Sie sind auf Seite 1von 24

DEFIview

2
PC software for MEDUCORE Standard and MEDUCORE Standard

Operating and installation instructions


Contents

Contents
1 Introduction 3
1.1 Product description................................................................ 3
1.2 System requirements.............................................................3
2 Safety 3
2.1 Safety information..................................................................3
3 Installation 4
4 Preparation and operation 8
4.1 Starting DEFIview...................................................................8
4.2 Opening session data............................................................8
4.3 Defining the settings.............................................................. 9
4.4 Using the SD card................................................................13
4.5 Menu structure......................................................................14

WM 67401a 01/2018

2 EN DEFIview
1 Introduction

1 Introduction
1.1 Product description
The DEFIview PC software can be used to export and analyze the
2
MEDUCORE Standard and MEDUCORE Standard sessions.
If you have any queries or feedback, please contact our

Technical Service team:

Phone: +49 (0)40- 88 18 96 - 122

E-mail: TechnischerService@weinmann-emt.de

1.2 System requirements


To use DEFIview you will require a PC with the
following specifications:
® ®
• Operating system: Microsoft Windows
® ®
10, Microsoft Windows 7
(Microsoft, Microsoft Windows, Windows, Windows 10,
and Windows 7 are registered trademarks of the Microsoft
Corporation in the USA and/or other countries.)
• Adobe Acrobat Reader

• SD card slot

• Graphics card: minimum resolution 1,920 x 1,080 pixels

2 Safety
2.1 Safety information
WM 67401a 01/2018

Read these instructions for use carefully. Also observe the


respective instructions for use for the devices, the
components, and the accessories.

DEFIview EN 3
3 Installation

3 Installation
1. Insert the USB stick in the PC.

2. Run the DEFIview2.exe file.


The installation wizard starts up.

3. If necessary: Click Run.

4. Click Next.

WM 67401a 01/2018

4 EN DEFIview
3 Installation

5. If necessary: select the destination directory in which


you would like DEFIview to be saved.
6. Click Next.
WM 67401a 01/2018

7. If necessary: select the Start Menu folder.

8. Click Next.

DEFIview EN 5
3 Installation

9. If necessary: place a check next to the file


associations Weinmann DefiView file (*.wmd).
10. Click Next.

WM 67401a 01/2018

11. If necessary: place a check next to Create a desktop icon.

12. Click Next.


The installation wizard continues with the installation.

6 EN DEFIview
3 Installation

13. Click Finish to complete the installation.

Result The installation is complete.


DEFIview can now be used.
WM 67401a 01/2018

DEFIview EN 7
4 Preparation and operation

4 Preparation and operation


4.1 Starting DEFIview

1. Start the software:


• by double-clicking the desktop symbol DEFIview

or

• via Start | All Programs | DEFIview |

DEFIview Result DEFIview starts up.

4.2 Opening session data


Requirement The program has started and the session data are already
saved on the PC.
When the program starts, the Archive window opens.

1. Open the desired session data in the Archive window by:

• double-clicking the correct line


WM 67401a 01/2018

or
• single-clicking the correct line.

• Confirm your selection by clicking Open

8 EN DEFIview
4 Preparation and operation

or

• click Import.

• Select the destination folder where the session data


you require are saved.
• Select the session data.

• Confirm your selection by clicking Open.

Intermediate result A new line with the required data appears in the Archive window.
Open the required session data by:

• double-clicking the correct line

or
• single-clicking the correct line.

• Confirm your selection by clicking

Open. Result The session data are opened.

4.3 Defining the settings


1. Select Menu | File | Settings.

Result The Settings window opens.


WM 67401a 01/2018

DEFIview EN 9
4 Preparation and operation

You can change the basic settings for the PC software in the
Settings menu. The following parameters can be set:

Parameter Possible values Description


Database Here you can select the type of database
configuration (default selection: H2 Standalone).
Database Data source Here you can select the destination
directory in which the database is saved.
Test Tests whether there is a link to the
database.
2
5
10 Here you can set the speed of the ECG
mm/sec 20 curve display and thus increase the
25 temporal resolution.
40
50
1
2 Here you can set the amplitude scaling
mm/mV 5 and thus the amplitude of the ECG
10 curve.
20
Screen Background Here you can select the background
color of the grid.
Signal Here you can select the color of the ECG
curves.
Grid Here you can select the color of the grid
lines.
Text Here you can select the font color.
Standard settings Here you can reset all the screen
parameters to the factory settings.
You can use the controls on the x and y
Grid spacing axes to set the height and width of the
grid cells.
Here you can select whether the
WM67401a01/2018

Print settings Title page printout should be preceded by a title


page.

10 EN DEFIview
4 Preparation and operation

Parameter Possible values Description


Event list Here you can select whether events are

displayed on the printout.

