Beruflich Dokumente
Kultur Dokumente
Manual of Petroleum
Measurement Standards
Chapter 5—Metering
Section 6—Measurement of Liquid
Hydrocarbons by Coriolis Meters
Manual of Petroleum
Measurement Standards
Chapter 5—Metering
FOREWORD
This standard may involve hazardous materials, operations, and equipment. This standard does
not purport to address all of the safety problems associated with its use. It is the responsibility of
the user of this standard to establish appropriate safety and health practices and determine the
applicability of regulatory limitations prior to use.
API publications may be used by anyone desiring to do so. Every effort has been made by the
Institute to assure the accuracy and reliability of the data contained in them; however, the
Institute makes no representation, warranty, or guarantee in connection with this publication and
hereby expressly disclaims any liability or responsibility for loss or damage resulting from its use
or for the violation of any federal, state, or municipal regulation with which this publication may
conflict.
Suggested revisions are invited and should be submitted to Measurement Coordination,
American Petroleum Institute, 1220 L Street, N.W., Washington, D.C. 20005.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
4
Introduction
API Boiler Plate.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
5
TABLE OF CONTENTS
Chapter 5—Metering
1 Scope
This standard is intended to be a guide for the specification, installation, verification and operation
of Coriolis meters used to dynamically measure liquid hydrocarbons. API MPMS Ch. 5.6 also
includes information that will assist in troubleshooting and improving the performance of the
meters.
Use of a Coriolis meter as a stand-alone density meter is not addressed by the standard.
1.1 Field of Application
The standard describes methods to achieve custody transfer measurement of liquid hydrocarbons
quantities (mass or volume) using a Coriolis meter. Coriolis meters can provide outputs for mass
flow rate, volumetric flow rate and density. Guidance for the measurement of density is found in
API MPMS Ch. 9.
This document provides guidance for the application, installation, proving, mass calculations,
auditing, reporting and security requirements for Coriolis meters.
Some of the guidance provided within also applies to allocation measurement. API MPMS Ch.
20, Allocation Measurement allows for different performance requirements.
2 Normative References
The following referenced documents are indispensable for the application of this document. For
dated references, only the edition cited applies. For undated references, the latest edition of the
referenced document (including any amendments) applies.
API Manual of Petroleum Measurement Standards, (MPMS) Chapter 1, Vocabulary
API MPMS Ch. 21.2, Electronic Liquid Volume Measurement Using Positive Displacement
and Turbine Meters
API MPMS Ch. 21.2-A1, Addendum 1 to Flow Measurement Using Electronic Metering
Systems, Inferred Mass
API TR 2573, Standard Guide for Sediment and Water Determination in Crude Oil (ASTM
D7829)
For the purpose of this document, the following definitions apply. Refer to API Manual of
Petroleum Measurement Standards (MPMS) Chapter 1 for the definition of additional terms
used in this standard.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
7
3.1.1
calibration
The process of utilizing a reference standard to determine a coefficient which adjusts the output
of the Coriolis transmitter to bring it to a value which is within the specified accuracy tolerance of
the meter over a specified flow range. This process is normally conducted by the manufacturer.
3.1.2
cavitation
Phenomenon related to and following flashing if the pressure recovers and the vapor bubbles
collapse (implode).
3.1.3
Mass (Coriolis), M
The indicated mass from the meter corrected with an applied meter factor and the flow pressure
effect compensation factor.
3.1.4
Coriolis meter
A Coriolis meter is a transmitter and a sensor.
3.1.5
meter factor (Coriolis), mass or volume (MF, MFm, MFv)
A dimensionless number obtained by dividing the actual quantity of fluid passed through the meter
(as determined by proving), by the quantity registered by the meter. For subsequent metering
operations, the actual quantity is determined by multiplying the indicated quantity by the meter
factor.
3.1.6
transmitter (Coriolis)
The electronics associated with a Coriolis meter which interpret the phase shift signal from the
sensor, converts it to a meaningful mass flow rate (represented in engineering units), and
generates an output signal representing mass flow rate. Manufacturers can also use it to drive
the flow tube(s), determine fluid density, and calculate and output a volumetric flow rate.
3.1.7
sensor (Coriolis)
A mechanical assembly consisting of: A. Housing: the means of providing environmental
protection. This may or may not provide secondary containment. B. Measurement sensor (s):
Sensors to monitor oscillations and to detect the effect of Coriolis forces. These are also referred
to as pickups or pickoffs. C. Support structure: A means for supporting the flow tube(s). D. Flow
tube: Oscillating flow tube(s) or channel through which the fluid to be measured flows. E. Vibrating
drive system: The means for inducing the oscillation of the flow tube(s).
3.1.8
indicated mass (Coriolis) IM
The registered indicated mass quantity accumulated during the transaction period.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
8
3.1.9
manufacturer density calibration factor
Numerical factor(s) that can be used to compensate for the density sensitivity of each Coriolis
sensor. They are unique to the sensor and derived during sensor calibration. When programmed
into the transmitter, the density calibration factor(s) cause the meter density measurement
performance to meet its stated specifications.
NOTE The manufacturer density calibration factor should not be confused with Density Meter
Factor (DMF).
3.1.10
manufacturer flow calibration factor
A numerical factor that can be used to compensate for the flow sensitivity of each Coriolis sensor.
It is unique to the sensor and derived during sensor calibration. When programmed into the
transmitter, the flow calibration factor(s) causes the meter flow rate measurement performance to
meet its stated specifications.
NOTE The manufacturer flow calibration factor should not be confused with K-factor or Meter
Factor (MF).
3.1.11 net mass; NM
The mass (M) of the fluid without Sediment and Water (S&W).
3.1.12
pressure drop (pressure loss)
The difference between upstream and downstream pressures due to the frictional and inertial
losses associated with fluid motion in the entrance, exit, and internal passages of the flow meter
or other specified systems and equipment.
3.1.13 sediment and water mass: SWM
The mass of Sediment and Water during a measurement period.
3.1.14
zeroing
A procedure that sets the stored zero value.
NOTE The zeroing operation should not be confused with resetting the totalizer.
3.1.15
zero offset, observed
The difference between the current observed zero value and the previous observed zero value.
3.1.16
zero stability
The variability of the observed zero value during the process of determining the value.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
9
NOTE This is a systematic uncertainty, which can be present over the working range of the
meter.
3.1.17
zero value, observed
The average mass or volume flow rate observed reading under zero (no) flow conditions, without
inhibited flow indications (i.e., low-flow cutoff disabled and bidirectional flow enabled).
3.1.18
zero error
The observed zero value expressed as a percentage of normal flow rate.
3.1.19
zero value, stored
The value within the transmitter representing a no flow condition.
3.1.20
zero verification
The process of determining if the observed zero value is within the operators zero error limit (a
tolerance).
3.1.21
zero value, as-found
The observed zero value before a meter is zeroed.
3.1.22
zero value, as-left
The observed zero value after a meter has been zeroed.
MFm = meter factor when the Coriolis meter is configured to indicate mass
4 General
A Coriolis meter consists of a sensor and a transmitter. A sensor has one or more flow tubes
through which the fluid flows. The flow tube or tubes are made to vibrate at their natural or
harmonic frequencies by means of an electromagnetic driving mechanism. The fluid flowing
through the tube(s) generates a Coriolis force that is directly proportional to the mass flow rate of
the fluid. The magnitude of the Coriolis force is measured and used to determine a mass flow
rate. The transmitter powers the sensor, processes the output from the sensor in response to
mass flow, and generates signals for the tertiary device(s).
