revision date: September 28, 2006 expiry date: 29 September 2010

Laboratorium Kalibrasi TELKOM MSC - Surabaya


Accreditation Number : LK-015-IDN
Address : Jl gayungan PTT VII No. 17-19 Surabaya
Telephone : 031-8283678
Facsimile : 031-8283388
Contact Person : Rudi Wibowo,ST
Electrical

Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)

in-house procedure:
1 DC ammeter 0.0001 A ~ 1A 0.1 mA/A
SCL-SOP-CUR 002
1A ~ 10 A 0.15 mA/A kalibrasi DC ammeter
SCL-SOP-DMM-001
kalibrasi multimeter
3 AC ammeter 0.0001 A ~ 0.1 A in-house procedure
f =40 Hz ~ 1 kHz 0.18 mA/A SCL-SOP-CUR-003
f = 1 kHz ~ 10 kHz 1.8 mA/A kalibrasi AC Ammeter
0.1 A ~ 1A SCL-SOP-DMM-001
f =40 Hz ~ 1 kHz 0.18 mA/A kalibrasi multimeter
f = 1 kHz ~ 10 kHz 1.6 mA/A
1A ~ 10 A
f = 40 Hz-1 kHz 0.34 mA/A
f = 1 kHz-10 kHz 1.8 mA/A
in-house procedure
4 DC voltmeter 0.0001 V ~ 10 V 15 µV/V
SCL-SOP-VOL-002
10 V ~ 1000 V 10 µV/V kalibrasi DC Voltmeter
SCL-SOP-DMM-001
kalibrasi multimeter
5 AC Voltmeter 0.0001 V ~ 0.2 V in-house procedure
f = 40 Hz-1 kHz 4.3 mV/V SCL-SOP-VOL-004
f = 1 kHz-100 kHz 4.7 mV/V kalibrasi AC Voltmeter
f = 100 kHz-100 MHz 6.1 mV/V SCL-SOP-DMM-001
0.2 V ~ 20 V kalibrasi multimeter
f = 40 Hz-1 kHz 0.18 mV/V
f = 1 kHz-100 kHz 0.68 mV/V
f = 100 kHz-100 MHz 0.98 mV/V
10 V ~ 1000 V
f = 40 Hz-1 kHz 0.30 mV/V
f = 1 kHz-30 kHz 0.09 mV/V
6 Ohmmeter 1Ω ~ 10 Ω 95 µΩ/Ω in-house procedure
10 Ω ~ 10 kΩ 43 µΩ/Ω SCL-SOP-RES-003
10 kΩ ~ 100 kΩ 15 µΩ/Ω kalibrasi ohmmeter
100 kΩ ~ 1 MΩ 15 µΩ/Ω SCL-SOP-DMM-001
1 MΩ ~ 10 MΩ 65 µΩ/Ω kalibrasi multimeter
10 MΩ ~ 100 MΩ 224 µΩ/Ω

1 of 4
revision date: September 28, 2006 expiry date: 29 September 2010
Laboratorium Kalibrasi TELKOM MSC - Surabaya
Accreditation Number : LK-015-IDN
Address : Jl gayungan PTT VII No. 17-19 Surabaya
Telephone : 031-8283678
Facsimile : 031-8283388
Contact Person : Rudi Wibowo,ST
Electrical (Cont'd)
Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)

