MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 2
The VCSn Module is the host analytical module within the UniCel DxH 800 that
functions to control sample preparation and delivery of prepared sample to the
flow cell for cellular analysis of the WBC differential, reticulocytes and NRBC.
The VCSn Module includes the following submodules: the Air Mix and
Temperature Control (AMTC), the Multi-transducer Module (MTM) and the
following assemblies: the VCSn panel, Flow-Cell Waste Assembly and
Distribution Valve Assembly.
The VCSn Module The VCSn Module prepares the WBC diff, reticulocyte and NRBC samples
Overview for flow cytometric analysis, generates the signals for all VCSn measurements
and handles the waste generated by the VCSn analyses. The integration of the
electronics, temperature control and hardware functionality for NRBC, WBC diff
and reticulocyte processes also resides within the VCSn Module. This includes
the reagent pumps, sample and sheath tanks, pressure regulators, sensors,
solenoid valves, sample distribution valve, flow cell waste chamber, the VCS
waste chamber, and the respective solenoid valves and manifolds.
2
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 3
Sample Flow The whole blood sample used by the VCSn Module is delivered by the Sample
Aspiration Module. The Sample Aspiration probe delivers whole blood samples
to the temperature-controlled NRBC and WBC diff mix and Retic stain
chambers in the AMTC. The whole blood sample is delivered into the mix
chambers below the level of the reagents to ensure the blood sample is
combined with temperature controlled reagents (Figure 1).
Once sample preparation is complete, the prepared samples are directed via
the distribution valve from the mix chambers in the AMTC submodule to the
sample line. The sample line is then pressurized, sending the sample to the flow
cell for comprehensive flow cytometric cellular analysis.
3
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 4
~582 μL WB/Elyse +
210 – 250 μL Stabilyse
Start Acquisition
1:23 dilution
Stabilyse temp at ~24°C
~34 μL WB +
~548 μL Erythrolyse
1:16 dilution
Ar Mix and Elyse temp ~24°C
4
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 5
22 μL WB + – 400 μL Diluent
1:18 dilution
Ar Mix temp ~24°C
Blood in diluent for 0.96 – 1.04 sec
The VCSn Module hosts the VCSn panel, the flow cell waste assembly, the
Submodules and distribution valve, the air mix temperature control module (AMTC) and the
Assemblies Multi-transducer Module (MTM). These submodules and assemblies are
described below.
5
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 6
6
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 7
Stain
Chamber
Retic
Chamber
Diff Chamber
Heat Exchange Block
with Insulation
NRBC
Peltier Heater Chamber
Assembly
7
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 8
– VCSn block Peltier receives input from the heat exchange block sensors
and directs the heating and cooling of the heat exchange block to maintain
a controlled sample preparation temperature.
• The Multi-transducer Module (Figure 10) functional components include:
– The MTM optical system includes a 5 mW diode red laser, beam shaping
optics and associated opto-mechanical components. The benefits of the
new design include:
• Custom-designed laser, proprietary to Beckman Coulter
• Ultra-low-noise laser operation for increased signal-to-noise ratio
• Enhanced laser power stability
• Newly designed opto-mechanical components to precisely position the
focused light within the sensing zone of the flow cell.
• Temperature-controlled hardware and sample analysis
8
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 9
• Thermal control has been implemented to ensure that the MTM maintains a
stable temperature regardless of environmental variances. This ensures
mechanical and optical stability of the MTM, increasing the integrity of the
system alignment. The MTM Peltier controls the MTM hardware temperature at
approximately 24° C (75° F).
9
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 10
Conclusion Since the release of VCS technology in the 1980s, Coulter Corp. and Beckman
Coulter have utilized the most advanced technology for the analysis of the WBC
differential and reticulocytes. The UniCel DxH 800 expands the parameter
capabilities of VCS technology to include NRBC. Advanced hardware and
algorithm technology elevates the comprehensive analysis of the WBC
differential, reticulocytes and NRBC to a new level. The VCSn Module within the
UniCel DxH 800 ensures consistent sample results through uniform sample
preparation. Temperature control of the reagents and the cytochemical reactions
in the newly designed mixing chambers alleviates external environmental
variables for more reportable results. The design includes new system hardware
elements; unique flow cell design, the distribution valve and MTM opto-
mechanicals and electronics, which were engineered to provide the most
consistent, reliable and accurate sample measurements for use by the UniCel
DxH 800 advanced algorithms. The hardware innovations within the VCSn
Module, in conjunction with the advanced algorithm applications, provide
the foundation for the science of Flow Cytometric Morphology unique to the
UniCel DxH 800.
10
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 11
11
MG-12087A Flow Cytometric Morphology Monograph.qxd:Anemia Brochure 10/28/08 1:28 PM Page 12
Australia, Gladesville (61) 2 9844 6000 Canada, Mississauga (1) 905 819 1234 China, Beijing (86) 10 6515 6028
Czech Republic, Prague (420) 272 01 73 32 Eastern Europe, Middle East, North Africa, South West Asia: Switzerland, Nyon (41) 22 365 3707
France, Villepinte (33) 1 49 90 90 00 Germany, Krefeld (49) 2151 33 35 Hong Kong (852) 2814 7431 India, Mumbai (91) 22 3080 5101
Italy, Cassina de’ Pecchi, Milan (39) 02 953921 Japan, Tokyo (81) 3 5530 8500 Korea, Seoul (82) 2 404 2146 Latin America (1) (305) 380 4709
Mexico, Mexico City (001) 52 55 9183 2800 Netherlands, Mijdrecht (31) 297 230630 Puerto Rico (787) 747 3335 Singapore (65) 6339 3633
South Africa/Sub-Saharan Africa, Johannesburg (27) 11 564 3203 Spain, Madrid (34) 91 3836080 Sweden, Bromma (46) 8 564 85 900
Switzerland, Nyon (41) 0800 850 810 Taiwan, Taipei (886) 2 2378 3456 Turkey, Istanbul (90) 216 570 17 17 UK, High Wycombe (44) 01494
441181 USA, Brea, CA (1) 800 352 3433, (1) 714 993 5321
B2008-9236-PO-5K www.beckmancoulter.com © 2008 Beckman Coulter, Inc. DIAG-PRINTED IN U.S.A.