Beruflich Dokumente
Kultur Dokumente
Instructions
Dear users: Thank you for purchasing the BF Series Automatic Hematology Analyzer (Model: BF-6880).
Please read the User Manual thoroughly prior to operation so as to properly use the instrument. Improper
operation may diminish accuracy and precision of measurement or pose a danger to your personal safety.
Warning
● Please dispose of reagents, waste solution, waste samples and consumables according to the national or local
regulations.
● Please dispose of waste solution and instrument consumables according to the regulations about medical
wastes, infective wastes and industrial wastes. Blood in wastes may have been contaminated by pathogens.
● Avoid contact with the puncture probe, because the blood, Control and Calibrator on the puncture probe
have potential bio-infectivity. When the puncture probe is aspirating a sample, there should be a certain
distance between the probe tip and the wall of the container to avoid splashing blood. Or the accuracy of
aspiration may be affected.
Caution
● The instrument should be operated by professional medical inspection specialists, or physicians, nurses or
lab assistants who are specially trained.
● The hospital or inspection bodies using the instrument should compile maintenance plans, and carry out
maintenance according to the plans. Otherwise, instrument malfunction may occur.
● The instrument must be controlled by the dedicated software. Install the software specified by our company.
The installation of other software and hardware may affect the normal operation of the system. Do not run
other software when the instrument is operating.
● For the use and storage of the reagents, refer to the reagent instructions.
● Expired reagent cannot be used. The reagent should be protected from dusts, dirt and bacteria once opened.
● Pine oil and benzene cannot be used to clean the outside of the instrument; otherwise it may cause color and
shape change. Soft cloth or gauze can be used for cleaning. Diluted detergent and alcohol can be used if
necessary.
● When the transportation or storage temperature is low or the relative humidity is more than 75%. The
instrument should be kept in the normal work environment for at least 24 hours before power on.
● An independent power supply is a must. If the instrument and other electrical instruments share one power
outlet, electromagnetism interference will affect the accuracy of the test results.
● Do not pull out the electrical wire with wet hands, or there is a risk of electrical shock.
● Do not trample, twist, or drag the wire and cable, or it may cause a fire. The damaged wire and cable cannot
be used.
● The input power should conform to the instrument requirement. Use specified fuses.
● Make sure the switch is in the [O] state before connecting the power cable.
● Do not touch moving parts when the instrument is testing to avoid accidents.
● Non-professionals cannot open the left, right and upper cover of the instrument when the main power is ON.
● Make sure the instrument is used under the specified conditions in the user manual. In improper conditions,
the instrument may not work well, the results may be inaccurate, instrument components may be damaged and
personal security is endangered.
● The protective measure may not be effective if the instrument is not operated according to the user manual.
Statement
The manufacturer is responsible for the security, reliability and performance of the instrument after the
following requirements are met:
(1)Installation, debugging and repair are conducted by professionals from the manufacturer.
Contents
I
User Manual
II
User Manual
III
User Manual
IV
User Manual
1.1 Introduction
BF Series Automatic Hematology Analyzer (Model: BF-6880) is a highly integrated instrument with
high-capability. It characterize in accurate test result, easy operation, low consumable. The instrument can test
quantitative analysis result of 25 blood parameters. Instrument is connected with computer to conduct
operations.
The scope of the product: BF Series Automatic Hematology Analyzer is an in vitro diagnostic medical
instrument used by professionals for screening. It is used for testing red blood cell (RBC), platelet (PLT),
white blood cell (WBC) number and volume distribution and concentration of hemoglobin, and it offers
scattergram of white blood cell to help clinical diagnosis.
1.3 Specifications
Model
Specifications
BF-6880
WBC, BAS#, NEU#, EOS#, LYM#, MON#, BAS%, NEU%, EOS%,
Test Item LYM%, MON%, RBC, HGB, MCV, MCH, MCHC, RDW-CV,
RDW-SD, HCT, PLT, MPV, PDW, PCT, P-LCR, P-LCC
Carry out five-part differential to WBC counting result; the instrument
Parameters can test 25 parameters, offer 1 scattergram and 4 histograms; it will alarm
Basic when the abnormal condition of pathology and morphology occurs
Features Throughput 80T/H
Sample Feeding Method Manual or automatic sample feeding.
Storage 100000 results (maximum)
Sample Barcode Automatic identification or manually input.
Reagent Type Four types in total, including diluent and 3 types of Lyses
Reagent Hemoglobin Test Test hemoglobin utilizing the reagent without cyanide
System Alarm Alarms when there is no reagent or the reagent expires.
Reagent Barcode External barcode reader or manual input.
WBC Classification Cell chemical staining technology, sheath flow technology, optical
Analysis Principle analysis method
System Utilizes impedance technology for WBC, RBC, PLT counting;
Counting Method
colorimetric method for HGB counting.
Data Interface RJ45
System Connect to LIS/HIS Can be connected to LIS/HIS
Weight 60kg
Whole
instrument Dimensions 606mm×690mm×550mm(L×W×H)
system
Power 175VA
1-1
User Manual
1-2
User Manual
1-3
User Manual
1-4
User Manual
1-5
User Manual
1.4.4.8 Lymphocytes
LYM # = WBC × LYM%
1.4.4.9 Neutrophil
NEU # = WBC × NEU%
1-6
User Manual
1.4.4.10 Monocyte
MON # = WBC × MON%
1.4.4.11 Eosinophil
EOS# = WBC × EOS%
1-7
User Manual
HGB
MCH =
Mean hemoglobin content RBC
HGB
MCHC = ×100
Mean hemoglobin concentration HCT
RBC Unit: 1012/L, MCV Unit: fL, HGB Unit: g/L
(4)RBC Distribution Width Variation Coefficient
RDW-CV is obtained through RBC distribution histogram. The volume distribution variation coefficient is in
the form of percentage.
(5)RBC Distribution Width Standard Deviation
RDW-SD is histogram width relative to RBC distribution histogram peak (20%), unit fL, as shown in Figure
1-4-10.
Figure 1-4-10
(6)RBC Distribution Histogram
The RBC volume distribution histogram is offered when the result is obtained. The graph that can indicate the
distribution of cell population is RBC distribution histogram. Histogram abscissa is RBC size (unit: fL),
vertical axis is RBC relative number (unit: 1012/L). After each counting, RBC distribution histogram can be
obtained in analysis result area of counting interface. RBC distribution histogram can also be obtained through
entering search interface.
1.4.6.4 PLT Parameter
(1)PLT Number (PLT)
PLT number is obtained through testing corresponding electrical pulse number, unit 109/L
PLT = n × 109 / L
(2)Mean PLT Volume (MPV)
Calculate mean PLT volume according to PLT distribution histogram. Unit: fL
(3)PLT Distribution Width (PDW)
PDW is obtained through PLT distribution histogram, which is geometric deviation limit of PLT volume
(10GSD)
(4)PCT
Calculate PCT according to the following formula, unit%; PLT unit 109/L unit ; MPV unit fL
PLT ´ MPV
PCT =
10000
(5)PLT Ratio(P-LCR)
PLT Ratio is obtained through PLT histogram.
(6)PLT Distribution Histogram.
The PLT volume distribution histogram is offered when the result is obtained. The graph that can indicate the
distribution of cell population is PLT distribution histogram. Histogram abscissa is PLT volume (unit: fL),
vertical axis is PLT relative number (unit: 109/L). After each counting, PLT distribution histogram can be
obtained in analysis result area of counting interface. PLT distribution histogram can also be obtained through
entering search interface.
1-8
User Manual
1.4.7 Rinsing
The instrument rinse itself automatically in each counting process, to ensure no residual sample exists.
(1)Internal and external wall of puncture probe should be rinsed with diluent;
(2)Counting pool should be rinsed with diluent;
(3)Flow cell should be rinsed with diluent.
1.5 Structure
BF Series Automatic Hematology Analyzer is composed of mechanical motion, pipeline, optical, electronic
control, software, etc.
Mechanical motion system: it mainly contains automatic sample injection unit and automatic sample aspiration
unit;
Pipeline system: it mainly contains liquid running pipeline, syringe unit and pump unit;
Optical system: it mainly contains laser shaping, light split and receiving unit;
Electronic control system: it contains main board, drive board and monitoring and control unit;
Software system: it contains embedded software and upper machine software.
The functional interface of front, rear and right of the instrument are as follows:
1.5.1 Front Picture
1 Front Door 2 Indicator (yellow: alarm indicator, green: running status indicator, red: power indicator from left to right)
3 STAT Key 4 Sample Loader 5 Emergency Rack 6 Tube Rack 7 Mode Key 8 Count Key
Figure 1-5-1 Front Picture
1-9
User Manual
1 Upper Cover 2 Right Door Assembly 3 Fan 4 Back Cover 5 Fan 6 Fan 7 Waste Liquid Inlet
8 Diluent Inlet 9 FDT Lyse Inlet 10 FDO Lyse Inlet 11 SLS-I Lyse Inlet 12 Detergent
13 Left Door Assembly
Figure 1-5-2 Rear Picture
1-10
User Manual
1.7 Symbol
Symbol Meaning
BIOLOGICAL RISKS
ALTERNATING CURRENT
BATCH CODE
USE BY
SERIAL NUMBER
DATE OF MANUFACTURE
PROTECTIVE GROUND
MANUFACTURER
CATALOGUE NUMBER
“ON”(POWER)
“OFF”(POWER)
The above symbols are included on the instrument, reagents, Controls and Calibrators.
1-11
User Manual
1-12
User Manual
Chapter 2 Installation
● To ensure the normal working of the instrument after installation, the initial installation of the
instrument should be carried out by authorized professionals of the manufacturer.
● Dedicated computer software should be used for controlling. It is recommended that the software and
database should not be installed in the system disk.
The result will be unreliable if the room temperature or power cannot meet the requirements. Or cause
instrument damage and endanger personal safety.
2.2 Unpacking
2.2.1Unpacking Procedures
After receiving the instrument, check whether there is physical damage onto the packaging case. Contact the
2-1
User Manual
manufacturer or the local distributor if any damage is found. Open the case if no external damage is found
according to the following steps:
(1)Place the package in the direction of the arrow upward.
(2)Open the instrument and accessory case, and check the items according to the packing list. Contact The
manufacturer or the local distributor if anything is missing.
2.2.2 Transportation Method
(1)Remove the reagent bottle and waste tank at the back of the analyzer prior to transportation.
(2)Ensure that the used Analyzer has gone through “Empty Pipeline” before transportation.
(3)Utility cart and the like are available for smooth and short-distance transportation.
(4)When moving and transporting, prevent the display on the front panel and the puncture probe from being
applied with external force, contacting other goods and damaging.
(5)When moving and transporting, always keep the Analyzer upright. Inclination and side laying are not
allowed.
(6)During transportation, avoid vibration as far as possible. After transport, examine and debug the Analyzer
before usage.
1 Detergent inlet 2 SLS-I Lyse inlet 3 FDO Lyse inlet 4 FDT Lyse inlet
5 Diluent inlet 6 Waste liquid inlet
Figure 2-3-1 Instrument and Reagent Connection
(1)Connection of Detergent, Lyse, Diluent, and Waste Solution
Put the bottles of FDT Lyse, FDO Lyse, SLS-I Lyse and Detergent at the back of the instrument. Connect
them according to Figure 2-3-1. Put the diluents barrel and waste liquid barrel under the working table.
Connect according to Figure 2-3-1.
(2)Connection of Liquid Level Sensor
Connect the other end of the accompanying liquid level sensor to the sensor port on the right side plate of the
instrument. Connect the liquid level sensor to the corresponding reagent bottle, diluent and waste liquid barrel,
according to the marks on the lead.
2-2
User Manual
● The water discharge system should conform to the waste water disposal requirements for local
medical institutions.
(3)Computer Connection
Connect the “Net Port” of the computer to “RJ45” on the right side plate of the instrument.
(4)Power Cable Connection
Connect one end of the accompanying power cable to the plug on the right side of the instrument. And connect
the power cable of the computer main case, display and printer.
The light beam of the barcode reader may hurt eyes; therefore, do not stare at it.
(6)Printer Connection
Connect the printer and the computer through a data cable, and check
a)whether the printer driver is installed.
b)the specifications of the printing paper.
During transportation, to avoid damaging puncture probes and mechanical arms, use screws, fixing
frames and nylon ties to fix the moving parts when leaving the factory. The screws, fixing frames, and
nylon ties must be taken down before power on, or it may cause damage to puncture probes and
mechanical arms. The positions for screws, frames and nylon ties are shown below:
Screw 1
Figure 2-3-2
2-3
User Manual
Fixing frame
Screw 2, 3
Figure 2-3-3
Screw 4
Figure 2-3-4
Tie
Figure 2-3-5
2-4
User Manual
Screw 1, 2, 3 and 4, the frame, and the nylon tie of the mechanical arm must be removed before power on.
Otherwise, it may cause damage to the puncture probe and mechanical arm.
2.3.2 Software Installation
The software has been installed before the instrument leaves the factory. The user should not uninstall the
software except when abnormity occurs. In case of a necessary un-installation, please follow the following
steps.
2.3.2.1 Operating System: Windows 7
Start “SetDoNet” after putting the software CD in the CD driver of the computer. “NET Framework4.0”
installation assembly will pop up, as Figure 2-3-6 shows:
Figure 2-3-6
Click “Install” in the above figure, and the following figure will pop up:
Figure 2-3-7
Click “OK” in the above figure, after restart the Microsoft SQL Server 2005 can AutoRun Installation, as
Figure 2-3-8 shows:
2-5
User Manual
Figure 2-3-8
Click “Next” in the following pages until the page – Figure 2-3-9 shows:
Figure 2-3-9
Continue to click “Next” until the installation completes. After install SQL2005, it will auto-run the
installation of VC++2008, as Figure 2-3-10 shows:
2-6
User Manual
Figure 2-3-10
VC++ installation completes, as Figure 2-3-11 shows:
Figure 2-3-11
At last, appearing the following tip illustrates the PC system environment is installed successfully. Then you
can install the upper computer software, as Figure 2-3-12 shows:
Figure 2-3-12
2-7
User Manual
When the above progress is finished, run upper computer installation program, the setup wizard will pop up, as
Figure 2-3-13 shows:
Figure 2-3-13
Click “Next” in the above figure, and select the installation path. The default path is C:\Program Files\, (It is
suggested that the program should not be installed in C:\, it can be installed in other disks.), meanwhile, all the
users have the permission to install the program by default. as Figure 2-3-14 shows:
Figure 2-3-14
Click “Next” in the above figure, the “Confirm” interface will pop up, as Figure 2-3-15 shows:
2-8
User Manual
Figure 2-3-15
Click “Next” in the above figure, the “Installing” interface will pop up, as Figure 2-3-16 shows:
Figure 2-3-16
Click “Next” after the progress is finished, and Figure 2-3-17 will pop up:
2-9
User Manual
Figure 2-3-17
Click “Close” in the above figure to finish the installation of the software. The shortcut of the program will be
displayed on the table after the installation.
