Beruflich Dokumente
Kultur Dokumente
METHOD DEVELOPMENT:
Analytic methods are intended to establish the identity, purity, physical characteristics and potency
of the drugs that we use. Methods are developed to support drug testing against specifications
during manufacturing and quality release operations, as well as during long term stability studies.
Methods may also support safety and characterization studies or evaluations of drug performance.
The three critical components for a HPLC method are:
sample preparation- % organic, pH, shaking/sonication, sample size, sample age,
analysis conditions- %organic, pH, flow rate, temperature, wavelength, and column age,
and
standardization- integration, wavelength, standard concentration, and response factor
correction.
STEPS FOR HPLC METHOD DEVELOPMENT:
1. Information on sample
2. Defining separation goals
3. Special procedure requirement, sample pretreatment, if any
4. Detector selection and setting
5. Separation conditions optimization
6. Checking for problems or special procedure requirements
7. Recovery of purified material
8. Quantitative calibration/ Qualitative method
9. Method validation for release to laboratories
HPLC CONDITIONS:
Buffer selection:
Choice of buffer is typically governed by the desired pH. The typical pH range for reserved-phase
on silica-based packing is pH 2 to 8. It is important that the buffer has a pKa close to the desired
pH since buffer controls pH best at their pKa. A rule is to choose a buffer with pKa value < 2 units
of the desired mobile phase pH.
Buffer concentration:
Generally, a buffer concentration of 10-50 mM is adequate for small molecules. Generally, no
more than 50% organic should be used with a buffer. This will depend on the specific buffer as
well as its concentration. Phosphoric acid and its sodium or potassium salts are the most common
buffer systems for reversed-phase HPLC.
Selection of detector:
Selection of detector depends on the chemical nature of analytes, potential interference, limit of
detection required availability and cost of detector. UV-Visible detector is versatile, dual
wavelength absorbance detector for HPLC. UV-Visible detectors are typical in many laboratories
as they can detect a wide array of compounds. Others detectors used in HPLC instrument include
photodiode array detector, fluorescence detector, conductivity detector, refractive index detector,
electrochemical detector, mass spectrometer detector and evaporative light scattering detector.
Column selection:
The heart of a HPLC system is the column. Changing a column will have the greatest effect on the
resolution of analytes during method development. Generally, modern reverse phase HPLC
columns are made by packing the column housing with spherical silica gel beads which are coated
with the hydrophobic stationary phase. In general, the nature of stationary phase has greatest effect
on capacity factor, selectivity, efficiency and elution. Silica matrices are robust, easily derivatized,
manufactured to consistent sphere size, and does not tend to compress under pressure. C18
columns are the commonly used columns in HPLC method analysis. C8 or Octyl bonded phases
are also used occasionally. Like C18, they are non-polar, but not as hydrophobic. Therefore,
retention times for hydrophobic compounds are typically shorter. Also, they may show somewhat
different selectivity than C18 due to increased base silica exposure unique selectivity results in
proton interaction of the bonded phase with electron deficient functional groups of solute
molecules.
Column dimensions:
This refers to the length and internal diameter of the packing media bed within the column tube.
Short columns (30-50mm) offer short run times, fast equilibration, low back pressure and high
sensitivity. Long columns (250-300mm) provide higher resolving power, but create more
backpressure, lengthen analysis times and use more solvent. Narrow column (2.1mm and smaller)
beds inhibit sample diffusion and produce narrower, taller peaks and a lower limit of detection.
They may require instrument modification to minimize distortion of the chromatography. Wider
columns (10-22mm) offer the ability to load more sample.
Mobile phase composition (or solvent strength) plays an important role in RP-HPLC separation.
Acetonitrile (ACN), methanol (MeOH) and tetrahydrofuran (THF) are commonly used solvents in
RP -HPLC having low UV cut- off of 190, 205 and 212nm respectively. These solvents are
miscible with water. Mixture of acetonitrile and water is the best initial choice for the mobile phase
during method development.
Change in the mobile phase pH can also improve column efficiency because it alters both the
ionization of the analyte and the residual silanols and it also minimizes secondary interactions
between analytes and the silica surface that lead to poor peak shape. To achieve optimum
resolution, it requires change in the pH of mobile phase. Method development can proceed by
investigating parameters of chromatographic separations first at low pH and then at higher pH until
optimum results are achieved.
Column temperature:
Separation of many samples can be enhanced by selecting the right column temperature. Higher
column temperature reduces system backpressure by decreasing mobile phase viscosity, which in
turn allows use of longer columns with higher separation efficiency. However, an overall loss of
resolution between mixture components in many samples occurs by increasing column
temperature. The optimum temperature is dependent upon the nature of the mixture components.
The overall separation can be improved by the simultaneous changes in column temperature and
mobile phase composition.
