Beruflich Dokumente
Kultur Dokumente
Anesthesia Units
Purpose
Scope of this Product Comparison Anesthesia units dispense a mixture of gases and
This Product Comparison covers anesthesia sys- vapors and vary the proportion to control a patient’s
tems that can have the following components: level of consciousness and/or analgesia during surgical
mainframes, hanger yokes and gauges, flowme- procedures. Basically, anesthesia units perform the
ters, vaporizers, flush valves, carbon dioxide following four functions:
(CO2) absorbers, ventilators, scavenging sys-
• Provide oxygen (O2) to the patient
tems, monitors, and alarms. Not included are
separate analyzers designed to measure concen- • Blend gas mixtures that can include, besides O2, an
trations of halogenated anesthetics and gases anesthetic vapor, nitrous oxide (N2O,) air, and other
supplied to the unit or to detect levels present in medical gases
the operating room; also not included are sepa-
rate stand-alone physiologic monitoring systems.
For information on any of these devices, see the
following Product Comparisons:
• Halogenated Anesthetics Analyzers
• Multiple Medical Gas Monitors, Respired/
Anesthetic
• Oxygen Monitors
• Physiologic Monitoring Systems, Acute Care;
Neonatal; ECG Monitors
• Pressure Monitors, Airway
• Spectrometers, Mass, Respiratory/Anesthetic
Gas Monitoring
UMDNS information
This Product Comparison covers the following de-
vice term and product code as listed in ECRI’s
Universal Medical Device Nomenclature System™
(UMDNS™):
• Anesthesia Units [10-134]
• Facilitate ventilation with these gas mixtures, re- regulator when the central gas supply is used because
gardless of whether ventilation is spontaneous, con- the pressure is already at about 50 psi.
trolled, or assisted
Most anesthesia machines have an O2 supply fail-
• Reduce, if not eliminate, anesthesia-related risks to ure device and alarm that protect the patient from
the patient and clinical staff inadequate O2 supply. If the O2 supply pressure drops
below about 25 to 30 psi, the unit decreases or shuts
The patient is anesthetized by inspiring a mixture
off the flow of the other gases and activates an alarm.
of O2, the vapor of a volatile liquid halogenated hydro-
carbon anesthetic, and, perhaps, N2O and other gases. The flow of each gas in a continuous-flow unit is
Because normal breathing is routinely depressed by controlled by a valve and indicated by a flowmeter. The
the anesthetic agents and by muscle relaxants admin- flowmeter can be a purely mechanical arrangement,
istered in conjunction with the anesthetic agents, res- with a flow tube in which a bobbin moves up and down
piratory assistance — either with an automatic depending on the flow, or the flowmeter can be an
ventilator or by manual compression of the reservoir electronic sensor with an LCD (liquid crystal display).
bag — is usually necessary to deliver the breathing gas After the gases pass through the control valve and
to the patient. flowmeter, enter the low-pressure system, and pass
through a vaporizer if required, they are administered
Principles of operation to the patient. On machines sold in the United States,
the N2O and O2 flow controls are interlocked so that
An anesthesia machine comprises four basic sub-
the proportion of O2 to N2O can never fall below a
systems: a gas supply and control circuit, a breathing
minimum value (nominal 0.25) to produce a hypoxic
and ventilation circuit, a scavenging system, and a set
breathing mixture. An O2 monitor, located on the
of system function and breathing circuit monitors
inspiratory side of the breathing circuit or analyzing
(e.g., inspired O2 concentration and breathing circuit
gas sampled from the Y-piece of the patient’s breathing
integrity). Also included in some anesthesia systems
circuit, displays O2 concentration in volume percent.
are a number of monitors and alarms that indicate
O2 monitors should sound an alarm when the O2 falls
levels and variations of several physiological vari-
below the preset limit.
ables and parameters associated with cardiopulmon-
ary function and/or gas and agent concentrations in If the flow of anesthetic gases to the patient must
breathed-gas mixtures. Manufacturers typically offer be stopped for any reason, an O2 flush valve can be
a minimum combination of monitors, alarms, and activated to provide a large flow of central-source O2
other features that customers must purchase to meet to purge the breathing circuit of anesthetic vapors. The
standards and ensure patient safety. To meet the O2 flush flow bypasses the flowmeters and vaporizers.
minimum standard of care in the United States, an- In some units, the anesthetic gas flow momentarily
esthesia machines must monitor O2 concentration, shuts off.
airway pressure, and either the volume of expired gas
(Vexp) or the concentration of expired CO2. Vaporizers
Gas supply and control Because the inhaled anesthetic agents, with the
exception of N2O, exist as liquids at room temperature
Because O2 and N2O are used in large quantities, and sea-level ambient pressure, they must be evapo-
they are usually drawn from the hospital’s central gas rated by a vaporizer. Vaporizers add a controlled
supplies. In the United States, cylinders containing amount of anesthetic vapor to the gas mixture. Some
compressed O2, N2O, and sometimes other gases are anesthesia units can accommodate up to three vapor-
mounted on yokes attached to the anesthesia machine izers. Most units have a lockout mechanism that pre-
and can serve as an emergency gas supply in case vents the use of more than one vaporizer at a time.
central supplies fail. The requirements in other coun- There are several types of vaporizers, including vari-
tries throughout the world may not necessarily call for able bypass (conventional), heated blender, measured
cylinders to be attached to the anesthesia machine as flow, and draw-over. Variable bypass vaporizers can
an emergency backup. The United States requires now be either mechanically controlled or electronically
indexing pins, which are intended to prevent acciden- controlled.
tal mounting of a gas cylinder on the incorrect yoke.
Each gas entering the system from a cylinder flows Variable bypass and heated blender vaporizers are
through a filter, a one-way check valve, and a regulator concentration calibrated and thus can deliver a prese-
that lowers the pressure to approximately 45 pounds lected concentration of vapor under varying condi-
per square inch (psi). There is no need for a separate tions. In a variable bypass vaporizer, such as those
2 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 3
Healthcare Product Comparison System
C866UN7A-01
Ventilator Bag
Scavenging
To Atmosphere System
gas, to the patient. In the T-piece circuit, most of the depends on fresh-gas flow and occurs in direct propor-
exhaled gas is vented out of the system, but the portion tion to that flow. This system, though adaptable to a
rebreathed depends on the fresh-gas flow rate. variety of anesthetic procedures, is used most often in
pediatric anesthesia.
In the circle system, fresh gas from the anesthesia
machine enters the inspiratory limb of the breathing Circle systems offer advantages over T-piece sys-
circuit and mixes with gas in the system before the tems in that they conserve a greater proportion of the
resulting mixture flows through a one-way valve to the anesthetic gases and conserve body heat and moisture
patient. Expired gas flows from the patient, through from the patient. The advantages of T-piece systems
a second (expiratory) limb of the circuit, passing an- include a lower circuit compliance, easier circuit ster-
other one-way valve, and flows into either a reservoir ilization, and a less complex design requiring fewer
bag or a ventilator bellows. When positive pressure is valves and no CO2 absorber (although one can be used
generated in the system, either by a manual squeeze with it).
of the reservoir bag or by compression of the bellows
by a mechanical ventilator, collected gas that does not Because excess pressure imposed on the patient’s
escape via an adjustable pressure-limiting (APL) valve lungs can cause serious lung damage, either an APL
to the scavenging system is driven through a CO2 valve or a valve in the ventilator allows excess gas to
absorption canister and back to the patient. The can- escape when a preset pressure is exceeded. There are
ister contains either soda lime or barium hydroxide two types of APL valves: spring-loaded and needle
lime that removes CO2 from the rebreathed gases. In valves. The spring tension in spring-loaded APL valves
circle breathing systems, a fresh-gas flow of 1 L/min can be adjusted to control the pressure at which the
or less typically is considered low-flow anesthesia (4 to valve will open. At pressures below this, the valve is
10 L/min is typically considered the usual fresh-gas closed. The needle-valve type is always open except
flow rate). A fresh-gas flow of 0.5 L/min is generally when fully closed. The pressure in the breathing sys-
considered minimal-flow anesthesia. In situations tem maintained by the needle valve depends on the
where the cost of anesthetic agents is high, low-flow flow through the valve. Therefore, when the valve is
anesthesia may be the preferred option. not fully closed, gas will always leak from the system.
The minimum exhaust pressure required to refill a
Machines with T-piece design have corrugated tub- ventilator bellows is usually 1 to 2 cm H2O; for maxi-
ing in which fresh gas and some expired gas mix before mum pressure, both types of valve are fully closed.
entering the patient at each inhalation. Partial re- Because many APL valves do not have calibrated
breathing is controlled by the supply rate of fresh gas, markings, the anesthetist must adjust them empiri-
and the exhaled anesthetic mixture leaves the circuit cally to give a desired peak inspired pressure. Circle
through an APL valve. Elimination of rebreathed CO2 systems and T-piece systems also include a pressure
4 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 5
Healthcare Product Comparison System
6 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
agents and commonly used CO2 absorbents can pro- the efficacy of viral filters that attempt to reduce viral
duce CO if the CO2 absorbent is excessively dry. Drying contamination of breathing systems is not established.
out can occur when (1) an anesthesia machine has been Frequent replacement of disposable filters can prevent
sitting idle (e.g., over a weekend), and (2) there is a inadequate gas delivery due to clogging, and some
continuous flow of medical gas (which is very dry) filters can be sterilized and reused.
through the CO2 absorber. When dry, the absorbent
becomes highly reactive in the presence of certain The piping connections for O2 and N2O within the
halogenated agents, resulting in the production of CO hospital walls can be accidentally interchanged during
as the agent flows through the machine’s CO2 ab- installation or repair of medical gas systems, with the
sorber. ECRI recommends that the absorbent material potential for causing patient injuries or deaths. After
in both canisters of an absorber be replaced whenever any such work, careful inspection and testing with an
there is reason to believe that a machine has been left O2 analyzer are vital. Gas lines should also be checked
idle with gas flowing for an undetermined time. Fresh for liquid, gaseous, solid particulate, and microorgan-
absorbent materials are sufficiently hydrated and nor- ism contamination after such work and periodically
mally remain hydrated by exhaled water vapor in the thereafter.
circle system, thereby preventing reaction with halo- In the United States, a diameter index safety sys-
genated agents. tem (DISS) is used to prevent the connection of gas
hoses from the machine to the wrong wall outlet, and
Some anesthesia system malfunctions can cause de-
a pin index safety system (PISS) is used to prevent the
livery of gas with excessive CO2 concentration, inade-
connection of the wrong cylinders to the yokes in the
quate or excessive anesthetic agent, or dangerously high
anesthesia machine. The PISS employs pins protrud-
pressure. Hypoventilation, compromised cardiac output,
ing from the yoke that correspond to holes in a specific
air in the pleural cavity (pneumothorax), and asphyxi-
type of gas cylinder post. Only a cylinder post with the
ation are possible consequences of such problems.
corresponding holes can fit properly onto the yoke.
Improperly calibrated vaporizers can result in the Other countries have similar requirements to ensure
delivery of the wrong concentration of anesthetic agent the proper connection of all medical gas hoses to the
to the patient. Removing some vaporizers from the anesthesia machine.
anesthesia machine and transporting them can dis-
Faulty or inoperative scavenging systems are re-
turb their calibration and could eventually cause de-
sponsible for most anesthetic gas pollution in the op-
livery of too much or too little anesthesia. The output
erating room; other causes include improper
of anesthesia vaporizers should be tested each time one
anesthesia administration technique and leaks in an-
is removed from a system and each time it is returned
esthesia equipment. Common sources of leaks include
to service. Each vaporizer should be inspected and the
worn hose connectors, the CO2 absorber, the APL
calibration verified at least twice a year.
valve, and the endotracheal tube or mask. Current
Contamination of any part of the anesthesia breathing scientific and epidemiologic studies have shown that
circuit, including the breathing tubes, Y-connector, face exposure to trace levels of anesthetic gases continually
mask, and reservoir bag, may lead to nosocomial infec- present in the operating room can cause adverse
tions. Reported cases include infections of the upper health effects in operating room personnel, such as an
respiratory tract or the lungs and, in one instance in increased incidence of spontaneous abortion and con-
Australia, transmission of hepatitis C. The Centers for genital anomalies in offspring. In addition, trace gas
Disease Control and Prevention (CDC) and the American levels in the air may have a slight anesthetizing effect
Association of Nurse Anesthetists (AANA) recommend on the anesthetist and surgeon.
the single use of disposables or high-level disinfection of The increased interest in low-flow anesthesia to re-
reusables or disposables between patients to prevent duce costs has increased the potential danger of leaks in
cross-contamination. There has been some controversy the anesthesia unit. Because low-flow anesthesia re-
concerning the use of disposable bacteria filters to pre- quires very little fresh gas flow, a leak in the equipment
vent patient cross-infections (Berry and Nolte 1991; can result in inadequate delivery of O2 and anesthetic
Brooks et al. 1991; Dorsch and Dorsch 1994; Snowdon gases. Regular testing of the anesthesia equipment us-
1994; Hogarth 1996; Komesaroff 1996). CDC has not ing standard leak tests should minimize the risk of leaks
made a definitive recommendation concerning the use of during the administration of anesthesia.
bacterial filters with anesthesia machines. Possible haz-
ards, such as the increased impedance to gas flows and Inadequate evacuation of some scavenging systems
obstruction of the circuit, are associated with these fil- can cause pressure to build up in the breathing circuit,
ters. Also, because many viruses are difficult to culture, with the potential for pneumothorax.