If the upper computer is installed in the C:\, it is needed to modify the properties of the installation directory. If
the upper computer is installed in the other disks, the properties do not need to be modified. Details are shown
as follows:
Right-click the installation directory, and select properties, as the following figure shows:
Figure 2-3-18
Click “Security” button in the above figure, and the following figure pops up:
2-10
User Manual
Figure 2-3-19
Click “Edit…” button in the above figure, and the following figure pops up:
Figure 2-3-20
Click “Users ...” in the above figure, and add the “Full Control” permission, that is to say, tick “Allow” in the
“Full Control” line under “Permissions for Users”, as the following figure shows:
2-11
User Manual
Figure 2-3-21
Then click “Apply”, “OK” button to save the setting.
2.3.2.2 Set the Computer
(1)IP Setup
Click【Start】→ 【Control Panel】→ 【Network and Sharing Center】, then select 【Change adapter settings】
from the left column, choose 【Local Area Connection】 and click 【Properties】, and then double-click
【Internet Protocol Version 4(TCP/IPv4) 】. Refer to the figure below:
Figure 2-3-22
The undefined number “XXX” in IP Address “172.16.100.XXX” includes, but is not limited to, the number
that the above figure shows. You can input any number between【1~ 225】except 138.
2-12
User Manual
when the computer sleeps】from the left column, and set【Never】 for when to “Put the computer to sleep”.
Refer to the figure below:
Figure 2-3-23
Figure 2-4-1
Turn on the power switch of the instrument first, and then log in the application software.
2.5.1 Login Software
Double-click the instrument application software icon on the desktop, or click “Start”, find the
software from “Program” window and single-click it to enter the “System Login” window, as Figure 2-5-1 and
Figure 2-5-2 show:
2-13
User Manual
Figure 2-5-1
Figure 2-5-2
Input username, password, the initial user name is Admin (cannot be modified), the initial password is 1). If
the input username or password is wrong, the screen will display login error, as Figure 2-5-3 shows:
Figure 2-5-3
Input username, password in Figure 2-5-2, and click “ ” to enter the “Component Maintenance”
interface as shown in figure 2-5-4.
2-14
User Manual
Figure 2-5-4
In the initial login, the checkbox “Automatically turn off after 3 seconds” in the “Component
Maintenance” screen is selected as default, you can cancel selection as required; If cancel, you need to
manually close the window. These operation will take effect when next logon.
When the “Component Maintenance” interface was closed, the screen will display “System self-testing…”.
Then the instrument starts to check the flow path, temperature and background as the figures shown below:
Figure 2-5-5
Figure 2-5-6
Click in Figure 2-5-6 after self-test to enter into main interface, as Figure 2-5-7 shows:
2-15
User Manual
Figure 2-5-7
Public Info. Area
2-16
User Manual
: Printer is disconnected.
: Printer is working.
2.5.2.3 Public Information Area
Bottom of the counting interface is the public information area:
Auto rinse Sample No. Sample Info. Running Status System Time
Figure 2-5-8
(1)Failure Information Area
The corresponding failure information will be displayed in this area when failure occurs. Click this area, the
failure dialog box will pop up, as the following figure shows:
Figure 2-5-9
Click the corresponding information, the detailed solution will be displayed in “Detail Information”
(2)X-B QC Switch Status
Use icon to indicate the switch status of X-B QC. With X-B is ON, without X-B is OFF.
(3)Sample Status: Display the analysis mode and test mode.
Analysis Mode: There are four modes. Auto-Whole Blood, Microscale-Whole Blood (No.2 position) or Close
-Whole Blood (No.3 position), Close-Prediluted (No.4 position).
Test Mode: There are two modes. “CBC” and “CBC+DIFF”.
(4)Test Status: Indicates the test status.
(5)Rinse Times: 22 in 22/200 indicating the tested sample number from the last time rinsing. 200 represents
the rinsing interval (set in Auto clean Settings).
2-17
User Manual
Click in Figure 2-5-7 to enter the “exit” window, as the following figure shows:
Figure 2-5-10
Click in the above figure to exit. Click to return to the interface continue tests.
● Log off is recommended when the user is at rest. For avoiding non-user damage software or modify
the data. Periodical database backup is recommended to avoid data lose caused by unforeseen
circumstances.
● Input the initial user name and password in the first login. Set the user name, permission and
password in “Setting” after login for next login.
2-18
User Manual
Chapter 3 Setting
3.1 Introduction
The system parameters of the instrument have been set before delivery. The interface of the first power on is
system default. In order to meet the different needs of practical application, two permissions are provided(user
permission and administrator permission), the user can reset some parameters.
Turn on the instrument, log in the software, and enter the count interface. Click in Figure 2-5-7, and
the sample information interface will pop up as the figure shows below:
Figure 3-2-1
Analysis Mode: There are four modes. Microscale-Whole Blood mode and Close-prediluted mode are manual
modes. Auto Whole Blood mode is an automatic analysis mode.
Equalize pressure automatically: It will equalize the pressure in the vacuum tube and the atmospheric pressure.
If “Equalize pressure automatically” is not clicked, there is pressure in the vacuum tube, the aspirated volume
may be not accurate, which may affect the accuracy of the test result.
Test Mode: There are two modes. “CBC” conducts counting without diff upon WBC. The counting results
include RBC, PLT histograms and other parameters. “CBC+DIFF” conducts counting and 5-differential upon
WBC, includes 25 parameters, 1 scattergram, and 4 histograms.
Tube Rack No.: When the auto-whole blood mode is selected and “Identify tube rack automatically” is not
selected, the Rack No. and Tube No. are in the available status, as Figure 3-2-2 shows:
Figure 3-2-2
Input the new rack No. and tube No. in the input box of rack No. and tube No.
Close counting mode: When close-whole blood or close-prediluted is selected, close counting method can be
3-1
User Manual
Figure 3-2-3
Sample No.: The sample number input in this box is the next tested sample. If the entered number repeats a
number of the intraday sample database, the following figure will pop up:
Figure 3-2-4
● Only under the analysis mode of “auto whole blood” and “close-whole blood (No.3 position)” have the
function of auto balance pressure.
● When entering a sample number, “-” can be entered. “.” cannot be entered. (“Sample No. Error” will
be prompted if there is “.”).
● The entered number of the sample that is to be tested cannot repeat the number of an already tested
or edited sample within the day.
● Reference Range: Click the pull-down menu of “Reference”, general, man, woman, child, newborn
and user-defined can be selected.
Figure 3-3-1
3-2
User Manual
Figure 3-3-2
(1)Default Value: This value is default when editing other sample information.
(2)Remember: If , the last edited sample information will be remembered when editing other
information.
3-3
User Manual
Figure 3-3-3
Input the corresponding mnemonic in the input box of “System Code Selection”, or click the line of the
corresponding item (“Setting” of main menu→ “Information” to set the commonly used ID). Click
.The reference range of test items includes “General”, “Man”, “Woman”, “Child”, “Newborn” and 5
“User Define”s. 10 options in total.
Or click the pull-down menu after “Reference”. Make selection from the popping-up figure:
Figure 3-3-4
(4)Sampling Time; Delivery Time: Select “Current Time” or “Blank” from the drop-down list of “Sample
Time” and “Deliver Time” in Figure 3-3-2. If “Current Time” is selected, click the corresponding time; manual
input is also available.
3.3.1.2 Patient Information
(1)Case No.: Input the case number in Figure 3-3-2.
(2)Name: Input the patient name in Figure 3-3-2
(3)Sex: Double-click the input box behind “Sex” in Figure 3-2-2, the following box will pop up:
3-4
User Manual
Figure 3-3-5
Input the corresponding mnemonic in the input box of “System Code Selection”, or click the line of the
corresponding item(“Setting” of main menu→ “Information” to set the commonly used ID). Click .
Or click the pull-down menu behind “Sex”, and make selection from the popping-up figure:
Figure 3-3-6
(4)Age: Input the patient age and double-click the input box of the age unit. The following box will pop up:
Figure 3-3-7
Input the corresponding mnemonic in the input box of “System Code Selection”, or click the line of the
corresponding item (“Setting” of main menu→ “Information” to set the commonly used ID). Click .
Or click the pull-down menu behind “Age Unit”, and make selection from the popping-up figure:
3-5
User Manual
Figure 3-3-8
The reference value cooperates with age and sex. After selecting sex, inputting age and selecting age unit, the
reference value will be adjusted automatically. Infant≤28 days, 28 days<child≤13 years old, 13 years old<
general. If the input age is more than 13 and sex is male, the reference value is male by default, if the input age
is more than 13 and sex is female, the reference value is female by default,
Only round numbers are allowed to be entered for the age. “.” cannot be centered.
(5)Cost Category: Enter the name of a cost category in the “Cost Category” box. Double-click the “Cost
Category” input box, and the following figure pops up:
Figure 3-3-9
Input the corresponding mnemonic in the input box of “System Code Selection”, or click the line of the
corresponding item (“Setting” of main menu→ “Information” to set the commonly used ID). Click .
Or click the pull-down menu behind “Cost Category”, and make selection from the popping-up figure:
Figure 3-3-10
(6)Department: Input the department name in the department box. Double-click “Department” input box, and
the following box will pop up:
3-6
User Manual
Figure 3-3-11
Input the corresponding mnemonic in the input box of “System Code Selection”, or click the line of
corresponding item (“Setting” of main menu→ “Information” to set the commonly used ID).Click .
Or click the pull-down menu behind “Dept.”, and make selection from the popping-up figure:
Figure 3-3-12
(7)Bed No.: Click the input box of “Bed No.” in Figure 3-2-2 and enter a bed number directly.
(8)Delivery Doctor: Input the name of a delivery doctor directly or double-click the input box, and the
following figure will pop up:
Figure 3-3-13
Input the corresponding mnemonic in the input box of “System Code Selection”, or click the line of
corresponding item (“Setting” of main menu→ “Information” to set the commonly used ID).Click .
Or click the pull-down menu behind “Dept.”, and make selection from the popping-up figure:
3-7
User Manual
Figure 3-3-14
(9)Remarks: Clinical diagnostic information after tests can be entered (can be input directly).
Click after finishing entering all the above information. When entering information, items for
which “Ignore” has been selected can be skipped. Click or to enter patient
information of other samples.
3.4 Setting
3.4.1 Normal User
Click “Setting” in the main menu to enter the setting interface, as Figure 3-4-1 shows:
Figure 3-4-1
3.4.1.1 Date Format Setting
There are three date formats:
YYYY-MM-DD, MM-DD-YYYY, DD-MM-YYYY, select one of them and press , the following
information will be displayed:
Figure 3-4-2
Press to save the changed date format. Press to exit the interface.
The changed date will be displayed in all positions with time (such as delivery time, sampling time, etc.).
3-8
User Manual
Figure 3-4-3
Click the pull-down menu behind “Select Language”, select the language, click , the tip “Saving
Succeed” will pop up, click to finish the language setting.
3.4.1.3 User Setting
Click “User Settings” in Figure 3-4-3, as Figure 3-4-4 shows:
Figure 3-4-4
Only the password can be modified when the user logs in as a general user. Click in Figure
3-4-4, and Figure 3-4-5 pops up:
3-9
User Manual
Figure 3-4-5
Input a new password and the old password in the above figure (enter the same new password twice), and click
to finish the modification.
User edition, addition and deletion cannot be conducted when the user logins as a general user.
3.4.1.4 Reagent Validity Setting
Single-click “Reagent Validity Setting” in Figure 3-4-4, as Figure 3-4-6 shows:
Figure 3-4-6
Click the pull-down menu of the corresponding item, select the validity according to the reagent instructions,
and press . The tip “Saving Succeed” will pop up, click to finish the setting. Click
to exit the interface.
3-10
User Manual
Figure 3-4-7
Figure 3-4-8
Control method, calculation method and range of L-J/X-bar QC can be selected from Figure 3-4-7.
ON/OFF and sample No./group of X-B QC can be set in Figure 3-4-8.
3.4.1.6 Print Setting
Click “Print Setting” in Figure 3-4-8. Printer name, page header icon, title and format of sample and QC can
be set in Figure 3-4-9.
3-11
User Manual
Figure 3-4-9
(1)Printer: Select a printer from its drop-down list.
(2)Report Default Icon: Report page header icon can be set. Click , and the dialog box of a
picture path will pop up. Select a picture (picture size is 34*34; format can be BMP and JPG).
Click to clear the page header icon. The icon will not be displayed in report preview.
(3)Sample: Headline, page footer and print report format can be input in this unit.
Headline: Delete the original headline and input new headline if headline need to be changed.
Page Footer: The page footer is user-defined, e.g. “This report is only responsible for the delivered sample.”
Print Report Format: Select the format in its drop-down list. Click to preview the format.
(4)QC: The headline and print report format can be input in this unit.
Headline: Delete the original headline and input new headline to change it.
Print Report Format: Select the print format in its pull-down menu, and click to preview the
format.
(5)Report Designer: User can design report format.
Refer to “Appendix B” for use.
(6)Print Research Parameter: when “Print Research Parameter” is ticked, “Input microscopic examination
result” will display in the interface, as following figure shows:
3-12
User Manual
Figure 3-4-10
(7)Print clogged aperture mark: in case that “Print clogged aperture mark” is ticked, if the report data is with
clogged aperture, there is the clogged aperture mark (?) on the printed report; in case that it is not ticked, if the
report data is with clogged aperture, there is no the clogged aperture mark (?) on the printed report.
