Marshall, Indirect tensile
and Universal tests

MANUALE DI ISTRUZIONI
INSTRUCTION MANUAL
Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

Contents
1. INTRODUCTION 1
1.1 Icons appearing in the manual ......................................................................................... 2
1.2 Manual revision history .................................................................................................... 3

2. DESCRIPTION 4

3. INSTALLATION AND SETUP 5


3.1 Computer specifications .................................................................................................... 5
3.2 Computer settings and connections ................................................................................. 5
3.2.1 Numerical format ............................................................................................................. 5
3.2.2 User Account Control (UAC) settings ............................................................................. 7
3.2.3 Wifi settings ..................................................................................................................... 8
3.2.4 IP Address ...................................................................................................................... 11
3.2.5 Energy saving settings ................................................................................................... 11
3.2.6 Connecting the PC to the testing device......................................................................... 13
3.3 Installing the software ..................................................................................................... 14
3.4 Running the software ...................................................................................................... 18

4. SOFTWARE COMPONENTS 20
4.1 Overview of the software................................................................................................. 20
4.2 CMU Main panel ............................................................................................................. 20
4.2.1 Viewing the panel .......................................................................................................... 21
4.2.2 Info button ...................................................................................................................... 21
4.2.3 Close button ................................................................................................................... 21
4.2.4 Date and time ................................................................................................................. 21
4.2.5 Connection button .......................................................................................................... 21
4.2.6 Option button ................................................................................................................. 21
4.2.7 Test Modules .................................................................................................................. 21
4.2.8 Connection Status .......................................................................................................... 21
4.2.9 System Status ................................................................................................................. 21
4.2.10 Message bar ................................................................................................................... 21
4.3 System options panel ....................................................................................................... 22
4.3.1 Viewing the panel .......................................................................................................... 22
4.3.2 Password panel............................................................................................................... 22
4.3.3 Dictionary button ........................................................................................................... 22
4.3.4 Language selection cell .................................................................................................. 23
4.3.5 Logo for print out ........................................................................................................... 23
4.3.6 Device type control ........................................................................................................ 23
4.3.7 Save button..................................................................................................................... 23
4.3.8 Close button ................................................................................................................... 23
4.4 Selection of the application ............................................................................................. 24
4.5 CBR test ............................................................................................................................ 25
4.5.1 Viewing the panel .......................................................................................................... 25
4.5.2 Results button................................................................................................................. 25

82-SW/CMU Rev.1 EN i 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.5.3 Save test button .............................................................................................................. 25


4.5.4 Print button ..................................................................................................................... 26
4.5.5 Local setting button ........................................................................................................ 26
4.5.6 Convert to Excel button .................................................................................................. 27
4.5.7 New test button............................................................................................................... 28
4.5.8 Close button ................................................................................................................... 28
4.5.9 Open multiple certificates button ................................................................................... 28
4.5.10 Results button ................................................................................................................. 28
4.5.11 Save test button .............................................................................................................. 28
4.5.12 Add test button ............................................................................................................... 28
4.5.13 Cancel button.................................................................................................................. 29
4.5.14 Clear table button ........................................................................................................... 29
4.5.15 Print button ..................................................................................................................... 29
4.6 CBR graph window......................................................................................................... 30
4.6.1 Viewing the window ...................................................................................................... 30
4.6.2 Maximum load and Penetration reading ......................................................................... 30
4.6.3 CBR index ...................................................................................................................... 30
4.6.4 Load setting reference for CBR ...................................................................................... 30
4.6.5 Load/Penetration table .................................................................................................... 32
4.6.6 Graph .............................................................................................................................. 32
4.6.7 Graph tools ..................................................................................................................... 32
4.7 CBR description window ................................................................................................ 33
4.7.1 Viewing the window ...................................................................................................... 33
4.7.2 Test type control ............................................................................................................. 34
4.7.3 Test parameters .............................................................................................................. 34
4.7.4 Sample information ........................................................................................................ 34
4.7.5 Report information ......................................................................................................... 34
4.7.6 Calendar button .............................................................................................................. 34
4.8 MARSHALL test ............................................................................................................ 35
4.8.1 Viewing the panel........................................................................................................... 35
4.8.2 Results button ................................................................................................................. 35
4.8.3 Save test button .............................................................................................................. 35
4.8.4 Print button ..................................................................................................................... 36
4.8.5 Local setting button ........................................................................................................ 36
4.8.6 Convert to Excel button .................................................................................................. 36
4.8.7 New test button............................................................................................................... 36
4.8.8 Close button ................................................................................................................... 36
4.9 MARSHALL Graph window ......................................................................................... 37
4.9.1 Viewing the window ...................................................................................................... 37
4.9.2 Stability, Flow, Rigidity, Corrected stability and Tangent flow reading ........................ 37
4.9.3 Load/Displacement table ................................................................................................ 37
4.9.4 Graph .............................................................................................................................. 37
4.9.5 Graph tools ..................................................................................................................... 38
4.10 MARSHALL description window ................................................................................. 39
4.10.1 Viewing the window ...................................................................................................... 39
4.10.2 Test type control ............................................................................................................. 39
4.10.3 Test parameters .............................................................................................................. 40
4.10.4 Report information ......................................................................................................... 40

14.03.2014 ii 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.10.5 Automatic calculation of the density of the sample ....................................................... 40


4.10.6 Calendar button .............................................................................................................. 40
4.11 INDIRECT TENSILE test .............................................................................................. 41
4.11.1 Viewing the panel .......................................................................................................... 41
4.11.2 Results button................................................................................................................. 41
4.11.3 Save test button .............................................................................................................. 41
4.11.4 Print button..................................................................................................................... 41
4.11.5 Local setting button ........................................................................................................ 42
4.11.6 Convert to Excel button ................................................................................................. 42
4.11.7 New test button .............................................................................................................. 42
4.11.8 Close button ................................................................................................................... 42
4.12 INDIRECT TENSILE Graph window .......................................................................... 43
4.12.1 Viewing the window ...................................................................................................... 43
4.12.2 Load, Deformation and ITS reading .............................................................................. 43
4.12.3 Load/Displacement table................................................................................................ 43
4.12.4 Graph ............................................................................................................................. 43
4.12.5 Graph tools ..................................................................................................................... 44
4.13 INDIRECT TENSILE description window................................................................... 45
4.13.1 Viewing the window ...................................................................................................... 45
4.13.2 Test type control ............................................................................................................ 45
4.13.3 Test parameters .............................................................................................................. 46
4.13.4 Report information ......................................................................................................... 46
4.13.5 Automatic calculation of the density of the sample ....................................................... 46
4.13.6 Calendar button .............................................................................................................. 46
4.14 CUSTOM test ................................................................................................................... 47
4.14.1 Viewing the panel .......................................................................................................... 47
4.14.2 Results button................................................................................................................. 47
4.14.3 Save test button .............................................................................................................. 47
4.14.4 Print button..................................................................................................................... 47
4.14.5 Local setting button ........................................................................................................ 48
4.14.6 Convert to Excel button ................................................................................................. 48
4.14.7 New test button .............................................................................................................. 48
4.14.8 Close button ................................................................................................................... 48
4.15 CUSTOM Graph window ............................................................................................... 49
4.15.1 Viewing the window ...................................................................................................... 49
4.15.2 Load, Displacement reading........................................................................................... 49
4.15.3 Load/Displacement table................................................................................................ 49
4.15.4 Graph Convert to Excel button ...................................................................................... 49
4.15.5 Graph tools ..................................................................................................................... 50
4.16 CUSTOM description window ....................................................................................... 51
4.16.1 Viewing the window ...................................................................................................... 51
4.16.2 Test type control ............................................................................................................ 51
4.16.3 Test parameters .............................................................................................................. 51
4.16.4 Report information ......................................................................................................... 52
4.16.5 Calendar button .............................................................................................................. 52

5. USE OF THE SOFTWARE 53


5.1 Registering the software .................................................................................................. 53

82-SW/CMU Rev.1 EN iii 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.1.1 Registration troubleshooting .......................................................................................... 55


5.2 Managing the language................................................................................................... 56
5.2.1 Changing the language of the software .......................................................................... 56
5.2.2 Editing the text of the software ...................................................................................... 57
5.2.3 Adding a new language .................................................................................................. 58
5.3 Setting the password ....................................................................................................... 59
5.4 Running a test.................................................................................................................. 60
5.4.1 CBR, Marshall, Indirect tensile test or Custom .............................................................. 60
5.4.2 Marshall test ................................................................................................................... 65
5.4.3 Indirect Tensile test ........................................................................................................ 70
5.4.4 Custom Test ................................................................................................................... 75
5.5 Managing test results in CBR test ................................................................................. 76
5.5.1 Deleting a test ................................................................................................................. 76
5.5.2 Reviewing a single test ................................................................................................... 77
5.6 Managing test results in Marshall test .......................................................................... 78
5.6.1 Deleting a test ................................................................................................................. 78
5.6.2 Reviewing a single test ................................................................................................... 79
5.7 Managing test results in Indirect Tensile test ............................................................... 80
5.7.1 Deleting a test ................................................................................................................. 80
5.7.2 Reviewing a single test ................................................................................................... 81
5.8 Managing test results in Custom test ............................................................................ 82
5.8.1 Deleting a test ................................................................................................................. 82
5.8.2 Reviewing a single test ................................................................................................... 83
5.9 Closing the software ........................................................................................................ 84

6. APPENDIX 85
6.1 Calculations used in the software .................................................................................. 85
6.1.1 CBR test module ............................................................................................................ 85
6.1.2 Marshall test module ...................................................................................................... 87
6.1.3 Indirect Tensile test module ........................................................................................... 90
6.2 Example test reports ....................................................................................................... 91
6.2.1 CBR test module ............................................................................................................ 91
6.2.2 Marshall test module ...................................................................................................... 94
6.2.3 Indirect Tensile test module ........................................................................................... 96

This instruction manual is an integral part of the machine and should be read before using the machine and be
safely kept for future reference.
CONTROLS reserves all rights of this manual; no part or whole can be copied without the written permission
of CONTROLS.
The proper use of this machine must be strictly adhered to, any other use must be considered as incorrect.
The manufacturer cannot be held responsible for damage caused by incorrect use of the machine.
The machine must not be tampered with for any reason. In case of tampering, the manufacturer declines any
responsibility of functioning and safety of the machine.
This manual is published by CONTROLS.
CONTROLS reserves the right to update its manuals without notification in order to correct possible typing
errors, mistakes, updating of information and/or updating of programs and/or accessories.
Such changes will be inserted in the latest edition of the current manual.
This manual in English is the original version. Printed in Italy.

