1.1 Principle
The DILAS FT2000 is a digital, high resolution, non-contact laser measurement sensor. It measures on all
non-shiny surfaces.
The operation of the DILAS FT2000 is based on time of flight measurement. The sensor calculates the
distance of the target surface using the time of flight of visible laser light pulses.
E1421/t 01/16 3
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
1.3.2 Detection
2.1 Location
For the choice of the location, try to protect as far as possible the optical window against eventual deposits
of dust and to place the sensor away from hot billet, bloom or slab.
See also § 2.5. precautions for laser safety.
4 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
2.2 Mounting
See mechanical specifications and dimensions § 10.
The DILAS FT2000 is fixed on his stand with a screw (∅ 12 mm). The foot is adjustable in 2 axes and allows
an easy alignment.
The stand should be sufficiently rigid to maintain the required alignment for the measuring range and
accuracy.
The DILAS FT2000 has a water cooling plate and a hood with air purging for protection of the window glass,
thus permitting the use in harsh environment.
However in some severe cases, it will be absolutely necessary to protect the sensor with an auxiliary heat
shield (see § 11.1 our heat shield proposed as accessory), or to remove excessive water vapour with
additional ventilation, etc...
Note: Unused outputs should be insulated or connected to free terminals. If the serial link is not used, it is recommended
to strap the 2 wires of the link (pin 4A and 4B) through a 100 Ω resistor.
Note: The sensor is supplied with a protective cover on the connector. Do not remove until the mating connector is
plugged in as it protects the contacts from corrosion and improves the hermetic sealing of the sensor.
E1421/t 01/16 5
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
It is necessary to cool the sensor if the ambient temperature is over 50 °C (120 °F).
Cooling water specifications:
• Clean industrial quality water at 25 °C (77 °F), maximum temperature 35 °C (95 °F).
• Pressure 1-2 bars, maximum recommended: 4 bars (approximately 60 PSI).
• Flow rate: 1-2 l/mn (approximately 0.5 gal/mn) (see picture above).
• Material: stainless steel (inside Ø 6 mm) cast in the cooling plate.
• Connections: rubber hose, Ø 10 mm.
2.4.2 Air purging
A supply of instrument air may be connected to the protective hood. The system is designed to provide
positive pressure in front of the window glass.
Air purging specifications:
• Air quality : the air should be dry and free from oil or dust. If these conditions cannot be met it is
better not to use air to avoid deposit of oil on the window glass.
• Air pressure : 50 to 200 g/cm² (1 to 3 PSI).
• Flow rate : 4 to 6 l/min (0.15 to 0.6 CFM).
• Connection : rubber tubing 10 mm diameter (0.4).
2.5 Precaution for laser safety
As per IEC 60825-1: 2007, the DILAS FT2000 is equipped with a class 2 laser diode with an emitted power
below 1 mW, and the DILAS FT2300 is equipped with a class 3R laser diode with an emitted power below
5 mW.
Following precautions are required:
• It is necessary to prevent from a continuous viewing in the laser beam.
• A momentary exposure (0.25 s) which could arrive during an accidental viewing situation is not
considered as dangerous. However it is not allowed to aim deliberately the laser beam towards
people. The use of optical instruments (glasses for instance) could be dangerous.
• Every DILAS FT2000, FT2300 has a warning logotype fixed on his housing as shown below:
6 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
3 INSTALLATION
4 CONFIGURATION
E1421/t 01/16 7
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
File menu
• Load settings: upload the settings from a file into the device.
• Save settings: write device settings into a file
• Reset to factory settings: Uploads factory settings into device (with exception of address
that remains unchanged)
Help menu
• Operation Manual: a PDF file is opened.
• About: a pop up with:
o software version of DELTAConf
o Type of sensor
o Serial number of sensor
o Both software versions of sensor.
Buttons:
• List of Errors: opens a window that lists all error codes with their signification.
• Button Apply: Uploads new data and saves data in the non-volatile memory of device.
• Button Cancel: Reads the current settings from the device.
4.3.1 Status
This box gives the status of the sensor.
8 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
4.3.2 Laser
Laser Off, Laser On: switch on/off the laser
4.3.3 Parameters
Distance at lower limit of Analog output: distance for 4 mA (0V)
Distance at upper limit of analog output: distance for 20 mA (10V).
Lower distance for Product presence: minimum distance for product presence (relay output)
Upper distance for Product presence: maximum distance for product presence (relay output)
Note 1: Measured position corresponds to the average position during the measuring time. In case of a relatively
constant speed we can consider that distances provided by the sensor are delayed by half the measuring time.