3.4.1.7 Clogged aperture re-test switch
Click “Clogged aperture re-test switch” in Figure 3-4-10, as the following figure shows:
Figure 3-4-11
3-13
User Manual
Figure 3-4-12
Click , then click at the lower right corner to exit the setting interface.
If is selected, the instrument will re-test automatically when the aperture is clogged in the process of
test. If is selected, the instrument will alarm “The aperture is clogged, remove the clog or rinse the
aperture.” when the aperture is clogged in the process of test.
3.4.1.8 Automatic sleep time
Click “Automatic sleep time” in Figure 3-4-11, the automatic sleep time of the instrument can be set, the range
is [10, 120], as Figure 3-4-13 shows:
Figure 3-4-13
Admin is system default administrator username. Default password is 1. The password can be modified,
but cannot be deleted.
3-14
User Manual
Figure 3-4-14
3.4.2.1 User Setting
Click “User Setting” in Figure 3-4-14, as Figure 3-4-15 shows:
Figure 3-4-15
(1)Add User:
Click in Figure 3-4-15, and the following figure pops up:
Figure 3-4-16
3-15
User Manual
Input the user name needing to be added in the input box behind “Username”. To grant the audit permission to
the new user, put a tick in the “Audit Permission” box. To grant the new user the admin permission, tick the
“Admin Permissions” box. Click to finish the user adding.
Click after username input if this user does not need audit permission or admin permission.
● The initial password of the new user is “1”, which can be modified after login.
● The name of a new user cannot be blank or the same as others.
(2)Delete User:
Click the line of general user except admin in Figure 3-4-15, the selected line will turn blue, the interface is
like 3-4-17:
Figure 3-4-17
Click in the above Figure, and the following prompt will pop up:
Figure 3-4-18
3-16
User Manual
Figure 3-4-19
The audit permission of a user can be edited again. Tick the check box in front of “Audit Permission” to grant
the permission to the user. Deselect the box and the user does not have the audit permission. The administrator
permission of a user can be edited. Select the “Admin Permissions” box to grant the administrator permission.
Deselect the box and the user becomes a general user.
(4)Modify Password:
The password of the administrator and general user can be modified by administrator.
Click the line needs to be modified in Figure 3-4-17, click , the interface is as follows:
Figure 3-4-20
Input the new password twice, and press to finish the operation.
3.4.2.2 Net Setting
Click “Net Setting” in Figure 3-4-17, and select “Device Setting” from the expansion menu, as Figure
3-4-21shows:
Figure 3-4-21
The IP of the instrument is displayed but cannot be modified by the user.
LIS Setting: Select “LIS Setting” from Figure 3-4-21, as Figure 3-4-22 shows:
3-17
User Manual
Figure 3-4-22
IP address and port of LIS computer can be set. ON/OFF status of auto-communication can be selected.
The user can select whether to transmit results (On/Off) if the aperture is clogged.
3.4.2.3 Unit Setting
Click “Unit Setting” in Figure 3-4-22, as Figure 3-4-23 shows:
Figure 3-4-23
Single-click the input boxes behind items and select the units. Click after input. Click
when “Saving Succeed” is prompted. Click “Default Value” to restore the setting when the unit has been
modified accidentally. “Default value setting succeed” indicates successful setting.
The units are as follow:
3-18
User Manual
Table 3-4-1
103/μL ***.**
WBC
2
10 /μL ****.*
/nL ***.**
106/μL **.**
RBC
104/μL ****
/pL **.**
mmol/L **.*
mmol/L ***.*
% **.*
HCT
L/L *.*** Default Unit
/nL ****
3-19
User Manual
Figure 3-4-24
Upper, lower Limit of Reference Input:
Click the input box of upper and lower limit to input the values.
Click after input. Click after the prompting of “Saving succeed” to finish the saving.
The following box will pop up if the input lower limit is more than the upper limit, or the input reference is not
within the set range.
Figure 3-4-25
Input again after checking if the above box is prompted.
Click “Default” to restore the setting when the reference has been modified accidentally. “Default value setting
succeed” indicates successful setting.
Figure 3-4-26
3.4.2.5 Information Setting
(1)Department Information Setting:
3-20
User Manual
Click “Information” in Figure 3-4-24, and select “Department Information” from the drop-down list of
“Information Type”, as Figure 3-4-27 shows:
Figure 3-4-27
a)Add Department Information: Input the department name in the input box behind Name. Input the
commonly used mnemonic in the input box behind ID, click .
b)Delete Department Information: Select the items needing to be deleted from the department information
list, click .
(2)Doctor Information Setting:
Select the “Doctor Information” from the drop-down list of “Information Type”, as Figure 3-4-28 shows:
Figure 3-4-28
a)Add Doctor Information: Input the doctor name in the input box behind name. Input the commonly used
mnemonic in the input box behind ID, click .
b)Delete Doctor Information: Select the items need to be deleted in the doctor information list, click
.
3-21
User Manual
(3)Reference Setting:
Select “Reference” from the pull-down menu of “Information Type”, as Figure 3-4-29 shows:
Figure 3-4-29
Select “General”, “Man”, “Woman”, “Child”, “Newborn”, “User Define 1”, “User Define 1”,
“User Define 2”, “User Define 3”, “User Define 4” and “User Define 5” in reference list. Only the commonly
used ID of reference can be modified here, but not the name or add other reference.
(4)Sex Setting:
Select “Sex” from the pull-down menu of “Information Type”, as Figure 3-4-30shows:
Figure 3-4-30
Select “Male” or “Female” from the sex list. Only the commonly used sex ID can be changed here.
(5)Age Unit Selection:
Select “Age” from the pull-down menu of “Information Type”, as Figure 3-4-31 shows:
3-22
User Manual
Figure 3-4-31
Select “Years Old”, “Month”, “Day” or “Hour” from the list of age unit. Only the commonly used age unit ID
can be changed here.
(6)Cost Category Setting
Select “Cost Category” from the pull-down menu of “Information Type”, as Figure 3-4-32 shows:
Figure 3-4-32
a)Add Cost Category: enter a cost category type in the “Name” input box. Enter a commonly used
mnemonic in the “ID” input box. Click to finish the addition.
b)Delete Cost Category: Select a cost category to be deleted. Click to delete the cost
category.
3.4.2.6 Backup Setting
Select “Backup Setting” from Figure 3-4-31, as Figure 3-4-33shows:
3-23
User Manual
Figure 3-4-33
(1)Backup:
Periodical database backup can prevent data loss.
Single click to backup the document into the folder. The default saving path is “backup”of
software installation. The folder name is the current date+time+*.bak. Click for saving.
The backup function can only back up the contents involved in “Setting”.
(2)Data Recovery:
The backup data can be used to recover the previous data if software cannot be used for some reason. Select
the saving path of backup folder before recovery. And select the document according to the backup date and
time.
3.4.2.7 Abnormal Mark Setting
Click “Abnormal Mark Setting” in Figure 3-4-33, and select “WBC” from the expansion menu, as Figure
3-4-34 shows:
3-24
User Manual
Figure 3-4-34
Select “RBC/PLT”, as Figure 3-4-35 shows:
Figure 3-4-35
The prompt range of WBC, RBC/PLT abnormal alarm information can be set in this interface. Click
for saving after input, click “Default” to recover, click to exit the interface.
3.4.2.8 Automatic Clean Setting
Click “Auto Clean Setting” in Figure 3-4-34, as Figure 3-4-36 shows:
3-25
User Manual
Figure 3-4-36
(1)Input automatic clean interval times in “Count Times” box (within 10—200). The set number in the figure
is 200, which means automatic clean will be conducted after each 200 sample tests. The count will return to 0
after clean.
(2)Input rinse DIFF pool interval times in the input box of rinse DIFF pool “Count Times” (between10-600).
The set number is 10 in the above figure, which means when the test times is no less than 10, DIFF pool
rinsing will be conducted before sleep (The software displays “Rinsing the instrument…”), then the instrument
will enter into sleep mode (in case of not conducting shutdown and manual rinse of the instrument in the
process of test).
(3)Input rinse probe interval times in the input box of rinse probe “Count Times” (between10-100). The set
number is 50 in Figure 3-3-34, which means when the function of auto balance pressure is used for 50 times
successively, the instrument will rinse probe automatically. Only “Enable” is ticked, the function of “rinse
probe” is activated.
3-26
User Manual
Chapter 4 Calibration
4.1 Introduction
Calibration of the instrument is to ensure the accuracy of the results. Calibration must be carried out before
test.
Click “Calibration” of the top menu, and select “Calibrator Calibration” from , as Figure
4-3-1 shows:
Figure 4-3-1
4-1
User Manual
Figure 4-3-2
(2)Click in the above interface. The close sample loader will be available, rotate the loader to make
position 3 aligned with the testing position (Click indicates right positioning), and place the prepared calibrator
in position 3, as Figure 4-3-3 shows:
Figure 4-3-3
(3)Press the “Count” key of the instrument. The following box will pop up when pushing back the loader.
Figure 4-3-4
(4)Start Sample Aspiration→Aspiration Complete→Testing DIFF→Testing RBC/WBC→Data Analysis
→Ready will be displayed in sequence.
● The STAT loader will be opened automatically for the next test.
● After clicking , the position turns into . In the status, close
the loader and click “Count” button on the front panel to conduct tests.
(5)If the sample is not placed in position 3 of the loader, the following box will pop up when pushing the
loader back:
4-2
User Manual
Figure 4-3-5
The loader will be opened when the above box is prompted. Rotate it to the right position and then push it
back.
If “click” is not heard when rotating the loader, the following box will pop up when pushing back the loader.
Figure 4-3-6
Carry out tests after right positioning.
4.3.1.3 Saving Calibration Results
After obtaining 3 or more counting results, the instrument will carry out CV and calibration coefficient
calculation and save the calibration coefficient results. Saving after 5 times of counting is recommended. With
the increase in calibration times, CV and calibration coefficient will be updated.
The following box will pop up if the results obtained exceed the lower limit set by the software.
Figure 4-3-7
Click to close the box and clear the result of this counting. The result will be displayed on the
calibration interface only if the counting result is valid.
When saving the counting result after obtaining 3 or more counting results, if the calibration coefficient of a
certain parameter is not within 75%—125%, click , the following box will pop up:
Figure 4-3-8
4-3
User Manual
The operator should check the reference value input, if the reference value input is correct, the operator should
delete calibration results, re-run the calibration count.
4.3.1.4 Deleting Calibration Results
If any count result has a large deviation from the previous one, click “Test Result” on the left of the screen
after the test, and there will be a “×” mark (namely ), indicating that this result is not counted
during CV, mean, and calibration coefficient calculation. If this result does not need deletion after the test,
click “Test Result” again, and there will be a “×” mark. (namely ), indicating that this test result
is valid and that it can be involved in calculation.
If this result is indeed invalid, click to delete it.
4.3.1.5 Saving Calibration Coefficient
Click in Figure 4-3-1 after 5 times counting, the following box will pop up.
Figure 4-3-9
Click to save the calibration coefficient. It will be saved in “Calibration Coefficient” of “Manual
Calibration” interface, as Figure 4-3-10 shows:
Figure 4-3-10
4-4
User Manual
Figure 4-3-11
(1)Select the number of fresh blood calibration from the pull-down menu behind “Sample No.” Reselect its
“No.” after each fresh blood sample is tested for 5 times.
(2)Put the prepared fresh blood in the instrument to test 3 times, and calculate the mean value. The mean value
will be taken as reference value and input in Figure 4-3-11.
(3)Click in Figure 4-3-11, and the following interface will pop up:
Figure 4-3-12
(4)Click in the above interface. The close sample loader will be available, rotate the loader to make
position 3 aligned with the testing position (“Click” indicates right positioning), and place the prepared fresh
blood in position 3. Press the counting key of the instrument. The following box will pop up when pushing
back the loader.
Figure 4-3-13
The results will be displayed after tests.
4-5
User Manual
Figure 4-3-14
Click to close the box and clear the result of this counting. The result will be displayed on the
calibration interface only if the counting result is valid.
(3)When saving the counting result after obtaining 3 or more counting results, if the calibration coefficient of a
certain parameter is not within 75%~125%, click , and the following box will pop up:
Figure 4-3-15
The operator should check the reference value input. If the reference value input is correct, the operator should
delete calibration results, and re-run the calibration count.
4.3.2.4 Deleting Calibration Results
If any count result has a large deviation from the previous one, click “Test Result” on the left of the screen
after the test, and there will be a “× “ mark (namely ), indicating that this result is not counted
during CV, mean, and calibration coefficient calculation. If this result does not need deletion after the test,
click “Test Result” again, and there will be a “×” mark. (namely ), indicating that this test
result is valid and that it can be involved in calculation.
If this result is indeed invalid, click to delete it.
4.3.2.5 Saving Calibration Coefficient
Click in Figure 4-3-11 after 5 times counting, the following box will pop up.
4-6
User Manual
Figure 4-3-16
Click to save the calibration coefficient. It will be saved in “Calibration Coefficient” of “Manual
Calibration” interface.
4.3.3 Manual Calibration
The calibration coefficient displayed on the screen is the calibration coefficient saved after calibration with
calibrator or fresh blood calibration.
Click “Manual Calibration” in Figure 4-3-11 to enter the manual calibration interface, as the following figure
shows:
Figure 4-3-17
Click the corresponding line under “Calibration Coefficient” when the calibration coefficient needs adjustment,
and input the calibration coefficient. Click . “Save succeed” will pop up. Click .
4-7
User Manual
Figure 4-4-1
The dates, operators, modes, calibration methods and detailed test values of the latest 100 times calibration
will be displayed. If calibration counting is more than 100 times, the previous calibration results will be
covered.
Query:The operator can query the calibration history according to the calibration date, username and
calibration method.
4-8
User Manual
5.1 Introduction
After long-term use, error may occur, which may lead to unreliable results. Quality control (QC) provides an
effective way to detect error. The accuracy and precision of daily sample results can be checked through QC
tests.
In order to ensure the reliability of the sample results. Low, medium and high level Control should be used
respectively to conduct QC for the instrument once every day.