14.03.2014 iv 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

1. INTRODUCTION
The program 82-SW/CMU is written to run in MS Windows® for data acquisition and
processing of CBR, Marshall, Indirect tensile and generally of load/displacement tests.

The software is designed to be used specifically with new DIGIMAX TOUCH 76-P60R02,
with 76-B3802 and with 34-V1172, 70-T1082, 70-T1182 and 70-T1192 testing machines.

The testing device is connected to the PC using a high-speed ethernet link, passing data and
commands to/from the software in real-time, allowing the device to be controlled by the PC,
and live mesaured and calculated test data to be displayed on the screen.

This manual contains installation and setup information (Section 3), reference sections
detailing the components of the test modules software (Section 4) and a user guide covering
all the basic functions of the software (Section 5).

Main features of the program are:


• Suitable for running CBR, Marshall, Indirect Tensile and universal
load/displacement tests
• Data are presented numerically and graphically in real-time
• Saving and management of tests through single files
• Single and multiple test result management and printout
• Data export to MS Excel®
• Language selection: English, French, Spanish and Italian plus any other custom language
• Connection to Controls machine with high speed Ethernet connection

WARNING:
Improper use of testing machines can cause injury to personnel and/or damage to
equipment. Refer to the relevant testing machine instruction manuals for safety
information and procedures for correct use of the machines.
Any use of the equipment other than that described in the relevant instruction manuals
will void the warranty terms of the equipment.

82-SW/CMU Rev.1 EN 1 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

1.1 Icons appearing in the manual

This icon indicates a NOTE; please read the items marked by this icon
carefully.

This icon indicates a WARNING message; the items marked by this icon refer
to the safety aspects of the operator and/or of the service engineer.

14.03.2014 2 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

1.2 Manual revision history

Revision / Date Change description


Software version
Rev. 1 First manual release
14th March 2014
SW ver. 1.4.3.1

82-SW/CMU Rev.1 EN 3 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

2. DESCRIPTION
The software comprises two parts:

The CMU is the base software, which communicates with the testing device
and is used for managing all the system settings and options.
The TEST MODULES are test modules that are run from within the CMU
and are used to view and manage current and previous test. The TEST
MODULUS do not function as a stand-alone product.

14.03.2014 4 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

3. INSTALLATION AND SETUP


3.1 Computer specifications
The minimum requirements of the PC are as follows:

• Pentium 4 ® CPU 3GHz


• 1Gb of free hard disk space
• Microsoft Windows ® XP or higher operating system
• RAM memory:
- for Windows XP or Vista: minimum 1GB; recommended 2GB
- for Windows 7 and 8: minimum 2GB; recommended 4GB
• CD-ROM drive
• One free RJ45 network port
• Screen resolution of 1024 x 800 pixels with colour quality set to 32 bit
• MS Excel 2003 or higher (optional)

3.2 Computer settings and connections


Before installing the software, it is recommended that you check the following settings of the
computer (procedures and screen shots may be vary for different versions of Windows):

3.2.1 Numerical format


The SW/CMU uses a point symbol as a decimal separator. Check the setting on the system
PC as follows:

1. From the Windows Start menu, open the Control Panel and select Regional and
Language Options from the list.

82-SW/CMU Rev.1 EN 5 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

2. In the window that appears, click the Customize button.

3. In the next window, make sure that the settings are the same as shown below. Note that
the Decimal symbol is ‘.’ and the Digit grouping symbol is blank.

4. Click Apply if any changes were made before clicking OK to close the window.

5. Click OK to close the previous window and then close the Control Panel.

14.03.2014 6 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

3.2.2 User Account Control (UAC) settings


If the system PC has Windows 7 and 8 operating system*, the user account control settings
must be configured as follows:

NOTE:
It is important that the UAC settings are configured as described below. If the settings
are not correct, it will not be possible to register the software and the test results database
and language customisation will not function.

1. From the Windows Start menu, open the Control Panel and select Action Center.

2. Select Change User Account Control Settings.

82-SW/CMU Rev.1 EN 7 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

3. Set the level to Never notify.

4. Click OK, close Control Panel and restart the PC.


If the UAC settings are changed after the software has been installed, the software must
be re-installed otherwise the changes will not be effective.

3.2.3 Wifi settings


If the system PC has Windows XP operating system* with a wireless internet connection, the
connection must be disabled. The procedure is as follows:

1. From the Windows Start menu, open the Control Panel and select Network
Connections.

14.03.2014 8 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

2. In the window that appears, select the wireless network and right-mouse click. From
the menu, select Disable.

3. Check that the connection status is ‘Disabled’. Close the Network Connections panel
and then close Control Panel.

82-SW/CMU Rev.1 EN 9 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

* To check which operating system the PC has, open Control Panel and select System.
Details of the operating system can be found on the General tab.

14.03.2014 10 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

3.2.4 IP Address
The properties of the LAN card must be set as shown in the picture below. If needed,
instructions for opening this panel can be found at:
‘http://windows.microsoft.com/en-US/windows7/Change-TCP-IP-settings’ (for windows 7)
or
‘http://windows.microsoft.com/en-US/windows-vista/Change-TCP-IP-settings’ (for
Windows Vista)
Tick the ‘Validate settings upon exit’ box, if there is one.

3.2.5 Energy saving settings


The energy saving settings must be switched off as follows:

1. From the Windows Start menu, open the Control Panel and select Power Options.

82-SW/CMU Rev.1 EN 11 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

2. Windows XP operating system - in the Power Options panel, select the Power
Schemes tab and select ‘Never’ for the all the available power off settings. Click OK to
save the changes and close the panel and then close the Control Panel.

3. Windows 7, 8 or Vista operating system – in the Power options window, select the
power plan marked as ‘recommended’ and then click ‘Change plan settings’.

Select ‘Never’ for all the available power off settings and then click ‘Save changes’.
Clos the Control panel.

14.03.2014 12 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

3.2.6 Connecting the PC to the testing device


The system PC is connected to the testing device via a LAN connection. Plug a network
cable into the LAN port of the PC and connect the other end to the LAN port of the unit (see
next figure as example for the connection to the DIGIMAX TOUCH unit).

Fig. 3-1

NOTE:
The test unit must connect directly to the LAN port of the PC. It must not be connected
to the PC via a hub.
If the PC is already connected to a network, the testing ubit must be connected to a
second, separate LAN card installed in the PC. If the PC network connection is wireless
a second LAN card is not necessary, but refer to Section 3.2.3 on Wifi settings.

82-SW/CMU Rev.1 EN 13 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

3.3 Installing the software


1. Close any programs running on the PC.

2. Insert the CD provided into the CD-ROM drive of the PC. After a few seconds the
installation program starts up automatically.
If it does not start automatically, run the file ‘setup.exe’ from the CD in Windows
Explorer.

Fig. 3-2

3. Follow the instructions given on screen:


If necessary, you can return to the previous window at any point during the
installation by clicking Back.
The default destination for the installation files is C:\Program Files\CMU. If you
want to change the location, click Browse and select a new location. Click Next to
continue.

Read the text about the software license and if you accept the terms, select the ‘I
accept the License Agreement’ check box and click Next.

14.03.2014 14 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

Read the text about the National Instruments license and if you accept the terms,
select the ‘I accept the License Agreement’ check box and click Next.

Click Next to start the installation.

82-SW/CMU Rev.1 EN 15 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

The next window shows the progress of the installation.

When the installation is complete, click Finish.

14.03.2014 16 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

The installation is now finished, and a new folder containing all the necessary files
will have been created in the destination directory that you specified at the start of
the process.
If requested, in the final window, click Restart to reboot the PC.

4. After the PC has restarted, the software is ready to use. If you want to create a
shortcut to run the software from an icon on the desktop, follow the steps below.

5. On the desktop screen of the PC, right-mouse click and select New Shortcut from
the menu.

6. Use Browse to select the location of the file ‘RTM.exe’ and click Next.

7. Enter a name for the shortcut and click Finish.

82-SW/CMU Rev.1 EN 17 14.03.2014


Manuale di Istruzione
struzione SW/CMU
SW/CMU Instruction Manual

8. The shortcut will now be available on the desktop.

Fig. 3-3

3.4 Running the software


1. Make sure the control console is connected to the PC as described in Section 3.2.6
and then switch on the testing device as described in the machine instruction manual.
manual

2. Set the testing device to display the MAIN MENU.

Fig. 3-4

3. Double-click
click on the CMU icon on the desktop of the PC, or select CMU CMU
from the Program list in the Windows Start menu.