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 7
Healthcare Product Comparison System
As mentioned previously, anesthesia units that lack Modular systems can be less expensive than precon-
integrated monitors and alarms can cause confusion figured systems, especially if the monitors are already
by sounding numerous alarms simultaneously. While owned.
integrated monitors and alarms are becoming more
Hospitals can purchase customized modular sys-
widespread, both modular and integrated systems are
tems, assembled from standard components, or they
subject to the confusion caused by false alarms. A false
can assemble their own modular systems. These sys-
alarm, caused by accidental patient movement or other
tems must meet all national and regional safety stan-
nonphysiological reasons, can confuse operating room
dards. Advantages of the modular approach include
staff and possibly draw attention away from other
flexibility in choosing and upgrading monitors and
alarms that may truly indicate a change in the pa-
ease of service; drawbacks include assembling a sys-
tient’s physiologic condition. Currently, there is no
tem that may not be successfully integrated, thereby
quantitative agreement on the prevalence of false
having multiple alarms, multiple displays, etc.
alarms in anesthesia units (Block and Schaaf 1996).
Ensuring that the alarm limits are properly set and Anesthesia units and patient monitoring systems
positioning sensors and electrodes in such a way as to should be carefully chosen to ensure that all the essential
minimize artifacts can reduce the incidence of false monitoring functions recommended by the American
alarms. Society of Anesthesiologists (ASA) are obtained and to
ensure optimal integration and an adequate standard of
The magnetic fields created by magnetic resonance
care. For legal reasons, the level of monitoring and
imaging (MRI) equipment may interfere with the func-
anesthesia delivery capabilities for each anesthesia sta-
tion of conventional anesthesia units and electronic
tion should be uniform so that all patients receive the
monitoring equipment when used in proximity to such
same standard of care for the same surgical procedures.
equipment. Conversely, magnetic materials and elec-
tronic monitors may interfere with MRI scanner func- Integrated anesthesia workstations, along with the
tion and degrade image quality. Rao et al. (1988) have gas/vapor dispensing subsystem and individual physi-
reported changing all ferromagnetic materials to non- ologic and equipment monitors, may also include a
ferromagnetic (including support structures, casters, device for automatically dispensing injectable drugs.
and gas cylinders) to allow an anesthesia machine to Consequently, the anesthesia workstation can be
be used in the MRI suite. A few suppliers offer MRI- viewed as an integrated monitoring system that dis-
compatible anesthesia machines, and a line of MRI- penses anesthetic drugs.
compatible monitors is available.
Hospitals should also consider the standardization
Users should be careful not to hang any extraneous of anesthesia equipment; that is, purchasing systems
materials (e.g., polyethylene garbage bag) or equip- that are compatible with equipment already in operat-
ment off anesthesia units. If accidently bumped, the ing rooms or other areas of the hospital (e.g., intensive
hanging objects may compromise the anesthesia unit’s care units). The purpose of standardization is to allow
stability, causing the hanging objects to be sucked into a reduced parts inventory, minimize the number of
the receiving end of the anesthesia unit. This might suppliers and service personnel, and reduce confusion
cause the full negative pressure to be transmitted to among the staff.
the patient breathing system, collapsing the reservoir
Pulse oximeters noninvasively measure O2 satura-
bag.
tion of blood hemoglobin (SpO2) and, along with O2
As anesthesia machines become increasingly elec- monitors and CO2 monitors, are increasingly being
tronically controlled, there is the risk of electrical fire required for anesthesia units by state law. Some U.S.
in the operating room. Current anesthetic agents will states have specified their own requirements for anes-
not ignite, but the presence of pressurized O2 and N2O thesia units. Hospitals should check with their state’s
could increase the severity of a fire initially caused by department of health for any regulations that may
electrical problems with the anesthesia unit. apply to their area. CO2 monitors measure end-tidal
CO2 and can help identify leaks and misconnections as
Purchase considerations well as indicate when the trachea has not been prop-
erly intubated.
Some anesthesia units require stand-alone physi-
ologic monitors (modular approach) and/or anesthetic Many features of anesthesia systems are optional,
agent monitors, while others have integrated monitors allowing hospitals to choose the ones that best fit their
(preconfigured approach). The advantages of precon- needs. Among anesthesia units with essentially
figured monitoring include convenience and electroni- equivalent mechanical gas/vapor dispensing subsys-
cally integrated displays and prioritized alarms. tems, the monitors included in the system and the
8 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
ways in which information is integrated and displayed • Recording and storing anesthesia-related data
are often the primary features distinguishing anesthe-
• Disposables
sia systems.
• Utilities
Microprocessor-controlled equipment could be af-
fected by the change from the year 1999 to 2000 if it When selecting a vaporizer, consider the type of
uses a real-time clock and its design does not include anesthetic agent required for the hospital’s patient mix
provision for a change in century — for instance, it in conjunction with the type of procedures being per-
does not use a four-digit data field for the year. Fur- formed. Users should ask the supplier if the anesthetic
thermore, any incompatibilities in the way different gas monitor will be able to identify and measure all
devices handle the year 2000 might have adverse ef- anesthetic agents used (i.e., some models may not
fects. Even if a device is unaffected, it may affect or be recognize sevoflurane).
affected by other devices through device and informa-
Hospitals can purchase service contracts or service
tion systems interfaces (see the December 1998 Health
on a time-and-materials basis from the supplier. Ser-
Devices citation below).
vice may also be available from a third-party organiza-
Facilities purchasing new equipment should add a tion. The decision to purchase a service contract should
specification to the request for proposal or other bid be carefully considered. Most suppliers should provide
documents stating that the device will not be affected routine software updates, which enhance the system’s
by the change to the year 2000 and that the supplier performance, at no charge to service contract custom-
will provide written certification of this fact. Such a ers. Purchasing a service contract also ensures that
precaution will help prevent costly downtime or major preventive maintenance will be performed at regular
software changes after purchase. Facilities should also intervals, thereby eliminating the possibility of unex-
consider whether any existing information systems pected maintenance costs. Also, many suppliers do not
with which a new system will be interfaced are already extend system performance and uptime guarantees
compliant. beyond the length of the warranty unless the system is
covered by a service contract. Hospitals that plan to
Cost containment service the anesthesia units in-house should inquire
Because anesthesia systems entail ongoing mainte- about the availability and cost of service training and
nance and operational costs, the initial acquisition cost the availability and cost of replacement parts.
does not accurately reflect the total cost of ownership. ECRI recommends that, to maximize bargaining
The anesthetic agents are the biggest ongoing expense leverage, hospitals negotiate pricing for service con-
associated with anesthesia units. Therefore, a pur- tracts before the system is purchased. Additional ser-
chase decision should be based on issues such as life- vice contract discounts may be negotiable for
cycle cost (LCC), local service support, discount rates, multiple-year agreements, or for service contracts that
and non-price-related benefits offered by the supplier. are bundled with contracts on other similar equipment
An LCC analysis should be conducted to determine in the department or hospital. Discounts will depend
the cost-effectiveness of all the units that meet the on the hospital’s negotiating skills, knowledge of dis-
users’ needs. counts offered to other customers (from services such
as ECRI’s online cost database or SELECT™ service),
Although many of the following costs may be similar the system configuration and model to be purchased,
for a number of anesthesia units, they should still be previous experience with the supplier, and the extent
carefully considered to determine total LCC for budget of concessions granted by the supplier, such as ex-
purposes: tended warranties, fixed prices for annual service con-
• Maintenance, service, and inspections tracts, and guaranteed on-site service response.
Buyers should make sure that applications training is
• Accessories, such as monitoring equipment, neces- included in the purchase price of the system. Some
sary to comply with standards suppliers offer more extensive on- or off-site training
• Optional accessories programs for an additional cost.
• Vaporizers (some have been offered at discounted
prices or at no cost upon the introduction of a new Stage of development
anesthetic agent) Efforts to improve the design of anesthesia units
center on gas supply and proportioning systems,
• Gases, including O2, N2O, and anesthetic agents
breathing circuits, gas scavenging and humidification
• Anesthesia circuits devices, gas monitors, ventilators, vaporizers, and data
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 9
Healthcare Product Comparison System
handling (display, processing, and reporting) software. Brooks JH, Gupta B, Baker D. Anesthesia machine
There is also an effort to decrease the overall size of contamination [abstract]. Anesthesiology 1991 Sep;
anesthesia units. 75(3A):A874.
Several U.S. manufacturers already offer versions Centers for Disease Control and Prevention. Draft
of anesthesia workstations in their product lines. The guideline for prevention of nosocomial pneumonia;
Committee for European Standardization (CEN) is notice of comment period. Fed Regist 1994 Feb
working on a European draft standard that establishes 2;59(22):4985.
the essential requirements for these workstations (see
Standards and Guidelines). The standard will outline Chant K, Kociuba K, Munro R, et al. Investigation of
the requirements for individual modules that together possible patient-to-patient transmission of hepati-
constitute a complete anesthesia workstation; the tis C in a hospital. New South Wales Public Health
draft applies the same standards to each module as to Bull 1994 May;5(5):47-51.
the system as a whole. A committee of the American
Davey A, Moyle JT, Ward CS. Ward’s anaesthetic
Society for Testing and Materials (ASTM) is drafting
equipment. 3rd ed. London: W.B. Saunders; 1992.
anesthesia workstation standards in the United
States. Dorsch JA, Dorsch SE. Understanding anesthesia
equipment: construction, care and complications.
Some anesthesia systems now use microprocessors
3rd ed. Baltimore: Williams & Wilkins; 1994.
to control certain parts of the unit. The microprocessor
may control and mix the flows of gases used in the Ehrenwerth J, Eisenkraft JB, eds. Anesthesia equip-
anesthesia unit. Computer-controlled systems are ex- ment: principles and applications. St. Louis: Mosby-
pected to become more common in the near future. Year Book; 1993.
The potential risk that current anesthetic agents Eisenkraft JB, Leibowitz AB. Ventilators in the oper-
pose to the ozone layer has led some regulatory agen- ating room. Int Anesthesiol Clin 1997 Winter;
cies to call for the phasing out of current anesthetic 35(1):87-108.
agents. However, phasing out the current agents
would necessitate finding acceptable alternative anes- Elliot B, Chestnut J. Dangers of alarms [letter]. Anaes-
thetic agents, and none have yet been found. thesia 1996 Aug;51(8):799-800.
Bromley HR, Tvorinsky S. An uncommon leak in the Komesaroff D. Disposable and autoclavable anaes-
anesthesia breathing circuit [letter]. Anesth Analg thetic circuits: the future is now. Anaesth Intensive
1997 Sep;85(3):707. Care 1996 Apr;24(2):173-5.
10 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
McMahon DJ. A synopsis of current anesthesia ma- Specification for minimum performance and safety
chine design. Biomed Instrum Technol 1991 May- requirements for components and systems of anes-
Jun;25(3):190-9. thetic gas monitors [standard]. ASTM Committee
F29 on Anesthetic and Respiratory Equipment.
Petty C. The anesthesia machine. New York: Churchill F1452-92(1992). 1992.
Livingstone; 1987.
Canadian Standards Association. Breathing systems
Petty WC. New anesthetic requires new vaporizers for for use in anaesthesia [standard]. Z168.9-92. 1986
safety. APSF Newsl 1995-96 Winter:46. (reaffirmed 1992).
Rao CC, McNiece WL, Emhardt J, et al. Modification Tracheal tube connectors [standard]. CAN/CSA-
of an anesthesia machine for use during magnetic Z7228-94. 1993.
resonance imaging [letter]. Anesthesiology 1988 Tracheal tubes [standard]. CAN/CSA-Z5361-94.
Apr;68(4):640-1. 1994.
Rogers S, Davies MW. My anaesthetic machine’s on European Committee for Standardization. Breathing
fire [letter]. Anaesthesia 1997 May;52(5):505. tubes intended for use with anaesthetic apparatus
and ventilators [draft standard]. prEN 12342. 1989.
Schreiber P. Safety guidelines for anesthesia systems.
Test methods for breathing system filters [stan-
Boston: North American Drager; 1985.
dard]. CEN/TC 215 N 326. 1989.
Sivalingam P, Hyde RA, Easy WR. An unpredictable International Organization for Standardization.
and possibly dangerous hazard of an anaesthetic Breathing tubes intended for use with anaesthetic
scavenging system [letter]. Anaesthesia 1997 Jun; apparatus and ventilators [standard]. 3rd ed. ISO
52(6):609-10. 5367:1991. 1991.
Snowdon SL. Hygiene standards for breathing sys- Inhalational anaesthesia systems — part 2: anaes-
tems? [editorial] Br J Anaesth 1994 Feb;72(2):143-4. thetic circle breathing systems [standard]. 1st ed.
ISO 8835-2:1993. 1993.
Somprakit P, Soontranan P. Low pressure leakage in
Anesthesia equipment
anaesthetic machines: evaluation by positive and
negative pressure tests. Anaesthesia 1996 May; American Association of Nurse Anesthetists. Infection
51(5):461-4. control guide [guideline]. 1993 (revised 1997).
