Beruflich Dokumente
Kultur Dokumente
Page 4
#1 Highlight (Shabbir)
Cathode ray oscilloscope (CRO) is test equipment that facilitates display of waveforms and measurement of its parameters
#2 Highlight (Shabbir)
Cathode Ray Oscilloscope is a very versatile tool widely used in electronic measurement and testing
Page 5
#1 Highlight (Shabbir)
The electron gun generates electrons and focuses them into a narrow beam
#2 Highlight (Shabbir)
The deflection system moves the beam horizontally and vertically across the screen.
#3 Highlight (Shabbir)
The screen is coated with a phosphorous material that glows when struck by the electrons
#4 Highlight (Shabbir)
Cathode ray tubes used in oscilloscopes consist of an electron gun, a deflection system, and a fluorescent screen
#5 Highlight (Shabbir)
The electron gun, deflection system and florescent screen are enclosed in an evacuated glass tube
#6 Highlight (Shabbir)
#7 Highlight (Shabbir)
The control grid is a hollow metal tube placed over the cathode
#8 Highlight (Shabbir)
the control grid is maintained at a negative potential relative to the cathode to keep the electrons bunched together
#9 Highlight (Shabbir)
The brightness, or intensity, can be adjusted by varying the voltage on the control grid
The first anode, known as focusing anode, focuses the electrons into a narrow beam
A
fixed positive voltage of several thousand volts is connected to
the second anode, known as accelerating anode, for
accelerating the electrons in the
direction of the screen
The deflection system moves the electron beam horizontally and vertically
The deflection system uses electrostatic fields to change the path of the beam.
Page 7
#1 Highlight (Shabbir)
#2 Highlight (Shabbir)
#3 Highlight (Shabbir)
More cycles can be displayed by decreasing the frequency of the saw tooth signal.
#4 Highlight (Shabbir)
X-amplifier and Y-amplifier can be used to adjust the scales on the X- and Y-axes.
#5 Highlight (Shabbir)
If AC coupled, the DC
component of a signal where the AC variation is superimposed
on an average DC value will be removed and the pure AC signal will
displayed
Page 9
#1 Highlight (Shabbir)
Dual (multi-channel) oscilloscopes display two or more waveforms on the CRT screen for comparison purposes
#2 Highlight (Shabbir)
Dual trace oscilloscopes use one electron gun and one deflecting arrangement to display two traces on the screen by sharing time
#3 Highlight (Shabbir)
In the alternate mode, each input signal is traced in alternate
horizontal sweeps
#4 Highlight (Shabbir)
#5 Highlight (Shabbir)
In the chop mode, the horizontal sweep is divided into a number
of time slots and at the end of one time slot the vertical trace
jumps from one signal to the
other
#6 Highlight (Shabbir)
Chop mode is preferred over the ALT mode for low frequency signals and low sweep rates
Page 11
#1 Highlight (Shabbir)
A dual trace oscilloscope has two channels, which facilitates tracing two signals on the screen simultaneously
#2 Highlight (Shabbir)
#3 Highlight (Shabbir)
in a CRO, For signal input, each channel has got a BNC port
#4 Highlight (Shabbir)
The scope (Oscilloscope) inputs are usually high-impedance with one-mega ohm resistance in parallel with 20pf capacitance
#5 Highlight (Shabbir)
ac coupling, through a capacitor, would block the dc components and trace the pure ac component with a time constant of about 0.1 second
#7 Highlight (Shabbir)
In GND position, the signal is disconnected from the scope, and the input is grounded
#8 Highlight (Shabbir)
#9 Highlight (Shabbir)
For voltage measurements, the variable gain knob must be at the calibrated position (CAL).
There are two ways to see both inputs simultaneously: alternate (ALT) mode and chop (CHOP) mode
In alternate mode, alternate inputs are displayed on successive sweeps of the trace
In chop mode, the trace jumps back and forth rapidly between the two signals
signals.
#15 Highlight (Shabbir)
in vertical controls, The POSITION selectors allow shifting of the image up and down
Page 13
#1 Highlight (Shabbir)
#2 Highlight (Shabbir)
The XY position allows a special feature called generating Lissajous figures, which is useful in phase comparison of two signals
#3 Highlight (Shabbir)
in horizonal controls, The POSITION selector allows shifting of the image left or right
#4 Highlight (Shabbir)
#5 Highlight (Shabbir)
#6 Highlight (Shabbir)
NORMAL mode produces a sweep only when the applied signal
crossed through the trigger point set by LEVEL, moving in the
direction set by SLOPE
#7 Highlight (Shabbir)
#8 Highlight (Shabbir)
displayed
#9 Highlight (Shabbir)
In AUTO position, the display will ‘free run’ in the absence of a signal
The external trigger (EXT) inputs are used in situations where synchronization with some external test signal or a clock signal is necessary
Page 15
#1 Highlight (Shabbir)
#2 Highlight (Shabbir)
HF REJ position puts a low-pass filter in front of the
trigger circuit preventing false triggering due to spikes
#3 Highlight (Shabbir)
#4 Highlight (Shabbir)
#5 Highlight (Shabbir)
#6 Highlight (Shabbir)
High impedance probes are used to minimize the effects on the circuit caused by the attachment of the CRO
#7 Highlight (Shabbir)
#8 Highlight (Shabbir)
When the probe capacitance is properly adjusted, a pure square wave without overshoot will be displayed
#9 Highlight (Shabbir)
Some probes feature a convenient choice of 1X or 10X attenuation, switchable at the probe tip