Factory setting: 0
E1421/t 01/16 9
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
Baud rate setting of sensor: this parameter sets the baud rate of the DILAS FT2000 sensor (9600, 19200
or 38400). The new Baud rate will be applied at the next power on.
Factory setting: 9600.
Error Hold Mode: this parameter defines a timeout when the sensor is not able to make a measure.
Error hold mode Sampling mode = 0 Sampling mode = 1 Sampling mode = 2 Sampling mode = 3
1
3.5 s 210 ms 560 ms 1050 ms
(factory setting)
Error hold mode = 0: no timeout, when the sensor is unable to measure, the relay output will immediately
switch to product absence state and the analogue output to 20 mA (10V).
Error hold mode = 1 or 2: the table above indicates the timeout. When the sensor is unable to measure, its
outputs will be kept as per the last valid measure. After the timeout is elapsed, the relay output will be
switched to product absence state and the analogue output to 20 mA (10V).
Factory setting : 1.
Alarm Output Mode: if set to 1, the DILAS FT2000 sensor produces a pulse from 24 V to 0 V on the alarm
output, for each new measurement of the laser module. This information can be used for synchronization.
Measurements of
laser module
Factory setting : 0.
4.3.6 Address
Address: is the slave address of the sensor for serial line protocol.
Factory setting : 7.
• Button Apply: Uploads new slave address.
10 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
5 DIAGNOSTIC
A 4 digits display at the rear of the sensor gives the measure in mm (or 1/10 or 1/100 of inch depending on
the configuration) or will display an error code if there is no measurement.
Error
Cause Action
message
• The parameters have been restored to factory
E101 Corruption of the user parameters default.
• Rewrite your application parameters.
Signal too high (target too • Check the target, it may be too shiny,
E256
reflective) • Install the sensor with a small angle.
Note 1: when one of these error code Exxx is active, the alarm output (pin 2B) is forced to 0.
Note 2: E101 for DILAS FT2000, FT2300, with serial number > 216300.
E1421/t 01/16 11
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
6 MAINTENANCE
Caution:
• Use an unpolished target.
• Never look towards laser but towards target.
12 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
7 SERIAL LINK
7.1 Hardware:
The hardware for the serial link is type EIA RS 485 on a twisted pair.
Asynchronous mode 9600, 19200 or 38400 bits/sec with : 1 start bit, 8 bits data, no parity, 1 stop bit. Factory
setting is 9600 bits/sec.
Note : It is necessary to put the terminating resistor of the serial link (100 Ω).
7.2 Protocol:
7.2.1 Message frame:
The master can address several individual slaves (DILAS FT2000). Slaves return a message (answer) to
queries that are addressed to them.
The slave answers to the query in about 5 ms. For new parameter allocation, the delay can be up to 30ms.
Note : For sampling mode, error hold mode, LASER ON and LASER OFF queries, and for first request for measurement (42hex)
after a “LASER OFF” state, wait 500 ms before sending next query.
Message frame:
Head = Ø2 Hex Address: 1 byte Control: 1 byte Data Checksum: 1 byte End = Ø3 Hex
Address (master to slave message) = Ø1 Hex to 1F Hex (ØØ Hex for general address). Factory setting is
address 07 Hex.
Address (answer from slave to master) = Address from request + 80 Hex.
Checksum = sum of bytes (Head +Address + Control + Data) (Last significant byte)
if checksum < 10 Hex, add 10 Hex.
Data:
The measures are given in mm, 1/10 or 1/100 inch, with 5 ASCII bytes:
Example: for a distance of 2502
In ASCII: 02502
Or in hexa: 30 Hex 32 Hex 35 Hex 30 Hex 32 Hex
Error message: when there is an error (see § 4.2.2) the data returned is in ASCII:
E0XXX where XXX corresponds to the error.
E1421/t 01/16 13
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
Control Note
Request for measurement 43 Hex
42 Hex If sensor is in <LASER OFF> state,
it will be put automatically in
<LASER ON> state
Allocation of an address 45 Hex Serial number on 5 or 6 bytes ASCII
number
LASER ON 4B Hex
LASER OFF 4C Hex
Identity request 4F Hex Serial number on 6 bytes ASCII
41 Hex Serial number on 5 bytes ASCII
Request for parameters 60 Hex
Allocation of new parameters 62 Hex
Queries 41 Hex and 45 Hex with serial number on 5 bytes are provided for compatibility reasons with older
sensors.