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
10
The frequency of the oscillating flow tube(s) can be measured and used to determine the density
of the fluid. Density determination is independent of the mass flow rate determination.
A meter typically can determine a volumetric flow rate through calculation by dividing the mass
flow rate with either the measured flowing density or a user programmed (fixed) density.
5 Application
Avoid cavitation or flashing in the meter and the prover. Cavitation will cause a measurement
error and can damage the sensor. Equation B.1 in Annex B provides a guideline to maintain
sufficient pressure downstream of the meter and the prover.
A provision for adequate draining, vapor elimination, and cleaning should be considered based
upon the fluid characteristics and sensor design. Sensors should be installed to avoid trapping
vapors, particulates, or liquids heavier than the process liquid.
Different flow tube geometry can be more or less likely to promote the accumulation of sediments
or coatings within the tube(s). Besides restricting flow, the accumulation of material within the flow
tube(s) can affect the observed zero, and the accuracy of the density and volume measurements.
Higher fluid velocities, when coupled with abrasive particles in the stream, can cause flow tube(s)
erosion and measurement error. The user should ensure that the flow tube material is compatible
with operating conditions and fluid properties including any corrosiveness or abrasiveness
characteristics.
5.1.2 Accuracy
Sensors have different accuracy specifications and sensitivities with regard to changes in ambient
and operating conditions. Consult the manufacturer for the performance envelope describing error
limits throughout these conditions and consider the limits with respect to the system requirements.
To address the effects of process pressure and temperature on the accuracy of the measurement
of mass, density and the corresponding determination of volume:
Temperature effect compensation for the flow tube(s) stiffness shall be integral to the
meter and performed real time in the meter.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
11
Pressure effect compensation for flow tube(s) stiffness is recommended for all applications
but is required when the effect is biasing quantity determination by more than 0.05 %.
Generalized equations for evaluation of the effect are provided in Annex D.
Note: This pressure effect compensation is not the correction for compressibility of liquid (CPL).
Pressure effect on flow tube stiffness can also affect density determination and its use
within volume calculations (e.g., CTL, CPL, CTPL). Refer to API MPMS Ch. 9, Density
Determination.
o Activating the pressure effect compensation mode in the transmitter (if available) and using
a live pressure input (analog or digital communication) to continuously adjust for pressure
change. Pressure values should be shared between the transmitter and tertiary device for
audit purposes. Or,
o Activating the pressure effect compensation mode in the transmitter and using a live
pressure measurement integral to the sensor. Pressure values should be shared between
the transmitter and tertiary device for audit purposes. This integral pressure value may not
be suitable for other calculations requiring a pressure value. Or,
o Calculating the pressure compensation in a tertiary device. See Annex D for details.
In lieu of real-time compensation, pressure compensation is turned off in the transmitter and:
o Prove the meter close enough to the average operating pressure to satisfy equation D.2 in
Annex D (while not affecting flow rate enough to affect meter performance), or;
o Correct the ticket external to the metering system (See Annex D and E).
Activating the pressure compensation in the transmitter with a fixed pressure shall not be allowed.
Any change in pressure compensation values or implementation of pressure compensation
requires proving. (Reference Section 7.2.7)
In both mass and volume measurement applications, this standard requires documentation of
flow tube pressure effect compensation in accordance to Annex D.
5.1.4 Documentation
Meter manufacturers provide calibration certificates. Other documentation for the meter such as
electrical area classification certification, hydrostatic test report, and material test reports can be
provided.
5.1.7 Crosstalk
With multiple sensors in close proximity, either in parallel or series, the vibrations generated by
each sensor could interfere (crosstalk) with each other, thereby causing measurement error.
Vibration isolation or dampening can be achieved by altering piping or supports. Consult the
manufacturer to reduce the possibility of crosstalk.
5.2.1 Environmental
Evaluate the temperature, humidity and corrosion extremes at the installation for appropriate
protection.
5.2.2 Electrical
Select the transmitter and sensor to meet the required electrical area classification. Consider the
power requirements for the sensor and transmitter. Design the electrical signal system to provide
appropriate fidelity and security. (Reference API MPMS Ch. 5.5)
The sensor, transmitter, and their interconnecting cables are all susceptible to Electromagnetic
Interference (EMI). Since the electrical signals of the meter are at relatively low power levels, care
should be taken to avoid interference generated from nearby electrical equipment and wiring.
Care should be taken when selecting the values to ensure that the following two conditions are
satisfied:
When fluid is flowing through the meter at maximum operational flow rate—the frequency
output by the transmitter should neither exceed 90 % of its rated maximum nor 90 % of the
maximum rated input frequency of the tertiary devices(s) receiving the signal.
When fluid is flowing through the meter at minimum operational flow rate —the frequency
output by the transmitter should be high enough to produce sufficient pulse resolution to
provide appropriate flow rate and quantity indications, especially for proving.
Low-flow or no flow cut-offs can be programmed in the transmitter or the tertiary device. These
values inhibit flow totalization for very low flow or true no flow conditions that can occur.
Coriolis meter based measurement system design considerations for custody transfer metering
can be found in Annex B and API MPMS Ch. 6.
6.1 Mounting
Follow the manufacturer recommendations for the proper mounting of the sensor. Consideration
should be given to the support of the sensor, the alignment of the inlet and outlet flanges with the
sensor, and the orientation of the sensor.
Piping should be supported or anchored to avoid transferring stresses from the piping to the
sensor. Meter performance, specifically observed zero offset, can be adversely affected by axial
bending and torsional stresses from pressure, weight, and thermal effects. These stresses and
associated loads can be minimized by utilizing properly aligned pipe and well-designed pipe
supports.
During the system construction phase, a pipe spool piece equal in length to the meter, should be
used in place of the meter to correctly align pipework.
Sensor orientation depends on the application and the geometry of the flow tube(s) and should
be recommended by the manufacturer. For example, solids settlement, plugging, coating, or
trapped gas can affect the meter performance.
6.2 Installation Conditions
6.3.1 Fill
To avoid damage to the sensor, a pipe spool should be installed in place of the meter each time
new piping or fluids are introduced into the piping system that can contain deleterious materials
from construction or maintenance activities.
Avoid cavitation, flashing, and fluid hammer when filling to prevent damage to the sensor. Avoid
shock loading caused by rapid opening or closing of valves.
An improperly zeroed meter or a shift in observed zero value can result in measurement bias. A
zero shift is usually caused by variations in temperature, pressure or density, or changes in the
mounting conditions as a result of poor installation practices. The bias is a constant offset in flow
rate. Thus, this constant offset will result in a percent error that increases as the flow rate
decreases.
6.4 Conditions Affecting Meter Performance
Performance can be affected by conditions that change the flexibility of the oscillating flow tube(s),
as well as, fluid properties, operating conditions, and installation conditions. If there are changes
in those properties or operating conditions, or there is an alteration to the sensor installation, a
change in meter factor can result.
Flow rate ----Fluid velocity can affect the density measurement. When the volumetric
flow rate is determined from meter’s measured density, higher velocities can produce
measurement errors, with the error increasing as the velocity increases.
Manufacturers can compensate for this effect in the transmitter.
As the flow rate approaches no flow, meter factor variation increases due to the zero
offset (a measurement bias primarily affecting the lower range of the meter) and zero
stability. See Figure 1 and 2.
Fluid temperature changes within the sensor and connecting piping can create piping
stresses which can change the observed zero value.
Flashing or cavitation within the sensor— Flow tube geometries and sensor designs
can create a pressure area within the sensor which is lower than the outlet pressure.