Transitional substances. In: London Revisions to the American Society for Testing and Materials. Particu-
Montreal Protocol Second Meeting of the Parties to lar requirements for anesthesia workstations and
the Montreal Protocol on Substances that Deplete their components [standard]. ASTM Committee
the Ozone Layer. Appendix C. 1990 Jun 27-29; Lon- F29.0109 on Anesthesia Workstations. F1850-98.
don. Ozone Secretariat Reports and Publications. 1998.
Available from Internet: http://www.unep.ch/ozone/ Specification for alarm signals in medical equip-
2mlonfin.htm. ment used in anesthesia and respiratory care
[standard]. ASTM Committee F29 on Anesthetic
Standards and guidelines and Respiratory Equipment. F1463-93. 1993.
Note: Although every effort is made to ensure that the Specification for anesthetic equipment — oro-
following list is comprehensive, please note that other pharyngeal and nasopharyngeal airways [stan-
applicable standards may exist. dard]. ASTM Committee F29 on Anesthetic and
Anesthesia breathing circuits Respiratory Equipment. F1573-95. 1995.
Standard specification for minimum performance and
American Society for Testing and Materials. Particu-
safety requirements for components and systems of
lar requirements for anesthesia workstations and
anesthesia gas machines [standard]. ASTM Commit-
their components [standard]. ASTM Committee
tee F29 on Anesthetic and Respiratory Equipment.
F29.0109 on Anesthesia Workstations. F1850-98.
F1161-88(1994). 1988 (reapproved 1994).
1998.
Specification for anesthetic breathing tubes [stan- American Society of Anesthesiologists. Anesthesia ap-
dard]. ASTM Committee F29 on Anesthetic and paratus checkout recommendations. 1993.
Respiratory Equipment. F1205-88(1993). 1988 (re- Recommendations for infection control for the practice
approved 1993). of anesthesiology. Committee on Occupational Health
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 11
Healthcare Product Comparison System
of Operating Room Personnel and Subcommittee on Medical electrical equipment — part 1: general re-
Infection Control Policy. 30503-5PB. 1992. quirements for safety. Amendment 1 [standard].
Standards for basic anesthetic monitoring. 1986 IEC 60601-1-am1 (1991-11). 1991.
(revised 1995). Medical electrical equipment — part 1: general re-
Association of Operating Room Nurses. Cleaning and quirements for safety. Amendment 2 [standard].
processing anesthesia equipment [recommended IEC 60601-1-am2 (1995-03). 1995.
practice]. 1977 (revised 1995). Medical electrical equipment — part 1: general re-
British Standards Institution. Anaesthetic and anal- quirements for safety. Section 2. Collateral standard:
gesic machines. Specification for continuous flow electromagnetic compatibility — requirements and
anaesthetic machines [standard]. BS 4272:Part 3. tests. IEC 60601-1-2 (1993-04). 1993.
1989. Medical electrical equipment — part 2: particular
Anaesthetic and analgesic machines. Specification requirements for the safety of anesthetic worksta-
for intermittent (demand) flow analgesic machines tions [standard]. IEC 60601-2-13 (1998-05). 1998.
for use with 50/50% (V/V) nitrous oxide [standard]. International Organization for Standardization. An-
BS 4272:Part 2. 1968 (revised 1996). aesthesia and respiratory care alarm signals —
Specification for oxygen analyzers for monitoring part 1: visual alarm signals [standard]. 1st ed. ISO
patient breathing mixtures. BS 5724:Section 2.27. 9703:Part 7-1:1997. 1992.
1989 (revised 1997). Anaesthesia and respiratory care alarm signals —
Canadian Anaesthetists’ Society. Guidelines to the part2: auditory visual alarm signals [standard]. 1st
practice of anaesthesia. 1996 (revised 1997). ed. ISO 9703-2:1997. 1997.
Canadian Standards Association. Continuous-flow in- Anaesthetic and respiratory equipment — conical
halation anaesthetic apparatus (anaesthetic ma- connectors — part 1: cones and sockets [standard].
chines) for medical use [standard]. CAN3-Z168. 1st ed. ISO 5356-1:1987. 1987.
3-97. 1984. Anaesthetic and respiratory equipment — conical
Continuous-flow inhalation anaesthetic apparatus connectors — part 1: cones and sockets: amend-
(anaesthetic machines) for medical use: supplement ment 1 [standard]. ISO 5356:1-1987/amend 1:1993.
no. 1 [standard]. CAN/CSA-Z168.3S1-M91. 1991. 1993.
Pressure regulators, gauges, and flow-metering de- Anaesthetic and respiratory equipment — conical
vices for medical gases [standard]. CAN/CSA- connectors — part 2: screw-threaded weight-bear-
Z305.3-M87. 1987 (revised 1997). ing connectors [standard]. 1st ed. ISO 5356-2:1987.
1987.
Centers for Disease Control and Prevention. Guideline
for handwashing and hospital environmental con- Anaesthetic and respiratory equipment — heat and
trol. Am J Infect Control 1986 Jun;14(3):110-29. moisture exchangers for use in humidifying re-
spired gases in humans [standard]. ISO 9360:1992.
Guideline for prevention of nosocomial pneumonia.
1992.
Hospital Infection Control Practices Advisory Com-
mittee, National Centers for Infectious Dieseases. Anaesthetic machines for use with humans [stan-
Respir Care 1994 Dec;39(12):1191-236. dard]. 2nd ed. ISO 5358:1992. 1992.
European Committee for Standardization. Anaes- International Task Force on Anaesthesia Safety. In-
thetic and respiratory equipment — conical connec- ternational standards for a safe practice of anaes-
tors — part 1: cones and sockets [standard]. EN thesia. Eur J Anaesthesiol 1993 Jan;10(Suppl
1281-1:1997. 1994 (revised 1997). 7):12-5.
Anaesthetic and respiratory equipment — conical con- Standards Association of Australia. Anaesthetic ma-
nectors — part 2: screw-threaded weight-bearing con- chines — non-electrical — for use with humans
nectors [draft standard]. prEN 1281-2:1994. 1994. [standard]. AS/NZS 4059 1996. 1996.
Medical electrical equipment — anaesthetic work- U.S. Food and Drug Administration. Anesthesiology
stations and their modules — particular require- devices. 21 CFR Part 868. 1995.
ments [draft standard]. prEN 740. 1989. Underwriters Laboratories. Electrically conductive
International Electrotechnical Commission. Medical equipment and materials for use in flammable anes-
electrical equipment — part 1: general requirements thetizing locations [standard]. 3rd ed. 1067. 1987
for safety [standard]. IEC 60601-1 (1988-12). 1988. (revised 1997).
12 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 13
Healthcare Product Comparison System
(There are many state rules and regulations in the Nitrogen distribution systems [hazard]. 1989 Feb;
United States regarding anesthesia machines; consult 18(2):85.
ECRI’s Healthcare Standards Directory or your state Anesthesia units and breathing systems [standard].
for more information.) 1989 Oct;18(10):363.
Citations from other ECRI publications Gas monitoring and the standard of care in the OR
[clinical perspective]. 1990 Jul;19(7):207-8.
Health Devices
Multiple medical gas monitors [evaluation]. 1990
North American Drager Narkomed 2A anesthesia Jul;19(7):209-35.
units [User Experience Network™]. 1986 May;
Oxygen regulator fire caused by use of two yoke wash-
15(5):142-3.
ers [hazard]. 1990 Nov;19(11):426-7.
Using an oxygen monitor with a pulse oximeter during
Monitoring and anesthesia systems: integration and a
anesthesia. [User Experience Network™]. 1986
new option. 1991 Mar-Apr;20(3-4):131.
Sep-Oct;15(9-10):294-5.
North American Drager Narkomed 4. 1991 Mar-
Anesthesia units with a flowmeter-controlled vapor-
Apr;20(3-4):131.
izer [hazard]. 1986 Dec;15(12):336-7.
Ohmeda CD. 1991 Mar-Apr;20(3-4):131.
North American Drager anesthesia vaporizers [User
Experience Network™]. 1986 Dec;15(12):332. North American Drager Narkomed 4 software update
and revised calibration procedure [User Experience
North American Drager halothane vaporizers [User Network™]. 1993 Dec;22(12):601.
Experience Network™]. 1987 Feb;16(2):55.
Use of inadequate (old) anesthesia scavenger inter-
Concentration calibrated vaporizers [hazard]. 1987 faces [hazard]. 1993 Dec;22(12):592-3.
Mar-Apr;16(3-4):112.
Risk of barotrauma and/or lack of ventilation with
Pre-use testing prevents “helpful” reconnection of anes- ventilatorless anesthesia machines [hazard]. 1994
thesia components [hazard]. 1987 May;16(5):178-9. Jan-Feb;23(1-2):54-5.
Anesthesia systems [evaluation]. 1988 Jan;17(1):3-29, Desflurane (Suprane) [guidance article]. 1994 Apr;
32-4. 23(4):131-9.
Automatic record keeping in anesthesia. 1988 Jan; False CO2 readings from disposable anesthesia
17(1):30-1. breathing circuits with an internal gas-sampling
Sodasorb PrePak CO2 absorption cartridges [hazard]. line [hazard]. 1995 Apr;24(4):160-1.
1988 Jan;17(1):35-6. Fires from oxygen use during head and neck surgery
Dryden anesthesia breathing circuits [hazard]. 1988 [hazard]. 1995 Apr;24(4):155-7.
Feb;17(2):66. Anesthesia systems [evaluation]. 1996 May-Jun;25(5-
Who should service anesthesia equipment? [User Ex- 6):158-211.
perience Network™]. 1988 Feb;17(2):70-1. Anesthesia ventilators with descending bellows: the
Dryden 10100 disposable CO2 absorbers [hazard]. need for appropriate monitoring [hazard]. 1996
1988 Sep;17(9):276. Oct;25(10):391-3.
Medical air compressors [User Experience Net- Leaching of the plasticizer from PVC tubing in heart-
work™]. 1988 Sep;17(9):279. lung bypass unit tubing circuits [User Experience
Network™]. 1996 Oct;25(10):393-5.
Pre-use anesthesia check fails to find faults [hazard].
Medical devices and the year 2000 problem [guidance
1988 Sep;17(9):274-5.
article]. 1997 Dec;(26)12:448-56.
Pre-use checklist for anesthesia units. 1988 Sep;
Anesthesia systems [update evaluation]. 1998
17(9):275.
Jan;27(1):4-27.
Barotrauma from anesthesia ventilators [hazard].
Carbon monoxide exposures during inhalation anes-
1988 Nov;17(11):354.
thesia: the interaction between halogenated anes-
Anesthesia systems [evaluation update]. 1988 Dec; thetic agents and carbon dioxide absorbents [hazard
17(12):366-7. report]. 1998 Nov;27(11):402-4.
New monitoring standards for anesthesia. 1988 Dec; Y2K compliance issues [special section]. 1998 Dec;
17(12):374. 27(12):412-23.
14 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Health Devices Alerts 31887 Caplan RA, Vistica MF, Posner KL, et al.
Adverse anesthetic outcomes arising from gas delivery
This Product Comparison lists Health Devices Alerts
equipment. Anesthesiology 1997 Oct;87(4):741-8.
(HDA) citations published since the last update of this
report. Each HDA abstract is identified by an Acces- 31888 Yasumoto M, Sakuragi T, Maruta S, et al.
sion Number. Recalls and hazard reports include de- Activation of apnea alarm by a surgical theater light
scriptions of the problem involved; abstracts of other during ophthalmological surgery [letter]. Anesth
published articles are referenced by bibliographic in- Analg 1997 Dec;85(6):1414-5.
formation. HPCS subscribers can call the Hotline for 32210 Osterud A. A non-rebreathing coaxial anaes-
additional information on any of these citations or to thesia system [letter]. Anaesthesia 1997 Nov;
request more extensive searches of the HDA database. 52(11):1123.
A3347 FDA designated Class II Recall No. Z-863-7 of 32385 Mizuno K, Sumiyoshi R. Air contamination of
certain North American Drager anesthesia units. The a closed anesthesia circuit. Acta Anaesthesiol Scand
oxygen flush button on some machines may stick in the 1998 Jan;42(1):128-30.
open position if the affected product contains small
32475 Primiano FP Jr. Open adjustable pressure
burs at the set screw holes that may contact the button.
limiter valve [letter]. Anesthesiology 1998 Feb;
The manufacturer initiated a recall by certified letter
88(2):552.
dated June 11, 1997. Users should verify that you
received the June 11, 1997, letter from North Ameri- 32785 Cowling M. Cato machine failure [letter].
can Drager containing a verification form, a small-di- Anaesth Intensive Care 1998 Jun;26(3):329-30.
ameter replacement button, replacement procedures, 32789 Tripathi M. A partial disconnection at the main
and an Allen wrench. If you want the manufacturer to stream CO2 transducer mimics “curare-cleft” capno-
replace the button or need further information regard- graph. Anesthesiology 1998 Apr;88(4):1117-9.
ing the button replacement, U.S. customers should
contact the North American Drager Technical Services 33147 McLaughlin AJ, Campkin NT. Electrical safety —
Department at (800) 543-5047, and international cus- a reminder [letter]. Anaesthesia 1998 Jun;53(6):608-9.
tomers should contact North American Drager Tech- 33516 Ti LK, Dhara SS. Another cause of a prolonged
nical Services at (215) 721-5400. Source: FDA downstroke on the capnograph [letter]. Anesthesiology
Enforcement Rep 1997 Aug 27; Manufacturer. 1998 Sep;89(3):801-2.
