Technical Information

Proline Promass 80A, 83A


Coriolis Mass Flow Measuring System
The single-tube system for highly accurate measurement of very
small flows

Application Features and benefits


The Coriolis measuring principle operates independently The Promass measuring devices make it possible to
of the physical fluid properties, such as viscosity and simultaneously record several process variables (mass/
density. density/temperature) for various process conditions
during measuring operation.
• Suitable for continuous measurement, filling and
dosing of very small flows. The Proline transmitter concept comprises:
• Extremely accurate measurement of liquids and gases • Modular device and operating concept resulting in a
such as emulsions, additives, flavouring, insulin, gases higher degree of efficiency
for high pressure and low pressure • Software options for batching and concentration
• Fluid temperatures up to +400°F (200°C) measurement for extended range of application
• Process pressures up to 5800 psi (400 bar) • Diagnostic ability and data back-up for increased
process quality
Approvals for hazardous area:
• ATEX, FM, CSA, TIIS The Promass sensors, tried and tested in over 100,000
applications, offer:
Approvals in the food industry/hygiene sector:
• Multivariable flow measurement in compact design
• 3-A, FDA, EHEDG
• Insensitivity to vibrations thanks to balanced single-
Connection to all common process control systems: tube measuring system
• HART®, Profibus® PA/DP, FOUNDATION • Efficient protection against forces from piping thanks
fieldbus™, MODBUS® to robust construction
Relevant safety aspects: • Easy installation without taking inlet and outlet runs
• Pressure Equipment Directive, SIL-2 into account

TI054D/24/ae/04.06
Proline Promass 80A, 83A

Table of contents

Function and system design. . . . . . . . . . . . . . . . . . . . . 3 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 21


Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Design / dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Material load curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Input signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Unified control concept for both types of transmitter: . . . . . . . . . 31
Language groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 31
Low flow cut off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 C-Tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Switching output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
FOUNDATION Fieldbus certification . . . . . . . . . . . . . . . . . . . . . 31
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
PROFIBUS DP/PA certification . . . . . . . . . . . . . . . . . . . . . . . . . 32
Electrical connection Measuring unit . . . . . . . . . . . . . . . . . . . . . . 9
MODBUS certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Electrical connection, terminal assignment . . . . . . . . . . . . . . . . . 10
Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Electrical connection Remote version . . . . . . . . . . . . . . . . . . . . . 11
Pressure measuring device approval . . . . . . . . . . . . . . . . . . . . . . 32
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Functional safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Remote version
cable specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Potential equalisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Registered trademarks. . . . . . . . . . . . . . . . . . . . . . . . 33
Performance characteristics. . . . . . . . . . . . . . . . . . . . 12
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 34
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Influence of medium temperature . . . . . . . . . . . . . . . . . . . . . . . . 14
Influence of medium pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Operating conditions: Installation . . . . . . . . . . . . . . . 14


Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Inlet and outlet run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Length of connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
System pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Operating conditions: Environment. . . . . . . . . . . . . . 18


Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Shock resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . 18

Operating conditions: Process . . . . . . . . . . . . . . . . . . 19


Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Medium pressure range (nominal pressure) . . . . . . . . . . . . . . . . . 19
Limiting flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

2 Endress+Hauser
Proline Promass 80A, 83A

Function and system design


Measuring principle The measuring principle is based on the controlled generation of Coriolis forces. These forces are always present
when both translational and rotational movements are superimposed.
FC = 2 · ∆m (v · ω)
FC = Coriolis force
∆m = moving mass
ω = rotational velocity
v = radial velocity in rotating or oscillating system

The amplitude of the Coriolis force depends on the moving mass ∆m, its velocity v in the system, and thus on
the mass flow. Instead of a constant angular velocity ω, the Promass sensor uses oscillation.
The measuring tube, through which the medium flows, oscillates. The Coriolis forces produced at the
measuring tube cause a phase shift in the tube oscillations (see illustration):
• At zero flow, i.e. when the fluid is at a standstill, the oscillation registered at points A and B is in phase, i.e.
there is no phase difference (1).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the outlet (3).

A A A
B B B

1 2 3

a0003383

The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the tube
oscillations at the inlet and outlet.
Compared to two-tube systems, other constructive solutions are required for the system balance for single-tube
systems. For this purpose, Promass A has an internal reference mass.
The measuring principle operates independently of temperature, pressure, viscosity, conductivity and flow
profile.
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and thus the
density of the oscillating system (comprising measuring tube and fluid) results in a corresponding, automatic
adjustment in the oscillation frequency. Resonance frequency is thus a function of fluid density. The
microprocessor utilises this relationship to obtain a density signal.
Temperature measurement
The temperature of the measuring tube is determined in order to calculate the compensation factor due to
temperature effects. This signal corresponds to the process temperature and is also available as an output.

Endress+Hauser 3
Proline Promass 80A, 83A

Measuring system The measuring system consists of a transmitter and a sensor. Two versions are available:
• Compact version: transmitter and sensor form a mechanical unit.
• Remote version: transmitter and sensor are mounted physically separate from one another.

Transmitter

Promass 80 • Two-line liquid-crystal display


• Configuration using key operation

a0003671

Promass 83 • Four-line liquid-crystal display


• Operation with “Touch control”
• Application-specific Quick Setup
• Mass flow, volume flow, density and temperature measurement as well as
calculated variables (e.g. fluid concentrations)

a0003672

Sensor

A • Single-tube system for highly accurate measurement of Documentation No.


very small flows TI 054D/24/ae
• Nominal diameters 1/24", 1/12", 1/8" (DN 1 to 4)
• Measuring tube made of 904L SS or Alloy C-22

a0003679

Further sensors in separate documentations

F • Universal sensor for fluid temperatures up to 400°F Documentation No.


(200°C) TI 053D/24/ae
• Nominal diameters 3/8" to 10" (DN 8 to 250)
• Measuring tubes made of 904L SS, 316L SS or
Alloy C-22

a0003673

F (High-temperature) • Universal high-temperature sensor for fluid temperatures Documentation No.


up to 660°F (350°C) TI 053D/24/ae
• Nominal diameters 1", 2" or 3" (DN 25, 50, 80)
• Tube material: Alloy C22

a0003675

M • Robust sensor for extreme process pressures, high Documentation No.


secondary containment requirements and fluid TI 053D/24/ae
temperatures up to 300°F (150°C)
• Nominal diameters 3/8" to 3" (DN 8 to 80)
• Tube material: titanium

a0003676

4 Endress+Hauser
Proline Promass 80A, 83A

H • Single bent tube. Low pressure loss and chemically Documentation No.
resistant material TI 052D/24/ae
• Fit-and-forget
• Nominal diameters 3/8" to 2" (DN 8 to 40)
• Tube material: zirconium

a0003677

I • Straight single-tube instrument. Minimal shear stress on Documentation No.


fluid, hygienic design, low pressure loss. TI 052D/24/ae
• Fit-and-forget: No special supports required for
installation.
• Nominal diameters 3/8" to 3" (DN 8 to 80)
• Tube material: titanium
a0003678

E • General purpose sensor, ideal replacement for volumetric Documentation No.


flowmeters. TI 061D/24/ae
• Nominal diameters 3/8" to 2" (DN 8 to 50)
• Tube material: 904L SS

a0002271

Input
Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
• Fluid density (proportional to resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)

Measuring range Measuring ranges for liquids

Nominal Size Range for full scale values (liquids) gmin(F)...gmax(F)

1/24" (1) 0 to 0.73 lb/min (0 to 20 kg/h)

2 0 to 3.7 lb/min (0 to 100 kg/h)

4 0 to 16.5 lb/min (0 to 450 kg/h)

Measuring ranges for gases


The full scale values depend on the density of the gas. Use the formula below to calculate the full scale values:
gmax(G) = gmax(F) ⋅ ρ(G) / 2 [lb/ft3]
gmax(G) = max. full scale value for gas [lb/min]
gmax(F) = max. full scale value for liquid [lb/min]
ρ(G) = gas density in [lb/ft3] at process conditions
Here, gmax(G) can never be greater than gmax(F)

Calculation example for gas:


• Measuring device: Promass A, 1/12" nominal
• Gas: air with a density of 0.473 lb/ft3 (at 68°F and 145 psi)
• Measuring range: 3.7 lb/min
Max. possible full scale value:
gmax(G) = gmax(F) ⋅ ρ(G) / 2 (lb/ft3] = 3.7 lb/min ⋅ 0.44 lb/min / 2 lb/ft3 = 1.4 lb/min

Recommended measuring ranges:


See information in the “Limiting flow” Section → Page 19 ff.