Sufficient backpressure shall be maintained on the sensor and prover to prevent
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
16
Erosion of the flow tube—Abrasive solids can reduce the flow tube thickness, which in
severe cases can lead to meter and density factor shifts and sensor tube failure.
Volume meter factors are impacted more than mass meter factors.
Corrosion of the sensor—Flow tube material compatibility with the fluid is essential for
reliable service.
Density— Changes in fluid density can affect measurement and could require a new
proving of the meter.
Viscosity— Change in the fluid viscosity can affect meter performance and meter
factors and at some point, can require proving of the meter to minimize the effect.
Some meters can compensate for the changes in viscosity. The manufacturer
recommendation for viscosity compensation should be followed. Additionally,
boundary layer thickening inside the tube caused by increasing viscosity, affects
operating flow range of the meter.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
17
6.5 Maintenance
7 Proving
7.1 General
This section addresses proving issues related specifically to Coriolis meters. Detailed
requirements for operations of provers and meter factor determination requirements are provided
in API MPMS Ch. 4.8.
Coriolis meters are unique in that they can be used to indicate in either mass or volume. A meter
is calibrated by the manufacturer to determine one or more calibration factors that are entered
into the transmitter. These factors, although adjustable, should remain unchanged. A proving shall
be accomplished after changes to any calibration factors within the transmitter.
7.2 Proving Considerations
7.2.1 Conditions
Proving conditions should be as close to the actual metering conditions as practical. See API
MPMS Ch. 4.8 for additional details.
The meter should be proved on a similar fluid and under similar operating and installation
conditions as encountered in normal operations. If operating or installation conditions change, the
meter should be proved under the new condition as soon as practical.
7.2.2 Density
Density stability is critical during proving for either mass or volume outputs. For proving
calculations, it is important to distinguish between flowing, observed, and base density, how each
is determined, and when and where each are applied to a proving.
There are various ways to determine density. Density measurements can be made online or off-
line via a representative sample as recommended in API MPMS Ch. 9.
Density method determination can be calculated from fluid composition, published equations or
be contractually controlled. The appropriate API physical property standards should be referenced
for more information on density determination and calculation.
Record density values as recommended in API MPMS Ch. 12.2 and Annex F (mass calculation).
7.2.4 Reproducibility
Meter factor reproducibility is the ability of a meter to generate results over a period of time where
the range of variation of (change in) pressure, temperature, flow rate, and physical properties of
the liquid is negligibly small. The expected reproducibility should be determined by the operating
company based on financial risk and experience with each individual meter and proving system
or meter linearity.
For inferred mass proving, where density is to be determined at the prover, the random uncertainty
and reproducibility of the mass meter factor is affected by meter performance, the method of
prover density determination, the number of density samples or readings taken, and the density
meter proving. Assessment of the mass meter factor should include evaluating the prover density
uncertainty. See API MPMS Ch.4.8 for additional information.
Determining acceptable meter factor reproducibility is an operating company responsibility. When
the reproducibility of a meter factor is outside the operating company tolerances, the meter zero
(observed zero value) needs to be verified. See Annex A for zero verification procedures.
Needless zeroing should be avoided in order to minimize potential errors associated with meter
factor reproducibility.
Industry practice is to allow a greater combined reproducibility tolerance for inferred mass
measurement systems which use volume (e.g., turbine) and density meters. The reproducibility
for these systems is the sum of the actual meter reproducibility of the volume meter and density
meter. Thus, the two meters factors can cancel out each other or produce a larger reproducibility
tolerance (larger MF shift). This greater tolerance should be extended to Coriolis meters that are
proved by inferred mass method. Common practice for custody transfer applications is to accept
new meter factors within 0.10 % to 0.50 % of the previous meter factor. Action and tolerance limits
will be defined individually by each operating company. Guidance on action and tolerance limits
is provided in API MPMS Ch. 13.2.
7.2.6.1 General
The methodologies used to prove a Coriolis meter are direct mass, inferred mass, and volumetric.
These methodologies differ significantly in the way they determine the reference quantity of fluid
(prover quantity) for a proving. The reference quantity should match the engineering units of the
meter’s output. See API MPMS Ch. 4.8 for details of the approved proving methods.
The result of a meter proving will be a new meter factor. All meter factors (MFm, MFv, DMF) shall
only be applied in devices that are compliant with API MPMS Ch. 21
Application of a meter factor in a transmitter that will change the pulse output is not allowed. No
factors or values (meter factor, calibration, or configuration factors) that affect the pulse output or
indicated quantity shall be changed in the transmitter after proving. When it is necessary to
change calibration or configuration factors or perform maintenance on the meter, perform a prove
to check for changes in meter factor. Reference API MPMS Ch. 4.8.
API MPMS Ch. 21.2 addresses auditing and reporting requirements for Electronic Liquid
Measurement (ELM) systems. The audit requirements for quantity (volume or mass) transactions
using a Coriolis meter are similar. Configuration and setup parameters factors contained in the
transmitter and tertiary devices shall be recorded in these logs and records. The information in
the transmitter shall be secured. The information in the transmitter shall be auditable.
In addition to the items above, the event of zeroing of the Coriolis meter shall be recorded,
including the date and time the meter was zeroed, along with the as-found and as-left observed
zero value. Each zero value has a corresponding meter factor associated with it. Recording the
stored zero and zero stability values are recommended.
8.5 Alarm and Error Log
Alarm and Error Logs shall be recorded and provided as part of audit packages for sensors, their
transmitter and tertiary devices. See API MPMS Ch. 21.2 for guidance on Alarm Logs.
8.6 Security and Access
Transmitters shall be secured against tampering or unauthorized or un-documented changes to
any variable or parameter that can alter the flow or density measurement. This can be achieved
by several methods including passwords or tamper evident seals or locks. See API MPMS Ch.
21.2.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
24
ANNEX A
(Normative)
Zero Verification
Initially, a meter shall be installed and brought to line or process conditions prior to evaluating the
zero value. Failure to properly purge the sensor prior to evaluating the zero can lead to an
incorrect evaluation. Once the sensor is at operating conditions, observe and record the flow rate,
safely block in the sensor, observe and record the zero value.
As a part of the meter proving process or when meter factors shift beyond an operator tolerance,
observe the zero value and record it for evaluation. This observed zero value should be
determined at the same conditions under which the meter factor was determined.
The need for zeroing is based on an operator’s tolerance. The operator’s tolerance should be
compared to the effect (Err0) that the observed zero value has on the operating flow rate
indication. Use equation A.1 to estimate the percent error that can be contributed by the zero
value.
q
Err 0 = -----0 100
qf
(A.1)
where
Err0 = zero error (%)
q0 = observed zero value
qf = flow rate during normal operation
To determine if the meter needs zeroing, evaluate if the value of Err0 of equation A.1 is within
operator’s tolerance (e.g., 0.04 %). If the value is not in tolerance, zeroing should bring the zero
error within tolerance.
In the case of the initial verification (startup verification) the recorded zero value (as found) is
assumed to be “zero”. The meter shall be proved after the meter has been zeroed.
After performing a zero operation, the new observed zero value (as left, current) should be
compared to the recorded zero value (as found, previous).
Equation A.2 can be used to estimate the shift in the meter factor relative to the difference in “as
found” and “as left” zero values.
where
If an observed zero value cannot be obtained within an acceptable tolerance an investigation into
the cause should occur.
In order to verify the observed zero value of the meter, there shall be no flow through the sensor.
The sensor shall be filled with the liquid to be measured at normal operating conditions. The
procedure to verify the zero value is as follows:
1. Open bypass valve if so equipped.
2. Stop flow through the sensor by closing the downstream double block-and-bleed valve
and ensure seal integrity.
