Sie sind auf Seite 1von 6

Millipore Steam

Sterilization & Integrity


Testing Procedures

Steaming-in-place and The intent of this technical brief is to provide a Standard Operating Procedure
integrity testing of (SOP) for the steam sterilization and integrity testing of a filter assembly typically
a sterilizing-grade filter used in aseptic processing (see figure 1). Because steam sterilization is the most
common source of damaging a filter, it is recommended to test the integrity of
assembly
sterilizing-grade filters after sterilization, before the filtration process.1,2,3
Therefore, the SOP includes both the Steaming-In-Place (SIP) and the post-SIP,
pre-use integrity testing of the vent and product filters.

The operations described in this SOP 3. SIP of the hydrophilic sterilizing-


B R I E F

should be performed in the following grade filter assembly, like


sequence: Millipore’s Durapore® CVGL
10 inch, LAGL 4 inch or MCGL
1. SIP of the hydrophobic filter
devices, used for the sterile filtration
assembly, like Millipore’s Aervent ®
of the product (product filter).
CTGR 5 inch or 10 inch or LAGR
T E C H N I C A L

4 inch devices, used for the venting 4. Post-SIP, pre-use integrity testing
of a sterile holding tank (vent filter). of the product filter.
2. Post-SIP, pre-use integrity testing The recommended post-SIP, pre-use
of the vent filter. filter integrity test for hydrophobic vent

Sterile Filter Assembly: Product Filter,


Sterile Tank and Vent Filter
Integrity Tester

Steam P
Safety
valve
Integrity Transfer P
Tester line Air/N2
P
WFI
Air/N2 T2

WFI/Product
P

Steam
T3 T1

Â
filters is the HydroCorr™ test. Unlike Specifications
classical testing procedures like
diffusion and bubble point for these SIP Parameters
filters, which require alcohol/water Minimum required steam supply pressure: >1.2 barg
mixtures as a wetting agent, the
HydroCorr™ test requires only clean Minimum required compressed air supply: > 1.5 barg (300 mbar
water. As this test is performed at the above steam pressure)
upstream side of the vent filter and no Minimum temperature in the coldest points: > 121.1 °C
downstream manipulation is required, it
is perfectly suited to assess the integrity Maximum reverse differential pressure across
of the filter and to demonstrate that the the filter during SIP: < 100 mbar
SIP procedure has been adequately Maximum forward differential pressure across
conducted, without damaging the the filter during cooling: < 350 mbar
vent filter.
The recommended post-SIP, pre- Minimum sterilization time: 30 min. at 121.1 °C
use filter integrity test for hydrophilic Cooling time: 30 min. (approximately)
product filters is the Enhanced Bubble
Point test.
Integrity testing involves wetting differential pressure is 100 mbar minimum bubble point values for non-
of the filter with the standard wetting throughout the entire SIP cycle. specified wetting fluids (see Millipore’s
medium and is a possible source of Millipore’s Durapore® filters are Application Note No. AN1505EN00).
breaking sterility of the sterilized sterilizing grade filters that are The recommended pressure
system. From the product filter to constructed with a PVDF membrane. for filling the housing prior to the
the vent filter on the sterile tank is These filters have been qualified to HydroCorr test is 1 barg. Table 2
a sealed, closed and sterile system. withstand 5 to 30 SIP cycles at 135 °C shows the HydroCorr integrity test
Water used to wet the filter before for 30 minutes in the forward direction criteria for Aervent type of filters.
integrity testing cannot be drained (see the Validation Guide of the
downstream of the closed filter system, respective filters). The maximum differ-
and will remain in the sterile product ential pressure allowed during SIP in
tank. Using the product to be filtered the forward direction is 350 mbar.
as the wetting agent facilitates pre-
Filter type Wetting volume (L)
use integrity testing and avoids the Integrity Testing Parameters
downstream evacuation of the testing The minimum bubble point speci- Millidisk® 20 5
medium since the product can be fication for hydrophilic sterilizing-
Millidisk 40 10
directly routed to the holding tank. grade (0.22 µm) Durapore filters
Product bubble point test is perfectly is 3450 mbar, after wetting with Optiseal® 4 inch 10
suited to assess the integrity of the pure water for 5 min. at 200 mbar Durapore 5 inch 18
filter as well as to demonstrate that differential pressure.
Durapore 10 inch 35
the SIP procedure has been adequately The minimum product bubble point
conducted, without damaging the is determined by the bubble point Table 1: Recommended volume for
product filter. ratio laboratory scale study. The wetting Durapore filters prior to Bubble
bubble point ratio (BPR) approach is a Point testing.
Filter Characteristics proven method used for determining
Millipore’s Aervent filters are sterilizing-
grade vent filters that are constructed Type of Aervent Filter Catalogue number Specification
with a PTFE membrane. These filters
have been qualified to withstand at 4 inch Optiseal LAGR04TP6 < 0.20 mL/min. @ 2620 mbar
least 40 SIP cycles at 135 °C for 5 inch Cartridge CTGR75S01 < 0.38 mL/min. @ 2620 mbar
30 minutes in the forward direction,
as well as 40 cycles in the reverse 10 inch Cartridge CTGR_1TP1 < 0.75 mL/min. @ 2620 mbar
direction (see the Validation Guide 20 inch Cartridge CTGR_2TP1 < 1.50 mL/min. @ 2620 mbar
of the respective filters).
The maximum differential pressure 30 inch Cartridge CTGR_3TP1 < 2.25 mL/min. @ 2620 mbar
allowed during SIP in the forward
Table 2: HydroCorr integrity test criteria for Aervent filters.
direction is 350 mbar. For reverse
SIP, the recommended maximum

