CX21
All rights reserved, Reproduction in whole or in part without written permission is prohibited.
CX21 MAINTENANCE MANUAL
INTRODUCTION
The purpose of this manual is to satisfy any requirements for maintenance material that maybe
considered as necessary to professionals in the maintenance field, such as Maintenance engineer
in Hospitals.
It is intended to be used as a tool for performing basic maintenance procedures if needed or when
required as per the recommended maintenance schedule.
The sections from this manual only cover procedure pertaining maintenance ’s that are considered
to be easily performed. For more involved maintenance’s or repairs, it is recommended that you
contact a qualified service engineer from your local Authorized Olympus dealer.
Maintenance parts, grease, and other items specified in the manual can be ordered from your
Authorized Olympus dealer, and subject to change without notice.
CONTENTS
CHAPTER 1
MAINTENANCE PROCEDURE
1. Maintenance of Microscope
1) Fundamental handling
a. Read the instruction manual thoroughly, handle the microscope correctly.
b. Be sure to make a usual cleaning, especially after every use of microscope.
c. When handling the microscope, do not expose it to shock, moisture, heat and dust.
d. If the problem occurs, do not treat it in self-judgement.
Temperature
Dry blue aspergillus Degree(C)
40
30
20
10
Aspergillus
Humidity % 60 70 80 90 100
-1-
CX21 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
2. Guide to Maintenance
2-1 Overview of maintenance
1) 2) 3)
Loosen the screw securing eyepiece.
7
5
7
0
6
0
5 0
OLYMPUS OLYMPUS
CX21
S
CX 21
10
OLY MPU
0
10 1
0
0
0
1 12
0
12 10 13
0 0 0
0
3
1
140
15
0
0
1
120
130
CX 21 14
0
15
0
1 4 16
0 0
160 1
7
1 7 150
1 6
0
7
1
30
3
0
20
02
1
0
1
03
0
0
0
0
01
0
Set your correct interpupillary distance. Sweep off dust on the outer surfaces Remove the eyepieces and objectives,
Note any areas suggesting a need for with the soft brush. If there are stains on then clean their lenses with cleaning
mechanical and/or optical maintenance the microscope frame, clean them with solution. (When removing the objective,
by operating it or observing a specimen. neutral detergent. use the rubber band.)
4) 5) 0
7
7 5
6) 7
5
5 0 6 0
5
0
OLYMPUS
100
0
1
OLYMPUS
10 1 2 1 3
0 0
CX 21
0
1 0 140
CX 21
12
0
13 1 5
0 0
140 1 6
0
15 7
1
0
16
0
1 7
03
02
03
01
2
0
0
Top lens
01
0
(A)
(glass) glass
Plastic Plastic
Remove the condenser downward. Take off the mechanical stage from the Mount the removed components and
Wipe off any dirt spots on the surface of plane stage by removing two screws. mechanical stage to the frame.
top lens, filter. The plastic parts should Clean the surface of stage and filter on
be cleaned with netural detergent. the base cover.
7) 8) 9)
PUS
S OLYM
OLY MPU
CX21
CX21
100
110
10
0
11
0
120
12
0
130
140
13
0
1 5
14 0
0
15
0
160
16
0
1 7
1 7 S
OLY MPU
CX 21
100
3
0
110
02
120
3
1 3
0
2
0
0
0
0
140
01
1 5
0
0
160
1 7
3
0
20
1
0
0
Polish all plastic components and the Return the interpupillary distance to Do a final check following the inspection
frame with silicon cloth. original condition and prepare for the sheet in this manual. (See page 7)
inspection. (see page 6)
Tool required:
Lens tissue Neutral detergent Cleaning solution Cotton swab Blower Silicon cloth
(For plastic part or (For lens or filter or tweezers (For finishing)
frame) etc.) etc.
-2-
CX21 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
Before cleaning: Lightly brush the lens surface or blow with the blower before wiping with tissue.