CAUTION: Isolating a fluid-filled system (blocking it in) can result in elevated pressures if the
temperature rises.
4. If a threshold is set in the meter to suppress the indication (low-flow cutoff), then this
value should be set to zero during this procedure. Some transmitters allow the user to
view the actual flow rate bidirectionally and independently of the low-flow cutoff setting.
5. The observed zero value will not normally be constant. Small variations of the observed
zero value are expected. A properly zeroed meter will indicate a small flow rate
fluctuation between positive and negative flow rates within the specified zero stability for
that meter when observed with bidirectional indication and the low flow cutoff set to zero.
If the observed zero value is found to be outside the operator expectation for error (see
Equation A.1) evaluate the installation for the conditions listed in Section 6.4 and the
following:
possible pipe stress
leaking valves
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
26
If the observed zero value is found to be within the operating company expectation for error, reset
the low-flow cutoff and flow direction to original settings.
For examples of how to determine acceptable observed zero value reference Section A.2.
A log or graph should be maintained of all meter factors and zero adjustments (observed zero
values – as found and as left) so that trending of the meter performance can be conducted per
API MPMS Ch. 13.2.
The following subsections are informative examples of how to evaluate zero values in order to
determine if the meter needs to be zeroed. Normally, a zero shift can be resolved by properly
installing the meter, purging the line and bringing the meter to line conditions.
Proving the meter may not resolve a zero shift problem (error) if:
the meter requires maintenance or repair, or
the flow rate changes significantly from the proving flow rate.
If the flow rate at operating conditions is less than 10 % of maximum meter range, the operator’s
tolerance may need to be increased to accommodate inherent zero stability or poor stability due
to installation effects.
A new meter was installed. Since there is no initial “observed” zero value for this new installation
the assumed value for ZV would be “zero” (as found).
The new meter was brought to operating conditions and the observed zero value of the meter at
the current conditions was +0.7 lb/min. The difference in the current zero value from the factory
(0, as found)) is 0.7 lb/min. The expected flow rate of the line meter is 6005 lb/min, then:
Typically, the meter would not require zeroing because this observed zero value could only
contribute 0.0001 bias in meter factor if or when proved at 6005 lb/min flow rate.
A meter was initially proved at a flow rate of 5000 lb/min with a +2.5 lb/min zero value (as found).
A meter factor of 0.9995 was generated.
At the next proving the meter factor was 0.9968. The observed zero value of the meter after the
proving and at the current conditions was +17 lb/min. The difference in the current observed and
the recorded zero value (as found) is 14.5 lb/min. If the line meter is flowing at 5005 lb/min, then:
Typically, the meter should be zeroed because this zero shift could be the cause of the 0.0027
meter factor shift. After zeroing the new observed zero value should be recorded and flow error
calculated based on the same equation.
Prove the meter after zeroing and the new meter factor should be close to the original meter factor
of 0.9995 when the zero value was +2.5 lb/min. A new observed zero value should be obtained
after the prove and it becomes the “as found” zero value for the new meter factor.
A meter was initially proved at a flow rate of 7000 lb/min with a +2.5 lb/min zero value (as found),
and a meter factor of 0.9997 was generated.
At the next proving the meter factor was 0.9990. Typically, no action would be taken since the
meter factor only shifted 0.0007.
If the observed zero value was determined to be +3.5 lb/min., the difference to the recorded zero
value (as found) is 1 lb/min. If the flow rate for this proving was 7700 lb/min, then meter does not
need to be zeroed because this zero shift would only contribute to a 0.0001 shift in the meter
factor.
A meter was initially proved at a flow rate of 6000 BPH with a +2.5 BPH zero value (as found).
The meter factor of 0.9996 was generated.
At the next proving the meter factor was 0.9958. The observed zero value of the meter after the
proving and at the current conditions was +25 BPH. The difference in the current observed zero
and the recorded zero value (as found) is 22.5 BPH. If the flow rate for this proving was 6006 BPH,
then;
Typically, the meter should be zeroed. After zeroing a new observed zero value was 20 BHP. This
zero value will contribute to a 0.0029 meter factor shift from the 0.9996.
An investigation should occur to resolve the reason that an acceptable zero value could not be
obtained (it should be close to 2.5 BPH). A new proving with the 20 BPH should generate a meter
factor near 0.9971.
The new proving will only correct for this high zero value at this flow rate. This zero value will
contribute to more significant meter factor shifts when the meter is operated at flow rates lower
than 6,000 BPH. See next two examples.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
28
A meter was initially proved at several flow rates to determine a meter factor curve with a +2.5
BPH zero value (as found).
The observed zero value of the meter at the next verification was 7.5 BPH. The difference in the
observed zero and the as found value is 5 BPH.
If the meter is proved again at the same rates to verify the MF curve without zeroing, equation
A.2 can be used to estimate meter factor shift at any flow rate over the range of the meter. See
the Table A-1 below.
A meter was proved at 5 flow rates to establish a linearity curve. The “as found” zero was 2
lb/min.
The meter was zeroed and the observed zero was 0.25 lb/min. The meter was proven again at
similar rates to establish new meter factor linearity. The 0.25 lb/min observed zero value would
be the new value for the old curve above and becomes the “as found” value for the new meter
factor curve.
Figure A-2 – Meter Factor Curve “as left” 0.25 lb/min zero
Equation A.2 can be used to estimate and validate the shift in the linearity curve.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS
30
B.1 General
Consideration should be given to the following items before using Coriolis meters for custody transfer
measurements:
External vibrations at specific frequencies can cause measurement errors.
Two-phase flow (liquid/gas) is unacceptable.
Meter systems should comply with all applicable codes and regulations. A schematic diagram of a
typical meter installation is shown in Figure B-1.
B.2 System Design
The meter should be oriented in a position that will assure that the flow tube(s) are completely filled with
fluid under all flow and static conditions. Provisions should be made to not measure flow during no-flow
conditions if gas can accumulate in the flow tubes and cause false readings (e.g., spurious or ghost
counting).
Where the flow range or pressure drop is too great for one meter, the installation of a bank of meters in
parallel can be used. When more than one meter is installed in parallel, a means should be provided to
balance flow through the meters and isolate the meters for proving purposes.
Any condition that tends to contribute to vaporization or cavitation of the liquid stream should be avoided
by system design and by operating the meter within its specified flow range. Vaporization or cavitation
can be minimized or eliminated by maintaining sufficient pressure in and immediately downstream of the
meter. In lieu of actual test data to determine back-pressure requirements at the outlet of the meter
station, the following equation can be applied:
P b = 2p + 1.25 p e
(B.1)
where
Pb = minimum gauge back pressure at meter station outlet
Δp = pressure drop across the sensor at the maximum operating flow rate
NOTE For some dense-phase fluids, such as ethylene and high-purity ethane, these guidelines might
not be sufficient.
Two-phase flow (liquid/gas) affects meter performance and is unacceptable for custody transfer. A meter
installation should be equipped with air/vapor elimination equipment so that measurement accuracy is
not degraded by air or vapor.
For volumetric measurement, thermowells for the temperature transmitter and test well should be
installed near the sensor so that the measured temperature is representative of the fluid temperature in
the meter. For mass measurement it is a good operational practice for thermowells and a temperature
transmitter to be installed. Normal practice is to install thermowells downstream of the meter.
A pressure transmitter should be installed near the sensor. This pressure measurement is used to
compensate for pressure effects (pressure compensation) on flow tube(s) and for fluid compressibility
correction (CPL) in volume applications.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 32
Strainers or other protective devices should be installed upstream of the meter to remove foreign objects
which can cause measurement error.