In order to develop a HPLC method effectively, most of the effort should be spent in method
development and optimization as this will improve the final method performance.
METHOD VALIDATION:
According to ICH guideline Q2 Validation of analytical procedure is the process for proving that
an analytical procedure is suitable for its intended purpose. Results obtained from method
validation study can be used to judge the quality, reliability and consistency of analytical results.
PRE−VALIDATION REQUIREMENTS:
Chemicals, such as reagents and standards, should be available in sufficient quantities, accurately
identified, sufficiently stable and checked for purity. Other materials and consumables, for
example, chromatographic columns, should be qualified to meet the column’s performance
criteria. The validation experiments should also be carried out by an experienced analyst to avoid
errors due to inexperience. Validation on the analytical procedure should be performed with
homogeneous samples, and validation data should be obtained by repeat LOD analysing aliquots
of a homogeneous sample, each of which has been independently prepared according to the
analytical method procedure.
REVALIDATION:
Revalidation is necessary whenever a method is changed and the new parameter is outside the
operating range. The operating parameters need to be specified with ranges clearly defined. In case
of methods for quantitation of impurities, if a new impurity is found that makes the method
deficient in its specificity, it needs modification and revalidation. Changes in equipment or
chemical quality may also have critical effects on method. So any such change needs revalidation.
The various validation parameters include
Linearity
Accuracy
Precision
Ruggedness
Robustness
LOD
LOQ
selectivity or specificity
1. Linearity:
ICH definition: The linearity of an analytical procedure is its ability (within a given range) to
obtain test results that are directly proportional to the concentration (amount) of analyte in the
sample.
It is essential to determine the useful range at which the instrumental response is proportional to
the analyte concentration. Generally, a value of correlation coefficient (r) > 0.998 is considered as
the evidence of an acceptable ft of the data to the regression line. Significance of deviation of
intercept of calibration line from the origin can be evaluated statistically by determining
confidence limits for the intercept, generally at 95% level. Linearity is determined by a series of
three to six injections of five or more standards. Peak areas (or heights) of the calibration standards
are usually plotted on the Y-axis against the nominal standard concentration, and the linearity of
the plotted curve is evaluated through the value of the correlation coefficient (r). Because
deviations from linearity are sometimes difficult to detect, two additional graphical procedures can
be used to evaluate the linearity of the plot. The first one is to plot deviations from regression line
versus concentration or versus logarithm of concentration. For linear ranges, the deviations should
be equally distributed between positive and negative values. Another approach is to divide signal
data by their respective concentrations yielding the relative responses. A graph is plotted with the
relative responses on Y-axis and the corresponding concentrations on X-axis on a log scale. The
obtained line should be horizontal over the full linear range. At higher concentrations, there will
typically be a negative deviation from linearity. Parallel horizontal lines are drawn in the graph
corresponding to, for example, 95 % and 105 % of the horizontal line. The method is linear up to
the point where the plotted relative response line intersects the 95 % line.
2. Accuracy:
Definition: The accuracy of an analytical procedure expresses the closeness of agreement between
the value that is accepted either as a conventional true value or as an accepted reference value
and the value found.
It is a qualitative characteristic that cannot be expressed as a numerical value. It has an inverse
relation to both random and systematic errors, where higher accuracy means lower errors [35].
Accuracy is evaluated by analyzing test drug at different concentration levels. Typically, known
amounts of related substances and the drug substance in placebo are spiked to prepare an accuracy
sample of known concentration of related substance. Samples are prepared in triplicate. ICH
recommends accuracy evaluation using a minimum of nine determinations over a minimum of
three concentration levels covering the range specified. It is determined by comparing the found
concentration with the added concentration. The methods of determining accuracy include analysis
of analysis of known purity (reference material), comparison of results of the proposed analytical
procedure with those of a second well characterized procedure and standard addition method. The
accuracy may also be inferred once precision, linearity and specificity have been established.
3. Precision:
The precision of an analytical procedure expresses the closeness of agreement (degree of scatter)
between a series of measurements obtained from multiple samples of the same homogeneous
sample under prescribed conditions. Precision is usually investigated at three levels: repeatability,
intermediate precision, and reproducibility. For simple formulation it is important that precision
be deter- mined using authentic homogeneous samples. A justification will be required if a
homogeneous sample is not possible and artificially prepared samples or sample solutions are used.
Repeatability (Precision) Repeatability is a measure of the precision under the same operating
conditions over a short interval of time. It is sometimes referred to as intra assay precision. Two
assaying options are allowed by the ICH for investigating repeatability:
A minimum of nine determinations covering the specified range for the procedure (e.g.,
three concentrations/three replicates as in the accuracy experiment), or
A minimum of six determinations at 100% of the test concentration.
The standard deviation, relative standard deviation (coefficient of variation), and confidence
interval should be reported as required by the ICH. The true value of the samples and the variation
of the assay may be between 97.5 and 99.1%.