QC tests cannot be done until the instrument is calibrated correctly.
5.2 QC Method
The instrument provides three QC methods: L-J QC, ⎯x QC and X-B floating mean QC.
5.2.1 L-J QC
Under “L-J” QC, the operator can carry out quality control for 25 parameters. The instrument provides 12 QC
documents in order to save quality control parameters and results. Each quality control document can save up
to 400 groups of quality control results. When the running number of quality control is more than 400, the new
QC results will cover the old ones.
Click “QC” in the main interface and select “L-J/Xbar” from , as shown in Figure 5-2-1.
Figure 5-2-1
5.2.1.1 Quality Control Setting
(1)Lot No. Information Setting
Click “Setting” button to enter the QC information setting interface, as shown in Figure 5-2-2:
5-1
User Manual
Figure 5-2-2
a)Select File No.
Click the pull-down menu of “File No.” to select the file needs QC, the range is 1—12.
b)QC Lot No. Input
Input the corresponding QC lot No. in the “Lot No.” box according to the Control instructions.
c)Validity Setting:
Click the pull-down menu of “Validity” to set the validity according to the instructions.
d)QC Level Selection:
Select QC level (High, Mid, Low) from the pull-down menu of “Level”. Each lot No. corresponds to one
level.
e)Target/Deviation Limit
Input the corresponding target value and deviation limit according to the QC instructions.
Click when the above input is completed. “Saving Succeed” will pop up, and click
.
The input lot number and validity should be those marked on the instructions.
(2)Preset Value
The operator can select any QC results from the QC chart (at least 3), calculate the mean, SD, and CV% of the
QC results. Set them as preset values when editing QC.
(3)Deviation Limit Setting
If the display form of deviation limit, or the calculation method of deviation limit in pre-set value needs to be
adjusted, following these steps:
a)Click “Setting” in the main menu, and select “QC Setting” from the left, as Figure 5-2-3 shows:
5-2
User Manual
Figure 5-2-3
b)Calculation method and range selection of deviation:
When “Absolute Value” calculation method is selected, the input deviation limit will be displayed in
the form of an absolute value. “Range” will use two times the standard deviation (2SD) or 3 times the
standard deviation (3SD) as the deviation limit.
When “Percent” calculation method is selected, the input deviation limit will be displayed in the form
of a percent. “Range” will use 2 times the variation coefficient (2CV) or 3 times the variation coefficient
(3CV) as the deviation limit.
c)Click in Figure 5-2-3. The preset value is obtained according to the set method, and it is taken
as the target value and deviation limit of the current QC file. The corresponding position of each parameter
will be displayed in Figure 5-2-2.
(4)Removal of Abnormal QC Points
If any abnormal QC point occurs, the preset value can be obtained after deleting the abnormal point. The
operating steps are as follows:
a)In Figure 5-2-2, click at the right lower corner, and the following figure pops up:
Figure 5-2-4
b)Click in the above figure, and the screen will switch to the QC interface, as the following
figure shows:
5-3
User Manual
Figure 5-2-5
If the first point on the right is an abnormal point, click it, the red cursor line is on the first point on the right.
Click , and the selected QC point will become blue. To continue to remove other QC points,
click them, and the red cursor line will be on them. Click . When obtaining pre-set values for
calculation, removed QC points are not included (blue QC points).
Click “Calculation” after the points are deleted, the screen will return to "Setting" interface, and the result will
be displayed.
If the valid QC point number is less than 3, click “ ” button in Figure 5-2-2, or click
“ ” button in Figure 5-2-5. Figure 5-2-6 will pop up:
Figure 5-2-6
If the above operation is found to be wrong, click the removed point to move the red cursor line onto it, and
click “Add”. The blue QC point will turn red or green. The removed point will return to normality.
Valid QC points are taken as the reference and deviation limit when taking a preset value.
5.2.1.2 QC Counting
Click “ ” in Figure 5-2-5 to enter the QC count interface, as shown in Figure 5-2-7.
5-4
User Manual
Figure 5-2-7
● The specified Control of the manufacturer should be used to avoid QC result error.
● Refer to the instructions for the use of Control.
● The result will not be reliable when instrument failure occurs.
Click in Figure 5-2-7, and the following interface will pop up:
Figure 5-2-8
Click in the above interface. The close sample loader will be available, rotate the loader to make
position 1 aligned with the testing position (“Click” indicates right positioning), and place the prepared
Control in position 1, as Figure 5-2-9 shows:
Figure 5-2-9
5-5
User Manual
Press the count key of the instrument. The following box will pop up when pushing back the loader.
Figure 5-2-10
Start Sample Aspiration→Aspiration Complete→Testing DIFF→Testing RBC/WBC→Data Analysis→Ready
will be displayed in sequence.
● The STAT loader will be opened automatically for the next test.
● After clicking , the position turns into . In the status, close
the loader and click “Count” button on the front panel to conduct tests.
If the sample is not placed in position 1 of the loader, the following box will pop up when pushing the loader
back:
Figure 5-2-11
The loader will be opened when the above box is prompted. Rotate it to position 1 and then push it back.
If “Click” is not heard when rotating the loader, the following box will pop up when pushing back the loader.
Figure 5-2-12
Carry out tests after right positioning.
If the count result is lower than the lower limit set by the software, or higher than the display range, the
following box will pop up:
5-6
User Manual
Figure 5-2-13
Figure 5-2-14
(1)QC Graph Setting
Click in “QC Graph” interface to set the parameters or move the display position, as Figure
5-2-15 shows:
5-7
User Manual
Figure 5-2-15
a)QC Item Setting
Select parameter names from Figure 5-2-15, and click . Only the selected parameter will be
saved.
b)QC Item Moving:
If the parameters need to be ordered again, select a parameter needing to be adjusted (the selected line is
blue), click “ ” or ” ” to move the selected parameter.
(2)Explanation of QC Graph interface
a)If the QC counting number is less than 3, the right side of the QC graph will not display the QC result of
each parameter.
b)The abscissa of the QC graph indicates the QC count number. The ordinate indicates the QC count result
of each parameter.
c)Vertical lines make the count data of the same group. For each parameter, the QC graph can display up to
400 points.
d)For each parameter, the three data on the left of the QC graph correspond to the three boundary values of
the QC graph, as shown in Figure 5-2-5.
From top to bottom, they represent the upper limit, target value and lower limit.
Upper limit: Control reference value + deviation limit
Target value: Control reference value
Lower limit: Control reference value - deviation limit
e)For each parameter, the three data on the right of the QC graph represent Mean, SD and CV%
respectively.
(3)Point explanation of QC chart
The points are QC results, they are connected by lines.
“Green Point” indicates the QC result is within the range. “Red Point” indicates the QC result is not within the
range.
(4)If the point is not within the range, conduct the following steps.
5-8
User Manual
Figure 5-2-16
5-9
User Manual
Figure 5-2-17
(1)Delete All
The operator can delete all the QC count results of the current QC file.
Click in Figure 5-2-17, as and the following figure pops up:
Figure 5-2-18
Click “ ”, and the interface will display “Delete Succeed”, which indicates that all the QC results
are deleted.
(2)Delete
The operator can delete a certain QC result of the current QC file.
Select a QC data needs to be deleted from Figure 5-2-17, and click . A box like Figure 5-2-18will
pop up. Click “ ”, and the interface displays “Deleting Succeed”, which indicates the selected QC
result is deleted.
(3)Print
The operator can print the QC count result of the current QC file.
(4)Export
The operator can export all the QC count results of the current QC file.
(5)Communicate
The operator can transmit the QC count results of the current QC file to LIS.
(6)Import
The operator can import all the QC count results.
5-10
User Manual
5.2.2 ⎯x QC
⎯x QC: the mean value of two testing results is saved as one QC point in the QC chart.
The operation is same as “5.2.1 L-J QC “.
5.2.3 X-B QC
X-B floating mean method is proposed by Dr. BrianBull. Through monitoring the stability of red blood cell
parameter, such as MCV, MCH, MCHC, etc. to monitor instrument performance. It belongs to the quality
control without control. Together with the quality control with control, they all belong to instrument
performance monitoring method which can reflect instrument performance from different aspects. They are not
mutually replaceable. The X-B quality control is recommended when the daily sample is greater than 100 per
day. This quality control method requires the use of random sample, therefore, the classified sample is not
suitable. It offers the upper and lower limit to form a reference range. Observe the changing trend of result
within the reference range.
This instrument conduct X-B QC toward MCV, MCH, MCHC. The sample number in each group can be set as
20~200. The sample is come from normal counting result of the instrument, without classifying whole blood
and pre-dilution mode.
Calculation is needed before testing, because X-B reference value is obtained through analyzing a large
number of random sample.
5.2.3.1 QC Setting
The QC parameter need to be set before X-B QC analysis.
(1)Click “Setting” in main menu, and select “QC Setting” in the left menu, open “X-B” interface, as Figure
5-2-19 shows:
Figure 5-2-19
Sample No. selected for each X-B point can be 20~200, the recommended number is 20.
Click ON under “X-B QC”, click , “Saving succeed” will pop up, click .
(2)Click “QC” in the above menu, and select “X-B” from , as Figure 5-2-20 shows:
5-11
User Manual
Figure 5-2-20
Target and Deviation Limit Input:
Click the target and deviation limit to input them.
Target: Reference of different area may be different, so the sample number should reach certain
number(more than 500), the mean should be taken as target of X-B QC.
Deviation Limit: 3%~5% of target should be taken as deviation limit.
5-12
User Manual
Figure 5-2-21
(1)QC graph explanation
The abscissa is the counting result number. The ordinate is the result.
For each parameter, the three numbers on the left of the QC graph correspond to the three boundaries of QC
graph. From top to bottom, they represent the upper limit, target value and lower limit.
Upper limit: Control reference value + deviation limit
Target: Control reference value
Lower limit: Control reference value - deviation limit
For each parameter, the three number on the right of the QC graph represent Mean, SD, and CV% respectively.
(2)Point of QC graph explanation
The points are QC results, they are connected by lines.
“Green Point” indicates the QC result is within the range. “Red Point” indicates the QC result is not within the
range.
(3)QC Data Modification
If the point is not within the range, conduct the following steps.
Check the target value and deviation limit.
Check whether the background test is normal.
Contact the manufacturer’s after sales service department if abnormity remains after calibration.
(4)QC Data Query
The QC result can be queried through previous and next. Corresponding QC result will be displayed under the
parameter name. The position of the QC point and the total QC result number will be displayed in the form of
“Position/Total Number” at the lower left side of the interface.
5.2.3.4 QC List Review
Click QC list in Figure 5-2-21to enter the X-B QC list review interface, as shown in Figure 5-2-22:
5-13
User Manual
Figure 5-2-22
(1)Delete
Click the QC data behind corresponding No., the line will turn blue, click , as Figure 5-2-23
shows:
Figure 5-2-23
Click “ ” in above Figure, “Delete Succeed” will be displayed, the selected records will be deleted.
Figure 5-2-24
Click “ ” in the above Figure, “Delete Succeed” will be displayed, all the records will be deleted.
5-14
User Manual
(3)Export
Click to enter the following figure:
Figure 5-2-25
Input the document name in the input box behind “Document Name”, and select the saving path, click
. (Note: The exported document will be saved in EXCEL form.)
(4)LIS Transmission
Click , the data will be transmitted to LIS, as the following figure shows:
Figure 5-2-26
(5)Import: Click “Import”, the operator can import the QC counting result in the interface.
5-15
User Manual
5-16
User Manual
6.1 Introduction
This chapter describes the whole process of daily operation from start to shut down and focus on the operation
process of whole blood sample and pre-dilution (Peripheral Blood) sample testing.
As shown in Figure 6-1-1:
Figure 6-1-1
6.3 Daily QC
QC analysis should be conducted every day before sample analysis to ensure reliable result. For the
detailed operation, please refer to chapter 5.
● Use clean EDTA-K2 anticoagulant vacuum tube, silicified glass/plastic tube and 20μL silica glass
capillary.
6-1
User Manual
● The sample that needs WBC differential or platelet count should be stored at room temperature. It
should be used within 8 hour after collection.
● If the analysis results of PLT, MCV, or WBC 5 part differential count are not needed, the sample can
be stored at 2 ℃ - 8 ℃ in refrigerator for 24 hours. The stored sample should be used after placing at
room temperature for at least 30 minutes.
● The sample should be mixed again before use after a period of storing.
6.4.2 Add Diluent
Figure 6-4-1
(1)Click in shortcut area (6-4-1), as the Figure 6-4-2, Figure 6-4-3 show:
Figure 6-4-2
Figure 6-4-3
(2)Take a clean centrifuge tube, place it at position 4 of STAT loader, rotate the STAT loader to align position
4 with test position, close the door of STAT loader, as Figure 6-4-4 shows:
6-2
User Manual
Figure 6-4-4
Press aspiration key, the diluent will be dispensed into the tube of Position 4, as Figure 6-4-5 shows:
Figure 6-4-5
Move the centrifuge tube after the door of STAT loader bounced.
(3)After dispensing the diluents, dispense 20 μL peripheral blood into the centrifuge tube that contains diluents
immediately. Then mix them.
(4)Click in Figure 6-4-3, as Figure 6-4-6 shows:
Figure 6-4-6
Cancel diluent adding to complete a preparation of pre-dilution sample.
(5)If there are some samples need to be diluted, the first preparation of pre-dilution sample is completed, as
Figure 6-4-2 shows, place the clean centrifuge tube at position 4 of STAT loader, close the door of
STAT loader and press the aspirate sample key, the second preparation of pre-dilution sample is completed.
The rest can be done in the same manner and pre-diluted samples can be prepared up to 20 continuously.
● The operator can also use pipette to aspirate 180μL diluent and dispense 20μL peripheral blood into
the centrifuge tube immediately. Then mix them.
● When “pre-dilute” is selected, “Attention! 1:10 Dilute sample” will be prompted in sample
information.
● The diluent should be protected from dust, or the analysis error will occur.
● After mixing of peripheral blood and diluent, it should be placed for 3 minutes, and then remix before
use.
● The diluted sample should be used within 30 minutes, or the analysis result will not be reliable.