Provide easy access to the transmitter for servicing, display readout and proving connections. Provide
physical access for portable proving and service vehicles.
Avoid installations near sources of flow pulsation and vibration.
Special considerations should be given to bidirectional installations to minimize the effect of flow-limiting
devices on the meter performance.
The meter should be protected from pressure surges as well as from excessive pressures caused by
thermal expansion of the fluid when the installation is not operating. A relief valve, if used, should not be
installed between the prover and the meter.
B.6 Proving Facilities
Facilities should be provided for proving the meter under normal operating conditions.
Stability of temperature, pressure, flow rate, and product composition is necessary to achieve acceptable
proving repeatability.
Metering systems should be provided with either manual or automatic means to permit proving the meter
under conditions of flow rate, flow direction, pressure, temperature, and fluid characteristics that exist
during the normal operation of the meter.
Connections for proving should be installed so air or vapor is not trapped in the piping between the meter
and the prover. Adequate bleed-off connections should be provided (see API MPMS Ch. 4.8). Reducing
the distance between the meter and prover can minimize problems with proving results. It is
recommended that the sensor be located upstream of the proving connection.
Consider the location and distance between the proving computer and the transmitter. Unlike other meter
types, where the pulse generation is located at the primary element, the Coriolis meter pulses come from
the transmitter. If the transmitter is not located near the meter and proving computer, then a remote
termination junction box should be provided near the proving computer to provide access to the meter
pulses.
B.7 Diagnostic Measurements
The inherent design of a Coriolis meter can offer diagnostics that through a manual or automatic process
identify conditions that affect meter performance. Diagnostic methods can require the use of an external
tool or can be integrated into meter design.
The following lists examples of parameters or analysis measures that some meters provide for diagnostic
measurement via a local display or a digital interface (e.g. RS-232, RS-485):
EPROM checksum
Configuration change flag
Tube vibration power indication
Pickoff or signal amplitude
Temperature output(s)
Live zero flow (no flow) indication
Status and measurement quality indicators
Alarm and failure indicators
Flowing density or flow tube resonant frequency
Flow tube health indication
Flow tube balance or symmetry
Frequency output test
Digital status output test
Analog output test
NOTE Please consult manufacturer for the diagnostic parameters that are available.
To further optimize the use of diagnostics, the operator should establish a baseline of the meter diagnostic
indicators during initial installation and meter proving.
Deviations from baseline diagnostics are useful in establishing acceptance criteria. During the proving,
meter diagnostics can be monitored for alarm or an out of tolerance condition. If an advanced diagnostic
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 34
for flow tube health exists, this diagnostic test can be performed after proving is complete and results
reported. Actual diagnostic capabilities vary by design; the user should consult with the manufacturer to
determine an appropriate set of diagnostics for the particular design.
A meter log file generated at calibration can establish meter baseline data. Meter log data and the results
of an automated flow tube health diagnostic should be included to provide a baseline of the meter
performance at proving. Baseline data can be used to verify meter performance upon startup, during
operation, and after component changes. Baseline data can also be useful in conducting health checks
of the meter. It is recommended that the manufacturers identify the parameters that define the baseline
performance for their products.
B.8 Installation Baseline
Operators typically establish the field baselines during the initial installation and startup of the meter.
Parameters that can be used to baseline meter performance are drive current, pickoff signal amplitude,
and zero value. Manufacturers often offer diagnostics that infer change in meter performance or flow
factor. These baseline relationships are useful in establishing acceptance criteria for the various
relationships and the need for a flow performance test (proving) and adjustment (if necessary).
B.9 Maintenance
Users should follow manufacturer recommendations for maintenance. Monitoring diagnostics, trending
of performance indicators and conducting periodic meter verification procedures can identify if abnormal
conditions develop.
Maintenance procedures, like cleaning, should be condition-based (based upon meter factors, process
conditions, diagnostics, and/or meter usage). The monitoring of performance indicators can identify the
need for cleaning. For example, by monitoring the measured flowing density and comparing it to the
actual fluid density, it is possible to infer coating on the flow tube(s). Performance indicators available to
the user are design specific and the meter manufacturer should be consulted on performance indicators
available and their interpretation.
B.10 Meter Verification
The field verification of a Coriolis metering system consists of monitoring and evaluating metering
conditions, diagnostic indicators output by the transmitter, or ancillary devices of the metering system
designed to identify possible change in system performance and the cause. The evaluation of these
indicators might guide the operator in determining the need to re-zero the meter, execute a proving,
adjust maintenance intervals, and implement improvements, if necessary, to the metering system.
The operator should follow design-specific meter verification procedures recommended by the
manufacturer and as a minimum the following general meter verification procedures should be
performed:
Diagnostic indicators—Verify all transmitter diagnostic indicators to be in the normal state.
Zero check—Perform a zero verification. See Annex A.
Transmitter calibration factors—Verify the flow and density calibration factors and any correction
factors in the configuration of the transmitter to be unchanged from most recent proving. The meter
transmitter verification should coincide with the zero verification.
Sensor Verification —Sensor diagnostics can be available that continuously, on-command, or
procedurally verify the performance of the sensor or infers change in measurement performance.
Users should consult the meter manufacturer for the availability of these types of diagnostics.
Annex C
(Informative)
Factory Calibration
During factory calibration, the output of a Coriolis meter under test is compared with a standard of lower
uncertainty to establish an initial calibration factor. Coriolis meters are typically factory calibrated on
gravimetric flow stands that are traceable to a national standard or by comparison to other traceable flow
standards.
Gravimetric flow stands typically flow water through the meter and then collect the water in a tank on a
weigh scale (see Figure C-1). The mass readout from the meter is compared to the weigh scale mass
indication, corrected for buoyancy effect. From this comparison a calibration factor is developed which
converts the sensor output to a flow rate in desired engineering units. Multiple tests at different flow rates
establish a repeatable calibration factor within the meter accuracy tolerance.
A separate calibration is performed on the density measurement of the meter. Density (ρ) of the fluid
contained within the vibrating flow tube is inversely proportional to the flow tube frequency (ƒ) of the
sensor assembly. This relationship is represented by the following equation.
1
---2
f (C.1)
where
ρ = density
ƒ = flow tube frequency, resonant frequency of the sensor
Factory calibration of density involves measurement of the flow tube frequency from the sensor and the
density value of the calibration fluid while the meter is full of fluid, either in a static or flowing state. Most
density meters are calibrated with two or more fluids (generally air and water) whose density values are
well defined and relatively stable. Density values for the fluid(s) can be determined either by using
equations of state or calculations from laboratory data or by incorporating a transfer standard to determine
the density of the fluid during calibration.
All the factory calibration references should be measured with instruments traceable to National Institute
of Standards and Technology (NIST) or another internationally recognized national metrology institute
(NMI).
The factory calibration produces the manufacturer density calibration factor(s) that defines the density
response for each density meter. An equation is then linearly interpolated or extrapolated by the
transmitter to determine fluid density measurement.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 36
Annex D
(Normative)
Pressure Effect Compensation for Tube Stiffness
D.1 General
Section 5.1.2 of this standard recommends the use of a Pressure Effect Compensation for the effects of
tube stiffness for all Coriolis meter applications, but is a requirement when the effect is biasing quantity
determination by more than 0.05%.
Pressure effect compensation for tube stiffness is applicable to both mass and density measurements of
a Coriolis meter. Therefore, it applies to both mass and inferred volume determination. Operators should
consult Manufacturers on the applicability and details of this correction for each specific meter design.
