► Referenced documents
Document Title
DD+DIS230.10E Chapter 3.8, Software Releases, Patches
NOTE:
Dispose the enclosure document after finishing your task
Please observe the confidentiality note below.
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Revision 1
04-2014 printed in Germany Document Node ID: 35167060
Enclosure_e_template_v07_modified
Agfa Company Confidential Copyright © 2014 Agfa HealthCare N.V.
A604525.2 Enclosure
► Manufacturer
Agfa HealthCare N.V.
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2014 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium, or its
affiliates.
All other trademarks mentioned in this document are held by Agfa HealthCare N.V. or
the respective owners and are used in an editorial fashion with no intention of
infringement.
Nothing contained in this legal notice nor in any text in this document shall be
construed as granting by implication, estoppel or otherwise, any license or right to use
any of the trademarks, service marks, trade names or logos appearing in this document
without the express prior written consent of their respective owner.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" prior to attempting any operation, repair or
maintenance task on the equipment.
Refer to Document ID 11849633, Agfa Intranet / Agfa Portal via Internet.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare N.V or one of its affiliates
or who are otherwise authorized by Agfa HealthCare N.V. or one of its affiliates to
provide such services.
LIST OF CONTENTS
1 INTRODUCTION ......................................................................................................................5
2 SOFTWARE UPGRADE VIA USB FLASH DRIVE...................................................................5
2.1 Checking current software version ...........................................................................................6
3 APPENDIX..............................................................................................................................15
3.1 Software Upgrade via Browser ...............................................................................................15
Subject Software installation instructions for software CIRM2002 and higher versions
SCOPE OF DELIVERY:
ZIP File CIRM2002*
* or higher version
NOTE:
Use this software at CR 30-Xm only. CR 30-Xm
The software will not be accepted at CR 30-X 2nd.
m
Figure 1
REQUIRED TOOLS:
Service PC with USB port
USB flash drive with at least 10 MB free disk space.
Recommendation: As the Digitizer is not able to read all kinds of USB flash
drives, use the USB flash drive delivered with the Digitizer.
The USB flash drive including software can also be ordered via order number*:
CM+604518 4
* The last digit in the spare part number indicates the spare part revision at release of this
document. When ordering, the actual revision of the spare part is delivered.
REQUIRED TIME:
Approximately 15 minutes for the installation via USB flash drive
NOTE:
Register a software upgrade in the local service management system as modification,
e.g. in SMS as “MOD”.
1 Introduction
Perform the software upgrade via USB flash drive in following steps:
# Step Section
1 Check current software version. 2.1
2 Uninstall the old Error Viewer. 2.2
3 Take out USB flash drive. 2.3
4 Prepare the USB flash drive. 2.4
5 Perform the software upgrade. 2.5
6 Install the new Error Viewer. 2.6
7 Perform a stall calibration for the Drawer Unit Motor (M4) 2.7
8 Create a backup. 2.8
9 Put back the USB flash drive in its cover. 2.9
10 Verify the software upgrade. 2.10
NOTE:
If the Digitizer IP address is not known, look it up in the NX configuration tool:
Double-click the icon <Start NX Configurator> on the desktop.
Select: Load active configuration
Go to: Devices > Digitizer Configuration
This page shows the Digitizer IP address.
Quit the NX Configurator: Close without saving any changes.
Figure 2
Figure 4
B
5175_c35_00A.cdr
Figure 5
CAUTION:
Parts underneath the top cover can be damaged.
Carefully remove the top cover. Follow the instructions as shown in next step.
5175_c35_004.cdr
Figure 6
5175_c35_backup1.cd
Figure 7
5175_c35_404.cdr
Figure 8
NOTE:
For software which is delivered on a USB flash drive together with a spare part this
step has already been done in production.
(1) Extract the downloaded software zip file into a folder on the Service PC.
Contents of the ZIP file:
Software package CIRM2002.zip*
This document
Depending on the software version a readme.txt
Figure 9
IMPORTANT:
Do not switch off the Digitizer during software installation as the Handling Control
Board is re-programmed during the software installation procedure.
