Beruflich Dokumente
Kultur Dokumente
SAMPLE CONDITIONING
Even more serious is the potential for interference with the integrity of the
sample, due to the effects of filtration, condensation, leakage, evaporation, and
so on, As a result, the information on the composition of the process fluid is not
only delayed, but also degraded.
Consequently, the best solution is to eliminate all sampling systems and place
the analyzer directly into the process. Such "in-pipe" designs are becoming more
and more available, Particularly well suited for these designs are the various
radiant energy analyzers and the probe sensors.
Sample systems are rarely duplicated, and thus each system must be
"debugged" as a new entity. In deciding the design of a sampling system, the
following questions should be raised:
Shuvro Chakrabarty
Technical Notes
The piped-in sample system usually includes the hardware for calibrating the
analysis system. It also provides a tap for obtaining samples for laboratory
testing. Sample conditioning systems are usually costly to maintain.
The installation price for the sampling system frequently exceeds the cost of the
analyzer, but its importance overrides this economic consideration because a
second-class analyzer can still furnish usable data if it operates with an efficient
sample system, whereas a poor sample invalidates the entire measurement.
In the first part of this section a number of devices are discussed that are
components of analyzer sampling systems. These devices include gas sampling
probes, gas and liquid filters, bypass filters, liquid homogenizers, liquid grab
sample collectors, chemical reactor sampling systems, and solids samplers. In
the second part of this section single-stream and multistream process sampling
systems are described. The more components there are in a sampling system,
the less reliable it is likely to be. The mean time between failure and maintenance
needs of the overall system will improve as the number of pumps, ejectors,
regulating valves, coolers, heaters, filters, coalescers, dryers, knockout traps,
manifolds, timers, and other components that comprise the system are reduced.
When taking gas samples, the goal is to obtain representative samples with
minimum time lag, using short, small-volume sampling lines. Whenever possible
it is preferred to draw dry and clean samples in order to minimize the need for
filters, dryers, knockout traps, or steam tracing. The sample tap should be
located on the side of the pipeline to minimize liquid or dirt entrainment, and the
sample should be taken from the center of the pipe. If it is necessary to
periodically remove the probe for cleaning or to perform a sampling traverse, it is
desirable that the probe be inserted through a packing gland and block valve.
The volume of the sampling system should be kept to a minimum, while the
velocities through the sample lines should be high to protect against the settling
of entrained liquids or particulates that can cause plugging. Sample line tubing
Shuvro Chakrabarty
Technical Notes
can be small as ½ in. (3 mm) diameter, and sample flow velocities should be
between 5 and 10 fps (1.5 to 3 mls). When the sample is taken from a
combustion zone or other "dirty" processes, a filter is usually provided at the tip
of the sample probe and a high-pressure filter blowback line ts provided for
periodic filter cleaning.
When sampling hot, wet stack gas, a filter capable of withstanding the gas
temperature is installed in the stack at the beginning of the sample line to prevent
solids from entering the gas sample line. After the sample is cooled, a coalescing
filter is used to remove suspended liquids before the sample goes to the
analyzer. The sample flow direction is from inside to outside. Filter housings with
Pyrex glass bowls are often used in this application to permit visual check of the
liquid level in the filter housing. Since there is often a considerable amount of
liquid present in the sample, steps must be taken to drain the housing to ensure
that liquid does not build up and carry downstream to the analyzer.
The liquid coalescing filter should be located as close to the analyzer as possible
to minimize the chance of condensation between the filter and the analyzer.
Additional precautions that can be taken to avoid downstream condensation are
to cool the sample below ambient temperature upstream of the coalescing filter
and to heat the line gently between the filter and the analyzer.
Filters
Most analyzer sampling systems will include a filter with at least one wire mesh
strainer (100 mesh or finer) to remove larger particles that might cause plugging.
Available filter materials include cellulose, which should only be considered in
applications where it does not absorb components of interest. Sintered-metallic
filters can remove particles as fine as 2 microns; cellulose filters can remove
down to 3 microns; and ceramic or porous metallic elements can trap particles of
13 microns or larger. When the solids content is high, two filters can be installed
in parallel, with isolation valving on each. Motorized self-cleaning filters are also
available for such services.
Glass microfiber filter tubes efficiently separate suspended liquids from gases.
The filter tubes capture the fine droplets suspended in the gas and cause the
droplets to run together to form large drops within the depth of the filter tube. The
large droplets are then forced by the gas flow to the downstream surface of the
filter tube, from which the liquid drains by gravity. This process is called
coalescing. The coalesced liquid drains from the tubes at the same rate as liquid
droplets enter the tubes. Therefore the tubes have an unlimited life when
coalescing liquids from relatively clean gases. The filters operate at their initial
retention efficiency even when wet with liquid. The flow direction is inside-to-
Shuvro Chakrabarty
Technical Notes
outside to permit the liquid to drip from the outside of the filter to the housing
drain.