This removes particles that may scratch the lens surface. (to protect the lens
coating)
Moisten the tip of With a circular motion, Dip a new cotton swab Wipe the objective lens
cotton swab with wipe the top lens in the cleaning solution from the center
cleaning solution. surface with the cotton and shake vigorously towards the periphery,
swab, to thoroughly to remove any excess while rotating the lens.
remove any oil or dirt cleaning solution.
from the lens.
When the lens size is large and difference in level Cleaning the immersion objective:
between the lens and the lens frame is small:
Fold the lens tissue several times and moisten it Wipe off the immersion oil while absorbing it with
with cleaning solution. lens tissue that is not moistened. After that, clean
After that, apply the folded line edge to the center the lens as the same manner mentioned on the
of lens, push it with index finger and turn the left. When the top surface of lens frame is higher
objective by the other hand to clean the lens than that of lens and remained dirty portion on
while moving it from the center towards the the periphery of lens can not be wiped off, clean
periphery. the lens referring the above figures, 1 to 4.
-3-
CX21 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
Fold the lens tissue into two or three layers and moisten its shaded part with cleaning solution.
Hold the filter at its edge and fold the lens paper from the lens center to outside as illustrated.
Move the lens tissue gradually to outside while turning the filter by left hand.
Hold a sheet of lens tissue between your middle and index fingers, then fold and wrap it around your
index finger. Hold the tissue down with your thumb and moisten it with sufficient cleaning solution.
1) 2) 3) 4) 5) 6)
Wipe the prism surfaces from front to backward at a stroke, applying even pressure.
-4-
CX21 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
Wrap a sheet of lens tissue around a cotton swab Dip the wrapped lens tissue in the cleaning
as illustrated. If the area to be cleaned is large, solution, and wipe the eyepiece from the
wrap the lens tissue looser and thicker. Otherwise, center towards the periphery in a circular
make a thin, tight wrap. motion.
Important notes:
1) Never rub the lens surface strongly.
2) Do not use the same lens tissue to clean more than one lens.
3) Do not moisten the lens tissue with an excessive amount of cleaning solution.
4) When cleaning with tweezers, be careful not to protrude its tip from the lens tissue.
-5-
CX21 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
(E)
S
OLYMPU
CX21
100
10
120
130
140
150
160
17
30
(C)
20
10
0
(D)
(F) 4x10x
(A)
(B)
-6-
CX21 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
Remarks:
-7-
CX21 MAINTENANCE MANUAL INSPECTION PROCEDURE
CHAPTER 2
INSPECTION PROCEDURE
B Eyepiece C Prism
D Objective
A Specimen
B Condenser lens
B or A Glass
(Frosted : underside)
C Frosted glass
C Collector lens
-8-
CX21 MAINTENANCE MANUAL INSPECTION PROCEDURE
Eyepiece
Check it by rotating the eyepiece
after clamping screw is loosened.
(Wipe carefully and gently so as
not to damage the lens coating.)
PUS
OLYM
CX21
100
110
120
130
Objective
140
150
160
17
Condenser
Check the top lens and filter if
conspicuous dirt is not seen.
Note: If dirty portion can not be identified in the above, it is assumed that internal lens and
prism are contaminated.
In this case, please contact your Authorized Olympus dealer.
-9-
CX21 MAINTENANCE MANUAL INSPECTION PROCEDURE
Light
Dust becomes conspicuous when If there is a dirty part or a remained part that
looking at it with the lens is not cleaned completely, the haze of this
inclination changed. part will disappear slower than that of the
other part.
- 10 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
CHAPTER 3
REPAIR PROCEDURE
1. Optical Adjustment
PREPARATION Adjusting the left/right optical axis
70
Objective 10X
Specimen Position of
approx. 62mm
whose center is
identified
(concentric circles etc.)
Control knob
- 11 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
70
are also deviated.