Endress+Hauser 5
Proline Promass 80A, 83A

Operable flow range Greater than 1000 :1. Flow rates above the preset full scale value do not overload the amplifier, i.e. the totalizer
values are registered correctly.

Input signal Status input (auxiliary input):


U = 3 to 30 V DC, Ri = 5 kΩ, galvanically isolated.
Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start, batching
start/stop (optional)

Current input (only Promass 83)


Active/passive selectable, galvanically isolated, resolution: 2 µA
• Active: 4 to 20 mA, Ri ≤ 150 Ω, Uout = 24 V DC, short-circuit proof
• Passive: 0/4 to 20 mA, Ri ≤ 150 Ω, Umax = 30 V DC

Output
Output signal Promass 80

Current output:
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value
adjustable, Temperature coefficient: typ. 0.005% o.r./°C, resolution: 0.5 µA
• Active: 0/4 to 20 mA, RL < 700 Ω (at HART: RL ≥ 250 Ω)
• Passive: 4 to 20 mA; supply voltage VS 18 to 30 V DC; Ri ≥ 150 Ω

Pulse/frequency output:
Passive, open collector, 30 V DC, 250 mA, galvanically isolated.
• Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1, pulse width max. 2 s
• Pulse output: pulse value and pulse polarity can be selected, pulse width adjustable (0.5 to 2000 ms).

PROFIBUS PA interface:
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated
• Profile Version 3.0
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic) = 0 mA
• Data transmission rate: 31.25 kBit/s
• Signal encoding: Manchester II
• Function blocks: 4 x Analog Input, 1 x Totalizer
• Output data: Mass flow, Volume flow, Density, Temperature, Totalizer
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode,
Totalizer control
• Bus address can be set at the measuring device via miniature switches or the on-site display (optional)

6 Endress+Hauser
Proline Promass 80A, 83A

Promass 83

Current output:
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value
adjustable, Temperature coefficient: typ. 0.005% o.r./°C, resolution: 0.5 µA
• Active: 0/4 to 20 mA, RL < 700 Ω (at HART: RL ≥ 250 Ω)
• Passive: 4 to 20 mA; supply voltage VS 18 to 30 V DC; Ri ≥ 150 Ω

Pulse/frequency output:
active/passive selectable, galvanically isolated
• Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100 Ω
• Passive: open collector, 30 V DC, 250 mA
• Frequency output: full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ration 1:1, pulse width max.
2s
• Pulse output: pulse value and pulse polarity selectable, pulse width adjustable (0.05 to 2000 ms); the on/off
ratio is 1:1 as of a frequency of 1 / (2 x pulse width)

PROFIBUS DP interface:
• PROFIBUS DP in accordance with EN 50170 Volume 2, IEC 61158-2, galvanically isolated
• Profile Version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding: NRZ-Code
• Function blocks: 6 x Analog Input, 3 x Totalizer
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizer 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be set at the measuring device via miniature switches or the on-site display (optional)
• Available output combination→ Page 10

PROFIBUS PA interface:
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated
• Data transmission rate: 31.25 kBit/s
• Current consumption: 11 mA
• Permissible supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Signal encoding: Manchester II
• Function blocks: 6 x Analog Input, 3 x Totalizer
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizer 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be set at the measuring device via miniature switches or the on-site display (optional)
• Available output combination→ Page 10

MODBUS interface:
• MODBUS device type: slave
• Address range: 1 to 247
• Supported function codes: 03, 04, 06, 08, 16, 23
• Broadcast: supported with the function codes 06, 16, 23
• Physical interface: RS485 in accordance with EIA/TIA-485 standard
• Supported baudrate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
• Automatic data transmission rate recognition
• Transmission mode: RTU or ASCII
• Response times:
Direct data access = typically 25 to 50 ms
Auto-scan buffer (data range) = typically 3 to 5 ms
• Available output combination→ Page 10

Endress+Hauser 7
Proline Promass 80A, 83A

FOUNDATION Fieldbus interface:


• FOUNDATION Fieldbus H1, IEC 61158-2, galvanically isolated
• Data transmission rate: 31.25 kBit/s
• Current consumption: 12 mA
• Permissible supply voltage: 9 to 32 V
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Bus connection with integrated reverse polarity protection
• Signal encoding: Manchester II
• ITK Version 4.01
• Function blocks: 7 x Analog Input, 1 x Digital Output, 1 x PID
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density,
Temperature, Totalizer 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer reset
• Link Master function (LM) is supported

Signal on alarm Current output:


Failsafe mode selectable (e.g. in accordance with NAMUR Recommendation NE 43)
Pulse/frequency output:
Failsafe mode selectable

Status output:
“Non-conductive” in the event of a fault or if the power supply fails

Relay output (Promass 83):


“Dead” in the event of a fault or if the power supply fails

Load see “Output signal”

Low flow cut off Switch points for low-flow cutoff freely selectable.

Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.

Switching output Status output:


Open collector, max. 30 V DC / 250 mA, galvanically isolated
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction,
limit values

Relay output (only Promass 83):


Normally closed (NC or break) or normally open (NO or make) contacts available (factory setting:
relay 1 = NO, relay 2 = NC),
max. 30 V / 0.5 A AC; 60 V / 0.1 A DC, galvanically isolated.
Configurable for: error messages, Empty Pipe Detection (EPD), flow direction, limit values, filler valve 1 and 2
(optional).

8 Endress+Hauser
Proline Promass 80A, 83A

Power supply
Electrical connection
Measuring unit A B C

d d
g g
b b

a
a
a

b d/(g) (d)

HART* Profibus-PA*
FOUNDATION fieldbus

– 27 PA(–)/FF(–) 27
+ 26 PA(+)/FF(+) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 + 20

N (L-) 2 c N (L-) 2 c
L1 (L+)1 L1 (L+) 1
b b

Profibus-DP* Profibus-DP**
MODBUS RS485**

B (RS485) 27 B (RS485) 27
A (RS485) 26 A (RS485) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 g + 20 g

N (L-) 2 c N (L-) 2 c
L1 (L+) 1 L1 (L+) 1
b b

a0002441

Connecting the transmitter, cable cross-section: max. 18 AWG (2.5 mm2)


A View A (field housing)
B View B (stainless steel field housing)
C View C (wall-mount housing)
*) Fixed communication board
**) Flexible communication board
a Cover of the connection compartment
b Cable for power supply: 85 to 260 V AC, 20 to 55 V AC,16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
c Ground terminal for protective conductor
d Signal cable: see terminal assignment → Page 10
Fieldbus cable:
Terminal No. 26: DP (A) / PA (+) / FF (+) / MODBUS RS485 (A) / (PA, FF: with reverse polarity protection)
Terminal No. 27: DP (B) / PA (-) / FF (-) / MODBUS RS485 (B) / (PA, FF: with reverse polarity protection)
e Earth terminal, signal cable screen / fieldbus cable / RS485 line
f Service connector for connecting service interface FXA 193 (Fieldcheck, ToF Tool - Fieldtool Package)
g Signal cable: see terminal assignment → Page 10
Cable for external termination (only for PROFIBUS DP with fixed communication board):
Terminal No. 24: +5 V
Terminal No. 25: DGND

Endress+Hauser 9
Proline Promass 80A, 83A

Electrical connection, Promass 80


terminal assignment
Terminal No. (inputs/outputs)

Order variant 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)

80***-***********A - - Frequency output Current output, HART

80***-***********D Status input Status output Frequency output Current output, HART

80***-***********H - - - PROFIBUS PA

Frequency output Current output Ex i


80***-***********S - -
Ex i, passive Active, HART

Frequency output Current output Ex i


80***-***********T - -
Ex i, passive Passive, HART

80***-***********8 Status input Frequency output Current output 2 Current output 1,


HART

Promass 83
The inputs and outputs on the communication board can be either permanently assigned (fixed) or variable
(flexible), depending on the version ordered (see table). Replacements for modules which are defective or
which have to be replaced can be ordered as accessories.