Pressure effect compensation applies to mass and volume quantities. It is a requirement to evaluate the
variation of operating pressure, its impact on mass and/or volume determination, and if pressure
compensation is necessary.
To evaluate the effect of pressure on mass measurement as required in Section 5.1.2, the following
generalized equations shall be used. Equation D.1 is used to evaluate the design stage and Equation
D.2 is used to evaluate the operational stage. Refer to the meter manufacturer for a meter’s mass
pressure effect (𝑃𝐸 ).
Where: 𝑃𝐸 = Pressure effect in percent (%) of mass flow rate per unit of gauge
pressure (manufacturer value)
To evaluate the effect of pressure on volume measurement as required in Section 5.1.2, the following
generalized equations shall be used. Equation D.3 is used to evaluate the design stage and Equation
D.4 is used to evaluate the operational stage. Refer to the meter manufacturer for a meter’s pressure
effect for mass (𝑃𝐸 ) and density (𝑃𝐸 ).
⎡ 𝟏 ∗ 𝑷𝒎𝒂𝒙_𝒐𝒑 𝑷𝒎𝒊𝒏_𝒐𝒑 ⎤
⎢ 𝝆𝒎 ∗ 𝑷𝒎𝒂𝒙_𝒐𝒑 𝑷𝒎𝒊𝒏_𝒐𝒑 ⎥
Design 0.05% ⎢ 𝟏 𝟏⎥ ∗ 𝟏𝟎𝟎 D.3
⎢ 𝟏 ∗ 𝑷𝒎𝒂𝒙𝒐𝒑 𝑷𝒎𝒊𝒏𝒐𝒑 ⎥
⎣ 𝟏𝟎𝟎 ⎦
⎡ 𝟏 ∗ 𝑷𝒘𝒂 𝑷𝒎 ⎤
⎢ 𝝆𝒎 ∗ 𝑷𝒘𝒂 𝑷𝒎 ⎥
Operational 0.05% ⎢ 𝟏 𝟏⎥ ∗ 𝟏𝟎𝟎 D.4
⎢ 𝟏 ∗ 𝑷𝒘𝒂 𝑷𝒎 ⎥
⎣ 𝟏𝟎𝟎 ⎦
Where: 𝑃𝐸 = Pressure effect in percent (%) of mass flow rate per unit of gauge pressure
(manufacturer value)
This standard prescribes four locations or options for applying pressure compensation. The methods are
identified by the Pressure Effect Compensation Indicator (PECI) status listed below:
A Mass Pressure Effect Compensation Factor (𝐶𝐹𝑃𝑆 ) is used to correct the mass determination for
the pressure effect on the flow tube(s).
Note: This factor is not the same as CPL for compressibility of the liquid.
Equation D.5 is a generalized equation that applies to the mass measurement of most Coriolis meters.
It can be used to quantify the correction needed.
1
𝐶𝐹𝑃𝑆 D.5
𝑃𝐸
1 ∗ 𝑃𝑓 𝑃
100
𝑃𝐸 = Mass flow pressure effect in percent (%) of rate per unit of gauge pressure
(manufacturer value)
𝑃 = Mass calibration pressure when PECI equals 1 (CT) or average meter proving
static gauge pressure when PECI equals 2 (TD) or 3 (OS)
The mass flow pressure effect correction algorithm of some Coriolis designs might not conform to the
above generalized equation. Operators should consult manufactures on the applicability of this algorithm
for the specific meter design to be used and method of implementation.
The actual pressure values or recordings vary for equation D.5. They vary by the choice of calculation
location or status indicator choice.
Pressure values for equation D.5 are based upon the following:
The PECI may be set to “4” if the operating pressure is maintained at or near proving pressure so that
any bias in the quantity determination from those pressure differences does not exceed ±0.05%. See
Section 5.1.2 and Section D.1.1, Equation D.1.
For more specific equations and details for each PECI status see Annex F Mass Calculations.
The Pressure Effect Compensation for volume determination is more complicated than for mass. The
level of complication is dependent upon the pressure effect on mass and density determination by the
meter. These effects can vary by meter design.
In volume determination the location for correction and PECI are limited to two choices.
This Indicator status is required for volume applications. A PECI value shall be recorded to the Quantity
Transaction Record(s) (QTR), measurement tickets, and proving reports.
API MPMS Ch. 9.4 is the appropriate standard to reference for stand-alone density measurement.
Pressure effect compensation for tube stiffness is applicable to density measurements of a Coriolis meter.
Operators should consult API MPMS Ch. 9.4 and manufacturers on the applicability and details of this
correction for each specific meter design.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 41
Annex E
(Informative)
Volume quantity determination and volume proving information are found various Chapters of the MPMS.
The intent of this annex is to provide a guide to supplement the other chapters for Coriolis meters used
to determine volume flow. Specific information related to Coriolis meters that might not be addressed in
the published standards are needed on tickets and proving reports to properly document or recreate
QTR.
Note: In general, the application of the pressure correction for tube stiffness, for both mass and density,
should be applied as a matter of course, irrespective of measurement mode.
E. 3.1 Transmitter.
When corrections are performed in the transmitter the status of the related values are:
PECI status is set to “1” and logged in the tertiary device.
A CFPS value of 1.000 is then used in tertiary devices and/or off-site measurement systems.
PCal for the calculations of both the mass and density corrections are provided during the factory
calibration of the sensor.
In addition to the information requirements for proving reports found in API MPMS 12.2.3 and 4.8 include
the following information for volume proving reports on Coriolis meters:
Annex F
(Normative)
Note to Voters: Quantity determinations are found in Chapter 12.2 of the MPMS. As of publication of
this document Chapter 12 Section 2 does not address the mass quantity calculations. The intent of this
annex is to provide a standard reference for mass calculations as related to Coriolis meters.
Note: This appendix is intended to be normative until a publication of Chapter 12 addresses mass
calculations. Then Chapter 12 will take precedence and Appendix F will be informative
F. 1 Mass Determination
This Annex prescribes guidelines for the determination of indicated mass, and mass. Strict adherence to
calculation sequence and application of corrections factors shall be followed. Chain calculations should
be performed without rounding or truncation.
Equations throughout this Annex will be presented in two forms, first in a dynamic form used to
continuously integrate quantities in electronic measurement devices over the Quantity Transaction
Record (QTR) or batch period. The second is the traditional static form of the equation.
Equations F.1 through F.3 are used for integrating indicated mass quantities in tertiary devices and/or
measurement systems, where the internal indicated mass in the tertiary device will accumulate at exactly
the same rate as in the Coriolis transmitter accumulators."
Equation F.1 defines how a meter’s indicated mass (IM) is continuously integrated over time in a tertiary
device using a frequency output (pulses) from a Coriolis transmitter.
𝐼𝑀 𝛥𝑁 ∗𝑞 (F.1)
Indicated Mass (IM) determined by snap shot readings (stop and start), external of transmitter using a
pulse counter. Typically, this equation is not used in real time.
𝑁 𝑁
𝐼𝑀 (F.2)
𝑁𝐾𝐹
Indicated Mass (IM) determined by snap shot reading, external from transmitter using a communication
link or manual reading of an accumulators or totalizers. Typically, this equation is not used in real time.
𝐼𝑀 𝐼𝑀
𝐼𝑀 (F.3)
1.000
A Pressure Effect Compensation Factor (𝐶𝐹𝑃𝑆 ) is used to compensate for the pressure effect on the
flow tube(s). This factor is not the same as CPL for compressibility of the liquid.
1
𝐶𝐹𝑃𝑆 (F.4)
𝑃𝐸
1 ∗ 𝑃𝑓 𝑃
100
The pressure effect correction algorithm of some Coriolis designs might not conform to the above
generalized equation; manufacturers should be consulted on the applicability of this algorithm for the
specific design and method of implementation if different.