4. The software will open showing the Main panel of the CMU software. See Section
4.2 for details.

14.03.2014 18 82
82-SW/CMU Rev.1 EN
Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5. When the software has started, it has to be connected to the device. The
screen of the software has an icon on the Main menu screen to show that it
has to be connected to the device and the Connection status of the software
should read on the bottom left of the main menu CONNECTED.

Fig. 3-5

6. When the software is run for the first time, the device type must be specified
as well has the language. These two procedure are described in Section 4.3.

82-SW/CMU Rev.1 EN 19 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4. SOFTWARE COMPONENTS
4.1 Overview of the software

The software comprises the CMU program. The CMU is the base software, which
communicates with the testing device and is used for managing all the system
settings and options.

4.2 CMU Main panel

This panel contains functions for setting and editing the language, changing
password, loading or saving system parameters and running available test module.
An information bar at the bottom of the panel displays the status of the system.
4.2.5 4.2.6 4.2.7 4.2.4 4.2.2 4.2.3

4.2.8 4.2.9 4.2.10

Fig. 4-1

14.03.2014 20 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.2.1 Viewing the panel


The panel is automatically displayed when the CMU software is started. The panel
cannot be accessed when the test module is running.
4.2.2 Info button
Opens a window showing company information and version numbers. To close
the window, click the close button in the top right corner of the window.
4.2.3 Close button
Closes the software.
4.2.4 Date and time
Displays the current date and time. The information is taken from the system
clock of the PC.
4.2.5 Connection button
Displays the connection status of the software 'connected' or 'disconnected' to the
testing device and reported in 4.2.8.
4.2.6 Option button
Opens the system option panel for managing the language and changing
password. See section 4.3.
4.2.7 Test Modules
This area of the CMU contains links for running the test modules that have been
installed with the software. Click on the relevant test module icon to open the test
module required for the test.
4.2.8 Connection Status
Indicates the status of the connection between the PC and the testing device.
4.2.9 System Status
Indicates the status of the testing system:
READY means that the system is ready to start a test;
TEST IN PROGRESS is shown when a test is running.
4.2.10 Message bar
Displays any messages from the software with the date and time that the incident
occurred.

82-SW/CMU Rev.1 EN 21 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.3 System options panel

This panel contains functions for setting and editing the language, changing the
passwords and loading or saving system parameters. See next figure.

4.3.2 4.3.3 4.3.4

4.3.5 4.3.6 4.3.7 4.3.8

Fig. 4-2

4.3.1 Viewing the panel


The panel is opened by clicking the Option button (Item 4.2.6) on the Main
panel. You will have to enter the level 1 password to gain access to the panel.
4.3.2 Password panel
Opens a panel for changing the password.
There is one level of password:
Level 1 is for the operator and must be known in order to access the system
options.
The default level 1 password is 'controls1'.
4.3.3 Dictionary button
Opens a panel for editing the text used in the software. See section 5.2 for more
information and instructions on editing the language.

14.03.2014 22 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.3.4 Language selection cell


For selecting the language to be used in the software.
The software comes with English and Italian languages. A custom language may
also be entered - see Section 5.2.2.
4.3.5 Logo for print out
Save and load the logo of your institution for future certificates with a maximum
dimension of 576 x 96 pixel.
A standard Windows save and open file dialogue box will open for you to find the
required file.
4.3.6 Device type control
For selecting the type of unit that is connected to the PC.
Click on the arrow to the right of the cell to select one of the following options:
• MULTISPEED if you have a multispeed connected
• DIGIMAX if you have a Digimax connected.

MULTISPEED
UNIFRAME

Models
34-V1172 - 34-V1174
70-T1082 – 70-T1084
70-T1182 - 70-T1184
70-T1192 - 70-T1194

DIGIMAX

Model
82-P60R02

All the other models are non compatible with the CMU software.

4.3.7 Save button


Saves any changes made to the settings in the panel.
4.3.8 Close button
Closes the panel.
If you close the panel and have not saved any changes, you will be asked to
confirm that you want to close without saving.

82-SW/CMU Rev.1 EN 23 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.4 Selection of the application

Once the software has been configured, it is possible to select the application
desired. The applications included in the software are:
• CBR test
• Marshall test
• Indirect Tensile test
• Customised test
The test may be selected by clicking on the Main menu (see next figure).

CBR
test

Marshall test

Indirect Tensile
test

Customised test

Fig. 4-3

14.03.2014 24 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.5 CBR test


The main display of the program will appear. This panel contains buttons for
opening results database, saving and creating reports of the current test, saving
test parameters. Also on the panel are tabs for viewing live results (the Graph
window) and entering test information (Description window).

4.5.2 4.5.3 4.5.4 4.5.9 4.5.5 4.5.6 4.5.7 4.5.8

Fig. 4-4

4.5.1 Viewing the panel


The panel is always visible when the test is running.
4.5.2 Results button
Opens the Results panel for managing, reviewing and printing previous tests.
4.5.3 Save test button
Saves the current test.

82-SW/CMU Rev.1 EN 25 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.5.4 Print button


Prints a test report for the current test on the default printer.
The report presents the test information from the Description window and a copy
of the graph on the Graph window. An example report is included in the
Appendix, Section 6.2.1.

4.5.5 Local setting button


Opens a panel for editing test type.
In this part of the software the user can setup: the communication channels
between the test device and the PC; enable test saving in file format text, Word
and Excel; the decimation (the number of a new sequence comprising only every nth
element of the original sequence).

Fig. 4-5

The communication channel shall be setup before every testing, checking the right
connections of the load cell and displacement transducers on the device controller
(Multispeed or Digimax).

The decimation shall be increased if the time-duration of the test is going to be


short (time duration of the order of seconds or few minutes) and needs to record
an high number of data. Otherwise if the test is going to be long (hours or days),
the decimation shall be high (i.e. above 50 or more).

Test Decimation
CBR
If test lasts less than 10 seconds, set 2-5.
Marshall
Otherwise set 10-20
Indirect Tensile Test
As above.
Custom
For tests longer than 15 minutes, set 50

14.03.2014 26 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.5.6 Convert to Excel button


Creates a report for the current test in Excel and automatically saves it as a
spreadsheet.
You will be prompted to choose a name and location for the file before it is
created. The report presents the test information from the Description window
and a copy of the graph on the Graph window.

Fig. 4-6

82-SW/CMU Rev.1 EN 27 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.5.7 New test button


Opens a new test, clearing all the data input in the Description panel.

4.5.8 Close button


Closes the test module.

4.5.9 Open multiple certificates button


Opens a panel containing summary data for the completed test, with a button to
create a comparison chart between tests.

4.5.10 4.5.11 4.5.12 4.5.13 4.5.14 4.5.15

Fig. 4-7

4.5.10 Results button


Opens the Results panel for comparing previous tests.
4.5.11 Save test button
Saves the current comparison to the database.
4.5.12 Add test button
Opens a panel to import a test for comparison.

14.03.2014 28 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.5.13 Cancel button


Cancels the test comparison from the panel.
4.5.14 Clear table button
Clear the two tables in the panel.
4.5.15 Print button
Prints a test report for the current test comparison on the default printer.

Fig. 4-8 comparison between two different tests

In the "Test Comparison" panel the user can compare two or more tests at the same time. The
graphs can represent two different analysis: Load vs. Penetration or CBR index vs. % of load.

82-SW/CMU Rev.1 EN 29 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.6 CBR graph window


This window displays live test data. See next figure.
4.6.1 4.6.2 4.6.3

4.6.4 4.6.5 4.6.6 4.6.7


Fig. 4-9

4.6.1 Viewing the window


To view the window, click the 'Graph' tab.
4.6.2 Maximum load and Penetration reading
Displays the live reading of the load cell transducer (in kilonewtons) and
penetration (millimetres) during a test.
4.6.3 CBR index
Displays the calculated CBR index at different percentage, with the possibility to
change it.
4.6.4 Load setting reference for CBR
Opens a new panel for setting the control parameters for each CBR index. See
next figure.

14.03.2014 30 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

The password is not necessary to access to this menu. The values should only be
changed by technical personnel authorised and trained by Controls.

Fig. 4-10

The values to be used are given in this following table.

2.0 11.50000
2.5 13.20000
EN 13286-47 / 4.0 17.60000
BS 1377: part 4 5.0 20.00000
6.0 22.20000
8.0 26.30000
2.0 11.50000
2.5 13.34835
4.0 17.60000
ASTM D1883
5.0 19.92579
6.0 22.20000
8.0 26.30000
2.0 11.50000
2.5 13.20000
4.0 17.60000
AS 1289.6.1.1
5.0 19.80000
6.0 22.20000
8.0 26.30000
2.0 11.50000
2.5 13.20000
4.0 17.60000
Test Method Q103A
5.0 19.80000
6.0 22.20000
8.0 26.30000
2.0 11.50000
2.5 13.33704
4.0 17.60000
CNR UNI 10009
5.0 20.00557
6.0 22.20000
8.0 26.30000

82-SW/CMU Rev.1 EN 31 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.6.5 Load/Penetration table


Displays a table showing Load and Penetration for each step.

4.6.6 Graph
Displays a live plot of penetration against load.

4.6.7 Graph tools


Three tools are available for manipulating what can be seen on the graph:
Default - for viewing the graph normally with zoom and panning off
Zoom - for zooming in and out on the graph or a selected area of the graph
Panning - drags the contents of the graph around within the axes

The right click of the mouse in the middle of the graph enables a second graph
tool which can give the possibility to manipulate the results directly on the graph
during the test.

