmitted to the host computer as the result value. But, if the
measurement result satisfied the condition formula for the level 1 flag, spaces or
zero value is transmitted as the result. It is defined by #ERR TRANS FORM on the
PARAMETER screen of the analyzer which, spaces or zero, is transmitted in this
case.

Figure 6-6 PARAMETER screen (#ERR TRANS FORM)

TOSOH CORPORATION Bioscience Division Page 24/25


HOST CONNECTION SPECIFICATIONS HLC-723GX Rev.2

#ERR TRANS FORM ZERO : When the result satisfied the condition formula
for the level 1 flag, zero is transmitted as the
result value (percent value).
#ERR TRANS FORM SPACE : When the result satisfied the condition formula
for the level 1 flag, spaces are transmitted as
the result value (percent value).

The number of digits used for each percent value is subject to the set value by
PERCENT DECIMAL on the PARAMETER screen. A communication example when
the measurement result satisfied the condition formula for the level 1 flag is shown
below.

Communication example, #ERR TRANS FORM set to “SPACE” (The part shown in
gray: filled with spaces)
(Analyzer) (Host)
<STX>HLC-723GXAAA120110260001AAA0.911A1.011A0.911A1.71110.51186.
6AAA0.0 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0AA12.4AAA0.0AA10.5AAA0.0AAA
0.0AAA0.0AAA0.0AAA0.0AAA9201AA01AAAABAR-CODE-ID111111111<ETX>
<BCC> Æ
Å <ACK>
<EOT> Æ

Communication example, #ERR TRANS FORM set to “ZERO”


(Analyzer) (Host)
<STX>HLC-723GXAAA120110260001AAA0.011A0.011A0.011A0.011A0.011A0.
0AAA0.0 AA0.0 AA0.0 AA0.0 AA0.0 AA0.0AAA0.0AAA0.0AAA0.0AAA0.0AAA
0.0AAA0.0AAA0.0AAA0.0AAAA001AA01AAAABAR-CODE-ID111111111<ETX>
<BCC> Æ
Å <ACK>
<EOT> Æ

TOSOH CORPORATION Bioscience Division Page 25/25