AC10CFIIP.

C18158 / 002
Cone number AC10CFIIP.C18158 Client Geomil Asia Sdn Bhd
Kind of cone Compression 124, Jalan Kapar 27/89
Calibration date 25-Jan-2019 Section 27, Taman Alam Megah
40400 Shah Alam
Malaysia
Channel 1 Cone resistance (qc ) Channel 2 Local sleeve friction (fs ) Channel 3 Pore pressure (u)
qc = Q c / A c fs = F s / A s
Range 0 ... 100 kN Range 0 ... 15 kN Range 0 ... 20 bar
Ac 1000 mm² As 15000 mm²
Zero load reading 184 mV Zero load reading 189 mV Zero load reading 174 mV
a-factor 0.8 b-factor 0
Offset 80 mm

Qc Load Eqv. qc Output Fs Load Eqv. fs Output Pressure Eqv. u Output


(kN) (MPa) (mV) (kN) (MPa) (mV) (bar) (MPa) (mV)
0 0 0 0.0 0.0 0 0 0.0 0
10 10 850 1.5 0.1 794 2 0.2 808
20 20 1695 3.0 0.2 1634 4 0.4 1626
30 30 2542 4.5 0.3 2478 6 0.6 2444
40 40 3386 6.0 0.4 3309 8 0.8 3261
50 50 4232 7.5 0.5 4160 10 1.0 4078
60 60 5078 9.0 0.6 4976 12 1.2 4895
70 70 5922 10.5 0.7 5810 14 1.4 5710
80 80 6764 12.0 0.8 6619 16 1.6 6525
90 90 7603 13.5 0.9 7425 18 1.8 7340
100 100 8443 15.0 1.0 8275 20 2.0 8153
90 90 7605 13.5 0.9 7483
80 80 6764 12.0 0.8 6667
70 70 5923 10.5 0.7 5840
60 60 5083 9.0 0.6 5009
50 50 4238 7.5 0.5 4213
40 40 3388 6.0 0.4 3356
30 30 2545 4.5 0.3 2512
20 20 1696 3.0 0.2 1683
10 10 846 1.5 0.1 854
0 0 -1 0.0 0.0 -4

Zero load error 0.01 % Zero load error 0.05 % Zero load error 0.04 %
Max. linearity 0.20 % Max. linearity 0.91 % Max. linearity 0.09 %
Max. hysteresis 0.07 % Max. hysteresis 0.73 %

Page 1 of 2
AC10CFIIP.C18158 / 002

Channel 4 Inclination X Channel 5 Inclination Y Channel 6 None

Range -20 ... 20 ° Range -20 ... 20 °

Angle (°) Output Angle (°) Output


(mV) (mV)
-20 2509 -20 2476
-15 2580 -15 2546
-10 2654 -10 2616
-5 2729 -5 2687
0 2804 0 2760
5 2882 5 2836
10 2956 10 2908
15 3029 15 2979
20 3102 20 3049

Calibration instrument(s) Certificate number(s) Date(s)


GCU1000/1-170214-011/1 2251724.00501.1 12-Nov-2018

Remark
We declare that the electrical cone with serial number AC10CFIIP.C18158 has been calibrated and that the specifications are according to
the ISO 22476-1:2012 (Geotechnical investigation and testing – Field testing - Part 1: Electrical cone and piezocone penetration test).
The calibrations are traceable to national and international standards.
Date 25-Jan-2019 Date 25-Jan-2019
Calibrated by Tom van Arnhem Approved by Marijn Kints

Signature Signature

Page 2 of 2