NOTE:
The Digitizer status LED activity sequence during software installation is as follows:
Slow red blinking: Approx. 0,5 minutes
Off: Approx. 0,5 minutes
Slow red blinking: Approx. 1 minute
Fast red blinking or constant green*: After approx. 2 minutes
If the status LED does not get off for approx. 0,5 minutes during software upgrade,
this is already an indication, that the software upgrade failed. See also section 2.10,
Verifying the software upgrade.
* Digitizer LED only turns constant green if the right Error Viewer is installed. If the Digitizer software does
not detect the Error Viewer or detects a version that does not fit, the LED is fast red blinking.
NOTE:
By keeping the default settings of the Error Viewer installation wizard it is
guaranteed, that the Error Viewer access rights for the different Windows users
(administrators / users) are set properly. See Figure 12 and Figure 13.
Between 1 to 15 seconds after installation of the Error Viewer the Digitizer status
LED turns green.
Figure 12 Figure 13
NOTE:
This step is required only if the previous software was version CIRM1102 or
CIRM1201.
The movement of the drawer has been improved in software CIRM1304 in order to limit
possible retry cycles. This advanced motion sequence requires a new Stall Calibration
for the Drawer Motor M4 after software update.
Figure 14
Figure 15
(1) Put the USB flash drive back in its cover and mount the screw.
(2) Mount the Top Cover.
(3) Close the Cassette Unit.
NOTE:
For new features and solved bugs in this software refer to the Digitizer chapter 3.8,
Software Releases, Patches, of the Digitizer service manual.
3 Appendix
NOTE:
Software upgrade via browser can be performed from any PC in the network.
It is tested with Firefox and Internet Explorer.
NOTE:
If the Digitizer IP address is not known, look it up in the NX Configuration Tool:
Double-click the icon <Start NX Configurator> on the desktop.
Select: Load active configuration
Go to: Devices > Digitizer Configuration
This page shows the Digitizer IP address.
Quit the NX Configuration Tool: Close without saving any changes.
Figure 16
Figure 17
NOTE:
After software installation on a new CF card the Digitizer LED will usually not get
green. Reason: The destination information is not available anymore. This is only
available after restoring the latest backup.
Figure 18
See the IMPORTANT note next page for following exceptional cases:
(A) A backup of the Digitizer data is not available.
(B) The Digitizer service menu cannot be accessed.
IMPORTANT:
(A) Proceeding if a Backup file is not available:
The Backup is a file with following syntax:
BAC_CR30-5179_<DigitizerSN>_<DDMMMYYYY>_<HH-MM-SS>.zip.
If the Backup is not available, following data are lost:
o Digitizer configuration, except Digitizer network settings. The network settings
are stored in an EPROM on the CPMI board.
o Logfiles
o Infocounter
In this case it is required to perform the Digitizer configuration. For details refer to
chapter 1, Controls, Connections and Set-up Procedure, of the Digitizer service
manual.
(B) Proceeding if the Digitizer service menu cannot be accessed:
The service menu cannot be accessed anymore, if the Digitizer network settings
are lost. This may happen, if CF card and CPMI board are replaced at the same
time. A restore is not possible in this case, as the browser connection is not
available anymore. Solving this issue has to be made in following steps:
Reset the Digitizer network settings by double-clicking file
RestoreDefaultAddress.cmd in directory CIRM2002* on the USB flash
drive and booting with the USB flash drive (for details see chapter 3.3,
Troubleshooting, of the Digitizer Service Manual).
Set the network IP4 address of the service PC to 192.192.192.193 with
subnet mask 255.255.255.0.
Connect the Service PC directly to the Digitizer and start the browser.
Enter IP address 192.192.192.192.
In the service menu select <Restore> and restore the latest Backup
from USB flash drive.
Reconnect the Digitizer to the NX workstation
Switch off/on the Digitizer.
* or higher version