Terminal No. (inputs/outputs)

Order variant 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)

Fixed communication boards (permanent assignment)

Current output
83***-***********A - - Frequency output
HART

Current output
83***-***********B Relay output Relay output Frequency output
HART

83***-***********F - - - PROFIBUS PA, Ex i

FOUNDATION
83***-***********G - - -
Fieldbus Ex i

83***-***********H - - - PROFIBUS PA

83***-***********J - - +5V (ext. termination) PROFIBUS DP

- FOUNDATION
83***-***********K - -
Fieldbus

83***-***********Q - - Status input MODBUS RS485

83***-***********R - - Current output 2 Current output 1


Ex i, active Ex i active, HART

Frequency output Current output Ex i


83***-***********S - -
Ex i, passive Active, HART

Frequency output Current output Ex i


83***-***********T - -
Ex i, passive Passive, HART

83***-***********U - - Current output 2 Current output 1


Ex i, passive Ex i passive, HART

Flexible communication boards

Current output
83***-***********C Relay output 2 Relay output 1 Frequency output
HART

Current output
83***-***********D Status input Relay output Frequency output
HART

Current output 1
83***-***********E Status input Relay output Current output 2
HART

Current output
83***-***********L Status input Relay output 2 Relay output 1
HART

10 Endress+Hauser
Proline Promass 80A, 83A

Terminal No. (inputs/outputs)

Order variant 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)

Current output
83***-***********M Status input Frequency output 2 Frequency output 1
HART

83***-***********N Current output Frequency output Status input MODBUS RS485

83***-***********P Current output Frequency output Status input PROFIBUS DP

83***-***********V Relay output 2 Relay output 1 Status input PROFIBUS DP

Current output 1
83***-***********W Relay output Current output 3 Current output 2
HART

Current output 1
83***-***********0 Status input Current output 3 Current output 2
HART

Current output 1
83***-***********2 Relay output Current output 2 Frequency output
HART

83***-***********3 Current input Relay output Current output 2 Current output 1


HART

83***-***********4 Current input Relay output Frequency output Current output


HART

83***-***********5 Status input Current input Frequency output Current output


HART

83***-***********6 Status input Current input Current output 2 Current output


HART

83***-***********7 Relay output 2 Relay output 1 Status input MODBUS RS485

Electrical connection Remote


version
S1 S1 S2 S2 TM TM TT TT
+ + GND + +

4 5 6 7 8 9 10 11 12 41 42

4 5 6 7 8 9 10 11 12 41 42
+ + + +
S1 S1 S2 S2 TM TM TT TT
GND
a0003681

a Wall-mount housing: Non-hazardous area and ATEX II3G (zone 2)


b Wall-mount housing: ATEX II2G (zone 1)
c Remote sensor flanged version
Wire color:
Terminal no.: 4/5 = grey; 6/7 = green; 8 = yellow; 9/10 = pink; 11/12 = white; 41/42 = brown

Supply voltage 85 to 260 V AC, 45 to 65 Hz


20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC

Cable entries Power-supply and signal cables (inputs/outputs):


• Cable entry M20 x 1.5 (8 to 12 mm)
• Thread for cable entries, 1/2" NPT, G 1/2"

Endress+Hauser 11
Proline Promass 80A, 83A

Connecting cable for remote version:


• Cable entry M20 x 1.5 (8 to 12 mm)
• Thread for cable entries, 1/2" NPT, G 1/2"

Remote version • 6 x 18 AWG (0.38 mm2) PVC cable with common shield and individually shielded cores
cable specifications • Conductor resistance: ≤ 50 Ω/km
• Capacitance: core/shield: ≤ 420 pF/m
• Cable length: max. 65 ft (20 m)
• Operating temperature: max. +220°F (105°C)
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC
requirements of EN 61326/A1, and NAMUR recommendation NE 21/43.

Power consumption AC: <15 VA (including sensor)


DC: <15 W (including sensor)
Switch-on current
• Max. 13.5 A (< 50 ms) at 24 V DC
• Max. 3 A (< 5 ms) at 260 V AC

Power supply failure Promass 80


Lasting min. 1 power cycle
• EEPROM saves measuring system data if the power supply fails
• S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial number,
calibration factor, zero point, etc.)

Promass 83
Lasting min. 1 power cycle:
• EEPROM or T-DAT save measuring system data if the power supply fails.
• S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial number,
calibration factor, zero point, etc.)

Potential equalization No measures necessary.

Performance characteristics
Reference operating Error limits following ISO/DIS 11631:
conditions • 68 to 86°F, 30 to 60 psi (20 to 30°C, 2 to 4 bar)
• Calibration systems as per national standards
• Zero point calibrated under operating conditions
• Field density calibrated (or special density calibration)

Maximum measured error The following values refer to the pulse/frequency output. The additional measured error at the current output
is typically ±5 µA.

Mass flow (liquid):

Promass 80A:
±0.15% ± [(zero point stability / measured value) x 100]% o.r.

Promass 83A:
±0.10% ± [(zero point stability / measured value) x 100]% o.r.

Mass flow (gas):

Promass 80A, 83A:


±0.50% ± [(zero point stability / measured value) x 100]% o.r.

12 Endress+Hauser
Proline Promass 80A, 83A

Volume flow (liquid)

Promass 80A, 83A:


±0.25% ± [(zero point stability / measured value) x 100]% o.r.
o.r. = of reading

Zero point stability (Promass A):

DN Max. full scale Zero point stability


value lb/min (kg/h)
lb/min (kg/h)

1/24" (1) 0.73 (20) 0.000036 (0.0010)

1/12" (2) 3.7 (100) 0.00018 (0.0050)

1/8" (4) 16.5 (450) 0.0008 (0.0225)

Sample calculation

[%]
±1.0

Promass 83
±0.5
Promass 80

±0.2

0
0 2 3.66 lb/min
(100 kg/h)

a0003401

Max. measured error in % of measured value (example: Promass 80A, 83A / 1/12" sensor)

Calculation example (mass flow, liquid):


Given: Promass 83A / 1/12", measured value flow = 1.47 lb/min
Max. measured error: ±0.10% ± [(zero point stability / measured value) x 100]% o.r.
Max. measured error → ±0.10% ±0.00018 lb/min ÷ 1.47 lb/min ⋅ 100% = ±0.012%

Density (liquid)

Standard calibration (1g/cc = 1 kg/l):


±0.02 g/cc

Special density calibration (optional)


±0.002 g/cc

After field density calibration or under reference conditions:


±0.0010 g/cc

Temperature
±33°F ±0.005 x T (T = medium temperature in °F)

Endress+Hauser 13
Proline Promass 80A, 83A

Repeatability Mass flow (liquid):


±0.05% ± [1/2 x (zero point stability / measured value) x 100]% o.r.

Mass flow (gas):


±0.25% ± [1/2 x (zero point stability / measured value) x 100]% o.r.

Volume flow (liquid):


±0.10% ± [1/2 x (zero point stability / measured value) x 100]% o.r.
o.r. = of reading
Zero point stability: see “Max. measured error”
Calculation example (mass flow, liquid):
Given: Promass 83A / 1/12", measured value flow = 1.47 lb/min
Repeatability: ±0.05% ± [1/2 x (zero point stability / measured value) x 100]% o.r.
Repeatability → ±0.05% ±1/2 ⋅ 0.00018 lb/min ÷ 1.47 lb/min ⋅ 100% = ±0.056%

Density measurement (liquid)


±0.0005 g/cc

Temperature measurement
±32°F ±0.0025 x T (T = medium temperature in °F)

Influence of medium When there is a difference between the temperature for zero point adjustment and the process temperature,
temperature the typical measured error of the Promass sensor is ±0.0001% of the full scale value / °F.

Influence of medium pressure A difference in pressure between the calibration pressure and the process pressure does not have any effect on
the accuracy.

Operating conditions: Installation


Installation instructions Note the following points:
• No special measures such as supports are necessary. External forces are absorbed by the construction of the
instrument, for example the secondary containment.
• The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring
system is not influenced by pipe vibrations.
• No special precautions need to be taken for fittings which create turbulence (valves, elbows, T-pieces, etc.),
as long as no cavitation occurs.
• For mechanical reasons and to protect the pipe, support is recommended for heavy sensors.