D3770 FDA has designated Class II Recall No. Z-220-9 33591 Kendell J, Barthram C. Revised checklist for
of certain Core-M (manufacturer) and North American anaesthetic machines. Anaesthesia 1998 Sep;53(9):
Drager (distributor) anesthesia units. If the monitors 887-90.
are not warmed up properly before calibration, the
34073 Anesthesia machines: product was not defec-
values stored for calibration may be invalid. The dis-
tive. Spec Law Dig Health Care Law 1998
tributor initiated a recall by letter dated October 7,
Dec;(236):31.
1998. The firm states that customers have received a
new user’s manual and have been instructed to affix a 34253 Goto T, Saito H, Nakata Y, et al. Effects of
sticker to the affected machines warning of the prob- xenon on the performance of various respiratory flow-
lem. No further action is required by the user. meters. Anesthesiology 1999 Feb;90(2):555-63.
Source: FDA Enforcement Rep 1998 Dec 9; Distributor. Health Devices Inspection and Preventive Maintenance
31323 Eltringham RJ, Varvinski A. The Oxyvent: an System
anesthetic machine designed to be used in developing
Anesthesia unit vaporizers. 436-0595. 1995
countries and difficult situations. Anaesthesia 1997
Jul;52(7):668-72. Anesthesia unit ventilators. 461-0595. 1995.
31420 Lee HT. Ideal placement for the suction cathe- Anesthesia units. 400-0595. 1995.
ter [letter]. Am J Anesth 1997 Jul-Aug;24(4):214. Capnometers and multiple medical gas monitors. 450-
31729 Aarhus A, Soreide E, Holst-Larsen H. Mechani- 0595. 1995.
cal obstruction in the anaesthesia delivery-system
Healthcare Risk Control
mimicking severe bronchospasm [case report]. Anaes-
thesia 1997 Oct;52(10):992-4. Surgery and anesthesia. 1996;3:Surgery and anesthe-
31884 Marchionni L, Agro F, Favaro R, et al. The sia:14.
flexible laryngeal mask as a nasal airway [letter]. Overview of anesthesia liability. 1996;4:Surgery and
Anesth Analg 1997 Nov;85(5):1179. anesthesia:2.
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 15
Healthcare Product Comparison System
Waste anesthetic gas. 1996;4:Surgery and anesthe- AMS (Advanced Medical Systems) Ltd [356053]
sia:15. Kazim Karabekir Cad
Pre-use checklist for anesthesia units (machines and 95/95 06060 Iskitler
accessories). 1996;12:Surgery and anesthesia:2. Ankara
Turkey
Operating Room Risk Management Phone: 90 (312) 3840520
Fax: 90 (312) 3423307
Automated record keeping in anesthesia. 1992 Jul; E-mail: ams@ams.com.tr
1:Anesthesia:8. Internet: http://www.ams.com.tr
Patient monitoring in the OR: Vigilance, monitoring,
and the standard of care. 1992 Jul;1:Anesthesia:1.
Anamed
Pre-use checklist for anesthesia units (machines and
accessories). 1992 Jul;1:Anesthesia:3. The Sinus TR is marketed in Europe, the Far East,
Selecting and using multiple medical gas monitors. the Middle East, and Russia
1992 Jul;1:Anesthesia:5.
Anamed GmbH [280486]
Selecting and using physiologic monitors. 1992 Jul; Arzbacher Rutenstrasse 80
1:Anesthesia:4. D-56130 Bad Ems Bremen
Germany
Anesthesia malpractice: an overview. 1993 Jul;1:Anes-
Phone: 49 (2603) 96460
thesia:2.
Fax: 49 (2603) 964696
Desflurane (Suprane). 1994 Oct;1:Anesthesia:11. E-mail: ABILO@hul.de
Internet: http://hul.de
NIOSH alert: Controlling exposures to nitrous oxide
during anesthetic administration. 1994 Oct;1:Anes-
thesia:9.
Blease
Supplier information Marketed worldwide, except for the United States
AIKA Dameca
Marketed in Asia Marketed worldwide, except for North America
AIKA Medical Group
Emergency Medical Equipment Div [290437] Dameca A/S [156977]
3-15-9 Hongo Islevdalvej 211
Bunkyo-Ku DK-2610 Roedovre Koebenhaven
Tokyo 113 Denmark
Japan Phone: 45 (44) 913480
Phone: 81 (3) 38134415 Fax: 45 (44) 916941
Fax: 81 (3) 38134330 E-mail: info@dameca.com
16 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 17
Healthcare Product Comparison System
Kimura Oxigel
Marketed worldwide, except for the United States Marketed in Africa and Latin America
Kimura Medical Instrument Co Ltd [152416] Oxigel Materiais Hospitalares Industria Comercio
17-5 Yushima 2-chome Ltda [152447]
Bunkyo-ku Rua Engenheiro Jorge Oliva 131
Tokyo 113 Vila Mascote Caixa Postal 21212
Japan 04362-060 Sao Paulo-SP
Phone: 81 (3) 38144481 Brazil
Fax: 81 (3) 38145304 Phone: 55 (11) 55646999
Megamed Fax: 55 (11) 55645013
E-mail: oxigel@originet.com
Marketed worldwide, except for the United States
Internet: http://www.originet.com/oxigel
Megamed AG [171243]
Hinterbergstrasse 11 Penlon
CH-6330 Cham Marketed worldwide
Switzerland Penlon Ltd [139281]
Phone: 41 (41) 7415151 Radley Rd
Fax: 41 (41) 7415144 Abingdon, Oxfordshire OX14 3PH
Mercury Medical England
Marketed in North America Phone: 44 (1235) 554222
Fax: 44 (1235) 555900
Mercury Medical [102449]
E-mail: export@penlon.co.uk
11300 A 49th St N
Internet: http://www.penlon.com
Clearwater FL 33762-4800
Phone: (813) 573-0088, (800) 237-6418 Royal Medical
Fax: (813) 573-7938 Marketed worldwide
Normeca Royal Medical Co Ltd [157039]
Marketed worldwide, except for North America 2 Floor Sung Hwa Bldg
Normeca A/S [162653] 463-3 Seo Kyo-dong Mapo-ku
Postboks 404 Seoul 121-210
N-1473 Skarer Republic of Korea
Norway Phone: 82 (2) 71711513
Phone: 47 (67) 927600 Fax: 82 (2) 3363328
Fax: 47 (67) 971766
Samed
Normeca Asia [321497] Marketed worldwide, except for the United States
Kanda-Blanca Bldg 502
2-18-16 Iwamoto Chiyoda Samed Elettromedicali srl [187040]
Tokyo 101-0032 Uffici e Magazzini
Japan Strada Provinciale 181
Phone: 81 (3) 56873899 I-20067 Merlino MI
Fax: 81 (3) 56873911 Italy
E-mail: normasta@mb.infoweb.or.jp Phone: 39 (02) 90658787
Fax: 39 (02) 90658795
North American Drager
Marketed worldwide, except for the Fabius, which is Siare
marketed worldwide except for North America, and Marketed worldwide, except for the United States
the Narkomed 6000, which is marketed in North Siare Hospital Supplies srl [152520]
America Via G Pastore 18
North American Drager [101982] I-40056 Crespellano BO
3135 Quarry Rd Italy
Telford PA 18969 Phone: 39 (051) 969802
Phone: (215) 721-5400, (800) 462-7566 Fax: 39 (051) 969366
Fax: (215) 721-9561 E-mail: mail@siare.it
Internet: http://www.nad.com Internet: http://www.siare.it
18 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 19
Healthcare Product Comparison System
About the chart specifications Continuing press alarm: Senses sustained elevated air-
The following terms are used in the charts: way pressure that does not periodically fall below a
selectable threshold level.
Pipeline gas inlets: Many listed systems include diame-
Installation: How the anesthesia unit is installed in the
ter-indexed safety system (DISS) pipeline connec-
location of its intended use.
tions and pin-indexed cylinders and yokes.
Abbreviations:
Suction system: The ability to provide suction from the
anesthesia machine. AGSS — Anesthesia gas scavenging system
O2 fail-safe: A safety device that protects the patient ANSI — American National Standards Institute
from inadequate O2. If the O2 supply pressure drops
below about 25 to 30 psi, the unit decreases or shuts APL — Adjustable pressure-limiting
off the flow of other gases and activates an alarm. ASTM — American Society for Testing and Materials
Hypoxic mixture fail-safe: A safety device that prevents BB — Blood bank, black box
an anesthesia machine from being set to dispense a
hypoxic mixture. BIS — Bispectral Index monitor
Bellows: Automatic ventilator bellows have an ascend- BP — Blood pressure
ing or descending design. The bag-in-bottle principle
is based on an equilibrium between the tank pressure BS — British Standard
and the bottle pressure during inspiration; during BSI — British Standards Institution
expiration, the patient exhales into the bag while the
tank is refilled for the following inspiration. CE mark — Conformite Europeene mark
Ventilation modes: The modes of ventilation available CMV — Controlled mechanical ventilation
with the anesthesia unit (e.g., manual ventilation,
CPAP — Continuous positive airway pressure
controlled mechanical ventilation, intermittent
mandatory ventilation). CRT — Cathode ray tube
Tidal volume, range, cc: A control adjusts the volume CSA — Canadian Standards Association
of individual breaths within this range.
DEKRA — Institut fuer Sicherheit, Umweltschutz
Minute volume, range, L/min: A control adjusts the total und Energie
inspiratory volume per minute delivery from the
bellows. DIN — Deutsches Institut fuer Normung
Frequency, bpm: The respiratory frequency can be set DIS — Draft international standard
within the listed range of breaths per minute.
ECG — Electrocardiogram
Inspiratory flow, L/min: The flow range of gas that the
ventilator is capable of delivering to the patient. EEG — Electroencephalogram
20 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
PVC — Pressure/volume control For those models whose prices were supplied to us
in currencies other than U.S. dollars, we have also
ROM/RAM — Read-only memory/random access listed the conversion to U.S. dollars to facilitate com-
memory parison among models. However, keep in mind that
exchange rates change often.
SCMV — Synchronized controlled mechanical ven-
tilation Need to know more?
SEV — Schweizerischer Electrotechnischer Verein For further information about the contents of this
Product Comparison, contact the HPCS Hotline at +1
SIMV — Synchronized intermittent mandatory (610) 825-6000, ext. 5265; +1 (610) 834-1275 (fax); or
ventilation hpcs@ecri.org (e-mail).
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 21
Healthcare Product Comparison System
GAS CYLINDER YOKES 2 (O2, N2O) 2 (O2, N2O) 2 (O2, N2O) 3 (2 O2, N2O)
VAPORIZERS
Agents Sevoflurane, iso- Sevoflurane, iso- Sevoflurane, iso- Isoflurane, halo-
flurane, enflurane flurane, halothane, flurane, halothane, thane, enflurane,
enflurane enflurane sevoflurane
Type Variable bypass Variable bypass Variable bypass Variable bypass
Number 3 3 3 2
Interlock Yes Yes Yes Yes
SUCTION SYSTEM Optional Optional Optional No
O2 FAIL-SAFE Yes Yes Yes Full gas cutoff
HYPOXIC MIXTURE
FAIL-SAFE 30% O2 30% O2 30% O2 25% O2
Primary controls
Ventilation modes Not specified Not specified Not specified Assist/cont, P,
preset limit
Tidal volume Yes Yes Yes Yes
Range, cc 200-1,000 200-1,300 200-900 : 0-2,660 50-1,200
Minute volume Yes Yes Yes : Not specified No
Range, L/min 1-20 2-33 1-20 : Not specified NA
Inspiratory flow,
L/min 3-648 5-100 5-65 : 3-40 10-50
I:E ratio 1:1 to 1:3 1:1 to 1:3 1:0.5 to 1:5 : 1:0.5 to 1:4
1:0.1 to 1:9.9
Inspiratory pause Not specified Not specified Not specified Adjustable
Pressure-limit,
cm H2O Not specified Not specified 40 : 15-65 5-55
PEEP, cm H2O 0-20 0-20 0-20 0-20
Other controls None specified None specified ACE-3000 has an Electronic flow
inspired time control
control
22 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
MONITORS
Airway pressure Yes Yes No : Yes Yes
Where measured Not specified Not specified NA : Not specified Ventilator, valve
Hi-pressure alarm Yes Yes NA : Not specified Yes
Subatmospheric
pressure alarm 5-60 cm H2O 5-60 cm H2O NA : Yes Yes
Continuing press
alarm No No NA : No No
Response time, sec Not specified Not specified Not specified 15 (90% of the time)
CO2 concentration No No Yes No
Apnea alarm NA NA Not specified NA
N2O No No Not specified No
Agent monitors Quartz Optional Optional No
Type of agents Sevoflurane, iso- Sevoflurane, iso- Sevoflurane, iso- NA
flurane, halothane, flurane, halothane, flurane, halothane,
enflurane enflurane enflurane
Auto ID Yes No No NA
Agent conc alarm Yes Not specified Not specified NA
ECG No No No No
Heart rate NA NA NA NA
ST segment NA NA NA NA
Noninvasive BP No No No No
Invasive BP No No No No
Temperature No No No No
Pulse oximeter Not specified Not specified Not specified No
Other monitors None specified None specified None specified None
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 23
Healthcare Product Comparison System
PRIORITIZED ALARMS No No No No
PHYSICAL FEATURES
H x W x D, cm (in) 148 x 63 x 79 142 x 71 x 74 148 x 60 x 65 158 x 100 x 72
(58.3 x 24.8 x 31.1) (56.9 x 27.9 x 29.1) (58.3 x 23.6 x 25.6) (62.2 x 39.4 x 28.3)
Delivery time, ARO Not specified Not specified Not specified 30 days
OTHER SPECIFICATIONS Auxiliary shelves. Auxiliary shelves. Auxiliary shelves. None specified.