Data
Status : 1 byte Measure: 5 bytes ASCII
Note : In <LASER OFF> state, the measure will be 00000 ASCII. The sensor remains in <LASER OFF>
state even after a request for measurement. Status byte is refreshed in case of a default. This query can be
used for Status monitoring even in <LASER OFF> state. The first measure values after a « LASER OFF »
state will be 00000 (ASCII) and up to 4 s in difficult measuring conditions (reflecting or absorbing surfaces).
Status : Byte = 1 0 • • • • • •
Critical detection
New measurement
Critical detection : 1
New measurement : 1
OK : 0
Old value : 0
Internal temperature
default Failure
Default : 1 Failure: 1
OK : 0 OK : 0
14 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
Example:
Query
Ø2 Hex Ø7 Hex 43 Hex 4C Hex Ø3 Hex
Answer
Ø2 Hex 87 Hex 43 Hex A1 Hex 02502: 5 bytes ASCII 66 Hex Ø3 Hex
Note : Status bit new measurement : this bit is set after each new measurement of the laser module. This bit is reset after
each answer of a request for measurement (42 Hex or 43 Hex ) on serial link.
Data
Serial number: 6 bytes ASCII Address: 1 byte
No answer
Example for product serial number 101215 with new allocated address = 8
Query
Ø2 Hex ØØ Hex 45 Hex 101215: 6 bytes ASCII Ø8 Hex 79 Hex Ø3 Hex
No answer
Check can be done with identity request at address 8.
4B Hex : LASER ON
This command will not operate if LASER OFF is selected by hardware (pins 3A and 3B connected).
After power on the laser is on (if LASER OFF is not selected by hardware).
Status: A0 Hex = 1010 0000 = Laser On, no default, Old value, Product absence
E1421/t 01/16 15
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
Data
Reference : 4 bytes ASCII Serial number : 6 bytes ASCII
Data:
Parameter number: 1 byte ASCII
Parameter Definition
number (ASCII)
1 Upper distance for product presence
2 Lower distance for product presence
3 Distance at upper limit of analog output
4 Distance at lower limit of analog output
5 mm or inch units / Sampling mode
6 Baud rate of serial link
7 Error hold mode
8 Alarm output mode
16 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
Data :
Parameter number: 1 byte ASCII Parameter value : 5 bytes ASCII
Data :
Parameter number: 1 byte ASCII Parameter value : 5 bytes ASCII
E1421/t 01/16 17
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
Sampling Mode
Example 1:
Query to set the upper distance for product presence at 12000 mm (472 inches)
Ø2 Hex Ø7 Hex 62 Hex 31 Hex 12000: 5 bytes ASCII 8F Hex Ø3 Hex
Answer
Ø2 Hex 87 Hex 62 Hex 31 Hex 12000: 5 bytes ASCII 1F Hex Ø3 Hex
Example 2:
Query to set the measuring units in 1/100 inches and sampling mode 2.
Ø2 Hex Ø7 Hex 62 Hex 35 Hex 00112: 5 bytes ASCII 94 Hex Ø3 Hex
Answer
Ø2 Hex 87 Hex 62 Hex 35 Hex 00112: 5 bytes ASCII 14 Hex Ø3 Hex
Other queries:
Following queries can be used for compatibility reasons with older sensors :
18 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
Data
Reference : 4 bytes ASCII Serial number : 5 bytes ASCII
Query
Ø2 Hex Ø7 Hex 41 Hex 4A Hex Ø3 Hex
Answer
Ø2 Hex 87 Hex 41 Hex 2007: 4 bytes ASCII 01215: 5 bytes ASCII 8C Hex Ø3 Hex
Note : the first measure values after a « LASER OFF » state will be 00000 (ASCII) and up to 4 s
in difficult measuring conditions (reflecting or absorbing surfaces).
Data
Serial number: 5 bytes ASCII Address: 1 byte
No answer
Example for product serial number 101215 at address 8. The five last characters are used.