14 Endress+Hauser
Proline Promass 80A, 83A

Mounting location
Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors.
For this reason, avoid the following mounting locations in the pipe:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream from a free pipe outlet in a vertical pipeline.

a0003605

Mounting location

Notwithstanding the above, the installation proposal below permits installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the
sensor running empty while measurement is in progress.

a0003597

Installation in a down pipe (e.g. for batching applications)


1 Supply tank
2 Sensor
3 Orifice plate, pipe restriction (see Table)
4 Valve
5 Batching tank

Nominal Size 1/24" 1/12" 1/8"

∅ Orifice plate, pipe restriction 0.03" 0.06" 0.12"

Endress+Hauser 15
Proline Promass 80A, 83A

Orientation
Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of flow
(direction of fluid flow through the pipe).

Vertical
Recommended orientation with direction of flow upwards. When fluid is not flowing, entrained solids will sink
down and gases will rise away from the measuring tube. The measuring tubes can be completely drained and
protected against solids build-up.

Horizontal
When installation is correct the transmitter housing is above or below the pipe. This means that no gas bubbles
or solids deposits can form in the bent measuring tube (single-tube system).
The sensor cannot be installed suspended (i.e. without support or holder) in the pipe. This prevents excessive
material load in the area of the process connection. The base plate of the sensor housing allows bench, wall or
post mounting.

a0003606

Vertical and horizontal orientation (Promass A)

Heating
Some fluids require suitable measures to avoid heat transfer at the sensor. Heating can be electric, e.g. with
heated elements, or by means of hot water or steam pipes made of copper.
Caution!
• Risk of electronics overheating! Consequently, make sure that the adapter between the sensor and
transmitter and the connection housing of the remote version always remain free of insulating material. Note
that a certain orientation might be required, depending on the fluid temperature .
• If using an electric trace heating system whose heating is regulated via phase angle control or pulse packages,
influence on the measured values cannot be ruled out due to magnetic fields (i.e. for values that are greater
than the values approved by the EN standard (sine 30 A/m)). In such instances, it is necessary to
magnetically screen the sensor.
The secondary containment can be screened with tin plate or electric sheets without privileged direction
(e.g. V330-35A) with the following properties:
– Relative magnetic permeability µr ≥ 300
– Plate thickness d ≥ 0.014" (0.35 mm)
• Information on permitted temperature ranges → Page 19
Special heating jackets, which can be ordered separately from Endress+Hauser as an accessory, are available
for the sensors.

16 Endress+Hauser
Proline Promass 80A, 83A

Zero point adjustment


All Promass devices are calibrated to state-of-the-art technology. The zero point determined in this way is
imprinted on the nameplate. Calibration takes place under reference conditions. → Page 12 ff.
For this reason, Promass generally does not require zero point adjustment!
Experience shows that the zero point adjustment is advisable only in special cases:
• To achieve highest measuring accuracy also with very small flow rates.
• Under extreme process or operating conditions (e.g. very high process temperatures or very high-viscosity
fluids).
Please note the following before carrying out the adjustment:
• The adjustment can only be performed with fluids that have no gas or solid contents.
• Zero point adjustment is performed with the measuring tubes completely filled and at zero flow (v = 0 m/
s). This can be achieved, for example, with shut-off valves upstream and/or downstream of the sensor or by
using existing valves and gates.
– Normal operation → valves 1 and 2 open
– Zero point adjustment with pump pressure → valve 1 open / valve 2 closed
– Zero point adjustment without pump pressure → valve 1 closed / valve 2 open

a0003601

Zero point adjustment and shut-off valves

Inlet and outlet run There are no installation requirements regarding inlet and outlet runs.

Length of connecting cable Max. 65 ft (20 meters), remote version

System pressure It is important to ensure that cavitation does not occur, because it would influence the oscillation of the
measuring tube. No special measures need to be taken for fluids which have properties similar to water under
normal conditions.
In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction lines, it is
important to ensure that pressure does not drop below the vapor pressure and that the liquid does not start to
boil. It is also important to ensure that the gases that occur naturally in many liquids do not outgas. Such effects
can be prevented when system pressure is sufficiently high.
Consequently, it is generally best to install the sensor:
• Downstream from pumps (no risk of partial vacuum)
• At the lowest point in a vertical pipe

Endress+Hauser 17
Proline Promass 80A, 83A

Operating conditions: Environment

Ambient temperature range Standard: -4 to +140°F (–20 to +60°C), sensor and transmitter
Optional: -40 to +140°F (–40 to +60°C), sensor and transmitter
Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below -4°F (–20°C) the readability of the display may be impaired.

Storage temperature -40 to +175°F (–40 to +80°C, preferably +68°F (20°C)

Degree of protection Standard: NEMA 4X (IP 67) for transmitter and sensor

Shock resistance In accordance with IEC 68-2-31

Vibration resistance Acceleration up to 1 g, 10 to 150 Hz, following IEC 68-2-6

Electromagnetic compatibility To EN 61326 / A1 and NAMUR recommendation NE 21


(EMC)

18 Endress+Hauser
Proline Promass 80A, 83A

Operating conditions: Process

Medium temperature range Sensor


-60 to +400°F (–50 to +200°C)

Seals:
(Only for mounting kits with threadedconnections)
Viton 5 to 400°F (–15 to 200°C); EPDM -40 to +320°F (–40 to +160°C); silicone -75 to +400°F (–60 to
+200°C); Kalrez -4 to +528°F (–20 to +275°C)

Medium pressure range Threaded joints:


(nominal pressure)
Max. 2300 psi (160 bar), standard versions; max. 5800 psi (400 bar), high pressure versions

Flanges:
ANSI Cl 150, Cl 300 / DIN PN 40 to 100 / JIS 10K

Pressure ranges of secondary containment:


375 psi (25 bar)
Warning!
In case a danger of measuring tube failure exists due to process characteristics, e.g. with corrosive process
fluids, we recommend the use of sensors whose secondary containment is equipped with special pressure
monitoring connections (ordering option). With the help of these connections, fluid collected in the secondary
containment in the event of tube failure can be bled off. This is especially important in high pressure gas
applications. These connections can also be used for gas circulation and/or detection. Dimensions

Limiting flow See information in the “Measuring range” Section → Page 5


Select nominal diameter by optimizing between required flow range and permissible pressure loss. An overview
of max. possible full scale values can be found in the “Measuring range” Section.
• The minimum recommended full scale value is approx. 1/20 of the max. full scale value.
• In most applications, 20 to 50% of the maximum full scale value can be considered ideal.
• Select a lower full scale value for abrasive substances such as fluids with entrained solids,
flow velocity < 3.3 ft/s (m/s).
• For gas measurement the following rules apply:
– Flow velocity in the measuring tubes should not be more than half the sonic velocity (0.5 Mach).
– The maximum mass flow depends on the density of the gas: formula → Page 5

Endress+Hauser 19
Proline Promass 80A, 83A

Pressure loss Pressure loss depends on the fluid properties and on the flow rate. The following formulae can be used to
approximately calculate the pressure loss:

4·g
Reynolds number Re =
·d··
a0003381

Re ≥ 23001) p = K · 0.25 · g1.75 · –0.75


a0003380

Re < 2300 p = K1 ·  · g
a0003379

3
∆p = pressure loss [mbar] ρ = density [kg/m ]
v = kinematic viscosity [m2/s] d = inside diameter of measuring tubes [m]
g = mass flow [kg/s] K to K1 = constant (depends on nominal diameter)

1) To compute the pressure loss for gases, always use the formula for Re ≥ 2300.