in-house procedure: SCL-


7 DC Current Source 0.0001 A ~ 1A 0.1 mA/A
SOP-CUR 001
1A ~ 10 A 0.15 mA/A kalibrasi AC/DC current source
8 AC Current Source 0.0001 A ~ 10 A in-house procedure:
f = 10 Hz ~ 20 Hz 0.35 mA/A SCL-SOP-CUR 001
f = 20 Hz ~ 40 Hz 0.35 mA/A kalibrasi AC/DC current source
f =40 Hz ~ 1 kHz 0.34 mA/A
f = 1 kHz ~ 5 kHz 0.52 mA/A
f = 5 kHz-10 kHz 0.56 mA/A
9 DC Voltage Source 0.1 V ~ 1000 V 6 µV/V in-house procedure
SCL-SOP-VOL-003
kalibrasi AC/DC Voltage source
10 AC Voltage Source 0.001 V ~ 1000 V in-house procedure
f = 20 Hz-20 kHz 0.56 mV/V SCL-SOP-VOL-003
f = 20 kHz-100 kHz 0.71 mV/V kalibrasi AC/DC Voltage source
11 Resistance 1Ω ~ 100 Ω 12 µΩ/Ω in-house procedure
100 Ω ~ 1 kΩ 12 µΩ/Ω SCL-SOP-RES-002
1 kΩ ~ 10 kΩ 12 µΩ/Ω kalibrasi DC Resistance
10 kΩ ~ 100 kΩ 12 µΩ/Ω
100 kΩ ~ 1 MΩ 12 µΩ/Ω
1 MΩ ~ 10 MΩ 53 µΩ/Ω
10 MΩ ~ 100 MΩ 103 µΩ/Ω
12 Capacitance f = 1 kHz; V = 1 Volt in-house procedure
1 pF ~ 1 pF 8.4 mF/F SCL-SOP-IMP-003
10 pF ~ 100 pF 1.6 mF/F kalibrasi standard capacitor
100 pF ~ 10 µF 6.2 mF/F
10 µF ~ 100 µF 9.5 mF/F
13 Inductance f = 1 kHz; V = 1 Volt in-house procedure
1 mH ~ 1 mH 3.2 mH/H SCL-SOP-IMP-004
1 mH ~ 100 mH 4.6 mH/H kalibrasi standard inductor
100 mH ~ 1H 8.6 mH/H
14 Capacitance Meter f = 1 kHz; V = 1 Volt in-house procedure
1 pF 7.9 mF/F SCL-SOP-IMP-001
10 pF 1.2 mF/F kalibrasi LCR meter
100 pF 1 mF/F
1 µF 4.4 mF/F
10 µF 6 mF/F
100 µF 12 mF/F
15 Inductance Meter f = 1 kHz; V = 1 Volt in-house procedure
100 µH 6.5 mH/H SCL-SOP-IMP-001
1 mH 3.1 mH/H kalibrasi LCR meter
10 mH 8 mH/H
100 mH 3.8 mH/H
1H 8 mH/H
10 H 4 mH/H

2 of 4
revision date: September 28, 2006 expiry date: 29 September 2010
Laboratorium Kalibrasi TELKOM MSC - Surabaya
Accreditation Number : LK-015-IDN
Address : Jl gayungan PTT VII No. 17-19 Surabaya
Telephone : 031-8283678
Facsimile : 031-8283388
Contact Person : Rudi Wibowo,ST
Electrical (Cont'd)
Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)
16 Wheatstone Bridge 0.1 Ω ~ 1Ω 0.72 mΩ/Ω in-house proc
1Ω ~ 10 Ω 0.79 mΩ/Ω SCL-SOP-RES-004
10 Ω ~ 1 kΩ 0.4 mΩ/Ω kalibrasi wheatstone bridge
1 kΩ ~ 10 kΩ 0.4 mΩ/Ω
10 kΩ ~ 100 kΩ 0.4 mΩ/Ω
0.1 MΩ ~ 1 MΩ 0.4 mΩ/Ω
1 MΩ ~ 10 MΩ 0.4 mΩ/Ω
10 MΩ ~ 100 MΩ 1 mΩ/Ω
17 DC Power Meter 0.0001 kVA ~ 10 kVA 0.8 mVA/VA in-house procedure
SCL-SOP-POW-001
18 AC Power Meter 0.0001 kVA ~ 10 kVA 1.5 mVA/VA kalibrasi Wattmeter
f = 45 Hz ~ 10 kHz
19 DC Clamp Meter 10 A ~ 550 A 3.4 mA/A in-house procedure
20 AC Clamp Meter 11 A ~ 551 A 3.5 mA/A SCL-SOP-CUR - 005
f = 40 Hz-10 kHz kalibrasi Clampmeter
21 High Resistance Meter 1 MΩ ~ 10 MΩ 1 mΩ/Ω in-house procedure
10 MΩ ~ 200 MΩ 1 mΩ/Ω SCL-SOP-RES-005
200 MΩ ~ 1000 MΩ 14 mΩ/Ω kalibrasi High Resistance meter
1000 MΩ ~ 10000 MΩ 15 mΩ/Ω
22 DC High Voltage Source 1 kV ~ 10 kV 6 mV/V in-house procedure
23 AC High Voltage Source 1 kV ~ 10 kV 6 mV/V SCL-SOP-VOL- 005
f = 50 Hz ~ 60Hz kalibrasi High Voltage Source
Time and Frequency
Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)
in-house procedure
1 Rb. Freq.std. 5 MHz 1.2 nHz/Hz
SCL-SOP-STD-0003
10 MHz 1.2 nHz/Hz kalibrasi Rb frequency std
in-house procedure
2 Frequency Counter 1E-08 GHz ~ 20 GHz 1.3 nHz/Hz SCL-SOP-FRE-001
kalibrasi frequency counter