Query
Ø2 Hex ØØ Hex 45 Hex 01215: 5 bytes ASCII Ø8 Hex 48 Hex Ø3 Hex
E1421/t 01/16 19
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
8 TECHNICAL SPECIFICATIONS
Model FT20•• FT23••
Emitted power (EN 60825-1) ≤ 1 mW , class 2 ≤ 5 mW , class 3R
0.3 - 30 m 0.3 - 50 m
Measuring Range (1)
up to 100 m on white mat surfaces up to 150 m on white mat surfaces
Maximum target temperature 1000 °C / 1830 °F 1300 °C / 2370 °F
±3 mm (up to 30 m, +15°C... +30°C)
Accuracy (2) (3)
±5 mm (up to 100 m, full temp. range)
6mm at 10m
Beam diameter 15mm at 25m
60mm at 100m
Mode 0 : 0.25 to 1 s, automatic, depending on Mode 0 : 0.1 to 1 s, automatic, depending on
surface (default mode) surface (default mode)
Sampling time
Mode 1 : 0.1 s on white mat surfaces only, Mode 1 : 0.1 s on white mat surfaces only,
Mode 2 : 0.25 s, Mode 3 : 0.50 s Mode 2 : 0.25 s, Mode 3 : 0.50 s
Wavelength 630 - 670 nm
(1) Distances are referred to the front plan of the sensor (see dimensions § 10).
(2) On not moving targets. Statistical spread: 95%
(3) Accuracy can be deteriorated by an intense ambient light, a reflecting or an absorbing Laser Class 2
surface, air turbulences. Under such conditions, accuracy is +/- 5mm. Laser Class 3R
Outputs
Model FT2•06 FT2•07 FT2•26 FT2•27
0 - 10 V 4 - 20 mA 0 - 10 V 4 - 20 mA
Measurement output (1)
(10 mA max.) (500 Ω max.) (10 mA max.) (500 Ω max.)
Communication protocol Serial link (RS485) Profibus-DP
Digital outputs: Product presence(2) Relay : Single pole changeover, Switching capacity : 550 VA: 250 V max, 5 A max
Digital outputs: Alarm (3) Static : 0-24 VDC 20 mA (24 V if no failure detected)
Digital inputs: Laser beam Off/On 0-24 V 15 mA
(1) Linearity 0.1% - Temperature drift: 50 ppm/°C.
The range of the analogue output can be set with the configuration kit or serial line (0 - 10 m default). For sensors with Profibus, the range is fixed (0 - 30 m).
(2) The product presence distance can be set with the configuration kit or serial line (0 - 10 m default). For sensors with Profibus, the range is fixed (0 - 30 m).
(3) Power supply default, internal temperature over 55 °C, Error Code.
Other Data
220 V (-10%) to 230 V (+10%) - 50 / 60 Hz or
Operating voltage 110 V (-10%) to 115 V (+10%) - 50 / 60 Hz or
24 VDC (±10%)
10 VA for AC versions
Power consumption
5 W for 24 VDC version
Connector fitted with silicone cable with protective steel braid
Cable
Standard length of 2 m (other length : 3, 5 or 8 m)
Weight 7 kg
Protection rating IP 66 (cast aluminium case)
Pollution degree 3
Air Purging Protection of the optic with clean air : 50 to 200 g/cm², 4 to 16 l/mn
-10 °C to 50 °C (14 °F to 122 °F)
Operating temperature Up to 120 ° C (250 ° F) with water cooling: industrial quality water at about 25 °C (77 °F),
pressure 1-2 bar and flow 1-2 l/mn
Storage temperature -40 to 80 °C / (-40 to 176 °F)
Operating relative humidity 98% max.
Operating altitude 0 to 2000 m
20 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
9 ORDER REFERENCES
FT 2 .0.
10 DIMENSIONS
E1421/t 01/16 21
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
11 ACCESSORY
12 CONNECTION
AC SUPPLY DC SUPPLY
22 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
E1421/t 01/16 23
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
DILAS FT2000
4. (*) Window glass support with glass and gasket .............................................................................. 7592061
8. Printed Circuit Board for DILAS FT2000, FT2300 (without laser module) :
FT2006 115 V or 230 V AC ..................................................... 7592995
24 V DC .................................................................... 7540205
24 E1421/t 01/16
DELTA – Aéroparc 2 - 12 rue des Cigognes - 67960 ENTZHEIM - FRANCE – TEL : +33 388 78 21 01 – FAX : +33 388 76 02 29 – www.deltasensor.eu
TIME OF FLIGHT
LASER DISTANCE MEASUREMENT
DILAS FT2000
D E L T A
E 1421/t
Aéroparc 2 - 12 rue des cigognes - 67960 ENTZHEIM - FRANCE
TEL. +33 388 78 21 01 - FAX. +33 388 76 02 29 - www.deltasensor.com
Ref. E 1421/t