Beruflich Dokumente
Kultur Dokumente
1
Ludwig Boltzmann Institute for Experimental and Clinical Trauma-
tology, Vienna, Austria,
2
Department of Anesthesiology and Intensive Care Medicine, Lorenz
Boehler Trauma Center, Vienna, Austria.
Received: July 4, 2002; revision accepted: February 20, 2003
Tables 1a and 1b. Results by the LUMItestPCT® versus the PCT-Q® test. patients and the highly increased procalcitonin levels (>
The results counting each sample separately (a) were similar to the
2 ng/ml) found in systemic inflammatory response syn-
results weighted by the number of samples per patient (b).
drome.
a) Results counting each sample separately, LUMItestPCT® versus PCT-Q® The total number of samples measured by both tests
test (n = 167) is not identical with the total patient number
Increased Normal Total (n = 15). However, the study is focused upon the com-
LUMItestPCT® LUMItestPCT®
parison between two different tests in individual sam-
Increased PCT-Q® test 14 2 16
ples gained only from severely traumatized patients, i.e.,
Normal PCT-Q® test 24 127 151 from a comparable group. Thus, the number of samples
Total 38 129 167 measured by both tests in severely traumatized patients
is relevant, rather than the actual number of patients.
High procalcitonin levels were detected in 14 of 38 samples by the PCT-
Q® test. Normal procalcitonin levels were detected in 127 of 129 samples Since the total number of samples measured per patient
by the PCT-Q® test. varied, clustering of samples was taken into considera-
tion in the statistical analysis. At the increased procalci-
b) Results weighted by the number of samples per patient, LUMItestPCT®
versus PCT-Q® test tonin level (> 0.5 ng/ml), the total number of highly
Increased Normal Total
increased samples (> 2 ng/ml) was unfortunately very
LUMItestPCT® LUMItestPCT® small. Nevertheless, it should be pointed out that the
PCT-Q® test only picked up 36.8% of all increased pro-
Increased PCT-Q® test 1.8 0.3 2.1 calcitonin levels (50% of the highly and 34.4% of the
Normal PCT-Q® test 2.3 10.6 12.9 moderately increased procalcitonin levels). For the low
Total 4.1 10.9 15.0
procalcitonin category, it might have appeared reason-
To adjust for the different numbers of samples per patient, the values we- able to measure procalcitonin in healthy volunteers
re weighted by the reciprocal individual sample size. The results weighted using both the PCT-Q® test and the LUMItestPCT®.
by the number of samples per patient were similar to the results shown in
Table 1a, which counted each sample separately. However, we elected not to do so since the PCT-Q® test
is not sensitive enough to pick up procalcitonin in
Table 2. Sensitivity (44%) and specificity (97%) of the PCT-Q® test com- healthy volunteers.
pared to the LUMItestPCT® with raw (Raw CI) and simultaneous 95% Apart from the fact that the PCT-Q® test is semi-
confidence intervals (Sim CI). quantitative and immunochromatographic while the
LUMItestPCT® is quantitative and immunoluminomet-
Parameter Raw CI Sim CI
ric, there are principal differences between the two
Sensitivity (%) 44 8–69 6–72 tests, both with regard to cost and to handling. In theo-
Specificity (%) 97 90–100 89–100 ry, the LUMItestPCT® is roughly 25% cheaper per sin-
gle test than the PCT-Q® test. Furthermore, the
LUMItestPCT® takes twice as long to perform (1 h) as
test picked up normal procalcitonin levels in 127 (nor- the PCT-Q® test. In practice, however, the LUMItest-
mal LUMItestPCT® and normal PCT-Q® test) of 129 PCT® requires double measurements and standard
samples (Table 1a). The true value can be expected in curves as well as a luminometer. Furthermore, the
the range from 6% to 72% with a probability of 95% LUMItestPCT® requires experienced personnel since it
(Table 2). The specificity of the PCT-Q® test was 97%. is far more complicated to perform than the PCT-Q®
The true value can be expected in the range from 89% test. Thus, in practice, the LUMItestPCT® is only cheap-
to 100% with a probability of 95% (Table 2). The results er for laboratories with access to a luminometer and
weighted by the number of samples per patient (Table enough experienced personnel. Nevertheless, the PCT-
1b) were similar to the results counting each sample sep- Q® test, though quicker and very simple to perform, is
arately (Table 1a). also less accurate. We consider this lack of accuracy of
the PCT-Q® test a serious drawback in the trauma inten-
Discussion sive care unit.
The PCT-Q® test was less sensitive than the LUMItest Since the PCT-Q® test is highly specific, increased
PCT® at both the moderately increased procalcitonin procalcitonin readings can be considered reliable, i.e, it
levels (0.5–2 ng/ml) found in some multiple trauma is unlikely that the PCT-Q® test yields falsely increased