Pressure loss coefficients for Promass A

DN d [m] K K1

1 1.1 ⋅ 10 -3
1.2 ⋅ 10 11
1.3 ⋅1011

2 1.8 ⋅ 10-3 1.6 ⋅ 1010 2.4 ⋅ 1010

4 3.5 ⋅ 10-3 9.4 ⋅ 108 2.3 ⋅ 109

High pressure version

2 1.4 ⋅ 10-3 5.4 ⋅ 1010 6.6 ⋅ 1010

4 3.0 ⋅ 10-3 2.0 ⋅ 109 4.3 ⋅ 109

[mbar]
10000
DN 2 DN 4

DN 1

1000

100

10

1
0.1 1 10 100 1000 [kg/h]

1
2
a0003595

Pressure loss diagram for water


1 Standard version
2 High pressure version

20 Endress+Hauser
Proline Promass 80A, 83A

Mechanical construction
Design / dimensions Dimensions: Wall-mount housing (non hazardous area and II3G / zone 2)

6.27"
Esc
(159.5)
- + E

9.84"
(250)

3.56"
(90.5)

8.46" 3.54" 1.77"


(215) (90) (45)
5.31"
(135)

3.18" 2.08" 3.18"


> 1.97" (81) (53) (81)
(50)

4.01"
(102)

2.08"
(53)

3.74"
3.20" (95)
(81.5)

0.45" 7.55" 0.45"


(11.5) (192) (11.5)

a0001150

Dimensions: Remote field housing (II2G / zone 1)

10.4"
(265)
9.53" *
(242)
9.45” 8.11”
(240) (206)
8.54" * 7.32"
(217) (186)

6.57"
(167)

13.9"
(355)

7.40"
3.94" 5.24" (188)
(100) (133)

* Blind version  0.34"


3.94"
(8.6 M8)
(100)
4.84"
(123)

a0002128

Endress+Hauser 21
Proline Promass 80A, 83A

Dimensions: stainless steel field housing

8.66” 6.02"
(220) (153)

Esc
6.73"
(171)
- + E

a0002245

Dimensions: stainless steel field housing

Dimensions: Remote version

4.67" 5.41"
(118.5) (137.5)
Nominal size T
1/24" 4.72" (120)
1/12" 4.72" (120)
T 1/8" 5.12" (130)

T = Dimension "B - 6.02” (153 mm)" for the corresponding


nominal diameter from the compact version dimensions.
a0002516

22 Endress+Hauser
Proline Promass 80A, 83A

Dimensions: 4-VCO-4-connection (welded)

8.94" 7.68"
(227) (195)
G U
8.15" * 6.61"
(207) (168)

6.30"
(160)

B A

di 4 x Æ 0.26"
(6.5)

F
P
E C
K
+0.06” (1.5)
-0.08” (2.0)
M L

* Blind version

a0003165

4-VCO-4-connection: 1.4539/904L SS, Alloy C-22

A
Size B C E F K L M P U / di
in G
(DN) in (mm) in (mm) in (mm) in (mm) in (mm) in (mm) in (mm) in (mm) in (mm)
(mm)

1/24" 12 10.7 1.26 8.98 6.30 5.71 11.4 6.50 4.72 0.04
AF 11/16"
(1) 1) (305) (273) (32) (228) (160) (145) (290) (165) (120) (1.1)

1/12" 12 10.7 1.26 12.2 6.30 5.71 14.6 6.50 4.72 0.07
AF 11/16"
(2) 1) (305) (273) (32) (310) (160) (145) (372) (165) (120) (1.8)

1/12" 12 10.7 1.26 12.2 6.30 5.71 14.6 6.50 4.72 0.06
AF 11/16"
(2) 2) (305) (273) (32) (310) (160) (145) (372) (165) (120) (1.4)

1/8" 12.4 11.1 1.26 17.1 8.67 6.89 19.6 7.68 5.90 0.14
AF 11/16"
(4) 1) (315) (283) (32) (435) (220) (175) (497) (195) (150) (3.5)

1/8" 12.4 11.1 1.26 17.1 8.67 6.89 19.6 7.68 5.90 0.12
AF 11/16"
(4) 2) (315) (283) (32) (435) (220) (175) (497) (195) (150) (3.0)
1)
3-A version can be supplied (Ra ≤ 16 µin / 0.4 µm). Only for 1.4539/904L SS
2) High pressure version

Endress+Hauser 23
Proline Promass 80A, 83A

Dimensions: 1/2" Tri-Clamp connection (welded)

8.94" 7.68"
(227) (195)
8.15" * 6.61"
(207) (168)
U G

6.30"
(160)

B A

di 4 x Æ 0.26"
(6.5)

F
P C
E
K
+0.06” (1.5)
M L -0.08” (2.0)

* Blind version

a0003183

1/2" Tri-Clamp connection / 3-A version 1):1.4539/904L SS

A B C E F G K L M P U di
Size
in in in in in in in in in in in in
(DN)
(mm) (mm) (mm) (mm) (mm) (mm) (mm) (mm) (mm) (mm) (mm) (mm)

1/24" 12.0 10.7 1.26 8.98 6.30 0.98 5.71 11.7 6.50 4.72 0.37 0.04
(1) (305) (273) (32) (228) (160) (25) (145) (296) (165) (120) (9.5) (1.1)

1/12" 12.0 10.7 1.26 12.2 6.30 0.98 5.71 14.9 6.50 4.72 0.37 0.007
(2) (305) (273) (32) (310) (160) (25) (145) (378) (165) (120) (9.5) (1.8)

1/8" 12.4 11.1 1.26 17.1 8.67 0.98 6.89 19.8 7.68 5.90 0.37 0.14
(4) (315) (283) (32) (435) (220) (25) (175) (503) (195) (150) (9.5) (3.5)
1)
3-A version (Ra ≤ 32 µin / 0.8 µm, option: Ra ≤ 16 µin / 0.4 µm).

24 Endress+Hauser
Proline Promass 80A, 83A

Dimensions 4-VCO-4-connection with mounting kit:


DN 15 flange EN 1092-1 (DIN 2501), JIS or 1/2" flange ANSI

8.94" LK 7.68"
(227) G (195)
8.15" * 6.61"
(207) U (168)
S

6.30"
(160)

B A

di 4 x Æ 0.26"

F
P
E C
K +0.06” (1.5)
-0.08” (2.0)
M L

* Blind version

a0003184

Mounting kit DN 15 flange EN 1092-1 (DIN 2501) PN 40: 1.4539/904L SS, Alloy C-22

DN A B C E F G K L M N P S LK U di

1 305 273 32 228 160 95 145 393 165 4 x Ø14 120 28 65 17.3 1.1

2 305 273 32 310 160 95 145 475 165 4 x Ø14 120 28 65 17.3 1.8

4 315 283 32 435 220 95 175 600 195 4 x Ø14 150 28 65 17.3 3.5

Loose flanges (not wetted) made of stainless steel 1.4404/316L

Mounting kit DN 15 flange (JIS) 10K: 1.4539/904L SS, Alloy C-22

DN A B C E F G K L M N P S LK U di

1 305 273 32 228 160 95 145 393 165 4 x Ø15 120 28 70 15.0 1.1

2 305 273 32 310 160 95 145 475 165 4 x Ø15 120 28 70 15.0 1.8

4 315 283 32 435 220 95 175 600 195 4 x Ø15 150 28 70 15.0 3.5

Loose flanges (not wetted) made of stainless steel 1.4404/316L

Mounting kit DN 15 flange (JIS) 20K: 1.4539/904L SS, Alloy C-22

DN A B C E F G K L M N P S LK U di

1 305 273 32 228 160 95 145 393 165 4 x Ø15 120 14 70 15.0 1.1

2 305 273 32 310 160 95 145 475 165 4 x Ø15 120 14 70 15.0 1.8

4 315 283 32 435 220 95 175 600 195 4 x Ø15 150 14 70 15.0 3.5

Endress+Hauser 25
Proline Promass 80A, 83A

Mounting kit 1/2" flange (ANSI) Cl 150:1.4539/904L SS, Alloy C-22 (dimensions in inches)

Size A B C E F G K L M N P S LK U di

1 1/24" 12.0 10.7 1.26 8.98 6.30 3.5 5.71 15.5 6.50 4 x Ø0.62 4.72 0.70 2.38 0.62 0.04

2 1/12" 12.0 10.7 1.26 12.2 6.30 3.5 5.71 18.7 6.50 4 x Ø0.62 4.72 0.70 2.38 0.62 0.07

4 1/8" 12.4 11.1 1.26 17.1 8.67 3.5 6.89 23.6 7.68 4 x Ø0.62 5.90 0.70 2.38 0.62 0.14

Loose flanges (not wetted) made of stainless steel 1.4404/316L

Mounting kit 1/2" flange (ANSI) Cl 300:1.4539/904L SS, Alloy C-22 (dimensions in inches)

Size A B C E F G K L M N P S LK U di

1 1/24" 12.0 10.7 1.26 8.98 6.30 3.7 5.71 15.5 6.50 4 x Ø0.62 4.72 0.81 2.62 0.62 0.04

2 1/12" 12.0 10.7 1.26 12.2 6.30 3.7 5.71 18.7 6.50 4 x Ø0.62 4.72 0.81 2.62 0.62 0.07

4 1/8" 12.4 11.1 1.26 17.1 8.67 3.7 6.89 23.6 7.68 4 x Ø0.62 5.90 0.81 2.62 0.62 0.14