This standard prescribes four methods for applying pressure compensation. The methods are identified
by the Pressure Effect Compensation Indicator (PECI) status listed below:
The status should be set in the tertiary device to indicate where and how 𝐶𝐹𝑃𝑆 is being applied. The
PECI shall be recorded to the Quantity Transaction Record(s) (QTR), measurement tickets, and
proving tickets.
Pressure values for Equation F.4 are based upon the following:
The PECI may be set to “4” if the operating pressure is maintained at or near proving pressure so that
any bias in the quantity determination from those pressure differences does not exceed 0.05%. See
Annex D.
Mass (𝑀) is calculated by multiplying the Indicated Mass (𝐼𝑀), with a Mass Meter Factor (𝑀𝐹 ) and mass
pressure compensation factor (𝐶𝐹𝑃𝑆 ). Since this standard disallows meter factor implementation in the
transmitter the measured mass determination can only be calculated in a tertiary device or off-site
measurement system.
It is critical when determining mass that the user understand where and how the pressure effect
compensation is being applied to prevent its omission or correcting for it twice. PECI in Section F.3
describes the status of the pressure compensation.
The following sections describe the requirements of implementing pressure effect compensation (𝐶𝐹𝑃𝑆)
at various auditing locations to prevent its omission or double correction.
F 4.1.1 Transmitter
1) PECI status shall be set to “1” and logged in the tertiary device
2) A CFPS value of 1.000 shall be used in tertiary devices and off-site measurement systems
3) 𝑃 shall equal the factory calibration pressure of the sensor
4) The pressure measurement (𝑃 ) and mass meter factor (𝑀𝐹 shall be flow-weight averaged over
the QTR and/or accounting period and logged in the tertiary device.
1
𝐶𝐹𝑃𝑆 (F.6)
𝑃𝐸
1 ∗ 𝑃 𝑃
100
(F.7)
𝑀 𝐼𝑀 ∗ 𝑀𝐹 ∗ 𝐶𝐹𝑃𝑆
Where: 𝐶𝐹𝑃𝑆 = Flow pressure effect compensation factor over the 𝑖th sampling interval
𝑃𝐸 = Flow pressure effect in percent (%) of rate per unit of gauge pressure
𝑃 = Flowing static gauge pressure or flow weighted average pressure at the meter over
the 𝑖th sampling interval
𝑃 = Calibration or meter proving average static pressure (psig)
𝑀 = Mass
𝐼𝑀 = Indicated mass over the 𝑖th sampling interval
𝑀𝐹 = Mass meter factor over the 𝑖th sampling interval.
𝑖 = ith sampling interval
𝑛 = Number of sampling intervals over the QTR or batch time period
1
𝐶𝐹𝑃𝑆 (F.8)
𝑃𝐸
1 ∗ 𝑃 𝑃
100
𝑀 𝐼𝑀 𝐼𝑀 ∗ 𝑀𝐹 ∗ 𝐶𝐹𝑃𝑆 (F.9)
𝑃 = Flow weighted average calibration or meter proving average static pressure (psig)
𝑀 = Mass
𝐼𝑀 = Closing indicated mass
𝐼𝑀 = Opening indicated mass
𝑀𝐹 = Flow weighted mass meter factor
𝐶𝐹𝑃𝑆 = Flow weighted average flow pressure effect compensation factor
Where: 𝑀 = Mass
𝐼𝑀 = Indicated mass over the 𝑖th sampling interval
𝑀𝐹 = Mass meter factor over the 𝑖th sampling interval.
𝐶𝐹𝑃𝑆 = Flow weighted average flow pressure effect compensation factor
𝑖 = ith sampling interval
𝑛 = Number of sampling intervals over the QTR or batch time period
F5 Mass Proving
There are two methods of mass proving of a Coriolis meter. These methods are inferred mass by means
of volumetric (displacement or tank) prover and direct mass by means of Coriolis master meter.
Only the “Average Meter Factor Method” shall be used for meter factor determination for both Inferred
Mass and direct mass. See API MPMS Ch.12 for “Average Meter Factor Method” details.
The generalized equations below illustrate these proving methods and API MPMS Ch. 4 provides
operational guidelines.
The following equations illustrate the determination of an intermediate meter factor (IMF) for inferred
mass, direct mass and final meter factor.
(F.11)
𝐼𝑛𝑓𝑒𝑟𝑟𝑒𝑑 𝑀𝑎𝑠𝑠 𝐷𝑒𝑟𝑖𝑣𝑒𝑑 𝑓𝑟𝑜𝑚 𝐷𝑖𝑠𝑝𝑙𝑎𝑐𝑒𝑚𝑒𝑛𝑡 𝑜𝑟 𝑇𝑎𝑛𝑘 𝑃𝑟𝑜𝑣𝑒𝑟
𝐼𝑀𝐹
𝐼𝑛𝑑𝑖𝑐𝑎𝑡𝑒𝑑 𝑀𝑎𝑠𝑠 𝑓𝑟𝑜𝑚 𝐿𝑖𝑛𝑒 𝑀𝑒𝑡𝑒𝑟
(F.13)
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 48
∑ 𝐼𝑀𝐹
𝑀𝐹
𝑛
𝐾 = Number of:
Passes per run for unidirectional
Round trips per run for bidirectional
𝐶𝑇𝑆 = BPV correction factor for the effect of temperature on steel
𝐶𝑃𝑆 = BPV correction factor for the effects of pressure on steel
𝑃𝐼𝑀 = Inferred Mass of the liquid in the prover
𝑅𝐻𝑂 = Prover density at measured temperature and pressure
𝐾 = Units conversion to match the mass units output by the meter under test
Additional corrections in the 𝑃𝑟𝑜𝑣𝑒𝑟 𝐼𝑛𝑓𝑒𝑟𝑟𝑒𝑑 𝑀𝑎𝑠𝑠 determination are required if a pressured tank
prover is used. A correction factor for the effects of pressure on steel along with vapor space volume,
and density should be determined and included in the 𝑃𝑟𝑜𝑣𝑒𝑟 𝐼𝑛𝑓𝑒𝑟𝑟𝑒𝑑 𝑀𝑎𝑠𝑠.
1. When pressure compensation effect is not applied in the Coriolis transmitter (PECI status is 2 or
3), no CFPS correction to the line meter pulses 𝑁 𝑜𝑟 𝑁𝑖 or the register readings shall be made
in a tertiary device or off-site measurement system that will be used for the proving calculations.
2. The average proving pressure becomes the new Pcal for subsequent CFPS calculations when
PECI status is 2 or 3.
Dynamic
∑ 𝑁 𝑜𝑟 𝑁𝑖
𝐼𝑛𝑑𝑖𝑐𝑎𝑡𝑒𝑑 𝑀𝑎𝑠𝑠 𝐿𝑖𝑛𝑒 𝑀𝑒𝑡𝑒𝑟 (F.20)
𝑁𝐾𝐹
Discrete 𝑁 𝑜𝑟 𝑁𝑖
𝐼𝑛𝑑𝑖𝑐𝑎𝑡𝑒𝑑 𝑀𝑎𝑠𝑠 𝐿𝑖𝑛𝑒 𝑀𝑒𝑡𝑒𝑟 (F.21)
𝑁𝐾𝐹
The “Adjusted Tank Volume” 𝐵𝑃𝑉 shall be substituted in equations F.24 and F.25 if a pressured tank
prover is used, see Annex F 5.1, equations F.16, F.17, and F.18 for details.