These commands are:


− Copy Data
− Description and Tip - the user can add a small description to the graph
− Visible Items - the user can visualize on the graph the legend, X scale, Y
scale,...
− Clear Graph - enable the user to clear all the data on the graph
− Create Annotation on the graph
− Autoscale (X or Y) - autoscaling may be set individually on the x or y axis to
resize the graph screen automatically
− Smooth Updates - using smooth updates avoids the flicker that erasing and
drawing causes
− Optional plane - gives the possibility to add new planes on the graph
− Export - the user can export the data to Excel, Clipboard,...

Fig. 4-11

14.03.2014 32 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.7 CBR description window

This window is for entering information about the test and the sample. See next
figure.

4.7.1 Viewing the window


To view the window, click the 'Report' tab.

4.7.1 4.7.2 4.7.3

4.7.4 4.7.5 4.7.6


Fig. 4-12

82-SW/CMU Rev.1 EN 33 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.7.2 Test type control


For selecting which type of test is being carried out. The standard options offered
with the software are as follow:
• EN 13286-47 / BS 1377: part 4
• ASTM D1883
• AS 1289.6.1.1
• Test method Q103A
• CNR UNI 10009
To change the test type, click the arrow to the right of the cell and select an option
from the drop-down list. Some of the test parameters will change depending on
what test type is selected.
4.7.3 Test parameters
For entering information about the test sample and how the test will be carried
out. Some of the test parameters will change depending on what test type is
selected. The parameters are descriptive test data (test organisation, testing
apparatus), volumetric details of the sample (weight, dimensions) and test
parameters related to the test device (Test speed, pre load, dry mould weight,...).

4.7.4 Sample information


For entering information related to the test, the specimen, operator and test date.
4.7.5 Report information
For entering information for a report, certificate number, client, certificate date
and other references.
4.7.6 Calendar button
Displays a calendar from which the required date can be selected and entered into
the adjacent cell. See next figure.

Fig. 4-13

14.03.2014 34 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.8 MARSHALL test


The main display of the program will appear. This panel contains buttons for
opening results database, saving and creating reports of the current test, saving
test parameters. Also on the panel are tabs for viewing live results (the Graph
window) and entering test information (Description window).

4.8.2 4.8.3 4.8.4 4.8.5 4.8.6 4.8.7 4.8.8

Fig. 4-14

4.8.1 Viewing the panel


The panel is always visible when the test is running.
4.8.2 Results button
Opens the Results panel for managing, reviewing and printing previous tests.
4.8.3 Save test button
Saves the current test.

82-SW/CMU Rev.1 EN 35 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.8.4 Print button


Prints a test report for the current test on the default printer.
The report presents the test information from the Description window and a copy
of the graph on the Graph window. An example report is included in the
Appendix, Section 6.2.2.
4.8.5 Local setting button
Opens a panel for editing test type.
In this part of the software the user can setup: the communication channels
between the test device and the PC; enable test saving in file format text, Word
and Excel; the decimation.
4.8.6 Convert to Excel button
Creates a report for the current test in Excel and automatically saves it as a
workbook.
You will be prompted to choose a name and location for the file before it is
created. The report presents the test information from the Description window
and a copy of the graph on the Graph window.
4.8.7 New test button
Opens a new test, clearing all the data input in the Description panel.
4.8.8 Close button
Closes the test module.

14.03.2014 36 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.9 MARSHALL Graph window


This window displays live test data. See next figure.
4.9.1 4.9.2

4.9.3 4.9.4 4.9.5

Fig. 4-15

4.9.1 Viewing the window


To view the window, click the 'Graph' tab.
4.9.2 Stability, Flow, Rigidity, Corrected stability and Tangent flow reading
Displays the live readings of the load cell transducer (in kilonewtons), flow in
millimetres, rigidity of the specimen in kN/mm, the corrected stability in kN and
tangent flow in millimetres.
4.9.3 Load/Displacement table
Displays a table showing Load and Displacement for each step.
4.9.4 Graph
Displays a live plot of penetration against load.

82-SW/CMU Rev.1 EN 37 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.9.5 Graph tools


Three tools are available for manipulating what can be seen on the graph:
Default - for viewing the graph normally with zoom and panning off
Zoom - for zooming in and out on the graph or a selected area of the graph
Panning - drags the contents of the graph around within the axes

The right click of the mouse in the middle of the graph enables a second graph
tool which can give the possibility to manipulate the results directly on the graph
during the test.

These commands are:


− Copy Data
− Description and Tip - the user can add a small description to the graph
− Visible Items - the user can visualize on the graph the legend, X scale, Y
scale,...
− Clear Graph - enable the user to clear all the data on the graph
− Create Annotation on the graph
− Autoscale (X or Y) - autoscaling may be set individually on the x or y axis to
resize the graph screen automatically
− Smooth Updates - using smooth updates avoids the flicker that erasing and
drawing causes
− Optional plane - gives the possibility to add new planes on the graph
− Export - the user can export the data to Excel, Clipboard,...

Fig. 4-16

14.03.2014 38 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.10 MARSHALL description window


This window is for entering information about the test and the sample. See next
figure.
Many parameters, including the test type, can be customised.

4.10.1 Viewing the window


To view the window, click the 'Report' tab.

4.10.1 4.10.2 4.10.3

4.10.4 4.10.5 4.10.6


Fig. 4-17
4.10.2 Test type control
For selecting which type of test is being carried out. The standard options offered
with the software are as follow:
• ASTM D1559 / ASTM D6927 / AASHTO T245
• ASTM D5581
• EN 12697-34
To change the test type, click the arrow to the right of the cell and select an option
from the drop-down list. Some of the test parameters will change depending on
what test type is selected.

82-SW/CMU Rev.1 EN 39 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.10.3 Test parameters


For entering information about the test sample and how the test will be carried
out. Some of the test parameters will change depending on what test type is
selected. The parameters are volumetric details of the sample (weight,
dimensions) and test parameters related to the test device (Test speed, test date,...).
4.10.4 Report information
For entering information for a report, certificate number, client, certificate date
and other references.
4.10.5 Automatic calculation of the density of the sample
The density of the sample is calculated automatically (with volumetric method) by
pressing this button, but it is also possible to insert a value manually which can be
obtained by hydrostatic weighting or other method.
4.10.6 Calendar button
Displays a calendar from which the required date can be selected and entered into
the adjacent cell. See next figure.

Fig. 4-18

14.03.2014 40 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.11 INDIRECT TENSILE test


The main display of the program will appear. This panel contains buttons for
opening results database, saving and creating reports of the current test, saving
test parameters. Also on the panel are tabs for viewing live results (the Graph
window) and entering test information (Description window).

4.11.2 4.11.3 4.11.4 4.11.5 4.11.6 4.11.7 4.11.8

Fig. 4-19
4.11.1 Viewing the panel
The panel is always visible when the test is running.
4.11.2 Results button
Opens the Results panel for managing, reviewing and printing previous tests.
4.11.3 Save test button
Saves the current test.
4.11.4 Print button
Prints a test report for the current test on the default printer.
The report presents the test information from the Description window and a copy
of the graph on the Graph window. An example report is included in the
Appendix, Section 6.2.3.

82-SW/CMU Rev.1 EN 41 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.11.5 Local setting button


Opens a panel for editing test type.
In this part of the software the user can setup: the communication channels
between the test device and the PC; enable test saving in file format text, Word
and Excel; the decimation.
4.11.6 Convert to Excel button
Creates a report for the current test in Excel and automatically saves it as a
workbook.
You will be prompted to choose a name and location for the file before it is
created. The report presents the test information from the Description window
and a copy of the graph on the Graph window.
4.11.7 New test button
Opens a new test, clearing all the data input in the Description panel.
4.11.8 Close button
Closes the test module.

14.03.2014 42 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.12 INDIRECT TENSILE Graph window


This window displays live test data. See next figure.

4.12.1 4.12.2

4.12.3 4.12.4 4.12.5

Fig. 4-20

4.12.1 Viewing the window


To view the window, click the 'Graph' tab.
4.12.2 Load, Deformation and ITS reading
Displays the live readings of the load cell transducer (in kilonewtons),
deformation in millimetres and the ITS value in GPa.
4.12.3 Load/Displacement table
Displays a table showing Load and Displacement for each step.
4.12.4 Graph
Displays a live plot of penetration against load.

82-SW/CMU Rev.1 EN 43 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.12.5 Graph tools


Three tools are available for manipulating what can be seen on the graph:
Default - for viewing the graph normally with zoom and panning off
Zoom - for zooming in and out on the graph or a selected area of the graph
Panning - drags the contents of the graph around within the axes

The right click of the mouse in the middle of the graph enables a second graph
tool which can give the possibility to manipulate the results directly on the graph
during the test.

These commands are:


− Copy Data
− Description and Tip - the user can add a small description to the graph
− Visible Items - the user can visualize on the graph the legend, X scale, Y
scale,...
− Clear Graph - enable the user to clear all the data on the graph
− Create Annotation on the graph
− Autoscale (X or Y) - autoscaling may be set individually on the x or y axis to
resize the graph screen automatically
− Smooth Updates - using smooth updates avoids the flicker that erasing and
drawing causes
− Optional plane - gives the possibility to add new planes on the graph
− Export - the user can export the data to Excel, Clipboard,...