Loose flanges (not wetted) made of stainless steel 1.4404/316L

Dimensions 4-VCO-4-connection with mounting kit: 1/4" NPT-F

8.94" U 7.68"
(227) (195)
8.15" * 6.61"
(207) G
(168)

6.30"
(160)

B A

di 4 x Æ 0.26"
(6.5)

F
P
E C
K +0.06” (1.5)
-0.08” (2.0)
M L

* Blind version

a0003185

Mounting kit 1/4" NPT-F connection: 1.4539/904L SS, Alloy C-22 (dimensions in inches)

Size A B C E F G K L M P U di

1/24" 12.0 10.7 1.26 8.98 6.30 AF 3/4" 5.71 11.7 6.50 4.72 1/4" NPT 0.04

1/12" 12.0 10.7 1.26 12.2 6.30 AF 3/4" 5.71 14.9 6.50 4.72 1/4" NPT 0.07
1)
1/12" 12.0 10.7 1.26 12.2 6.30 AF 3/4" 5.71 14.9 6.50 4.72 1/4" NPT 0.06

1/8" 12.4 11.1 1.26 17.1 8.67 AF 3/4" 6.89 19.8 7.68 5.90 1/4" NPT 0.14
1)
1/8" 12.4 11.1 1.26 17.1 8.67 AF 3/4" 6.89 22.4 7.68 5.90 1/4" NPT 0.12
1)
High pressure version only available as 1.4539/904L SS

26 Endress+Hauser
Proline Promass 80A, 83A

Dimensions 4-VCO-4-connection with mounting kit: 1/8" or 1/4" SWAGELOK

8.94" 7.68"
(227) (195)
8.15" * 6.61"
(207) G
(168)

U
6.30"
(160)

B A

di 4 x Æ 0.26"

F
P C
E
K +0.06” (1.5)
-0.08” (2.0)
M L

* Blind version

a0003186

Mounting kit SWAGELOK connection: 1.4539/904L (dimensions in inches)

DN A B C E F G K L M P U di

1 12.0 10.7 1.26 8.98 6.30 SW 7/16" 5.71 14.2 6.50 4.72 1/8" 0.04

1 12.0 10.7 1.26 8.98 6.30 SW 9/16" 5.71 14.2 6.50 4.72 1/4" 0.04

2 12.0 10.7 1.26 12.2 6.30 SW 7/16" 5.71 17.4 6.50 4.72 1/8" 0.07

2 12.0 10.7 1.26 12.2 6.30 SW 9/16" 5.71 17.4 6.50 4.72 1/4" 0.07

21) 12.0 10.7 1.26 12.2 6.30 SW 7/16" 5.71 17.4 6.50 4.72 1/8" 0.06
1)
2 12.0 10.7 1.26 12.2 6.30 SW 9/16" 5.71 17.4 6.50 4.72 1/4" 0.06

4 12.4 10.7 1.26 17.1 8.67 SW 9/16" 6.89 22.5 7.68 5.90 1/4" 0.14

41) 12.4 10.7 1.26 17.1 8.67 SW 9/16" 6.89 22.5 7.68 5.90 1/4" 0.12
1)
High pressure version

Endress+Hauser 27
Proline Promass 80A, 83A

Purge connections / pressure vessel monitoring


Caution!
The pressure vessel is filled with dry nitrogen (N2). Do not open the purge connections unless the containment
can be filled immediately with a dry inert gas. Use only low gauge pressure to purge. Maximum pressure:
70 psi (5 bar).

L L

a0003187

Size L H G

1/24" 3.62" (92.0 mm) 4.02" (102.0 mm) 1/2" NPT

1/12" 5.12" (130.0 mm) 4.02" (102.0 mm) 1/2" NPT

1/8" 7.58" (192.5 mm) 7.58" (192.5 mm) 1/2" NPT

Weight • Compact version: see table below


• Remote version
– Sensor: see table below
– Wall-mount housing: 11 lbs (5 kg)

Promass A / Nominal size 1/24" 1/12" 1/8"

Compact version, lbs (kg) 22 (10) 24 (11) 33 (15)

Remote version, lbs (kg) 17 (8) 20 (9) 29 (13)

Weight information in lbs [kg].


All values (weight) refer to devices with 1/2" flanges.

Material Transmitter housing:


• Compact housing: stainless steel 1.4301/304
• Compact housing: powder coated die-cast aluminum
• Wall-mount housing: powder coated die-cast aluminum
• Remote field housing: powder-coated die-cast aluminum

Sensor housing / containment:


Acid and alkali-resistant outer surface
Stainless steel 1.4301/304

Connection housing, sensor (remote version):


• Stainless steel 1.4301/304

28 Endress+Hauser
Proline Promass 80A, 83A

Process connections
• Mounting kit for flanges EN 1092-1 (DIN 2501) / ANSI B16.5 / JIS B2238 → stainless steel 1.4539/904L,
Alloy C-22 2.4602/N 06022.
Loose flanges → stainless steel 1.4404/316L
• VCO coupling → stainless steel 1.4539/904L, Alloy C-22 2.4602/N 06022
• Tri-Clamp (OD tubes) (1/2") → stainless steel 1.4539/904L
• Mounting kit for SWAGELOK (1/4", 1/8") → stainless steel 1.4401/316
• Mounting kit for NPT-F (1/4") → stainless steel 1.4539/904L1.4539/904L,
Alloy C-22 2.4602/N 06022

Measuring tube(s):
• Stainless steel 1.4539/904L, Alloy C-22 2.4602/N 06022

Seals:
Welded process connections without internal seals

Material load curves Flange connections to EN 1092-1 (DIN 2501) (mounting kit)
Wetted parts (flange, measuring tube): 1.4539/904L, Alloy C-22
Loose flanges (not wetted): 1.4404/316L

[bar]

50

40
PN 40
30

20
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
a0003284

Flange connections to ANSI B16.5 (mounting kit)


Wetted parts (flange, measuring tube): 1.4539/904L, Alloy C-22
Loose flanges (not wetted): 1.4404/316L

600
ANSI Class 300
Pressure (psi)

400

200
ANSI Class 150

0
-80 -40 0 40 80 120 160 200 240 280 320 360 400
Temperature (°F) 392

a0003285

Flange connections to JIS B2238 (mounting kit)


Wetted parts (flange, measuring tube): 1.4539/904L, Alloy C-22
Loose flanges (not wetted): 1.4404/316L

[bar]

20

10
10K
0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
a0003312

Endress+Hauser 29
Proline Promass 80A, 83A

Promass A: Tri-Clamp process connection


The load limit is defined exclusively by the material properties of the outer clamp used. This clamp is not
included in the scope of delivery.

Process connection: 4-VCO-4, 1/4" NPT-F, SWAGELOK


• 4-VCO-4-coupling (welded): 1.4539/904L, Alloy C-22
• 1/4" NPT threaded adapter (screwed, mounting kit): 1.4539/904L, Alloy C-22
• 1/4" or 1/8" SWAGELOK threaded joint (screwed, mounting kit): 1.4401/316

6000

1/24", Alloy C-22

5000

4000 1/24", 904L SS


Pressure (psi)

3000

2000 1/12" - 1/8", 904L SS / Alloy C-22

1000
-80 -40 0 40 80 120 160 200 240 280 320 360 400
392
Temperature (°F)
a0003309

Process connections for high pressure version (DN 2...4)


• 4-VCO-4-coupling (welded): 1.4539/904L, Alloy C-22
• 1/4" NPT threaded adapter (screwed, mounting kit): 1.4539/904L
• 1/4" or 1/8" SWAGELOK threaded joint (screwed, mounting kit): 1.4401/316

6000
1/12" - 1/8", 904L SS
Pressure (psi)

5000

4000
-80 -40 0 40 80 120 160 200 240 280 320 360 400
Temperature (°F) 392

a0003310

Process connections • Welded process connections: 4-VCO-4 coupling, 1/2" Tri-Clamp


• Threaded on process connections: flanges EN 1092-1 (DIN 2501), ANSI, JIS;
1/4" NPT threaded adapter; 1/8" or 1/4" SWAGELOK threaded joints

30 Endress+Hauser
Proline Promass 80A, 83A

Human interface
Display elements • Liquid-crystal display: backlit, two lines (Promass 80) or four lines (Promass 83) with 16 characters per line
• Selectable display of different measured values and status variables
• At ambient temperatures below -4°F (–20°C) the readability of the display may be impaired.