The filter tube grade should be selected for maximum liquid drainage rate, rather
than maximum filtration efficiency rating, because a liquid drainage rate
decreases with increasing filtration efficiency rating. If liquid is carried into the
filter in slugs rather than dispersed as droplets in the gas, a filter that is properly
sized for steady-state conditions can be flooded and permit liquid carryover. If
slugging of liquid is expected, a filter with a relatively large bowl should be
selected to provide adequate liquid holding capacity, and provisions should be
made to drain the liquid automatically from the bowl of the housing as it
accumulates. An automatic float drain can be used if the pressure is in the 10 to
400 psig (28 bars) range. Above 400 psig, which is the upper limit for
commercially available float drains, the possibilities are: constant bleed drain, a
valve with an automatic timed actuator, or an external reservoir with manual
valves. The external reservoir can be constructed of pipe or tubing with sufficient
volume to hold all the liquid that is expected to be collecting during any period of
unattended operation. To drain liquid while the filter is operating at pressure or
vacuum conditions, the reservoir inlet valve is closed and the outlet valve is
opened.
Packed towers can also be used as scrubbers. In these designs the gas sample
bubbles up through a nonreactive packing of beads which are wetted by a liquid
Foaming and liquid carryover are problems in packed towers, as in spargers.
When the impurities in the process gas stream are both solids and liquids, such
as in particulate matter and mist-carryover problems in chlorine plants-the fiber
mist eliminator should be considered. The liquid particles form a film on the fiber
surface, and the drag of the gas moves this film and the dissolved solids
Shuvro Chakrabarty
Technical Notes
Shuvro Chakrabarty
Technical Notes
can be started with 25-micron filter tubes and with inside-out flows at very low
flow rates. If the suspended liquid is lighter than the continuous phase, the
housing should be oriented so that the drain port is up.
Glass microfiber filter tubes readily remove suspended gas bubbles from liquid,
eliminating the need for deaeration tanks, baffles, or other separation devices.
Flow direction through the filter is outside-to-inside, and the separated gas
bubbles rise to the top of the housing and are vented through the drain port. If
slipstream sampling is used, the separated bubbles are swept out of the housing
with the bypassed liquid. Filter tubes rated at 25 microns are a good choice for
gas bubble separation.
For the removal of small amounts of polymer dust in vapor samples there are
melt filters with removable, heated metallic surfaces which melt and collect the
polymer dust from the sample.
Liquid analyzer sample streams usually have a high solids content. In addition,
the analyzers are often located in remote areas of the plant and are infrequently
serviced. Therefore, the sample filter system must have a long life between filter
tube changes even when solids loading is high. The recommendation for this
type of application is a two-stage filter system, a 75 micron prefilter followed by a
Shuvro Chakrabarty
Technical Notes
25 micron final filter. The filters should be oversized as much as possible without
causing excessive lag time. Plastic filter housings are usually a good choice.
Solids Sampling
When solids in bins or silos need to be sampled, one of the various solids
feeders can be considered. When solids are to be sampled while flowing by
gravity or while pneumatically conveyed under pressure, the choice of sampling
devices becomes more limited. The screw-type solids sampler can be used on
these applications but it is limited in the amount of pressure or vacuum it can seal
against, and it does not provide an intermixed, representative sample. Screw
samplers also introduce large transportation lags and can not easily return the
sample into the process.
Sample conditioning begins with the location of a suitable sample take-off point.
To obtain a representative sample, the take-off is usually located at the side of a
process line, especially in the case of liquid samples where there is the possibility
of vapor on the top of a horizontal line and dirt or solids on the bottom of the line.
For sampling vapors, the connection may be located in the side or top of the
process line, but in both cases with due consideration to accessibility for
maintenance.
Shuvro Chakrabarty
Technical Notes
Ideally, the sample at the appropriately selected take-off point will require little or
no conditioning; however, it is good practice to install a sampling probe for most
applications as a precautionary measure to prevent particulates from entering the
sample transport system. Sampling processes that are still reacting chemically or
pyrolysis gases may require reaction quenching, or fractionation, at the sample
take-off. This is done by cooling or backflushing with an inert gas or liquid to keep
the sample take-off clean and reliably active, while drawing off a reproducible
sample for analysis.
Sample Transport
Single-line transport is the most direct approach and is used when the sample
line volume is small in relation to the analyzer sample consumption so that the
transport time lag is reasonably short. It is usually used when the analyzer is
field-mounted close to the sample point and sample exhaust facilities are
available.
Bypass-return fast loop is the most commonly used transport loop for analyzers
mounted away from the sample takeoff point. By circulating a continuous high
velocity across a device, to create a differential pressure, and drawing off a
bypass stream to the analyzer, a fast loop is obtained, which is adjustable with no
waste in the product. In most cases, such devices as pumps, control valves, and
process equipment exist in the process for this purpose. Otherwise it may be
necessary to install an orifice in the process line or a circulating sample pump.
With a circulation sample pump, care must be taken to prevent cavitation by
locating the pump close to the sample take-off.