Fig. 4-21

14.03.2014 44 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.13 INDIRECT TENSILE description window


This window is for entering information about the test and the sample. See next
figure.
Many parameters, including the test type, can be customised.

4.13.1 Viewing the window


To view the window, click the 'Report' tab.

4.13.1 4.13.2 4.13.3

4.13.4 4.13.5 4.13.6

Fig. 4-22
4.13.2 Test type control
For selecting which type of test is being carried out. The standard options offered
with the software are as follow:
• AASHTO T283
• EN 12697-23
To change the test type, click the arrow to the right of the cell and select an option
from the drop-down list. Some of the test parameters will change depending on
what test type is selected.

82-SW/CMU Rev.1 EN 45 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.13.3 Test parameters


For entering information about the test sample and how the test will be carried
out. Some of the test parameters will change depending on what test type is
selected. The parameters are volumetric details of the sample (weight,
dimensions) and test parameters related to the test device (Test speed, test date,...).
4.13.4 Report information
For entering information for a report, certificate number, client, certificate date
and other references.
4.13.5 Automatic calculation of the density of the sample
The density of the sample is calculated automatically (with volumetric method) by
pressing this button, but it is also possible to insert a value manually which can be
obtained by hydrostatic weighting or other method.
4.13.6 Calendar button
Displays a calendar from which the required date can be selected and entered into
the adjacent cell. See next figure.

Fig. 4-23

14.03.2014 46 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.14 CUSTOM test


The main display of the program will appear. This panel contains buttons for
opening results database, saving and creating reports of the current test, saving
test parameters. Also on the panel are tabs for viewing live results (the Graph
window) and entering test information (Description window).

4.14.2 4.14.3 4.14.4 4.14.5 4.14.6 4.14.7 4.14.8

Fig. 4-24
4.14.1 Viewing the panel
The panel is always visible when the test is running.
4.14.2 Results button
Opens the Results panel for managing, reviewing and printing previous tests.
4.14.3 Save test button
Saves the current test.
4.14.4 Print button
Prints a test report for the current test on the default printer.
The report presents the test information from the Description window and a copy
of the graph on the Graph window.

82-SW/CMU Rev.1 EN 47 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.14.5 Local setting button


Opens a panel for editing test type.
In this part of the software the user can setup: the communication channels
between the test device and the PC; enable test saving in file format text, Word
and Excel; the decimation.
4.14.6 Convert to Excel button
Creates a report for the current test in Excel and automatically saves it as a
workbook.
You will be prompted to choose a name and location for the file before it is
created. The report presents the test information from the Description window
and a copy of the graph on the Graph window.
4.14.7 New test button
Opens a new test, clearing all the data input in the Description panel.
4.14.8 Close button
Closes the test module.

14.03.2014 48 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.15 CUSTOM Graph window


This window displays live test data. See next figure.

4.15.1 4.15.2

4.15.3 4.15.4 4.15.5

Fig. 4-25

4.15.1 Viewing the window


To view the window, click the 'Graph' tab.
4.15.2 Load, Displacement reading
Displays the live readings of the load cell transducer (in kilonewtons),
displacement in millimetres.
4.15.3 Load/Displacement table
Displays a table showing Load and Displacement for each step.
4.15.4 Graph Convert to Excel button
Displays a live plot of penetration against load.

82-SW/CMU Rev.1 EN 49 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.15.5 Graph tools


Three tools are available for manipulating what can be seen on the graph:
Default - for viewing the graph normally with zoom and panning off
Zoom - for zooming in and out on the graph or a selected area of the graph
Panning - drags the contents of the graph around within the axes

The right click of the mouse in the middle of the graph enables a second graph
tool which can give the possibility to manipulate the results directly on the graph
during the test.

These commands are:


− Copy Data
− Description and Tip - the user can add a small description to the graph
− Visible Items - the user can visualize on the graph the legend, X scale, Y
scale,...
− Clear Graph - enable the user to clear all the data on the graph
− Create Annotation on the graph
− Autoscale (X or Y) - autoscaling may be set individually on the x or y axis to
resize the graph screen automatically
− Smooth Updates - using smooth updates avoids the flicker that erasing and
drawing causes
− Optional plane - gives the possibility to add new planes on the graph
− Export - the user can export the data to Excel, Clipboard,...

Fig. 4-26

14.03.2014 50 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4.16 CUSTOM description window


This window is for entering information about the test and the sample. See next
figure.
Many parameters, including the test type, can be customised.
4.16.1 Viewing the window
To view the window, click the 'Report' tab.

4.16.1 4.16.2 4.16.3

4.16.4 4.16.5
Fig. 4-27

4.16.2 Test type control


For inserting which type of test is being carried out.
4.16.3 Test parameters
For entering information about the test sample and how the test will be carried
out. Some of the test parameters will change depending on what test type is
selected. The parameters are volumetric details of the sample (weight,
dimensions) and test parameters related to the test device (Test speed, test date,...).

82-SW/CMU Rev.1 EN 51 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

4.16.4 Report information


For entering information for a report, certificate number, client, certificate date
and other references.
4.16.5 Calendar button
Displays a calendar from which the required date can be selected and entered into
the adjacent cell. See next figure.

Fig. 4-28

14.03.2014 52 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5. USE OF THE SOFTWARE


5.1 Registering the software
The software must be registered in order to use it. A trial period of 31 days begins when the
software is installed. After the trial period ends, it will not be possible to use the software
without applying to CONTROLS for a registration code.
All the functions of the software are fully operational during the trial period.

NOTE:
The software uses the system clock of the PC to work out how many days of a trial period
remain. If the PC date or time are changed after the software is installed, it will no longer
be possible to register the software on the PC.

During the trial period, the registration window is displayed every time the software is
started. To use the software without registering, click the Skip button. The number of days
remaining in the trial period are displayed in red at the top of the window.
The software can still be registered even if the trial period has expired.

Fig. 5-1

The procedure to register the software is as follows:

1. Install and run the software according to the instructions given in Section 3 of this
manual.

2. In the registration window that appears when the software starts, you will see a
Registration number.

3. Contact CONTROLS After Sales Service (preferably by emailing service@controls.it)


with the Registration number and ask for an Unlock code for the software 82-
SW/CMU.

82-SW/CMU Rev.1 EN 53 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

NOTE:
You must tell CONTROLS your Registration number and the name of the software you
want to register in order to receive an Unlock code.
You should also include the acknowledgement (ACK) number of your purchase and the
name of your company in any correspondance.

To continue using the software while you wait for the Unlock code, click the Skip button on
the registration window.

4. When you receive the Unlock code, click the Register button on the registration
window.

5. In the next window, type the Unlock code into the cells provided.
The unlock code should have 16 characters: enter 4 characters in each cell; do not
enter any hyphens or spaces. For example, this unlock code: 9526-6452-7042-4126 is
shown entered below.

Fig. 5-2

6. Click the Apply button. A message will tell you if registration was successful.

Fig. 5-3

7. Click the OK button to open the software.

14.03.2014 54 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.1.1 Registration troubleshooting


If you followed the steps in Section 5.1 but the software registration failed, check the
following things before contacting CONTROLS:

1. Did you provide CONTROLS with the correct Registration number?


If you made a mistake in the number you gave to CONTROLS, the Unlock code will be
wrong. Check the number and repeat steps 3 to 6 of Section 5.1.

2. Did you enter the Unlock code correctly?


If you enter the wrong code, make a mistake in the code or enter too few or too many
characters, the code will not work and you will see the error message shown below.
Click the Retry button on the error message to enter the code again, or click the Skip
button to continue using the software on the trial period.

Fig. 5-4

3. Is your Registration number ‘0’?


This means that the User Account Controls have not been set correctly and it is not
possible to register the software on the PC. See Section 3.2.

4. Has the clock or calendar of the PC been changed since the software was installed?
Changing the time or date of the PC will lock the software and it will not be possible to
register the software on the PC.

5. Has someone attempted to bypass the registration procedure?


Attempting to bypass or hack the registration procedure will lock the software and it
will not be possible to register the software on the PC.

82-SW/CMU Rev.1 EN 55 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.2 Managing the language


5.2.1 Changing the language of the software
The software comes with English and Italian languages and a custom option (see Section
5.2.3). To change the language used in the software, follow the steps below:

1. Click on the button on the CMU Main panel.


If you have a test module open you will have to close it to access the Main panel.

2. Enter the level 1 password to access the System Option Panel.

3. From the pull down menu, select the language you require.

Pull down menu

Fig. 5-5

4. Click the Save button to save the change.

5. Click the Close button to close the panel.

6. The new language will be effective immediately.

14.03.2014 56 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.2.2 Editing the text of the software


The text used in the software can be edited as follows:

1. Follow steps 1 and 2 of Section 5.2.1.

2. Click on the button


bu to open the Language panel.

Reference language cell Edit language cell

Page cell

Fig. 5-6

3. Select the language you want to edit from the Edit language cell and, if required, the
language you want to use as a reference
r from the Reference language cell.
cell
Use the arrows to the left of the cell to scroll through the options or click on the cell to
select from a list showing all the options.

4. Select a section of the software from the Page cell and then edit
dit the text in the right-
hand column of the table. Repeat for other sections of software as necessary.
The left-hand
hand column displays the selected reference language.

5. Click the Save button to save the changes.


To close the panel without saving the changes, click the
the Close button and then
click the Confirm button on the message.