Unified control concept for Promass 80:


both types of transmitter:
• Local operation with three keys (–, +, E)
• Quick Setup menus for straightforward commissioning

Promass 83:
• Local operation with three optical keys (–, +, E)
• Application specific Quick Setup menus for straightforward commissioning

Language groups Language groups available for operation in different countries:


• Western Europe and America (WEA):
English, German, Spanish, Italian, French, Dutch and Portuguese
• Eastern Europe/Scandinavia (EES):
English, Russian, Polish, Norwegian, Finnish, Swedish and Czech
• South and Eastern Asia (SEA):
English, Japanese, Indonesian
Only Promass 83:
• China (CN):
English, Chinese
The language group is changed using the “ToF Tool - Fieldtool Package” operating program.

Remote operation Promass 80:


Operation via HART, PROFIBUS PA

Promass 83:
Operation via HART, PROFIBUS PA/DP, FOUNDATION Fieldbus

Certificates and approvals


CE mark The measuring system is in conformity with the statutory requirements of the EC Directives.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.

C-Tick mark The measuring system is in conformity with the EMC requirements of the Australian Communications
Authority (ACA).

Ex approval Information about currently available Ex versions (ATEX, FM, CSA) can be supplied by your Endress+Hauser
Sales Center on request. All explosion protection data are given in a separate documentation which is available
upon request.

Sanitary compatibility • 3-A approval


• EHEDG-tested

FOUNDATION Fieldbus The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification Fieldbus FOUNDATION. The device thus meets all the requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1.
• Interoperability Test Kit (ITK), revision status 4.0 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus FOUNDATION

Endress+Hauser 31
Proline Promass 80A, 83A

PROFIBUS DP/PA The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification PNO (PROFIBUS User Organisation). The device thus meets all the requirements of the following
specifications:
• Certified to PROFIBUS Profile Version 3.0 (device certification number: available on request)
• The device can also be operated with certified devices of other manufacturers (interoperability)

MODBUS certification The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
“MODBUS/TCP Conformance Test Policy, Version 2.0”. The measuring device has successfully passed all the
test procedures carried out and is certified by the “MODBUS/TCP Conformance Test Laboratory” of the
University of Michigan.

Other standards and • EN 60529


guidelines Degrees of protection by housing (IP code).
• EN 61010
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory
Procedures.
• EN 61326/A1 (IEC 1326)
“Emission in accordance with Class A requirements”.
Electromagnetic compatibility (EMC requirements).
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NAMUR NE 43
Standardisation of the signal level for the breakdown information of digital transmitters with analogue output
signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics

Pressure measuring device Flow meters with a nominal diameter smaller or equal to 1" (DN 25) are covered by Art. 3(3) of the European
approval directive 97/23/EC (Pressure Equipment Directive) and are designed according to sound engineering practice.
For larger nominal diameters, optional approvals according to Cat. II/III are available when required (depends
on fluid and process pressure).

Functional safety SIL-2:


In accordance with IEC 61508/IEC 61511-1 (FDIS)
4 to 20 mA output according to the following order code:
Promass 80***-***********A
Promass 80***-***********D
Promass 83***-***********A
Promass 83***-***********B

Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor. The Endress+Hauser service organisation can provide detailed information on request.

32 Endress+Hauser
Proline Promass 80A, 83A

Documentation
❑ Technical Information Promass 80F, 80M, 83F, 83M (TI053D/24/ae)
❑ Technical Information Promass 80H, 80I, 83H, 83I (TI052D/24/ae)
❑ Technical Information Promass 80E, 83E (TI061D/24/ae)
❑ Operating Instructions Promass 80 (BA 057D/06/en)
❑ Operating Instructions Promass 80 PROFIBUS PA (BA072D/06/en)
❑ Operating Instructions Promass 83 (BA 059D/06/en)
❑ Operating Instructions Promass 83 FOUNDATION Fieldbus (BA065D/06/en)
❑ Operating Instructions Promass 83 PROFIBUS DP/PA (BA063D/06/en)
❑ Operating Instructions Promass 83 MODBUS (BA107D/06/en)
❑ Description of Device Functions Promass 80 (BA 058D/06/en)
❑ Description of Device Functions Promass 80 PROFIBUS PA (BA073D/06/en)
❑ Description of Device Functions Promass 83 (BA 060D/06/en)
❑ Description of Device Functions Promass 83 FOUNDATION Fieldbus (BA066D/06/en)
❑ Description of Device Functions Promass 83 PROFIBUS DP/PA (BA064D/06/en)
❑ Description of Device Functions Promass 83 MODBUS (BA108D/06/en)
❑ Supplementary documentation on Ex-ratings: ATEX, FM, CSA
❑ Functional safety manual Promass 80, 83 (SD077D/06/en)

Registered trademarks
KALREZ® and VITON®
Registered trademarks of Firma E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP ®
Registered trademark of Ladish & Co., Inc., Kenosha, USA
SWAGELOK®
Registered trademark of Swagelok & Co., Solon, USA
HART®
Registered trademark of HART Communication Foundation, Austin, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organisation, Karlsruhe, Germany
FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA
MODBUS®
Registered trademark of the MODBUS Organisation
HistoROM™, S-DAT®, T-DAT™, F-CHIP®, ToF Tool - Fieldtool® Package, Fieldcheck®,
Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH

Endress+Hauser 33
Proline Promass 80A, 83A

Ordering Information

NOTE: Endress+Hauser reserves the right to change or modify product, specifications, and ordering information at any time without notice.
Please consult Endress+Hauser or your local representative for the most recent information.

Promass 80 A

1 2 3 4 5 6 7 8 9 10 11
Promass 80 A -

Nominal Diameter 7 Protection


1 01 1/24”, 0.73 lb/min max. full scale A NEMA 4X (IP 67) compact aluminum housing
02 1/12”, 3.7 lb/min max. full scale B NEMA 4X (IP 67) compact stainless steel housing (only for
04 1/8” 16.5 lb/min max. full scale approvals A and R)
2 Measuring system C NEMA 4X (IP 67) wall-mounted housing with 30 ft. (10 m) of cable,
A Standard version, 904L SS measuring tube remote version (only for approvals A and R)
B Standard version, Alloy C-22 measuring tube D NEMA 4X (IP 67) wall-mounted housing, 65 ft (20 m) of cable,
C 3.1B wetted parts, 904L SS measuring tube remote version
D 3.1B wetted parts, Alloy C-22 measuring tube (only for approvals A and R)
G High pressure version, 904L SS measuring tube (not for 1/24” size) E NEMA 4X (IP 67) aluminum field housing, 30 ft. (10 m) of cable,
K High pressure version, 3.1B wetted parts, 904L SS measuring tube remote version
(not for 1/24” size) (not for approvals A and R)
F NEMA 4X (IP 67) aluminum field housing, 65 ft. (20 m) of cable,
3 Process Connections
remote version
SVW Cajon 4-VCO-4 couplings
1 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact,
PPD Cajon 4-VCO-4 couplings, 240 grit (not for Alloy C-22 tube)
aluminum field housing (only for approvals A, N, P and R)
STA 1/2” Tri-clamp couplings, 150 grit, 3-A version (not for Alloy C-22
2 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact,
tube or high pressure version)
stainless steel housing (only for approvals A and R)
SPD 1/2” Tri-clamp couplings, 240 grit, 3-A version (not for Alloy C-22
3 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67), remote version,
tube or high pressure verison)
wall mount housing, 65 ft (20 m) cable (only for approvals A and R)
4 Test / Treatment 7 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67) remote field
(Test pressure of the measuring system to 1.5 times nominal pressure housing, 65 ft (20 m) cable (only for approval N, P)
rating of the selected process connections)
8 Cable entries
A No special test or treatment
B 1/2” NPT
B Pressure tested, 2.3 pressure test certificate and type test
L 1/2” NPT fieldbus connector (only for approval A)
of secondary containment
C Oil and grease-free process wetted surface 9 Power Supply / Display
D Oil and grease-free process wetted surface, 2.3 pressure 7 85 to 260 VAC, without display, remote configuration only (not for
test certificate and wall mount or SS housing)
type test of secondary containment 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
F Containment with purge connection, 2.3 pressure test certificate and only (not for wall mount or SS housing)
type test of secondary A 85 to 260 VAC, with display, push button operation (language: EN,
ES, FR, IT, NL, PT, DE)
5 Calibration
B 20 to 55 VAC / 16 to 62 VDC, with display, push button operation
A 0.15% flow, standard density calibration (language: EN, ES, FR, IT, NL, PT, DE)
B 0.15% flow, 0.002 g/cc density calibration X Sensor only
C 5-point calibration, 0.15% flow (specify flow range), standard
10 Software
density calibration
A Standard software
D 5-point calibration, 0.15% flow (specify flow range), 0.002 g/cc
X Sensor only
density calibration
G A2LA/SCS 5-point, 0.15% flow; 0.02 g/cc density, with flow 11 Outputs / Inputs
certificate traceable according to ISO 9000 (ISO/EIC 17025) A Current HART, frequency
Specify flow range (not for 1/24” sensor) D Current HART, frequency, status output, status input
H A2LA/SCS 5-point, 0.15% flow; 0.002 g/cc density, with flow H PROFIBUS-PA (approvals A and R only)
certificate traceable according to ISO 9000 (ISO/EIC 17025) S Current HART, frequency; active I.S. (not for approvals A and R)
Specify flow range (not for 1/24” sensor) T Current HART, frequency; passive I.S. (not for approvals A and R)
8 Current HART, current output, status input, frequency output
6 Approvals
X Sensor only
A General purpose, for use in non-hazardous locations
N FM explosion proof, Class I, Div. 1, Grps A-D; CSA approved Class I,
Div. 1, Grps A-D and zone 1 (only for aluminum field housing)
R FM non-incendive Class I, Div. 2, Grps A-D; CSA approved Class I,
Div. 2, Grps A-D and zone 2

