Sie sind auf Seite 1von 2

Topic 4: Image Acquisition Errors (Most CR systems include algorithms designed

to detect abrupt edges of exposure and non-


The data in the imaging plate is located within the
collimated edges in order to develop a histogram exposure, which are assumed to represent
collimator edges.)
Data clipping
 Most CR systems do not send all the data Field recognition error
acquired by the image receptor to the post-
acquisition system because the quantity of  The computer cannot find the edges of
data is far greater than the display system collimation
can provide for viewing.  Can include off focus and scatter in the
 the software processing the data coming
histogram
from the image receptor is programmed to
recognize that some data represents Result:
exposure below the diagnostic imaging  Dark, light, or low contrast images
range,whereas other data represents  Inaccurate exposure indicator number
exposure far above the diagnostic range.
Essentially, the algorithm eliminates data  No collimation
that is clinically irrelevant to a particular  There will be too much exposure in the
examination. histogram, that will also cause a
 The information that is important on the histogram analysis error and will cause
histogram will be determined by the values the exposure indicator to reflect a large
of interest (VOI), which will be different for exposure to the plate.
each body part.  Too much collimation
 The algorithm also has an LUT that has the  can also cause histogram analysis errors
appropriate contrast for each body part. and lead to poor image quality
Therefore, the algorithm will vary
depending on the body part.
 If the histogram has excess exposure from
 This is why CR systems demand to know poor collimation techniques, this could
what examination is being performed before cause an error in locating the VOI, which
the image can be acquired, processed, and can cause rescaling errors that result in poor
displayed density and contrast on the image.
Histogram analysis error
 It is also possible to have an abrupt straight
 The computer has reference histograms for line of tissue density when there are
the body parts prosthetic devices or other parts with large
 The computer uses the shape of the density differences that may lead to a
histogram and algorithm to locate the VOI histogram analysis error
and determine the exposure indicator
 Another histogram analysis error will occur
 When an image is obtained that does not fit when an area of increased or decreased
the parameters that were used for the attenuation is located in the body where it is
reference histogram, it may result in a not normally located
histogram analysis error  When multiple exposures are done on one
What happens when a histogram analysis plate, symmetric part placement and
exposure field borders will give the best
error occurs? results.
• The image will not display appropriately.  Lack of symmetry may result in poor image
• The appropriate density and contrast will not quality due to histogram analysis errors. If
be displayed because the computer does not the computer has exposure field recognition
recognize the data. errors. Each exposure of multiple exposures
• It can give an incorrect exposure indicator must also have the appropriate technique,
number. because the computer will average the
density and contrast between all exposures.
 Another problem is that when obtaining Increased Thickness, Size and Density – Increased
cross-table images, often the part is close to Scatter Radiation
one side of the plate, which can also cause
exposure field recognition errors and lead to 3. BEAM-RESTRICTING DEVICE/ COLLIMATOR
rescaling errors. Increased Collimation – Decreased Scatter
Radiation
Low intensity radiation response
DEVICES THAT REDUCE SCATTERED
 PSP stores background exposure RADIATION
 Imaging Plate responds to an exposure as
low as 60 µR 1.) Beam Restrictors
 Background is 40 µR/day to 80 µR/day 2.) Grids
 Plates unused for more than 48 hours
should be erased RADIOGRAPHIC GRID
Scatter control  Device used to reduce the intensity of scatter
radiation in the remnant x-ray beam
 It is recommended that kVp levels not  Principal Function: to improve image
contrast
exceed 80 for any non-grid radiography,  Secondary function: to absorb scattered
including the chest. radiation
 It is recommended that when scatter ratios  Used when:
approach 50 percent, grids should become  Anatomical part >10 cm
mandatory. Failure to do so greatly reduces  >60 kVp is used
image contrast due to the scatter radiation
reaching the image receptor. • Advantage: improved image contrast
• Disadvantages:
 Higher technical factors used
 High patient dose
• Position: between the patient & IR

GRID CUTOFF
• The undesirable absorption of primary x-ray
by the grid
• Cause: improper grid position
• Primary Radiographic Effects: Further
reduction in the number of photons
reaching the IR
• Decrease in radiographic density
• Most common to parallel grid

SECONDARY RADIATION/FOG Short dimension grid VS Long dimension grid

Increased kVp – Increased Scattered Radiation – Long dimension grids


• have lead lines that run parallel to the long
Increased Fog
axis
• can be used in both portrait and landscape
FACTORS THAT CONTROL SCATTER RADIATION
orientation
1.KVp Short dimension grids

Increased kVp – Increased Scatter Radiation– • have lead lines run parallel to the short
Increased Fog – Increased Density axis
• should only be used in landscape
2.PATIENT THICKNESS, SIZE AND DENSITY

Das könnte Ihnen auch gefallen