60
50
(2) Aligning the cross center of eyepiece with the specimen center
75
70
60
50
75
70
60
50
- 12 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
3. Firmly tighten the screws which secure Image at the end of adjustment
the left sleeve.
75
70
60
50
- 13 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
2. Mechanical Adjustment
2-1 Preparation for the tension adjustment of X/Y-wire
If a specimen image is moved when the stage is brought into the desired position of
specimen, it is necessary to adjust the wire tension of stage.
1) Take off the mechanical stage (A) from the plane stage (B)
by removing the two screws (*1).
75
70
60
50
(B)
(A)
*1
10
0
10
(B) OLYMPU
S
12
0 13
0
140
CX21
(A)
150
160
1
7
30
20
10
0
(a)
2) Remove the specimen holder (C) from the 3) Remove the Y-cover (E).
inner guide (D). Screw: CUTB3X8SA (*3), 3pcs.
Screw: 3PSK2X4SA (*2), 3pcs. <indicated by arrows>
*2 Y-wire
(D)
X-wire
(E)
(C)
*3
* For the tension adjustment of X-wire, * For the tension adjustment of Y-wire,
refer to next page. refer to next page.
- 14 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
- 15 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
- 16 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
(A)
sliding surface
A002
Grease
OLYMPUS
(OT2008)
CX21
clearance
in gear contact point
attachment part Grease of spring
a (OT2012)
Applied part: Grease
(F) (about 20mm) a (OT2006)
Grease
(G)
(OT2006) (I)
groove (H)
and whole
surface
(character (D)
indication (C)
surface)
(E)
Grease
(OT2006)
(B)
A003
- 17 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
(Inlet) U1
INPUT L X1/X2 : Binder
Circuit Board
X3 : Tube
100-120/ E
220-240VAC XDS1 : Socket
50/60Hz N X3 XDS1
1 6V20W
2 Halogen Lamp
X1 X2
If the lamp is not lit, check if the halogen bulb (6V20W) is broken or lamp socket is burned and
also check that the voltage is being outputted each from (1), (2), (3) using multimeter to identify
the defective part. (Refer to the figure below.)
In case where there is a problem in the circuit board, replace the circuit board as a whole because
the components can not be supplied. Since the rheostat ass’y is incorporated in the circuit board,
the voltage adjustment is not necessary. ( It has been already adjusted: Min. 1.5V or less, Max. 4.5V +/-0.3 )
- 18 -
CX21 MAINTENANCE MANUAL REPAIR PROCEDURE
Grease;
OT2008 *2
(D)
Left and right Front side
dovetails (A)
: Pushing direction
sliding (F)
surface Grease;
(E) OT2008 A009
sliding
(B) surface
(1) Loosen the stopper (A) and remove the condenser (B) downward by turning
the knob (F).
(2) Remove the left dovetail (C) as seen from the front side.
Screws : AB3X8SA, 2pcs. (*1)
Note on assembly :
1) Apply grease to the portions shown as the above figure.
2) The right dovetail (D) is mounted by pushing it in the arrow directions.
3) The left dovetail (C) is mounted by pushing it in the upward direction.
At this time, adjust the position of dovetail (C) in the left and right directions so that
the condenser moves smoothly without a play (vertical movement).
- 19 -
CX21 MAINTENANCE MANUAL JIGS AND TOOLS /GREASES
CHAPTER 4
JIGS AND TOOLS / GREASES
2. List of Greases
No. Description Ref. page
OT2006 Grease (heavy) 17
OT2008 Grease (medium) 17, 19
OT2012 Mo grease 17
- 20 -
CX21 MAINTENANCE MANUAL MAINTENANCE PARTS
CHAPTER 5
MAINTENANCE PARTS
H The index No. are shown on the figures in “REPAIR PROCEDURE” instead of parts No.
*1 Contact your Authorized dealer.
- 21 -
OLYMPUS OPTICAL CO., LTD
2-3-1, Nishi shinjuku shinjuku-ku, Tokyo, Japan