Transport Lag
After selecting the appropriate sample transport method for each analyzer
system, a calculation of the sample time lag should be made, using conventional
flow equations based on the following:
Shuvro Chakrabarty
Technical Notes
2. Total length of the fast loop from the sample take-off point to the
analyzer location and back to the sample return point. Any
restrictions on this loop should be excluded from available
differential pressure.
Sample Conditioning
The extracted sample begins conditioning at the take-off point, continues through
the transport, and finishes conditioning at the analyzer location prior to entering
the analyzer. All samples require some form of conditioning to make them
suitable for the analyzer and to assure reliable on-stream operation. The
conditioning is done at the appropriate location in the sample system loop in
order to maintain the integrity of the sample.
Vaporizing Samples
Removal of Entrainments
This is a routine matter in liquid and gas sampling systems, which normally starts
with the minimum provisions of locating a proper sample tap and providing a
sample probe at the take-off point. Filtration is normally used for gases and
liquids. It removes both liquid and particulate entrainments from gases and
primarily particulate matter from liquids.
Shuvro Chakrabarty
Technical Notes
For gases requiring further conditioning of heavy loading, cyclone filters can be
used if the sample has adequate velocity. For liquids, coalescers are frequently
used to remove both undesirable gases and liquids by gravity. The removal of
free water from a hydrocarbon stream is usually accomplished by passing the
liquid through a hydrophilic element, causing the water droplets to accumulate on
the element. The hydrocarbon stream is then passed through a hydrophobic
element that rejects the water, removing it from the bottom with a hydrocarbon
bypass stream. Another method of removing moisture from a stream uses a
selective permeation device with a drying medium that creates differential
pressure to drive the water through the permeable material, thus removing it from
the flowing stream.
Some sample streams may contain corrosive gases or water vapor that influence
the accuracy of the measurement or potentially damage the analyzer. Removal of
such undesirable matter or components can be accomplished by passing the
sample through a packed bed of solid chemicals and/or desiccant. Further, a
liquid treating agent may be applied with provisions that gas streams be broken
up into small bubbles to assure proper contact between the liquid and gas
phases. Care should be taken in such systems to avoid alteration of the sample
and to condition the sample to a desired and reproducible form for analysis.
Shuvro Chakrabarty
Technical Notes
4. Tubing sizes are critical, especially where low flows are used because they
limit the amount of increase in the adsorption/desorption phenomenon. A rule of
thumb is to use the smallest possible tubing to achieve maximum flow to
accommodate the sample loop design.
6. Filtration of the sample in ppm analysis can create significant problems unless
the filter is totally inert. Therefore, stainless steel filters using a high flow rate or
dense fluorocarbon inert materials are recommended for such applications.
8. When it is necessary to provide an aid for transporting the sample from its
sample point to the sample system through the analyzer, as is frequently done in
ambient monitoring systems, a sample pump, ejector, or aspirator is necessary.
In such cases, the pump must be a diaphragm. If practical, an ejector to
pressurize the sample or an aspirator to aspirate it through the measuring device
can be used. Both are more desirable than a sample pump.
Multistream Switching
3. The loss of information from more than one analysis in the case
of analyzer failure
Shuvro Chakrabarty
Technical Notes
5. Maintenance requirements
After reviewing the above, one can decide if multistreaming is feasible and
whether it should be manual or automatic. Whether manual or automatic,
multistreaming requires good quality valves for stream switching.
One common and important requirement in all such systems is that a continuous
bypass be provided for each sampling point to avoid dead-ended sample lines.
The sample system should be laid out in such a manner that contamination
between streams is avoided. This is best accomplished by arranging the solenoid
valve in a double-block-double-bleed arrangement, which is rather expensive.
More often a three-way solenoid valve is used for each stream with the venting
port always at low pressure to create a relief in case of a leak. To prevent
contamination, dead volume of the sample system should be considered, as well
as equalization of the pressures upstream of the three-way valves. The problem
is more severe in ppm sampling systems because of the adsorption / desorption
effects, and careful consideration should be given to the design criteria described
above.
The reliability of each analyzer is measured by its ability to check the analyzer
calibration as recommended by the manufacturer. Therefore, each system must
include provisions for testing and calibration. A means of isolating the inlet is
needed in order to allow a calibration sample to be introduced manually or
automatically.
Shuvro Chakrabarty
Technical Notes
Sample Disposal
Sample disposal is a critical area, both from the economic point of view that
would preserve any quantity of sample not used and from the environmental side
that would prevent the emission of most hydrocarbons into the air.
When there is an economic justification for saving the sample, for example,
liquids in boiling point and viscometer analyzers, a sample collection and return
system must be furnished to collect the sample at atmospheric pressure and
pump it back at high pressure into the process. For gases with no sample return
point, the sample can be pressurized back into the process, or as is most
frequently done, the sample can be vented into the flare system. However,
except in rare cases, venting is done directly into the atmosphere. When this is
not possible, extreme caution should be taken to control the backpressure when
venting the sample into a flare or other collection system with varying pressures.
Ambient Considerations
Shuvro Chakrabarty
Technical Notes
Shuvro Chakrabarty