6-3
User Manual
● The sample should be re-mixed after placing for a period of time before use.
● The stability of the analysis result on pre-dilution mode should be evaluated according to their own
sample number, sample collection method and technical level.
Patient information can be edited or modified if the lock key is in the status of , as Figure 6-5-1 shows:
Figure 6-5-1
(1)Sample Information Edition: refer to 3.2.1.1
(2)Patient Information Edition: refer to 3.2.1.2
6.5.2 Sample Analysis Procedures
In main interface, if the current mode is “Close-Whole Blood”, testing can be carried out directly. Or set the
current mode as “Close-Whole Blood”.
● Set the reference range in setting interface, otherwise, incorrect alarm may be prompted.
● Conduct sample test after working mode selection, the default reference range is “General”. The
alarms after test is in accordance with the reference range of General.
6.5.2.1 Test Interface
6-4
User Manual
Figure 6-5-2
6.5.2.2 Analysis Mode and Sample No. Setting
(1)In the dialogue of “Analysis Mode”, select Microscale-Whole Blood (No.2 position) or Close-Whole Blood
(No.3 position).
(2)In the dialogue of “Test Mode”, select CBC or CBC+DIFF.
“CBC” mode: conduct counting without WBC differential. The result includes 14 parameters and RBC,
PLT histogram. “CBC+DIFF” mode: conduct counting and WBC differential. The result includes 25
parameters, scattergram, histogram and 10 research parameters.
(3)In the dialogue of “Closed Counting Mode”, select Press counting key to test or Close STAT loader to test.
(4)In the dialogue of “Sample No.” edit sample No.
(5)In the drop-down list of Reference, select the reference range for next sample analysis, and default as
“General”.
● The upper limit of sample No. is 15, “—” is allowed while “.” is not allowed, or the software will
prompt invalid input.
● The input sample No. cannot be the same with the daily tested sample No. or edited sample No., or the
new record will replace the tested or edited one.
6-5
User Manual
Figure 6-5-3
(2)On ready status of counting (the green indicator is lighted), shake the tube up and down lightly to mix it
well and place it in position 2 or 3 of STAT loader, close the door of STAT loader. Press “Count” to test (If
“Close STAT Loader to Test” is selected, test will start directly after closing the door of STAT loader). Result
will be saved and displayed after testing. The STAT loader will be available after aspirating the sample.
(3)Place the next sample in position 2 or 3 of STAT loader, close the door of STAT loader. Press “Count” to
test (If “Close STAT Loader to Test” is selected, test will start directly after closing the door of STAT loader).
(4)Click in main interface to print the report. Or click to enter into data list, then
press or to print the report.
(5)If the sample is placed in STAT loader, the position (Figure 6-5-3) is not in accordance with the set position
(select in Figure 6-5-2), the following dialogue will pop up when and then close the door of STAT loader.
Figure 6-5-4
The door of STAT loader will open again when the above dialogue prompts. Rotate it to the right position and
then close the door and test.
If “Click” is not heard when rotating the STAT loader, the following dialogue will pop up when close the door
of STAT loader.
Figure 6-5-5
Test after right positioned.
● The disposable vacuum blood collection tube can be used for 5 times at most. Because fragments
produced by repeatable puncture may lead invalid test result.
6-6
User Manual
● In the process of testing, the interface of Picture and of Research Parameter/microscopic examination
result can be checked. But no operation can be conducted. The data will be saved into the data review
after test automatically.
● In analyzing, if aperture is clogged, and there is “?” behind the test result, meanwhile, “aperture is
clogged, please remove clog or rinse aperture” will prompt, this result is unreliable. Please test after clog
removal.
● The result will be inaccurate if the temperature exceeds the normal working temperature range of the
instrument.
6.5.3 Picture Check
In shortcut area of Figure 6-5-1, click , click “Picture” to check the pictures of the test results, as
Figure 6-5-6 shows:
Figure 6-5-6
Histogram of RBC, PLT and WBC and histogram and scattergram of WBC 4 diff will be displayed.
6.5.4 Research Parameter Check
Click “Research Parameter/microscopic examination result” in Figure 6-5-6to check 10 research parameters,
as Figure 6-5-7 shows:
6-7
User Manual
Figure 6-5-7
“Please input microscopic result” will prompt if print the research parameter directly without inputting
microscopic result.
6-8
User Manual
Figure 6-6-1
(3)Change analysis mode and sample No.: the detailed operation procedures are the same as 6.5.2.2
6.6.2 Sample Analysis
(1)After selection, click in Figure 6-5-1, the mode will switch to “Close-Prediluted” mode. And the
door of STAT loader will open automatically, then rotate the STAT loader to align position 4 with testing
position (“Click” indicates right positioning), as Figure 6-6-2 shows:
Figure 6-6-2
(2)On ready status of counting (the green indicator is lighted), shake the tube up and down lightly to mix it
well and place it in position 4 of STAT loader, close the door of STAT loader. Press “Count” to test (If “Close
STAT Loader to Test” is selected, test will start directly after closing the door of STAT loader). Result will be
saved and displayed after testing. The STAT loader will be available after aspirating the sample.
(3)Place the next sample in position 4 of STAT loader, close the door of STAT loader. Press “Count” to test (If
“Close STAT Loader to Test” is selected, test will start directly after closing the door of STAT loader).
(4)Click in main interface to print the report. Or click to enter into data list, then press
or to print the report.
● In the process of testing, the interface of Picture and of Research Parameter/microscopic examination
result can be checked. But no operation can be conducted. The data will be saved into the data review
after test automatically.
● The result will be inaccurate if the temperature exceeds the normal working temperature range of the
instrument.
6-9
User Manual
Width
Figure 6-7-1
6.7.1.3 Barcode Cutting
The starting blank and ending blank should be no less than 3mm (Figure 6-7-1).
6.7.1.4 Barcode Label Paste Requirement
(1)Paste the barcode smoothly.
(2)Paste the barcode in the right position, the lower edge of barcode (including starting blank) should run
along the lower edge of tube cover. And guarantee the barcode can be seen through the longitudinal notch of
tube rack after the tubes are placed in the tube rack.
6.7.2 Sample Information Edition
Patient information can be edited or modified if the lock key is in the status of , as Figure 6-7-2 shows:
6-10
User Manual
Figure 6-7-2
6.7.2.1 Sample Information Edition:
refer to 3.3.1.1.
6.7.2.2 Patient Information Edition:
refer to 3.3.1.2.
6.7.3 Analysis Mode and Sample No. Setting
In main interface, press the . As shown in Figure 6-7-3:
Figure 6-7-3
(1)In the dialogue of “Analysis Mode”, select Microscale-Whole Blood (No.2 position) or Close-Whole blood
(No.3 position ).
(2)In the dialogue of “Test Mode”, select CBC or CBC+DIFF.
“CBC” mode: conduct counting without WBC differential. The result includes 14 parameters and RBC,
PLT histogram. “CBC+DIFF” mode: conduct counting and WBC differential. The result includes 25
6-11
User Manual
● The upper limit of sample No. is 15, “—” is allowed while “.” is not allowed, or the software will
prompt invalid input.
● The input sample No. cannot be the same with the daily tested sample No. or edited sample No., or the
new record will replace the tested or edited one.
(5)In the drop-down list of Reference, select the reference range for next sample analysis, and default as
“General”.
(6)Rack No.: If “Automatic Rack No. Identification” is selected, test will be conducted from No.1. Tube No. is
the sequence tube No. on the rack. If “Automatic Rack No. Identification” is not selected, test will be
conducted from the input rack No.
(7)Do get barcode from scanner?: In auto-whole blood testing, internal barcode reader can be selected, if use it,
sample barcode in data browse /list is the barcode of test tube; if do not use it, sample barcode in data browse
/list is the edited barcode in sample register.
(8)Click OK in Figure 6-7-3 to save the above settings.
6.7.4 Sample Analysis Procedures
● Set the reference range in setting interface, otherwise, incorrect alarm may be prompted.
● Conduct sample test after working mode selection, the default reference range is “General”. The
alarms after test is in accordance with the reference range of General.
● The disposable vacutainer can be used for 5 times at most. Because fragments produced by repeatable
puncture may lead invalid test result.
6.7.4.1 Analysis Interface
Figure 6-7-4
6.7.4.2 Sample Analysis
On ready status of counting (the green indicator is lighted), place the prepared sample tube on the tube rack,
6-12
User Manual
and place the tube rack on the right side of the sample conveyor (there is a groove on the bottom of the tube
rack, and the groove should be toward right side of the sample conveyor), click in Figure 6-7-3 or press
“Count” on the front cover of the instrument to test. In the process of testing, the running status displays: Start
Sample Aspiration→Aspiration Complete→Testing DIFF→Testing RBC/WBC→Data Analysis, meanwhile,
the green indicator on the front cover of the instrument is twinkling.
According to the registered sample information, label the corresponding sample No. on the sample tube,
then place these sample tubes on tube rack based on the sequence of sample No.
(1)The result of each sample will be saved to query interface. The racks will be loaded to the left side of the
autosampler after testing. The operator can take away the racks.
(2)Click to print the report; or click in main interface to enter into data list interface,
and then click or to print the report. Refer to Chapter 7.
6.7.5 Picture Check
Click “Picture” in Figure 6-7-4 to check the pictures of the test results, as Figure 6-7-5 shows:
Figure 6-7-5
Histogram of RBC, PLT and WBC; histogram and scattergram of WBC 4 diff will be displayed.
6.7.6 Research Parameter Check
Click “Research Parameter/Microscopic examination result” in Figure 6-7-5 to check 10 research parameters,
as Figure 6-7-6 shows:
6-13
User Manual
Figure 6-7-6
Figure 6-8-1
The default emergency sample No. starts from 0-1:
(2)Sample Position Selection: For 2mL vacutainer, select “Close-Whole Blood (No.3 Position)”. For 0.5mL
vacutainer, select “Microscale -Whole Blood (No.2 Position)”. For prediluted sample, select “Close-Prediluted
(No.4 Position)”. Click in Figure 6-8-1. And the door of STAT loader will open automatically,
rotate the STAT loader to align position 3, 2 or 4 with testing position (“Click” indicates right positioning), as
Figure 6-8-2 shows:
6-14
User Manual
Figure 6-8-2
(4)The door of STAT loader will open automatically after test. Click to stop emergency test. Switch the
mode to “Auto-Whole Blood” for routine sample test.
(5)Click in main interface to enter into data list interface. Edit patient information compared with
Sample No., Rack No. and Tube No. (both Rack No. and Tube No. of emergency sample are 0), and in
accordance with 3.3.1.1 and 3.3.1.2. Click or to print the report. Refer to
Chapter 7.
(6)If the sample is placed in STAT loader, the position (Figure 6-8-2) is not in accordance with the set position
(select in Figure 6-8-1), the following dialogue will pop up when and then close the door of STAT loader.
Figure 6-8-3
The door of STAT loader will open again when the above dialogue prompts. Rotate it to the right position and
then close the door and test.
If “Click” is not heard when rotating the STAT loader, the following dialogue will pop up when close the door
of STAT loader.
Figure 6-8-4
6-15
User Manual
● In the process of testing, other interfaces can be checked. But no operation can be conducted. The data
will be saved into the data review after test automatically.
● The result will be inaccurate if the temperature exceeds the normal working temperature range of the
instrument.
WBC scattergram abnormal ? WBC scattergram abnormal DIFF channel scattergram abnormal
6-16
User Manual
Abnormal RBC distribution Abnormal RBC histogram distribution Abnormal RBC histogram
Bimodality RBC bimodality distribution RBC histogram has two or more peaks.
6.10 Sleeping
When it reaches the set sleep time, the instrument will enter into sleeping status automatically, and the status
information will be displayed in failure area. Users can press Count key to wake up the instrument, and
proceed with the instrument.
6.12 Shutdown
6-17
User Manual
In order to ensure the stability and accuracy test result tested by the instrument. Shutdown is required
after 24 hours continuous working. The operator must follow these steps to shut down the instrument.
Shutdown includes instrument shutdown and software exit.
6.12.1 Instrument Shutdown
Figure 6-12-1
(2)Click , “Shutdown Instrument” will pop up, the instrument will conduct pipe and counting pool
rinsing and soaking in the process of shutdown, the following dialogue will pop up after the previous steps:
Figure 6-12-2
(3)Switch to “O” on the right of the instrument to finish shutdown.
(4)Empty the waste barrel and deal with the waste solution after shutdown.
(1)Click ,or select “Menu”→“Exit System”, the following dialogue will pop up:
Figure 6-12-3
6-18
User Manual
7.1 Introduction
The instrument will store the result into database automatically after each test. The maximum storage capacity
is 100,000. The operator can query the sample result stored in the database. The result will be covered if the
storage number reaches the maximum.
Click in main interface to enter into query interface, as Figure 7-1-1 shows:
Figure 7-1-1
The results displayed in the database are in testing sequence (the last result is at the first place). The default
mode is the list mode.
If the screen cannot display all results, press “Page Down” or “Page Up” to switch the pages. The position of
the current result and the total number of database will be displayed in the form of “Pos/Total” at the bottom of
the interface.
Figure 7-2-1
(2)Input the “Start No.” and “End No.” need to be queried in Figure 7-2-1, click ”. The sample
7-1
User Manual
result will be displayed on the current screen and will be selected with (there is a in front of sample
position No.).
The input sample position should be within the range of sample database, otherwise, “Input range error,
please input again” will be prompted.
(3)If “Cancel Selection” is selected in drop-down list , the selection will be canceled.
7.3 Print
The sample results can be printed in sample query interface.
7.3.1 Single Sample Printing
Click in Figure 7-1-1, if “Preview before print or not” ( ) is
selected in “Print Setting”, as the following figure shows:
Figure 7-3-1
Connect with printer correctly, and click in above figure to print the report.
After printing the selected report, the current sample print status is , switch to the counting
interface, the print status on the printed report is “printed”.
7.3.2 Bulk Print
Click in Figure 7-1-1, as Figure 7-3-2 shows:
7-2
User Manual
Figure 7-3-2
Select the test date of the report, and input the start and end sample No., if “Audited” ( ) is
selected and “Preview before print or not” ( ) is selected, the report can be
previewed, as the following figure shows:
Figure 7-3-3
The selected & audited samples will be printed. If “Audited” ( ) is not selected, all selected
samples will be printed.