24 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
GAS CYLINDER YOKES Optional 2 (O2, N2O) 2 (O2, N2O) 3 max : 4 max :
5 max *
VAPORIZERS
Agents Isoflurane, haloth- Isoflurane, haloth- Isoflurane, haloth- Sevoflurane, iso-
ane, enflurane, des- ane, enflurane, des- ane, enflurane, des- flurane, halothane,
flurane, sevoflurane flurane, sevoflurane flurane, sevoflurane enflurane
Type Temperature Variable bypass Variable bypass Plenum-type variable
compensated bypass, fully
compensated
Number 2/active : 3/active 1 2 1
Interlock Yes Yes Yes Fully compatible
SUCTION SYSTEM Optional Optional Optional Opt, direct/venturi
O2 FAIL-SAFE Yes Acoustic with N2O Yes Multigas cutoff,
shutoff full alarms **
HYPOXIC MIXTURE
FAIL-SAFE Yes Ratio system O2 >20% Yes Gear-driven ratio
in fresh gas system ***
AUTOMATIC VENTILATOR Optional Optional Yes Blease 8200s
Inspiratory pause Not specified 50% maximum 50% maximum Not specified
Pressure-limit,
cm H2O 10-65 adjustable 10-60 mbar 10-60 mbar 10-70 adjustable
PEEP, cm H2O 3-20 variable 0-16 mbar 0-16 mbar 0-20 variable
Other controls None specified Adjustable pressure, Adjustable pressure, Adult and pediatric
limitable EVK, compensation, modes, standby in
compliance, both modes, spirom-
volume control etry MV/TV selection
System checks None specified Leak test Leak test Self-verification
test
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 25
Healthcare Product Comparison System
MONITORS
Airway pressure Yes Yes Yes Peak and mean
Where measured Y-piece, bellows blk Inspiratory side Not specified Y-piece
Hi-pressure alarm Adjustable 10-60 mbar Yes 10-70 cm H2O,
adjustable
Subatmospheric
pressure alarm Yes No No -10 cm H2O, fixed,
internal
Continuing press
alarm Yes 2.5-60 mbar Yes PEEP referenced
26 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Weight, kg (lb) 121 (267) : 22 (48.5) 110 (242.6) 85 (187) : 104 (229)
153 (337) : 130 (287) **
Shelves, cm [56.5 : 34] x 34 None 27 x 61 [50:62:75] x 44 (2);
[45:47:74] x 50 (1)
Drawers, cm 10.5 x 37 x 37 (3) None 57 x 60 (3) 35 x 10 x 35
(4 maximum)
Writing shelf, cm 37 x 37 None 42 x 57 26 x 35 (1) pull-out
writing table
Installation Mobile Mobile Mobile Mobile, wall,
ceiling
POWER REQUIRED, VAC 110/220 115-230 115-230 110/120/220/240
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 27
Healthcare Product Comparison System
WHERE MARKETED Worldwide, except Worldwide, except Worldwide, except Worldwide, except
North America North America North America North America
FDA CLEARANCE No No No No
CE MARK (MDD) Yes Yes Yes Yes
YEAR 2000 COMPLIANT Yes Yes Yes Yes
PIPELINE GAS INLETS 3 or 4 3 or 4 3 or 4 3 or 4 (O2, N2O,
air, vacuum)
GAS CYLINDER YOKES 2 or 3 2 (O2, N2O) 1, 2, 3, 4, 5 3 (O2, N2O, CO2)
VAPORIZERS
Agents Sevoflurane, enflur- Sevoflurane, enflur- Sevoflurane, enflur- Sevoflurane, enflur-
ane, halothane, iso- ane, halothane, iso- ane, halothane, iso- ane, halothane, iso-
flurane, desflurane flurane, desflurane flurane, desflurane flurane, desflurane
Type Temperature Temperature Temperature Temperature
compensated compensated compensated compensated
HYPOXIC MIXTURE
FAIL-SAFE Yes Yes Yes Yes
Inspiratory flow,
L/min 2-15 2-15 2-15 2-15
I:E ratio 1:1, 1:2, 1:3 1:2 1:2 1:1, 1:2, 1:3
Inspiratory pause Not specified Not specified Not specified Not specified
Pressure-limit,
cm H2O 25-85 adjustable 25-85 adjustable 25-85 adjustable 25-85
PEEP, cm H2O Optional Optional Optional Optional
Other controls None specified None specified None specified None specified
System checks None specified None specified None specified None specified
28 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
MONITORS
Airway pressure Yes Optional Optional No
Where measured Y-piece Not specified Not specified NA
Hi-pressure alarm Yes Optional Optional NA
Subatmospheric
pressure alarm Yes Optional Optional NA
Continuing press
alarm Yes Optional Optional NA
Other features Stored trends None specified None specified None specified
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 29
Healthcare Product Comparison System
PHYSICAL FEATURES
H x W x D, cm (in) 148 x 80 x 64 160 x 40 x 40 148 x 80 x 60 160 x 80 x 64
(58.2 x 31.5 x 25.2) (63 x 15.75 x 15.75) (58.2 x 31.5 x 23.6) (63 x 31.5 x 25.2)
Shelves, cm 55 x 45 40 x 30 55 x 40 55 x 40
Drawers, cm 55 x 35 40 x 30 45 x 35 55 x 40
Writing shelf, cm 40 x 35 20 x 30 27 x 37 20 x 30
Delivery time, ARO Not specified Not specified Not specified Not specified
OTHER SPECIFICATIONS Gas mixer; Gas mixer; Gas mixer; Gas mixer;
3 color options of 3 color options of 3 color options of 3 color options of
red, white, or blue; red, white, or blue; red, white, or blue; red, white, or blue;
manual ventilation manual ventilation manual ventilation manual ventilation
through optional through optional through optional through optional
bellows. Meets bellows. Meets bellows. Meets bellows. Meets
requirements of IEC requirements of IEC requirements of IEC requirements of IEC
601-1 and ISO 5358. 601-1 and ISO 5358. 601-1 and ISO 5358. 601-1 and ISO 5358.
30 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
AESTIVA 3000 AS/3 Anesthesia Excel 110 SE (Model Excel 110 SE (Model
Delivery Unit (ADU) 7000 ventilator) 7800 & 7900 vents)
WHERE MARKETED Worldwide Worldwide Asia, Canada, Latin Worldwide
America, Japan, USA
FDA CLEARANCE Yes Yes Yes Yes
CE MARK (MDD) Yes Yes No Yes
YEAR 2000 COMPLIANT Yes Yes Yes Yes
PIPELINE GAS INLETS 3 (O2, N2O, air) 3 (O2, N2O, air) 2 (O2, N2O) 2 (O2, N2O)
GAS CYLINDER YOKES Up to 5 (O2, N2O, 3 (2 O2, N2O) 3 (O2, N2O, air) 3 (O2, N2O, air)
air, CO2, heliox)
VAPORIZERS
Agents Sevoflurane, enflur- Sevoflurane, enflur- Sevoflurane, enflur- Sevoflurane, enflur-
ane, halothane, iso- ane, halothane, iso- ane, halothane, iso- ane, halothane, iso-
flurane, desflurane flurane, desflurane flurane, desflurane flurane, desflurane
Type See footnote * Electronically See footnote * See footnote *
controlled variable
bypass
Number 2 or 3 1 active, 3 storage 2 2
Interlock Yes NA Yes Yes
SUCTION SYSTEM Optional Optional Optional Optional
O2 FAIL-SAFE Pneumatic Yes Pneumatic Pneumatic
HYPOXIC MIXTURE
FAIL-SAFE Mechanical limit Maintains ~25% Mechanical limit Mechanical limit
O2 of total gas flow
AUTOMATIC VENTILATOR AESTIVA Smartvent Yes Model 7000 Model 7800 :
Model 7900
Bellows Yes Yes Yes Yes
Size 1,500 mL Adult/pediatric 1,500 mL 1,500 mL
Type Ascending (standing) ** Ascending ** Ascending (standing) ** Ascending (standing) **
Primary controls
Ventilation modes Volume, pressure Volume/pressure con- Volume Volume : Volume and
trol, SIMV, manual pressure
Tidal volume Yes Yes No Yes
Range, cc 20-1,500 20-1,400 NA 50-1,500 : 20-1,500
Minute volume No Yes Yes No
Range, L/min NA 0.1-30 2-30 NA
Inspiratory flow,
L/min NA 80 maximum NA 10-100 : NA
I:E ratio 2:1 to 1:8 1:4.5 to 2:1 1:1 to 1:3 NA : 2:1 to 1:8
System checks Manual pre-use test, Circuit leakage & Manual pre-use test, Manual pre-use test,
ventilator self-test compliance, hypoxic ventilator self-test ventilator self-test
mixture control
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 31
Healthcare Product Comparison System
AESTIVA 3000 AS/3 Anesthesia Excel 110 SE (Model Excel 110 SE (Model
Delivery Unit (ADU) 7000 ventilator) 7800 & 7900 vents)
SCAVENGING SYSTEM Active or passive Active or passive Active or passive Active or passive
MONITORS
Airway pressure Yes Yes Yes * Yes
Where measured Inspiratory limb Y-piece, bellows blk Inspiratory limb Inspiratory limb
Hi-pressure alarm Yes Adjustable Yes Yes
Subatmospheric
pressure alarm Yes Yes Yes Yes
Continuing press
alarm Yes Yes Yes Yes
32 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
AESTIVA 3000 AS/3 Anesthesia Excel 110 SE (Model Excel 110 SE (Model
Delivery Unit (ADU) 7000 ventilator) 7800 & 7900 vents)
DISPLAYS AESTIVA Smartvent Yes Monitoring pod 7800 : 7900
Number 1 1 1 1
Type ICE Color LCD LCD LCD : ICE
Integrated Yes Yes Yes Yes
Interface w/others RS232 Yes HP ACMS, Omnicare 24 HP ACMS, Omnicare 24
DATA INPUT No Membrane switches, No No
ComWheel, flow knobs
PRIORITIZED ALARMS Yes 3 (caution, advi- No Yes
sory, warning)
PHYSICAL FEATURES
H x W x D, cm (in) 135.8 x 75 x 83 154 x 73 x 80 166.4 x 50.7 x 68 166.4 x 50.7 x 68
(53.5 x 29.5 x 32.7) (60.6 x 28.7 x 31.5) (65.5 x 20 x 26.8) (65.5 x 20 x 26.8)
Weight, kg (lb) 136 (300) 110 (242); 130 (286) 122 (269) 122 (269)
with AS/3 Monitoring
Shelves, cm [48/68/88/] x 35 or 62 x 40 (1) 54.5 x 36 (1); 54.5 x 36 (1);
[19/27/35] x 14 28 x 30 (1) 28 x 30 (1)
Drawers, cm 10.5 x 38.5 x 26; 11 x 42 x 36 (2, 39.4 x 26.7 x 10.2 39.4 x 26.7 x 10.2
14.5 x 38.5 x 26 opt optional) (2) (2)
Writing shelf, cm 38.5 x 26 51 x 30 (1) No No
Installation Mobile with locking Mobile with locking Mobile with locking Mobile with locking
casters, pendant casters, pendant casters casters
POWER REQUIRED, VAC 100/120; optional 110/240 100/120; optional 100/120; optional
220/240 220/240 220/240
Auxiliary outlets 4 3 4-6 4-6
BACKUP BATTERY In ventilator Yes No In ventilator
Type Gel cell Lead acid NA Ni-Cd; gel cell
Use per charge, hr 30 min ~0.5 NA 0.5
PURCHASE INFORMATION
Price $39,340-52,000 Not specified $12,200-35,000 $12,200-35,000
Delivery time, ARO 4 weeks Not specified 4-8 weeks 4-8 weeks
OTHER SPECIFICATIONS Options include 2 or Trends of fresh gas, Options include Options include
3 in-line vapori- agent, & ventilator auxiliary O2 flow- auxiliary O2 flow-
zers, pendant or settings; measures meter, freestanding meter, freestanding
trolley, choice of fresh-gas & agent Bain circuit Bain circuit
ventilator display consumption; total adapter, GMS Bain adapter, GMS Bain
alarms, breathing fresh gas flow & O2/ circuit adapter, GMS circuit adapter, GMS
system bag alarms, N2O percentages dis- absorber, IV pole, absorber, IV pole,
number of cylinder played; automatic cable management cable management
yokes, auxiliary switch to secondary arm, shelf angle arm, shelf angle
common gas outlet, driving gas of vent- blocks, suction reg- blocks, suction reg-
CO2 bypass, gain ilator; software ulators, vaporizer ulators, vaporizer
module, integrated selects driving gas; storage brackets, storage brackets,
suction, O2 flowmet- service menu w/error mounting for mounting for
er, auxiliary light, log; service soft- monitors. Meets monitors. Meets
IV posts, mounting. ware. Meets requirements of requirements of
Meets requirements requirements of ASTM ASTM F1850, CSA, ASTM F1850, CSA,
of ASTM F1850, CSA, F29.01.09, ISO/DIS JIS, and UL. EN 740, JIS, and UL.