6. Close the panel by clicking the Close button.

82-SW/CMU Rev.1 EN 57 14.03.2014


Manuale di Istruzione
struzione SW/CMU
SW/CMU Instruction Manual

5.2.3 Adding a new language


A new language can be added as follows:

1. Follow steps 1 and 2 of Section 5.2.1.

2. Click on the button to open the Language panel.

3. Click the Add language button.


Reference language cell

Page cell

Fig. 5-7
4. Select the language you want to use as a reference from the Reference language cell. cell
Use the arrows to the left of the cell to scroll through the options or click on the cell to
select from a list showing all the options.
5. Select the first section of the software from the Page cell and then edit
dit the text in the
right-hand
hand column of the table. Repeat for all the other sections of software.
The left-hand
hand column displays the selected reference language.

6. Click the Save button to save the changes.


To close the panel without saving the
t changes, click the Close button and then
click the Confirm button on the message.

7. Close the panel by clicking the Close button.

8. The new language will now be available in the pulldown menu.

Fig. 5-8

14.03.2014 58 82
82-SW/CMU Rev.1 EN
Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.3 Setting the password

1. Click on the button on the CMU Main panel.


If you have a test module open you will have to close it to access the Main panel..

2. Enter the level 1 password to access the System Option Panel.

3. Click the Password button.

4. Click the relevant button for the level of password that you want to change, 1, 2 or 3,
and in the window that appears, enter the current password for the level you selected.

5. In the next window, enter the new password in both cells.

Fig. 5-9

6. Click the Confirm button to save the new password.


If you decide not to save the new password, click the Cancel button.

NOTE:
If you lose or forget the new password it cannot be recovered. It is recommended to keep
a list of the passwords in a safe place.

7. Click the Close button on the System options panel to close it.
82-SW/CMU Rev.1 EN 59 14.03.2014
Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.4 Running a test


5.4.1 CBR, Marshall, Indirect tensile test or Custom
This procedure describes how to run a test with the software. Refer also to the instruction
manual of the machine that will be used for the test.

1. Run the software as described in Section 3.4.

2. Run the test module by clicking on the icon on the CMU Main panel.

Fig. 5-10

3. Click the Test Setting button in the Graph window.

Fig. 5-11

14.03.2014 60 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4. In the Test setting panel,


panel select the Load Channel and Displacement Channel of the
testing frame that will be running the test. Check
heck that suitable values are entered for the
Decimation (low values
val for short tests (i.e 10), bigger values for tests that require a
long testing time (i.e 50 or more).
more) If required, change the method to save the file report.
report
See Section 4.5.5 for more information.

Fig. 5-12

5. If you made changes, click


c the Save button to save them
m and close the panel.
panel
To close the panel without saving the changes, click the Close button and then
click the Confirm button on the message.

6. If you want to see the tests saved, click the Open tab button to open. A window
will be opened and in this one you will find the results saved previously.
Then click
lick the Close Button to close the panel.

Fig. 5-13

82-SW/CMU Rev.1 EN 61 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

7. Click the ‘Open’ tab to view the file results.

8. Select the type of test you are performing from the list of standards.

9. Enter the sample information and measurements, test information and client details in
the cells. Make sure that you enter the correct dimensions so that the specimen strength
can be calculated correctly.
If you saved the report data (see step 10) when carrying out previous tests on the
same batch of samples, you can click the Save button.
You can enter data in as many or as few of the fields as you require, depending on what
information you want to see on the test report. The only fields that must have a value
are specimen dimensions and certificate number. If you do not enter a certificate
number, you will be prompted to enter one when the test finishes before the test can be
saved. If you don’t enter specimen dimensions, the test will not start.
Test information

Fig. 5-14

10. If you want to save the values you have entered so that they can be loaded in
quickly another time when you are testing the same batch of samples, click the
Save button

11. Mount the test sample in the machine and set the machine to the correct position for
starting a test (refer to the instruction manual of the machine).

12. Click the ‘Graph’ tab to view the results.

14.03.2014 62 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

13. Then live load, strength and elapsed time readings will be displayed in the Graph
window and strength or load will be plotted against time on the graph, depending on
what you selected for the y-axis units in the Report window. The icon of the Start test
button will change to a stop symbol while the test is running. The system status will
display the message ‘TEST IN PROGRESS’ with a green background.

Fig. 5-15

14. The test will stop automatically when specimen failure is detected*, or if the load limit
or time limit is reached and the system will unload.
*The failure detection parameters are specified in the PID settings (see Item 21).

If you want to stop the test manually before failure, click the Start test button
(which will be displayed as a stop icon during the test).

15. Remove the test sample from the testing frame.

16. Save the test.

17. If you want to create an Excel test report, click the Excel Button. In the
standard Windows ‘Save file’ dialogue box that appears, choose a location and
enter a name for saving the file and click ‘Save’.

NOTE:
The system PC must have MS Excel installed to use the Save to Excel function.

Excel will open and a report containing all the information from the Report window
and a copy of the graph from the Graph window will be generated and automatically
saved using the filename and location that you entered.

82-SW/CMU Rev.1 EN 63 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

18. To perform another test of the same type, repeat the above from step 9.

19. If you want to print a test report, click the Printer button.
A report for each test containing all the information from the Report window and a
copy of the graph from the Graph window will be printed on the PC’s default printer.
A summary report will also be printed containing a summary table of all the test results
and average values for breaking load and strength.

14.03.2014 64 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.4.2 Marshall test


This procedure describes how to run a test using the software. Refer also to the instruction
manual of the testing frame that will be used for the test.

1. Run the software as described in Section 3.4.

2. Run the test module by clicking on the icon on the CMU Main panel.

Fig. 5-16

3. Click the Test Setting button in the Graph window.

Fig. 5-17

82-SW/CMU Rev.1 EN 65 14.03.2014


Manuale di Istruzione
struzione SW/CMU
SW/CMU Instruction Manual

4. In the Test setting panel,


panel select the Load Channel and Displacement Channel of the
testing frame that will be running the
t test. Check that suitable values are entered for the
Decimation (low valuse for short tests (i.e 10),
10) bigger values for tests that require a
long testing time (i.e 50 or more).
more). If required, change the method to save the file report.
See Section 4.5.5 for more information.

Fig. 5-18

5. If you made changes, click


c the Save button to save them
m and close the panel.
panel
To close the panel without saving the changes, click the Close button and then
click the Confirm
irm button on the message.

6. If you want to see the tests saved, click the Open tab button to open. A window
will be opened and in this one you will find the results saved previously.
Then click the Close Button to close the panel.

Fig. 5-19

14.03.2014 66 82
82-SW/CMU Rev.1 EN
Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

7. Click the ‘Open’ tab to view the file results.

8. Select the type of test you are performing from the list of standards.

9. Enter the sample information and measurements, test information and client details in
the cells. Make sure that you enter the correct dimensions so that the specimen strength
can be calculated correctly.
If you saved the report data (see step 10) when carrying out previous tests on the
same batch of samples, you can click the Save button.
You can enter data in as many or as few of the fields as you require, depending on what
information you want to see on the test report. The only fields that must have a value
are specimen dimensions and certificate number. If you do not enter a certificate
number, you will be prompted to enter one when the test finishes before the test can be
saved. If you don’t enter specimen dimensions, the test will not start.

Test information

Fig. 5-20

10. If you want to save the values you have entered so that they can be loaded in
quickly another time when you are testing the same batch of samples, click the
Save button

11. Mount the test sample in the machine and set the machine to the correct position for
starting a test (refer to the instruction manual of the machine).

12. Click the ‘Graph’ tab to view the results.

82-SW/CMU Rev.1 EN 67 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

13. Then live load, strength and elapsed time readings will be displayed in the Graph
window and strength or load will be plotted against time on the graph, depending on
what you selected for the y-axis units in the Report window. The icon of the Start test
button will change to a stop symbol while the test is running. The system status will
display the message ‘TEST IN PROGRESS’ with a green background.

Fig. 5-21

14. The test will stop automatically when specimen failure is detected*, or if the load limit
or time limit is reached and the system will unload.
*The failure detection parameters are specified in the PID settings (see Item 21).

If you want to stop the test manually before failure, click the Start test button
(which will be displayed as a stop icon during the test).

15. Remove the test sample from the testing frame.

16. Save the test.

17. If you want to create an Excel test report, click the Excel Button. In the
standard Windows ‘Save file’ dialogue box that appears, choose a location and
enter a name for saving the file and click ‘Save’.

NOTE:
The system PC must have MS Excel installed to use the Save to Excel function.

Excel will open and a report containing all the information from the Report window
and a copy of the graph from the Graph window will be generated and automatically
saved using the filename and location that you entered.

18. To perform another test of the same type, repeat the above from step 9.

14.03.2014 68 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

19. If you want to print a test report, click the Printer button.
A report for each test containing all the information from the Report window and a
copy of the graph from the Graph window will be printed on the PC’s default printer.
A summary report will also be printed containing a summary table of all the test results
and average values for breaking load and strength.

82-SW/CMU Rev.1 EN 69 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.4.3 Indirect Tensile test


This procedure describes how to run a test using the software. Refer also to the instruction
manual of the testing frame that will be used for the test.

1. Run the software as described in Section 3.4.

2. Run the test module by clicking on the icon on the CMU Main panel.

Fig. 5-22

3. Click the Test Setting button in the Graph window.

Fig. 5-23

14.03.2014 70 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

4. In the Test setting panel,


panel select the Load Channel and Displacement Channel of the
testing frame that
hat will be running the test. Check that suitable values are entered for the
Decimation (low valuse for short tests (i.e 10),
10) bigger values for tests that require a
long testing time (i.e 50 or more).
more). If required, change the method to save the file report.
repor
See Section 4.5.5 for more information.