After printing the selected report, the current sample print status is , switch to the counting
interface, the print status on the printed report is “printed”.
7.4 Query
The data can be queried through different conditions. Click in Figure 7-1-1, and select
7-3
User Manual
Figure 7-4-1
Users can select different query conditions to query, including: sample No., barcode No., test time, test mode,
analysis mode, case No., bed No., name, gender, department, deliver doctor, auditor and tester. The results can
be queried according to one condition or multi-condition.
7.4.1 Query According to Sample No.
Input the sample No. in the box behind “Sample No.” need to be queried in Figure 7-4-1, if the input sample
No. is “14”, click , the sample results with No. “14” will be displayed, as Figure 7-4-2shows:
Figure 7-4-2
7-4
User Manual
will be displayed.
7.4.4 Query According to Case No.
Input the Case No. in the box behind “Case No.” need to be queried in Figure 7-4-1, click , the
results will be displayed.
7.4.5 Query According to Bed No.
Input the Bed No. in the box behind “Bed No.” need to be queried in Figure 7-4-1, click , the results
will be displayed.
7.4.6 Query According to Name and Sex
Input the name and select the sex of the patient in the box behind “Name” need to be queried in Figure 7-4-1,
click , the results will be displayed.
7.4.7 Query According to Department and Deliver Doctor
Select the department of the sample in the drop-down list behind “Department”, and select the deliver doctor
of the sample in the drop-down list behind “Deliver Doctor” need to be queried in Figure 7-4-1, click
, the results will be displayed.
7.4.8 Query According to Auditor
Select the auditor of the sample in the drop-down list behind “Auditor” need to be queried in Figure 7-4-1,
click , the results will be displayed.
7.4.9 Query According to Tester
Select the tester of the sample in the drop-down list behind “Tester” need to be queried in Figure 7-4-1, click
, the results will be displayed.
If “Perfect Match” ( ) is selected in above query, only perfect match results will be
displayed. If “Perfect Match” ( ) is not selected in above query, all results that can
meet the input condition will be displayed.
Figure 7-5-1
7-5
User Manual
Figure 7-5-2
Figure 7-5-3
● The repeatability of the results is invalid if any parameter of the selected sample is invalid.
7-6
User Manual
Figure 7-5-4
User can edit the deviation limits corresponding to the parameters in the trend chart based on the requirements,
and click after editing. Click in Figure 7-5-3, as the following figure shows:
Figure 7-5-5
User can set the item need to be checked based on the requirements, and select in front of the parameter,
or click , , , , modify the position of parameters in the trend
chart, and click to exit after setting.
7-7
User Manual
Figure 7-6-1
Select the test date in the drop down list of “Test Date”, and input the start sample No. and end sample No.,
click , the auditor information of audited sample result will be displayed behind auditor.
7.7 Communication
7.7.1 Communication Interface
Click list review interface, and select proper records in the list , and data
transmission can be conducted. The interface is shown as follow:
Figure 7-7-1
Transmission Succeed is displayed, as Figure 7-7-2 shows:
Figure 7-7-2
7-8
User Manual
Select “Sample Application” In the drop down list , The interface is shown as follow:
Figure 7-7-3
Input start and end sample No., click , and apply patient information from LIS system, and store the
applied patient information into the corresponding interface of the record of the software.
7.8 Delete
Click in Figure 7-1-1, and select the records need to be deleted in its drop-down
Figure 7-8-1
7.9 Export
The exported data can be saved.
Click in Figure 7-4-2, select the data needs to be exported in its drop-down
7-9
User Manual
Figure 7-9-1
Input the file name in the input box behind “File Name”, and select the saving path, click , the
following box will pop up:
Figure 7-9-2
Indicate successful data export.
7.12 Recheck
Select the record which is suspected and right click in Figure 7-1-1, as Figure 7-12-1 shows:
7-10
User Manual
Figure 7-12-1
Click recheck in Figure 7-12-1, the status becomes recheck, then, testing (run the regular test). last, completed
(finish the recheck).
In the process of recheck, the status of counting interface is the same with what of result query.
Figure 7-13-1
The sample result position and the total number in sample database will be displayed in the lower side of the
interface in “Pos/No.” form.
Current data can be saved, deleted, audited and print in this interface.
7-11
User Manual
Figure 7-14-1
Click review compare in Figure 7-14-1, the following figure prompts:
Figure 7-14-2
7-12
User Manual
Figure 7-14-3
(1)Item setting:
Select the parameter items need to be displayed in Figure 7-14-3, click when the check box
displays , the selected parameter items will be displayed.
(2)Item moving:
Select the parameter items need to be adjusted (the line is blue) if the parameter items need to be reordered,
click or to carry it out.
7.14.2 Compare List
Input name “Li” and do not input case No. in Figure 7-14-2. Select a test time, and click “Query”, the
following figure prompts:
Figure 7-14-4
7-13
User Manual
The test result of Case No.456 which is in the selected Test Time will be displayed.
Input Case No. “456” Name Li and Test Time in Figure 7-14-2, click Query, the following figure prompts:
Figure 7-14-5
The test result of Case No.456 and Name Li which is in the selected Test Time will be displayed.
Input Case No. “456” Name Wang Hong and Test Time in Figure 7-14-2, select Exact Search click Query, the
following figure prompts:
Figure 7-14-6
The test result of Case No.456 and Name Wang Hong which is in the selected Test Time will be displayed.
7.14.3 Tendency
Click Tendency after the above Query, the following figure prompts:
7-14
User Manual
Figure 7-14-7
Select different parameter items in the drop-down list of Parameter, for instance, select WBC to carry out the
tendency of WBC. The corresponding WBC values are displayed in the bottom of Figure 7-14-7.
By default, “ ”is ticked in the compare list, users can check the corresponding graph in the tendency
interface. If only one item is ticked, select corresponding parameter in the tendency interface, and only one
point will display in the graph.
Figure 7-15-1
There are 3 types of test data backup:
(1)Click Selected Record to select the records need to be backed up. Selected Records refer to the items
with .
(2)Click All Records to backup all the items.
7-15
User Manual
(3)Click Input Range, select the start and end date after Test Date, and backup the items that are in the selected
range.
Click OK, the following figure prompts:
Figure 7-15-2
Select the path of backup, input File Name, and click save, the following figure prompts:
Figure 7-15-3
Click OK to exit after the operation of backup.
Figure 7-15-4
In addition, QC data backup, Calibration coefficients backup and Set data backup can be selected, all the
backup contents will be recovered if necessary.
7-16
User Manual
Figure 7-16-1
Select the path of backup, input File Name, and click open, the following figure prompts:
Figure 7-16-2
The following figure prompts after data recovery:
Figure 7-16-3
Click OK to exit, and the backup data will display in the database.
7.17 Help
Click “Help” in the menu to enter into Component Maintenance interface, and check the service condition of
the components, as following figure shows:
7-17
User Manual
Figure 7-17-1
All the buttons should be dimmed. If any button is highlighted, the corresponding test item should be checked
or the corresponding component should be replaced. Please contact the after-sales service of the manufacturer
and ask the professionals for maintenance advice.
7-18
User Manual
Chapter 8 Service
8.1 Introduction
In order to ensure the normal running of the instrument, routine maintenance is required. The instrument will
prompt the user to conduct maintenance after testing a certain number of sample or a continuous working
period. The service menu in the instrument offers routine maintenance methods and failure solutions, but users
should make their own maintenance plan according to daily sample number, operating environment, running
time, etc. to reduce the impact of various factors and ensure the safe, stable and effective running of the
instrument.
● Don’t spill reagents, samples, waste liquid, etc. over the mechanical or electrical parts of the
instrument, so as to avoid instrument damage.
● While working, the operator should take protective action, such as wearing protective gloves and
putting on labor clothes. Otherwise he/she may be infected when in contact with contaminated area or
liquid. When in contact with contaminated liquid, wash with water immediately and perform
sterilization.
● During cleaning and maintenance, pay attention to the parts holding liquid. The failure may lead to
danger.
● The instrument includes a laser device. Pay attention to the alarm labels during maintenance. Don’t
stare at the light beam directly or through optical devices.
● Improper maintenance may damage the instrument. The manual must be followed in maintenance.
● Please contact with the after-sales service department of the manufacturer Company if the problems
that are not explained in the instruction occur. The professional staff will provide maintenance
suggestions.
(2)If the instrument is on for 24 hours, daily “Instrument Rinsing” should be conducted. Click in the
main interface and the instrument will execute instrument rinsing.
(3)If the instrument is on for 24 hours, the interval times of automatic rinse should be set: 10-200, and the
instrument will execute automatic rinse when the test times reaches the set test times.
(4)If the instrument is on for 24 hours, daily “WBC Pool Rinsing”, “RBC Pool Rinsing”, “DIFF Pool Rinsing”
operation should be conducted (service-maintenance-rinsing).
8.2.1.2 Weekly
If normal shutdown operation is conducted every day, “Detergent Soaking”(WBC Pool, RBC Pool, DIFF Pool)
should be conducted weekly.(service-maintenance).
8.2.1.3 Monthly
Swab rinsing should be conducted monthly if daily shutdown is conducted (Service-Maintenance-Rinsing).
For the first use of the 10mL detergent, take the 10mL detergent bottle out of the package, open the
cover, take out the rubber plug and pour 5mL -10mL detergent (about half a bottle of detergent) from
the 500mL detergent bottle to the 10mL detergent bottle. Then open the front door of the instrument,
put the 10mL detergent bottle in the reagent bottle slot which is in the middle of the counting unit (with
black cover). For the continuous use, when changing 500mL detergent, the 10mL detergent also needs to
be changed in accordance with the above steps.
8-1
User Manual
Figure 8-3-1
The version of software and algorithm library are displayed:
Figure 8-3-2
The versions shown in the figure above is just for reference. The version No. will change as the program
8-2
User Manual
upgrades.
Select “Instrument Information” in its drop down menu, as Figure 8-3-3 shows:
Figure 8-3-3
It has been set before delivery. Therefore, it can only be checked.
8.3.2 Basic Status
Select “Basic Status” in Figure 8-3-1, the screen will display as shown in Figure 8-3-4:
Figure 8-3-4
Status explanation:
(1)Temperature: Display the real-time temperature of and the normal range of reaction pool, working
environment and laser to help the operator judge and locate faults. When the room temperature is lower than
15℃ and is higher than the set lowest temperature or the room temperature is higher than 30℃ and is lower
than the set highest temperature, “the room temperature is out of the normal range” will prompt at the bottom
right of the software interface.
8-3
User Manual
(2)Pressure: Display the pressure reducing valve and normal pressure range.
(3)Voltage: Display 58V, 5V, voltage and the normal range of WBC aperture and of RBC aperture.
(4)Current: Display the normal range of and power adjust valve of laser current.
Figure 8-4-1
Valve detection and voltage detection cannot be conducted at the same time.
8-4
User Manual
normal range, Click in front of “Maintain” in Figure 8-5-1, and select “Replace/Prime”, as Figure 8-5-1
shows:
Figure 8-5-1
Execute when first install the instrument or all reagent are used up. If single reagent need to
be changed, its operation is as follow:
(1)FDO Lyse Replacement
Click “Replace FDO Lyse” when bubble exits in FDO Lyse pipeline, or FDO reagent is polluted, or FDO
reagent has been used up.
Figure 8-5-2
Figure 8-5-3
After replacing FDO Lyse, the following box will pop up:
8-5
User Manual
Figure 8-5-4
Figure 8-5-5
“Operation Completed” will pop up after changing the detergent, click to finish the replacement.
If the operation needs to be canceled, click in the box of “Yes/No Replace Detergent” to cancel
detergent replacement.
(6)FDO Lyse Priming
8-6
User Manual
Click “Prime FDO Lyse” when FDO Lyse being emptied. The following box will pop up:
Figure 8-5-6
Figure 8-5-7
The following box will pop up after FDO Lyse priming.
Figure 8-5-8
In the “fault information area”, clear the corresponding error message, the instrument automatically
conduct FDO Lyse priming when the software prompt FDO Lyse insufficiency.
(7)FDT Lyse Priming
Click “Prime FDT Lyse” after FDT Lyse is emptied. And the operation should be conducted as the prompts.
Or: In the “fault information area”, clear the corresponding error message, the instrument automatically
conduct FDT Lyse priming when the software prompt FDT Lyse insufficiency.
(8)SLS-I Lyse Priming
Click “Prime SLS-I Lyes” after SLS-I Lyse is emptied. And the operation should be conducted as the prompts.
Or: In the “fault information area”, clear the corresponding error message, the instrument automatically
conducts SLS-I Lyse priming when the software prompt SLS-I Lyse insufficiency.
(9)Diluent Priming
Click “Prime Diluent” after Diluent is emptied. And the operation should be conducted as the prompts. Or: In
the “fault information area”, clear the corresponding error message, the instrument automatically conduct
Diluent priming when the software prompt Diluent insufficiency.
8.5.2 Rinsing
Select “Rinse” in Figure 8-5-1, as Figure 8-5-9 shows:
8-7
User Manual
Figure 8-5-9
(1)Conduct WBC pool rinsing when the background test value of WBC or HGB parameter is abnormal. Click
in Figure 8-5-9. As shown in Figure 8-5-10.
Figure 8-5-10
Figure 8-5-11
“Operation Complete” will pop up after WBC pool rinsing, click to finish the whole process.
Click in Figure 8-5-10 to cancel WBC pool rinsing.
(2)Conduct RBC pool rinsing when the background test value of RBC or PLT parameter is abnormal. Click
in Figure 8-5-9. Confirm box will pop up, click , progress bar will pop up.
“Operation Complete” will pop up after RBC pool rinsing, click to finish the whole process.
Click to cancel RBC pool rinsing in confirm box.
(3)Conduct DIFF pool rinsing when the WBC-diff parameter is abnormal. Click
8-8
User Manual
in Figure 8-5-9. Confirm box will pop up, click , progress bar will pop up. “Operation Complete”
will pop up after DIFF pool rinsing, click to finish the whole process.