F 5.2 Intermediate Meter Factor (IMF) Determination by Direct Mass Proving by Coriolis Meter
Direct mass proving is the method in which the proving of a line meter is performed by means of a mass
meter (Coriolis Meter) used as a master meter prover.
General Requirements:
1. Direct mass proving reports shall document PECI status and proving pressure.
2. PECI status for mass master meters shall be PECI 1 or 2. Master meter pressure effect
compensation is performed in the IMF equations for PECI 2 and in the transmitter for PECI 1. PECI
3, or 4 is not allowed on the master meter prover.
3. For the line meter all PECI status are valid and shall be documented on prove report.
𝐼𝑀 𝐼𝑀 ∗ 𝐶𝐹𝑃𝑆 ∗ 𝑀𝑀𝐹
𝐼𝑀𝐹 (F.27)
𝐼𝑀 𝐼𝑀
F 5.3 Meter Factor Determination for Inferred Mass and Direct Mass Proving
When the intermediate meter factors (𝐼𝑀𝐹 ) meet acceptable uncertainty criteria, a final average meter
factor (𝑀𝐹 ) shall be calculated as follows:
∑ 𝐼𝑀𝐹
𝑀𝐹 (F.28)
𝑛
Reference API MPMS Chapters 4, and 13 for intermediate meter factor evaluation and acceptability
criteria.
The following is a list of the information necessary to document and determine the transaction quantity
(QTR).
PECI Status [Include the following when the PECI status is 2 (TD) or 3(OS)]
𝑃
𝑃 (Flow weighted average pressure at the meter)
𝑃
Mass
This section provides the necessary information to be contained in mass proving reports. API MPMS
Ch. 12 should be referenced for additional information or clarification of terms.
The following is a list of the minimum measurement information necessary to document an inferred
mass prove.
General Data:
o Customer
o Operator
o Location
o Meter ID
o Date
o Time
Meter Data:
o Meter Type
o Meter Manufacturer
o Flange Size
o Model No.
o Sensor Serial Number
o Transmitter Serial Number
o Flow Calibration Factor(s)
o Nominal K-Factor
o Non-Resettable Totalizer Value
o Zero Value, as found (in metered mass units
o Date of last meter zero
o Zero performed before this prove (Y/N)
o Zero Value, as left
o PECI Status
Prover data:
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 54
o Manufacturer
o Type
o Serial Number
o Base Prover Volume
o Wall Thickness
o Internal Diameter
o Flow Tube Material
o Detector bar material
o Thermal Expansion Coefficients
o Calibration date
Product Data:
o Name
o Prover flowing density (RHOtpp) in meter mass units / prover volume units, (i.e. lbs/Bbl)
o Flow Density Determination Method
When the density is determined from a spot-sample or spot-measurement, then record the following
information:
o Base Density
o Equilibrium Vapor Pressure
o Observed Density value
Temperature
Pressure
o Equation of state
Prove Data:
Meter
o temperature (optional)
o pressure
Intermediate Meter Factor (IMF)
CPSp
BPV
Inferred Prover Mass
Flow rate
Meter:
Pulses (N (avg))
Average pressure of prove runs (Pcal)
Indicated mass (meter)
o Average Meter factor (average of the intermediate meter factors).
o Uncertainty of runs
o Repeatability of runs (optional method)
Optional Meter Factor Comparison Data:
o Technician Company
o Witness Company
The following is a list of the minimum measurement information necessary to document a direct mass
prove.
General Data:
o Customer
o Operator
o Location
o Meter ID
o Date
o Time
Meter Data:
o Meter Type
o Meter Manufacturer
o Flange Size
o Model No.
o Sensor Serial Number
o Transmitter Serial Number
o Flow Calibration Factor(s)
o Nominal K-Factor
o Non-Resettable Totalizer Value
o Zero Value “As found” (in metered mass units)
o Date of last meter zero
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 56
Prove Data:
o Per run
Total Meter Pulses (N or Ni)
Total Prover Pulses per run (N or Ni)
Prover
o Temperature (Tp)
o Pressure (Pp)
o Flow rate
Prover Master Meter Factor (MMF)
Master Meter Mass
Meter mass
Intermediate Meter Factor (IMF)
Meter:
Pulses (N (avg))
Indicated mass (IMm)
o Technician Company
o Witness Company
F7 Discrimination
Irrespective of how the observed values are obtained and their level of discrimination, the values used
in mass calculations shall be to the discrimination levels in the tables below.
US Customary SI Units
lb/gal* lb/bbl kg/l kg/m3*
Density (Absolute, A.xxxx ABC.xx ABC.xx ABC.xx
flowing)
*Density discriminations are per GPA 2145-16
Table F-1 Discrimination Levels for Density Values used in Mass Calculations
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 58
Correction Factor
CFPS CFPS
X.xxxxx X.xxxxx
All other discriminations are in accordance with the appropriate API standard
Table F-2 Discrimination Levels for Mass Values used in Mass Calculations
F.8 Rounding
Annex G
(Informative)
Determination of Net Mass
The non-merchantable sediment and water mass (SWM) quantity can be determined by taking the
difference between the M and NM such that:
%𝑆&𝑊
𝐶𝑆𝑊 1.0 (G.1)
100
𝑁𝑀 𝑀 ∗ 𝐶𝑆𝑊 (G.2)
SWM M NM (G.3)
Where:
CSWM = Correction of Sediment & Water by Mass
%S&WM= Percent of Sediment & Water by (M) Mass,
SWM = Mass of sediment and water during the measurement period
NM = Net mass during the measurement period
M = mass of fluid measured during the measurement period
Methods for determination of sediment and water (S&W) are presented in API MPMS Ch. 10.
G.3.1 Percent Sediment and Water for Mass quantities by Mass test method.
API MPMS Ch. 10.7 (ASTM D4377) and 10.9 (ASTM D4928) both describe determination of water by
mass, the limits for these methods are 2.0% and 5.0% respectively. API 10.1 (ASTM D473) and 10.8
(ASTM D4807) both describe determination of sediment by mass, the limits for these methods are
0.40% and 0.15% respectively.
Per API TR 2573 (ASTM D7829), “To obtain the total sediment and water results when the two are
determined by separate methods: First, add the raw, unrounded results together, and then round the sum
to the correct significant figure.” The resulting is the sediment and water percent by mass (%S&WM).
G.3.2 Percent Sediment and Water for Mass quantities by volume test method.
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 60
Determining sediment and water percent by centrifuge is a commonly used lab and field procedure. The
API MPMS Ch. 10.3 and API MPMS Ch. 10.4 standards describe the determination of sediment and
water by volume. The following equation from API MPMS Ch. 10.9 (ASTM D4928) converts percent
sediment and water (%S&Wv) by Volume to a percent by Mass (%S&WM).
𝐷𝑒𝑛𝑠𝑖𝑡𝑦 𝑜𝑓 𝑊𝑎𝑡𝑒𝑟
%𝑆&𝑊 %𝑆&𝑊 ∗ (G.4)
𝐷𝑒𝑛𝑠𝑖𝑡𝑦 𝑜𝑓 𝑂𝑖𝑙 𝐸𝑚𝑢𝑙𝑠𝑖𝑜𝑛
Note: In accordance with API MPMS Ch. 10.9 (ASTM D4928) a density of water 999.016 kg/m3 will be
used unless otherwise determined by empirical or analytical methods.
Where:
%S&Wv = Volume Percent sediment and water determined by volume test method
SECTION 6—MEASUREMENT OF LIQUID HYDROCARBONS BY CORIOLIS METERS 61
Bibliography