EN 740, JIS, and UL. 8835-1, and EN 740.
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 33
Healthcare Product Comparison System
Number 2 2 2 3
Interlock Yes Yes Yes Yes
SUCTION SYSTEM Optional Optional Optional Optional
O2 FAIL-SAFE Pneumatic Pneumatic Pneumatic Pneumatic
HYPOXIC MIXTURE
FAIL-SAFE Mechanical limit Mechanical limit Mechanical limit Mechanical limit
Primary controls
Ventilation modes Volume Volume : Volume and NA Volume control
pressure
Tidal volume No Yes NA Yes
Range, cc NA 50-1,500 : 20-1,500 NA 50-1,500
Minute volume Yes No NA No
Range, L/min 2-30 NA NA NA
Inspiratory flow,
L/min NA 10-100 : NA NA 10-100
I:E ratio 1:1 to 1:3 NA : 2:1 to 1:8 NA NA
System checks Manual pre-use test, Manual pre-use test, Manual pre-use test Manual pre-use test,
ventilator self-test ventilator self-test system self-test,
ventilator self-test
34 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Subatmospheric
pressure alarm Yes Yes NA Yes
Continuing press
alarm Yes Yes NA Yes
Other features Trends, printing, Trends, printing, None specified Alarm management
networking available networking available options
with AS/3 monitoring with AS/3 monitoring
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 35
Healthcare Product Comparison System
PHYSICAL FEATURES
H x W x D, cm (in) 158.1 x 69.2 x 69.9 158.1 x 69.2 x 69.9 158 x 73.6 x 76 158 x 94 x 71
(62.3 x 27.3 x 27.5) (62.3 x 27.3 x 27.5) (62.2 x 29 x 29.9) (62.2 x 37 x 28)
Weight, kg (lb) 116 (255.8) 116 (255.8) 93.5 (206.1) 210 (463.1)
Shelves, cm 61.6 x 38.1 (1); 61.6 x 38.1 (1); 59.7 x 36 (1); 92.7 x 38 (1);
38.1 x 39.3 (1) 38.1 x 39.3 (1) 29 x 31 (1) 37 x 30 (1)
Drawers, cm 47.6 x 34.3 x 14 (up 47.6 x 34.3 x 14 (up 39.4 x 26.7 x 10.2 36 x 38 x 10.1 (2);
to 3 optional) to 3 optional) (2) 36 x 38 x 20.3 (1)
Writing shelf, cm 46.5 x 36.8, 46.5 x 36.8, No No
optional optional
Installation Mobile with locking Mobile with locking Mobile with locking Mobile with locking
casters casters casters casters
POWER REQUIRED, VAC 100/120; optional 100/120; optional None 100/120; optional
220/240 220/240 220/240
Auxiliary outlets 4-6 4-6 NA 4
BACKUP BATTERY No In ventilator No UPS; min system func
Type NA Ni-Cd; gel cell NA Gel cell
Use per charge, hr NA 0.5 NA See footnote *
PURCHASE INFORMATION
Price $14,300-40,000 $14,300-40,000 $20,500 $62,900-65,100
Delivery time, ARO 4-8 weeks 4-8 weeks 4-8 weeks 4-8 weeks
OTHER SPECIFICATIONS Options include Options include Options include Alert zones and
auxiliary O2 flow- auxiliary O2 flow- auxiliary O2 flow- trend advisories;
meter, freestanding meter, freestanding meter, freestanding 3.5" disk drive;
Bain circuit Bain circuit Bain circuit optional absorber
adapter, GMS Bain adapter, GMS Bain adapter, GMS Bain with latching fresh-
circuit adapter, GMS circuit adapter, GMS circuit adapter, GMS gas inlet; tilting
absorber, IV pole, absorber, IV pole, absorber, IV pole, top shelf; battery
cable management cable management add-on drawers, backup/failure
arm, shelf angle arm, shelf angle auxiliary shelves. alarm; output panel;
blocks, suction reg- blocks, suction reg- guarded O2 flush;
ulators, vaporizer ulators, vaporizer integral mounting
storage brackets, storage brackets, rail/footrest;
mounting for mounting for options: auxiliary
monitors. Meets monitors. Meets O2 flowmeter, free-
requirements of requirements of stand Bain circuit
ASTM F1850, CSA, ASTM F1850, CSA, adapter, GMS
JIS, and UL. EN 740, JIS, and UL. absorber, IV pole,
auxiliary shelves. **
36 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Number 3 3 1 or 2 1 or 2
Interlock Yes Yes Yes Yes
SUCTION SYSTEM Optional Optional Optional Yes
O2 FAIL-SAFE Pneumatic Pneumatic Acoustic alarm, Acoustic alarm,
N2O lock N2O lock
HYPOXIC MIXTURE
FAIL-SAFE Mechanical limit Mechanical limit O2 ratio controller O2 ratio controller
Primary controls
Ventilation modes Volume control Volume : Volume and IPPV, SIMV, manual, IPPV, SIMV, manual,
pressure PCV, spontaneous PCV, spontaneous
Tidal volume Yes Yes Yes Yes
Range, cc 50-1,500 50-1,500 : 20-1,500 10-1,400 10-1,400
Minute volume No No Yes Yes
Range, L/min NA NA 0.06-25 0.06-25
Inspiratory flow,
L/min 10-100 10-100 : NA 75 maximum 75 maximum
I:E ratio NA NA : 2:1 to 1:8 1:3 to 2:1 1:3 to 2:1
System checks Manual pre-use test, Manual pre-use test, Leakage Leakage
system self-test, ventilator self-test
ventilator self-test
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 37
Healthcare Product Comparison System
Subatmospheric
pressure alarm Yes Yes Mean pressure Mean pressure
or airway pressure or airway pressure
Continuing press
alarm Yes Yes Yes Yes
Other features Alarm management Trends, printing, None specified None specified
options networking available
with AS/3 monitoring
38 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Weight, kg (lb) 220 (485.1) 139.3 (309.5) 140 (308.7) 200 (441)
Shelves, cm 92.7 x 38 (1); 80.6 x 37 (1); Not specified (2) Not specified (2)
37 x 30 (1) 43 x 40 (1)
Drawers, cm 36 x 38 x 10.1 (2); 14 x 47.6 x 34.3 (up Not specified (1) Not specified (1)
36 x 38 x 20.3 (1) to 3 optional)
Writing shelf, cm No 47.6 x 36.8 (1, 37 x 42 37 x 42
optional)
Installation Mobile with locking Mobile with locking Mobile, pendant Mobile, pendant
casters casters
POWER REQUIRED, VAC 100/120; optional 110/120; optional 110/230 110/230
220/240 220/240
Auxiliary outlets 4 6 3 4
BACKUP BATTERY UPS; min system func In ventilator Optional Optional
Type Gel cell Ni-Cd; gel cell Not specified Not specified
Use per charge, hr See footnote * 0.5 1 1
PURCHASE INFORMATION
Price $75,750-77,950 $21,000-47,000 Not specified Not specified
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 39
Healthcare Product Comparison System
Number 1 or 2 2 3 1
Interlock Yes Yes NA Yes
SUCTION SYSTEM Optional Yes Optional Optional
O2 FAIL-SAFE Acoustic alarm, Acoustic alarm, Acoustic alarm, Acoustic with N2O
N2O lock N2O lock N2O lock shutoff
HYPOXIC MIXTURE
FAIL-SAFE O2 ratio controller O2 ratio controller Minimal setting Ratio system, O2
O2 20% >20% in fresh gas
AUTOMATIC VENTILATOR Electronically Electronic controls, Computer controlled Yes
driven/controlled pneumatically driven
Bellows Yes Yes Yes Yes
Size Adult/pediatric Adult/pediatric Adult/pediatric Adult/pediatric
Type Piston unit Descending, bag in 4-membrane chamber Descending, bag in
bottle bottle
Primary controls
Ventilation modes IPPV, manual, spont, PCV, IPPV, manual, IPPV, PCV, manual, Manual, CMV
pressure limited spontaneous spontaneous
Tidal volume Yes Yes Yes Yes
Range, cc 50-1,400 50-1,400 20-2,000 50-1,500
Minute volume Yes Yes Yes No
Range, L/min 0.3-25 0.3-45 0.12-25 NA
Inspiratory flow,
L/min 75 maximum 5-75 adjustable 9-90 0.5-40
I:E ratio 1:3 to 2:1 1:4 to 2:1 1:4 to 4:1 1:3 to 2:1
System checks None specified Compliance, leaks System test, leakage None specified
compensation,
compromise, volume
40 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
MONITORS
Airway pressure Piezoresistive Piezoresistive Yes Yes
Where measured System pressure Gas inlet System Inspiratory side
Hi-pressure alarm 5-78 mbar 5-98 mbar 0-75 cm H2O 20-80 mbar
Subatmospheric
pressure alarm Airway pressure Mean pressure or Not specified No
airway pressure
Continuing press
alarm Yes Yes Yes 2.5-60 mbar
Other features None specified Stored trends, hard- Quantitative anes- None
copy printouts thesia; end-tidal
control, ALICE, BB
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 41
Healthcare Product Comparison System
42 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Number 2 3 maximum 2 2
Interlock Yes Not specified Yes Yes
SUCTION SYSTEM Optional Yes No Not specified
O2 FAIL-SAFE Acoustic with N2O Yes Pneumatic Yes
shutoff
HYPOXIC MIXTURE
FAIL-SAFE Ratio system, O2 Pneumatic Yes Minimum 30% O2
>20%
AUTOMATIC VENTILATOR Yes Yes Optional (AV-500) Yes
Inspiratory flow,
L/min 1-80 Yes No 1.2-15
I:E ratio 1:1, 1:1.5, 1:3, 3:1 to 1:9 1:1 to 1:4 1:1.5, 1:2, 1:3
2:1, 3:1
Inspiratory pause 5 sec Yes No Not specified
Pressure-limit,
cm H2O 20-80 adjustable Yes 7 fixed Not specified
PEEP, cm H2O 0-20 4-20 No Optional
Other controls Plateau sigh, 30 sec None specified Float-type flowmeter None specified
expiratory pause,
volume constant
ventilation
System checks Not specified Complete circuit, None None specified
pre-use test for
leaks
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 43
Healthcare Product Comparison System
Subatmospheric
pressure alarm Yes (pressure-relief No No No
valve included)
Continuing press
alarm Yes Yes No No
44 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Delivery time, ARO 4-6 weeks 6 weeks Not specified Not specified
OTHER SPECIFICATIONS Automatic compensa- SIMV; PCV; total None specified. None specified.
tion for patient circuit compliance.
system compliance;
O2 flush; fresh-gas
decoupling; auto
Vt constant; patient
circuit; integrated
compact block
heating device to
avoid condensation;
low-flow ability;
minimal-flow
ability.
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 45
Healthcare Product Comparison System
WHERE MARKETED Worldwide, except Worldwide, except North America Worldwide, except
USA USA North America
FDA CLEARANCE No No Yes Not specified
CE MARK (MDD) Yes Yes Yes Not specified
YEAR 2000 COMPLIANT Yes Yes Yes Not specified
PIPELINE GAS INLETS 3 2 or 3 3 (O2, N2O, air) Draw-over :
2 (O2, N2O)
GAS CYLINDER YOKES 2 (O2, N2O) 2 (O2, N2O) 4 (2 O2, N2O, air) No
VAPORIZERS
Agents Sevoflorane, enflur- Sevoflorane, enflur- Sevoflorane, enflur- Halothane,
ane, halothane, iso- ane, halothane, iso- ane, halothane, iso- enflurane, trilene,
flurane, desflurane flurane, desflurane flurane, desflurane isoflurane
Type Variable bypass Variable bypass Variable bypass, Variable bypass
heated desflurane
Number 1 1 3 1
Interlock Yes Yes Yes No
SUCTION SYSTEM Optional Optional Yes No
O2 FAIL-SAFE Yes Yes Acoustic with N2O No
shutoff
HYPOXIC MIXTURE
FAIL-SAFE With alarm With alarm Ratio system O2, No
<20% in fresh gas
AUTOMATIC VENTILATOR Yes Yes Yes No
Inspiratory flow,
L/min 0-90 Not specified 0.5-80 NA
I:E ratio 1:3 to 2:1 1:3 to 3:1 1: 4, 1:3, 1:2, 1:1, NA
2:1, 3:1
Inspiratory pause Yes 0-30% of insp time 20/30% insp time NA
Pressure-limit,
cm H2O 5-90 0-80 10-100 NA
PEEP, cm H2O 0-20 mechanical 0-20 mechanical 0-20 electronic NA
46 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Subatmospheric
pressure alarm No No 2 mbar NA
Continuing press
alarm Yes Yes Yes NA
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 47
Healthcare Product Comparison System
Weight, kg (lb) 120 (264.6) 138 (304.3) 170 (375) 6.6 (14.5) :
10.8 (23.8)
Shelves, cm 54 x 39 (1) 54 x 30 (1) 652" No : 19 x 36 (1)
Auxiliary outlets 4 4 4 NA
BACKUP BATTERY No Yes Yes NA, pneumatic
Type NA Lead acid Lead acid NA
Use per charge, hr NA 1 30 min NA
PURCHASE INFORMATION
Price Not specified Not specified Not specified Not specified
Delivery time, ARO Not specified Not specified Not specified 4-8 weeks
OTHER SPECIFICATIONS Proportional Microprocessor Fresh gas decoupled; Specially designed
flowmeter for O2 controlled; tidal volume cons- for use in field
(ORC); one bellows proportional flow- tant; compliance- hospitals and
type for adults and meter for O2 (ORC); compensated system similar locations;
children; pressure- one bellows type for and circuit; inte- MP-1 is a basic
controlled IPPV. adults and children; grated absorber/ draw-over machine;
Meets requirements SIMV and assist mode bellows; heated MP-2 includes by-
of DEKRA, IEC 601-1, ventilation patient module pass valve and
MedGV, SEV, TUV, and included. Meets (inspiratory/expira- rotameter box.