Click to cancel DIFF pool rinsing in confirm box.
(4)In order to avoid sampling component contamination, swab rinsing should be conducted after one month’s
running.
Click in Figure 8-5-9. The confirm box will pop up, click , as Figure 8-5-12 shows:
Figure 8-5-12
“Operation Complete” will pop up after swab rinsing, click to finish the whole process.
Click ” to cancel swab rinsing in confirm box.
(5)Bubble may exists in flow cell when magnified cell mass exists in scattergram and the background testing
value of WBC parameter is higher than normal value. “Sheath Pool Rinsing” should be conducted at this time.
The operation is as follow:
Click in Figure 8-5-9. Confirm box will pop up, click , progress bar will pop
up. “Operation Complete” will pop up after sheath pool rinsing, click to finish the whole process.
Click to cancel sheath pool rinsing in confirm box.
(6)If bubble exists in sample aspiration pump or the test values are lower after reagent replacement. “Eliminate
syringe pump bubble” should be conducted.
Click in Figure 8-5-9. Confirm box will pop up, click , progress
bar will pop up. “Operation Complete” will pop up after eliminating syringe pump bubble, click to
finish the whole process.
Click to cancel eliminating syringe pump bubble in confirm box.
8.5.3 Maintenance
Click “Maintain” in Figure 8-5-9, as shown in Figure 8-5-13.
8-9
User Manual
Figure 8-5-13
8.5.3.1 Emptying
In order to avoid liquid spill during pipeline maintenance, empty should be conducted first. Taken waste liquid
buffer bottle as an example:
(1)Click to empty the corresponding bottle and the pipeline. The following box will pop up.
Figure 8-5-14
Click in above figure to conduct empty WC. Progress bar will pop up. “Operation Complete” will
pop up after emptying WC, click to finish the whole process.
Click button on Figure 8-5-14 to cancel the emptying WC in confirm box.
The operation of “Empty PK”, “Empty WBC Pool”, “Empty RBC Pool”, “Empty DIFF Pool”, “Empty FDO
Lyse”, “Empty FDT Lyse”, “Empty SLS-I Lyse” and “Empty Diluent” is the same as “Empty WC”.
(2)Once instrument will not be used for 3 days-2 weeks, click button on Figure 8-5-13, as
Figure 8-5-15 shows:
Figure 8-5-15
8-10
User Manual
Figure 8-5-16
Take out reagent pipelines from reagent bottles and click “ ”, the progress bar of “Emptying
Pipeline About 6 Minutes…”will display. If the operation of emptying pipeline has been completed,
“Operation Complete” will pop up. After that, instrument should be placed at a clean place.
(3)Once instrument will not be used for 2 weeks, please replace reagent with distilled water, click
button on Figure 8-5-13, as Figure 8-5-17shows:
Figure 8-5-17
Click button on above figure, the progress bar of “Package Clean Pipe, About 33 Minutes…” will
display and if the operation of package clean pipe has been completed, “Operation Completed” will pop up.
After that, instrument should be placed at a clean place.
Then break-off distilled water and click button on Figure 8-5-13, as Figure 8-5-18 shows:
Figure 8-5-18
Figure 8-5-19
Take out reagent pipelines from distilled water bottles and click “ ”, the progress bar of “Package
Clean Pipe, About 33 Minutes…” will display, and if the operation of package clean pipe has been completed,
8-11
User Manual
“Operation Completed” will pop up. After that, instrument should be placed at a clean place.
8.5.3.2 Aperture
Aperture clog remove, zap and backflush should be conducted in order to clear the debris in aperture.
Click button on Figure 8-5-13, the following box will pop up:
Figure 8-5-20
Click in above figure to clean the aperture through high-voltage direct current. The progress bar
will pop up meanwhile.
Clear the debris in aperture to prevent and remove RBC aperture clog.
Click , the confirm box will pop up, click to flush it. The progress bar will pop up
meanwhile.
Together with zap, the RBC aperture clog can be removed.
Click , the confirm box will pop up, click to zap and flush the aperture. The
progress bar will pop up meanwhile.
8.5.3.3 Soaking with Detergent
In order to ensure the accuracy of the test result, detergent soaking of WBC pool, RBC pool and DIFF pool
assembly should be conducted in the following conditions:
Scattergram of test result is abnormal.
Aperture clog failure.
(1)RBC Pool Soaking
RBC pool soaking should be conducted every other week, the operation is as follow:
Click “RBC Pool” in Figure 8-5-13, as Figure 8-5-21shows:
Figure 8-5-21
Figure 8-5-22
8-12
User Manual
Figure 8-5-23
Click “Next” in Figure 8-5-23 to empty RBC pool, as the following figure shows:
Figure 8-5-24
Operation has been finished after emptying, as the following figure shows:
Figure 8-5-25
Figure 8-5-26
8-13
User Manual
Figure 8-5-27
Click “Next” to empty WBC pool, as the following figure shows:
Figure 8-5-28
Figure 8-5-29
Figure 8-5-30
The interface will pop up after soaking, as Figure 8-5-31shows:
8-14
User Manual
Figure 8-5-31
Click “Next”, as the following figure shows:
Figure 8-5-32
“Operation Complete” will pop up after emptying.
8.5.4 Reagent Registration
Click “Reagent Register” in Figure 8-5-13, the screen display as Figure 8-5-33 shows:
Figure 8-5-33
8.5.4.1 Input the barcode information manually
Manual barcode information input can be conducted if the external barcode reader is not connected.
Click barcode input box in Figure 8-5-33 to input the information according to the reagent package (click
if input incorrectly). Click after input.
Left times: Scan a barcode information for each, the remaining number will change in accordance with the
reagent in proportion to the size of bottle.
8-15
User Manual
Figure 8-5-34
Re-input the information after checking if the above condition occurs. Contact with the manufacturerCompany
or its distributor for continuous failure.
The following box will pop up if the barcode has been used:
Figure 8-5-35
Change another bottle of reagent if the above condition happened.
8.5.4.4 Reagent Replacement
Conduct reagent replacement in “Replace/Prime” of “Maintain” after reagent registration
Do not stare at the scanning beam during instrument running to avoid eye injury.
8-16
User Manual
Figure 8-6-1
Figure 8-6-3
8-17
User Manual
Swab
Figure 8-6-4
Unscrew upper joint of puncture needle and put the new puncture needle, as Figure 8-6-5 shows:
Figure 8-6-5
Insert the new puncture needle into swab vertically (as Figure 8-6-4 shows) Fix the bolt well.
Figure 8-7-1
The operator can also select “Error Log” and “Failure Log” in its drop-down list for checking.
8-18
User Manual
Testing the sample in case of failure may cause inaccurate result. If alarm prompts in the process of
testing, failure solution should be adopted first.
9.1 Introduction
Please refer to the following relative failure solutions if failures occur in the process of testing; please contact
with the after-sales service department of the manufacturer if the failures still exist.
Alarm information will be prompted in failure information area if failure occurs. Click the corresponding
failure, Figure 9-1-1 will pop up:
9-1
User Manual
Alarm
Alarm Statement Solution
Code
1.Please check whether SLS Lyse is sufficient.
2.Please replace new reagent if there is no reagent and clear the
[E00005] SLS Lyse Empty
failure.
3.Check the floater switch if the reagent is sufficient.
Data Collection Board Verification 1.Click Clear to clear this failure.
[E00022]
Error 2.Please check the data collection board if this failure remains.
1.Click Clear to clear this failure.
2.Please check whether the data collection board has been
Data collection board is
[E00021] connected if this failure remains.
disconnected
3.Please check whether the data collection board has been
damaged.
1.Click Clear to clear this failure.
2.If this failure remains. please check whether the heater has
been damaged. replace it if it is damaged.
[E10004] Laser temperature is too high. 3.Please check whether the thermistor has been damaged. replace
it if it is damaged.
4.Please check whether the overheat protector has been damaged.
replace it if it is damaged.
1.Click Clear to clear this failure.
2.If this failure remains. please check whether the heater has
been damaged. replace it if it is damaged.
[E10005] Laser temperature is too low. 3.Please check whether the thermistor has been damaged. replace
it if it is damaged.
4.Please check whether the overheat protector has been damaged.
replace it if it is damaged.
1.Click Clear to clear this failure.
2.Please check the temperature sensor connection if this failure
Laser temperature sensor
[E10006] remains.
disconnected.
3.Please check whether the temperature sensor has been
damaged. replace it if it is damaged.
9-2
User Manual
Alarm
Alarm Statement Solution
Code
1.Click Clear to clear this failure.
2.Please check the temperature sensor connection if this failure
WBC pool temperature sensor
[E10003] remains.
disconnecting
3.Please check whether the temperature sensor has been
damaged. replace it if it is damaged.
1.Click Clear to clear this failure.
[E00010] Test Motor Error 2.Please check whether test motor failure occur if this failure
remains.
1.Click Clear to clear this failure.
[E00016] Test Motor Sensor Error 2.Please check whether test motor sensor failure occur if this
failure remains.
1.Click Clear button to automatic clear failure.
[E00023] Sheath motor error
2.Check whether sheath motor failure occur.
1.Click Clear button to automatic clear failure.
[E00026] Sheath motor limit error
2.Check whether sheath motor failure occur.
1.Empty waste barrel or replace it with new waste barrel.
2.Please check the connection between sensor and mainframe if
[E00003] Waste Barrel Full
this failure remains.
3.Check the floater sensor.
1.Click Clear to clear this failure.
[E00013] Sampling Motor Error
2.Please check the sampling motor if this failure remains.
1.Click Clear to clear this failure.
[E00019] Sampling Motor Sensor Error
2.Please check the sampling motor sensor if this failure remains.
[E00001] Waste Bottle Full 1.Empty waste barrel or replace it with new waste barrel.
1.Click Clear to clear this failure.
[E00009] x-Axis Motor Error
2.Please check X axis motor if this failure remains.
1.Click Clear button to automatic clear failure.
[E00015] x-Axis Motor Sensor Error 2.If this failure still exist. check if failure occure on the sensor
of X-axis motor.
1.Click Clear to clear this failure.
[E00027] x-Axis Motor First Sensor Error
2.Please check X axis motor sensor 1 if this failure remains.
1.Click Clear to clear this failure.
[E00028] x-Axis Motor Second Sensor Error
2.Please check X axis motor sensor 2 if this failure remains.
1.Click Clear to clear this failure.
[E00029] x-Axis Motor Third Sensor Error
2.Please check X axis motor sensor 3 if this failure remains.
[E00020] Front door is opened 1.Click Clear to clear this failure automatically.
9-3
User Manual
Alarm
Alarm Statement Solution
Code
1.Please check whether the diluent has been contaminated.
HGB background voltage exceeds
[E10016] 2.Click Clear if diluent has not been contaminated to clear this
range
failure.
Tube rack feed to home position 1.Check if there is barrier on mobile pipeline. if any. remove it.
[E20001]
error(posteroanterior) 2.Applies lubed on mobile bearing.
9-4
User Manual
Alarm
Alarm Statement Solution
Code
1.Please check whether there is diluent in the diluent bottle.
2.Replace a new diluent if there is no diluent in the bottle. Click
[E00002] Diluent Bottle Empty Clear to prime diluent.
3.Check the floater switch if the diluent is sufficient.
4.Check whether No.6 or 12 valve is working normally.
[E20012] Left side tube racks are full. Check if tube racks in left side of sampler are full
[E20009] Tube rack has been moved Check K5, K6 switch
[E20002] Have no tube rack Check if tube racks in right side of sampler are full
Move tube rack from home position 1.Apply lubed on mobile bearing.
[E20010]
error 2.Please check K9 switch
1.Click Clear to clear this failure.
[E00014] Y Axis Motor Sensor Error
2.Check whether Y axis motor sensor failure occur.
1.Please check S1 optocoupler.
[E20018] S1 optocoupler error.
2. Please check S1 optocoupler catch.
[E30002] Data colloct board reset error. Change data colloct board.
Temperature controller board reset
[E30003] Change temperature controller.
error.
[E30004] Sampler board reset error Change sampler board.
1. Please check waste liquid pump .
[E40001] Row of waste liquid fault error.
2. Please check waste liquid valve .
9-5
User Manual
9-6
User Manual
10.1 Transportation
The transportation of the instrument should be in accordance with the contract stipulation. It should avoid
violent collision and should be protected from corrosive materials.
10.2 Storage
The packaged instrument should be kept in a well-ventilated room without corrosive gas, the temperature
should be -10℃~40℃, the relative humidity should be no more than 75%.
10-1
User Manual
10-2
User Manual
A-1
User Manual
MSH –message head: this message segment is required item, includes HL7 message basic information,
message separator value, message type and message coding method and so on, it is each HL7 message’s first
message segment.
Information Example:
MSH|^~\&|XXX|1234567890|||20100419104618||ORU^R01|361|P^S|2.4|||||CHN|UTF-8<cr>
Serial HL7 Advised
Field Name Length Explanation Example
NO. Length
Include the first field separator after message
Field
1 1 1 segment, used for regulating other message |
Separator
field separator value
Coded Include discreteness separator, repeat
2 4 4 ^~\&
Character separator, ESC, sub-discreteness separator
Send Send terminal apply program
3 7 180 XXX
Program value: XXX
Instrument Sending terminal instrument, value: 12345678
4 10 180
Code instrument code 90
Message created time (form
20110310
7 Send Time 14 26 As YYYY[MM[DD[HH[MM[SS]]]]]) ,
144704
Take system time value
Message type, form as “information
Message ORU^R0
9 7 7 type^event type”, value: ORU^R01(Sample)
Type 1
OUL^R21 (LJ/X, XB QC)
Message Message control ID is used for only mark one
10 20 20 361
Control ID message, value :PID
This field is used for decide on whether to
transact HL7 operation program’s(7th layer)
Transact transact rule definition information.
11 3 3 P^S
ID NO. Value:P^ message type(Type
Value:S-sample, LJ-LJ /X barQC, XB-XB
QC)
HL7
Agreements adopt HL7 version No. Value:
12 Version 3 60 2.4
2.4
NO.