Shocks Here you can select whether shocks are


displayed on the printout.

Print the ECG No Do not print any ECG leads.


Channel II Print channel II of the session duration
overview of
session duration PAD channel Print PAD channel of the session
duration
Complete view Print the whole session.
Length Current page Print the current page.
Selected events Print selected events.
Print settings No Do not print any parameter curves.
Grid Here you can select whether a grid

should be included on the printout.

Screen
2
5 Here you can set the speed of the ECG
10
mm/sec curve display and thus increase the
20
temporal resolution.
25
40
50
Screen
1 Here you can set the amplitude scaling
WM 67401a 01/2018

2
mm/mV and thus the amplitude of the ECG
5
curve.
10
20

DEFIview EN 11
4 Preparation and operation

Parameter Possible values Description


Print settings Printer Here you can specify the printer / output
form for the session data.
Here you can set the language.
• Thai • Thai
• Français • French
• Español • Spanish
Language* • Arabic • Arabic
• Dutch • Dutch
• Polish • Polish
Language and • Deutsch • German
localization • Russian • Russian
• English • English (US)
Height unit cm You can set the height unit for the curve
in display here.
Weight unit kg You can set the unit for the patient's
lbs weight here.
Network Europe (50 Hz) Here you can select the network
frequency USA (60 Hz) frequency of the power supply system.
*The language will be changed the next time you restart
the software.

2. Click OK.
Intermediate result The settings are saved.

3. Close and restart DEFIview.


Result The settings are adopted by DEFIview. 01/2018
WM 67401a

12 EN DEFIview
4 Preparation and operation

4.4 Using the SD card


Loss of data due to incorrect SD card!
SD cards not purchased from Weinmann may have
reduced functionality or result in the loss of data.
⇒ Only use SD cards from Weinmann.
⇒ Do not use the SD card for third-party files.

4.4.1 Inserting the SD card


Requirement The session data are on the SD card.

1. Connect SD card to the PC:

• via the SD card slot of the PC

or
• via an external card reader connected to the PC.

2. The files saved on the SD card are transferred to the PC.

Result The session data from the SD card can be used on the PC
with DEFIview.
Further information on using the SD card in the device can
2
be found in the MEDUCORE Standard or MEDUCORE
Standard instructions for use.

4.4.2 Deleting data from the SD card


Requirement There is an SD card in the PC's card slot or in an external
card reader connected to the PC.
1. Select all the files to be deleted.

2. Delete the files:

• via the Del key

or

• clicking the right mouse button.


WM 67401a 01/2018

3. Confirm the prompt.


All data on the SD card are erased.
Result The SD card is empty.

DEFIview EN 13
4 Preparation and operation

4.5 Menu structure

4
2

4-1 Complete menu

No. Designation Description


1 Menu and icon bar Contains command keys and patient and settings
information (see "4.5.1 Menu bar", page 15).
2 Event list Displays the events within a session (see "4.5.3 Event
list", page 17).
3 Comment window Input field for adding comments to a session.
4 Curve area Displays the ECG leads (I, II, III, aVR, aVL, aVF, Pads
and Plet) (see "4.5.4 Curve area", page 18).
WM 67401a 01/2018

14 EN DEFIview
4 Preparation and operation

4.5.1 Menu bar

1 2 3

4-2 Menu bar

No. Designation Description


Opens the archive with an overview of individual
Open sessions. Here you can export and delete sessions and
import new sessions.
Patient data The user can enter supplementary patient data and
comments here.
Shows the last passed or failed function checks. The
Device data user is given information about enabled, activated or
deactivated options. Available only for MEDUCORE
2
Standard sessions.
1 File Print Opens the print settings for printing session data.
Saves session data in wmd format. Alternatively,
Save session data can be saved in zip format and unpacked.
This delivers one file in edf format and one file in xml
format.
Opens the settings for sending by e- mail. This
Send by e-mail generates a pdf file appended to an e-mail as an
attachment. Supported e-mail programs are Microsoft
Outlook and Mozilla Thunderbird.
Settings Opens Settings.
Exit program Exits and closes DEFIview.
Default Sets curve view to ECG leads II, Pads and to the
2 View plethysmogram.
6-lead ECG Sets curve view to ECG leads I, II, III, aVR, aVL and aVF.
WM 67401a 01/2018

Opens the Instructions for Use in the relevant


Instructions for Use language set under Menu | Settings | Language
3 Help and localization.
About DEFIview Opens information and contact details for the
manufacturer.