VDE. requirements of tory); autoclavable;
DEKRA, IEC 601-1, software driven; MS-
MedGV, SEV, TUV, and DOS-based ventilator
VDE. design; sample gas
recirculation inlet;
low-flow capability.
Meets requirements
of CSA and
UL 2101-1.
48 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
HYPOXIC MIXTURE
FAIL-SAFE No Yes Yes Yes
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 49
Healthcare Product Comparison System
Subatmospheric
pressure alarm NA Yes Yes Yes
Continuing press
alarm NA Yes Yes Yes
Other features None specified None specified Stored trends log, Stored trends log,
low fresh gas alarm, low fresh gas alarm,
compliance trend compliance trend
50 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
Drawers, cm No 2 1 1
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 51
Healthcare Product Comparison System
Number 1 1 1 2
Interlock NA NA NA Yes
SUCTION SYSTEM Optional Optional Optional Yes
O2 FAIL-SAFE Yes Yes Yes Servomatic type
HYPOXIC MIXTURE
FAIL-SAFE Yes Yes Yes SF-25
Primary controls
Ventilation modes IPPV, manual, IPPV, manual, IPPV, manual, Manual, mechanical/
spontaneous spontaneous spontaneous spontaneous
Tidal volume Yes Yes Yes Yes
Range, cc 50-1,500 20-1,500 50-1,500 0-1,000
Minute volume Yes Yes No Yes
Range, L/min Not specified Not specified Not specified 2-40
Inspiratory flow,
L/min 10-100 10-100 10-100 Not specified
I:E ratio 4:1 to 1:4.5 4:1 to 1:4.5 4:1 to 1:4.5 Not specified
System checks Pre-use system check Pre-use system check None Manometers with
O2 alarm illuminated
52 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
MONITORS
Airway pressure Yes Yes Yes Yes (O2)
Where measured Absorber/Y-piece Absorber/Y-piece Absorber Not specified
Hi-pressure alarm Yes Yes Yes Not specified
Subatmospheric
pressure alarm Yes Yes Yes Not specified
Continuing press
alarm Yes Yes Yes Not specified
Response time, sec Not specified Not specified Not specified Not specified
CO2 concentration Not specified Not specified No Not specified
Apnea alarm Not specified Not specified NA Not specified
N2O Not specified Not specified No Not specified
Agent monitors No No No Not specified
Type of agents NA NA NA Not specified
Other features None specified None specified None specified None specified
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 53
Healthcare Product Comparison System
PRIORITIZED ALARMS Warning, caution, Warning, caution, Warning, caution, Not specified
advisory advisory advisory
PHYSICAL FEATURES
H x W x D, cm (in) 53 x 128 x 41 60 x 136 x 62 80 x 135.2 x 68.6 145 x 68 x 58
(20.9 x 50.4 x 16.1) (23.6 x 53.5 x 24.4) (31.5 x 53.2 x 27) (57 x 26.8 x 22.8)
Writing shelf, cm Not specified Not specified Not specified Not specified
POWER REQUIRED, VAC 100-240, 50/60 Hz 100-240, 50/60 Hz 100-120, 200-240, 110, 220
50/60 Hz
Auxiliary outlets No No No 4
BACKUP BATTERY Yes Yes Yes No
Type Lead acid Lead acid Lead acid NA
Use per charge, hr 90 min 90 min 6 NA
PURCHASE INFORMATION
Price Not specified Not specified Not specified $7,100
54 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
VAPORIZERS
Agents Halothane, sevo- Halothane, sevo- Halothane, sevo- Halothane, sevo-
flurane, isoflurane, flurane, isoflurane, flurane, isoflurane, flurane, isoflurane,
enflurane enflurane enflurane enflurane
Type Plenum Plenum Plenum Plenum
Number 1, 2, or 3 1, 2, or 3 1, 2, or 3 1, 2, or 3
Interlock Yes Yes Yes Yes
SUCTION SYSTEM Optional Optional Optional Optional
O2 FAIL-SAFE Yes Yes Yes Yes
HYPOXIC MIXTURE
FAIL-SAFE Mechanical Mechanical Mechanical, paramag- Mechanical, paramag-
netic, or fuel cell netic, fuel cell
AUTOMATIC VENTILATOR Optional (V200) Optional (AV800) * Optional (AV800) * Optional (AV800) *
Primary controls
Ventilation modes Flow, inspir/exp Flow, inspir/exp Volume, pressure Volume, pressure
time time
Tidal volume Yes Yes Yes Yes
Range, cc 10-2,000 (adult) 50-2,000 (adult) 20-1,600 (adult) 20-1,600 (adult)
Minute volume 1-30 1-30 No No
Range, L/min Not specified Not specified NA NA
Inspiratory flow,
L/min 0.25-1 L/sec 0.25-1 L/sec NA NA
I:E ratio Variable Variable 1:0.3 to 1:6 1:0.3 to 1:6
Other controls None specified None specified Standby, spontan- Standby, spontan-
eous mode, print eous mode, print
button, spirometry button, spirometry
System checks None specified None specified Ventilator self-test Ventilator self-test
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 55
Healthcare Product Comparison System
SCAVENGING SYSTEM Optional vacuum or Optional vacuum or Optional vacuum or Optional vacuum or
exhaust exhaust exhaust exhaust
AUTO RECORD KEEPER No No No No
ANESTHESIA DATA
MANAGEMENT Optional Optional Optional Optional
MONITORS
Airway pressure Yes Yes Yes Yes
Where measured Inspiratory limb Inspiratory limb Inspiratory limb Inspiratory limb
Hi-pressure alarm Yes Yes Variable Variable
Subatmospheric
pressure alarm Yes Yes Yes Yes
Continuing press
alarm Yes Yes Yes Yes
Auto ID NA NA NA NA
Agent conc alarm NA NA NA NA
ECG Not specified Not specified Not specified Not specified
Heart rate Not specified Not specified Not specified Not specified
ST segment Not specified Not specified Not specified Not specified
Noninvasive BP Not specified Not specified Not specified Not specified
Invasive BP Not specified Not specified Not specified Not specified
Temperature Not specified Not specified Not specified Not specified
Pulse oximeter Not specified Not specified Not specified Not specified
Other monitors None specified None specified None specified None specified
Other features None specified None specified None specified None specified
56 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
PHYSICAL FEATURES
H x W x D, cm (in) Not specified 148 x 62 x 23 148 x 45 x 62 148 x 58 x 62
(58.3 x 24.4 x 9.1) (58.3 x 17.7 x 24.4) (58.3 x 22.8 x 24.4)
POWER REQUIRED, VAC Not specified Not specified 110/240, universal 110/240, universal
power supply power supply
Auxiliary outlets 4 4 4 4
BACKUP BATTERY None No Yes Yes
Type NA NA Sealed lead-acid Sealed lead-acid
Use per charge, hr NA NA 0.5 0.5
PURCHASE INFORMATION
Price Not specified Not specified Not specified Not specified
Delivery time, ARO 4-8 weeks 4-8 weeks 4-8 weeks 4-8 weeks
OTHER SPECIFICATIONS Modular system with Modular system with Modular system with Modular system with
monitoring options; monitoring options; monitoring options; monitoring options;
options include options include options include options include
vaporizer, A100 vaporizer, A100 vaporizer, A100 vaporizer, A100
absorber, coaxial absorber, coaxial absorber, coaxial absorber, coaxial
circuits, IV pole, circuits, IV pole, circuits, IV pole, circuits, IV pole,
cable management cable management cable management cable management
system, suction system, suction system, suction system, suction
regulator and regulator and regulator and regulator and
receivers. Meets the receivers. Meets the receivers. Meets the receivers. Meets the
requirements of BS, requirements of BS, requirements of BS, requirements of BS,
CSA, DIN, ISO, CSA, DIN, ISO, CSA, DIN, ISO, CSA, DIN, ISO,
JIS, and TUV. JIS, and TUV. JIS, and TUV. JIS, and TUV.