Nation
17 3 3 Nation code mark, refer to HL 7 2.4 CHN
Code
Character ISO/IEC 10646-1-1993 International
18 10 10 UTF-8
Set character standard value: UTF-8
A-2
User Manual
A-3
User Manual
A-4
User Manual
A-5
User Manual
(1)Patient Sample
<SB> MSH|^~\&|XXX||||20110310150421||ORU^R01|8|P^S|2.4|||||CHN|UTF-8 <cr>
PID||1234567890|||Wang Sanqiang|||Male<cr>
PV1|||clinic^^235689||||Doctor Wang|Zhang San|Li Si<cr>
OBR||2|12345|1001^ Count Results||20110310112251|20110310112409|||||| |20110310 112251||||0|0 <cr>
OBX|1|IS|2001^MODE||0||||||F<cr>
OBX|2|IS|2002^MODE_EX||1||||||F<cr>
OBX|3|IS|2003^Ref||0||||||F<cr>
OBX|4|IS|2004^Age||17|age|||||F<cr>
OBX|5|ST|2005^Note||note position||||||F<cr>
OBX|6|NM|2007^V_WBC||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|7|NM|2008^V_BAS_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|8|NM|2009^V_NEU_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|9|NM|2010^V_EOS_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|10|NM|2011^V_LYM_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|11|NM|2012^V_MON_c||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|12|NM|2013^V_BAS_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|13|NM|2014^V_NEU_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|14|NM|2015^V_EOS_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|15|NM|2016^V_LYM_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|16|NM|2017^V_MON_p||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|17|NM|2018^V_RBC||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|18|NM|2019^V_HGB||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|19|NM|2020^V_MCV||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|20|NM|2021^V_MCH||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|21|NM|2022^V_MCHC||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|22|NM|2023^V_RDW_CV||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|23|NM|2024^V_RDW_SD||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|24|NM|2025^V_HCT||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|25|NM|2026^V_PLT||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|26|NM|2027^V_MPV||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|27|NM|2028^V_PDW||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|28|NM|2029^V_PCT||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|29|NM|2030^V_P_LCR||4.63|10*9/L|11.00-12.00|L|||F<cr>
OBX|30|ED|2101^RBC Scattergram.BMP||……BMP binary system data change to BASE64
code……||||||F<cr>
OBX|31|ED|2102^PLT Scattergram.BMP||……BMP binary system data change to BASE64
code……||||||F<cr>
OBX|32|ED|2103^WBC Scattergram.BMP||……BMP binary system data change to BASE64
code……||||||F<cr>
OBX|33|ED|2034^DIFF Scattergram.BMP||……BMP binary system data change to BASE64
code……||||||F<cr>
OBX|34|ED|2104^WBC Scattergram.BMP||……BMP binary system data change to BASE64
code……||||||F<cr>
<EB><CR>
(2)L-J/Xbar QC
A-6
User Manual
<SB>MSH|^~\&|XXX||||20110311091016||OUL^R21||P^LJ|2.4|||||CHN|TUF-8<cr>
OBR||2|123 |1002^ LJ QC||20100819 |20110217131356|||||| |||||0|0<cr>
OBX|1|IS|2006^Level||0||||||F<cr>
OBX|2|NM|2007^V_WBC||4.63||||||F<cr>
OBX|3|NM|2008^V_BAS_c||4.63||||||F<cr>
OBX|4|NM|2009^V_NEU_c||4.63||||||F<cr>
OBX|5|NM|2010^V_EOS_c||4.63||||||F<cr>
OBX|6|NM|2011^V_LYM_c||4.63||||||F<cr>
OBX|7|NM|2012^V_MON_c||4.63||||||F<cr>
OBX|8|NM|2013^V_BAS_p||4.63||||||F<cr>
OBX|9|NM|2014^V_NEU_p||4.63||||||F<cr>
OBX|10|NM|2015^V_EOS_p||4.63||||||F<cr>
OBX|11|NM|2016^V_LYM_p||4.63||||||F<cr>
OBX|12|NM|2017^V_MON_p||4.63||||||F<cr>
OBX|13|NM|2018^V_RBC||4.63||||||F<cr>
OBX|14|NM|2019^V_HGB||4.63||||||F<cr>
OBX|15|NM|2020^V_MCV||4.63||||||F<cr>
OBX|16|NM|2021^V_MCH||4.63||||||F<cr>
OBX|17|NM|2022^V_MCHC||4.63||||||F<cr>
OBX|18|NM|2023^V_RDW_CV||4.63||||||F<cr>
OBX|19|NM|2024^V_RDW_SD||4.63||||||F<cr>
OBX|20|NM|2025^V_HCT||4.63||||||F<cr>
OBX|21|NM|2026^V_PLT||4.63||||||F<cr>
OBX|22|NM|2027^V_MPV||4.63||||||F<cr>
OBX|23|NM|2028^V_PDW||4.63||||||F<cr>
OBX|24|NM|2029^V_PCT||4.63||||||F<cr>
OBX|25|NM|2030^V_P_LCR||4.63||||||F<cr>
OBX|26|ED|2031^RBC Histogram.BIN||……BIN binary system data change to BASE64 code……||||||F<cr>
OBX|27|ED|2032^PLT Histogram. BIN||……BIN binary system data change to BASE64 code……||||||F<cr>
OBX|28|ED|2034^DIFF Scattergram.BMP||……BMP binary system data change to BASE64
code……||||||F<cr>
OBX|29|ED|2035^BASO Scattergram.BMP||……BMP binary system data change to BASE64
code……||||||F<cr>
<EB><CR>
(3)X-B QC
<SB>MSH|^~\&| XXX||||20110311091040||OUL^R21||P^XB|2.4|||||CHN| UTF-8<cr>
OBR||||1004^ XB QC|||20071207160000||||||||||||<cr>
OBX|1|NM|2079^XB_Num||20||||||F<cr>
OBX|2|NM|2073^m_MCV_R||12.204||||||F<cr>
OBX|3|NM|2074^m_MCH_R||0.258||||||F<cr>
OBX|4|NM|2075^m_MCHC_R||12.445||||||F<cr>
OBX|5|NM|2076^m_MCV_L||45.859||||||F<cr>
OBX|6|NM|2077^m_MCH_L||1.258||||||F<cr>
OBX|7|NM|2078^m_MCHC_L||2.36||||||F<cr>
OBX|8|NM|2020^V_MCV||4.63||||||F<cr>
A-7
User Manual
OBX|9|NM|2021^V_MCH||4.63||||||F<cr>
OBX|10|NM|2022^V_MCHC||4.63||||||F<cr>
<EB><CR>
A-8
User Manual
B.1.1 “TextBox”
Rectangular box which contain multiple lines of text. Type, color and width of frame, font attribute, text
alignment and font direction (vertical or horizontal) can be set. Use “Text” and “Frame” tools to set the
attribute of the object, as Figure B.1 shows:
Figure B.1
Text box object includes: text, variables, data fields or any combination of these. Font formatting will be
applied to all text included in text object.
TextBox Modification:
Click on the left of Report Edit Designer, click the mouse after rectangular icon appears, as shown in
Figure B.2:
Figure B.2
Clipboard operation;
B-1
User Manual
Word wrap;
Cancel button;
OK button.
data included in the database includes patient information, sample test information, the corresponding
set of specific field will be detailed in the following chapter.
B.1.2 “Picture Frame”
Picture can be inserted in the report. The format of the picture is BMP, WMF, ICO.
Picture frame modification:
Click on the left of Report Edit Designer, click the mouse after rectangular icon appears, as shown in
Figure B.3:
Figure B.3
Click “Select…” in Figure B.3, and click “OK”, the picture can be inserted into the report.
B.1.3 “Line”
The horizontal or vertical line can be inserted in the report. In separate statement of the report, straight-line
makes it easy to be read. The line thickness and color can be adjusted by using the drawing toolbar.
Click , drag the mouse in the current page, the cursor will turn into a pencil to draw a straight line. Click
the mouse to begin the line, release it when the line is finished. The line can be modified.
Line modification: select the corresponding button in “frame toolbar” to modify the line.
(1)“Standard” toolbar
(2)“Format” toolbar
(3)“Frame” toolbar
B-2
User Manual
(4)“Alignment” toolbar
Figure B.4
Select the paper size of current printer in the drop down list of paper size.
If the current printing support self-defined paper format, select "Self-defining", and then input the width and
length of the paper format.
Not all printer drive or printer support self-defining paper format (e.g. printer drive “HP LaserJet 6L”
does not support 76mm*127mm size; printer drive “HP LaserJet 4L” does not support all self-defining
size)
B.2.2 Paper Source
Click “Paper Source” in Figure B.4, as Figure B.5 shows:
Figure B.5
Select the commonly used paper source.
B.2.3 Margins
Click “Margin” in Figure B.4, as Figure B.6 shows:
B-3
User Manual
Figure B.6
If the “Extend to the printer” option is selected, the page form of designer will not display the border area. All
regions of the page will be printed correctly. But the size is different in different printing.
If this option is canceled, and all margin set to be 0, then the margin will be automatically set to the selected
printer's maximum print area. When designed report switch from one printer to other printer (the printable area
of ink jet printer is smaller than that of stylus printer), this function very useful.
If the margin set to be non 0, margins will be reflected in a page from of the designer (marked with gray lines).
If you use a dot matrix printer, first preview whether the print content is within the print area (some stylus
printer will not print the content beyond printing scope, and other printer prompts the beyond print scope). In
this case, set the margins manually.
B.2.4 Other
Click “Margin” in Figure B.4, as Figure B.7 shows:
Figure B.7
Set the number of columns and column spacing according to the page width. If the "print to front page" option
is selected, it allows print the remaining area in the new page.
B.3 Users create their own report sheet
The report template we provided, list all data of patient test report and L-J QC report in detail. Open all
selected objects of the report before create report sheet. The report template is in software installation directory,
\ Print \ language\Sample: is patient sample report template; \ Print \language\ QC: is L-J QC report template.
Concrete operation: Open select report edit Select All New Report Page SetupPaste. In this
new report, the position, font and the letter of text can be modified.
B.3.1 Title
Set the sample test report and QC report in “System Setting”. For example: × × × hospital, the title is × × ×
hospital test report. The “Test Report” can be set hereby. It can be modified into “Blood test report”, “LJ
report”, etc. The title can be modified into static text.
B.3.2 Paper
Users also can modify the size of print paper. If the user use inkjet or laser printer and A4 paper, set paper A4.
If the user use stylus printer (such as: epson 300K, 1600K, etc.), and use 80 column printing paper, set the
paper to be self-defining. If a 80 column paper need to print three reports, set the paper to be 9.34 (length). If a
B-4
User Manual
80 paper need to print two reports, set the paper to be 14, paper width is 22. The default margin is 0.
B.3.3 Select the object need to be modified
Click the object need to be modified with mouse (points around the selected object will appear), press Shift,
more than one objects can be selected. Press Ctrl, and move the mouse meanwhile, then the mouse moved area
will be selected.
(1)Modify the letter of the static text: select the text need to be modified, and double click the textbox, input
the modified letter in the corresponding box, and press “Enter”.
(2)Modify the data field: If it is used to display certain data, only one textbox need to be added to appropriate
position, double-click the text box, input the data value, refer to the existing template for the value.
(3)Move object: select the object (more than one can be selected), then press the four buttons around “Move”
to move the object, arrow keys on the keyboard can also be used.
(4)Change size: select the object, press the up and down buttons of “high” or “width” to increase or decrease
the height or width.
(5)Change the font: Select the object, and then select the font size or bold, italic and so on.
(6)Undo: error happened during modification, undo operation for one or more times, the report will return back
to the style before modification.
(7)Save: press the “Save” button after all operation is completed.
If it is patient sample report, please store in the software installation directory \ Print \ language\Sample
folder; save quality control report into \ Print \language\ QC folder of software installation directory
(8)Open the system setting in data management software, select the modified or created report in print setting.
B-5
User Manual
B-6
User Manual
If you need any technical support, please contact us according to the following address and telephone:
Register/Manufacturer: DIRUI INDUSTRIAL CO.,LTD.
Register/Manufacturer Address:
95 Yunhe Street, New & High Tech. Development Zone, Changchun, Jilin 130012, the People’s Republic of
China
Headquarter Address:
3333 Yiju Street, New & High Tech. Development Zone Changchun, Jilin 130103, the People’s Republic of
China
Factory Address:
95 Yunhe Street, New & High Tech. Development Zone Changchun, Jilin 130012, the People’s Republic of
China
Tel: 400 811 6695, 400 811 6605
Website: http://www.dirui.com.cn
E-mail: dirui@dirui.com.cn
For complaint: +86(431)81935326 85177245
Fax: +86(431)85173354
Domestic customer service hotline: 400 811 6695, 400 811 6605
Domestic fax: +86(431)85100405
Domestic customer service e-mail: service.ch@dirui.com.cn
C-1
User Manual
C-2
User Manual
Operation Consumption
Sleep(do not use the instrument for 2 or more hours) Diluent 26mL
Wake-up(use the instrument after sleep(within 3h)) Diluent 76mL
Wake-up(use the instrument after sleep(3—10h)) Diluent 90mL
Wake-up(use the instrument after sleep(over 10h)) Diluent 136mL
Detergent 14mL
Diluent 245mL
Manual rinse FDO Lyse 3mL
FDT Lyse 0.48mL
SLS-I Lyse 1mL
Detergent 6.8mL
Diluent 200mL
Automatic rinse(Set times: 10—200) FDO Lyse 3mL
FDT Lyse 0.48mL
SLS-I Lyse 1mL
Detergent 7.4mL
Diluent 185mL
Automatic rinse before sleep (Set times 10—600) FDO Lyse 3mL
FDT Lyse 0.48mL
SLS-I Lyse 1mL
Diluent 130mL
FDO Lyse 2mL
Power on
FDT Lyse 0.22mL
SLS-I Lyse 0.5mL
Diluent 170mL
Power off
Detergent 11mL
D-1
User Manual
D-2
User Manual
D-3
User Manual
D-4
User Manual
E-1
User Manual
(2)WBC≤3.5%;
(3)HGB≤2.0%;
(4)PLT≤5.0%.
E.7 Throughput
The test speed of the analyzer is 80t/h.
E-2
User Manual
F-1
User Manual
F-2