Intermediate (Precision): Intermediate precision is defined as the variation within the same
laboratory. The extent to which intermediate precision needs to be established depends on the
circumstances under which the procedure is intended to be used. Typical parameters that are
investigated include day-to-day variation, analyst variation, and equipment variation. Depending
on the extent of the study, the use of experimental design is encouraged. Experimental design will
minimize the number of experiments that need to be performed. It is important to note that the
ICH allows exemption from doing intermediate precision when reproducibility is proven. It is
expected that the intermediate precision should show variability that is in the same range or less
than repeatability variation. The ICH recommended the reporting of standard deviation, relative
standard deviation (coefficient of variation), and confidence interval of the data.
5. Specificity:
ICH definition: The quantitation limit of an individual analytical procedure is the lowest amount
of analyte in a sample that can be determined quantitatively with suitable precision and accuracy.
The detection limit of an individual analytical procedure is the lowest amount of analyte in a
sample that can be detected but not necessarily quantitated as an exact value.
Two types of approaches can be used to determine the quantitation limit or detection limit, as
described below.
Signal-to-Noise Approach. Quantitation limit is defined as the concentration of related substance
in the sample that will give a signal-to-noise (S/N) ratio of 10: 1. Detection limit (LOD)
corresponds to the concentration that will give a signal-to-noise ratio of 3: 1. The quantitation limit
of a method is affected by both the detector sensitivity and the accuracy of sample preparation at
such a low concentration. In practice, the quantitation limit should be lower than the corresponding
ICH reporting limit.
To investigate the effect of both factors (i.e., sample preparation and detector sensitivity),
solutions of different concentrations near the ICH reporting limits are prepared by spiking known
amounts of related substances into excipients. Each solution is prepared according to the procedure
and analyzed repeat LOD to determine the S/N ratio. The average S/N ratio from all analyses at
each concentration level is used to calculate the LOQ or LOD. The following equation can be
used to estimate the LOQ at each concentration level. Since different concentration levels give
different LOQs, typically the worst-case LOQ will be reported as the LOQ of the method.
LOQ= 10 * concentration (in%related substance)/ (S/N)
Alternatively, the spike solution can be diluted serially to lower concentrations. The S/N ratio at
each concentration level is determined. The concentration level (in percent related substance) that
gives an S/N value of about 10 will be reported as the LOQ.
Standard Deviation Approach. The following equations can be used to determine quantitation limit
and detection limit by standard deviation of the response at low concentrations:
LOQ= 10 * SD/ S
where SD is the standard deviation of the response near LOQ and S is the slope of the linearity
curve near LOQ.
There are two ways to determine SD:
1. Using experiments similar to those given for the signal-to-noise approach, determine the standard
deviation of the responses by repeat analysis of a solution near the targeted LOQ.
2. Construct a calibration curve near the targeted LOQ:
a. Determine the residual standard deviation of the regression line of calibration, or
b. Determine the standard deviation of the y-intercept.
7. System suitability:
System suitability testing (SST) is an integral part of many analytical procedures. The tests are
based on the concept that the equipment, analytical operations and samples are the integral part of
the system that can be evaluated as such. System suitability test provide the added assurance that
on a specific occasion the method is giving, accurate and precise results. System suitability test
are run every time a method is used either before or during analysis. The results of each system
suitability test are compared with defined acceptance criteria and if they pass, the method is
deemed satisfactory on that occasion. In case of HPLC methods, system suitability tests ensure the
adequacy for performing the intended application on daily basis. The primary SST parameters
considered are resolution (Rs), repeatability (% RSD of peak response and retention time), column
efficiency (N), and tailing factor (Tf). The other SST parameters include retention factor (k) and
separation factor (α).
SST Limits:
Resolution (Rs) >2.0
Repeatability (RSD) <1.0% for five replicates
Plate count (N) >2000
Tailing factor (Tf) ≤2.0
Separation factor (α) >1.0
REFERENCES:
1. Lee Y.C., Analytical Method Validation and Instrument Performance Verification, John
Wiley & Sons Inc., 2004, Chapter 5, “Method Validation for HPLC Analysis of Related
Substances in Pharmaceutical Products”, Pg no. 27-51.
2. Azim S., Mitra M., Bhasin P.,” HPLC Method Development and Validation: A Review”,
International Research Journal of Pharmacy, 2013, ISSN 2230-8407, Pg no. 39-46.
Accessed on 09.10.2019 at 19.30.
3. Paithankar H.V., “HPLC Method Validation for Pharmaceuticals: A Review”,
International Journal of Universal Pharmacy and Bio Sciences, 2013, ISSN 2319-8141, Pg
no. 229-240. Accessed on 90.10.2019 at 19.30.
PURVI NEEMA
2019H1460628P
M. Pharm (Sem 1)