Fig. 5-24

5. If you made changes, click


c the Save button to save them
m and close the panel.
panel
To close the panel without saving the changes, click the Close button and then
click the Confirm button on the message.

6. If you want to see the tests saved, click the Open tab button to open. A window
will be opened and in this one you will find the results saved previously.
Then click the Close Button to close the panel.

Fig. 5-25

82-SW/CMU Rev.1 EN 71 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

7. Click the ‘Open’ tab to view the file results.

8. Select the type of test you are performing from the list of standards.

9. Enter the sample information and measurements, test information and client details in
the cells. Make sure that you enter the correct dimensions so that the specimen strength
can be calculated correctly.
If you saved the report data (see step 10) when carrying out previous tests on the
same batch of samples, you can click the Save button.
You can enter data in as many or as few of the fields as you require, depending on what
information you want to see on the test report. The only fields that must have a value
are specimen dimensions and certificate number. If you do not enter a certificate
number, you will be prompted to enter one when the test finishes before the test can be
saved. If you don’t enter specimen dimensions, the test will not start.

Test information

Fig. 5-26

10. If you want to save the values you have entered so that they can be loaded in
quickly another time when you are testing the same batch of samples, click the
Save button

11. Mount the test sample in the machine and set the machine to the correct position for
starting a test (refer to the instruction manual of the machine).

12. Click the ‘Graph’ tab to view the results.

14.03.2014 72 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

13. Then live load, strength and elapsed time readings will be displayed in the Graph
window and strength or load will be plotted against time on the graph, depending on
what you selected for the y-axis units in the Report window. The icon of the Start test
button will change to a stop symbol while the test is running. The system status will
display the message ‘TEST IN PROGRESS’ with a green background.

Fig. 5-27

14. The test will stop automatically when specimen failure is detected*, or if the load limit
or time limit is reached and the system will unload.
*The failure detection parameters are specified in the PID settings (see Item 21).

If you want to stop the test manually before failure, click the Start test button
(which will be displayed as a stop icon during the test).

15. Remove the test sample from the testing frame.

16. Save the test.

17. If you want to create an Excel test report, click the Excel Button. In the
standard Windows ‘Save file’ dialogue box that appears, choose a location and
enter a name for saving the file and click ‘Save’.

NOTE:
The system PC must have MS Excel installed to use the Save to Excel function.

Excel will open and a report containing all the information from the Report window
and a copy of the graph from the Graph window will be generated and automatically
saved using the filename and location that you entered.

18. To perform another test of the same type, repeat the above from step 9.

82-SW/CMU Rev.1 EN 73 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

19. If you want to print a test report, click the Printer button.
A report for each test containing all the information from the Report window and a
copy of the graph from the Graph window will be printed on the PC’s default printer.
A summary report will also be printed containing a summary table of all the test results
and average values for breaking load and strength.

14.03.2014 74 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.4.4 Custom Test

1. Follow steps 1 to 19 of Section 5.4.3

Fig. 5-28

82-SW/CMU Rev.1 EN 75 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.5 Managing test results in CBR test


5.5.1 Deleting a test

1. Run the CMU - CBR test module.

2. Click the Open Tab button A window will be opened and in this one you will
find the results saved previously.

Fig. 5-29

3. Find the test file that you want to delete in the window and click anywhere on the row
to highlight it.

4. To delete the test, click the Del button and then Yes

NOTE:
The test will be permanently deleted.

14.03.2014 76 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.5.2 Reviewing a single test

1. Run the CMU CBR test module.

2. Click the Open Tab button A window will be opened and in this one you will
find the results saved previously.

Fig. 5-30

3. Find the test in the window and double-click on it.

4. If you want to create an Excel report of the test, click the Excel button. In the
standard Windows ‘Save file’ dialogue box, choose a location and enter a name for
saving the file and click ‘Save’.
Excel will open and a report for the test will be generated and automatically saved
using the filename and location that you entered. An example report is included in the
Appendix, Section 6.2.

5. If you want to print a test report, click the Print button.


A report for the test will be printed on the PC’s default printer. An example report is
included in the Appendix, Section 6.2.

6. When you’ve finished, click the Close button to close the Review panel.

7. Click the Close button on the Results panel to return the CMU Main
panel.

82-SW/CMU Rev.1 EN 77 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.6 Managing test results in Marshall test


5.6.1 Deleting a test

1. Run the CMU - Marshall test module.

2. Click the Open Tab button A window will be opened and in this one you will
find the results saved previously.

Fig. 5-31

3. Find the test file that you want to delete in the window and click anywhere on the row
to highlight it.

4. To delete the test, click the Del button and then Yes

NOTE:
The test will be permanently deleted.

14.03.2014 78 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.6.2 Reviewing a single test

5. Run the CMU Marshall test module.

6. Click the Open Tab button A window will be opened and in this one you will
find the results saved previously.

Fig. 5-32

7. Find the test in the window and double-click on it.

8. If you want to create an Excel report of the test, click the Excel button. In the
standard Windows ‘Save file’ dialogue box, choose a location and enter a name for
saving the file and click ‘Save’.
Excel will open and a report for the test will be generated and automatically saved
using the filename and location that you entered. An example report is included in the
Appendix, Section 6.2.

9. If you want to print a test report, click the Print button.


A report for the test will be printed on the PC’s default printer. An example report is
included in the Appendix, Section 6.2.

10. When you’ve finished, click the Close button to close the Review panel.

11. Click the Close button on the Results panel to return the CMU Main
panel.

82-SW/CMU Rev.1 EN 79 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.7 Managing test results in Indirect Tensile test

5.7.1 Deleting a test

1. Run the CMU - Indirect Tensile test module.

2. Click the Open Tab button A window will be opened and in this one you will
find the results saved previously.

Fig. 5-33

3. Find the test file that you want to delete in the window and click anywhere on the row
to highlight it.

4. To delete the test, click the Del button and then Yes

NOTE:
The test will be permanently deleted.

14.03.2014 80 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.7.2 Reviewing a single test

5. Run the CMU Indirect Tensile test module.

6. Click the Open Tab button A window will be opened and in this one you will
find the results saved previously.

Fig. 5-34

7. Find the test in the window and double-click on it.

8. If you want to create an Excel report of the test, click the Excel button. In the
standard Windows ‘Save file’ dialogue box, choose a location and enter a name for
saving the file and click ‘Save’.
Excel will open and a report for the test will be generated and automatically saved
using the filename and location that you entered. An example report is included in the
Appendix, Section 6.2.

9. If you want to print a test report, click the Print button.


A report for the test will be printed on the PC’s default printer. An example report is
included in the Appendix, Section 6.2.

10. When you’ve finished, click the Close button to close the Review panel.

11. Click the Close button on the Results panel to return the CMU Main
panel.

82-SW/CMU Rev.1 EN 81 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.8 Managing test results in Custom test


5.8.1 Deleting a test

1. Run the CMU - Custom test module.

2. Click the Open Tab button A window will be opened and in this one you will
find the results saved previously.

Fig. 5-35

3. Find the test file that you want to delete in the window and click anywhere on the row
to highlight it.

4. To delete the test, click the Del button and then Yes

NOTE:
The test will be permanently deleted.

14.03.2014 82 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

5.8.2 Reviewing a single test

5. Run the CMU Custom test module.

6. Click the Open Tab button A window will be opened and in this one you will
find the results saved previously.

Fig. 5-36

7. Find the test in the window and double-click on it.

8. If you want to create an Excel report of the test, click the Excel button. In the
standard Windows ‘Save file’ dialogue box, choose a location and enter a name for
saving the file and click ‘Save’.
Excel will open and a report for the test will be generated and automatically saved
using the filename and location that you entered. An example report is included in the
Appendix, Section 6.2.

9. If you want to print a test report, click the Print button.


A report for the test will be printed on the PC’s default printer. An example report is
included in the Appendix, Section 6.2.

10. When you’ve finished, click the Close button to close the Review panel.

11. Click the Close button on the Results panel to return the CMU Main
panel.

82-SW/CMU Rev.1 EN 83 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

5.9 Closing the software


1. Click the Close button on the Test module panel to exit the test module.

2. Click the Close button on the CMU Main menu to close the software.

14.03.2014 84 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

6. APPENDIX
6.1 Calculations used in the software
6.1.1 CBR test module

a. EN 13286-47 / BS 1377 Part 4


− “Test data”
− Test speed: X.XX mm/min (input, default: 1.27)
Preload: XXX N (input, default 44)
− “Sample data”
− Sample height: XXX mm (input, default =0) (A)
− Dry mould weight: XXXX.XX g (input, default =0) (B)
− Dry mould + sample weight: XXXX.XX g (input, default =0) (C)
− Mould volume: XXXX.XX cc (input, default =0) (F)
− Compacted sample density X.XXX t/m3 (calculation, =(C-B)/F) (E)
− Swelling time: XX.X h (input, default =0) (Z)
− Final swelling deformation: XX.XX mm (input, default =0) (G)
− Swelling: XX.XX% (calcolato, = G/A*100) (H)
− Initial Dry density: XXXX.XX t/m3 (calculation, =100*E/(100+M)) (L)
− Water content: XX.XX% (input, default =0) (M)
− Compactive effort: TEXT (input)
− Age of specimen: TEXT (input)

b. ASTM D1883
“Test data”
− Test speed: X.XX mm/min (input, default: 1.27)
− Preload: XXX N (input, default 44)
“Sample data”
− Sample height: XXX mm (input, default =0) (A)
− Dry mould weight: XXXX.XX g (input, default =0) (B)
− Dry mould + sample weight: XXXX.XX g (input, default =0) (C)
− Dry Sample weight: XXXX.XX g (calculation, =C-B, which can be edited) (D)
− Mould volume: XXXX.XX cc (input, default =0) (N)
− Swelling time: XX.X h (input, default =0) (E)
− Final swelling deformation: XX.XX mm (input, default =0) (F)
− Swelling: XX.XX% (calcolato, = F/A*100) (G)
− Dry density before soaking: XXXX.XX kg/m3 (input, default =0) (H)
− Dry density after soaking: XXXX.XX kg/m3 (input, default =0) (I)
− Water content before compaction: XX.XX% (input, default =0) (L)
− Water content after compaction: XX.XX% (input, default =0) (M)
− Water content in top 1-in after soaking: XX.XX% (input, default =0) (N)
− Water content average after soaking: XX.XX% (input, default =0) (O)
− Compactive effort: TEXT (input)
− Age of specimen: TEXT (input)