DEFIview EN 15
4 Preparation and operation

4.5.2 Icon bar

1 2 3 4 5 6 7

8 9 10
4-3 Icon bar

No. Designation Description


1 Patient list Opens archive of saved session data
(see "4.2 Opening session data", page 8).
2 Save Opens window for saving session data.
3 Print Opens "Print settings" window (see "4.3 Defining the
settings", page 9).
4 Edit patient Opens the "Patient data" window (see "4.5.5 Edit
patient data", page 19).
Allows selection of the curves to be displayed:
• I, II, III
5 ECG lead • aVR, aVL, aVF
• II, Pads, Plet (plethysmogram)
• I, II, III, aVR, aVL, aVF, Pads, Plet
6 Start/stop Allows the ECG curves to run in real time if a pulse
oximetry sensor is connected.
7 Speed Here you can set the speed of the ECG curve display
and thus increase the temporal resolution.
8 Amplitude scaling Here you can set the amplitude scaling and thus the
amplitude of the ECG curve.
Here you can minimize ECG disturbances caused by
9 ECG filter the power supply system:
• No filter
• Line filter (50 Hz)
WM 67401a 01/2018

16 EN DEFIview
4 Preparation and operation

No. Designation Description


Displays the following parameters:
• Heart rate (HR)
10 Parameter fields • Pulse rate (only with "SpO2" option)
• Oxygen saturation (only with "SpO2" option)
• Systolic/diastolic blood pressure (MEDUCORE
2
Standard only)

4.5.3 Event list

1 2

4-4 Event list

No. Designation Description


Displays the exact time of the event in the format
1 Timestamp Hour:Minute:Second.
You can sort events in ascending or descending order
by clicking on Timestamp.
Displays the event.
2 Event You can sort events in ascending or descending
alphabetical order by clicking on Event.
WM 67401a 01/2018

3 Scroll bar Allows you to reach the events in the list.

DEFIview EN 17
4 Preparation and operation

4.5.4 Curve area

1 5

4-5 Curve area

No. Designation Description


1 Date Indicates the session date.
2 ECG calibration mark Shows the curve corresponding to 1 mV of the ECG
signal.
Provides the name of the ECG lead
• I
• II
• III
3 ECG lead • aVR
• aVL
• aVF
• Pads: Shows the lead of the defibrillation
electrode.
• Plet: plethysmogram
WM67401a01/2018

4 Curve display Displays the course of an ECG lead.


5 Event Marks an event in the curve using a vertical, red line.

18 EN DEFIview
4 Preparation and operation

4.5.5 Edit patient data

1. Click on the Patient data button


or
2. select Menu | File | Patient data.

Result The Edit window opens.


Here you can enter or edit the patient and session data and
any other comments.

3. Edit data.

4. Click OK.
Result Data are edited and saved.

4.5.6 Measuring the ECG amplitudes


Requirement The session data are opened (see "4.2 Opening session
WM 67401a 01/2018

data", page 8).


1. Select a starting point in the ECG leads with the cursor.

2. Click the left mouse button once.

DEFIview EN 19
4 Preparation and operation

3. Draw the cursor across the range to be measured.

Intermediate result A red coordinate triangle appears.


The size of the coordinate triangle can be altered as required
and in all four quadrant directions by moving the mouse. The
horizontal side measures the time difference between the
starting point and end point in milliseconds (ms); the vertical
side measures the amplitude in millivolts (mV).
4. Select the end measuring point.

5. Click the left mouse button once.

6. Read off the measured values next to the

selection. Result ECG amplitudes measured.

4.5.7 ECG lead displayed individually.


Requirement The session data are opened (see "4.2 Opening session
data", page 8).
WM 67401a 01/2018

1. Click on the Magnifying glass symbol.

Result The required ECG lead is displayed individually.

20 EN DEFIview
4 Preparation and operation

4.5.8 Hide ECG lead


Requirement The session data are opened (see "4.2 Opening session
data", page 8).

Click on the Close magnifying glass symbol.

Result The required ECG lead is hidden.

4.5.9 Add and hide ECG leads


Requirement Session data are open (see "4.2 Opening session data", page
8).

1. Use the PC mouse to right-click on the left-hand edge of


the page.

Intermediate result A list of ECG leads appears.

2. Use the PC mouse to left-click on any ECG

leads. Result The desired ECG leads are added or hidden.


WM 67401a 01/2018

DEFIview EN 21
Manufacturer Center for Production, Logistics, Service
WEINMANN Emergency WEINMANN Emergency
Medical Technology GmbH + Co. KG Medical Technology GmbH + Co. KG
Frohbösestraße 12 Siebenstücken 14
22525 Hamburg 24558 Henstedt-Ulzburg
GERMANY GERMANY
T: +49 40 88 18 96-120
E: customerservice@weinmann-emt.de
WM 67401a 01/2018 EN

Simply Professional Made in Germany WEINMANN-Emergency.com