2
Steam-In-Place Procedure Standard Operating Procedure 6. When the temperature throughout
for a Sterile Tank Equipped 1. Check that the steam supply and the system reaches over 121.1 °C,
With a Vent Filter compressed gas pressures are set as measured by the temperature
The manual operations that are up at the required values. gauges T1, T2 and T3, the timer
described in this SOP should be is started. Sterilization time should
2. Respectively open V2, V3, V4, V5,
performed respecting the given be at least 30 minutes or longer if
and V1 to introduce steam to the
sequence. For automatic SIP validation has decreed so.
system and to purge air from the
procedures refer to Millipore’s During the sterilization phase both
system.
Technical Brief No. TB011EN00. pressure and temperature should
3. Partially close bleed valves V2 and be recorded regularly.
Prior to commencing the procedure,
V5 to build tank pressure to at least
the following is assumed: 7. When the required sterilization
0.5 bar and wait for the temper-
• Tank and adjacent piping is clean time has been achieved, close the
ature gauges T1 and T3 to indicate
(e.g. by means of CIP) and empty; steam supply valve V1 and slowly
more than 100 °C.
open V9 to introduce sterile
• Filter housing is installed and the 4. Then, slowly open V6 to introduce compressed gas into the system.
correct vent filter is put in place; steam to the vent filter. Crack open
CAUTION: Make sure that the
• All connections are checked for bleed valves V7 and V8 to establish
system remains under positive
proper fitting; a steady flow of steam and allow
pressure (as indicated by pressure
• System has been checked for leak- for condensate drainage and air
gauges P1, P2 and P3) and control
tightness by means of a pressure removal from the filter housing.
that the delta-P over the filter does
hold test; Note: It is of utmost importance to
not exceed 350 mbar.
control the difference between
• All valves are closed and silicone pressure gauges P1 and P2 and 8. Allow for steam purge from all
tubing are attached to bleed valves keep the delta-P over the filter to a bleed valves and close valves V7
and directed to a condensate drain; maximum of 100 mbar. For reverse and V8 to increase the flow of
• Use caution to avoid contact direction SIP, use an Optiseal filter sterile gas through the system.
with steam or hot stainless steel or a code 7* filter. Do not use a Maintain the gas flow to cool
surfaces; code 0** filter. down the system until the temper-
ature gauges T1, T2 and T3
• Wear protective glasses at all times 5. Ensure all air and condensate are
indicate approximately 40 °C.
and heat resistant protective gloves effectively removed by keeping V2,
when necessary. V5, V7 and V8 cracked open so 9. Respectively close valves V5,
that a 15 cm wisp of steam and V2 and V4, and keep V6 and
a continuous drip of water can be V9 open to maintain a positive
seen exiting. pressure into the sterile system
while it is not in use.
* 2–226 O-ring locking outlet with spear assembly ** 2–222 O-ring outlet

Recommended SIP Process for a Sterile Tank Equipped with a Vent Filter
Initial Setup Integrity Tester SIP Cycle Cooling & Drying
V12 V12
V12
Steam P2 Steam P2
Steam P2 V7 V7
V7 Safety Safety
Safety
valve valve
valve
P1 P1
P1 Transfer V3 Air V3 Air
V3 Air N2 Transfer
Transfer V6 V9 V6 V9 N2
V6 V9 N2 line line
line WFI WFI
WFI V8 V10 V8 V10
V8 V10
T2 T2
T2