82-SW/CMU Rev.1 EN 85 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

c. AS 1289.6.1.1
“Test data”
− Test speed: X.XX mm/min (input, default: 1.00)
− Preload: XXX N (input, default 50)
“Sample data”
− Sample height: XXX mm (input, default =0) (A)
− Dry mould weight: XXXX.XX g (input, default =0) (B)
− Dry mould + sample weight: XXXX.XX g (input, default =0) (C)
− Wet mould + sample weight: XXXX.XX g (input, default =0) (S)
− Dry Sample weight: XXXX.XX g (calculation, =C-B, which can be edited) (D)
− Dry sample weight after swelling: XXXX.XX g (calculation, =(C-B)/(1+(M/100)),
which can be edited) (R)
− Mould volume: XXXX.XX cc (input, default =0) (F)
− Sample volume: XXXX.XX cc (input, default =0) (U)
− Swelling time: XX.X h (input, default =0)
− Final swelling deformation: XX.XX mm (input, default =0) (G)
− Swelling: XX.XX% (calculation, =G/A*100) (H)
− Dry density: XXXX.XX g/cc (calculation, =(1/F)*(C-B)/(1+(M/100))(L)
− Moisture content before compaction : XX.XX% (input, default =0) (M)
− Moisture content after soaking : XX.XX% (calculation, =M+(S-C)/D*100) (T)
− Moisture content first 30mm: XX.XX% (input, default =0)
− MDD (Maximum dry density): XXXX.XX cc (input, default =0) (N)
− LDR (Laboratory density ratio): XX.XX % (calculation, =L/N*100) (P)
− OMC (Optimum moisture content): XX.XX% (input, default =0) (Q)
− LMR (Laboratory moisture ratio) : XX.XX% (calculation, =M/Q*100) (R)
− Sample volume after soaking: XXXX.XX cc (calculation, =U*(100+H)/100) (V)
− Sample density after soaking: XXXX.XX g/cc (calculation, =D/V) (Z)
− Compactive effort: TEXT (input)
− Age of specimen: TEXT (input)

d. Q103A
“Test data”
− Test speed: X.XX mm/min (input, default: 1.00)
− Preload: XXX N (input, default 45)
“Sample data”
− Sample height: XXX mm (input, default =0) (A)
− Dry mould weight: XXXX.XX g (input, default =0) (B)
− Dry mould + sample weight: XXXX.XX g (input, default =0) (C)
− Mould volume: XXXX.XX cc (input, default =0) (F)
− Sample volume: XXXX.XX cc (input, default =0) (U)
− Swelling time: XX.X h (input, default =0) (F)
− Final swelling deformation: XX.XX mm (input, default =0) (G)
− Swelling: XX.XX% (calcolato, = G/A*100) (H)
− CCD (Calculated Dry Density): XXX.XX t/m3 (calculation, =(1/F)*(C-
B)/(1+(M/100))(L)
− IMC (Initial Moisture Content): XX.XXX% (input, default =0) (M)
− APD (Apparent Particle Density): XXX.X t/m3 (input, default =0) (T)
− IMC at zero air voids: XX.XX% (calculation =((1/L)-(1/T))*100)
− MDD (Maximum dry density): XXXX.XX cc (input, default =0) (N)
− OMC (Optimum moisture content): XX.XX% (input, default =0) (Q)

e. CNR UNI 10009


All the parameters are the same according to EN 13287-47 / BS 1377 Part 4

14.03.2014 86 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

6.1.2 Marshall test module

a. ASTM D1559/ ASTM D6927/ AASHTO T245

Calculation
Laboratory molded specimens shall satisfy the thickness requirement of 63.5 6 2.5 mm (2.50
6 0.10 in.). Specimens within the thickness tolerance may be corrected based on specimen
volume or thickness. Stabilities determined on field cores with large variation in volume or
thickness shall also be corrected. However, results with larger corrections should be used
with caution. Correction factors (correlation ratios) are given in Table 1. The correlation
ratio is used in the following manner.

A=BxC
where:
A = corrected stability,
B = measure of stability (load), and
C = correlation ratio from Table 1.

82-SW/CMU Rev.1 EN 87 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

The report shall include the following information:


− Type of sample tested (laboratory mixed sample, plant mixed sample, or pavement
core specimen).
− If available, the nature of bituminous mixture, including aggregate type and grading,
binder grade, and binder content.
− Individual and average specimen bulk specific gravities.
− Height of each test specimen in millimetres (or inches) to the nearest 0.25 mm (0.01
in.).
− Individual and average values of Marshall stability (uncorrected and corrected if
required) to the nearest 50 N (or 10 lbf).
− Individual and average value of Marshall flow in units of 0.25 mm (0.01 in.) or in
units of mm directly, where Flow (0.01 in.) = 4 x Flow (mm), as well as the method
used for determining flow (peak or tangent offset).
− Test temperature to the nearest 0.2°C (0.4°F).

b. ASTM D5581

Calculation
Specimens within the thickness tolerance may be corrected based on specimen volume or
thickness. Stabilities determined on field cores with large variation in volume or thickness
shall also be corrected. However, results with larger corrections should be used with caution.
Correction factors (correlation ratios) are given in Table 1. The correlation ratio is used in
the following manner.

A=BxC
where:
A = corrected stability,
B = measure of stability (load), and
C = correlation ratio from Table 1.

The report shall include the following information:


− Type of sample tested (laboratory sample or pavement core specimen),
− The height of each test specimen in inches (or millimetres),
− Average maximum load in pounds-force (or newtons) of three specimens, corrected
when required,
− Average flow value, in hundredths of an inch (twenty-five hundredths of a
millimetre) of three specimens, and
− Test temperature.

14.03.2014 88 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

c. EN12697-34

Calculation
The maximum load attained represents the stability of the mix only if the specimen height
equals the required 63,5 mm. A corrected stability shall be reported by multiplying the
maximum load with the correction factor calculated from the following equations using the
volume of the specimen as determined in accordance with with EN 12697-6:

c = 5,2 e –0,025 9 h = 5,2 e –3,2 ⋅ 10 (–6) ν

where
c is the correction factor;
h is the height of the specimen (in millimetres);
ν is the volume of the specimen determined in 6.1 (in cubic millimetres).
The stability, S, is reported to the nearest 0,1 kN.

To obtain the flow value F, measure on the graph of force versus deformation the distance
from the intersection of the tangent and the base line, A, to the point where the maximum
load is achieved, M. The flow values Ft and FT are similarly obtained.

Record the flow values, F, Ft and FT, to the nearest 0,1 mm.

Test results shall be considered reliable if both the variation in stability between specimens,
VS, is less than 15 % and the variation in flow between specimens, Vf, is less than 20 %.
Reject the test specimen with the largest variation and obtain the average value of the other
specimens if either:
− the stability value from one or more specimens differs from the average stability by
more than 15 %, or
− the flow value from one or more specimens differs from the average flow by more
than 20 %.

Furthermore, if any value differs from the new average by more than 15 % for stability or 20
% for flow, repeat the test.

The Marshall quotient shall be obtained by the calculation of S/F and reporting the value
obtained to 0,1 kN/mm.

82-SW/CMU Rev.1 EN 89 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

6.1.3 Indirect Tensile test module

a. EN 12697-23

Calculation
ITS [GPa]= ( 2 * Peak load [kN] / π * diameter [mm] * thickness [mm])

b. AASHTO T283

Calculation
ITS [kPa]= ( 2000 * Peak load [N] / π * diameter [mm] * thickness [mm])

14.03.2014 90 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

6.2 Example test reports


6.2.1 CBR test module

An Excel test report:

82-SW/CMU Rev.1 EN 91 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

A printed test report:

14.03.2014 92 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

An Excel report:

82-SW/CMU Rev.1 EN 93 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

6.2.2 Marshall test module

An Excel test report:

14.03.2014 94 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

A printed test report:

82-SW/CMU Rev.1 EN 95 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

6.2.3 Indirect Tensile test module

An Excel test report:

14.03.2014 96 82-SW/CMU Rev.1 EN


Manuale di Istruzioni SW/CMU
SW/CMU Instruction Manual

A printed test report:

82-SW/CMU Rev.1 EN 97 14.03.2014


Manuale di Istruzione SW/CMU
SW/CMU Instruction Manual

Notes:

14.03.2014 98 82-SW/CMU Rev.1 EN