in-house procedure
3 Oscilator 1 MHz ~ 10 MHz 4.2 nHz/Hz SCL-SOP-FRE-003
kalibrasi frequency generator
4 Signal Generator in-house procedure
Frequency 1E-08 GHz ~ 26.5 GHz 4.2 nHz/Hz SCL-SOP-FRE-003

Level 20 dB ~ 70 dB 1 dB kalibrasi frequency generator

FM Modulation 10 kHz (dev) ~ 400 kHz (dev) 29 mHz/Hz


AM Modulation 0.01 %mod ~ 99 %mod 3 %mod

in-house procedure
5 Time Mark Generator 2.00E-09 s ~ 5s 0.23 ms/s SCL-SOP-FRE-002
kalibrasi time mark generator

3 of 4
revision date: September 28, 2006 expiry date: 29 September 2010
Laboratorium Kalibrasi TELKOM MSC - Surabaya
Accreditation Number : LK-015-IDN
Address : Jl gayungan PTT VII No. 17-19 Surabaya
Telephone : 031-8283678
Facsimile : 031-8283388
Contact Person : Rudi Wibowo,ST
Time and Frequency

Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)

in-house procedure
6 RF Power meter -70 dBm ~ 10 dBm 1 dB
SCL-SOP-FRE-006/006A
10 MHz ~ 18 GHz kalibrasi RF power meter
in-house procedure
7 RF Attenuator -70 dBm ~ 10 dBm 1 dB
SCL-SOP-FRE-011
10 MHz ~ 18 GHz kalibrasi RF attenuator
8 Spectrum Analyzer in-house procedure
frequency 1E-06 GHz ~ 20 GHz 60 nHz/Hz SCL-SOP-FRE-004
Level 0 dBm ~ 100 dBm 1 dB kalibrasi spectrum analyzer
9 Osciloscope in-house procedure
Vertical deflection (V/division) 0.001 V ~ 130 V 1.3 mV/V SCL-SOP-FRE-002
Horizontal defelction (time/division) 2.00E-09 s ~ 5s 0.9 ms/s kalibrasi oscilloscope
10 Bit Error Tester (BER) in-house procedure
Amplitude 1.00E-03 Vpp ~ 60 Vpp 13 mV/V SCL-SOP-FRE-005
Frequency 1.00E-09 GHz ~ 20 GHz 4.3 nHz/Hz kalibrasi BER tester
11 AM/FM Modulation Meter in-house procedure
FM Modulation 1.00E+01 kHz (dev) ~ 400 kHz (dev) 46 mHz/Hz SCL-SOP-FRE-015

AM Modulation 0 % mod ~ 99 %mod 3.0 %mod kalibrasi AM/FM Mod

Optical
Best Measurement
No. Instrument to be calibrated Measurement Range Methods/Specifications Notes
Capability *)
1 Optical Time Domain 2138.4 m ~ 22226.1 m 0.92 m in-house procedure SCL-
Reflectometer SOP-OPT-004 kalibrasi
OTDR
2 Optical Power Meter -60 dBm ~ 5 dBm 1.7 % of in-house procedure
reading SCL-SOP-OPT-003
wavelength 1310 nm and 1550 nm kalibrasi optical power meter
3 Optical Light Source -60 dBm ~ 9 dBm 0.21 dB in-house procedure
SCL-SOP-OPT-003
wavelength 750 nm ~ 1800 nm kalibrasi optical light source
4 Optical Attenuator -60 dBm ~ 9 dBm 0.21 dB in-house procedure
SCL-SOP-OPT-002
wavelength 1310 nm and 1550 nm kalibrasi optical attenuator
Notes:
Best Measurement Capability is stated as Expanded Uncertainty at approximately 95% confidence level, calculated using
coverage factor of k = 2, includes uncertainty contribution due to reference standards used by the laboratory, measurement process
at the laboratory, and contribution due to typical characteristic of nearly ideal device under calibration respective to the laboratory

The actual characteristic of device under calibration shall also be considered for individual calibration certificate issued
by the laboratory

4 of 4