P3 P3
P3 V11 V4 V11 V4
V11 V4
V5 V2 V5 V2
V5 V2 T1 T1
T1 T3 T3
T3

= Open Valve = Closed Valve

3
Post SIP, Pre-Use Vent Filter Integrity
Test Procedure Recommended Post SIP Pre-use Filter Integrity Test
1. Close the compressed gas valve for Hydrophobic Vent Filters
V9, keep V6 open, and open V7
to vent the system. Wait for the HydroCorr Testing Draining
Integrity Tester
pressure as measured by P1 and V12
V12 Steam P2
P2 to drop to atmospheric pressure. Steam P2 V7
V7
Safety
Safety valve V1
2. Fill the pressure vessel with clean valve V1
P1
pure water and attach the inlet of P1 V3 Air
V3 Air V9 N2
V6
the vessel to a compressed gas V6 V9 N2
Transfer
supply at 1 bar. Attach the outlet Transfer line V8 V10 WFI
line V8 V10 WFI T2
tubing of the pressure vessel to V10. T2

P3
3. Slowly open V10 to have water P3 V11 V4
V11 V4
entering the filter housing. Ensure V5 V2
V5 V2 T1
that the filling pressure does not T3 T1
T3

exceed 1 bar and that air cannot


enter the housing (e.g. empty = Open Valve = Closed Valve
pressure vessel). Continue filling
until water is seen exiting the hose Steam-In-Place Procedure Standard Operating Procedure
attached to V7. for Product Filter 1. Check that the steam supply and
Note: Should the filter housing The manual operations that are compressed gas pressures are set
be installed on top of a rather described in this SOP should be up at the required values.
tall tank, it may prove useful to performed respecting the given
increase the pressure to adjust for 2. Open MV1 and MV2 and purge
sequence. For automatic SIP
gravity influence while the filling the steam line until complete
procedures refer to Millipore’s
operation commences. absence of condensate.
Technical Brief No. TB011EN00.
4. Close V10 and bleed air from Prior to commencing the procedure, 3. Fully open MV4 and MV5 to allow
the pressure vessel by slowly the following is assumed: for subsequent air and condensate
opening the pressure relief valve evacuation.
• Filter housing is installed and the
on top of the vessel until atmos- correct product filter is put in place; 4. Slowly open MV3 to progressively
pheric conditions are reached. introduce steam and heat up the filter.
• Product filter is dry;
5. Close V7, open V12 and attach an 5. Partially close bleed valves MV2,
• All valves are closed and silicone
automatic filter integrity tester to V12. MV4 and MV5 so that a wisp of
tubing are attached to bleed valves
6. Double check that V6 is open and steam and a continuous drip of
and directed to a condensate drain;
that V7, V8, V9 and V10 are all fully water can be seen exiting.
• Use caution to avoid contact with
closed and run the HydroCorr test. 6. Respectively open V11 and crack
steam or hot stainless steel surfaces;
7. When the test is finished and a open bleed valve V5 to establish
• Wear protective glasses at all times a steady flow of steam and allow
positive result (i.e. pass) is obtained, and heat resistant protective gloves
close V6 and V12 and detach the for condensate drainage and air
when necessary. removal from the filter housing.
filter integrity tester.
8. Open V7 and V8 to drain water Note: It is of utmost importance
to control the difference between
from the housing. The draining can Recommended SIP pressure gauges P3 and P4 and
be facilitated by carefully opening
V9 and applying pressure until the
for the Product Filter keep the delta-P over the filter to
system is empty. P4 MV5 a maximum of 350 mbar.
MV6
Air/N2 MV8 7. Ensure all air and condensate are
9. Fully open V9 to allow for drying of MV7
the filter over a period of 30 min.
WFI/
P3
effectively removed by keeping
Product MV1 MV3 V11 V4
To tank MV2, MV4, MV5, and V5 cracked
10.Close V7 and V8, open V6 Steam
MV2 MV4 V5 open so that a 15 cm wisp of
and keep V9 open to build and T3 steam and a continuous drip of
maintain a positive pressure into water can be seen exiting.
the system while it is not in use.

4
Recommended SIP and Cooling Process for Hydrophilic Product Filter
P2
SIP Cycle P2 MV5 Cooling Cycle MV5
MV6 MV6
MV8 MV8
Air/N2
MV7 MV7
WFI/ WFI/
Product P3 Product P3
MV1 MV3 V11 V4 MV1 MV3 V11 V4
Steam To tank Steam To tank
MV2 MV4 MV2 MV4
VS VS

T3 T3

= Open Valve = Closed Valve

Recommended Post SIP Pre-use Filter Wetting and Integrity Test


for Hydrophilic Product Filter

Wetting with Product P2 MV5 Bubble Point Test Integrity tester P2 MV5
MV6 MV6
Air/N2 MV8 Air/N2 MV8

MV7 MV7
WFI/ WFI/
Product P3 Product P3
MV1 MV3 V11 V4 MV1 MV3 V11 V4
Steam To tank Steam To tank
MV2 MV4 MV2 MV4
VS VS
T3 T3