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 57
Healthcare Product Comparison System
VAPORIZERS
Agents Halothane, sevo- Halothane, sevo- Halothane, sevo- Halothane, sevo-
flurane, enflurane, flurane, enflurane, flurane, enflurane, flurane, enflurane,
isoflurane isoflurane isoflurane isoflurane
Type Plenum Variable bypass Variable bypass Variable bypass
Number 1, 2, or 3 2 or 3 1 or 2 optional 2 or 3
Interlock Yes Yes Yes Optional
SUCTION SYSTEM Optional Optional Optional Optional
O2 FAIL-SAFE Yes Audible alarm Audible alarm Audible alarm
HYPOXIC MIXTURE
FAIL-SAFE Mechanical, paramag- Oxygen ratio Oxygen ratio Oxygen ratio
netic, fuel cell controller controller controller
AUTOMATIC VENTILATOR Optional (AV800) * Yes Yes Yes
Inspiratory flow,
L/min NA Not specified Not specified Not specified
I:E ratio 1:0.3 to 1:6 1:1 to 1:3 1:1 to 1:3 1:1 to 1:3
Other controls Standby, spontan- Float-type flowmeter Float-type flowmeter Float-type flowmeter
eous mode, print
button, spirometry
System checks Ventilator self-test Tidal volume compli- Tidal volume compli- Tidal volume compli-
ance, pre-use check ance, pre-use check ance, pre-use check
for leaks for leaks for leaks
58 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
SCAVENGING SYSTEM Optional vacuum or Active or passive Active or passive Active or passive
exhaust
AUTO RECORD KEEPER No No No No
ANESTHESIA DATA
MANAGEMENT Optional No No No
MONITORS
Airway pressure Yes No No No
Where measured Inspiratory limb NA NA NA
Hi-pressure alarm Variable NA NA NA
Subatmospheric
pressure alarm Yes NA NA NA
Continuing press
alarm Yes NA NA NA
O2 concentration Yes No No No
Type of sensor Paramagnetic, fuel NA NA NA
cell
Response time, sec 10-20 NA NA NA
CO2 concentration No No No No
Apnea alarm NA NA NA NA
N2O No No No No
Agent monitors No No No No
Type of agents NA NA NA NA
Auto ID NA NA NA NA
Agent conc alarm NA NA NA NA
ECG Not specified No No No
Heart rate Not specified NA NA NA
ST segment Not specified NA NA NA
Noninvasive BP Not specified No No No
Invasive BP Not specified No No No
Temperature Not specified No No No
Pulse oximeter Not specified No No No
Other monitors None specified No No No
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 59
Healthcare Product Comparison System
PHYSICAL FEATURES
H x W x D, cm (in) 148 x 70 x 62 159 x 64 x 68.5 142 x 45 x 60 138 x 62.5 x 23.6
(58.3 x 27.5 x 24.4) (62.6 x 25.2 x 27) (55.9 x 17.7 x 23.6) (54.3 x 24.6 x 9.29)
60 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
VAPORIZERS
Agents Halothane, sevo- Halothane, Sevoflurane, enflur- Sevoflurane, enflur-
flurane, enflurane, enflurane, ane, halothane, iso- ane, halothane, iso-
isoflurane isoflurane flurane, desflurane flurane, desflurane
Type Variable bypass Variable bypass Variable bypass Variable bypass
Number 2 or 3 3 maximum 1, 2 1
Interlock Optional Yes Optional No
SUCTION SYSTEM Optional Optional Optional Optional
O2 FAIL-SAFE Audible alarm Audible Audible Audible and visual
HYPOXIC MIXTURE
FAIL-SAFE Yes No Audible and visual Audible and visual
Inspiratory flow,
L/min Not specified 3-75 1-90 0.2-70
I:E ratio 1:1 to 1:3 1:4 to 2:1 1:4 to 4:1 1:4 to 4:1
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 61
Healthcare Product Comparison System
SCAVENGING SYSTEM Active or passive Optional, venturi, Active optional Optional active
O2 pump
AUTO RECORD KEEPER No No No No
ANESTHESIA DATA
MANAGEMENT No No No No
MONITORS
Airway pressure No Analogic Electronic Electronic
Where measured NA Not specified Y-piece Y-piece
Hi-pressure alarm NA Yes 0-100 cm H2O 0-80 cm H2O
Subatmospheric
pressure alarm NA No No No
Continuing press
alarm NA Yes No No
62 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
PHYSICAL FEATURES
H x W x D, cm (in) Not specified 53 x 53 x 170 63 x 153 x 69 140 x 65 x 65
(20.9 x 20.9 x 67) (24.8 x 60.2 x 27.2) (55.1 x 25.6 x 25.6)
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 63
Healthcare Product Comparison System
GAS CYLINDER YOKES 2 (O2, N2O) 4 (2 O2, N2O, air) 4 (2 O2, N2O, air) 4 (2 O2, N2O, air)
VAPORIZERS
Agents Sevoflurane, enflur- Sevoflurane, enflur- Sevoflurane, enflur- Sevoflurane, enflur-
ane, halothane, iso- ane, halothane, iso- ane, halothane, ane, halothane,
flurane, desflurane flurane, desflurane isoflurane isoflurane
Type Variable bypass Not specified Injection vaporizer Injection vaporizer
Number 1 2 3 4*
Interlock No Yes Yes NA
SUCTION SYSTEM Optional Optional Optional Optional
O2 FAIL-SAFE Audible Yes Audible and visual Yes
alarm
HYPOXIC MIXTURE
FAIL-SAFE Audible and visual Yes Continuous monitor Yes
with audible alarm
AUTOMATIC VENTILATOR Yes Yes Yes Yes
Inspiratory flow,
L/min 1-90 4-72 180 max 0.5-96
I:E ratio 1:4 to 4:1 1:3, 1:2, 1:1 1:3 to 4:1 4:1 to 1:4
64 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
MONITORS
Airway pressure Electronic Yes Yes Yes
Where measured Y-piece Inspiratory limb Insp and exp limbs Insp and exp limbs
Hi-pressure alarm 0-100 cm H2O Yes Yes Yes
Subatmospheric
pressure alarm No No No No
Continuing press
alarm No No Yes Yes
Response time, sec Not specified Not specified Not specified Not specified
CO2 concentration Optional See footnote * Yes See footnote *
Apnea alarm Yes Not specified Yes Not specified
N2O Optional Not specified Yes Not specified
Agent monitors Optional See footnote * Yes See footnote *
Type of agents Sevoflurane, enflur- Sevoflurane, enflur- Sevoflurane, enflur- Sevoflurane, enflur-
ane, halothane, iso- ane, halothane, iso- ane, halothane, ane, halothane,
flurane, desflurane flurane isoflurane isoflurane
Auto ID Yes See footnote * Yes See footnote *
Agent conc alarm Yes See footnote * Yes See footnote *
ECG Yes See footnote * Yes See footnote *
Heart rate Yes See footnote * Yes See footnote *
ST segment Yes See footnote * Yes See footnote *
Noninvasive BP Yes See footnote * Yes See footnote *
Invasive BP Optional See footnote * Yes See footnote *
Temperature Optional See footnote * Yes See footnote *
Pulse oximeter Yes Not specified Yes Not specified
Other monitors None specified Cardiac output, resp Cardiac output, resp Cardiac output, resp
loops, tcO2, tcCO2 loops, tcO2, tcCO2 loops, tcO2, tcCO2
Other features None specified Stored trends, up to Stored trends, up to Stored trends, up to
8 channels of hemo- 8 channels of hemo- 8 channels of hemo-
dynamic monitoring dynamic monitoring dynamic monitoring
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 65
Healthcare Product Comparison System
POWER REQUIRED, VAC 110/220 100/110, 220/240, 100-240, 50/60 Hz 100, 110, 220, 240,
50/60 Hz 50/60 Hz
Auxiliary outlets No 4 optional 4 optional 4 optional
BACKUP BATTERY No Yes Yes Yes
Type NA Lead acid Lead acid Lead acid
Use per charge, hr NA 1 2 1
PURCHASE INFORMATION
Price Not specified Not specified Not specified Not specified
Delivery time, ARO 20-30 days Not specified Not specified Not specified
OTHER SPECIFICATIONS Low-flow system. Close-attached Modular system; same Servo feedback
Meets requirements breathing system bellows used for control of volume
of IEC 601-1. with low compression adults and children. and pressure;
volume; optional pressure support and
agent monitoring. control; assist
(900 C), volume, and
manual control
modes; computer
interface (print,
plot, and trend);
optional lung
mechanics calculator
(compliance and
resistance); agent
monitoring.
66 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
WHERE MARKETED Worldwide, except Worldwide, except Worldwide, except Worldwide, except
USA USA USA USA
FDA CLEARANCE No No No No
CE MARK (MDD) Yes Yes Yes Yes
YEAR 2000 COMPLIANT Yes Yes Yes Yes
PIPELINE GAS INLETS 2 (O2, N2O) 3 (O2, N2O, air) 3 (O2, N2O, air) 3 (O2, N2O, air)
VAPORIZERS
Agents Halothane, sevo- Halothane, sevo- Halothane, sevo- Halothane, sevo-
flurane, isoflurane, flurane, isoflurane, flurane, isoflurane, flurane, isoflurane,
enflurane enflurane enflurane enflurane
Type Variable bypass Variable bypass Variable bypass Variable bypass
HYPOXIC MIXTURE
FAIL-SAFE Yes Yes Yes Yes
Tidal volume No No No No
Range, cc 5-2,000 5-2,000 5-2,000 5-2,000
Minute volume No No No No
Range, L/min NA NA NA NA
Frequency, bpm 5-50; 8-60 5-50; 8-60 5-50; 8-60 5-50; 8-60
Inspiratory flow,
L/min 0-60 0-60 0-60 0-60
I:E ratio Infinitely Infinitely Infinitely Infinitely
adjustable adjustable adjustable adjustable
Inspiratory pause No No No No
Pressure-limit,
cm H2O 60 60 60 60
PEEP, cm H2O Optional Optional Optional Optional
System checks None specifed None specified None specifed None specified
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 67
Healthcare Product Comparison System
MONITORS
Airway pressure With ventilator With ventilator With ventilator With ventilator
Where measured Ventilator output Ventilator output Ventilator output Ventilator output
Hi-pressure alarm Yes Yes Yes Yes
Subatmospheric
pressure alarm No No No No
Continuing press
alarm Optional Optional Optional Optional
Expiratory vol/flow No No No No
Type of sensor NA NA NA NA
Where measured NA NA NA NA
Rate alarm NA NA NA NA
Apnea alarm NA NA NA NA
O2 concentration No No No No
Type of sensor NA NA NA NA
Auto ID NA NA NA NA
Agent conc alarm NA NA NA NA
ECG No No No No
Heart rate NA NA NA NA
ST segment NA NA NA NA
Noninvasive BP No No No No
Invasive BP No No No No
Temperature No No No No
Pulse oximeter No No No No
Other monitors No No No No
68 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
DISPLAYS No No No No
Number NA NA NA NA
Type NA NA NA NA
Integrated NA NA NA NA
Interface w/others NA NA NA NA
DATA INPUT No No No No
PRIORITIZED ALARMS No No No No
PHYSICAL FEATURES
H x W x D, cm (in) 50 x 40 x 26 140.5 x 52 x 44 50 x 60 x 26 150 x 56 x 53
(19.7 x 15.7 x 10.2) (55.3 x 2 x 17.3) : (19.7 x 23.6 x 10.2) (60 x 22 x 20.9)
146 x 67 x 67
(57.5 x 26.4 x 26.4)
Weight, kg (lb) 14 (30.9) 30 (66) : 75 (165) 16 (35.3) 60 (132)
Shelves, cm No 52 x 42 : 58 x 31 60 x 24 47 x 40
Delivery time, ARO 60 working days 60 working days 60 working days 60 working days
OTHER SPECIFICATIONS Portable unit Mobile, modular Wall-mounted unit; Designed for use in
with handles; unit; use in smaller for use in MRI suite; mobile.
pipeline gauges. ORs, casualty anesthesia rooms, Meets requirements
Meets requirements departments, field casuality depart- of BS 4272 and
of BS 4272 and hospitals, and ments, maternity ISO 5369.
ISO 5369. anesthetic rooms : suites, and plaster
Modular, rooms. Meets
transportable unit requirements of
designed for appli- BS 4272 and
cations within OR, ISO 5369.
induction
rooms, and casualty
departments.
Both meet the
requirements of BS
4272 and ISO 5369.
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 69
Healthcare Product Comparison System
WHERE MARKETED Worldwide Asia, Europe Asia, Latin America Asia, Latin America
Primary controls
Ventilation modes CMV, manual/ CMV, SCMV, PVC, IMV, CMV, IMV CMV, SMV, PCV
spontaneous SPVC, IPPV, CPAP
Tidal volume Yes Yes Yes Yes
Range, cc 100-1,500; 10-150 0-1,500; 0-400 ped 50-1,000 20-600
Minute volume Yes Yes Yes Yes
Range, L/min 0-20 0.5-451 2-50 2-75
Inspiratory flow,
L/min Not specified 4.0-1001 50 75
I:E ratio 4:1 to 1:4 1:4 to 2:1 1:1 to 1:5 2:1, 1:3
System checks Leaks, resistance, Electronic system None specified Complacency correct-
compliance, check of ventilator ion, leak check
compensation
70 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
MONITORS
Airway pressure Yes Yes No Yes
Where measured Patient circuit Y-piece, gas inlet NA Not specified
Hi-pressure alarm 10-100 cm H2O Yes NA Yes
Subatmospheric
pressure alarm Measures real-time No NA Yes
Continuing press
alarm Measures Yes NA Yes
continuously
Low pressure/apnea Yes Yes NA Yes
Other press alarms No No No None specified
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 71
Healthcare Product Comparison System
PHYSICAL FEATURES
H x W x D, cm (in) 158 x 90 x 82 140 x 60 x 75 147.8 x 63 x 56.3 150 x 59 x 58
(62.2 x 35.4 x 32.3) (55.1 x 23.6 x 29.5) (58.2 x 24.8 x 22.2) (59.1 x 23.2 x 22.8)
: 121 x 50 x 55 : 143 x 62.5 x 58.5
(47.6 x 19.7 x 21.6) (56.3 x 24.6 x 23)
Weight, kg (lb) 162 (357.2) without 85 (187) : Not 112 (246.4) : 140 (308)
patient monitor specified 64 (140.8)
Shelves, cm Not specified 52 x 30 (1) : 37 x 54.6 (1) 59 x 39 (1)
48 x 35 (1)
Drawers, cm 1 41 x 50 (3) : 35.4 x 33.2 x 10.5 Not specified
41 x 40 (3) (3)
Writing shelf, cm 1 45 x 32 (1) : None No No
72 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
VAPORIZERS
Agents Halothane, ethrane,
isoflurane, sevo-
flurane
Type Variable bypass
Number 1
Interlock Yes
SUCTION SYSTEM Optional
O2 FAIL-SAFE Pneumatic with
N2O cutoff
HYPOXIC MIXTURE
FAIL-SAFE Ratio control,
25% O2
AUTOMATIC VENTILATOR Yes
Bellows Yes
Size Adult/infant
Type Ascending, bag in
bottle
Primary controls
Ventilation modes Manual, CMV, IMV,
SIMV, PEEP, CPAP
Tidal volume Yes
Range, cc 50-1,300
Minute volume Yes
Range, L/min 2-32
Inspiratory flow,
L/min 2-100
I:E ratio 0.5:1, variable up
to 1:4
Inspiratory pause Not specified
Pressure-limit,
cm H2O 2-80 adjustable
PEEP, cm H2O 2-30
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 73
Healthcare Product Comparison System
MONITORS
Airway pressure Yes
Where measured Not specified
Hi-pressure alarm Yes
Subatmospheric
pressure alarm Below 6 cm H2O
or as selected
Continuing press
alarm Yes (noncycle)
O2 concentration No
Type of sensor NA
Auto ID NA
Agent conc alarm NA
ECG No
Heart rate NA
ST segment NA
Noninvasive BP No
Invasive BP No
Temperature No
Pulse oximeter No
Other monitors No
Other features No
74 ©2000 ECRI. Duplication of this page by any means for any purpose is prohibited.
Anesthesia Units
DISPLAYS Yes
Number 1
Type Color LCD
Integrated Yes
Interface w/others Yes
DATA INPUT Touchscreen,
keyboard
PRIORITIZED ALARMS Yes
PHYSICAL FEATURES
H x W x D, cm (in) 72 x 65 x 140
(28.3 x 25.6 x 55.1)
Auxiliary outlets 4
BACKUP BATTERY Yes
Type Not specified
Use per charge, hr 1
PURCHASE INFORMATION
Price AU$42,000-48,000
(US$27,325-31,223)
Warranty 1 year
Service contract Not specified
©2000 ECRI. Duplication of this page by any means for any purpose is prohibited. 75