Beruflich Dokumente
Kultur Dokumente
CHAPTER
04-11-05 AERO NUSANTARA INDONESIA III
REVISION NDT PROCEDURE MANUAL SECTION PAGE
2 1 --
Chapter III
Eddy Current Inspection
ISSUED DATE
CHAPTER
05-07-07 AERO NUSANTARA INDONESIA III
REVISION NDT PROCEDURE MANUAL SECTION PAGE
2 1 1
1. Scope
1.1 This procedure is applicable in measurement of electrical conductivity of
aluminum alloy.
1.2 This procedure is applicable in detection of surface and near surface
discontinuities in metal parts.
2. Reference
2.1 ASTM E215- 98 Standard practice for Standardizing Equipment for
Electromagnetic Examination of Seamless Aluminum-
Alloy Tube.
2.2 MIL-STD-1537C Electrical Conductivity Test for verification of heat
treatment of aluminum alloys, Eddy current methods.
2.3 ASTM-E-1004 Eddy current Measurements of Electrical Conductivity
3. Definitions
3.1 Volume resistivity (ρ)
Is a value of resistivity of a known volume of nonferrous metal expressed in
ohm (centimeters)2 per centimeter or simply ohm centimeters, where the area
is of uniform cross-section A in sq-cm and the resistance R in ohms is the
value for a known length L in cm of the area A, or expressed in equation form:
4. General Requirements
ISSUED DATE
CHAPTER
05-07-07 AERO NUSANTARA INDONESIA III
REVISION NDT PROCEDURE MANUAL SECTION PAGE
2 1 3
1) Equipments
Direct reading instruments that used for conductivity measurements this
procedure must be within ± 0.5 %IACS of the value being measured
a. Instrument sensitivity
Reading instruments shall have sensitivity such that changes of at
least 0.5 %IACS are clearly distinguishable over the conductivity
range of the aluminum alloys under test.
b. Instrument Accuracy
Reading instruments shall be capable of determining the conductivity
of aluminum alloys as a percentage of the IACS with an accuracy of
±1.0 % IACS or better through electrically non-conductive film and
coatings at least 0.003 inch (0.076 mm) thick.
2) Instrument Conductivity Standards
At least two instrument conductivity standards are required.
The high and low value instrument standards will have a minimum
difference of 10 % IACS. At least one standard will cover the 25 to 32 %
IACS range and one standard will cover the 38 to 62 % IACS range.
b. Standardization frequency
During continuous operation, calibration of the conductivity instrument
shall be conducted at 15 minutes intervals.
c. Lift-off effect
Lift off effect is determined by comparison of readings taken when
placing the probe on a bare standard followed by placing a 0.003 inch
(0.076 mm) thick non conducting flat shim between the probe and the
standard. The difference in the two readings shall not exceed 0.5
ISSUED DATE
CHAPTER
05-07-07 AERO NUSANTARA INDONESIA III
REVISION NDT PROCEDURE MANUAL SECTION PAGE
2 1 4
%IACS.
4) Test Article
a. Surface
The areas to be tested shall be free of dirt, grease, oil or other
contaminants. The surface finish of the areas to be tested shall be
150 RHR maximum, RMS 165 maximum or less than a maximum
established by objective evidence, for instrument in use, that the
required accuracy is maintained.
b. Edge Effect
To insure readings are not distorted from edge effect, the specimen
area to which the eddy current instrument probe is applied shall be at
least as large as the outside diameter of the test probe or per
manufacturer's specifications. The coil shall be centered on all parts
whose dimensions approach this limitation.
1) Equipments
Selection of the eddy current equipment for a particular inspection
is at the discretion of the using facility provided that the following
are accomplished.
a. The inspection system shall consist of an instrument and one or
more test coils with an operating frequency suitable for
conducting the specified eddy current inspection and capable of
obtaining the desired information.
b. If equipment is designed for lift-off compensation, all test shall
be performed with the instrument in the balanced for lift-off
condition. Unless otherwise specified the system shall be
balanced for 0 to 3 mil lift off.
c. The sensitivity of the system shall be such that a repeatable
change in signal response of the magnitude specified here in
can be detected when the appropriate test coil encounters a
known specified natural or artificial discontinuity in a reference
standard.
Note:
Any probes and instruments found not function or defect
Will be removed from service area or used repair tag to
attach to the instrument.
3) Equipment Calibration
4) Test Article
5. Detail Requirements
5.1. Eddy current Conductivity Testing
1) Lift off correction factors
a. Establish lift-off correction factors for material or parts having a
nonconductive coating exceeding 0.003 inches (0.0076 mm) as
follows:
1. Obtain plastic shim stock or a paper of the same thickness as the
nonconductive layer on the parts to be tested. Where paint
thickness is not known, it will be necessary to remove or measure
the paint thickness.
2. Place the shim or paper a bare specimen of the same nominal
conductivity as the parts to be tested.
3. Note the difference in readings of conductivity between test run
when the probe is placed directly on the surface of the bare
specimen and when the probe is placed on the paper or plastic
shim stock on top of the specimen. Use this difference as the lift
off correction factor for reading on the parts to be tested.
2) Operating Precautions
a. When necessary to rub or slide probe tip against a surface, do it
gently to save wear on the probe surface
b. Don't tip bang probe pressure is desirable
c. A firm consistent probe pressure is desirable.
d. Clean the surface of the material to be measured, if required.
e. Keep the probe tip clean by wiping often with a clean lint-free cloth or
a suitable commercial wiper.
3) Inspection Procedure
a. Connect the required probe coil to the instrument
b. Switch on the instrument and allow it to warm up for at least the
length of time recommended by the manufacturer.
c. Make all necessary setup and control adjustments in accordance with
the manufacturer's recommendation.
d. Calibrate the measurement system as follows:
ISSUED DATE
CHAPTER
05-07-07 AERO NUSANTARA INDONESIA III
REVISION NDT PROCEDURE MANUAL SECTION PAGE
2 1 7
4) Interpretation of Results
a. The results of eddy current conductivity testing are based on the
comparison of an unknown with one or more standards.
b. Electrical conductivity testing is most successfully used on flat objects
with a clean surface. For most instruments, the test object should
have an electrical conductivity that is between the values of the two
standards used to calibrate the instrumentation.
c. The value obtained in the measurement of the test object is
applicable only at the temperature of the test object at the time of
measurement.
In order to determine the electrical conductivity at other temperatures,
the temperature coefficient of conductivity of the test object must be
determined.
5). Report
The written report of an electrical conductivity measurement should
contain any information about the test set up that will be necessary to
duplicate at the same or some other location.