Introduction
The Agilent Technologies LCR meter
makes fast measurements on compo-
nents. It is optimized for applications
that require precision and versatility.
The instrument's performance ranges
from general bench-top impedance
measurements to complex trans-
former, coil and electrolytic capacitor
measurements. The LCR meter offers
fast, reliable, and versatile testing at
a low cost.
Open/short/load:
Correct for test fixture and cable
errors
AC test signal levels: Option 4263B-001 adds transformer Built-in system features:
20m–1Vrms, 5mVrms steps measurement functions: turns-ratio, GPIB and handler interfaces
mutual-inductance and dc-resistance
Measurement time (typical):
Cable length settings: 25 ms at best conditions
0, 1, 2, 4 meters
Contact check time (typical):
5 ms per measurement
2
High-quality results System features for test automation Evaluate transformers and coils
• See five digits of data • Maximize accuracy with error with Option 4263B-001
• Make precise measurements with correction • Measure turns-ratio, mutual
0.1% basic accuracy • Use performance specified with inductance and dc-resistance
• Select from 11 impedance 0, 1, 2, and 4 meter cables • Easily make connections with
parameters • Test device contact failure with 16060A transformer test fixture
• Verify device performance at contact check function • Measure parameter responses with
simulated operating conditions • Automate testing with GPIB variable signal levels
• Monitor actual test signal voltage interface
and current levels • Reduce ground-loops with isolated
handler interface
• Continue testing after ac power
loss with continuous memory
• Perform pass/fail testing with
comparator function
(High/In/Low)
3
Specifications
Measurement accuracy
4
Measurement conditions Other Specifications Other instrument functions
Measurement parameters/ranges Test signal level monitor:
1. Warm-up time: ≥ 15 min. Parameter Range Voltage, current
2. Ambient temperature: 23 ±5 ˚C |Z|, R, X 1 mΩ to 100 MΩ
3. Test signal voltage: 1 Vrms |Y|, G, B 10 nS to 1000S Error Correction: Open, Short, Load
4. Test cable length: 0 meter C 1 pF to 1 F
5. Open and short corrections performed L 10 nH to 100 kH Comparator: HIGH, IN, and LOW for each
6. Measurement time: Medium or Long D 0.0001 to 9.9999 displayed parameter
(Other test condition data is available in the Q 0.1 to 9999.9
operation manual.) u -180˚ to + 180˚ Save/recall: 10 instrument states from
∆ -999.99% to 999.99% non-volatile memory
For |Z|, |Y|, L, C, R, X, G, and B accuracy
(Ae), refer to Table 1. Table 1 equations yield Option 4263B-001: DC resistance 1 mΩ to Front-end Protection:
accuracy based on frequency and DUT 100 MΩ Vmax = √ 8/C @ Vmax ≤ 250 V
characteristic impedance (Zm). Zm is Vmax = √ 2/C @ Vmax ≤ 1000 V
from Figure 1, Conversion Diagram. Mutual inductance 1 µH to 100 H C in Farads
(typical)
D accuracy(De) = ± Ae/100 Handler interface: Negative logic and isolated.
Turns-ratio 0.9 to 200 (typical) Signals are HIGH/IN/LOW, No-Contact,
Q accuracy(Qe) = ± (Qm)2xDe EOM, Index, Alarm, Keylock, Ext. Trigger.
where(Qm x De < 1) 1-/+(QmxDe) Measurement conditions and functions
Test frequency: 100 Hz, 120 Hz, GPIB interface: Instrument control, TALK-
u accuracy (ue) = 0.573 x Ae 1 kHz, 10 kHz, 100 kHz. (Option 4263B-002 only mode for LISTEN-only printers using
adds 20 kHz.) GPIB or Centronics/GPIB converter
Ae = Accuracy of |Z|, |Y|, L, C, R, X, G,
and B AC test signal level: 20 m - 1 Vrms, Physical characteristics
De = D accuracy 5 mVrms steps Power: 90-132 Vac or 198-264 Vac.
Dm = Measured value of D 47-66 Hz. 45 VA typical.
Bias:
Qe = Q accuracy
Internal: + 1.5 and +2.0 Vdc Operating temperature: 0 to 45 °C
Qm = Measured value of Q
External: 0 to + 3.0 Vdc
ue = u phase angle accuracy
Dimensions: 320 (W) x 100 (H) x
Zm = DUT impedance at test frequency in Ranging: Auto and Hold 300 (H) mm
Hertz
Trigger: Internal, Manual, and External Weight: 4.5 kg (typical)
Measurement time:
SHORT MEDIUM LONG
25ms 65ms 500ms
5
Test Fixtures/Accessories for the Agilent 4263B
16060A transformer test fixture 16065C external bias adapter 16089C Kelvin IC clip leads
Allows fast connections to For external dc bias of DUT. Vmax IC package clip. 1 meter length.
transformers ≤ 40 Vdc.
16064B LED display/trigger box 16089A Kelvin clip leads 16034G Test fixture
Displays comparator status. Large clip. 1 meter length. For SMD components.
1.5 meter cable. External trigger.
16089B Kelvin clip leads Component dimensions (L x W):
Medium clip. 1 meter length. 0.6 mm x 0.3 mm to 5.0. mm x 1.6 mm
6
Ordering information 1 Test fixtures and accessories 16065A 200-Vdc external voltage
Agilent 4263B LCR Meter bias fixture
Furnished accessory: power cable 16034E/G/H SMD component test fixture
16065C 40-Vdc external voltage bias
3
Options 16043A/B 3-terminal SMD test fixture adapter