34 Endress+Hauser
Proline Promass 80A, 83A

Promass 83 A
1 2 3 4 5 6 7 8 9 10 11

Promass 83 A -
Nominal Diameter
1 01 1/24”, 0.73 lb/min max. full scale
02 1/12”, 3.7 lb/min max. full scale
04 1/8” 16.5 lb/min max. full scale
2 Measuring system
A Standard version, 904L SS measuring tube 8 Cable entries
B Standard version, Alloy C-22 measuring tube B 1/2” NPT
C 3.1B wetted parts, 904L SS measuring tube L 1/2” NPT fieldbus connector (only for approval A)
D 3.1B wetted parts, Alloy C-22 measuring tube X Sensor only, without cable entries
G High pressure version, 904L SS measuring tube (not for 1/24” size) 9 Power Supply / Display
K High pressure version, 3.1B wetted parts, 904L SS measuring tube 7 85 to 260 VAC, without display, remote configuration only (not for
(not for 1/24” size) wall mount or SS housing)
3 Process Connections 8 20 to 55 VAC / 16 to 62 VDC, without display, remote configuration
SVW Cajon 4-VCO-4 couplings only (not for wall mount or SS housing)
PPD Cajon 4-VCO-4 couplings, 240 grit (not for Alloy C-22 tube) A 85 to 260 VAC, with display, touch-control operation (language: EN,
STA 1/2” Tri-clamp couplings, 150 grit, 3-A version (not for Alloy C-22 ES, FR, IT, NL, PT, DE)
tube or high pressure version) B 20 to 55 VAC / 16 to 62 VDC, with display, touch-control operation
SPD 1/2” Tri-clamp couplings, 240 grit, 3-A version (not for Alloy C-22 (language: EN, ES, FR, IT, NL, PT, DE)
tube or high pressure verison) X Sensor only
4 Test / Treatment 10 Software
(Test pressure of the measuring system to 1.5 times nominal pressure A Standard software
rating of the selected process connections) C Concentration measurement functions
A No special test or treatment E Advanced diagnostics functions
B Pressure tested, 2.3 pressure test certificate and type test of J Batching functions
secondary containment L Concentration measurement and batching functions
C Oil and grease-free process wetted surface N Batching functions and advanced diagnostics functions
D Oil and grease-free process wetted surface, 2.3 pressure test R Concentration measurement and advanced diagnostics functions
certificate and type test of secondary containment T Concentration measurement, batching and advanced diagnostics functions
F Containment with purge connection, 2.3 pressure test certificate X Sensor only
and type test of secondary containment 11 Outputs / Inputs
5 Calibration Fixed I/O modules
A 0.1% flow, standard density calibration A Current HART, frequency
B 0.1% flow, 0.002 g/cc density calibration B Current HART, frequency, 2 relays
C 5-point calibration, 0.1% flow (specify flow range), standard F Profibus-PA, I.S. (not for approvals A and R)
density calibration G Foundation Fieldbus, I.S. (not for approvals A and R)
D 5-point calibration, 0.1% flow (specify flow range), 0.002 g/cc H Profibus-PA (for approvals A and R only)
density calibration J Profibus-DP (for approvals A and R only)
G A2LA/SCS 5-point, 0.1% flow; 0.02 g/cc density, with flow K Foundation fieldbus (for approvals A and R only)
certificate traceable according to ISO 9000 (ISO/EIC 17025) Q Modbus RS485, status input
Specify flow range (not for 1/24” sensor) R Current HART, current; active I.S. (not for approvals A and R)
H A2LA/SCS 5-point, 0.1% flow; 0.002 g/cc density, with flow S Current HART, frequency; active I.S. (not for approvals A and R)
certificate traceable according to ISO 9000 (ISO/EIC 17025) T Current HART, frequency; passive I.S. (not for approvals A and R)
Specify flow range (not for 1/24” sensor) U Current HART, current; passive I.S. (not for approvals A and R)
6 Approvals Flexible I/O modules
A General purpose, for use in non-hazardous locations C Current HART, frequency, 2 relays
N FM explosion proof, Class I, Div. 1, Grps A-D D Current HART, frequency, relay, status input
CSA approved Class I, Div. 1, Grps A-D and zone 1 E Current HART current, relay, status input
R FM non-incendive Class I, Div. 2, Grps A-D L Current HART, 2 relays, status input
CSA approved Class I, Div. 2, Grps A-D and zone 2 M Current HART, 2 frequency, status input
N Modbus RS485, current output, frequency output, status input
7 Protection
P Profibus-DP, current HART, frequency output, status input
A NEMA 4X (IP 67) compact aluminum housing
V Profibus-DP, 2 relay outputs, status input
B NEMA 4X (IP 67) compact stainless steel housing (only for
W Current HART, 2 current outputs, relay
approvals A and R)
0 Current HART, 2 current outputs, status input
C NEMA 4X (IP 67) wall-mounted housing with 30 ft. (10 m) of cable,
2 Current HART, frequency, current, relay
remote version (only for approvals A and R)
3 Current HART, current, relay, current input
D NEMA 4X (IP 67) wall-mounted housing, 65 ft (20 m) of cable, 4 Current HART, frequency, relay, current input
remote version (only for approvals A and R) 5 Current HART, frequency, current input, status input
E NEMA 4X (IP 67) aluminum field housing, 30 ft. (10 m) of cable, 6 Current HART, current, current input, status input
remote version (not for approvals A and R) 7 Modbus RS485, 2 relay outputs, 1 status input
F NEMA 4X (IP 67) aluminum field housing, 65 ft. (20 m) of cable, X Sensor only
remote version (not for approvals A and R)
1 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact,
aluminum field housing (only for approvals A, N, P and R)
2 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67), compact,
stainless steel housing (only for approvals A and R)
3 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67), remote version,
wall mount housing, 65 ft (20 m) cable (only for approvals A and R)
7 -40°F (-40°C) ambient temperature, NEMA 4X (IP 67) remote field
housing, 65 ft (20 m) cable (only for approval N, P)

Endress+Hauser 35
Proline Promass 80A, 83A

United States Canada Mexico

Endress+Hauser, Inc. Endress+Hauser Canada Endress+Hauser, México, S.A. de C.V.


2350 Endress Place 1075 Sutton Drive Av. Gustavo Baz No. 43
Greenwood, IN 46143 Burlington, ON L7L 5Z8 Fracc. Bosques de Echegaray
Tel. 317-535-7138 Tel. 905-681-9292 Naucalpan de Juárez, C.P. 53310, Estado
Sales 888-ENDRESS 800-668-3199 de México
Service 800-642-8737 Fax 905-681-9444 México
fax 317-535-8498 www.ca.endress.com Tel: (52) 55-5371-1110
inquiry@us.endress.com Fax (52) 55-5371-1128
www.us.endress.com eh.mexico@mx.endress.com

TI054D/24/ae/04.06
© 2006 Endress+Hauser