= Open Valve = Closed Valve

8. When the temperature downstream 10. Allow for steam purge from all 2. Maintain MV3 open, close the
of the product filter, as measured bleed valves and close valves MV2 compressed gas supply valve MV6,
by the temperature gauge T3, and MV4 to increase the flow of and open MV5 to vent the system.
reaches over 121.1 °C, the timer gas through the system. Maintain Wait for the pressure as measured by
is started. Sterilization time should the gas flow to cool down the P4 to drop to atmospheric pressure.
be at least 30 minutes or longer, system until the temperature gauge 3. If possible, set the inlet product
as established during validation. T3 indicates approximately 30 °C. pressure at 2.8 bar. Gradually
During the sterilization phase both 11. Respectively close valves V5, V11 open MV7 to fill the filter housing
pressure and temperature should be and MV5, and keep MV6 and with product and vent the filter
recorded regularly. MV3 open to maintain a positive housing from MV5, until all
9. At completion of the sterilization pressure into the sterile filter system upstream air has been released.
cycle, close the steam supply valve while it is not in use. 4. When product is seen exiting the
MV1 and slowly open MV6 to hose attached to MV5, close the
introduce compressed gas into Post SIP, Pre-Use Filter Integrity Test
Procedure vent valve MV5 and continue to
the system. maintain the 2.8 bar pressure for
1. Ensure that V3, V6 and V7 on
CAUTION: Make sure that the the vent filter are open and that at least one minute to dissolve any
system remains under positive the sterile tank downstream of residual gas within the filter and
pressure (as indicated by pressure the product filter is vented at the ensure membrane wetting.
gauges P3 and P4) and control atmospheric pressure.
that the delta-P over the filter does
not exceed 350 mbar.
5
5. Fully open the downstream valve References assembled and sterilized prior to
V11 and gradually open V4 to set 1. Revision of annex 1 to EC Guide use. More importantly, however,
the differential pressure (P4–P3) at to GMP for Sterile Medicinal such testing should be conducted
approximately 200 mbar. Products; 1997. "The integrity of after the filter is used in order to
6. Continue to flow product through the sterilised filter should be verified detect any filter leaks or perforations
the filter to the sterile tank at before use and should be confirmed that may have occurred during
appropriate pressure differential immediately after use by an appro- filtration. Forward flow, bubble
for at least five minutes. priate method such as a bubble point and pressure hold tests are
point, diffusive flow or pressure acceptable integrity tests.”
7. Then close MV7 and MV3 to
hold test." 3. PDA Technical Report No. 26,
isolate the filter and fully open V4.
2. FDA Guideline on Sterile Drug Sterilizing Filtration of Liquids;
8. Open MV8 and attach an 1998. “Integrity testing of
Products Produced by Aseptic
automatic filter integrity tester. sterilizing grade production filters
Processing; 1987. “Normally,
9. Double-check that MV8, V11, V4, integrity testing of the filter is pre and post use is a fundamental
V3, V6 and V7 are open and performed after the filter unit is element of sterility.”
that MV3 and MV4 are all fully
closed and run the enhanced
bubble point test.
10. When the test is finished and a
positive result (i.e. pass) is obtained
close MV8 and detach the filter
integrity tester.
11. Open MV7, MV3 and MV5 to re-
start the filtration of the product.
12.When product is seen through Discover the More in Millipore™
MV5, close MV5 and continue In every application, every step and every scale, count on Millipore to be
the filtration of the product. everywhere for you—from monoclonals to vaccines, from clinical through
pilot to full-scale manufacturing. Our technologies are used by most of the
world’s major biopharmaceutical companies. But we deliver more than
advanced separation, purification, sterilization and quality control products.
With Millipore, you get services to optimize and validate your processes,
comprehensive resources to streamline and enhance your operation,
unmatched know how forged from nearly 50 years’ experience—
and solutions that integrate it all. For higher yields, improved process
economics and faster speed to market, discover the more in Millipore.

To Place an Order or Receive Technical Assistance


For additional information call your nearest Millipore office:
In the U.S. and Canada, call toll-free 1-800-MILLIPORE (1-800-645-5476)
In the U.S., Canada and Puerto Rico, fax orders to 1-800-MILLIFX
(1-800-645-5439)
Outside of North America contact your local office.
To find the office nearest you visit www.millipore.com/offices.
Internet: www.millipore.com
Technical Service: www.millipore.com/techservice

Millipore, Durapore and Aervent are registered trademarks of Millipore Corporation.


HydroCorr is a trademark of Millipore Corporation.

 Lit. No. TB015EN00 Rev. A 05/03 03-151 Printed in U.S.A.


© 2003 Millipore Corporation, Billerica, MA 01821 U.S.A. All rights reserved.

Das könnte Ihnen auch gefallen